Series HTTP 
Portable Transit Time Ultrasonic Flow Meter 
Operations & Maintenance 
Manual 
Rev 01/06 
Pdf thumbnail preview - SDK control API:C# PDF Thumbnail Create SDK: Draw thumbnail images for PDF in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support Thumbnail Generation with Various Options for Quick PDF Navigation
www.rasteredge.com
Pdf thumbnail preview - SDK control API:VB.NET PDF Thumbnail Create SDK: Draw thumbnail images for PDF in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support Thumbnail Generation with Various Options for Quick PDF Navigation
www.rasteredge.com
Rev.  01/06 
-1.1-                  
HTTP 
BEFORE OPERATING THE HTTP 
The HTTP flow meter is equipped with a Lead Acid Gel 
Cell battery.  This battery will require charging before 
initial operation. 
Apply power, utilizing the enclosed 12 Volt DC output line 
power converter or cigarette-style power cord, to the 
HTTP for a period of 16-24 hours prior to using the 
product for the first time
 The power converter connects to 
the DC IN +12V socket connection located on the side of 
the enclosure.  See Figure 1.1 A fully charged battery 
will provide up to 24 hours of continuous operation before 
recharging will be necessary.   
When the battery level has decreased to a point where 
recharging is required, the LOW BATTERY indicator will 
illuminate on the front panel.  At that point, the meter will 
only operate a short time more until it automatically turns 
itself off—preventing excessive battery discharge that can 
damage the Gel Cell battery.  The HTTP has an integral 
charging circuit that prevents overcharging.  The 
instrument can be permanently connected to AC line 
power without damaging the flow meter or the battery.  
Page  1.10  of  this  manual  contains  additional 
recommendations to preserve and maximize the power in 
the HTTP battery. 
If the HTTP is to be used for extended periods of 
operation, the AC power converter or the 12 volt cigarette 
converter can remain connected indefinitely. 
Important Notice! 
Figure 1.1 
SDK control API:How to C#: Preview Document Content Using XDoc.Word
With the SDK, you can preview the document content according to the preview thumbnail by the ways as following. C# DLLs for Word File Preview. Add references:
www.rasteredge.com
SDK control API:How to C#: Preview Document Content Using XDoc.PowerPoint
With the SDK, you can preview the document content according to the preview thumbnail by the ways as following. C# DLLs: Preview PowerPoint Document.
www.rasteredge.com
Rev.  01/06 
-1.2-                  
HTTP 
TABLE OF CONTENTS 
Pages 
Quick-Start Operating Instructions 
1.4-1.5 
Introduction 
General 
1.6 
Applications 
1.6 
Product Matrix 
1.7 
Product Specifications 
1.8 
Transmitter Connections 
1.9-1.10 
Battery Charging and Maintenance 
1.10-1.12 
Input/Output Connections and Options 
4-20 mA Output 
1.13 
optional Data Logger 
1.14 
optional Dual Control Relay 
1.14 
optional Rate Pulse Output 
1.15 
optional RS232C 
1.15 
optional RS485 
1.15 
Transducer Mounting 
Mounting Location 
2.1-2.2 
Transducer Mounting Method 
2.3-2.5 
Transducer Spacing - Keypad Entry 
2.6-2.13 
Transducer Spacing - UltraLink
2.13-2.14 
Pipe Preparation 
2.15 
Transducer Mounting 
2.15-2.21 
Part 1 -  
Introduction 
Connections 
Inputs and 
Outputs  
Part 2 -  
Transducer 
Installation  
SDK control API:How to C#: Set Image Thumbnail in C#.NET
VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel, VB.NET PowerPoint, VB How to C#: Set Image Thumbnail in C#.NET. To Preview Images in WinForm Application.
www.rasteredge.com
SDK control API:How to C#: Preview Document Content Using XDoc.excel
document in memory. With the SDK, you can preview the document content according to the preview thumbnail by the ways as following.
