Dear Customer, 
Congratulations on your purchase of KELLER Data Loggers DCX, the 
next generation Data Logger! 
Before installation and use of your new device as per your requirements, it is recommended that you 
read these instructions carefully. They contain important information for successful operation of the 
KELLER Logger DCX 4.0 software. 
Sections in the manual: 
Table of contents
Installation instructions
(Data Logger) 
Installation of the software
Configuration of the Data Logger
(Writer / WriterWL) 
Read-out of stored measured values
(Reader) 
Processing of measured values read
(Modules) 
Appendix
(Sample application, Pressure Conversion Table, Messages) 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 2 
How to C#: Set Image Thumbnail in C#.NET
PDF, C#.NET convert PDF to svg, C#.NET convert PDF to text, C#.NET convert PDF to images VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel, VB.NET Back Color.
pdf to powerpoint converter; pdf into powerpoint
VB.NET Word: Word Conversion SDK for Changing Word Document into
completed. To convert PDF back to Word document in VB.NET, please refer to this page: VB.NET Imaging - Convert PDF to Word Using VB.
converter pdf to powerpoint; how to add pdf to powerpoint presentation
Table of Contents 
Installation of Data Logger DCX
..........5
Dip pipe installation (DCX-22 AA, DCX-22 SG and DCX-22 VG).......................................................5
Additional recommendations:
..........6
Installation of the KELLER Logger DCX 4.0 software.......................................................................6
System Prerequisites:
................6
Installation of the software
................6
Files installed
.......................7
The KELLER Logger DCX 4.0 software package
8
The configuration program (Writer / WriterWL)
8
Initialization
......................9
The menus
..................9
File
.....................9
Port-Setup
................9
Settings
..................9
Options
...............10
Modem
..............10
Info
...................10
Configuration functions
...........11
Channels to record
........11
Start time
...............11
Use Fixed Save-Interval
.....11
Start Record at Event
.......11
Event Detect Interval
........11
Recording-Interval after Event
...11
Building Mean value after X measurements at 1 second intervals............................................12
Event
.................12
Endless (circular memory)
....14
Device-Identity (for file name)
.....14
Comment
............14
Actual value
.............14
Device type
.............15
Conversion to
............15
Starting conditions at installation
....15
Water density / Calculate water density
16
Set User-Values (only in Writer)
..16
Operation of Writer / WriterWL
.......17
Units
..................17
Read Configuration
........17
Stop Record
............17
Write Configuration
.........17
Cancel
................17
Exit
...................17
The Reader program (Reader)
....18
Initialization
....................18
The menus
................19
File
...................19
Port-Setup
..............19
Options
...............19
Modem
..............19
Info
...................19
Read-out of stored data
..........20
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 3 
C# PDF Page Rotate Library: rotate PDF page permanently in C#.net
control, RasterEdge XDoc.PDF, is a 100% clean .NET solution for C# developers to permanently rotate PDF document page and save rotated PDF document back or as
convert pdf slides to powerpoint; converting pdf to powerpoint
C# Image: Tutorial for Collaborating, Marking & Annotating
Besides, more annotations can be drawn and saved back to the database We are dedicated to provide powerful & profession imaging controls, PDF document, image to
export pdf into powerpoint; how to change pdf file to powerpoint
Setting a mark
................20
Description of individual modules
...21
Airpressure-compensation (Airpressure-Comp.)
