Rev.  01/06 
-4.3- 
HTTP 
Click on the button labeled Configuration for updating flow range, 
liquid, pipe and I/O operating information.  The first screen that 
appears after clicking the Configuration button is the BASIC tab.  
See Figure 4.2
BASIC TAB - see Figure 4.2 
• 
General Units allows selection of either English (U.S.) or Metric 
units of measure. If measurements of the pipe are to be entered 
in inches, select English. If measurements are to be entered in 
millimeters, select Metric. If the General Units are altered from 
those at instrument startup, then click on the Download button 
on the lower right-hand portion of the screen and re-cycle power 
to the HTTP. 
• 
Transducer Type selects the transducer that will be connected 
to  the  HTT  flow  meter.  Select from  HTTN,  HTTH,  or  HTTS  
models.  This selection will influence transducer spacing and 
flow meter performance. If you are unsure about the type of 
transducer to which the HTT will be connected, consult the 
shipment packing list or call the Hedland factory for assistance.  A 
change  of Transducer  Type will  cause a System  Configuration    
Error  (Error 1002)  to  occur. This error will  clear when the  
microprocessor is reset or power is cycled on the flow meter. 
Figure 4.2 
Basic Tab 
PART 4 - SOFTWARE UTILITIES 
Configuration 
Show pdf thumbnail in - software control project:C# PDF Thumbnail Create SDK: Draw thumbnail images for PDF in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support Thumbnail Generation with Various Options for Quick PDF Navigation
www.rasteredge.com
Show pdf thumbnail in - software control project:VB.NET PDF Thumbnail Create SDK: Draw thumbnail images for PDF in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support Thumbnail Generation with Various Options for Quick PDF Navigation
www.rasteredge.com
Rev.  01/06 
-4.4- 
HTTP 
• 
Transducer  Mount selects the orientation of the transducers 
on the piping system. See Part 2 of this manual and Table 2.2 
for detailed information regarding mounting modes for particular 
pipe and liquid characteristics. Whenever Transducer Mount is 
changed, a download command and subsequent microproc-
essor reset or flow meter power cycle must be conducted. 
• 
Transducer Spacing is a value calculated by the HTT flow 
meter  that  takes  into  account  pipe,  liquid,  transducer  and 
mounting information. This spacing will adapt as these parameters 
are modified. The spacing is given in inches for English units 
selection and mm for Metric. This value is the lineal distance 
that must be between the center lines on the transducer heads. 
• 
Pipe Material is selected from the pull-down list. If the pipe ma-
terial utilized is not located on the list, select Other and enter 
pipe material sound speed (much of this information is available 
via web sites such as www.ultrasonic.com) and relative rough-
ness (rms internal surface regularities/the pipe internal diame-
ter) of the pipe. 
• 
Pipe O.D. and  Wall  Thickness are based on the physical 
dimensions  of  the  pipe  on  which  the  transducers  will  be 
mounted.  Enter this value in inches for English units or millimeters 
for Metric units. 
• 
Liner  Material is selected from the pull-down list.  If the pipe 
liner material utilized is not located on the list, select Other and 
enter liner material sound speed (much of this information is 
available at web sites such as www.ultrasonic.com) and relative 
roughness  (rms  internal  surface  regularities/the  pipe  internal 
diameter) of the pipe liner. 
• 
Fluid Type is selected from a pull-down list. If the liquid is not 
located  on  the  list,  select  Other  and  enter  the  liquid  sound 
speed and viscosity into the appropriate boxes. Liquid Specific 
Gravity is required if mass measurements are to be made. 
PART 4 - SOFTWARE UTILITIES 
software control project:How to C#: Set Image Thumbnail in C#.NET
VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel, VB.NET PowerPoint, VB.NET Tiff, VB.NET Imaging, VB.NET OCR, VB Thumbnail item. Make the ToolBox view show.
www.rasteredge.com
software control project:C# Word - Header & Footer Processing in C#.NET
using RasterEdge.XDoc.Word; Create Footer & Header. The following C# sample code will show you how to create a header and footer in section.
www.rasteredge.com
Rev.  01/06 
-4.5- 
HTTP 
FLOW Tab - see Figure 4.3 
• 
Flow Rate Units are selected from the pull down lists. Select 
an appropriate rate unit and time from the two lists. 
• 
Totalizer  Units are selected from pull down lists. Select an  
appropriate totalizer unit and totalizer exponent. The totalizer  
exponents are in Scientific Notation and permit the eight digit 
totalizer to accumulate large values before the totalizer “rolls 
over” and starts again at zero. Table 4.1 illustrates the Scientific 
Notation values and their respective decimal equivalents. 
