Agilent Signal Generators Programming Guide
33
Using IO Interfaces
Using LAN
Using LAN 
The Agilent X-Series and MXG are capable of 100Base–T LAN communication. The ESG, PSG, and 
E8663B are designed to connect with a 10Base–T LAN. Where auto–negotiation is present, the ESG, 
PSG, and E8663B can connect to a 100Base–T LAN, but communicate at 10Base–T speeds. For more 
information refer to http://www.ieee.org.
The signal generator can be remotely programmed through a 100Base–T LAN interface or 10Base–T 
LAN interface and LAN–connected computer using one of several LAN interface protocols. The LAN 
allows instruments to be connected together and controlled by a LAN–based computer. LAN and its 
associated interface operations are defined in the IEEE 802.2 standard. For more information refer to 
http://www.ieee.org.
NOTE
For more information on configuring your signal generator for LAN, refer to the User’s Guide 
for your signal generator. Also, for the Agilent MXG, refer to www.agilent.com and search on 
the FAQs: Hardware Configurations and Installation.
The signal generator supports the following LAN interface protocols:
• VXI–11 (See page 44)
• Sockets LAN (See page 45)
• Telephone Network (TELNET) (See page 46)
• File Transfer Protocol (FTP) (See page 50)
• LXI (See page 52)
VXI–11 and sockets LAN are used for general programming using the LAN interface, TELNET is used 
for interactive, one command at a time instrument control, and FTP is for file transfer. LXI is used 
to communicate with multiple instruments through LAN events using precision time protocols. 
For more information on the LXI standards, refer to www.agilent.com/find/.
NOTE
For more information on configuring the signal generator to communicate over the LAN, 
refer to “Using VXI–11” on page 44.
The following sections contain information on selecting and connecting IO libraries and LAN interface 
hardware that are required to remotely program the signal generator through LAN to a LAN–based 
computer and combining those choices with one of several possible LAN interface protocols.
• “Setting Up the LAN Interface” on page 34
• “Verifying LAN Functionality” on page 38
Convert multiple page pdf to jpg - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
batch convert pdf to jpg online; pdf to jpeg
Convert multiple page pdf to jpg - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
advanced pdf to jpg converter; convert pdf file to jpg file
34
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Setting Up the LAN Interface
For LAN operation, the signal generator must be connected to the LAN, and an IP address must be 
assigned to the signal generator either manually or by using DHCP client service. Your system 
administrator can tell you which method to use. (Most modern LAN networks use DHCP.)
NOTE
Verify that the signal generator is connected to the LAN using a 100Base–T LAN or 
10Base–T LAN cable.
Configuring the VXI–11 for LAN (Agilent X-Series and MXG) 
NOTE
To communicate with the signal generator over the LAN, you must enable the VXI–11 SCPI service. Select 
VXI–11 until On is highlighted. (Default condition is On.)
For optimum performance, use a 100Base–T LAN cable to connect the signal generator to the LAN. 
For details on each key, use the key help. For information describing the key help, refer to “Getting Help (Agilent X-Series and MXG)” on 
page 21 and the User’s Guide. For additional SCPI command information, refer to the SCPI Command Reference.
Utility > IO Config
C# Create PDF from images Library to convert Jpeg, png images to
Batch convert PDF documents from multiple image formats, including Jpg image formats into one or multiple PDF file in able to be cropped and pasted to PDF page.
convert pdf file to jpg online; pdf to jpeg converter
VB.NET Create PDF from images Library to convert Jpeg, png images
Components to batch convert PDF documents in Visual Basic multiple image formats into one or multiple PDF file. Crop and paste specified image area to PDF page.
convert pdf to jpg for online; convert pdf to jpg batch
Agilent Signal Generators Programming Guide
35
Using IO Interfaces
Using LAN
Configuring the VXI–11 for LAN (ESG/PSG)
Manual Configuration
The Hostname softkey is only available when LAN Config Manual DHCP is set to Manual.
To remotely access the signal generator from a different LAN subnet, you must also enter the subnet 
mask and default gateway. See your system administrator for more information.
For more information on the manual configuration, refer to “Manually Configuring the Agilent 
X-Series and MXG LAN” on page 36 or to “Manually Configuring the ESG/PSG LAN” on page 36.
NOTE
To communicate with the signal generator over the LAN, you must enable the VXI–11 SCPI service. Select VXI–11 until On is highlighted. 
