Remote-Control Commands
NFC A/B/F
81
Operating Manual 1175.7268.02 ─ 05
Manual operation: See "Number Of Payload Bytes" on page 45
[:SOURce<hw>]:BB:NFC:CBLock<ch>:APGeneric:STEPresent
<StdFrameEodPres>
Selects if the EoD is present or not.
Parameters:
<StdFrameEodPres> 0 | 1 | OFF | ON
Manual operation: See "EoD (CRC)" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:ATIMeout <ATimeout>
Only used with PDU type "supervisory". Determines whether an "ATN" (Attention) or
"Timeout" supervisory PDU type is used.
Parameters:
<ATimeout>
ATN | TOUT
Manual operation: See "ATN or Timeout" on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BCCError <BCCError>
If enabled, an error is added intentionally to the BCC (checksum).
Parameters:
<BCCError>
0 | 1 | OFF | ON
Manual operation: See "BCC Error" on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BFANticollision <BFANticol>
Determines the bit frame Anticollision.
Parameters:
<BFANticol>
SDD0 | SDD2 | SDD1 | SDD4 | SDD8 | SDD16
Manual operation: See "Bit Frame SDD / Bit Frame Anticollision" on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BFSDd <BFSdd>
Determines Bit frame SDD.
Parameters:
<BFSdd>
SDD0 | SDD2 | SDD1 | SDD4 | SDD8 | SDD16
Manual operation: See "Bit Frame SDD / Bit Frame Anticollision" on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLOCk<st>:BDATa <BData>
Sets the value of "Block Data" .
Frame Configuration
Add and delete pages from pdf - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add page break to pdf; add a page to a pdf in reader
Add and delete pages from pdf - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add page to pdf preview; adding a page to a pdf in reader
Remote-Control Commands
NFC A/B/F
82
Operating Manual 1175.7268.02 ─ 05
Parameters:
<BData>
integer
Manual operation: See "Block List, Block Data, Block List Configuration"
on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLKSelection <BlockSel>
Selects a block to be read/written.
Parameters:
<BlockSel>
float
*RST: 
1
Manual operation: See "Block or Byte Selection (ADD)" on page 37
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLOCk<st>:BNUMber <BNumber>
Sets the block number in the block list.
Parameters:
<BNumber>
float
Manual operation: See "Block List, Block Data, Block List Configuration"
on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLOCk<st>:LEN <BLLength>
Sets the block length.
Parameters:
<BLLength>
LEN2 | LEN3
Manual operation: See "Block List, Block Data, Block List Configuration"
on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLOCk<st>:LOCKed <LControl>
Enables/disables status information on lock for the corresponding block ("BLOCK-1" to
"BLOCK-C").
Parameters:
<LControl>
0 | 1 | OFF | ON
Manual operation: See "Lock Control or Status" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BLOCk<st>:SLORder <SCLOrder>
Sets the service code list order.
Frame Configuration
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
In order to run the sample code, the following steps would be necessary. Add necessary references: How to VB.NET: Delete Consecutive Pages from PDF.
adding page numbers to a pdf in reader; add page numbers to pdf document
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
XDoc.PDF enables you to delete PDF page(s) with customized options, including a single page, a series of pages, and random pages to be Add necessary references
adding page numbers in pdf file; adding pages to a pdf document
Remote-Control Commands
NFC A/B/F
83
Operating Manual 1175.7268.02 ─ 05
Parameters:
<SCLOrder>
float
*RST: 
0
Manual operation: See "Block List, Block Data, Block List Configuration"
on page 36
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BNO <BNo>
Selects the block number to be read/write.
Parameters:
<BNo>
float
*RST: 
1
Example: 
BB:NFC:CBLock<CH>:BNO 78
Selects the block number 78 to be read/written
Manual operation: See "Block Number (BNo)" on page 37
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BTYPe <BType>
Selects the block type to be sent.
Parameters:
<BType>
TPI | TPR | TPS
Example: 
See chapter 6.1, "Programming Example", on page 66
.
Manual operation: See "Block Type" on page 37
[:SOURce<hw>]:BB:NFC:CBLock<ch>:BYTSelection <ByteSel>
Selects a byte to be read/written.
