Agilent Signal Generators Programming Guide
43
Using IO Interfaces
Using LAN
Figure 2-3 3 IO Config Form (Windows NT)
Check to see that the Default Protocol is set to Automatic. 
1. Run the IO Config program.
2. Click on TCPIP in the Configured Interfaces text box. If there is no TCPIP0 in the box, follow the 
steps shown in the section “Using VISA Assistant” on page 42.
3. Click the Edit button. 
4. Click the radio button for AUTO (automatically detect protocol).
5. Click OKOK to end the IO Config program.
Convert pdf page to jpg - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
change pdf file to jpg; change pdf to jpg file
Convert pdf page to jpg - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
convert multiple pdf to jpg; convert online pdf to jpg
44
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Using VXI–11 
The signal generator supports the LAN interface protocol described in the VXI–11 standard. VXI–11 
is an instrument control protocol based on Open Network Computing/Remote Procedure Call 
(ONC/RPC) interfaces running over TCP/IP. It is intended to provide GPIB capabilities such as SRQ 
(Service Request), status byte reading, and DCAS (Device Clear State) over a LAN interface. This 
protocol is a good choice for migrating from GPIB to LAN as it has full Agilent VISA/SICL support.
NOTE
It is recommended that the VXI–11 protocol be used for instrument communication over the 
LAN interface.
Configuring for VXI–11 
The Agilent IO library has a program, IO Config, that is used to setup the computer/signal generator 
interface for the VXI–11 protocol. Download the latest version of the Agilent IO library from the 
Agilent website. Refer to the Agilent IO library user manual, documentation, and Help menu for 
information on running the IO Config program and configuring the VXI–11 interface.
Use the IO Config program to configure the LAN client. Once the computer is configured for a LAN 
client, you can use the VXI–11 protocol and the VISA library to send SCPI commands to the signal 
generator over the LAN interface. Example programs for this protocol are included in “LAN 
Programming Interface Examples” on page 118 of this programming guide.
NOTE
To communicate with the signal generator over the LAN interface you must enable the 
VXI–11 SCPI service. For more information, refer to “Configuring the DHCP LAN (Agilent 
X-Series and MXG)” on page 37 and “Configuring the DHCP LAN (ESG/PSG)” on page 38.
If you are using the Windows NT platform, refer to “Windows NT and Agilent IO Libraries M 
(and Earlier)” on page 8 for information on using Agilent IO Libraries versions M or earlier 
to configure the interface.
For Agilent IO library version J.01.0100, the “Identify devices at run–time” check box must 
be unchecked. Refer to Figure 2-4.
C# Create PDF from images Library to convert Jpeg, png images to
Batch convert PDF documents from multiple image formats, including Jpg, Png, Bmp, Gif, Tiff Any piece of area is able to be cropped and pasted to PDF page.
change from pdf to jpg; convert pdf image to jpg image
C# Image Convert: How to Convert Adobe PDF to Jpeg, Png, Bmp, &
This demo code just converts first page to jpeg image. String inputFilePath = @"C:\input.pdf"; String outputFilePath = @"C:\output.jpg"; // Convert PDF to jpg.
batch convert pdf to jpg; change pdf to jpg on
Agilent Signal Generators Programming Guide
45
Using IO Interfaces
Using LAN
Figure 2-4 4 Show Devices Form (Agilent IO Library version J.01.0100)
Using Sockets LAN 
NOTE
Users with Windows XP operating systems and newer can use this section to better 
understand how to use the signal generator with port settings. For more information, refer to 
the help software of the IO libraries being used.
Sockets LAN is a method used to communicate with the signal generator over the LAN interface 
using the Transmission Control Protocol/Internet Protocol (TCP/IP). A socket is a fundamental 
technology used for computer networking and allows applications to communicate using standard 
mechanisms built into network hardware and operating systems. The method accesses a port on the 
signal generator from which bidirectional communication with a network computer can be 
established. 
Sockets LAN can be described as an internet address that combines Internet Protocol (IP) with a 
device port number and represents a single connection between two pieces of software. The socket 
can be accessed using code libraries packaged with the computer operating system. Two common 
versions of socket libraries are the Berkeley Sockets Library for UNIX systems and Winsock for 
Microsoft operating systems.
Your signal generator implements a sockets Applications Programming Interface (API) that is 
compatible with Berkeley socket for UNIX systems, and Winsock for Microsoft systems. The signal 
generator is also compatible with other standard sockets APIs. The signal generator can be controlled 
using SCPI commands that are output to a socket connection established in your program. 
C# Image Convert: How to Convert Tiff Image to Jpeg, Png, Bmp, &
RasterEdge.XDoc.Office.Inner.Office03.dll. RasterEdge.XDoc.PDF.dll. RasterEdge.XDoc. PowerPoint.dll. This demo code will convert first page to jpeg image.
bulk pdf to jpg converter online; change file from pdf to jpg on
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
Convert PDF to Jpg, Png, Bmp, Gif, Tiff and Bitmap in ASP.NET. Dim page As PDFPage = doc.GetPage(0) ' Convert the first PDF page to a PNG file.
to jpeg; best convert pdf to jpg
46
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Before you can use sockets LAN, you must select the signal generator’s sockets port number to use:
• Standard mode. Available on port 5025. Use this port for simple programming.
