Remote Control Basics
R&S®EX-IQ-BOX
215
User Manual 1409.5570.32 ─ 05
In case the R&S DigIConf and the controller are installed on the same PC, the IP address
127.0.0.1 or Local Host is used. Hence the valid visa resource string is:
TCPIP::127.0.0.1::INSTR
Identifying instruments in a network
If several instruments are connected in a network, each instrument has its own IP address
and associated resource string. The controller identifies these instruments by means of
the resource string.
Socket communication
An alternative way for remote control of the software is to establish a simple network
communication using sockets. The socket communication, also referred as “Raw Ether-
net communication” or "Raw socket", does not require a VISA installation on the remote
controller side. Connection can also be performed with "Win Socket" communication.
An alternative way to establish a socket communication is to use a build-in telnet program.
The telnet program is part of every operating system and supports a communication with
the software on a command-by-command basis. For better utilization and to enable auto-
mation by means of programs, user defined sockets can be programmed.
Socket connections are established on a specially defined port. The socket address is a
combination of the IP address or the host name of the instrument and the number of the
port configured for remote-control. R&S DigIConf uses port number 5026 for this purpose
by default, but can be set. The port is configured for communication on a command-to-
command basis and for remote control from a program.
8.1.3 Starting a Remote Control Session
The instrument and the controller have to be connected with the suitable cable and
switched on.
A remote control program must open a connection to R&S DigIConf (using Raw socket
or VISA functionality), before it can send commands to and receive device responses
from R&S DigIConf.
Refer to ​chapter 8.1.5, "Examples", on page 216 for practical examples on setting up of
a remote control link and starting of a remote control session.
8.1.4 Switching to Remote Control
When it is started, the software is always in the manual operating state and can be oper-
ated via the block diagram.
Starting remote control
Send a command from a controller to the signal generation software.
Remote Control Interfaces and Protocols
How to convert pdf to html code - Convert PDF to html files in C#.net, ASP.NET MVC, WinForms, WPF application
How to Convert PDF to HTML Webpage with C# PDF Conversion SDK
online convert pdf to html; converting pdf to html email
How to convert pdf to html code - VB.NET PDF Convert to HTML SDK: Convert PDF to html files in vb.net, ASP.NET MVC, WinForms, WPF application
PDF to HTML Webpage Converter SDK for VB.NET PDF to HTML Conversion
converting pdf to html format; embed pdf to website
Remote Control Basics
R&S®EX-IQ-BOX
216
User Manual 1409.5570.32 ─ 05
The software is in a remote control state as soon as it receives a command from the
controller.
8.1.5 Examples
These sections provides examples for setting up of remote control connection and start-
ing a remote control session over LAN interface.
This section assumes basic knowledge of programming and operation of the controller.
A description of the interface commands can be obtained from the relevant manuals.
8.1.5.1 Remote Control Over LAN Using Socket Protocol
Through the examples in this section, the program 'Measurement & Automation Explorer'
from National Instruments under Windows operating system is used for setting up a LAN
remote control link and starting a remote control session.
Configuring the controller
To enable the external controller to communicate with the software via TCP/IP protocol,
set up a remote control link as follow:
1. Connect the controller and the instrument to the network (network cable) and switch
them on.
2. Start the 'Measurement & Automation Control' program on the controller.
3. Select "Devices and Interfaces > Create New".
4. Select "VISA TCP/IP Resource" and confirm with "Next".
Remote Control Interfaces and Protocols
VB.NET PDF Convert to Tiff SDK: Convert PDF to tiff images in vb.
