Ver. Hospital & General (PRM board)
XXX.XXX.XXX.XXX
BAUD RATE 9600/19200
5.10
CALIBRATION FACTOR
slightly vary & by setting this value suitably one can get desired calibration accuracy. Range provided is
0.75 to 1.25 for calibration factor.
calibration factor, he may do so. Having set this & upon pressing ‘PROG’ key, the next menu option that
appears is SET UNIT.
CALIB. XX.XX
FACTOR
5.9
HARDWARE CHECK
Hardware check can be performed by or keys.
This checks LV supplies & EHT. If these are found to be alright, display prompts with an indication as
key.
This prompts a menu as ‘calibration factor’.
HW. CHK
5.7
IP ADDRESS
5.8
BAUD RATE
There are two options as indicated in LCD display.
One can select any of the two options for baud rate by using byor keys. Having done that, user can
go to next option by pressing ‘PROG’ key. The next option that appears in LCD display is hardware
check. Default setting is = 9600
5.11
SET UNIT (setting of Engineering unit)
User can select any one of the four units as required by his plant / facility. However by default, mR/hr unit
is set at factory. Selection is byor keys. The next menu option that will appear, on selection of
‘PROG’ key after this above function set, is ‘DEVICE ADDRESS’.
Default setting is = mR/hr
SET UNIT mR/hr / µSv/hr / cps / cpm
Convert multi page pdf to single jpg - Convert PDF to JPEG images in C#.net, ASP.NET MVC, WinForms, WPF project
How to convert PDF to JPEG using C#.NET PDF to JPEG conversion / converter library control SDK
convert pdf file to jpg format; changing file from pdf to jpg
Convert multi page pdf to single jpg - VB.NET PDF Convert to Jpeg SDK: Convert PDF to JPEG images in vb.net, ASP.NET MVC, WinForms, WPF project
Online Tutorial for PDF to JPEG (JPG) Conversion in VB.NET Image Application
bulk pdf to jpg converter online; convert pdf to jpg converter
8
Ver. Hospital & General (PRM board)
DEVICE
ADDRESS XXX
5.14
MAX. SCALE
Depending on the unit chosen i.e., mR/hr or µSv/hr or cps or cpm, the maximum permissible scales are
100, 1000, 2000 and 50,000 respectively. User can choose any value upto a max of this upper limit & a
value above the minimum value. 
saving settings) and default values will be loaded . In the display, chosen unit will appear & appropriate
max value desired can be entered by the user. Similar is the case for other units such as Sv/hr, cps or
cpm. After selection of max. scale, user can now select ‘PROG’ button to go to next menu option, as
follows. Default setting is = 100.00mR/hr
MAX mR/hr
(4-20mA)
XXX.XX
5.13
AUTO ACKNOWLEDGE
This is a selection option for alarm Ack
By or 
is = OFF
AUTO ACK
ON/OFF
(5 min)
5.12
DEVICE ADDRESS
Default settings is = 000
 or  keys to load this value.
Having completed this task press ‘PROG’ button for the next menu i.e., ‘AUTO ACK’
5.15
MIN. SCALE
Depending on the scale selected initially, appropriate unit (i.e typical mR/hr) appears in the display.
Any value
MIN mR/hr
(4-20mA)
XXX.XX
C# PDF Convert to Tiff SDK: Convert PDF to tiff images in C#.net
Using this .NET PDF to TIFF conversion control, C# developers can render and convert PDF document to TIFF image Both single page and multi-page Tiff image
convert multiple pdf to jpg; best pdf to jpg converter for
JPG to PNG Converter | Convert JPEG to PNG, Convert PNG to JPG
space & pixel depth; Ability to convert single-page images between JPEG How to Convert Single JPEG Image to PNG. Select "Convert to PNG"; Select "Start" to start
changing pdf to jpg on; convert pdf page to jpg
9
Ver. Hospital & General (PRM board)
Engineering
Unit Max. scale (4-20mA)
Min. scale (4-20mA)
mR/hr
100.00
000.00
Sv/hr
1000.0
0000.0
CPS
2000
0000
CPM
50000
00000
Note : 
Also, in the
visual LED display also the same scales are set. Any value exceeding this max. scale will show as over
rate, as per actual value.