(Default condition is On.)
Use a 10Base–T LAN cable to connect the signal generator to the LAN.Where auto–negotiation is present, the ESG or PSG can connect to 
100Base–T LAN, but will communicate at 10Base–T speeds. For more information refer to http://www.ieee.org.
For details on each key, use the Key and Data Field 
Reference. For additional SCPI command 
information, refer to the SCPI Command Reference.
Utility > GPIB/RS–232 
LAN
VB.NET PDF Image Extract Library: Select, copy, paste PDF images
able to perform image extraction from multiple page adobe PDF Extract multiple types of image from PDF file in Scan high quality image to PDF, tiff and various
convert pdf to 300 dpi jpg; change pdf to jpg
C# PDF Convert to Images SDK: Convert PDF to png, gif images in C#
conversions from PDF document to multiple image forms. can use this sample code to convert PDF file to PDFDocument(inputFilePath); // Get the first page of PDF
.pdf to jpg converter online; convert .pdf to .jpg
36
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Manually Configuring the Agilent X-Series and MXG LAN 
Manually Configuring the ESG/PSG LAN 
For details on each key, use the key help (described in User’s Guide). For additional SCPI command information, refer to the SCPI 
Command Reference.
Your hostname can be up to 20 characters long.
Utility > IO Config
SCPI commands:
:SYSTem:COMMunicate:LAN:CONFig MANual
:SYSTem:COMMunicate:LAN:CONFig?
For details on each key, use the Key and Data Field 
Reference.
For additional SCPI command information, refer 
to the SCPI Command Reference.
The Hostname softkey is available only when LAN Config Manual DHCP is set to 
Manual. Your hostname can be up to 20 characters long.
Utility > IO Config
SCPI commands:
:SYSTem:COMMunicate:LAN:CONFig MANual
:SYSTem:COMMunicate:LAN:CONFig?
C# WPF PDF Viewer SDK to convert and export PDF document to other
Image: Copy, Paste, Cut Image in Page. Link: Edit Create multiple pages Tiff file from PDF document. Convert PDF to image file formats with high quality, support
convert pdf into jpg; convert pdf to jpeg on
C# PDF Image Extract Library: Select, copy, paste PDF images in C#
C#.NET extract image from multiple page adobe PDF file Extract various types of image from PDF file, like Get JPG, JPEG and other high quality image files from
change pdf to jpg online; convert pdf to jpg 100 dpi
Agilent Signal Generators Programming Guide
37
Using IO Interfaces
Using LAN
DHCP Configuration
If the DHCP server uses dynamic DNS to link the hostname with the assigned IP address, the 
hostname may be used in place of the IP address. Otherwise, the hostname is not usable.
For more information on the DHCP configuration, refer to “Configuring the DHCP LAN (Agilent 
X-Series and MXG)” on page 37 or “Configuring the DHCP LAN (ESG/PSG)” on page 38.
AUTO (DHCP/Auto–IP) Configuration (Agilent X-Series and MXG)
DHCP and Auto–IP are used together to make automatic (AUTO) mode for IP configuration. 
Automatic mode attempts DHCP first and then if that fails Auto–IP is used to detect a private 
network. If neither is found, Manual is the final choice.
If the DHCP server uses dynamic DNS to link the hostname with the assigned IP address, the 
hostname may be used in place of the IP address. Otherwise, the hostname is not usable.
Auto–IP provides automatic TCP/IP set–up for instruments on any manually configured networks. 
For more information on the AUTO (DHCP/Auto–IP) configuration, refer to “Configuring the DHCP 
LAN (Agilent X-Series and MXG)” on page 37.
Configuring the DHCP LAN (Agilent X-Series and MXG)
For details on each key, use the key help (described in User’s Guide). For additional SCPI command information, refer to the SCPI 
Command Reference.
AUTO (DHCP/Auto–IP): Request a new IP address in the following sequence: 1) from the DHCP (server–based 
LAN), 2) Auto–IP (private network without a network administrator) or if neither is available, 3) Manual setting is 
selected.
DHCP: Request a new IP address from the DHCP server each power cycle.
Confirming this action configures the signal generator as a DHCP client. In DHCP mode, the signal generator 
will request a new IP address from the DHCP server upon rebooting to determine the assigned IP address.
