mvc export to pdf : Best pdf to jpg converter application Library utility azure asp.net .net visual studio E8251-9035534-part514

Agilent Signal Generators Programming Guide
333
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
Command for Bit File Downloads
Because the signal generator adds a 10–byte file header during a bit file download, you must use the 
SCPI command shown in Table6-7. If you FTP or copy the file for the initial download, the signal 
generator does not add the 10–byte file header, and it does recognize the data in the file (no data in 
the transmitted signal).
Bit files enable you to control how many bits in the file the signal generator modulates onto the 
signal. Even with this file type, the signal generator requires that all data be contained within bytes. 
For more information on bit files, see “Bit File Type Data” on page322.
8
Send the user file data stored in the array (databuffer) to the signal generator.
iwrite() sends the data specified in databuffer to the signal generator (session identifier 
specified in id).
The third argument of iwrite(), bytesToSend, contains the length of the databuffer in bytes. 
In this example, it is 2000.
The fourth argument of iwrite(), 0, means there is no END of file indicator in the data.
In many programming languages, there are two methods to send SCPI commands and data:
Method 1 where the program stops the data download when it encounters the first zero 
(END indicator) in the data.
Method 2 where the program sends a fixed number of bytes and ignores any zeros in 
the data. This is the method used in our program.
For your programming language, you must find and use the equivalent of method two. 
Otherwise you may only achieve a partial download of the user file data.
9
Send the terminating carriage (\n) as the last byte of the waveform data.
iwrite() writes the data “\n” to the signal generator (session identifier specified in id).
The third argument of iwrite(), 1, sends one byte to the signal generator.
The fourth argument of iwrite(), 1, is the END of file indicator, which the program uses to 
terminate the data download.
To verify the user file data download, see “Command for Bit File Downloads” on page 333 and 
“Commands for Binary File Downloads” on page 334.
Table6-7 Bit File Type SCPI Commands
Type
Command Syntax
Command
:MEM:DATA:BIT <"file_name">,<bit_count>,<block_data>
This downloads the file to the signal generator.
Line
Code Description—Download User File Data
Best pdf to jpg converter - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
advanced pdf to jpg converter; convert pdf to jpeg on
Best pdf to jpg converter - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
change pdf file to jpg file; change pdf to jpg online
334
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
Command Syntax Example
The following command downloads a file that contains 17 bytes:
:MEM:DATA:BIT "new_file",131,#21702%S!4&07#8g*Y9@7
Since this command is file specific (BIT), there is no need to add the file path to the file name.
After execution of this command, the signal generator creates a file in the bit directory (memory 
catalog) named “new_file” that contains 27 bytes. Remember that the signal generator adds a 
10–byte file header to a bit file. When the signal generator uses this file, it will recognize only 
131 of the 136 bits (17 bytes) contained in the file.
For information on downloading block data, see “Downloading User Files” on page330.
Commands for Binary File Downloads
To download a user file as a binary file type means that the signal generator, when the file is 
selected for use, sees all of the data contained within the file. For more information on binary files, 
see “Binary File Type Data” on page e 325. There are two ways to download the file: to be able to 
extract the file or not. Each method uses a different SCPI command, which is shown in Table 6-8.
Query
:MEM:DATA:BIT? <"file_name">
Within the context of a program this query extracts the user file data. Executing the query 
in a command window causes it to return the following information: 
<bit_count>,<block_data>.
Query
:MEM:CAT:BIT?
This lists all of the files in the bit file directory and shows the remaining non–volatile 
memory:
<bytes used by bit files>,<available non-volatile memory>,<"file_names">
Table6-8 Binary File Type Commands
Command 
Type
Command Syntax
For 
Extraction
SCPI
:MEMory:DATA:UNPRotected "bin:file_name",<datablock> 
This downloads the file to the signal generator. You can extract the file within the 
context of a program.
FTP
a
put <file_name> /user/bin/file_name
No 
extraction
:MEM:DATA "bin:file_name",<block data>
This downloads the file to the signal generator. You cannot extract the file.
Table6-7 Bit File Type SCPI Commands
Type
Command Syntax
Online Convert PDF to Jpeg images. Best free online PDF JPEG
Online PDF to JPEG Converter. Download Free Trial. Convert a PDF File to JPG. Drag and drop your PDF in the box above and we'll convert the files for you.
convert .pdf to .jpg online; pdf to jpeg
Online Convert Jpeg to PDF file. Best free online export Jpg image
Online JPEG to PDF Converter. Download Free Trial. Convert a JPG to PDF. You can drag and drop your JPG file in the box, and then start
change from pdf to jpg on; convert pdf file into jpg format
Agilent Signal Generators Programming Guide
335
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
File Name Syntax
There are three ways to format the file name, which must also include the file path:
• "BIN:file_name"
• "file_name@BIN"
• "/user/BIN/file_name"
Command Syntax Example
The following command downloads a file that contains 34 bytes:
:MEM:DATA "BIN:new_file",#2347^%S!4&07#8g*Y9@7.?:*Ru[+@y3#_^,>l
After execution of this command, the signal generator creates a file in the Binary (Bin) directory 
(memory catalog) named “new_file” that contains 34 bytes. 
