Description of individual modules 
Each module can be directly invoked from the reader (data is transferred automatically). If the module 
is invoked directly, the file to be processed can be read by clicking "File – Open...”. 
Airpressure-compensation (Airpressure-Comp.) 
In this module one or more (if invoked from the Reader) files are compensated for air pressure. The air 
pressure is taken either from a previously created "*.idc” File, or a fixed air pressure (constant) is used 
for compensation. 
If the air pressure data file read in contains measured values from a device of the type DCX-22 AA, 
then Channel P2 is the "Channel with airpressure”. 
Click Convert to compensate each file with the selected air pressure. 
Warning! 
No new files are created, instead the values in the current files are overwritten. In addition a mark is 
set preventing a second airpressure compensation. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 21 
Pdf page to powerpoint - C# Create PDF from PowerPoint Library to convert pptx, ppt to PDF in C#.net, ASP.NET MVC, WinForms, WPF
Online C# Tutorial for Creating PDF from Microsoft PowerPoint Presentation
convert pdf to ppt; picture from pdf to powerpoint
Pdf page to powerpoint - VB.NET Create PDF from PowerPoint Library to convert pptx, ppt to PDF in vb.net, ASP.NET MVC, WinForms, WPF
VB.NET Tutorial for Export PDF file from Microsoft Office PowerPoint
convert pdf document to powerpoint; convert pdf to powerpoint using
Conversion to Waterlevel values (WL-Converter) 
Since the data logger pressure values are always measured and recorded in bar, this module is 
necessary for conversion to water level values. The configuration settings specified in WriterWL are 
re-displayed for verification. 
Clicking Convert leads to all measured values being converted into water levels. The unit for 
converted values is always meter (m). 
Warning! 
No new files are created, instead values in the current files are overwritten. In addition a mark is set 
preventing a second calculation of water level values. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 22 
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
XDoc.PDF ›› VB.NET PDF: Delete PDF Page. using RasterEdge.Imaging.Basic; using RasterEdge.XDoc.PDF; How to VB.NET: Delete a Single PDF Page from PDF File.
how to convert pdf to ppt online; convert pdf pages into powerpoint slides
C# PDF Page Insert Library: insert pages into PDF file in C#.net
C# programmers are capable of adding and inserting (empty) PDF page or pages from various file formats, such as PDF, Tiff, Word, Excel, PowerPoint, Bmp, Jpeg
pdf to powerpoint slide; convert pdf slides to powerpoint online
Creation of text data (Text-Converter) 
For display and/or further processing of measured values if required, conversion into a text file is 
necessary. This module creates such a file in tab-delimited format with the suffix ".txt". The file can be 
opened with any desired editor. It is however recommended that the file be imported into a commonly 
used spreadsheet program such as Microsoft Excel for a better overview. 
The user has several options: 
Write specified channel-names and units into File: 
To extract a smaller set of data, specific channels can be selected. 
Other units can also be specified. A user-defined unit can be specified under "Options – 
Configuration – Units", which will be applied to the entire KELLER Logger DCX 4.0 – package. 
Write only recorded channels into File: 
channels are implemented). If this option is not selected, all 15 channels are written out, 
whether they contain data or not. Due to this fixed assignment, a user macro or other program 
can always access the desired measured values. If this option is checked off, only recorded 
channels are written to the file. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 23 
VB.NET PDF Page Insert Library: insert pages into PDF file in vb.
PDF Pages. |. Home ›› XDoc.PDF ›› VB.NET PDF: Insert PDF Page. Professional .NET PDF control for inserting PDF page in Visual Basic .NET class application.
pdf to powerpoint; chart from pdf to powerpoint
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
Pages. |. Home ›› XDoc.PDF ›› C# PDF: Delete PDF Page. Demo Code: How to Delete a Single PDF Page from PDF File in C#.NET. How
pdf to ppt; converting pdf to ppt
Write User-Values into File: 
Specify whether user-defined values created with the Writer are written into the file. Each 
value can also be given its own description. This description can in turn be used by a user 
macro or program to read in the required values.  