Utility > IO Config
SCPI commands:
:SYSTem:COMMunicate:LAN:CONFig DHCP|AUTO
:SYSTem:COMMunicate:LAN:CONFig?
C# Create PDF Library SDK to convert PDF from other file formats
Gratis control for creating PDF from multiple image formats such as tiff, jpg, png, gif, bmp Create and save editable PDF with a blank page, bookmarks, links
c# convert pdf to jpg; batch pdf to jpg converter
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
Convert PDF documents to multiple image formats, including Jpg VB.NET programming sample code to convert PDF file to inputFilePath) ' Get the first page of PDF
conversion pdf to jpg; pdf to jpg converter
38
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Configuring the DHCP LAN (ESG/PSG)
Setting up Private LAN 
You can connect the Agilent X-Series, MXG, ESG, or PSG directly to a PC using a crossover cable. To 
do this, you should either choose to set IP addresses of the PC and signal generator to differ only in 
the last digit (example: PC’s IP: 1.1.1.1 and Signal generator’s IP: 1.1.1.2); or you can use the DHCP 
feature or Auto–IP feature if your PC supports them. For more information go to www.agilent.com, 
and search on the Connectivity Guide (E2094–90009) or use the Agilent Connection Expert’s Help to 
see the Connection Guide.
Verifying LAN Functionality
Verify the communications link between the computer and the signal generator remote file server 
using the ping utility. Compare your ping response to those described in “LAN Ping Responses” on 
page 40.
NOTE
For additional information on troubleshooting your LAN connection, refer to “If You Have 
Problems” on page 39 and to the Help in the Agilent IO Libraries and documentation for 
LAN connections and problems.
For details on each key, refer to the Key and Data Field 
Reference. For additional SCPI command information, 
refer to the SCPI Command Reference.
NOTE
Use a 10Base–T LAN cable to connect the signal 
generator to the LAN. 
DHCP: Request a new IP address from the DHCP server each power cycle.
Confirming this action configures the signal generator as a DHCP client. In 
DHCP mode, the signal generator will request a new IP address from the 
DHCP server upon rebooting to determine the assigned IP address.
SCPI commands:
:SYSTem:COMMunicate:LAN:CONFig DHCP
:SYSTem:COMMunicate:LAN:CONFig?
Agilent Signal Generators Programming Guide
39
Using IO Interfaces
Using LAN
From a UNIX
®
workstation, type (UNIX is a registered trademark of the Open Group):
ping <hostname or IP address> 64 10  
where <hostname or IP address> is your instrument’s name or IP address, 64 is the packet size, 
and 10 is the number of packets transmitted. Type man ping at the UNIX prompt for details on the 
ping command. 
From the MS–DOS
®
Command Prompt or Windows environment, type:
ping -n 10 <hostname or IP address> 
where <hostname or IP address> is your instrument’s name or IP address and 10 is the number of 
echo requests. Type ping at the command prompt for details on the ping command.
NOTE
In DHCP mode, if the DHCP server uses dynamic DNS to link the hostname with the 
assigned IP address, the hostname may be used in place of the IP address. Otherwise, the 
hostname is not usable and you must use the IP address to communicate with the signal 
generator over the LAN.
If You Have Problems
If you are experiencing problems with the LAN connection on the signal generator, verify the rear 
panel LAN connector green LED is on.
For additional information on troubleshooting your LAN connection, refer to the Help in the Agilent 
IO Libraries and documentation for LAN connections and problems.
NOTE
The N5161A/62A signal generators have a LAN Reset hardkey on the front panel that can be 
pressed to reset the instrument to either the factory default address conditions or to the 
manually set address conditions. 
Press and release the LAN Reset hardkey once to reset the instrument to the factory default 
conditions, with a default IP address of 169.254.9.16. If there is a DHCP server on the 
network, the N5161A/62A will get an IP address from it first. If there is no DHCP server on 
the network, the the instrument will switch to Auto-IP mode and assign the default IP 
address of 169.254.9.16.
This is equivalent to sending the following SCPI commands:
:SYSTem:COMMunicate:LAN:DEFaults
:SYSTem:COMMunicate:LAN:RESTart
Press the LAN Reset hardkey for five seconds (until the LAN LED blinks) to reset the 
instrument to the manually set conditions.
If there are multiple N5161A/62A instruments operating on the same network, one 
MS–DOS, and Visual Basic are registered trademarks of Microsoft.
