Agilent Signal Generators Programming Guide
343
Creating and Downloading User–Data Files
User File Data (Bit/Binary) Downloads (E4438C and E8267D)
Real–Time Custom High Data Rates
Custom has two modes for processing data, serial and parallel. When the data bit–rate exceeds 
50 Mbps, the signal generator processes data in parallel mode, which means processing the data 
symbol by symbol versus bit by bit (serial). This capability exists in only the Custom format when 
using a continuous data stream. This means that it does not apply to a downloaded PRAM file type 
(covered later in this chapter).
In parallel mode, for a 256QAM modulation scheme, Custom has the capability to reach a data rate 
of up to 400 Mbps. The FIR filter width is what determines the data rate. The following table shows 
the maximum data rate for each modulation type. Because the signal generator’s maximum symbol 
rate is 50 Msps, a modulation scheme that has only 1 bit per symbol is always processed in serial 
mode.
The only external effect of the parallel mode is in the EVENT 1 output signal. In serial and parallel 
mode, the signal generator outputs a narrow pulse at the EVENT 1 connector. But in parallel mode, 
the output pulse width increases by a factor of bits–per–symbol wide, as shown in the following 
figure.
Modulation Type
Bit Rate Range for Internal Data (bit rate = symbol rate  bits per symbol)
16 Symbol Wide FIR 
Filter
32 Symbol Wide FIR 
Filter
64 Symbol Wide FIR 
Filter
BPSK, 2FSK, MSK 
1bps–50Mbps
1bps–25 Mbps
1bps–12.5Mbps
C4FM, OQPSK, 
4FSK
2bps–100Mbps
2bps–50Mbps
2bps–25Mbps
IS95 OQPSK, 
QPSK
P4DQPSK, 
IS95 QPSK
GRAYQPSK, 
4QAM
D8PSK, EDGE, 
8FSK, 8PSK
3bps–150Mbps
3bps–75Mbps
3bps–37.5Mbps
16FSK, 16PSK, 
16QAM
4bps–200Mbps
4bps–100Mbps
4bps–50Mbps
Q32AM
5bps–250Mbps
5bps–125Mbps
5bps–62.5Mbps
64QAM
6bps–300Mbps
6bps–150Mbps
6bps–75Mbps
128QAM
7bps–350Mbps
7bps–175Mbps
7bps–87.5Mbps
256QAM
8bps–400Mbps
8bps–200Mbps
8bps–100Mbps
Best convert pdf to jpg - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
convert pdf to jpg for; convert pdf document to jpg
Best convert pdf to jpg - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
batch convert pdf to jpg; convert pdf file to jpg on
344
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
NOTE
This section applies only to the E4438C with Option 001, 002, 601, or 602, and the E8267D 
with Option 601 or 602.
If you encounter problems with this section, refer to “Data Transfer Troubleshooting 
(N5162A, N5182A, E4438C and E8267D Only)” on page378.
To verify the SCPI parser’s responsiveness when remotely using the :MEM:DATA SCPI 
command to upload files, the file’s upload should be verified using the *STB? command. 
Refer to the SCPI Command Reference.
This section contains information to help you transfer user–generated PRAM data from a system 
controller to the signal generator’s PRAM. It explains how to download data directly into PRAM and 
modulate the carrier signal with the data. 
The control bits included in the PRAM file download, control the following signal functions:
• bursting
• timing signal at the EVENT 1 rear panel connector
• data pattern reset
PRAM data downloads apply to only real–time Custom and TDMA modulation formats. In the TDMA 
formats, PRAM files are available only while using the unframed data selection. The following table 
on page345 shows which signal generator models support these formats.
20 ns
32QAM (5 bits per symbol)
100 ns
10 Msps
10.000001 Msps
bit rate = bits per symbol x symbol rate
NOTE: The pulse widths values are only for example purposes. The actual width may vary from the above values.
Online Convert Jpeg to PDF file. Best free online export Jpg image
Online JPEG to PDF Converter. Download Free Trial. Convert a JPG to PDF. You can drag and drop your JPG file in the box, and then start
to jpeg; convert pdf to jpg for online
Online Convert PDF to Jpeg images. Best free online PDF JPEG
Download Free Trial. Convert a PDF File to JPG. Easy converting! We try to make it as easy as possible to convert your PDF files to JPG.
convert pdf into jpg format; best pdf to jpg converter
Agilent Signal Generators Programming Guide
345
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
PRAM files differ from bit and binary user files. 
