Agilent Signal Generators Programming Guide
63
Using IO Interfaces
Using RS–232 (ESG and PSG Only)
For More Information
For more information on using LXI see the Agilent website dedicated to LXI instrumentation:
www.agilent.com/find/lxi.
Agilent LXI Application Notes:
• Using LXI to go beyond GPIB, PXI and VXI (AN 1465–20) 
• 10 Good Reasons to Switch to LXI (AN 1465–21) 
• Transitioning from GPIB to LXI (AN 1465–22) 
• How to Use VXI and PXI in Your New LXI Test System (AN 1465–23) 
• Using Synthetic Instruments in Your Test System (AN 1465–24) 
• Migrating system software from GPIB to LAN/LXI (AN 1465–25) 
• Modifying a GPIB System to Include LAN/LXI (AN 1465–26) 
Using RS–232 (ESG and PSG Only)
CAUTION
For long strings of commands and waveform downloads, upgrading to Agilent IO 
Libraries 15.0 and above decreases RS–232 performance, resulting in an Error –310. 
NOTE
Because GPIB, LAN, and USB offer better communication performance, Agilent recommends 
that RS–232 only be used for interactive sessions or short commands.
The RS–232 serial interface is available on the ESG signal generators. 
The PSG AUXILIARY INTERFACE connector is compatible with ANSI/EIA232 (RS–232) serial 
connection but GPIB and LAN are recommended for making faster measurements and when 
downloading files. Refer to the User’s Guide.
The RS–232 serial interface can be used to communicate with the signal generator. The RS–232 
connection was once standard on most PCs but has now been replaced by USB. RS–232 can be 
connected to the signal generator’s rear panel connector using the cable described in Table 2-2 on 
page 66. Many functions provided by GPIB, with the exception of indefinite blocks, parallel polling, 
serial polling, GET, non–SCPI remote languages, SRQ, and remote mode are available using the 
RS–232 interface. 
The serial port sends and receives data one bit at a time, therefore RS–232 communication is slow. 
The data transmitted and received is usually in ASCII format with SCPI commands being sent to the 
signal generator and ASCII data returned.
Convert pdf images to jpg - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
convert multiple pdf to jpg online; .net convert pdf to jpg
Convert pdf images to jpg - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
convert from pdf to jpg; convert pdf to gif or jpg
64
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using RS–232 (ESG and PSG Only)
The following sections contain information on selecting and connecting IO libraries and RS–232 
interface hardware on the signal generator to a computer’s RS–232 connector.
• “Selecting IO Libraries for RS–232” on page 64
• “Setting Up the RS–232 Interface” on page 65
• “Verifying RS–232 Functionality” on page 67
Selecting IO Libraries for RS–232
The IO libraries can be downloaded from the National Instrument website, http://www.ni.com, or 
Agilent’s website, http://www.agilent.com. The following is a discussion on these libraries. 
CAUTION
For long strings of commands and waveform downloads, upgrading to Agilent IO 
Libraries 15.0 and above decreases RS–232 performance, resulting in an Error –310. 
Because of the potential for portability problems, running Agilent SICL without the VISA 
overlay is not recommended by Agilent Technologies.
HP Basic  
The HP Basic language has an extensive IO library that can be used to control the 
signal generator over the RS–232 interface. This library has many low level 
functions that can be used in BASIC applications to control the signal generator 
over the RS–232 interface.
VISA  
VISA is an IO library used to develop IO applications and instrument drivers that 
comply with industry standards. It is recommended that the VISA library be used 
for programming the signal generator. The NI–VISA and Agilent VISA libraries are 
similar implementations of VISA and have the same commands, syntax, and 
functions. The differences are in the lower level IO libraries used to communicate 
over the RS–232; NI–488.2 and SICL respectively. 
NI–488.2  
NI–488.2 IO libraries can be used to develop applications for the RS–232 interface. 
See National Instrument’s website for information on NI–488.2.
SICL  
Agilent SICL can be used to develop applications for the RS–232 interface. See 
Agilent’s website for information on SICL. 
