Remote Control Commands
R&S®EX-IQ-BOX
305
User Manual 1409.5570.32 ─ 05
Parameters:
<State>
OFF | ON
*RST: 
OFF
Example: 
SOUR1:EBOX:CPRI:HW:HSSS:OUTP:SFP2 ON
activates the SFP2 interface for signal output.
Manual operation: See "Outputs" on page 160
[:SOURce<hw>]:EBOX:CPRI:HW:HSSS:OUTPut:SMA[:STATe] <State>
Determines the SMA (SubMiniatur version A) interface for signal output, i.e. aligns the
transmitter type of the DUT. You can activate more than one interface for the signal output
simultaneously, e.g. set SFP1 for signal transmission to the DUT, and SMA for signal
monitoring.
Parameters:
<State>
OFF | ON
*RST: 
OFF
Example: 
SOUR1:EBOX:CPRI:HW:HSSS:OUTP:SMA ON
activates the interface for signal output.
Manual operation: See "Outputs" on page 160
[:SOURce<hw>]:EBOX:CPRI:HW:LBRate <Lbrate>
Selects the line bit rate for the communication link.
The available line speed depends on the CPRIbreakout board version, see ​"CPRI Line
Bit Rate" on page 157
Parameters:
<Lbrate>
LR2X | LR4X | LR5X | LRAuto | LR1X | LR8X | LR10X | LR16X
LR1X
1x614.4 Mbit/s
LR2X
2x1228.8 Mbit/s
LR4X
4x2457.6 Mbit/s
LR5X
5x3072.0 Mbit/s
LR8X
5x4915.2 Mbit/s
LR10X
5x6144.0 Mbit/s
LR16X
5x9830.4 Mbit/s
*RST: 
LR2X
CPRI Remote Control Commands
Converter pdf to html - Convert PDF to html files in C#.net, ASP.NET MVC, WinForms, WPF application
How to Convert PDF to HTML Webpage with C# PDF Conversion SDK
convert pdf into html online; convert pdf to webpage
Converter pdf to html - VB.NET PDF Convert to HTML SDK: Convert PDF to html files in vb.net, ASP.NET MVC, WinForms, WPF application
PDF to HTML Webpage Converter SDK for VB.NET PDF to HTML Conversion
adding pdf to html; how to convert pdf to html email
Remote Control Commands
R&S®EX-IQ-BOX
306
User Manual 1409.5570.32 ─ 05
Example: 
SOUR1:EBOX:CPRI:HW:LBR LR2X
sets the line bit rate 2x1228.8 Mbit/s.
Manual operation: See "CPRI Line Bit Rate" on page 157
[:SOURce<hw>]:EBOX:CPRI:HW:LOF?
Queries the CPRI frame delimiter K28.5.
Return values: 
<Lof>
0 | 1
*RST: 
0
Example: 
SOUR1:EBOX:CPRI:HW:LOF?
check if the frame delimiter K28.5 is detected.
Response: 1 (0 = no error)
the K28.5 frame delimiter is not found.
Usage: 
Query only
Manual operation: See "LOF (Loss of Frame)" on page 168
[:SOURce<hw>]:EBOX:CPRI:HW:LOS?
The command queries data synchronization.
Return values: 
<Los>
0 | 1
*RST: 
0
Example: 
SOUR1:EBOX:CPRI:HW:LOS?
check if data is synchronized.
Response: 1 (0 = no error)
data lost their synchronization.
Usage: 
Query only
Manual operation: See "LOS (Loss of Signal)" on page 168
[:SOURce<hw>]:EBOX:CPRI:HW:PROTocol <Protocol>
Determines the CPRI protocol version. The protocol version depends on the selected line
bit rate. See ​"Protocol Version" on page 158 for an overview.
Parameters:
<Protocol>
V1 | V2
Example: 
EBOX:CPRI:HW:PROT V2
Selects CPRI version 2.0.
Manual operation: See "Protocol Version" on page 158
CPRI Remote Control Commands
Online Convert PDF to HTML5 files. Best free online PDF html
Our PDF to HTML converter library control is a 100% clean .NET document image solution, which is designed to help .NET developers convert PDF to HTML webpage
convert pdf to website html; convert pdf to html open source
C#: How to Use SDK to Convert Document and Image Using XDoc.
You may use our converter SDK to easily convert PDF, Word, Excel, PowerPoint, Tiff, and Dicom files to raster images like Jpeg, Png, Bmp and Gif.
how to convert pdf into html; convert pdf to web form
Remote Control Commands
R&S®EX-IQ-BOX
307
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:CPRI:HW:R21T?
Displays the cable delay.