Parameters:
<ByteSel>
float
*RST: 
1
Manual operation: See "Block or Byte Selection (ADD)" on page 37
[:SOURce<hw>]:BB:NFC:CBLock<ch>:CFGType <ConfType>
Determines what platform or protocol the device in listen mode is configured for.
Parameters:
<ConfType>
T2 | T4A | NDEP | DT4A | OFF | 0 | ON | 1
Manual operation: See "Configuration Type " on page 38
Frame Configuration
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# Read: PDF Text Extract; C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image
add a page to a pdf; add page number to pdf hyperlink
C# PDF insert image Library: insert images into PDF in C#.net, ASP
C#.NET PDF SDK - Add Image to PDF Page in C#.NET. How to Insert & Add Image, Picture or Logo on PDF Page Using C#.NET. Add Image to PDF Page Using C#.NET.
add page number to pdf reader; adding a page to a pdf file
Remote-Control Commands
NFC A/B/F
84
Operating Manual 1175.7268.02 ─ 05
[:SOURce<hw>]:BB:NFC:CBLock<ch>:CHAining <Chaining>
Determines if chaining is applied.
Parameters:
<Chaining>
0 | 1 | OFF | ON
Manual operation: See "Chaining" on page 38
[:SOURce<hw>]:BB:NFC:CBLock<ch>:CID <CID>
Determines the value of CID.
Parameters:
<CID>
float
*RST: 
1
Manual operation: See "DID (DID field)/ CID (CID field)" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:CII4 <ConfType>
Determines what platform or protocol the device is configured for.
Parameters:
<ConfType>
T2 | T4A | NDEP | DT4A | OFF | 0 | ON | 1
[:SOURce<hw>]:BB:NFC:CBLock<ch>:CSUPported <CSupported>
Determines if CID is supported.
Parameters:
<CSupported>
0 | 1 | OFF | ON
Manual operation: See "DID Supported / CID Supported" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DATA <Data>
Selects the data source type.
Parameters:
<Data>
ZERO | ONE | PATTern | PN9 | PN11 | PN15 | PN16 | PN20 |
PN21 | PN23 | DLISt
*RST: 
PN9
Manual operation: See "Data List Management..." on page 20
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DATA:DSELection <DSelection>
Selects a Data List.
Parameters:
<DSelection>
string
Frame Configuration
VB.NET PDF Password Library: add, remove, edit PDF file password
passwordSetting.IsAssemble = True ' Add password to PDF file. These two demos will help you to delete password for an encrypted PDF file.
add and remove pages from pdf file online; adding a page to a pdf document
C# PDF Sticky Note Library: add, delete, update PDF note in C#.net
C#.NET PDF SDK - Add Sticky Note to PDF Page in C#.NET. Able to add notes to PDF using C# source code in Visual Studio .NET framework.
add pages to pdf preview; add page number to pdf preview
Remote-Control Commands
NFC A/B/F
85
Operating Manual 1175.7268.02 ─ 05
Manual operation: See "Data List Management..." on page 20
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DATA:LENGth <Length>
Determines the length of the transmitted user data / general data.
Parameters:
<Length>
float
Range: 
 to  65536
*RST: 
0
Example: 
See chapter 6.1, "Programming Example", on page 66
.
Manual operation: See "Data Length" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DATA:PATTern <Pattern>
Defines a bit pattern.
Parameters:
<Pattern>
integer
Example: 
See chapter 6.1, "Programming Example", on page 66
.
Manual operation: See "Data Source (Data)" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DEQD <DivEqDiv>
Determines if the same bit rate divisor for both directions is supported.
Parameters:
<DivEqDiv>
0 | 1 | OFF | ON
Manual operation: See "D(LISTEN->POLL)=D(POLL->LISTEN) / D(PICC-
>PCD)=D(PCD->PICC)" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DFOLlowing <DFollowing>
Determines if a DID is following.
Parameters:
<DFollowing>
0 | 1 | OFF | ON
Manual operation: See "DID following (I-block type, R-block type or S-block type)"
on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DID <DID>
Determines the value of DID (Device Identification Number).
