Agilent Signal Generators Programming Guide
203
Programming the Status Register System
Status Groups
Data Questionable Modulation Event Enable Register
The Data Questionable Modulation Event Enable Register lets you choose which bits in the Data 
Questionable Modulation Event Register set the summary bit (bit 7 of the Data Questionable 
Condition Register) to 1.
Command:
STATus:QUEStionable:MODulation:ENABle <value> where <value> is the sum of the decimal values 
of the bits you want to enable. 
Example:
Enable bit 4 and bit 3 so that whenever either of those bits are set to 1, the Data Questionable 
Modulation summary bit of the Data Questionable Condition Register is set to 1. Send the command 
STAT:QUES:MOD:ENAB 520 (16 + 8). 
Query:
STATus:QUEStionable:MODulation:ENABle?
Response:
Decimal value of the sum of the bits previously enabled with the 
STATus:QUEStionable:MODulation:ENABle <value> command.
Convert pdf to jpg file - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
conversion of pdf to jpg; convert pdf to jpeg on
Convert pdf to jpg file - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
.pdf to jpg; best pdf to jpg converter online
204
Agilent Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Data Questionable Calibration Status Group
NOTE
Some bits in this status group do not apply to the N5161A/62A/81A/82A/83A, E4428C, 
E8257D, E8663B, and the E8663D and return zero when queried. See Table 4-12 on page 205 
for more information.
The Data Questionable Calibration Status Group is used to determine the specific event that set bit 8 
in the Data Questionable Condition Register. This group consists of the Data Questionable Calibration 
Condition Register, the Data Questionable Calibration Transition Filters (negative and positive), the 
Data Questionable Calibration Event Register, and the Data Questionable Calibration Event Enable 
Register
This is the named 
status register for 
the E4438C. 
However, not all 
signal generator 
models use all of 
the shown events 
(i.e. some use only 
a subset of the 
E4438C’s status 
registers).
Online Convert Jpeg to PDF file. Best free online export Jpg image
Convert a JPG to PDF. You can drag and drop your JPG file in the box, and then start immediately to sort the files, try out some settings and then create the
change pdf to jpg format; best pdf to jpg converter for
Online Convert PDF to Jpeg images. Best free online PDF JPEG
Online PDF to JPEG Converter. Download Free Trial. Convert a PDF File to JPG. Drag and drop your PDF in the box above and we'll convert the files for you.
best way to convert pdf to jpg; convert .pdf to .jpg
Agilent Signal Generators Programming Guide
205
Programming the Status Register System
Status Groups
Data Questionable Calibration Condition Register
The Data Questionable Calibration Condition Register continuously monitors the calibration status of 
the signal generator. Condition registers are read only.
Data Questionable Calibration Transition Filters (negative and positive)
The Data Questionable Calibration Transition Filters specify which type of bit state changes in the 
condition register set corresponding bits in the event register. Changes can be positive (0 to 1) or 
negative (1 to 0).
Data Questionable Calibration Event Register
The Data Questionable Calibration Event Register latches transition events from the condition register 
as specified by the transition filters. Event registers are destructive read-only. Reading data from an 
event register clears the content of that register.
Table 4-12     Data Questionable Calibration Condition Register Bits
Bit
Description
0
a
a.In the N5161A/81A, E4428C, E8257D, E8663B, and the E8663D, this bit is set to 0.
I/Q Calibration Failure. A 1 in this bit position indicates that the I/Q modulation calibration experienced a failure.
1
b
b.In the N5161A/81A and N5162A/82A, this bit is set to 0.
DCFM/DCM Zero Failure. A 1 in this bit position indicates that the DCFM/DCM zero calibration routine has 
failed. This is a critical error. The output of the source has no validity until the condition of this bit is 0.
214
Unused. These bits are always set to 0.
15
Always 0.
Query:
STATus:QUEStionable:CALibration:CONDition?
Response:
The decimal sum of the bits set to 1.
Commands:
STATus:QUEStionable:CALibration:NTRansition <value> (negative transition), or 
STATus:QUEStionable:CALibration:PTRansition <value> (positive transition), where <value> is 
the sum of the decimal values of the bits you want to enable.
Queries:
STATus:QUEStionable:CALibration:NTRansition? 
STATus:QUEStionable:CALibration:PTRansition?
Query:
STATus:QUEStionable:CALibration[:EVENt]?
C# Image Convert: How to Convert Adobe PDF to Jpeg, Png, Bmp, &
C# sample code for PDF to jpg image conversion. This demo code convert PDF file all pages to jpg images. // Define input and output files path.
convert pdf to jpeg; batch pdf to jpg
C# Image Convert: How to Convert Dicom Image File to Raster Images
RasterEdge.XDoc.Office.Inner.Office03.dll. RasterEdge.XDoc.PDF.dll. RasterEdge.XDoc. PowerPoint.dll. This demo code convert dicom file all pages to jpg images.
change pdf file to jpg file; change from pdf to jpg
206
Agilent Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Data Questionable Calibration Event Enable Register
The Data Questionable Calibration Event Enable Register lets you choose which bits in the Data 
Questionable Calibration Event Register set the summary bit (bit 8 of the Data Questionable 
Condition register) to 1.