For information on downloading block data, see “Downloading User Files” on page330.
Selecting a Downloaded User File astheData Source
This section describes how to format SCPI commands for selecting a user file using commands from 
the GSM and Custom modulation formats. While the commands shown come from only two formats, 
the concept remains the same when making the data selection for any of the other real–time 
modulation formats that accept user data. To find the data selection commands for both framed and 
unframed data for the different modulation formats, see the signal generator’s SCPI Command 
Reference.
1. For TDMA formats, select either framed or unframed data:
:RADio:GSM:BURSt ON|OFF|1|0
ON(1) = framed
OFF(0) = unframed
Query
:MEM:DATA? "bin:file_name"
This returns information on the named file: <bit_count>,<block_data>. 
Within the context of a program, this query extracts the user file, provided it was 
download with the proper command.
Query
:MEM:CAT:BIN?
This lists all of the files in the bit file directory and shows the remaining 
non–volatile memory:
<bytes used by bit files>,<available non-volatile memory>,<"file_names">
a. See “FTP Procedures” on page337.
Table6-8 Binary File Type Commands
Command 
Type
Command Syntax
C# WPF PDF Viewer SDK to convert and export PDF document to other
Best PDF Viewer control as well as a powerful .NET WinForms application to image file formats with high quality, support converting PDF to PNG, JPG, BMP and
convert multiple pdf to jpg; convert pdf to 300 dpi jpg
C# PDF Convert to Images SDK: Convert PDF to png, gif images in C#
Best PDF converter SDK for Visual Studio .NET for converting PDF to image in C#.NET Support exporting PDF to multiple image forms, including Jpg, Png, Bmp
convert from pdf to jpg; convert pdf to jpg for online
336
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
2. Select the user file:
3. Configure the remaining signal parameters.
4. Turn the modulation format on:
:RADio:CUSTom:STATe On
Modulating and Activating the Carrier
Use the following commands to modulate the carrier and turn on the RF output. For a complete 
listing of SPCI commands, refer to the SCPI Command Reference.
:FREQuency:FIXed 2.5GHZ
:POWer:LEVel -10.0DBM
:OUTPut:MODulation:STATe ON
:OUTPut:STATe ON
Modifying User File Data
There are two ways to modify a file after downloading it to the signal generator:
•  Use the signal generator’s bit file editor. This works for both bit and binary files, but it converts 
a binary file to a bit file and adds a 10–byte file header. For more information on using the bit 
file editor, see the signal generator’s User’s Guide. You can also access the bit editor remotely 
using the signal generator’s web server. For web server information, refer to the Programming 
Guide.
• Use a hex editor program on your PC or UNIX workstation, as described below.
Modifying a Binary File with a Hex Editor
1. FTP the file to your PC/UNIX.
For information on using FTP, see FTP Procedures. Ensure that you use binary file transfers 
during FTP operations.
2. Modify the file using a hex editor program.
3. FTP the file to the signal generator’s BIN memory catalog (directory).
Unframed Data
:RADio:CUSTom:DATA "BIT:file_name"
:RADio:CUSTom:DATA "BIN:file_name"
Framed Data
:RADio:GSM:SLOT0|1|2|3|4|5|6|7:NORMal:ENCRyption "BIT:file_name"
:RADio:GSM:SLOT0|1|2|3|4|5|6|7:NORMal:ENCRyption "BIN:file_name"
C# Create PDF from images Library to convert Jpeg, png images to
Best and professional C# image to PDF converter SDK for Visual Studio .NET. Batch convert PDF documents from multiple image formats, including Jpg, Png, Bmp, Gif
convert pdf image to jpg online; convert .pdf to .jpg
VB.NET PDF - Convert PDF with VB.NET WPF PDF Viewer
Best WPF PDF Viewer control as well as a powerful PDF converter. PDF to image file formats with high quality, support converting PDF to PNG, JPG, BMP and GIF.
convert pdf into jpg; convert pdf to jpg file
Agilent Signal Generators Programming Guide
337
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
Modifying a Bit File with a Hex Editor
1. FTP the file to your PC/UNIX.
For information on using FTP, see FTP Procedures. Ensure that you use binary file transfers 
during FTP operations.