Return values: 
<r21>
float
Example: 
EBOX:CPRI:HW:R21T?
Response: '-117 ns'
Usage: 
Query only
Manual operation: See "R21" on page 157
[:SOURce<hw>]:EBOX:CPRI:HW:RCLock:CRLocked?
This command queries, if clock recovery is unlocked. Clock recovery must be locked in
CPRI REC test mode.
Return values: 
<Reclocked>
0 | 1
ON
The clock recovery is locked.
OFF
The clock recovery is unlocked.
*RST: 
1
Example: 
SOUR1:EBOX:CPRI:HW:RCL:CRL?
check if clock recovery is locked.
Response: 1
clock recovery is locked.
Usage: 
Query only
Manual operation: See "Clock Recovery Locked" on page 155
[:SOURce<hw>]:EBOX:CPRI:HW:RCLock:CSLocked?
This command queries, if the main PLL of the R&S EX-IQ-BOX is unlocked. This state
is fundamental and must be locked in every operating mode of the R&S EX-IQ-BOX.
Return values: 
<Slocked>
OFF | ON
ON
The PLL is locked.
OFF
The PLL is unlocked.
*RST: 
1
CPRI Remote Control Commands
C# Convert: PDF to Word: How to Convert Adobe PDF to Microsoft
After integrating XDoc.Converter for .NET SDK into your C# .NET project, you may directly use the following sample code to have a quick evaluation of PDF to MS
how to add pdf to website; convert pdf form to html form
XDoc.Converter for .NET, Support Documents and Images Conversion
file converter SDK supports various commonly used document and image file formats, including Microsoft Office (2003 and 2007) Word, Excel, PowerPoint, PDF, Tiff
convert pdf to html code; convert pdf to web page online
Remote Control Commands
R&S®EX-IQ-BOX
308
User Manual 1409.5570.32 ─ 05
Example: 
SOUR1:EBOX:CPRI:HW:RCL:CSL?
check if the PLL of the R&S EX-IQ-BOX is locked.
Response: 0
the PLL is unlocked.
Usage: 
Query only
Manual operation: See "Clock Synthesis Locked" on page 155
[:SOURce<hw>]:EBOX:CPRI:HW:RCLock:ROLocked?
This command checks, if the reference clock output is available.
A second synthesizer generates a 10 MHz reference clock selectable at one of the
GPIO interface connectors. The status LED turns blue, if the clock output is available.
Return values: 
<Reflocked>
0 | 1
*RST: 
1
Example: 
SOUR1:EBOX:CPRI:HW:RCL:ROL?
check if the reference clock output is available.
Response: 1
clock output is not available.
Usage: 
Query only
Manual operation: See "10MHz Reference Output Locked" on page 155
[:SOURce<hw>]:EBOX:CPRI:HW:RCLock:SOURce <Source>
Selects the reference signal source. In test mode CPRI RE, the R&S EX-IQ-BOX is
synchronized either external or internally. Test mode CPRI REC uses the reference clock
embedded in the CPRI link.
The reference clock is extracted from the CPRI link. This selection is mandatory in a CPRI
REC test setup, since the R&S EX-IQ-BOX operates in CPRI slave mode.
Applies to a CPRI REC test mode.
CPRI Remote Control Commands
VB.NET PDF Converter Library SDK to convert PDF to other file
generation, PDF content extraction and metadata editing if they integrate this VB.NET PDF converter control with For how to convert PDF to HTML document in
convert pdf to html file; convert pdf link to html
C# PDF Converter Library SDK to convert PDF to other file formats
Able to export PDF document to HTML file. Able to create convert PDF to SVG file. Remarkably, this PDF document converter control for C#.NET can
converting pdf to html; how to convert pdf to html code
Remote Control Commands
R&S®EX-IQ-BOX
309
User Manual 1409.5570.32 ─ 05
Parameters:
<Source>
RIN | INTernal | CRECovery
RIN
CPRI RE test setup with a connected R&S instrument.
The R&S EX-IQ-BOX synchronizes the CPRI communication link
with the reference signal provided by an R&S instrument. The 10
MHz signal is fed to the BNC connector REF IN on the back of the
R&S EX-IQ-BOX.
INTernal
CPRI RE test setup, if the R&S EX-IQ-BOX is working stand-
alone.
The R&S EX-IQ-BOX uses its internal reference oscillator for syn-
chronization.
CRECovery
CPRI REC test mode.
The reference clock is extracted from the CPRI link. This selection
is mandatory in a CPRI REC test setup, since the R&S EX-IQ-BOX
operates in CPRI slave mode.
*RST: 
RIN
Example: 
SOUR1:EBOX:CPRI:HW:RCL:SOUR RIN
set the internal clock source for reference.