Default setting is = 000.00mR/hr
5.16
LOAD DEFAULT SETTINGS
& chose option by pressing or key for default settings.
Followed by this selection & upon pressing ‘PROG’ key one would see the menu as given below.
LOAD DEFAULT
SETTINGS?
5.17
MODIFY PASSWORD
If user wants to change the pass word, he can do so by or key. Select no option to retain same
save settings. The next menu that appear as given below.
MODIFY
PASSWORD
YES/NO
5.18
SAVE SETTINGS
Use or  
dose rates and indicates in the display visually.
SAVE ?
SETTINGS
C# TIFF: C#.NET Code to Convert JPEG Images to TIFF
transform & convert various image forms, like Jpeg, Png, Bmp, Gif, and REImage object to single or multi-page Tiff image Use C# Code to Convert Jpeg to Tiff.
best way to convert pdf to jpg; batch pdf to jpg converter
JPG to Word Converter | Convert JPEG to Word, Convert Word to JPG
Open JPEG to Word Converter first; Load JPG images from local folders in "File" in toolbar Windows Explorer; How to Convert Single Word Image to JPEG.
convert pdf to 300 dpi jpg; .pdf to .jpg online
10
Ver. Hospital & General (PRM board)
CHAPTER - VI
SIDE PANEL CONNECTOR DETAILS
FOR INTERNAL ELECTRONIC UNIT:
a
Fuse
-
750mA
b.
3 pin MS Male connector (Mains)
Pin Number
Signal
1
Live
2
Neutral
3
Chasis/Earth/GND
FOR EXTERNAL ELECTRONIC UNIT :
a
Fuse
-
750mA
b.
3 pin MS Male connector (Mains)
Pin Number
Signal
1
Live
2
Neutral
3
Chasis/Earth/GND
c.
6 pin MS connector Male connector (To Detector Probe)
Pin Number
Signal
A
GND
B
+5V
C
RS-485 (“A”)
(For connecting to detector side
D
RS-485 (“B”)
RS485 Transceiver)
E
Fraction of HV
F
NC
6 pin MS (panel mount connector)
A
E
D
C
B
F
3 pin MS (panel mount connector)
1
2
3
3 pin MS (panel mount connector)
1
2
3
Internal Detector Electronic Unit
side panel view
External Detector probe
Electronic Unit side panel view
JPG to JBIG2 Converter | Convert JPEG to JBIG2, Convert JBIG2 to
Ability to preserve original images without any affecting; Ability to convert image swiftly between JPG & JBIG2 in single and batch mode;
pdf to jpeg; convert pdf file into jpg
JPG to DICOM Converter | Convert JPEG to DICOM, Convert DICOM to
Open JPEG to DICOM Converter first; Load JPG images from local folders in "File" in toolbar Windows Explorer; How to Convert Single DICOM Image to JPEG.
convert pdf to jpg file; convert multipage pdf to jpg
11
Ver. Hospital & General (PRM board)
CHAPTER - VII
CALIBRATION
Procedure for Calibration with Source
This unit namely Gamma Area Monitor Type GA 720 has been calibrated at Radiation Standard & Calibration Lab
of Nucleonix Systems (P) Ltd., using Gamma Survey Instruments Calibrator of AEA Technology, USA. It operates
Calibration Procedure with Gamma Source (Cs – 137 – 136.742mCi)
Associates. This comprises of a 164
mCi – 137 gamma source traceable to NIST.
The monitor is pre-loaded with a lookup table for computation of dose rate. To calibrate instrument with source,
Very Important Note : 
effect due to attenuator blocks at close distance (<50 cm)
When the manual is dispatched to the customer, calibration factor is to be mentioned below and also saved in the
instrument program.