RasterEdge.XDoc.TIFF.dll. Use corresponding namespaces; using RasterEdge.Imaging. Basic; using RasterEdge.XDoc.PDF; Demo Code to Convert PDF to Tiff Using VB.NET
how to convert pdf to html; embed pdf into website
VB.NET PDF Convert to Word SDK: Convert PDF to Word library in vb.
using RasterEdge.XDoc.PDF; Convert PDF to Word Document in VB.NET Demo Code. This is an example for converting PDF to Word (.docx) file in VB.NET program.
converting pdf into html; convert pdf into html
Remote Control Basics
R&S®EX-IQ-BOX
217
User Manual 1409.5570.32 ─ 05
5. Choose the type of TCP/IP resource you wish to add and select "Next".
6. Enter the IP address or the host name of R&S DigIConf and select "Next".
7. Enter the alias name if required.
Remote Control Interfaces and Protocols
Online Convert PDF to HTML5 files. Best free online PDF html
library control is a 100% clean .NET document image solution, which is designed to help .NET developers convert PDF to HTML webpage using simple C# code.
convert fillable pdf to html form; adding pdf to html page
C# PDF Convert to SVG SDK: Convert PDF to SVG files in C#.net, ASP
PDFDocument pdf = new PDFDocument(@"C:\input.pdf"); pdf.ConvertToVectorImages( ContextType.SVG, @"C:\demoOutput Description: Convert to html/svg files and
add pdf to website html; adding pdf to html
Remote Control Basics
R&S®EX-IQ-BOX
218
User Manual 1409.5570.32 ─ 05
The alias name must not be mistaken for the computer name. It is only used for
instrument identification within the program and displayed in the menu as an option
in case of an Ethernet link.
8. Confirm the settings with "Finish".
The instrument is configured and the settings are displayed in the "TCP/IP Settings"
tab.
Remote Control Interfaces and Protocols
C# PDF Convert to Jpeg SDK: Convert PDF to JPEG images in C#.net
Output.jpg"); Following demo code will show how to convert all PDF pages to Jpeg images with C# .NET. // Load a PDF file. String
convert pdf fillable form to html; convert pdf to web form
C# PDF Convert to Text SDK: Convert PDF to txt files in C#.net
other industry-leading methods to convert target PDF document to other editable file formats using Visual C# code, such as, PDF to HTML converter assembly, PDF
export pdf to html; attach pdf to html
Remote Control Basics
R&S®EX-IQ-BOX
219
User Manual 1409.5570.32 ─ 05
9. To test the connection, select "Validate".
A message indicates whether the link to the instrument can be set up or not.
If a connection cannot be set up, check whether the controller and the instrument are
connected to the network (network cable) and switched on. Correct spelling of the IP
address or the computer name can also be checked. For further error location, inform
the network administrator. In large networks, specification of additional addresses
may be required for link setup, e.g. gateway and subnet mask, which are known to
the network administrator.
R&S DigIConf is now registered in the program and can be addressed via the
resource string or alias name.
Starting a remote control over LAN
1. Start the 'Measurement & Automation Explorer' on the controller.
2. In the "Configuration" window, select "Device and Interfaces > VISA TCP/IP Resour-
ces", select the required instrument and select "Open VISA Test Panel".
Remote Control Interfaces and Protocols
C# PDF Convert to Images SDK: Convert PDF to png, gif images in C#
x86. C# Sample Code for PDF to Png in C#.NET. You can use this sample code to convert PDF file to Png image. // Load a PDF file.
converting pdf to html email; convert pdf to web page
C# PDF Annotate Library: Draw, edit PDF annotation, markups in C#.
This page will guide C# users how to add text comments on PDF page using C# demo code in Visual Stuodio .NET class. C#.NET: Add Text Box to PDF Document.
convert pdf fillable form to html; adding pdf to html
Remote Control Basics
R&S®EX-IQ-BOX
220
User Manual 1409.5570.32 ─ 05
3. In the "viWrite" tab, write the command to be send to the instrument and select "Exe-
cute".
Instrument responses are displayed on the "viRead" tab.
Remote Control Interfaces and Protocols
Remote Control Basics
R&S®EX-IQ-BOX
221
User Manual 1409.5570.32 ─ 05
For further program operation refer to the online help of the program.
Setting up a Telnet Connection
This chapter provides an example on how to establish a remote control connection over
telnet protocol and a simple sockets-based program example that can be further devel-
oped.
To control the software, only a telnet program is required. The telnet program is part of
every operating system.