Figure 4.3 
Flow Tab 
Flow Units 
Configuration 
PART 4 - SOFTWARE UTILITIES 
Exponent 
Display Multiplier 
E-1 
x 0.1 (divide by 10) 
E0 
x 1 (no multiplier) 
E1 
x 10 
E2 
x 100 
E3 
x 1,000 
E4 
x 10,000 
E5 
x 100,000 
E6 
x 1,000,000 
TABLE 4.1 - Totalizer Exponent Values 
software control project:VB.NET Image: How to Process & Edit Image Using VB.NET Image
How to Process Image Using VB.NET. In this section, we will show you all VB.NET Image Cropping Assembly to Crop Image, VB.NET Image Thumbnail Creator Control SDK
www.rasteredge.com
software control project:How to C#: File Format Support
PowerPoint Pages. Annotate PowerPoint. Text Search. Insert Image. Thumbnail Create. VB.NET How-to, VB.NET PDF, VB.NET Word, VB Microsoft Office 2003 PowerPoint Show
www.rasteredge.com
Rev.  01/06 
-4.6- 
HTTP 
• 
MIN  Flow is used by the HTT to establish filter settings in its 
operating system.  Enter a flow rate that is the minimum flow 
rate anticipated within the system.  For uni-directional systems, 
this value is typically zero.  For bi-directional systems this value 
is  set  to  a  negative  number  that  is  equal  to  the  maximum 
negative flow rate that is anticipated within the system.
• 
MAX Flow is used by the HTT to establish filter settings in its 
operating  system.    Enter  a  flow  rate  that  is  the  maximum, 
positive flow rate anticipated within the system. 
• 
The Damping value is increased to increase stability of the flow 
rate readings. Damping values are decreased to allow the flow 
meter to react faster to changing flow rates. 
• 
Low Flow Cutoff is entered as a percentage between MAX and 
MIN Flow. It influences how the meter will act at flows close to 
zero. Typically, an entry of 1% provides a stable zero indication, 
while providing a 100:1 turndown ratio for measurements. 
• 
Low  Signal Cutoff is a relative value that should be entered 
after a successful startup. For an initial value, enter 5% (Signal 
Strength indications below 3% are considered to be below the 
noise ceiling and should not be indicative of a successful flow 
meter startup.) The entry has three purposes. It provides an 
error indication—Low Signal Strength (Error 0010 on display) 
when liquid conditions within the pipe have changed to the point 
where flow measurements may not be possible. It warns if the 
pipe’s liquid level has fallen below the level of the transducers. It 
can signal  that  something with the flow meter  installation  or 
configuration may have  changed. For  example,  the  couplant 
used  to  mount  the  transducer  has  become  compromised,  a 
cable has been disconnected or a pipe size has been altered. 
• 
Substitute Flow is used to provide an indication and output that 
signifies that an error exists with the flow meter or its setup. It is 
set as a percentage between MIN Flow and MAX Flow. In a uni-
directional system, this value is typically set to zero to indicate 
zero flow while in an error condition. In a bi-directional system, 
the percentage can be set such that zero is displayed in a error 
condition. To calculate where to set the Substitute Flow value in 
a bi-directional system, perform the following operation: 
Substitute Flow =          100 x MAX Flow 
____  
MAX Flow + MIN Flow 
PART 4 - SOFTWARE UTILITIES 
software control project:XDoc.HTML5 Viewer for .NET, All Mature Features Introductions
By clicking a thumbnail, you are redirect to a to search text-based documents, like PDF, Microsoft Office Word In addition, you may customize to show or hide
www.rasteredge.com
software control project:VB.NET Image: VB.NET DLL for Image Basic Transforming in .NET
VB.NET demo code below will show you how to crop a local image by We are dedicated to provide powerful & profession imaging controls, PDF document, tiff files
www.rasteredge.com
Rev.  01/06 
-4.7- 
HTTP 
• 
Entry of data in the Basic and Flow tabs is all that is required to 
provide flow measurement functions to the flow meter.  If the 
user is not going to utilize input/output functions, click on the 
Download button to transfer the configuration to the HTT 
instrument. 
The Advanced TAB (see Figure 4.4) contains several filter settings 
for the HTT flow meter.  These filters can be adjusted to match 
response times and data “smoothing” performance to a particular 
application.  The factory settings are suitable for most installations. 
• 
Time Domain Filter adjusts the number of raw data sets (the 
wave forms viewed on the UltraLink
Diagnostics Screen) that 
are  averaged  together.    Increasing  this  value  will  provide 
greater damping of the data and slow the response time of the 
flow meter.  This filter is not adaptive—it is operational to the 
value set at all times. 
• 
Low  Signal  Cutoff is a duplicate entry from Page 4.6.  
Adjusting this value adjusts the value on the Flow TAB. 
• 
Substitute Flow is a duplicate entry from Page 4.6.  Adjusting 
this value adjusts the value on the Flow TAB. 