Over Load Test
Passed
(a)
10 Times Highest Range
Yes/No
(b)
50 Times Highest Range
Yes/No
(c)
100 Times Highest Range
Yes/No
Calibration Factor
:__________________ Calibrated by
:________________
Calibration Date
:__________________ Instrument S.No.:________________
Calibration Due
:__________________
Tested By
:
Date
:
Expected
Doserate (mR/hr)
0.0 mR/hr
1.0 mR/hr
5.0 mR/hr
10.0 mR/hr
20.0 mR/hr
50.0 mR/hr
100.0 mR/hr
Observed Doserate (mR/hr)
C# Imaging - Planet Barcode Generation Guide
bar codes on documents such as PDF, Office Word, Excel, PowerPoint and multi-page TIFF. BarcodeHeight = 200; barcode.AutoResize = true; //convert barcode to
change from pdf to jpg; .net pdf to jpg
JPEG to PDF Converter | Convert JPEG to PDF, Convert PDF to JPEG
Open JPEG to PDF Converter first; Load JPG images from local folders in "File" in toolbar Windows Explorer; How to Convert Single PDF Image to JPEG.
change file from pdf to jpg; convert pdf photo to jpg
12
Ver. Hospital & General (PRM board)
CHAPTER - VIII
AVAILING OF MAINTENANCE/ CALIBRATION SERVICES AND WARRANTY CLAUSE
(WITH IN INDIA)
8.1
GENERAL
unit back to our works by RPP/COURIER/SPEED POST PARCEL/GATI/XPS/door delivery.  We shall
.
The equipment is to be sent to :
The Servicing Department
NUCLEONIX SYSTEMS PRIVATE LIMITED
Plot No : 162 A & B, PHASE II, I.D.A.Cherlapally,
e-mail : info@nucleonix.com 
www
.nucleonix.com
warranty
glass construction we do not provide warranty. In case of failure of these components at cost price as per
Nucleonix price list user has to buy.
the equipment once a year at our works.
8.2
EQUIPMENT REPAIRS / SERVICING POLICY (WITH IN INDIA)
(a)
During Warrantee
availing services/ repairing
facility during warrantee period.
equipments are sent to our works only. For larger systems such as installed systems, networked systems,
PC vendor is to be contacted. Also for software related problems online support will be provided. Software
support doesn't include cleaning of virus problems etc.
or loss during transportation.
account. Only services are free.
Please indicate in your correspondence equipment model & serial number.
For Door Delivery Transportation contact XPS / GATI cargo in your city / town or a reliable courier service to
Road.
After servicing, equipments will be sent back by same mode of transport such as XPS / GATI / COURIER /
RPP.
JPG to GIF Converter | Convert JPEG to GIF, Convert GIF to JPG
Open JPEG to GIF Converter first; Load JPG images from local folders in "File" in toolbar Windows Explorer; How to Convert Single GIF Image to JPEG.
convert pdf file into jpg format; best pdf to jpg converter online
13
Ver. Hospital & General (PRM board)
All types of Radiation detectors, glass ware, PMTs etc which are fragile are not covered in warrantee, if the
However complete electronics is covered for 1 year warrantee.
You can also send the equipment personally to our works for repairs either during or after warrantee, after
person can stay for a day or two & get it repaired & or calibrated.
(B)
AFTER WARRANTEE SERVICES
works, you may please observe the following procedure.
Freight charges @ actual.
for repairs, you may do so by following the steps given below.
electronic circuit boards & enclosure.
services accordingly.
, to take time & also it will cost
more which includes Engineer's TA & DA etc., apart from basic service charges + cost of spares etc.
advise you to send them to our works.
problems, to enable our engineer to come prepared with adequate spares.
For any additional information send an e-mail to info@nucleonix.com, Atten: Customer support division.
8.3
EQUIPMENT REPAIRS / SERVICING POLICY (FOR EXPORTS)
Factory acceptance tests (FA
(a)
During & after warranty:
manual / servicing manual.
, user can take corrective
action or contact Nucleonix systems by email for help.
customer if required. Skype will be used for communication.
During & after warranty
both ways.
.
After warranty, services will be similar to that of services during warranty. However, customer will have to pay
installation & minor maintenance very soon.
14
Ver. Hospital & General (PRM board)
8.4
HOW TO AVAIL CALIBRATION SERVICES [APPLICABLE ONLY FOR ALL TYPES OF RADIATION
MONITORS PORTABLE & INSTALLED SUCH AS SURVEY METERS & AREA GAMMA MONITORS]
(FOR INDIAN CUSTOMERS)
for Nucleonix Systems manufactured products only, in general, as a company policy.