Write specified separators into File: 
Since number, date and time formats are country-specific, they have to be localized, which 
can be done using this option. The user can choose the desired character from an existing 
drop-down list, or specify another character. The mask used for the date and time formats can 
be changed (e.g. from dd.mm.yyyy to yyyy.mm.dd). Short or long names can also be selected, 
depending on the number of characters (mmm= short form of the month e.g. Sep., mmmm = 
month written out fully e.g. September). 
On clicking Convert the user is asked to change or confirm the default file name. The default file 
name is based on the following pattern:  
"(DeviceID+date+time+Index).txt" 
-  
date and time that the text data was created! 
Upon clicking Save, the file is created. 
Sample file created using the Text-Converter, displayed in Microsoft Excel 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 24 
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
PowerPoint, VB.NET Tiff, VB.NET Imaging, VB.NET OCR, VB.NET Twain, VB.NET Barcode Read, VB.NET Barcode Generator, view less. C# PDF - Modify PDF File and Page
convert pdf to editable ppt; convert pdf to powerpoint slides
How to C#: Overview of Using XDoc.PowerPoint
Create and Edit PowerPoint Page and File. Tell C# users Please click to see details. Annotate PowerPoint Page. Various annotation objects
conversion of pdf to ppt online; convert pdf to powerpoint
Display module (Viewer) 
This module is used for displaying the measured values read out. The user can select the following 
options to customize the view: 
Channels to be displayed 
Pressure and temperature units 
Zoom in and zoom out of the graph 
Display of cross-hair, grid and legend 
For better viewability, the pressure values can be inverted and displayed. This is especially 
recommended for files with water level values (virtual representation of the water height). 
Under "File – Save as Bitmap..." the graph is saved in the default format (*.bmp). 
The status bar displays the x- and y-values per unit respectively. 
Clicking this button resets the display to its starting state. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 25 
C# PowerPoint - Sort PowerPoint Pages Order in C#.NET
C#.NET. Support Customizing Page Order of PowerPoint Document in C# Project. Overview. splitting. C# DLLs: Move PowerPoint Page Position. Add
create powerpoint from pdf; drag and drop pdf into powerpoint
C# PowerPoint - Delete PowerPoint Document Page in C#.NET
C# DLLs: Delete Page from PowerPoint File. Delete a Single PowerPoint Page from PowerPoint File in C#. How to delete a single page from a PowerPoint document.
convert pdf into ppt online; convert pdf into powerpoint online
Files with water level values: 
If a file with converted water level readings is displayed, in older versions of the viewer, the pressure 
unit cannot be changed thereafter. This is indicated by the display of unit details [m] alongside the 
pressure channel. 
Example of a file converted into water level readings 
Hint: 
In Windows Explorer under "Extras – Folder options" specify Viewer as the program for opening files 
with the suffix "*.idc". Thereafter each .idc file created by the Reader can be displayed in the Viewer by 
double-clicking. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 26 
WISKI-Converter 
The design of this module is very similar to that of the Text-Converter. It is also used for creation of a 
text-file which can be used for further processing or for display.  
Under "Options – Configuration - General" there are two selectable formats. 
ZRXP-Format: 
If the data is converted into this format, it can be displayed and processed with 
the WISKI - TimeSeriesViewer from the firm Kisters (30-day demo version 
available here
). 
CSV-Format: 
This format is a widely-used general  semicolon-separated format with the 
following pattern: 
Date;Time;P1-P2 [bar];P1 [bar];P2 [bar];TOB1 [°C];TOB2 [°C] 
29.11.2002;13:15:05;0,0001;0,9628;0,9626;22,9336;23,3125 
... 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 27 
Overstort-calculation (Overstort) 
In some countries (e.g. Holland, Belgium) water consumption is measured with the help of an overstort 
wall. The volume of water consumed is calculated from the measured water level minus the Trigger-in-
value (Val1) configured with the WriterWL. This Trigger-in-value corresponds to the distance of the 
sensor from the overstort wall. Tr
to attain a turn-off hysteresis. 