Online Convert PDF to Jpeg images. Best free online PDF JPEG
it as easy as possible to convert your PDF Download and try RasterEdge.XDoc.PDF for .NET processing, converting, compressing and stroing images, documents and
.pdf to .jpg converter online; change pdf file to jpg online
Online Convert Jpeg to PDF file. Best free online export Jpg image
So, feel free to convert them too with our tool. Easy converting! It enables you to build a PDF file with one or more images.
convert pdf to jpg for; .pdf to jpg
Agilent Signal Generators Programming Guide
65
Using IO Interfaces
Using RS–232 (ESG and PSG Only)
Setting Up the RS–232 Interface
1. Setting the RS–232 Interface Baud Rate (ESG/PSG/E8663B)
NOTE
Configure your computer to use baud rates 57600 or lower only. Select the signal 
generator’s baud rate to match the baud rate of your computer or UNIX workstation or 
adjust the baud rate settings on your computer to match the baud rate setting of the 
signal generator.
The default baud rate for VISA is 9600. This baud rate can be changed with the 
“VI_ATTR_ASRL_BAUD” VISA attribute.
For details on each key, use the key help (described in User’s Guide). For additional SCPI command information, refer to the SCPI 
Command Reference.
Select a baud 
rate of 9600.
SCPI commands:
:SYSTem:COMMunicate:SERial:BAUD <number>
:SYSTem:COMMunicate:SERial:BAUD?
C# Create PDF from images Library to convert Jpeg, png images to
C# Create PDF from Raster Images, .NET Graphics and REImage File with XDoc Batch convert PDF documents from multiple image formats, including Jpg, Png, Bmp
batch pdf to jpg online; changing file from pdf to jpg
VB.NET PDF Convert to Images SDK: Convert PDF to png, gif images
VB.NET components for batch convert high resolution images from PDF. Convert PDF documents to multiple image formats, including Jpg, Png, Bmp, Gif, Tiff, Bitmap
convert pdf file into jpg format; convert pdf image to jpg
66
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using RS–232 (ESG and PSG Only)
2. Setting the RS–232 Echo Softkey
3. Connect an RS–232 cable from the computer’s serial connector to the ESG signal generator’s 
RS–232 connector or the PSG’s AUXILIARY INTERFACE connector. Refer to Table 2-2 for RS–232 cable 
information.
NOTE
Any 9 pin (male) to 9 pin (female) straight–through cable that directly wires pins 2, 3, 5, 7, 
and 8 may be used.
Table 2-2 2 RS–232 Serial Interface Cable
Quantity
Description
Agilent Part Number
1
Serial RS–232 cable 9–pin (male) to 9–pin (female)
8120–6188
For details on each key, use the key help (described in User’s Guide). For additional SCPI command information, refer to the SCPI 
Command Reference.
Toggle RS–232 Echo Off On until Off is 
highlighted. Selecting On echoes or returns 
characters sent to the signal generator and 
prints them to the display.
SCPI commands:
:SYSTem:COMMunicate:SERial:ECHO ON|OFF
:SYSTem:COMMunicate:SERial:ECHO?
C# Image Convert: How to Convert Adobe PDF to Jpeg, Png, Bmp, &
C# sample code for PDF to jpg image conversion. This demo code convert PDF file all pages to jpg images. // Define input and output files path.
change pdf file to jpg file; change pdf to jpg image
C# Image Convert: How to Convert Tiff Image to Jpeg, Png, Bmp, &
RasterEdge.XDoc.Office.Inner.Office03.dll. RasterEdge.XDoc.PDF.dll. RasterEdge.XDoc. PowerPoint.dll. This demo code convert TIFF file all pages to jpg images.
convert from pdf to jpg; change pdf to jpg image
Agilent Signal Generators Programming Guide
67
Using IO Interfaces
Using RS–232 (ESG and PSG Only)
Verifying RS–232 Functionality
You can use the HyperTerminal program available on your computer to verify the RS–232 interface 
functionality. To run the HyperTerminal program, connect the RS–232 cable between the computer 
and the signal generator and perform the following steps: 
1. On the PC click Start > Programs >  Accessories > Communications > HyperTerminal. 
2. Select HyperTerminal.
3. Enter a name for the session in the text box and select an icon.
4. Select COM1 (COM2 can be used if COM1 is unavailable).
5. In the COM1 (or COM2, if selected) properties, set the following parameters:
•  Bits per second: 9600 must match signal generator’s baud rate; for more information, refer to 
“Setting Up the RS–232 Interface” on page 65.