Bit and binary user files (see page321) download to non–volatile memory and the signal generator 
loads the user file data into PRAM (volatile/waveform memory) for use. The signal generator adds the 
required control bits when it generates the signal.
A PRAM file downloads directly into PRAM, and it includes seven of the required control bits for 
each data (payload) bit. The signal generator adds the remaining control bits when it generates the 
signal. You download the file using either a list or block data format. Programs such as MATLAB or 
MathCad can generate the data. 
This type of signal control enables you to design experimental or proprietary framing schemes.
After selecting the PRAM file, the signal generator builds the modulation scheme by reading data 
stored in PRAM, and constructing framing protocols according to the PRAM file data and the 
modulation format. You can manipulate PRAM data by changing the standard protocols for a 
modulation format such as the symbol rate, modulation type, and filter either through the front panel 
interface or with SCPI commands.
Understanding PRAM Files
The term PRAM file comes from earlier Agilent products, the E443xB ESGs. PRAM is another term 
for waveform memory (WFM1), which is also known as volatile memory. This means that PRAM files 
and waveform files occupy the same memory location. The signal generator’s volatile memory 
(waveform memory) storage path is /user/BBG1/waveform. For more information on memory, see 
“Signal Generator Memory” on page315.
The following figure shows a PRAM byte and illustrates the difference between it and a bit/binary 
user file byte. Notice the control bits in the PRAM byte.
Only three of the seven control bits elicit a response from the signal generator. The other four bits 
are reserved. Table 6-9 describes the bits for a PRAM byte.
E4438C ESG
E2867D PSG
Custom
a
a. For ESG, requires Option 001, 002, 601, or 602, for PSG requires Option 601 or 602.
TDMA
b
b. Real–time TDMA modulation formats require Option 402 and include EDGE, GSM, NADC, PDC, 
PHS, DECT, and TETRA.
Custom
a
Table6-9 PRAM Data Byte
Bit
Function
Value
Comments
0
Data
0/1
This is the data (payload) bit. It is “unspecified” when burst (bit 2) is set to 0.
User File Data Byte: 
MSB
Payload Bits
PRAM File Data Byte: 
1 1 0 1 0 1 0 1
Control bits
Payload bit
LSB
1 0 0 1 1 1 0 1
MSB
LSB
C# Create PDF from images Library to convert Jpeg, png images to
Best and professional C# image to PDF converter SDK for Visual Studio .NET. Batch convert PDF documents from multiple image formats, including Jpg, Png, Bmp, Gif
batch pdf to jpg; pdf to jpeg converter
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
Best adobe PDF to image converter SDK for Visual Studio .NET. Convert PDF documents to multiple image formats, including Jpg, Png, Bmp, Gif, Tiff, Bitmap
convert pdf file to jpg format; best pdf to jpg converter online
346
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
As seen in Table6-9, only four bits, shown in the following list, can change state:
• bit 0—data
• bit 2—bursting
• bit 6—EVENT 1 rear panel output
• bit 7—pattern reset
Because a PRAM byte has only four bits that can change states, there are only 15 possible byte 
patterns as shown in Table 6-10. The table also shows the decimal value for each pattern, which is 
needed for downloading data using the list format shown on page 350.
1
Reserved
0
Always 0
2
Burst
0/1
1 = RF on 
0 = RF off 
For non–bursted, non–TDMA systems, to have a continuous signal, set this bit to 1 for all 
bytes. For framed data, set this bit to 1 for on timeslots and 0 for off timeslots.
3
Reserved
0
Always 0
4
Reserved
1
Always 1
5
Reserved
0
Always 0
6
EVENT1 
Output
0/1
To have the signal generator output a single pulse at the EVENT 1 connector, set this bit 
to 1. Use this output for functions such as a triggering external hardware to indicate when 
the data pattern begins and restarts, or creating a data–synchronous pulse train by 
toggling this bit in alternate bytes.
