Metabolic Analysis with LabScribe
Introduction
Metabolic parameters are detected, and functions derived from those parameters are computed in 
LabScribe’s Metabolic Advanced Analysis Module.
The Metabolic Advanced Analysis Module requires a separate license. The first time you select 
Metabolic, you will be asked for a username and a serial number. Your username and serial number 
can be found on the packing slip that is included with the metabolic kit.
The Metabolic Advanced Analysis Module can both analyze online data as they are being 
collected, and perform offline analysis of previously recorded data files.
This document includes a Step by Step tutorial for using most of the features of the Metabolic 
Advanced Analysis Module, as well as a more detailed Reference section that covers the material 
in the tutorial, and adds additional context and detail. To use the Step by Step guide, you will need a 
recording with metabolic data. In order to use the online analysis part of the module, you will need to 
be recording these parameters as you proceed through the tutorial. This file can then be saved and 
used in the offline analysis tutorial.
For users not familiar with how to use iWorx hardware and software to generate a file with metabolic 
data, a checklist for setting up the metabolic cart, and configuring LabScribe to record metabolic data, 
follows the Step by Step and Reference sections of this document. Detailed hardware and software 
instructions are also included in the individual Human Exercise lab exercises that utilize iWorx gas 
analyzers. There are also instructional videos available at iworx.com illustrating the process.
Metabolic Analysis: Step by Step
When Metabolic is chosen from the Advanced menu, a submenu opens, displaying three options: 
Calibrate Flowhead, Online Calculations, and Offline Calculations.
Metabolic submenu
1
Convert pdf to text on - SDK application project:C# PDF Convert to Text SDK: Convert PDF to txt files in C#.net, ASP.NET MVC, WinForms, WPF application
C# PDF to Text (TXT) Converting Library to Convert PDF to Text
www.rasteredge.com
Convert pdf to text on - SDK application project:VB.NET PDF Convert to Text SDK: Convert PDF to txt files in vb.net, ASP.NET MVC, WinForms, WPF application
VB.NET Guide and Sample Codes to Convert PDF to Text in .NET Project
www.rasteredge.com
Complete instructions for using the Calibrate Flowhead dialog can be found in the Spirometer 
Calibration exercise found in the Human Exercise-GA200 and the Human Exercise-GA300 folders 
of the complete LabScribe installer.
The Online Calculations and Offline Calculations options provide analysis of metabolic 
experiments that use a mixing chamber such as the experiments in the Human Exercise-GA200 and 
Human Exercise-GA300 folders of the complete LabScribe installer.
Online Calculations
To use the Online Calculations:
To perform real-time analysis of mixing chamber metabolic data, LabScribe needs to be configured as 
in the experiments in the Human Exercise-GA200 and Human Exercise-GA300 folders. The 
procedure for setting up the iWorx hardware and software is also summarized after the Reference 
section of this manual. Using the Online Calculations, it is possible to generate metabolic 
calculations in real time.
1) Configure the hardware and LabScribe software to record mixing chamber metabolic data.
2) Record a sample, autoscaling all channels. Stop recording as you configure the online analysis.
3) Choose Online Calculations from the Metabolic submenu to display the Metabolic Toolbar 
above Channel 1.
Online Metabolic Toolbar
4) Click on the down arrow on the left side of the MetabolicToolbar to display a submenu with three 
choices: Setup, AutoSize, and Set Font Size.
Metabolic Online Setup Menu
5) Choose Setup to open the Online Calculation Setup dialog.
Online Metabolic Calculations Setup dialog.
2
SDK application project:Online Convert PDF to Text file. Best free online PDF txt
Download Free Trial. Convert a PDF to Text. Easy converting! We try to make it as easy as possible to convert your PDF files to Text.
www.rasteredge.com
SDK application project:VB.NET PDF Text Extract Library: extract text content from PDF
PDF ›› VB.NET PDF: Extract PDF Text. Advanced Visual Studio .NET PDF text extraction control, built in .NET framework 2.0 and compatible with Windows system.
www.rasteredge.com
6) From the CO2 Channel menu, choose the % CO
2
channel from your recording.
7) From the O2 Channel menu, choose the % O
2
channel from your recording.
8) From the Volume Channel menu, choose the breath volume channel from your recording.
9) In order to compensate for variation from breath to breath, it is possible for LabScribe to average 
the data over a segment of time. In the text box to the right of Average, indicate the segment 
duration you would like to average for analysis. The ACSM recommends an averaging time of 
twenty seconds.
10)In the text box next to Weight, enter the weight of your subject in kilograms.
11)In the text box to the right of O2 Conc. in Inhaled Air, indicate the oxygen concentration of 
inhaled air, usually 20.9%.
12)In the text box to the right of CO2 Conc. in Inhaled Air, indicate the carbon dioxide concentration 
of inhaled air, usually 0.04%.
13)From the Calculations list, control-click on those variables you would like to record in the data 
boxes of the Metabolic Toolbar. To remove a variable after you have selected it, control-click on 
that variable. Definitions of all variables can be found in the Metabolic Analysis: Reference 
section.