Manual operation: See "Source" on page 155
[:SOURce<hw>]:EBOX:CPRI:HW:SEED:RECeived?
Queries the scrambling seed used by the connected transmitter.
Return values: 
<Received>
string
Example: 
EBOX:CPRI:HW:SEED:REC?
Response: '3fffffff'
Usage: 
Query only
Manual operation: See "Received Seed" on page 158
[:SOURce<hw>]:EBOX:CPRI:HW:SEED:SCRambling <Scrambling>
Specifies the scrambling seed. This parameter applies to CPRI protocol versoin 2.0
Parameters:
<Scrambling>
integer
Range: 
depends on other settings  to  depends on other set-
tings
Example: 
EBOX:CPRI:HW:SEED:SCR #H3FFFFFFF
Manual operation: See "Scrambling Seed" on page 158
CPRI Remote Control Commands
C# PDF Convert to Text SDK: Convert PDF to txt files in C#.net
other industry-leading methods to convert target PDF document to other editable file formats using Visual C# code, such as, PDF to HTML converter assembly, PDF
convert pdf form to html; how to convert pdf file to html
C# PDF Convert to Jpeg SDK: Convert PDF to JPEG images in C#.net
C# PDF - Convert PDF to JPEG in C#.NET. C#.NET PDF to JPEG Converting & Conversion Control. Best and free adobe PDF to jpeg converter SDK for Visual Studio .NET
create html email from pdf; convert url pdf to word
Remote Control Commands
R&S®EX-IQ-BOX
310
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:CPRI:HW:SMA:DOSWing <Doswing>
Determines the differential output voltage swing of the CML (Current Mode Logic) trans-
mitter.
Parameters:
<Doswing>
float
Range: 
0.1  to  1.6
Increment: 0.1
*RST: 
0.8
Example: 
SOUR1:EBOX:CPRI:HW:SMA:DOSW 1.0
sets 1.0 Vpp differential output voltage of the CML.
Manual operation: See "Differential Output Swing" on page 159
[:SOURce<hw>]:EBOX:CPRI:HW:SMA:RXEGain <Rxegain>
The command sets the input gain of the equalizer. This equalizer is an active high-pass
filter, that equalizes the low-pass behaviour of the transmission medium, such as e.g. the
FR4 backplane. The gain, expressed in dB, relates to the freqeuncy of 2 GHz.
Fig. 9-4: CPRI - equalizer gain over frequency
Parameters:
<Rxegain>
float
Range: 
 to  15
Increment: 1
*RST: 
0
Default unit: dB
CPRI Remote Control Commands
Remote Control Commands
R&S®EX-IQ-BOX
311
User Manual 1409.5570.32 ─ 05
Example: 
SOUR1:EBOX:CPRI:HW:SMA:RXEG 10
sets the equalizer gain to 10 dB.
Manual operation: See "Rx Equalizer Gain" on page 159
[:SOURce<hw>]:EBOX:CPRI:HW:T14T <time>
Sets the frame timing difference between the output signal at R1 and the input signal at
R4.
Depending on the CPRI test mode you can specify either T14 or TOffset. The other
parameter is calculated accordingly.
Parameters:
<time>
float
Example: 
EBOX:CPRI:HW:T14T 180
sets T14=180 ns
Manual operation: See "T14" on page 157
[:SOURce<hw>]:EBOX:CPRI:HW:TIMing?
The command queris, if the R21 timing calculation is invalid (CPRI requirement 21).
Return values: 
<Timing>
0 | 1
*RST: 
0
Example: 
SOUR1:EBOX:CPRI:HW:TIM?
check if timing calculation is correct.
Response: 1 (0 = no error)
the R21 timing calculation is invalid.
Usage: 
Query only
[:SOURce<hw>]:EBOX:CPRI:HW:TOFFset <offset>
Sets the frame timing offset between the input signal at R2 and the output signal at R3.
Depending on the CPRI test mode you can specify either T14 or TOffset. The other
parameter is calculated accordingly.
Parameters:
<offset>
float
Example: 
EBOX:CPRI:HW:TOFF  10
sets 10 ns timing offset
Manual operation: See "TOffset" on page 157
CPRI Remote Control Commands
Remote Control Commands
R&S®EX-IQ-BOX
312
User Manual 1409.5570.32 ─ 05
9.4.4 CPRI L1 Analysis Subsystem
The L1 analysis (L1:) command subsystem contains the remote-control commands for
queryinng the link status.