7
Pattern Reset
0/1
0 = continue to next sequential memory address.
1 = end of memory and restart memory playback.
This bit is set to 0 for all bytes except the last byte of PRAM. To restart the pattern, set 
the last byte of PRAM to 1.
Table6-9 PRAM Data Byte
Bit
Function
Value
Comments
VB.NET PDF - Convert PDF with VB.NET WPF PDF Viewer
Best WPF PDF Viewer control as well as a powerful PDF converter. Convert PDF to image file formats with high quality, support converting PDF to PNG, JPG
convert pdf images to jpg; changing file from pdf to jpg
VB.NET Create PDF from images Library to convert Jpeg, png images
Best and professional image to PDF converter SDK Components to batch convert PDF documents in Visual Basic Support create PDF from multiple image formats in VB
.net pdf to jpg; convert pdf file into jpg
Agilent Signal Generators Programming Guide
347
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
Viewing the PRAM Waveform
After the waveform data is written to PRAM, the data pattern can be viewed using an oscilloscope. 
There is approximately a 12–symbol delay between a state change in the burst bit and the 
corresponding effect at the RF out. This delay varies with symbol rate and filter settings, and 
requires compensation to advance the burst bit in the downloaded PRAM file.
Table6-10 PRAM Byte Patterns and Bit Positions
Bit Function
P
a
t
t
e
r
n
R
e
s
e
t
E
V
E
N
T
1
O
u
t
p
u
t
R
e
s
e
r
v
e
d
(
B
i
t
=
0
)
R
e
s
e
r
v
e
d
(
B
i
t
=
1
)
R
e
s
e
r
v
e
d
(
B
i
t
=
0
)
B
u
r
s
t
R
e
s
e
r
v
e
d
(
B
i
t
=
0
)
D
a
t
a
Bit 
Pattern 
Decimal 
Value
Bit Position
7
6
5
4
3
2
1
0
---
Bit Pattern
1
1
0
1
0
1
0
1
213
1
1
0
1
0
1
0
0
212
1
1
0
1
0
0
0
1
209
1
1
0
1
0
0
0
0
208
1
0
0
1
0
1
0
1
149
1
0
0
1
0
0
0
1
145
1
0
0
1
0
0
0
0
144
0
1
0
1
0
1
0
1
85
0
1
0
1
0
1
0
0
84
0
1
0
1
0
0
0
1
81
0
1
0
1
0
0
0
0
80
0
0
0
1
0
1
0
1
21
0
0
0
1
0
1
0
0
20
0
0
0
1
0
0
0
1
17
0
0
0
1
0
0
0
0
16
C# WPF PDF Viewer SDK to convert and export PDF document to other
Best PDF Viewer control as well as a powerful .NET WinForms application Convert PDF to image file formats with high quality, support converting PDF to PNG
batch pdf to jpg converter online; bulk pdf to jpg
C# PDF Convert to Tiff SDK: Convert PDF to tiff images in C#.net
Best C#.NET PDF converter SDK for converting PDF to Tiff in Visual Studio .NET project. Also supports convert PDF files to jpg, jpeg images.
convert pdf file to jpg file; convert pdf to jpeg on
348
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
PRAM File Size
Because volatile memory resides on the baseband generator (BBG), the maximum PRAM file size 
depends on the installed baseband generator option, as shown in Table 6-11.
The maximum PRAM user file size in the table above refers to the maximum number of payload bits. 
After downloading, the signal generator translates each downloaded payload bit into a 32–bit word:
• 1 downloaded payload bit
• 7 downloaded control bits as shown in Table 6-9 on page e 345
• 24 bits added by the signal generator
The following table shows the maximum file size after the signal generator has translated the 
maximum number of payload bits into 32–bit words.
To properly size a PRAM file, you must determine the file size after the 32–bit translation process. 
The signal generator measures a PRAM file size in units of bytes; each 32–bit word equals 4 bytes.