Command:
STATus:QUEStionable:CALibration:ENABle <value>, where <value> is the sum of the decimal 
values of the bits you want to enable. 
Example:
Enable bit 1 and bit 0 so that whenever either of those bits are set to 1, the Data Questionable 
Calibration summary bit of the Data Questionable Condition Register is set to 1. Send the command 
STAT:QUES:CAL:ENAB 520 (2 + 1). 
Query:
STATus:QUEStionable:CALibration:ENABle?
Response:
Decimal value of the sum of the bits previously enabled with the 
STATus:QUEStionable:CALibration:ENABle <value> command.
C# Create PDF from images Library to convert Jpeg, png images to
C# Create PDF from Raster Images, .NET Graphics and REImage File with XDoc Batch convert PDF documents from multiple image formats, including Jpg, Png, Bmp
.pdf to .jpg converter online; pdf to jpeg converter
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
Turn multipage PDF file into image files in .NET WinForm application. Convert PDF to Jpg, Png, Bmp, Gif, Tiff and Bitmap in ASP.NET.
batch pdf to jpg online; convert pdf to jpg file
Agilent Signal Generators Programming Guide
207
Programming the Status Register System
Status Groups
Data Questionable BERT Status Group
NOTE
This status group applies only to the E4438C with Option UN7. Refer to Table 4-13 on 
page 208 for more information.
The Data Questionable BERT Status Group is used to determine the specific event that set bit 12 in 
the Data Questionable Condition Register. The Data Questionable Status group consists of the Data 
Questionable BERT Condition Register, the Data Questionable BERT Transition Filters (negative and 
positive), the Data Questionable BERT Event Register, and the Data Questionable BERT Event Enable 
Register
C# WPF PDF Viewer SDK to convert and export PDF document to other
Convert PDF to image file formats with high quality, support converting PDF to PNG, JPG, BMP and GIF. C#.NET WPF PDF Viewer Tool: Convert and Export PDF.
convert pdf file to jpg on; convert pdf image to jpg online
C# TIFF: C#.NET Code to Convert JPEG Images to TIFF
Use C# Code to Convert Jpeg to Tiff. jpg", @"C:\demo2.jpg", @"C:\demo3.jpg" }; // Construct List foreach (string file in imagePaths) { Bitmap tmpBmp = new Bitmap
batch convert pdf to jpg; convert pdf document to jpg
208
Agilent Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Data Questionable BERT Condition Register
The Data Questionable BERT Condition Register continuously monitors the hardware and firmware 
status of the signal generator. Condition registers are read only.
Data Questionable BERT Transition Filters (negative and positive)
The Data Questionable BERT Transition Filters specify which type of bit state changes in the 
condition register set corresponding bits in the event register. Changes can be positive (0 to 1) or 
negative (1 to 0).
Table 4-13Data Questionable BERT Condition Register Bits
Bit
Description
0
No Clock. A 1 in this bit position indicates no clock input for more than 3 seconds.
1
No Data Change. A 1 in this bit position indicates no data change occurred during the last 200 clock signals.
2
PRBS Sync Loss. A 1 is set while PRBS synchronization is not established. *RST sets the bit to zero. 
310
Unused. These bits are always set to 0.
11
Down conv. / Demod Unlocked. A 1 in this bit position indicates that either the demodulator or the down converter 
is out of lock.
12
Demod DSP Ampl out of range. A 1 in this bit position indicates the demodulator amplitude is out of range. The 
*RST
command sets this bit to zero (0). 
13
Sync. to BCH/TCH/PDCH. If the synchronization source is BCH, a 1 in this bit position indicates BCH 
synchronization is not established; it does not indicate the TCH/PDCH synchronization status. If the sync source is 
TCH or PDCH, a 1 in this bit position indicates that TCH or PDCH synchronization is not established. 
*RST
sets this 
bit to zero.
14
Waiting for TCH/PDCH. A 1 in this bit position indicates that a TCH or PDCH midamble has not been received. This 
bit is set when bit 13 is set. The bit is also set when the TCH or PDCH synchronization was once locked and then 
lost (in this case the front panel displays “WAITING FOR TCH (or PDCH)”). 
*RST
sets this bit to zero.
15
Always 0.
Query:
STATus:QUEStionable:BERT:CONDition?
Response:
The decimal sum of the bits set to 1.