2. Modify the file using a hex editor program.
If you need to decrease or increase the number of bits of interest, change the file header hex 
value.
3. FTP the file to the signal generator’s BIT memory catalog (directory).
FTP Procedures
CAUTION
Avoid using the *OPC? or *WAI commands to verify that the FTP process has been 
completed. These commands can potentially hang up due to the processing of other 
SCPI parser operations. Refer to the SCPI Command Reference.
NOTE
If you are remotely FTPing files and need to verify the completion of the FTP process, then 
query the instrument by using SCPI commands such as: ':MEM:DATA:', ':MEM:CAT', '*STB?', 
'FREQ?', '*IDN?', 'OUTP:STAT?'. Refer to the SCPI Command Reference.
80 Byte File From Signal Generator
02 80 hex = 640 bits designated as bits of interest
Modified File (80 Bytes to 88 Bytes)
02 bd hex = 701 bits designated as bits of interest
Added bytes
VB.NET Create PDF from images Library to convert Jpeg, png images
Best and professional image to PDF converter SDK for Visual Studio .NET. Support create PDF from multiple image formats in VB.NET, including Jpg, Png, Bmp
batch pdf to jpg online; change pdf file to jpg online
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
Best adobe PDF to image converter SDK for Visual Studio .NET. Convert PDF documents to multiple image formats, including Jpg, Png, Bmp, Gif, Tiff, Bitmap
convert pdf image to jpg image; pdf to jpg
338
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
There are three ways to FTP a file:
• use Microsoft’s ® Internet Explorer FTP feature
• use the signal generator’s internal web server (ESG firmware  C.03.76)
• use the PC or UNIX command window
Using Microsoft’s Internet Explorer
1. Enter the signal generator’s hostname or IP address as part of the FTP URL.
ftp://<host name> or <IP address>
2. Press Enter on the keyboard or Go from the Internet Explorer window.
The signal generator files appear in the Internet Explorer window.
3. Drag and drop files between the PC and the Internet Explorer window
Using the Signal Generator’s Internal Web Server
1. Enter the signal generator’s hostname or IP address in the URL.
http://<host name> or <IP address>
2. Click the Signal Generator FTP Access button located on the left side of the window.
The signal generator files appear in the web browser’s window.
3. Drag and drop files between the PC and the browser’s window
For more information on the web server feature, refer to the Programming Guide.
Using the Command Window (PC or UNIX)
1. From the PC command prompt or UNIX command line, change to the proper directory:
• When downloading from the signal generator, the directory in which to place the file.
• When downloading to the signal generator, the directory that contains the file.
2. From the PC command prompt or UNIX command line, type ftp <instrument name>. 
Where instrument name is the signal generator’s hostname or IP address.
3. At the User: prompt, press Enter (no entry is required).
4. At the Password: prompt, press Enter (no entry is required).
5. At the ftp prompt, type the desired command:
Microsoft is a U.S registered trademark of Microsoft Corporation.
To Get a File From the Signal Generator
get /user/<directory>/<file_name1> <file_name>
To Place a File in the Signal Generator
Agilent Signal Generators Programming Guide
339
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
• <file_name1> is the name of the file as it appears in the signal generator’s directory.
• <file_name> is the name of the file as it appears in the PC/UNIX current directory.
• <directory> is the signal generator’s BIT or BIN directory.
6. At the ftp prompt, type: bye
7. At the command prompt, type: exit
Understanding Framed Transmission For Real–Time TDMA
Specifying a user file as the data source for a framed transmission provides you with an easy method 
to multiplex real data into internally generated TDMA framing. The user file fills the data fields of 
the active timeslot in the first frame, and continue to fill the same timeslot of successive frames as 
long as there is more data in the file with enough bits to fill the data field. This functionality enables 
a communications system designer to download and modulate proprietary data sequences, specific PN 
sequences, or simulate multiframe transmission such as those specified by some mobile 
communications protocols. As the example in the following figure shows, a GSM multiframe 
transmission requires 26 frames for speech.
Figure6-1 GSM Multiframe Transmission
When you select a user file as the data source for a framed transmission, the signal generator’s 
firmware loads PRAM with the framing protocol of the active TDMA format. This creates a file named 
AUTOGEN_PRAM_1 in addition to a copy of the user file. For all addresses corresponding to active 
(on) timeslots, the signal generator sets the burst bit to 1 and fills the data fields with the user file 
put <file_name> /user/<directory>/<file_name1>
To Get a File From the Signal Generator
340
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
data. Other bits are set according to the configuration selected. For inactive (off) timeslots, the signal 
generator sets the burst control bit to 0, with the data being unspecified. 