Determining the File Size
The following example shows how to calculate a downloaded file size using a PRAM file that contains 
89 bytes (payload bits plus 7 control bits per payload bit):
89 bytes + [(89  24 bits) / 8] = 356 bytes
Because the file downloads one fourth of the translated 32–bit word, another method to calculate the 
file size is to multiply the downloaded file size by four:
89 bytes  4 = 356 bytes
See also “Signal Generator Memory” on page315 and “Checking Available Memory” on page319.
Table6-11 Maximum PRAM User File Size (Payload Bits Only)
Modulation 
Format
Baseband Generator Option
001, 601
002
602
Custom
TDMA
8 Mbits
a
a. File size with no other files residing in volatile memory.
32 Mbits
a
64 Mbits
a
Table6-12 Maximum File Size After Downloading
Modulation 
Format
Baseband Generator Option
001, 601
002
602
Custom
TDMA
32 MBytes
a
a. File size with no other files residing in volatile memory.
128 MBytes
a
256 MBytes
a
Agilent Signal Generators Programming Guide
349
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
Minimum File Size
A PRAM file requires a minimum of 60 bytes to create a signal. If the downloaded file contains less 
than 60 bytes, the signal generator replicates the file until the file size meets the 60 byte minimum. 
This replication process occurs after you select the file and turn the modulation format on. The 
following example shows this process using a downloaded 14–byte file:
• During the file download, the 14 bytes are translated into 56 bytes (fourteen 32–bit words).
14 bytes  4 = 56 bytes
• After selecting and turning the format on, the signal generator replicates the file contents to 
create the 60 byte minimum file size
60 bytes / 14 bytes = 4.29 file replications
The signal generator rounds this real value up to the next highest integer. In this example, the 
signal generator replicates the fourteen 32–bit words (56 bytes) by a factor of 5, which makes the 
final file size 280 bytes. This equates to a 70 byte file. 
14 bytes  5 = 70 bytes
70 + [(70  24) / 8] = 280 bytes
Or
56 bytes  5 = 280 bytes
File size increases
by a factor of 4
File size increases
by a factor of 5
350
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
SCPI Command for a List Format Download
Using the list format, enter the data in the command line using comma separated decimal values. 
This file type takes longer to download because the signal generator must parse the data. When 
creating the data, remember that the signal generator requires a minimum of 60 bytes. For more 
information on file size limits, see “PRAM File Size” on page e 348.
Command Syntax
:MEMory:DATA:PRAM:FILE:LIST <"file_name">,<uint8>[,<uint8>,<...>]
uint8  
The decimal equivalent of an unsigned 8–bit integer value. For a list of usable 
decimal values and their meaning with respect to the generated signal, see Table 
6-10 on page e 347.
Command Syntax Example
The following example, when executed, creates a new file in volatile (waveform) memory with the 
following attributes:
• creates a file named new_file
• outputs a single pulse at the EVENT 1 connector
• bursts the data pattern 1100 seven times over 28 bytes
• transmits 32 non–bursted bytes
• resets the data pattern so it starts again
:MEMory:DATA:PRAM:FILE:LIST <"new_file">,85,21,20,20,21,21,20,20,21,21,20,20,21,21,
20,20,21,21,20,20,21,21,20,20,21,21,20,20,16,16,16,16,16,16,16,16,16,16,16,16,16,16,
16,16,16,16,16,16,16,16,16,16,16,16,16,16,16,16,16,144
The following list defines the meaning of the different bytes seen in the command line:
SCPI Command for a Block Data Download
The IEEE standard 488.2–1992 section 7.7.6 defines block data. The signal generator is able to 
download block data significantly faster than list formatted data (see page 350), because it does not 
have to parse the data. When creating the data, remember that the signal generator requires a 
minimum of 60 bytes. For more information on file size limits, see “PRAM File Size” on page e 348.
85
Send a pulse to the EVENT 1 output, and burst the signal with a data bit of 1.
21
Burst the signal with a data bit of 1.
20
Burst the signal with a data bit of 0.
16
Do not burst the signal (RF output off), and set the data bit to 0.
144
Reset the data pattern, do not burst the signal (RF output off), and set the data bit to 0.