21
Conversion to Waterlevel values (WL-Converter)
22
Creation of text data (Text-Converter)
...23
Display module (Viewer)
.........25
Files with water level values:
.....26
WISKI-Converter
.............27
Overstort-calculation (Overstort)
.......28
APPENDIX
..................29
Sample application
.............29
Pressure Conversion Table
......30
Messages:
.................30
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 4 
C# TIFF: Merge and Split TIFF File(s) with C# Programming
C#.NET Demo Code for Merging & Splitting TIFF File(s). // split TIFF document into 2 parts and save them back to disk TIFFDocument.SplitDocument(sourceFilePath
convert pdf to powerpoint presentation; convert pdf into powerpoint
RasterEdge Product Refund Policy
the first step for you is to sign and send back RasterEdge Software We are dedicated to provide powerful & profession imaging controls, PDF document, image to
convert pdf into ppt; how to convert pdf into powerpoint on
Installation of Data Logger DCX 
Dip pipe installation (DCX-22 AA, DCX-22 SG and DCX-22 VG) 
Revolution in level metrology: 
Gone are the days of the costly measuring stations at the lakes and rivers. The new generation of 
level measuring stations presents itself inconspicuously and vandal-proof and can be set up at 
considerably lower costs compared to conventional systems. Here is an application example at a 
barrage: 
The dip pipe is firmly mounted at the measuring place. If the dip pipe is being sunk 
into the ground, openings have to ensure that water can enter the pipe. 
The level sensor, connected by a cable to the electronics housing, is immersed into 
the pipe. The housing incorporates the electronics, the battery and the air pressure 
sensor. 
The electronics housing is mounted onto the end of the pipe. 
The protective cap of the interface must be cleaned from dirt and water prior to 
removing. 
Via data cable, a Laptop is connected to the system on site. The stored measuring 
values with the programmed parameters can now be read out or the instrument can 
be reconfigured. 
The system is set up for a wireless data transfer via modem. 
A vandal-proof metal cap mounted at the end of the system prevents unauthorised 
access. 
The level sensor and the electronics housing form a leak-proof unit and are thus 
protected against flooding and brackish water. 
Warning! 
A check of pressure offset values is recommended after each installation with resetting if necessary 
(see Set Pressure Offset
in the description of Writer/WriterWL)! 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 5 
C# PDF: Start to Create, Load and Save PDF Document
can use PDFDocument object to do bulk operations like load, save, convert images/document to page in the document), you can save it back to a PDF file or
pdf to ppt converter; convert pdf to editable powerpoint online
C# Imaging - Linear ITF-14 Barcode Generator
Y to control barcode image area on PDF, TIFF, Word 14 barcode image fore and back colors in BarcodeHeight = 200; barcode.AutoResize = true; //convert barcode to
adding pdf to powerpoint; how to convert pdf to ppt for
Additional recommendations: 
- The battery capacity is not measured but calculated from the average current used. If the 
battery is replaced, as also upon Power-On-Reset, the display will be reset to 100%. A battery 
replacement is recommended if the battery capacity indicator displays a status of less than 
30%. Continued read-out of large volumes of data is thereby guaranteed. (During read-outs, 
the battery status indicator is not updated, consequently it is possible that a critical battery 
condition may not be recognized or may be recognized too late). 
- The measuring head should be checked at least once a year for dirt. 
Installation of the KELLER Logger DCX 4.0 software 
System Prerequisites: 
Processor 
minimum Pentium 75 MHz 
Screen resolution 
minimum 1024 x 768 
Main memory  
minimum 16 MB RAM 
Free disk space 
minimum 20 MB recommended 
Serial Port 
1 free serial port for communication with the Data Logger 
Operating system (tested) 
Windows9x, Windows NT, Windows 2000, Windows XP 
Installation of the software 
On our homepage www.KELLER-druck.com
you will always find the latest version of the Logger DCX 
software. 
Double-click the file "LoggerDCX4-xx.exe” to start the installation program. 
On the first screen, you can make your choice of language to be used during the installation. 
In the third screen, the Destination Directory can be specified (Default: "C:\Program Files\KELLER 
Druckmesstechnik”). 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 6 
How to C#: Create a Winforms Control
pages, VB.NET comment annotate PDF, VB.NET delete PDF pages, VB.NET convert PDF to SVG. VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel, VB.NET Back Color.
how to convert pdf slides to powerpoint presentation; embed pdf into powerpoint
VB.NET Image: VB.NET Codes to Add Antique Effect to Image with .
a touch of history to the image which can help bring back the sweet We are dedicated to provide powerful & profession imaging controls, PDF document, image to
change pdf to powerpoint online; convert pdf file to ppt online
Under Install Type, you may choose the type of installation: 
Typical and Compact:  
Default installation without additional modules  
Custom: 
Installation of additional modules WISKI-Conversion and Overstort 
Finally, you can specify the Program Folder to be created under Programs in the Start menu. 