14)Click OK to close the dialog.
15)Return to the Metabolic Toolbar menu (accessed by clicking the arrow) and choose AutoSize. 
The size of the data boxes and the titles will be adjusted to the number of variables you have 
chosen.
16)From the Metabolic Toolbar menu, choose Set Font Size. The font size of the data box values 
can be chosen from the dialog that opens.
17)Resume recording. The changing variables will be displayed in the data boxes of the toolbar.
Online Metabolic main window display.
18)Save the recording for offline analysis.
3
SDK application project:C# PDF Text Extract Library: extract text content from PDF file in
Text: Extract Text from PDF. |. Home ›› XDoc.PDF ›› C# PDF: Extract PDF Text. Enable extracting PDF text to another PDF file, TXT and SVG formats.
www.rasteredge.com
SDK application project:VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.
Convert PDF to HTML. Convert PDF to SVG. Convert PDF to Text. Convert PDF to JPEG. Convert PDF to Png, Gif, Bitmap Images. File &
www.rasteredge.com
Offline Calculations
The offline Metabolic Calculations dialog allows sophisticated offline analysis of previously
recorded metabolic data.
To perform offline analysis:
1) Open the recording from the online analysis or another file with previously recorded metabolic 
data.
2)Choose Offline Calculations from the Metabolic submenu to open the Offline Calculations 
dialog. The panels of this dialog can be resized by moving the mouse cursor over the 
boundaries until a double-headed arrow appears, and dragging the boundaries to resize the 
panels.
3) Familiarize yourself with the Offline Calculations dialog, pictured below.
• The tabbed configuration dialogs are on the upper left portion of the dialog.
• An XY graph window on the right displays the Metabolic Graph, showing metabolic data from 
the selected segments in the Segments Selected list.
• Between the configuration dialogs and the XY graph window is the Segments Selected list, an 
editable list of the segments that can be displayed and analyzed.
The Offline Calculations Dialog
4) The Data Table is located on the lower part of the dialog.
To configure the Channels:
1) Click on the Channels tab, opening the Channels configuration dialog.
4
SDK application project:C# PDF Convert to Jpeg SDK: Convert PDF to JPEG images in C#.net
C# PDF - Convert PDF to JPEG in C#.NET. C#.NET PDF to JPEG Converting & Conversion Control. Convert PDF to JPEG Using C#.NET. Add necessary references:
www.rasteredge.com
SDK application project:VB.NET PDF Convert to HTML SDK: Convert PDF to html files in vb.
Convert PDF to HTML. |. Home ›› XDoc.PDF ›› VB.NET PDF: PDF to HTML. Convert PDF to HTML in VB.NET Demo Code. Add necessary references:
www.rasteredge.com
The Channels configuration dialog.
2) From the CO
2
Channel menu, choose the % CO
2
channel from your recording.
3) From the O
2
Channel menu, choose the % O
2
channel from your recording.
4) From the Volume Channel menu, choose the breath volume channel from your recording.
5) From the Heart rate Channel menu, choose the calculated heart rate channel from your 
recording.
6) Optionally, choose an Energy Channel if there is one on the recording, typically a power output 
from an ergometer.
To configure the Settings:
1) Click on the Settings tab, opening the Settings configuration dialog.
The Settings configuration dialog.
2) Enter the subject’s name and the protocol being used in the Name and Protocol text boxes.
3) Enter the subject’s age and sex in the Age and Sex text boxes.
5
SDK application project:C# PDF Convert to HTML SDK: Convert PDF to html files in C#.net
easy for C# developers to convert and transform style that are included in target PDF document file original formatting and interrelation of text and graphical
www.rasteredge.com
SDK application project:VB.NET PDF Convert to Word SDK: Convert PDF to Word library in vb.
VB.NET PDF - Convert PDF to MS Office Word in VB.NET. VB.NET Tutorial for How to Convert PDF to Word (.docx) Document in VB.NET. Best
www.rasteredge.com
4) Select the amount of time to average per segment in the Average (sec) text box.
5) Enter the subject’s height (in centimeters or inches) and weight (in kilograms or pounds) in the 
Height and Weight text boxes.
6) Enter the maximum experimental heart rate your subject may attain in the Maximum Heart Rate 
text box. This is typically calculated as: 220 - the subject’s age.
7) Enter the concentrations of oxygen (usually 20.9%) and carbon dioxide (usually 0.04%) in inhaled 
air in the Insp. O2 (%) and Insp. CO2 (%) text boxes.
8) The subject’s profile can be saved for future experiments by clicking the Save Subject button at 
the top of the dialog. This profile can later be retrieved and entered by clicking the Load Subject 
button.
9) By clicking Calculate, the Metabolic Graph will appear in the XY graph window, and the Data 
Table will be populated with the metabolic parameters. Click Calculate whenever settings are 
updated.