How to avail calibration services:
It is best advised that each of the Radiation monitors including Area monitorsare calibrated once in a year.
When you want to send your Radiation monitor / Area monitor / Contamination monitor for calibration to our
works. You may send the equipment for calibration, by following the steps given below:
1.
meters, contamination monitors & Area Gamma Monitors) are Rs: 2500 + Packing + Freight charges. You
can email a ‘work order’ accepting these charges.
2.
details of mode of transport sent with docket particulars.
3.
calibration.
4.
You are requested to ensure good packing to avoid any transportation damages. Especially if there are
5.
/ GATI cargo / Courier / RPP / Speed Post parcel etc. Send the equipment on freight paid basis. (Equipments
sent by other methods such as Rail/Road etc will not be collected). Also you can cover for transit insurance
transportation both ways.
6.
post.
7.
mode of transportation as mentioned above.
8.5
HOW TO AVAIL CALIBRATION SERVICES (FOR FOREIGN CUSTOMERS)
if needed, for calibration. 
services.
Calibration Standards Lab & Facility:
We have two calibration labs.
i.Low Level Calibration Lab.
ii.High Dose Rate Calibration lab.
Low Level Calibration Lab:
from Amersham.
meters, contamination monitors, 
15
Ver. Hospital & General (PRM board)
Gamma Survey instruments calibrator has Cs-137 source 161.5 mCi as on 05 Aug 2002. It is basically a
gamma survey instruments calibrator procured from AEA Technologies UK / USA. Has NIST traceability
accuracy within +/- 7%
High Dose Rate Calibration Lab: This lab has a 8 Ci, Co-60 standard housed in a CRC-2 camera,
Area monitors
etc are calibrated in this lab. This CRC-2 camera is housed in a separate concrete building. All the radiation
, before they are
shipped / dispatched.
CRC-2 camera has Co-60 standard obtained from Bhabha Atomic Research h Centre, Mumbai. It is a
certified source.
8.5
ANNUAL MAINTENANCE CONTRACT (AMC)
Annual maintenance contract (AMC) services:
it is advised that customer enters into an economical Annual maintenance contract with Nucleonix system.
Detailed AMC proposal can be obtained from our customer support division (CSD), by giving required
inputs.
Inputs required by our CSD to send you AMC proposal:
Name, year & data of purchase, Sl. Nos. of equipments, Model No's, No. of equipments for which AMC is
required. 
indicated.
Advantage of entering into AMC:
Equipment services offered will be prompt & timely
which may become obsolete.
Obsolescence in electrons is quite rapid. If you enter into AMC guaranteed service for the period of AMC will
be the responsibility of Nucleonix Systems.
Nucleonix Systems will maintain Engineers at your disposal to attend to AMC calls on time
Without AMC prompt service calls are not guaranteed.
AMC.
Training on maintenance / servicing:
To a limited extent, we offer training on maintenance / repairs at our works to customers on chargeable
services.
16
Ver. Hospital & General (PRM board)
CHAPTER – IX
BLOCK DIAGRAM & DESCRIPTION
(a)
Detectors probe with associated electronics
(b)
DETECTOR PROBE ELECTRONICS
0.01 – 100.00 mR/hr. Detector probe electronics consists of a HV module, one energy compensated detector tail
pulse to RS485 converter electronics, for the detector. Detector probe receives +24V, LV supply and it also
receives L-ON signal from microcontroller, which is active 'LOW' during the time of low range detector being ON &
goes 'HIGH' for high range detector.
Depending upon the countrate these detectors are switched. Also HV fraction is taken from the detector probe into
main board ADC to indicate HV failure condition
MAIN PCB CIRCUITS
Low volt
.
Both these supplies are used internally to power-up various circuits, for their functionality.
A four channel ADC reads fraction of HV & LVs & reports to micro-controller under program control to indicate HV
& LV failures if any to the user.
User interface to GA70 is through keypad command buttons EHT, PROG, INC, DEC, ACK & RESET. These keys
, driven by
multiplexed display driver
selected unit (such as mR/hr or µSv/hr or cps or cpm)
The other interfaces to micro-controller include:
(a)
alarm & normal condition.
(b)
ACTIVE condition.
(c)
interfacing.