Files installed 
KELLER Logger DCX 4.0 for Windows consists of the following files:
- Writer (general configuration)* 
ter level sensors) 
(Air pressure compensation module) 
/ CSV – Format (D)) 
Non-water level applications such as the rain logger are configured with the Writer. 
- Writer WL (configuration for wa
- Reader (Reader program) 
- Viewer (Display program) 
- Air pressure compensation
- Water level converter (module for conversion to water level values
- Text-Converter (module for generation of text data) 
- WISKI-Converter (module for conversion into ZRXP 
- Overstort (module for generation of an Overstort report (NL)) 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 7 
The KELLER Logger DCX 4.0 software package  
The software consists of three main programs and additional modules. 
Both the configuration programs Writer and WriterWL are used to read and update the settings of the 
connected device, as also to start and end recording. WriterWL configures devices for water level 
applications, while the Writer is used for making general recordings. 
The Reader gives an overview of all records present in the connected device, and reads in the records 
selected. Each record read in is stored in a separate file with the suffix ".idc". These files can be read 
by all modules for further processing or display.  
Each module can be either executed directly from the Reader, or as a separate stand-alone program. 
The configuration program (Writer / WriterWL) 
These programs are used for the configuration and starting/stopping of recordings as desired. Since 
the only difference between these programs is the water level configuration feature additionally 
provided in WriterWL, it is the latter's operation that is primarily described in detail in this manual. 
WriterWL 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 8 
Initialization 
The Data Logger DCX needs to be initialized before communication can be established with it. This 
can be done by clicking Read Configuration. If a device is present, its settings are read in and 
displayed in WriterWL. The device type, serial number, device time and PC time as well as the (total) 
memory available in the device are listed in the status window. The status bar shows the current 
recording status as well as the current time. 
The menus 
File 
Save configuration
The current configuration is backed up to a file (*.cfx) 
Open configuration: 
A saved configuration file is opened and loaded.  
Exit:  
Exit the program  
Port-Setup 
Port  
Selection of the serial interface 
Modem:                         Upon activation, the communication protocol for data transfer via a modem is 
negotiated.  
Settings 
Editing the device identity: 
On this screen you may enter a description (max. 63 characters) by which the device can be uniquely 
identified. This identification, along with the current date/time and index, is used to form the name of 
the file that the reader creates after reading in recorded data. 
Setting the Pressure-Offset: 
It is possible to recalibrate the pressure offset of the 
existing sensors. This can become necessary if, for 
example, the original values do not tally exactly due to the 
measuring setup. Either an individual sensor, or both 
sensors simultaneously if installed, can be set to a new 
value.  
To reset a sensor to its factory setting, click Set Factory Zero respectively.  
To update the pressure values displayed, click Read Act Value(s). 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 9 
Options 
Configuration 
General: 
Setting the program language, activation/deactivation of auto detection of existing serial ports 
(changes will take place only upon program restart), and showing/hiding the picture. 
Units: 
The number of decimal places for the predefined units of pressure, temperature and length is specified 
here. In addition one user-defined pressure and length unit each can be specified, with details about 
conversion factors and decimal places. 
Channel-Names: 
The user can customize the descriptions of individual channels as per his requirements. 
Modem: 
This page is needed for specification of a telephone number and additional initialization commands.  
KELLER Logger DCX 4.0 program! 
Modem 
Connecting: 
initialization commands 
Disconnecting: Disconnect from the modem 
Info 
Information about the software version and contact details for the company KELLER AG für 
Druckmesstechnik is displayed. 
Clicking OK causes the values specified in the WriterWL to be updated and the configuration screen to 
be exited. CANCEL causes return to the main program without saving any changes. 
Operating Manual for Logger DCX 4.0 