•  Data bits: 8
•  Parity: None
•  Stop bits: 1
•  Flow Control: None 
NOTE    Flow control, through the RTS line, is driven by the signal generator. For the purposes of 
this verification, the controller (PC) can ignore this if flow control is set to None. 
However, to control the signal generator programmatically or download files to the signal 
generator, you must enable RTS–CTS (hardware) flow control on the controller. Note that 
only the RTS line is currently used.
Software Flow Control using XON and XOFF is not supported. Only RTS–CTS hardware 
flow is supported.
6. Go to the HyperTerminal window and select File > Properties.
7. Go to Settings > Emulation and select VT100.
8. Leave the Backscroll buffer lines set to the default value.
9. Go to Settings > ASCII Setup. 
10.Check the first two boxes and leave the other boxes as default values.
Once the connection is established, enter the SCPI command *IDN? followed by <Ctrl j> in the 
HyperTerminal window. The <Ctrl j> is the new line character (on the keyboard press the Cntrl key 
and the j key simultaneously). 
The signal generator should return a string similar to the following, depending on model:
Agilent Technologies    <instrument model name and number>, US40000001,C.02.00
VB.NET Create PDF from images Library to convert Jpeg, png images
Components to batch convert PDF documents in Visual Basic .NET class. NET converter control for exporting high quality PDF from images.
convert pdf to jpg converter; convert pdf into jpg format
C# TIFF: C#.NET Code to Convert JPEG Images to TIFF
Use C# Code to Convert Jpeg to Tiff. Define Jpeg images list. string[] imagePaths = { @"C:\demo1.jpg", @"C:\demo2.jpg", @"C:\demo3.jpg" }; // Construct List
convert pdf file to jpg file; reader convert pdf to jpg
68
Agilent Signal Generators Programming Guide
Using IO Interfaces
RS–232 Programming Interface Examples
Character Format Parameters
The signal generator uses the following character format parameters when communicating through 
RS–232:
• Character Length: Eight data bits are used for each character, excluding start, stop, and parity 
bits.
• Parity Enable: Parity is disabled (absent) for each character.
• Stop Bits: One stop bit is included with each character.
If You Have Problems
1. Verify that the baud rate, parity, and stop bits are the same for the computer and signal 
generator.
2. Verify that the RS–232 cable is identical to the cable specified in Table 2-2.
3. Verify that the application is using the correct computer COM port and that the RS–232 cable is 
properly connected to that port.
4. Verify that the controller’s flow control is set to RTS–CTS.
RS–232 Programming Interface Examples
NOTE
The portions of the programming examples discussed in this section are taken from the full 
text of these programs that can be found in Chapter 3, “Programming Examples.”
• “Interface Check Using HP BASIC” on page 69
• “Interface Check Using VISA and C” on page 69
• “Queries Using HP Basic and RS–232” on page 70
• “Queries for RS–232 Using VISA and C” on page 70
Before Using the Examples
Before using the examples: On the signal generator select the following settings:
• Baud Rate – 9600 must match computer’s baud rate
• RS–232 Echo – Off
The following sections contain HP Basic and C lines of programming removed from the programming 
interface examples in Chapter 3, Programming Examples,, these portions of programming demonstrate 
the important features to consider when developing programming for use with the RS–232 interface.
NOTE
For RS–232 programming examples, refer to “RS-232 Programming Interface Examples 
(ESG/PSG Only)” on page 157
Agilent Signal Generators Programming Guide
69
Using IO Interfaces
RS–232 Programming Interface Examples
Interface Check Using HP BASIC 
This portion of the example program “Interface Check Using HP BASIC” on page 69, causes the signal 
generator to perform an instrument reset. The SCPI command *RST will place the signal generator 
into a pre–defined state. 