Frame Configuration
VB.NET PDF insert image library: insert images into PDF in vb.net
with this sample VB.NET code to add an image to textMgr.SelectChar(page, cursor) ' Delete a selected As String = Program.RootPath + "\\" output.pdf" doc.Save
adding page numbers to pdf documents; add page to existing pdf file
C# PDF Password Library: add, remove, edit PDF file password in C#
passwordSetting.IsAssemble = true; // Add password to PDF file. These C# demos will help you to delete password for an encrypted PDF file.
add pages to pdf file; add page numbers to a pdf file
Remote-Control Commands
NFC A/B/F
86
Operating Manual 1175.7268.02 ─ 05
Parameters:
<DID>
float
*RST: 
1
Manual operation: See "DID (DID field)/ CID (CID field)" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DLP2 <TaDlp2>
Enables support of divisor 2 for LISTEN to POLL (Bit Rate Capability).
Parameters:
<TaDlp2>
0 | 1 | OFF | ON
Manual operation: See "D(L->P=8), D(L->P=4), D(L->P=2) " on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DLP4 <TaDlp4>
Enables support of divisor 4 for LISTEN to POLL (Bit Rate Capability).
Parameters:
<TaDlp4>
OFF | ON | 1 | 0
Manual operation: See "D(L->P=8), D(L->P=4), D(L->P=2) " on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DLP8 <TaDlp8>
Enables support of divisor 8 for LISTEN to POLL (Bit Rate Capability).
Parameters:
<TaDlp8>
0 | 1 | OFF | ON
Manual operation: See "D(L->P=8), D(L->P=4), D(L->P=2) " on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DLTPoll <Dltp>
In ATTRIB command, sets the divisor in the corresponding transmission direction.
Parameters:
<Dltp>
DIV1 | DIV2 | DIV4 | DIV8
Manual operation: See "Divisor (LISTEN to POLL), Divisor (POLL to LISTEN) /
Divisor (PCD to PICC), Divisor (PICC to PCD)" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPL2 <TaDpl2>
Enables support of divisor 2 for POLL to LISTEN (Bit Rate Capability).
Parameters:
<TaDpl2>
0 | 1 | OFF | ON
Manual operation: See "D(P->L=8), D(P->L=4), D(P->L=2) " on page 40
Frame Configuration
Remote-Control Commands
NFC A/B/F
87
Operating Manual 1175.7268.02 ─ 05
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPL4 <TaDpl4>
Enables support of divisor 4 for POLL to LISTEN (Bit Rate Capability).
Parameters:
<TaDpl4>
0 | 1 | OFF | ON
Manual operation: See "D(P->L=8), D(P->L=4), D(P->L=2) " on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPL8 <TaDpl83>
Enables support of divisor 8 for POLL to LISTEN (Bit Rate Capability).
Parameters:
<TaDpl83>
0 | 1 | OFF | ON
Manual operation: See "D(P->L=8), D(P->L=4), D(P->L=2) " on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPPCd <DPPCd>
In ATTRIB command, sets the divisor in the corresponding transmission direction.
Parameters:
<DPPCd>
DIV1 | DIV2 | DIV4 | DIV8
Manual operation: See "Divisor (LISTEN to POLL), Divisor (POLL to LISTEN) /
Divisor (PCD to PICC), Divisor (PICC to PCD)" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPPicc <Dpp>
In ATTRIB command, sets the divisor in the corresponding transmission direction.
Parameters:
<Dpp>
DIV1 | DIV2 | DIV4 | DIV8
Manual operation: See "Divisor (LISTEN to POLL), Divisor (POLL to LISTEN) /
Divisor (PCD to PICC), Divisor (PICC to PCD)" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DPTListen <Dptl>
In ATTRIB command, sets the divisor in the corresponding transmission direction.
Parameters:
<Dptl>
DIV1 | DIV2 | DIV4 | DIV8
Manual operation: See "Divisor (LISTEN to POLL), Divisor (POLL to LISTEN) /
Divisor (PCD to PICC), Divisor (PICC to PCD)" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DRI <Dri>
Sets DRI.