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​HDLC:​BRATe?​
...............................................................312
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​LOCal:​LOF?​
.............................................................312
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​LOCal:​LOS?​
.............................................................312
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​LOCal:​RAI?​
..............................................................313
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​REMote:​LOF?​
...........................................................313
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​REMote:​LOS?​
...........................................................313
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​REMote:​RAI?​
............................................................314
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​REMote:​RESet?​
........................................................314
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​RX:​REMote:​SAP?​
...........................................................314
[:​SOURce<hw>]:​EBOX:​CPRI:​L1:​VERSion?​
......................................................................315
[:SOURce<hw>]:EBOX:CPRI:L1:HDLC:BRATe?
Queries the HDLC bit rate value during link setup.
Return values: 
<rate>
float
Example: 
EBOX:CPRI:L1:HDLC:BRAT?
Response: 4800
Usage: 
Query only
Manual operation: See "HDLC Rate" on page 166
[:SOURce<hw>]:EBOX:CPRI:L1:RX:LOCal:LOF?
Queries whether the received signal is synchronized correctly at the CPRI breakout
board, i.e. LOF (Loss of Frame).
Return values: 
<lof>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:LOC:LOF?
queris if received data is synchronized.
Response: 1 (error), data lost synchronization.
Usage: 
Query only
Manual operation: See "Local Alarms" on page 168
[:SOURce<hw>]:EBOX:CPRI:L1:RX:LOCal:LOS?
Queries whether LOS (Loss of Signal) has occurred in the received signal at the CPRI
breakout board.
Return values: 
<lo>
0 | 1 | OFF | ON
CPRI Remote Control Commands
Remote Control Commands
R&S®EX-IQ-BOX
313
User Manual 1409.5570.32 ─ 05
Example: 
SOUR1:EBOX:CPRI:L1:RX:LOC:LOS?
check if loss of signal happened.
Response: 0 (no error)
Usage: 
Query only
Manual operation: See "Local Alarms" on page 168
[:SOURce<hw>]:EBOX:CPRI:L1:RX:LOCal:RAI?
Queries the RAI (Remote Arm Indication) state at the CPRI breakout board, i.e. if any
remote part of the end-to-end link has failed.
Return values: 
<rai>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:LOC:RAI?
check if a remote part of the end-to-end link has failed.
Response: 0 if OK, 1 if not OK
Usage: 
Query only
Manual operation: See "Local Alarms" on page 168
[:SOURce<hw>]:EBOX:CPRI:L1:RX:REMote:LOF?
Queries whether the received signal is synchronized correctly at the DUT, that means
without LOF (Loss oFrame).
Return values: 
<lof>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:REM:LOF?
queries if the received data is synchronized.
Response: 1 (error), data lost synchronization.
Usage: 
Query only
Manual operation: See "Remote Alarms" on page 169
[:SOURce<hw>]:EBOX:CPRI:L1:RX:REMote:LOS?
Queries whether LOS (Loss of Signal) has occurred in the received signal at the DUT.
Return values: 
<los>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:REM:LOS?
check if loss of signal happened.
Response: 0 (no error)
Usage: 
Query only
Manual operation: See "Remote Alarms" on page 169
CPRI Remote Control Commands
Remote Control Commands
R&S®EX-IQ-BOX
314
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:CPRI:L1:RX:REMote:RAI?
Queries the RAI (Remote Arm Indication) state at the DUT, i.e. if any remote part of the
end-to-end link has failed.
Return values: 
<rai>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:REM:RAI?
check if a remote part of the end-to-end link has failed.
Response: 0 if OK, 1 if not OK
Usage: 
Query only
Manual operation: See "Remote Alarms" on page 169
[:SOURce<hw>]:EBOX:CPRI:L1:RX:REMote:RESet?
Queries whether the DUT is in reset state.
Return values: 
<reset>
0 | 1 | OFF | ON
1
CPRI RE test mode: Confirms that a reset was performed.
CPRI REC test mode: Confirms a reset request.
0
CPRI RE test mode: No reset was performed.
CPRI REC test mode: No reset is requested.
Example: 
SOUR1:EBOX:CPRI:L1:RX:REM:RES?
check the reset state of the DUT.
Usage: 
Query only
Manual operation: See "Remote Alarms" on page 169
[:SOURce<hw>]:EBOX:CPRI:L1:RX:REMote:SAP?
Queries whether the data communication failed (SDI Service Access Point Defect Indi-
cation).
Return values: 
<sap>
0 | 1 | OFF | ON
Example: 
SOUR1:EBOX:CPRI:L1:RX:REM:SAP?
check the data communication.
Response: 1 (error)
data communication failed.
Usage: 
Query only
Manual operation: See "Remote Alarms" on page 169
CPRI Remote Control Commands
Documents you may be interested
Documents you may be interested