The serial interface address for the signal generator in this example is 9. The serial port used is 
COM1 (Serial A on some computers). Refer to “Using RS–232 (ESG and PSG Only)” on page 63 for 
more information.
The following program example is available on the signal generator’s Documentation CD–ROM as 
rs232ex1.txt. For the full text of this program, refer to “Interface Check Using HP BASIC” on 
page 157 or to the signal generator’s documentation CD–ROM.
170    CONTROL 9,0;1       ! Resets the RS-232 interface
180    CONTROL 9,3;9600    ! Sets the baud rate to match the sig gen
190    STATUS 9,4;Stat     ! Reads the value of register 4
200    Num=BINAND(Stat,7)  ! Gets the AND value
210    CONTROL 9,4;Num     ! Sets parity to NONE
220    OUTPUT 9;"*RST"     ! Outputs reset to the sig gen
Interface Check Using VISA and C
This portion of the example program “Interface Check Using VISA and C” on page 69, uses VISA 
library functions to communicate with the signal generator. The program verifies that the RS–232 
connections and interface are functional. In this example the COM2 port is used. The serial port is 
referred to in the VISA library as ‘ASRL1’ or ‘ASRL2’ depending on the computer serial port you are 
using.
The following program example is available on the signal generator Documentation CD–ROM as 
rs232ex1.cpp. For the full text of this program, refer to “Interface Check Using VISA and C” on 
page 158 or to the signal generator’s documentation CD–ROM.
int baud=9600;// Set baud rate to 9600
ViSession defaultRM, vi;// Declares a variable of type ViSession
// for instrument communication on COM 2 port
ViStatus viStatus = 0;   
// Opens session to RS-232 device at serial port 2
viStatus=viOpenDefaultRM(&defaultRM);
viStatus=viOpen(defaultRM, "ASRL2::INSTR", VI_NULL, VI_NULL, &vi);
viStatus=viEnableEvent(vi, VI_EVENT_IO_COMPLETION, VI_QUEUE,VI_NULL);
viClear(vi);// Sends device clear command
// Set attributes for the session
viSetAttribute(vi,VI_ATTR_ASRL_BAUD,baud); 
viSetAttribute(vi,VI_ATTR_ASRL_DATA_BITS,8);
70
Agilent Signal Generators Programming Guide
Using IO Interfaces
RS–232 Programming Interface Examples
Queries Using HP Basic and RS–232
This portion of the example program “Queries Using HP Basic and RS–232” on page 70, example 
program demonstrates signal generator query commands over RS–232. Query commands are of the 
type *IDN? and are identified by the question mark that follows the mnemonic.
Start HP Basic, type in the following commands, and then RUN the program:
The following program example is available on the signal generator Documentation CD–ROM as 
rs232ex2.txt. For the full text of this program, refer to “Queries Using HP Basic and RS-232” on 
page 160 or to the signal generator’s documentation CD–ROM.
190    OUTPUT 9;"*IDN?"           ! Querys the sig gen ID
200    ENTER 9;Str$               ! Reads the ID
210    WAIT 2                     ! Waits 2 seconds
220    PRINT "ID =",Str$          ! Prints ID to the screen
230    OUTPUT 9;"POW:AMPL -5 dbm" ! Sets the the power level to -5 dbm
240    OUTPUT 9;"POW?"            ! Querys the power level of the sig gen
Queries for RS–232 Using VISA and C
This portion of the example program “Queries for RS–232 Using VISA and C” on page 70, uses VISA 
library functions to communicate with the signal generator. The program verifies that the RS–232 
connections and interface are functional.