Figure 4.4 
Advanced Tab 
Downloading 
the 
Configuration 
Meter Filter 
Configuration 
PART 4 - SOFTWARE UTILITIES 
software control project:VB.NET Image: Sharpen Images with DocImage SDK for .NET
This guiding page will show you how to sharpen an image in a Visual Basic .NET image processing application. Besides, we would like
www.rasteredge.com
software control project:VB.NET Image: VB.NET Code to Create Watermark on Images in .NET
Add Watermark to Image. In the code tab below we will show you the We are dedicated to provide powerful & profession imaging controls, PDF document, tiff files
www.rasteredge.com
Rev.  01/06 
-4.8- 
HTTP 
• 
Short Pulse Duration is a function used on pipes larger than 2 
inches (50 mm). If the pipe has an outer diameter of 8 inches or 
more, make sure that the Auto Short Pulse box is checked. Set 
this value to zero to disable the function. 
• 
Flow  Filter  Damping establishes a maximum
adaptive filter 
value. Under stable flow conditions (flow that varies less than 
the Flow Filter Hysteresis  entry)  this  adaptive  filter  will 
increase  the  number  of  successive  flow  readings  that  are 
averaged together up to this maximum value. If flow changes 
outside of the Flow Filter Hysteresis window, the Flow Filter 
adapts  by  decreasing  and  allows  the  meter  to  react  faster.  
Increasing this  value tends  to provide smoother steady-state 
flow readings and outputs. 
• 
Flow Filter Hysteresis creates a window around the average 
flow measurement reading whereby if the flow varies within that 
window, greater Flow Filter Damping will occur. The filter also 
establishes a flow rate window where measurements outside of 
the window are captured by the Bad Data Rejection Filter.  
The value is entered as a percentage of actual flow rate. 
Example: 
If the average flow rate is 100 GPM and the  Flow  Filter 
Hysteresis is set to 5%, a filter window of 95-105 GPM is 
established.    Successive  flow  measurements  that  are 
measured within that window are recorded and averaged in 
accordance with the Flow Filter Damping setting.  Flow 
readings outside of the window are held up in accordance 
with the Bad Data Rejection Filter. 
• 
Flow Filter MinHysteresis sets a minimum hysteresis window 
that is invoked at sub 0.25 FPS (0.08 MPS) flow rates, where 
the  “of  rate” Flow Filter Hysteresis  is  very  small  and 
ineffective.    This  entry  is  entered  in  pico-seconds  and  is 
differential time.  If very small fluid velocities are to be meas-
ured, increasing the Flow Filter MinHysteresis value can 
increase reading stability. 
• 
Flow  Filter  Sensitivity allows configuration of how fast the 
Flow  Filter  Damping will adapt in the positive
direction.  
Increasing this  value  allows  greater  damping  to  occur faster 
than lower values.  Adaptation in the negative direction is not 
user adjustable. 
PART 4 - SOFTWARE UTILITIES 
software control project:C# TIFF: C#.NET Code to Create Windows TIFF Viewer | Online
400; option.DocViewerHeight = 400; // Show the thumbnail ThumbDock.Left; // Set the thumbnail height option & profession imaging controls, PDF document, image to
www.rasteredge.com
software control project:C# Word: How to Create C# Word Windows Viewer with .NET DLLs
400; option.DocViewerHeight = 400; // Show the thumbnail ThumbDock.Left; // Set the thumbnail height option & profession imaging controls, PDF document, tiff
www.rasteredge.com
Rev.  01/06 
-4.9- 
HTTP 
• 
Bad  Data  Rejection is a value related to the number of 
successive readings that must be measured outside of the Flow 
Filter  Hysteresis and Flow  Filter  MinHysteresis windows 
before the flow meter will use that flow value.  Larger values are 
entered into the Bad Data Rejection when measuring liquids 
that contain gas bubbles, as the gas bubbles tend to disturb the 
ultrasonic signals and cause more extraneous flow readings to 
occur.  Larger Bad Data Rejection values tend to make the flow 
meter more sluggish to rapid changes in actual flow rate. 
Entry of data in the Basic and Flow screens are all that is required 
to provide flow measurement functions to the flow meter. If the user 
is not going to utilize input/output functions or data logging, click on 
the Download  button  to  transfer the  configuration to  the  HTTP   
instrument. 
The entries made in the Output TAB (See Figure 4.5) establish 
input and output calibration and ranges for ISO-MOD modules in-
stalled in the HTTP flow meter. If a module was ordered from and 
installed at the Hedland factory, then the Output TAB will contain 
information and configuration for that module. Select the appropri-
ate  module  from  the  pull-down  menu  and  press  the  Download    
button. If a module has been changed from the factory setting, a 
Configuration  error  will  result.    This  error  will  be  cleared  by      
resetting the HTTP microprocessor from the Communications/
Output 
Configuration 
PART 4 - SOFTWARE UTILITIES 
Figure 4.5 
4-20mA Configuration 
Rev.  01/06 
-4.10- 
HTTP 
Commands/Reset Target button or by cycling power on the HTT 
flow meter. Once the proper output modules are selected and the 
microprocessor is reset, calibration and configuration of the modules 
can be completed. If a module slot is empty in the HTT enclosure, 
select NONE as the module type. 