10)The subject’s anaerobic threshold (AT) is calculated automatically, and is shown in the VCO2 vs 
VO2 XY graph. This value can be recalculated manually as described in the Metabolic Graph 
section below. Click Calculate AT to display the recalculated anaerobic threshold.
To display and analyze the Metabolic Graph:
1) Familiarize yourself with the Metabolic Graph, which will be displayed in the XY graph area for 
all the segments in the selection and is illustrated below.
• By default, the graph displayed is of VCO2, VO2, and RER over time. All the segments are 
displayed.
• Segments can be deselected (or selected) by clicking on the check box to the left of the 
segment number in the Segments Selected list to the left of the graph. The UP and DOWN 
arrows on the computer keyboard can be used to move quickly through the individual cycles.
• The specific parameters shown in the graph can be customized in the Custom Graph dialog.
The Metabolic Graph.
6
2) Click the arrow to the lower left of the XY graph window to open a menu with options for the 
displayed XY graph.
XY Graph menu.
3) Click Copy Graph to copy the current Metabolic Graph to the clipboard. It can then be pasted 
into the Journal or an external application.
4) View Metabolic displays the default graph.
5) VCO
2
vs VO
2
displays VCO
2
as a function of VO
2
. The subject’s anaerobic threshold (AT) is 
calculated automatically and is indicated on this graph, as well as the intervals over which the 
slopes before and after the aerobic/anaerobic break were determined. These intervals (indicated 
by the markers labeled 1 - 2 and 3 - 4) can be adjusted by moving the markers on the graph. 
Click Calculate AT in the Settings dialog to recalculate the anaerobic threshold based on the 
revised intervals.
6) VE vs VO
2
displays the minute ventilation as a function of VO
2
.
VCO
2
vs VO
2
XY graph with Anaerobic Threshold (AT) indicated.
7
7) VE vs VCO
2
displays the minute ventilation as a function of VCO
2
.
8) HR VCO
2
vs VO
2
displays the Heart Rate and VCO
2
as a function of VO
2
.
9) % Fat displays the calories from fat as a % of total calories burned over time.
10)VE vs Watts displays the minute ventilation as a function of Watts (requires an Energy channel).
11)HR VO
2
/HR vs Watts displays the Heart Rate and VO
2
/HR as a function of Watts (requires an 
Energy channel).
12)VO
2
VCO
2
vs Watts displays VO
2
and VCO
2
as a function of Watts (requires an Energy channel).
13)VE/VO
2
VE/VCO
2
vs Watts displays VE/VO
2
and VE/VCO
2
as a function of Watts (requires an 
Energy channel).
14)REE vs Watts displays REE as a function of Watts (requires an Energy channel).
15)Custom displays a custom graph as configured in the Custom Graph dialog.
To configure Zones:
1) Click on the Zones tab to open the Zones configuration dialog.
The 
Zones configuration dialog.
2) Different heart rate zones are displayed in different colors on the X-axis of the default graph and 
the VO
2
vs VCO
2
XY graph. The default zones are automatically calculated based on the 
subject’s maximum heart rate and will typically not need to be changed. If, however, you would 
like to change the zones from the default settings, it is possible to change an individual subject’s 
heart rate zones by changing the values in this dialog.
To configure the Custom Graph:
1) Click on the Custom Graph tab to open the Custom Graph dialog.
2) Choose the variable you would like displayed on the X-axis of the XY graph by choosing one of 
the variables in the menu next to X-axis.
3) Choose the variable(s) you would like displayed on the Y-axis by choosing them from the menus 
in the boxes of the left column next to Y-axis.
4) Choose the desired color for each Y variable, whether you would like the data displayed as dots 
or a line, and the Y-axis scale for each chosen variable.
8
The Custom Graph configuration dialog.
5) Click Plot to display your custom graph. Choosing Custom from the XY graph menu will display 
the most recently configured custom graph.
To configure the Report:
1) Click on the Report tab to open the Metabolic Report configuration dialog.
2) By clicking Create Report, a dated report is generated. This report includes subject information, 
the metabolic data in table form, and the current set of XY graphs.
The Report Configuration Dialog
3) Included in the report are data from previously saved subject data. When Save Subject is clicked 
in the Settings configuration dialog, the averaged data from the current experiment are saved as 
part of the subject profile and when Load Subject is clicked in the Settings dialog of a future 
experiment, these data are entered into the Results summary that is part of the report, as 
indicated in the figure below.
9
Results summary in Report including data from this subject’s prior experiments.
4) Click Print to print the report.
5) Click Save as HTML to save the HTML source code for the report.
To configure the Report Template:
1) Click on the Report Template tab to open the Metabolic Report Template configuration dialog. 
Here you can customize the HTML code for the report generated in the Report dialog.
The Report Template configuration dialog.
2) Edit the elements to be included in the report by deleting elements not desired, changing the 
order of the elements, and adding elements from the Insert menu.
3) Add text directly to the report template and customize the appearance of added text with the 
HTML code provided.
10