• TELNET mode. The telnet SCPI service is available on port 5023.
NOTE
For backward compatibility, on the E8663B, ESG, and PSG, the signal generator also accepts 
references to the Telnet SCPI service at port 7777 and sockets SCPI service at port 7778. 
Ports 7777 and 7778 are disabled on the Agilent MXG.
For the MXG, beginning with firmware versions A.01.51 and greater, the default telnet port 
is 5024. Telnet port 5023 is still available for backwards compatibility. Refer to the SCPI 
Command Reference.
An example using sockets LAN is given in “LAN Programming Interface Examples” on page 118 of 
this programming guide.
Using Telnet LAN 
Telnet provides a means of communicating with the signal generator over the LAN. The Telnet client, 
run on a LAN connected computer, will create a login session on the signal generator. A connection, 
established between computer and signal generator, generates a user interface display screen with 
SCPI> prompts on the command line. 
Using the Telnet protocol to send commands to the signal generator is similar to communicating with 
the signal generator over GPIB. You establish a connection with the signal generator and then send 
or receive information using SCPI commands. Communication is interactive: one command at a time.
NOTE
The Windows 2000 operating systems use a command prompt style interface for the Telnet 
client. Refer to the Figure 2-7 on page 49 for an example of this interface.
Windows XP operating systems and newer can use this section to better understand how to 
use the signal generator with port settings. For more information, refer to the help software 
of the IO libraries being used.
The following telnet LAN connections are discussed:
• “Using Telnet and MS–DOS Command Prompt” on page 46
• “Using Telnet On a PC With a Host/Port Setting Menu GUI” on page 47
• “Using Telnet On Windows 2000” on page 48
• “The Standard UNIX Telnet Command” on page 49
A Telnet example is provided in “Unix Telnet Example” on page 50.
Using Telnet and MS–DOS Command Prompt
1. On your PC, click Start > Programs > Command Prompt. 
VB.NET PDF - Convert PDF with VB.NET WPF PDF Viewer
Image from PDF Page. Image: Copy, Paste, Cut Image in Page. with high quality, support converting PDF to PNG, JPG, BMP and GIF. Convert PDF to Word DOCX document.
best pdf to jpg converter for; convert pdf to jpg 300 dpi
VB.NET Create PDF from images Library to convert Jpeg, png images
Components to batch convert PDF documents in Visual Basic .NET image formats in VB.NET, including Jpg, Png, Bmp Crop and paste specified image area to PDF page.
convert pdf into jpg format; change pdf into jpg
Agilent Signal Generators Programming Guide
47
Using IO Interfaces
Using LAN
2. At the command prompt, type in telnet.
3. Press the Enter key. The Telnet display screen will be displayed.
4. Click on the Connect menu then select Remote System. A connection form (Figure 2-5) is displayed. 
Figure 2-5     Connect Form (Agilent IO Library version J.01.0100)
5. Enter the hostname, port number, and TermType then click Connect.
• Host NameIP address or hostname
• Port5023
• Term Typevt100
6. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2-6 on page 48.
7. To signal device clear, press Ctrl–C on your keyboard.
8. Select Exit from the Connect menu and type exit at the command prompt to end the Telnet 
session.
Using Telnet On a PC With a Host/Port Setting Menu GUI
1. On your PC, click Start > Run. 
2. Type telnet then click the OK button. The Telnet connection screen will be displayed. 
3. Click on the Connect menu then select Remote System. A connection form is displayed. See Figure 
2-5.
4. Enter the hostname, port number, and TermType then click Connect.
• Host Namesignal generator’s IP address or hostname
• Port5023
• Term Typevt100
5. At the SCPI> prompt, enter SCPI commands. Refer to Figure 2-6 on page 48.
6. To signal device clear, press Ctrl–C.
7. Select Exit from the Connect menu to end the Telnet session.
C# WPF PDF Viewer SDK to convert and export PDF document to other
from PDF Page. Image: Copy, Paste, Cut Image in Page. Link: Edit with high quality, support converting PDF to PNG, JPG, BMP and Convert PDF to Word DOCX document
convert pdf pages to jpg; convert pdf pictures to jpg
C# TIFF: C#.NET Code to Convert JPEG Images to TIFF
and REImage object to single or multi-page Tiff image Use C# Code to Convert Jpeg to Tiff. string[] imagePaths = { @"C:\demo1.jpg", @"C:\demo2.jpg", @"C:\demo3
changing pdf file to jpg; best pdf to jpg converter online
48
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Figure 2-6 6 Telnet Window (Windows 2000)
Using Telnet On Windows 2000
1. On your PC, click Start > Run. 
2. Type telnet in the run text box, then click the OK button. The Telnet connection screen will be 
displayed. See Figure 2-7 on page 49 (Windows 2000).