www.rasteredge.com
Rev.  01/06 
-1.3-                  
HTTP 
TABLE OF CONTENTS 
Pages 
General Programming Information 
3.1-3.4 
Keypad Entry Detail 
3.4-3.25 
BASIC MENU 
3.4-3.13 
DATALOG OPERATION 
3.14-3.16 
OUT2 MENU 4-20 mA 
3.17-3.19 
Optional Input/Output 
3.20-3.24 
SECURITY MENU 
3.25 
SERVICE MENU 
3.26-3.27 
Signal Strength 
3.26 
Setting ZERO Flow 
3.27 
DISPLAY MENU 
3.28-3.29 
Software Utility Operation 
UltraLink
4.1-4.9 
Data Logger 
4.14-4.16 
Appendix 
Keypad Interface Map 
Fluid Characteristic Table 
Error Codes 
Digital Communications Protocol 
Pipe Dimension Chart: Cast Iron 
Pipe Dimension Chart: Ductile Iron 
Pipe Dimension Chart: ST, SS, PVC 
Velocity to Volumetric Conversion 
Customer Service 
Statement of Warranty  
Part 3 - 
Operation  
Part 4 -  
Software 
SDK control API:How to C#: Generate Thumbnail for Word
Preview Document. Conversion. Convert Word to PDF. Convert Word Convert Word to ODT. Convert PDF to Word. Text Search. Insert Image. Thumbnail Create. Thumbnail Create
www.rasteredge.com
SDK control API:How to C#: Generate Thumbnail for PowerPoint
Preview Document. Conversion. Convert PowerPoint to PDF. Convert PowerPoint to Pages. Annotate PowerPoint. Text Search. Insert Image. Thumbnail Create.
www.rasteredge.com
Rev.  01/06 
-1.4-                  
HTTP 
This manual contains detailed operating instructions for all 
aspects of the HTTP instrument.  The following 
condensed instructions are provided to assist the operator 
in getting the instrument configured and measuring as 
quickly as possible.  This pertains to basic operation only.  
If specific instrument features are to be used or if the 
installer is unfamiliar with this type of instrument, refer to 
the appropriate section in the manual for complete details.  
1. TRANSDUCER LOCATION 
A. In general, select a mounting location on the piping 
system with a minimum of 10 pipe diameters (10 
times the pipe inside diameter) of straight pipe 
upstream and 5 straight diameters downsteam.  
See Table 2.1 for detailed piping configurations and 
recommended lengths of straight pipe. 
B. Select a mounting method for the transducers 
based on pipe size and liquid characteristics.  See 
Figure 1.2 Select W-Mount for plastic pipes 
flowing clean, non-aerated liquids in the 2-3 inch 
(50-75 mm) internal diameter range.  Select V-
Mount for pipes of all materials and most liquids in 
pipe sizes from 2-10 inches (50-250 mm).  Select 
Z-Mount for pipes 10-100 inches (250-2540 mm) 
inches. 
Transducer 
Location 
QUICK-START OPERATING INSTRUCTIONS 
W-Mount 
V-Mount
Z-Mount
Figure 1.2 
SDK control API:C# Image: View & Operate Web Page Using .NET Doc Image Web Viewer
Support multiple document and image formats, like PDF and TIFF; Thumbnail images order of source document file using mouse dragging in thumbnail preview section;
www.rasteredge.com
SDK control API:How to C#: Generate Thumbnail for Excel
Preview Document. Conversion. Convert Excel to PDF. Convert Excel to Insert Image. Thumbnail Create. Thumbnail Create. |. Home ›› XDoc.Excel ›› C# Excel
www.rasteredge.com
Rev.  01/06 
-1.5-                  
HTTP 
C. Enter the following data into the HTTP transmitter      
via the integral keypad or UltraLink
software 
utility. 
D. Record the value calculated and displayed as 
Transducer Spacing/XDCR SPC. 
2. PIPE  PREPARATION  AND  TRANSDUCER 
MOUNTING 
A. The piping surface, where the transducers are to be 
mounted, needs to be clean and dry.  Remove 
loose scale, rust and paint to ensure satisfactory 
acoustical bonds. 
B. Apply a liberal amount of couplant grease onto the 
transducer faces.   