1. To establish a Telnet connection with R&S DigIConf, start the telnet program and
enter the socket address.
The socket address is a combination of the IP address or the host name of the R&S
DigIConf and the number of the port configured for remote-control via telnet.
Tip:
R&S DigIConf uses the port number 5026 for remote connection via Telnet.
The connection to the instrument is set up and remote-control commands can be
sent.
Remote Control Interfaces and Protocols
Remote Control Basics
R&S®EX-IQ-BOX
222
User Manual 1409.5570.32 ─ 05
2. Even if the cursor is not visible on the screen, enter blind a remote-control command
and confirm with Enter.
After the first remote-control command had been send, R&S DigIConf is in remote-
controlled state.
Instrument response is displayed after the command has been sent.
8.2 SCPI Command Structure
SCPI commands consist of a so-called header and, in most cases, one or more param-
eters. The header and the parameters are separated by a "white space" (ASCII code 0
SCPI Command Structure
Remote Control Basics
R&S®EX-IQ-BOX
223
User Manual 1409.5570.32 ─ 05
to 9, 11 to 32 decimal, e.g. blank). The headers may consist of several mnemonics (key-
words). Queries are formed by appending a question mark directly to the header.
The commands can be either device-specific or device-independent (common com-
mands). Common and device-specific commands differ in their syntax.
Restrictions in the use of SCPI commands in R&S DigIConf
R&S DigIConf does not support the entire functionality of the SCPI commands. Note the
following restrictions:
2 commands entered in a row are not provided (usually divided by a semicolon)
only a limited subset of Common Commands is available
no status reporting system
8.2.1 Syntax for Common Commands
Common (=device-independent) commands consist of a header preceded by an asterisk
(*) and possibly one or more parameters.
Example: 
*RST
RESET
Resets the instrument.
*IDN?
IDENTIFICATION QUERY
Queries the instrument identification string.
*OPT?
OPTION IDENTIFICATION QUERY
Queries the options included in the instru-
ment.
SCPI Command Structure
Remote Control Basics
R&S®EX-IQ-BOX
224
User Manual 1409.5570.32 ─ 05
8.2.2 Syntax of Device-Specific Commands
Optional note heading
Not all commands used in the following examples are necessarily implemented in the
instrument.
For demonstration purposes only, assume the existence of the following commands for
this section:
DISPlay[:WINDow<1...4>]:MAXimize <Boolean>
FORMat:READings:DATA <type>[,<length>]
HardCOPy:DEVice:COLor <Boolean>
HardCOPy:DEVice:CMAP:COLor:RGB <red>,<green>,<blue>
HardCOPy[:IMMediate]
HardCOPy:ITEM:ALL
HardCOPy:ITEM:LABel <string>
HardCOPy:PAGE:DIMensions:QUADrant[<N>]
HardCOPy:PAGE:ORIentation LANDscape | PORTrait
HardCOPy:PAGE:SCALe <numeric value>
MMEMory:COPY <file_source>,<file_destination>
SENSE:BANDwidth|BWIDth[:RESolution] <numeric_value>
SENSe:FREQuency:STOP <numeric value>
SENSe:LIST:FREQuency <numeric_value>{,<numeric_value>}
Long and short form
The mnemonics feature a long form and a short form. The short form is marked by upper
case letters, the long form corresponds to the complete word. Either the short form or the
long form can be entered; other abbreviations are not permitted.
Example: 
HardCOPy:DEVice:COLor ON is equivalent to HCOP:DEV:COL ON
Case sensitivity
Upper case and lower case notation only serves to distinguish the two forms in the man-
ual, the instrument itself is case-insensitive.
Numeric suffixes
If a command can be applied to multiple instances of an object, e.g. specific channels or
sources, the required instances can be specified by a suffix added to the command.
Numeric suffixes are indicated by angular brackets (<1...4>, <n>, <i>) and are replaced
by a single value in the command. Entries without a suffix are interpreted as having the
suffix 1.
SCPI Command Structure
Documents you may be interested
Documents you may be interested