Poleni's formula is used for flow calculations: 
Q = 1.7 x m x b x h
3/2
= flow 
1.7 
= fixed correction factor 
= construction factor 
= width of the overstort wall 
The total overstort-volume is calculated using all calculated and intermediary flows and the 
corresponding duration of the overstort.  
The module creates a detailed record, which can be printed as a report under "File – Print...”. 
This module is available only in Dutch and German. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 28 
APPENDIX 
Sample application 
Recording configuration: 
Fixed Save-Interval  = 1 hour 
Event  
= Save if delta Channel > Val1, Val1 = 0.5 bar 
Event-Detect Interval = 1 second 
Save-Interval during the Event: = deactivated, correspondingly also 1 second 
= Start Recording (1. measure for compare) 
p[bar]
= Recording with fixed Save interval (1 h.) 
= Event-Recording (delta > 0.5 bar) at 1 sec. int. 
4.0 
3.5 
3.0 
2.5 
2.0 
1.5 
1.0 
0.5 
t
12.30 
13.00 
13.30 
14.00 
14.30 
15.00 
15.30 
16.00 
16.30 
16.08.02 
17.00 
16.08.02 
12.00 
As soon as the configuration has been written into the device, a first measured value is recorded (in 
the sample:Start recording at 12.00 p.m.). As per configuration, this measured value is compared 
every second with the current measured value, till the difference is 0.5 bar. This first happens at 13.15. 
The next recording takes place at 13.30, etc...  Independent of that, each fixed Save-Interval (1 hour), 
an additional recording is done.  
At 17.00 both a recording at the fixed Save-Interval as well as an event recording occur 
simultaneously. Only one measured value is stored however. 
Warning! 
With the Event "Save if delta Channel > Val1” the interval "Recording interval after Event" is always 
taken into account. This means that in the following configuration example, the interval time is always 
1 second, that is both for event detection as well as for recording! 
Event Detect Interval  
= 5 seconds 
Recording interval after Event: = 1 second 
The time value specified under Fixed Save-Interval is however independent and in addition to the 
above mentioned interval. 
Operating Manual for Logger DCX 4.0 
KELLER AG 
Page 29 
Pressure Conversion Table  
bar 
kPa 
mWs/mH
2
O inchH
2
O mmHg(Torr) inchHg 
psi 
bar 
100 
10,1972 
401,463 750,062 
29,530 
14,5038 
kPa 
0.01 
0,101972 
4,01463 7,50062 
0,29530 
0,145038 
mWs/mH
2
O 9,8067*10
-2
9,8067 
39,3701 73,5559 
2,8959 
1,42233 
inchH
2
2,49089*10
-3
0,249089 2,540*10
-2
1,86832 
7,35559*10
-2
3,613*10
-2
mmHg(Torr) 1,33322*10
-3
0,133322 1,35951*10
-2
0,535240
3,9370*10
-2
1,9337*10
-2
inchHg 
3,38639*10
-2
3,38639 0,345316 
13,5951 25.40 
0,491154 
psi 
6,89476*10
-2
6,89476 0,70307 
27.68 
51,7149 
2,03602 
Messages: 
Message 
Significance 
REC-ERROR: Please perform a power-on reset ! 
A problem occurred during the recording.  
The battery must be removed and reinstalled.
BAT-ERROR: Please replace battery ! 
The battery must be replaced 
ACK-ERROR: Please perform a power-on reset ! 
There was a problem in saving during the 
recording. The battery must be removed and 
reinstalled. 
Device has changed, please read configuration first! ! The connected device was swapped out. 
It must be reinitialized with Read Configuration 
No communication! Please check connection and 
port-no. 
No communication with the connected device. 
The serial ports at the device and PC need to be 
checked. 
State: Record active 
Device is in recording mode 
State: Record prepared 
Device is being prepared for recording mode 
State: Record stopped 
Recording mode is stopped 
Memory full! 
The memory is completely full 
(only possible, if "Endless recording (circular 
memory)" has not been activated) 
Operating Manual for Logger DCX 4.0 
KELLER AG 