The following program example is available on the signal generator Documentation CD–ROM as 
rs232ex2.cpp. For the full text of this program, refer to “Queries for RS-232 Using VISA and C” on 
page 162 or to the signal generator’s documentation CD–ROM.
status = viOpenDefaultRM(&defaultRM);// Initializes the system
// Open communication with Serial Port 2
status = viOpen(defaultRM, "ASRL2::INSTR", VI_NULL, VI_NULL, &instr);
Agilent Signal Generators Programming Guide
71
Using IO Interfaces
Using USB (Agilent X-Series and MXG)
Using USB (Agilent X-Series and MXG)
CAUTION
USB cables are not industrial graded and potentially allows data loss in noisy 
environments.
USB cables do not have a latching mechanism and the cables can be pulled out of the 
PC or instrument relatively easily. 
The maximum length for USB cables is 30 m, including the use of inline repeaters.
NOTE
The USB interface is available only on Agilent X-Series and MXG signal generators.
The USB 2.0 interface supports USBTMC or USBTMC–USB488 specifications.
For more information on connecting instruments to the USB, refer to the Agilent Connection 
Expert in the Agilent IO Libraries Help.
USB 2.0 connectors can be used to communicate with the signal generator. The 
N5161A/62A/81A/82A is equipped with a Mini–B 5 pin rear panel connector (device USB). The 
N5171B/72B/81B/82B is equipped with a Type-B rear panel connector (device USB). Use a Type–A to 
Mini–USB 5 pin cable to connect the signal generator to the computer (Refer to “Setting Up the USB 
Interface” on page 72). Connect the Type–A front panel connector (host USB) can be used to connect 
a mouse, a keyboard, or a USB 1.1/2.0 flash drive (USB media). (Refer to the User’s Guide.) ARB 
waveform encryption of proprietary information is supported. Many functions provided by GPIB, 
including GET, non–SCPI remote languages, and remote mode are available using the USB interface.
NOTE
For a list of compatible flash drives to use with the USB external interface. Refer to 
http://www.agilent.com/find/mxg.
Do not use the Type A front panel USB to connect to a computer.
The following sections contain information on selecting and connecting I/O libraries and the USB 
interface that are required to remotely program the signal generator through the computer and 
combining those choices with one of several possible USB interface protocols.
• “Selecting I/O Libraries for USB” on page 72
• “Setting Up the USB Interface” on page 72
• “Verifying USB Functionality” on page 73
72
Agilent Signal Generators Programming Guide
Using IO Interfaces
Using USB (Agilent X-Series and MXG)
Selecting I/O Libraries for USB 
CAUTION
The Agilent X-Series and MXG’s USB interface requires Agilent IO Libraries Suite 14.1 
or newer to run properly. For more information on connecting instruments to the USB, 
refer to the Agilent Connection Expert in the Agilent IO Libraries Help.
The I/O libraries can be downloaded from the National Instrument website, http://www.ni.com, or 
Agilent’s website, http://www.agilent.com. The following is a discussion on these libraries.
NOTE
I/O applications such as IVI–COM or VXIplug&play can be used in place of VISA.
VISA  
VISA is an I/O library used to develop I/O applications and instrument drivers 
that comply with industry standards. It is recommended that the VISA library be 
used for programming the signal generator. The NI–VISA and Agilent VISA 
libraries are similar implementations of VISA and have the same commands, 
syntax, and functions. The differences are in the lower level I/O libraries used to 
communicate over the USB; NI–488.2 and SICL respectively.
NI–488.2  
NI–488.2 I/O libraries can be used to develop applications for the USB interface. 
See National Instrument’s website for information on NI–488.2.
SICL  
Agilent SICL can be used to develop applications for the USB interface. See 
Agilent’s website for information on SICL. 
CAUTION     Because of the potential for portability problems, running Agilent SICL without the 
VISA overlay is not recommended by Agilent Technologies.
Setting Up the USB Interface
Rear Panel Interface (Mini–B 5 pin)
To use USB, connect the USB cable (Refer to Table 2-3, “USB Interface Cable,” on page 72, for USB 
cable information.) between the computer and the signal generator’s rear panel Mini–B 5–pin USB 
connector. 
Table 2-3 3 USB Interface Cable
Quantity
Description
Agilent Part Number
1
USB cable Mini–B 5 pin to Type–A
82357–61601