In the last byte that contains the last user file data bit, the signal generator sets the Pattern Reset bit 
to 1. This causes the user file data pattern to repeat in the next frame.
NOTE
The data in PRAM is static. Firmware writes to PRAM once for the configuration selected 
and the hardware reads this data repeatedly. Firmware overwrites the volatile PRAM 
memory to reflect the desired configuration only when the data source or TDMA format 
changes.
For example, transmitting a 228–bit user file for timeslot #1 (TS1) in a normal GSM transmission 
creates two frames. Per the standard, a GSM normal channel is 156.25 bits long, with two 57–bit data 
fields (114 user data bits total per timeslot), and 42 bits for control or signalling purposes.The user 
file completely fills timeslot #1 for two consecutive frames, and then repeats. The seven remaining 
timeslots in the GSM frame are off, as shown in Figure 6-2
Figure6-2 Mapping User File Data to a Single Timeslot
NOTE
Compliant with the GSM standard, which specifies 156.25–bit timeslots, the signal generator 
uses 156–bit timeslots and adds an extra guard bit to every fourth timeslot.
For this protocol configuration, the signal generator’s firmware loads PRAM with the bits defined in 
the following table. (These bits are part of the 32–bit word per frame bit.) The Pattern Reset bit, bit 
7, is 0 for frame one and 1 for the last byte of frame two. 
Frame
Timeslot
PRAM Word 
Offset
Data Bits
Burst Bits
Pattern Reset Bit
1
0
0 -155
0/1 (don’t care)
0 (off)
0 (off)
1
1 (on)
156 - 311
set by GSM standard (42 bits) & first 
114 bits of user file
1 (on)
0
1
2
312 - 467
0/1 (don’t care)
0
0
Agilent Signal Generators Programming Guide
341
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
1
3
468 - 624
0/1 (don’t care)
0
0
1
4
625 - 780
0/1 (don’t care)
0
0
1
5
781 - 936
0/1 (don’t care)
0
0
1
6
937 - 1092
0/1 (don’t care)
0
0
1
7
1093 - 1249
0/1 (don’t care)
0
0
2
0
1250 - 1405
0/1 (don’t care)
0
0
2
1 (on)
1406 - 1561
set by GSM standard (42 bits) & 
remaining bits of user file
1 (on)
0
2
2 through 6
1562 - 2342
0/1 (don’t care)
0
0 (off)
2
7
2343 - 2499
0/1 (don’t care)
0
1 (1 in offset 
2499 only)
Frame
Timeslot
PRAM Word 
Offset
Data Bits
Burst Bits
Pattern Reset Bit
342
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
Event 1 output is set to 0 or 1 depending on the sync out selection, which enables the EVENT 1 
output at either the beginning of the frame, beginning of a specific timeslot, or at all timeslots (SCPI 
command, :RADio:GSM:SOUT FRAME|SLOT|ALL).
Because timeslots are configured and enabled within the signal generator, a user file can be 
individually assigned to one or more timeslots. A timeslot cannot have more than one data source 
(PN sequence or user file) specified for it. The amount of user file data that can be mapped into 
hardware memory depends on both the amount of PRAM available on the baseband generator, and 
the number and size of each frame. (See “Determining Memory Usage for Custom and TDMA User 
File Data” on page327.)
PRAM adds 31 bits to each bit in a frame, which forms 32–bit words. 
The following shows how to calculate the amount of PRAM storage space required for a GSM 
superframe:
NOTE
For the total PRAM memory usage, be sure to add the number of PRAM blocks that the user 
file occupies to the PRAM file size. For more information, see “Calculating Volatile Memory 
(PRAM) Usage for Framed Data” on page328.
Bits per superframe = normal GSM timeslot  timeslot per frame  speech multiframe(TCH) 
superframe
size of normal GSM timeslot = 156.25 
bits 
timeslots per frame = 8 timeslots.
speech multiframe(TCH) = 26 frames
superframe = 51 speech multiframes
1.
Calculate the number of bits in the superframe:
156.25  8  26  51 = 1,657,500 bits
2.
Calculate the size of the PRAM file:
1,657,500 bits  4 bytes (32–bit words) = 6,630,000 bytes
3.
Calculate how much memory the PRAM file will occupy
6,630,000 bytes / 1,024 bytes per PRAM block = 6,474.6 memory blocks
4.
Round the quotient up to the next integer value
6,475 blocks  1,024 bytes per block = 6,630,400 bytes