Agilent Signal Generators Programming Guide
351
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
Command Syntax
:MEMory:DATA:PRAM:FILE:BLOCk <"file_name">,<blockdata>
The following sections explain how to format the SCPI command for downloading block data:
• Command Syntax Example
• Command Syntax in a Program Routine
Command Syntax Example
This example conceptually describes how to format a block data download command (#ABC represents 
the block data):
:MEMory:DATA:PRAM:FILE:BLOCk <"file_name">,#ABC
<"file_name">  the file name as it will appear in the signal generator
#  
indicates the start of the block data
A  
the number of decimal digits present in B
B  
a decimal number specifying the number of data bytes to follow in C
C  
the PRAM file data in bytes
my_file  
the PRAM file name as it will appear in the signal generator’s WFM1 
memory catalog
#  
indicates the start of the block data
3  
B has three decimal digits
240  
240 bytes of data to follow in C
12%S!4&07#8g*Y9@7...  the ASCII representation of some of the block data (binary data) 
downloaded to the signal generator, however not all ASCII values are 
printable
In actual use, the block data is not part of the command line as shown above, but instead resides in 
a binary file on the PC/UNIX. When the program executes the SCPI command, the command line 
notifies the signal generator that it is going to receive block data of the stated size, and to place the 
file in the signal generator file directory with the indicated name. Immediately following the 
command execution, the program downloads the binary file to the signal generator. This is shown in 
the following section, “Command Syntax in a Program Routine”
Command Syntax in a Program Routine
This section demonstrates the use of the download SPCI command within the confines of a C++ 
program routine. The following code sends the SCPI command and downloads a 240 byte PRAM file 
to the signal generator’s WFM1 (waveform) memory catalog. This program assumes that there is a 
char array, databuffer, that contains the 240 bytes of PRAM data and that the variable numbytes 
stores the length of the array.
file_name
A
C
:MEMory:DATA:PRAM:FILE:BLOCk “my_file”,#324012%S!4&07#8g*Y9@7...
B
352
Agilent Signal Generators Programming Guide
Creating and Downloading User–Data Files
Pattern RAM (PRAM) Data Downloads (E4438C and E8267D)
Line
Code—Download PRAM File Data
1
2
3
4
5
6
7
8
9
int bytesToSend;
bytesToSend = numbytes;
char s[4];
char cmd[200];
sprintf(s, "%d", bytesToSend);
sprintf(cmd, ":MEM:DATA:PRAM:FILE:BLOCk \"FILE1\", #%d%d", strlen(s),bytesToSend);
iwrite(id, cmd, strlen(cmd), 0, 0);
iwrite(id, databuffer, bytesToSend, 0, 0);
iwrite(id, "\n", 1, 1, 0);
Line
Code Description—Download PRAM File Data
1
Define an integer variable (bytesToSend) to store the number of bytes to send to the signal 
generator.
2
Store the total number of PRAM bytes in the integer variable defined in line 1. numbytes 
contains the length of the databuffer array referenced in line 8.
3
Create a string large enough to hold the bytesToSend value as characters plus a null character 
value. In this code, string s is set to 4 bytes (3 characters for the bytesToSend value and one 
null character—one character equals one byte).
4
Create a string and set its length (cmd[200] ) to hold the SCPI command syntax and 
parameters. In this code, we define the string length as 200 bytes (200 characters).
5
Store the value of bytesToSend in string s. For this example, bytesToSend = 240; s = ”240”
6
Store the SCPI command syntax and parameters in the string cmd. The SCPI command prepares 
the signal generator to accept the data.
sprintf() is a standard function in C++, which writes string data to a string variable.
strlen() is a standard function in C++, which returns length of a string.
bytesToSend = 240, then s = “240” plus the null character, strlen(s) = 4, so 
cmd = :MEM:DATA:PRAM:FILE:BLOCk ”FILE1\” #3240.
7
Send the SCPI command stored in the string cmd to the signal generator contained in the 
variable id.
iwrite() is a SICL function in Agilent IO library, which writes the data (block data) specified 
in the string cmd to the signal generator.
The third argument of iwrite(), strlen(cmd), informs the signal generator of the number of 
bytes in the command string. The signal generator parses the string to determine the 
number of data bytes it expects to receive.
The fourth argument of iwrite(), 0, means there is no END of file indicator for the string. 
This lets the session remain open, so the program can download the PRAM file data. 