C. Attach the transducers to the pipe at the location(s) 
determined in Step 1.  Refer to Figure 1.2 for 
proper orientation. 
3. TRANSDUCER/POWER CONNECTIONS 
A. If additional transducer cable is required, utilize 
RG59 coaxial wire with 75 Ohm terminations. 
B. Refer to the WIRING DIAGRAM located on the 
inner door of the HTTP transmitter and Figure 1.3 
for proper power and transducer connections.   
4.   INITIAL SETTINGS AND POWER UP 
A. Press the ON button on the flow meter keypad. 
B. From the Service Menu, verify that signal strength 
is greater than 2.0%. 
C. Input proper units of measure and I/O data. 
Startup 
Connections 
QUICK-START OPERATING INSTRUCTIONS 
1.  Transducer mounting method 
2.  Pipe O.D. (Outside Diameter) 
3.  Pipe wall thickness 
4.  Pipe material 
5.  Pipe sound speed* 
6.  Pipe relative roughness* 
7.  Pipe liner thickness 
(if present)
8.  Pipe liner material 
(if present)
9.  Fluid type 
10. Fluid sound speed* 
11. Fluid viscosity* 
12. Fluid specific gravity* 
*Nominal values for these 
parameters are included 
within the HTTP operating 
system.  The nominal values 
may be used as they appear 
or may be modified if exact 
system values are known. 
Figure 1.3 
SDK control API:How to C#: Overview of Using XDoc.PowerPoint
document (ODP). Empower to navigate PowerPoint document content quickly via thumbnail. Able to you want. Create Thumbnail. See this
www.rasteredge.com
SDK control API:How to C#: Overview of Using XDoc.Word
Tell C# users how to: create a new Word file and load Word from pdf; merge, append, and split Word files; insert, delete, move, rotate, copy Create Thumbnail.
www.rasteredge.com
Rev.  01/06 
-1.6-                  
HTTP 
PART 1 - INTRODUCTION 
The HTTP ultrasonic flow meter is designed to measure 
the fluid velocity of liquid within closed conduit (pipe).  The 
transducers are a non-contacting, clamp-on type, which 
will provide benefits of non-fouling operation and ease of 
installation.  
HTTP transit time flow 
meters  utilize  two 
transducers that function 
as  both  ultrasonic 
transmitters  and 
receivers.    The 
transducers are clamped 
on the outside of a closed 
pipe at a specific distance from each other
. The 
transducers can be mounted in V-mode where the sound 
transverses the pipe two times, W-mode where the sound 
transverses the pipe four times, or in Z-mode where the 
transducers are mounted on opposite sides of the pipe 
and the sound crosses the pipe once. This selection is 
based on pipe and liquid characteristics. The flow meter 
operates by alternately transmitting and receiving a 
frequency modulated burst of sound energy between the 
two transducers and measuring the time interval that it 
takes for sound to travel between the two transducers.  
The difference in the time interval measured is directly 
related to the velocity of the liquid in the pipe. 
The HTTP flow meter can be successfully applied on a 
wide range of metering applications. The simple to 
program transmitter allows the standard product to be 
used on pipe sizes ranging from 2 - 100 inch (50 - 2540 
mm) internal diameters. A variety of liquid applications can 
be accommodated: ultrapure liquids, potable water, 
chemicals, raw sewage, reclaimed water, cooling water, 
river water, plant effluent, etc. Because the transducers 
are non-contacting and have no moving parts, the flow 
meter is not affected by system pressure, fouling or wear.  
Standard transducers are rated to 300 
°
F (150 
°
C). Higher 
temperatures can be accommodated.  Please consult the 
Hedland factory for assistance. 
General 
Application 
Versatility 
Rev.  01/06 
-1.7-                  
HTTP 
PART 1 - INTRODUCTION 
The HTTP employs modular construction and provides 
electrical safety for the operator.  The display face 
contains voltages no greater than 10 Vdc.  All user 
connections are made through sealed bulk-head plugs 
located on the side of the HTTP enclosure. 