40
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
instrument will use the default IP address value and the others will have randomly chosen 
IP address values. If you have not manually set the IP addresses of your N5161A/62A 
instruments (and you need to know the addresses), turn each N5161A/62A instrument on 
one at a time to set the default IP address. Then manually configure the IP address of each 
instrument using the Web-Enabled interface as described in “Using the Web Browser” on 
page 11. Repeat this process for each N5161A/62A instrument on the same network.
LAN Ping Responses
Using Interactive IO
Use the VISA Assistant utility available in the Agilent IO Libraries Suite to verify instrument 
communication over the LAN interface. Refer to the section on the “IO Libraries and Programming 
Languages” on page 4 for more information. 
The Agilent IO Libraries Suite is supported on all platforms except Windows NT. If you are using 
Windows NT, refer to the section below on using the VISA Assistant to verify LAN communication. 
See the section on “Windows NT and Agilent IO Libraries M (and Earlier)” on page 8 for more 
information. 
Normal Response for UNIX
A normal response to the ping command will be a total of 9 or 10 packets received with a 
minimal average round–trip time. The minimal average will be different from network to 
network. LAN traffic will cause the round–trip time to vary widely.
Normal Response for DOS or 
Windows
A normal response to the ping command will be a total of 9 or 10 packets received if 10 echo 
requests were specified.
Error Messages
If error messages appear, then check the command syntax before continuing with 
troubleshooting. If the syntax is correct, resolve the error messages using your network 
documentation or by consulting your network administrator.
If an unknown host error message appears, try using the IP address instead of the hostname. 
Also, verify that the host name and IP address for the signal generator have been registered 
by your IT administrator.
Check that the hostname and IP address are correctly entered in the node names database. To 
do this, enter the nslookup <hostname> command from the command prompt.
No Response 
If there is no response from a ping, no packets were received. Check that the typed address 
or hostname matches the IP address or hostname assigned to the signal generator in the 
System LAN Setup menu. For more information, refer to “Configuring the DHCP LAN (Agilent 
X-Series and MXG)” on page 37 or “Configuring the DHCP LAN (ESG/PSG)” on page 38.
Ping each node along the route between your workstation and the signal generator, starting 
with your workstation. If a node doesn’t respond, contact your IT administrator.
If the signal generator still does not respond to ping, you should suspect a hardware problem. 
Check the signal generator LAN connector lights
Verify the hostname is not being used with DHCP addressing
Intermittent Response
If you received 1 to 8 packets back, there maybe a problem with the network. In networks 
with switches and bridges, the first few pings may be lost until these devices ‘learn’ the 
location of hosts. Also, because the number of packets received depends on your network 
traffic and integrity, the number might be different for your network. Problems of this nature 
are best resolved by your IT department.
Agilent Signal Generators Programming Guide
41
Using IO Interfaces
Using LAN
NOTE
The following sections are specific to Agilent IO Libraries versions M and earlier and apply 
only to the Windows NT platform. 
42
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Using VISA Assistant
Use the VISA Assistant, available with the Agilent IO Library versions M and earlier, to communicate 
with the signal generator over the LAN interface. However, you must manually configure the VISA 
LAN client. Refer to the Help menu for instructions on configuring and running the VISA Assistant 
program.
1. Run the IO Config program.
2. Click on TCPIP0 in the Available Interface Types text box.
3. Click the Configure button. Then Click OK to use the default settings.
4. Click on TCPIP0 in the Configured Interfaces text box.
5. Click Edit...
6. Click the Edit VISA Config... button.
7. Click the Add device button.
8. Enter the TCPIP address of the signal generator. Leave the Device text box empty.
9. Click the OK button in this form and all subsequent forms to exit the IO Config program. 
If You Have Problems
1. Verify the signal generator’s IP address is valid and that no other instrument is using the IP 
address. 
2. Switch between manual LAN configuration and DHCP using the front panel LAN Config softkey and 
run the ping program using the different IP addresses. 
NOTE
For Agilent IO Libraries versions M and earlier, you must manually configure the VISA LAN 
client in the IO Config program if you want to use the VISA Assistant to verify LAN 
configuration. Refer to the IO Libraries Installation Guide for information on configuring IO 
interfaces. The IO Config program interface is shown in Figure 2-4 on page 45.