Frame Configuration
Remote-Control Commands
NFC A/B/F
88
Operating Manual 1175.7268.02 ─ 05
Parameters:
<Dri>
D1 | D2 | D8 | D4 | D16 | D32 | D64
Manual operation: See "DRI" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DSI <Dsi>
Sets DSI.
Parameters:
<Dsi>
D1 | D2 | D8 | D4 | D16 | D32 | D64
Manual operation: See "DSI" on page 40
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DSUPported <DSupported>
Determines if DID is supported.
Parameters:
<DSupported>
0 | 1 | OFF | ON
Manual operation: See "DID Supported / CID Supported" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:DWSelection <DWSelection>
Selects DESELECT or WTX.
Parameters:
<DWSelection>
DSEL | WTX
Manual operation: See "DESELECT or WTX" on page 39
[:SOURce<hw>]:BB:NFC:CBLock<ch>:EASupported <EAtqb>
Determines if Extended ATQB is supported.
Parameters:
<EAtqb>
0 | 1 | OFF | ON
Manual operation: See "Extended SENSB_RES / Extended ATQB" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:ESSupported <ESensbres>
Determines if Extended SENSB_RES is supported.
Parameters:
<ESensbres>
0 | 1 | OFF | ON
Manual operation: See "Extended SENSB_RES / Extended ATQB" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:FSC <Fsc>
Selects the maximum frame size in bytes.
Frame Configuration
Remote-Control Commands
NFC A/B/F
89
Operating Manual 1175.7268.02 ─ 05
Parameters:
<Fsc>
F16 | F24 | F32 | F40 | F48 | F64 | F96 | F128 | F256
Manual operation: See "FSC" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:FWI <Fwi>
Determines the FWI which is needed to calculate Frame Waiting Time (FWT).
Parameters:
<Fwi>
float
Manual operation: See "FWI" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:GBSelection <GBSelection>
Selects 8-byte block to be read/written.
Parameters:
<GBSelection>
float
*RST: 
1
Example: 
BB:NFC:CBLock<CH>:GBSelection 122
Selects GB number 122.
Manual operation: See "Global Block Selection (ADD)" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:GDAVailable <GDAvailable>
Determines if General data is available.
Parameters:
<GDAvailable>
0 | 1 | OFF | ON
Manual operation: See "General Data" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:IBNumber <IBNumber>
Indicates if a Valid I-block or a Valid R(ACK) block is received.
Parameters:
<IBNumber>
0 | 1 | OFF | ON
Manual operation: See "Block Number (for I-block type or R-block type)"
on page 37
[:SOURce<hw>]:BB:NFC:CBLock<ch>:KPARameter <KParameter>
Determines the number of historical bytes (T1 to Tk).
Parameters:
<KParameter>
float
*RST: 
1
Frame Configuration
Remote-Control Commands
NFC A/B/F
90
Operating Manual 1175.7268.02 ─ 05
Manual operation: See "k" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:LREDuction <LReduction>
Selects the length reduction (LR).
Parameters:
<LReduction>
LR64 | LR128 | LR192 | LR254
Manual operation: See "Length Reduction" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:MBLI <Mbli>
Determines the Maximum Buffer Length Index (MBLI).
Parameters:
<Mbli>
float
*RST: 
1
Manual operation: See "MBLI" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:MFSize <MFSize>
Selects the maximum frame size in bytes.
Parameters:
<MFSize>
F16 | F24 | F32 | F40 | F48 | F64 | F96 | F128 | F256
Manual operation: See "FSC" on page 41
[:SOURce<hw>]:BB:NFC:CBLock<ch>:MICHaining <MChaining>
Determines if more information (MI) is chained.
Parameters:
<MChaining>
0 | 1 | OFF | ON
Manual operation: See "MI (more information) Chaining" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:MTR0 <MTr0>
Sets the minimum value of TR0 supported.
Parameters:
<MTr0>
TR00 | TR01 | TR02
Manual operation: See "Minimum TR0, TR1, TR2" on page 42
[:SOURce<hw>]:BB:NFC:CBLock<ch>:MTR1 <MTr1>
Sets the minimum value of TR1 supported.