Non-volatile flash memory retains all user-entered 
configuration values in memory for several years, even if 
power is lost or the unit is turned off.  Data Logger values 
are stored in flash memory in the logger.  Password 
protection is provided as part of the Security menu and 
prevents inadvertent configuration changes or totalizer 
resets. 
The serial number and complete model number of your 
HTTP is located on the  inside of the transmitter’s front 
cover.  Should technical assistance be required, please 
provide the Hedland Customer Service Department with 
this information.  
User Safety 
Data Integrity 
Product 
Identification 
Product Matrix 
Power Supply 
A - 115 VAC 
B - 230 VAC  
(two round terminals) 
C - 100 VAC  
G - 230 VAC  
(three rectangular  
terminals) 
Output 2 
1 - 4-20 mA  
Output 1 
N - None (Std) 
6 - 200,000 event data logger ¹ 
Approvals 
N - Ordinary Area 
F - Class I Div 1, Groups C & D  
Options 
N - HTTN Transducer Package 
H - HTTH High Temperature Transducer Package 
2 - HTTS Small Pipe Transducer Package  
-pipe sizes less than 2" (refer to Form No. 312 for part number matrix for HTTS transducers) 
¹ The data logger records up to 30,000 points per file, with a maximum of 16 files. The total number 
of points that can be recorded on the logger is 200,000. 
HTTP - 
… … … 
A - 
… …
Replacement Parts
Part Number
HTTP Flow meter  
HTTP-001 
Data Logger, 200,000-event 
HTTP-004 
Padded carrying case 
HTTP-008 
Standard Transducer Set 
HTTP-012 
High Temp Transducer Set 
HTTP-016 
Transducer Cable set, 20 ft. (6m) 
HTTP-COAX-20 
Acoustic Grease, temporary mount 
HTTF-CPLNT2 
Mounting Track, w/measuring scale  HTTF-MTRK 
36 inch SS hose clamp 
HTTF-HCLP 
Power converter, 115 V U.S. 
HTTP-044 
Power converter, 230 V European 
HTTP-048 
Power converter, 230 V U.K. 
HTTP-052 
Power cord, 12 V  cigarette-style 
HTTP-060 
4-20 mA interconnect cable 
HTTP-064 
Infrared serial adapter 
HTTP-072 
USB to DB-9 adapter 
HTTP-076 
ULTRALink software CD 
HTTP-086 
Manual, HTTP flow meter 
HTTP-MANUAL 
Rev.  01/06 
-1.8-                  
HTTP 
PART 1 - SPECIFICATIONS 
Rev.  01/06 
-1.9-                  
HTTP 
PART 1 - HTTP TRANSMITTER CONNECTIONS 
After unpacking, it is recommended to save the shipping 
carton and packing materials in case the instrument is 
stored or re-shipped.  Inspect the equipment and carton 
for damage.  If there is evidence of shipping damage, 
notify the carrier immediately. 
When the HTTP is to be utilized for extended periods of 
time in one location, the enclosure should be placed in 
an area that is convenient for servicing, calibration or for 
observation of the LCD readout. 
1. Locate the transmitter within the length of transducer 
cable that was supplied with the HTTP system.  If this 
is not possible, additional cable should be RG59 co-
axial cable and terminations should be 75 Ohm. 
Longer cables are also available by contacting the 
Hedland factory. 
2. Place the HTTP transmitter in a location that is:  
♦ 
Where little vibration exists. 
♦ 
Protected from falling corrosive fluids. 
♦ 
Within ambient temperature limits -40 to 185 °F (-40 
to 85 °C). 
♦ 
Out of direct sunlight.  Direct sunlight may increase 
transmitter temperatures above maximum limit. 
3. If the transmitter will be subjected to a wet 
environment, it is recommended that the cover 
remain closed and the latches secured after 
configuration is completed.  The faceplate/keypad of 
the HTTP is watertight, but avoid letting water collect 
on the keypad area. 
Transmitter 
Location 
Considerations 
It is highly recommended that the internal battery in the 
HTTP be fully charged before using the meter for the first 
time.  Details covering this procedure are located on 
Page 1.1 of this manual. 