3. Type open at the prompt and then press the Enter key. The prompt will change to (to).
4. At the (to) prompt, enter the signal generator’s IP address followed by a space and 5023, which 
is the Telnet port associated with the signal generator. 
5. At the SCPI> prompt, enter SCPI commands. Refer to commands shown in Figure 2-6 on 
page 48.
6. To escape from the SCPI> session type Ctrl-].
7. Type quit at the prompt to end the Telnet session. 
Agilent Signal Generators Programming Guide
49
Using IO Interfaces
Using LAN
Figure 2-7 7 Telnet 2000 Window
The Standard UNIX Telnet Command
Synopsis
telnet [host [port]] 
Description
This command is used to communicate with another host using the Telnet protocol. When the 
command telnet is invoked with host or port arguments, a connection is opened to the host, and 
input is sent from the user to the host.
Options and Parameters
The command telnet operates in character–at–a–time or line–by–line mode. In line–by–line mode, 
typed text is echoed to the screen. When the line is completed (by pressing the Enter key), the text 
line is sent to host. In character–at–a–time mode, text is echoed to the screen and sent to host as 
it is typed. At the UNIX prompt, type man telnet to view the options and parameters available with 
the telnet command. 
NOTE
If your Telnet connection is in line–by–line mode, there is no local echo. This means you 
cannot see the characters you are typing until you press the Enter key. To remedy this, 
change your Telnet connection to character–by–character mode. Escape out of Telnet, and at 
the telnet> prompt, type mode char. If this does not work, consult your Telnet program's 
documentation.
50
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Unix Telnet Example
To connect to the instrument with host name myInstrument and port number 5023, enter the 
following command on the command line: telnet myInstrument 5023.
When you connect to the signal generator, the UNIX window will display a welcome message and a 
SCPI command prompt. The instrument is now ready to accept your SCPI commands. As you type 
SCPI commands, query results appear on the next line. When you are done, break the Telnet 
connection using an escape character. For example, Ctrl-],where the control key and the] are 
pressed at the same time. The following example shows Telnet commands:
$ telnet myinstrument 5023
Trying....
Connected to signal generator 
Escape character is ‘^]’.
Agilent Technologies, E44xx SN-US00000001
Firmware:
Hostname: your instrument
IP :xxx.xx.xxx.xxx
SCPI> 
Using FTP 
FTP allows users to transfer files between the signal generator and any computer connected to the 
LAN. For example, you can use FTP to download instrument screen images to a computer. When 
logged onto the signal generator with the FTP command, the signal generator’s file structure can be 
accessed. Figure 2-8 shows the FTP interface and lists the directories in the signal generator’s user 
level directory.
NOTE
File access is limited to the signal generator’s /user directory. 
Agilent Signal Generators Programming Guide
51
Using IO Interfaces
Using LAN
Figure 2-8 8 FTP Screen 
The following steps outline a sample FTP session from the MS–DOS Command Prompt:
1. On the PC click Start > Programs > Command Prompt.
2. At the command prompt enter: 
ftp < IP address > or < hostname >
3. At the user name prompt, press enter.
4. At the password prompt, press enter.
You are now in the signal generator’s user directory. Typing help at the command prompt will 
show you the FTP commands that are available on your system. 
5. Type quit or bye to end your FTP session.
6. Type exit to end the command prompt session.
52
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using LAN
Using LXI
NOTE
Full LXI–B feature implementation is only available on instruments with firmware >A.01.50. 
A license may be required to enable this feature and to download firmware versions 
>A.01.50. For information on new firmware releases, go to 
http://www.agilent.com/find/upgradeassistant.
This section assumes general familiarity with the LXI-C class of instruments and aims to clarify a 
number of general use cases for measurement synchronization between an MXA signal analyzer and a 
MXG signal generator using a LAN connection. Refer to http://www.lxistandard.org/home and to 
www.agilent.com/find/lxi.
This LXI section contains the following:
• “IEEE 1588” on page 52
• “Peer to Peer Messaging” on page 54
• “Configuring LXI Triggers” on page 55
• “Using the LXI Event Log” on page 57
• “Setting up and executing a list sweep measurement” on page 57
• “Synchronize a List Sweep Measurement Between an MXA and an MXG using Peer to Peer 
Messages” on page 58
• “For More Information” on page 63
IEEE 1588
The IEEE 1588 standard defines a Precision Time Protocol (PTP) for synchronizing various clocks 
connected by Ethernet. This will give your measurement instrumentation a common sense of time. To 
ensure that the MXA and MXG are operating with synchronized PTP clocks, follow the procedure 
outlined in the table below.
Time Synchronization using an MXA Signal Analyzer and a MXG Signal Generator
To ensure that the MXA and MXG are operating with synchronized PTP clocks, follow this procedure: 
1. Connect the MXA and MXG to the same Ethernet switch and the ensure that both instruments 
are configured for LAN operation.
2. Power on the MXA and MXG.
3. Send the following SCPI command to both the MXA and MXG:
:LXI:CLOCk:PTP:DOMain 1
This parameter may be any integer between 0 to 127. The default value is 0.