To configure the 4-20 mA output or data logger, click on the Out-
put tab (See Figure 4.5). The output menu allows selection, con-
figuration, calibration and testing of various input/output modules.  
Standard HTTP flow meters contain a single 4-20 mA output mod-
ule located in Module #2 position and a data logger located in the 
Module #1 position. The window will appear as shown in Figure 
4.5. (The 4-20 mA module is mounted internally in the flow meter 
and requires meter disassembly in order to replace the module.  
The data logger is located under the sealed front plate on the meter 
face. The logger is designed for repeated installation and removal.)  
Detailed  information  regarding  all  of  the  modules  available  and 
configuration options are available in section 3 of this manual. To 
disable the data logger, select None for Module #1, and select any 
other module for Module #2. 
4-20 mA Module Configuration 
If the 4-20 mA output has been installed, the screen shown in Figure 
4.5 will appear in UltraLink
at the OUTPUT tab: 
• 
Flow @4mA and Flow @20mA set the span of the 4-20 mA 
output. The entry is made in the same flow measurement units 
that were entered in the Flow Tab. The output can be set to 
span across zero (4 mA can be set to a negative flow value) so 
that the module will output bi-directional flow. For example, if a 
flow range spans from –100 to +100, the HTT will output 4 mA 
at –100 and 20 mA at +100 and output 12 mA (50% of the 
output) at 0. 
• 
Calibration/Test is used to adjust the factory calibration span 
of the 4-20 mA output and to test the output. The 4-20 mA out-
put is factory calibrated and should not require adjustment in the 
field. If the module is replaced or if recalibration is required, the 
following procedure is used to calibrate the span of the module: 
1.  Connect a milliamp meter serially within the 4-20 mA module 
output. 
2.  Check the Calibration/Test box. 
PART 4 - SOFTWARE UTILITIES 
Rev.  01/06 
-4.11- 
HTTP 
3.  Select the 4 mA Calibration box. 
4.  Adjust the count value to the right of the 4 mA button until 
the milliamp meter registers 4.00 mA. 
5.  Select the 20 mA Calibration box. 
6.  Adjust the count value to the right of the 20 mA button until 
the milliamp meter registers 20.00 mA. 
7.  Press the Test button. 
8.  Adjust the count value to 12. 
9.  Verify that the milliamp meter registers 12.00 mA. 
10. Uncheck the Calibration/Test box. 
Setting Zero and Calibration 
UltraLink
contains a powerful multi-point calibration routine that 
can  be  used  to  calibrate  the  HTT  flow  meter  to  a  primary 
measuring  standard  in  a  particular  installation.  To  initialize  the 
three-step calibration routine, click on the Calibration button located 
on the top of the UltraLink
Data Screen. The display shown in 
Figure 4.6 will appear. The first step (Page 1 of 3) in the calibration 
process is the selection of the engineering units with which the 
calibration will be performed. Select the units and click the Next 
button at the bottom of the window. 
PART 4 - SOFTWARE UTILITIES 
Figure 4.6 
Calibration Units 
Flow Meter 
Calibration 
Rev.  01/06 
-4.12- 
HTTP 
The second screen (Page 2 of 3) Figure 4.7, establishes a baseline 
zero flow rate measurement for the instrument. To zero the flow 
meter, establish zero flow in the pipe (turn off all pumps and close a 
dead-heading  valve).  Wait  until  the  delta-time  interval  shown  in 
Figure 4.7 is stable (and typically very close to zero). Click the Set 
button. Click the Next button when prompted, then click the Finish 
button on the Calibration Screen. 
NOTE: If the Set button was clicked, do not proceed with Flow 
Rate Calibration before clicking the Finish button to save the Zero 
setting. 
The screen shown in Figure 4.8 (Page 3 of 3) allows multiple actual 
flow rates to be recorded by the HTT. To calibrate a point, establish 
 stable,  known  flow  rate  (verified  by  a  real-time  primary  flow 
instrument), enter the actual flow rate in the Figure 4.8 window and 
click the Set button. Note: If only two points are to be used (zero 
and span), it is preferable to use the highest flow rate anticipated in 
normal operation as the calibration point. If an erroneous data point 
is collected, the point can be removed by pressing the Edit button, 
selecting the bad point and then selecting Remove. 
PART 4 - SOFTWARE UTILITIES 
Figure 4.7 
Setting Zero Flow 
Wait for Stable Reading 
Important! 