Commands:
STATus:QUEStionable:BERT:NTRansition <value> (negative transition), or 
STATus:QUEStionable:BERT:PTRansition <value> (positive transition), where 
<value> is the sum of the decimal values of the bits you want to enable.
Queries:
STATus:QUEStionable:BERT:NTRansition? STATus:QUEStionable:BERT:PTRansition? 
Agilent Signal Generators Programming Guide
209
Programming the Status Register System
Status Groups
Data Questionable BERT Event Register
The Data Questionable BERT Event Register latches transition events from the condition register as 
specified by the transition filters. Event registers are destructive read-only. Reading data from an 
event register clears the content of that register.
Data Questionable BERT Event Enable Register
The Data Questionable BERT Event Enable Register lets you choose which bits in the Data 
Questionable BERT Event Register set the summary bit (bit 3 of the Data Questionable Condition 
Register) to 1.
Query:
STATus:QUEStionable:BERT[:EVENt]?
Command:
STATus:QUEStionable:BERT:ENABle <value> where <value> is the sum of the decimal values of the 
bits you want to enable
Example:
Enable bit 11 and bit 2 so that whenever either of those bits are set to 1, the Data Questionable BERT 
summary bit of the Data Questionable Condition Register is set to 1. Send the command 
STAT:QUES:BERT:ENAB 520 (2048 + 4). 
Query:
STATus:QUEStionable:BERT:ENABle?
Response:
Decimal value of the sum of the bits previously enabled with the 
STATus:QUEStionable:BERT:ENABle <value> command.
210
Agilent Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Agilent Signal Generators Programming Guide
211
5
Creating and Downloading Waveform Files
NOTE
The ability to play externally created waveform data in the signal generator is available only 
in the N5172B/82B with Option 653 or 656, N5162A/82A with Option 651, 652 or 654, 
E4438C ESG Vector Signal Generators with Option 001, 002, 601, or 602, and E8267D PSG 
Vector Signal Generators with Option 601 or 602.
On the Agilent MXG, the internal baseband generator speed upgrade Options 670, 671, and 
672 are option upgrades that require Option 651 and 652 to have been loaded at the factory 
(refer to the Data Sheet for more information). Any references to 651, 652, or 654 are 
inclusive of 671, 672, and 674.
For the N5161A/62A the softkey menus and features mentioned in this chapter are only 
available through the Web–Enabled MXG or through SCPI commands. Refer to “Using the Web 
Browser” on page 11 and to the SCPI Command Reference.
This chapter explains how to create Arb–based waveform data and download it into the signal 
generator.
• “Overview of Downloading and Extracting Waveform Files” on page 212
• “Understanding Waveform Data” on page 213
• “Waveform Structure” on page 221
• “Waveform Phase Continuity” on page 223
• “Waveform Memory” on page 226
• “Commands for Downloading and Extracting Waveform Data” on page 232
• “Creating Waveform Data” on page 242
• “Downloading Waveform Data” on page 248
• “Loading, Playing, and Verifying a Downloaded Waveform” on page 254
• “Using the Download Utilities” on page 257
• “Downloading E443xB Signal Generator Files” on page 258
• “Programming Examples” on page 261
• “Troubleshooting Waveform Files” on page 311
212
Agilent Signal Generators Programming Guide
Creating and Downloading Waveform Files
Overview of Downloading and Extracting Waveform Files
Overview of Downloading and Extracting Waveform Files
The signal generator lets you download and extract waveform files. You can create these files either 
external to the signal generator or by using one of the internal modulation formats (ESG/PSG only). 
The signal generator also accepts waveforms files created for the earlier E443xB ESG signal generator 
models. For file extractions, the signal generator encrypts the waveform file information. The 
exception to encrypted file extraction is user–created I/Q data. The signal generator lets you extract 
this type of file unencrypted. After extracting a waveform file, you can download it into another 
Agilent signal generator that has the same option or software license required to play it. Waveform 
files consist of three items:
1. I/Q data
2. Marker data
3. File header
NOTE
This order of download is required, as the I/Q data downloads results in the overwriting of 
all of these three parts of the file.
The signal generator automatically creates the marker file and the file header if the two items are not 
part of the download. In this situation, the signal generator sets the file header information to 
unspecified (no settings saved) and sets all markers to zero (off).
There are three ways to download waveform files: FTP, programmatically or using one of three 
available free download utilities created by Agilent Technologies:
• N7622A Signal Studio Toolkit 2
http://www.agilent.com/find/signalstudio
• Agilent Waveform Download Assistant for use only with MATLAB 
http://www.agilent.com/find/downloadassistant
• Intuilink for Agilent PSG/ESG Signal Generators 
http://www.agilent.com/find/intuilink
NOTE
Agilent Intuilink is not available for the Agilent MXG.
Documents you may be interested
Documents you may be interested