QUADERA 
141 
Background Library 
The Background Library in the section of the calibration project displays the 
parameters required to determine the background. Changed parameters take 
effect in the next measurement task. 
The Background Library contains the corresponding background data for all 
measured mass numbers in tabular form. You can use various zero gases and 
perform consecutive background determinations. The results can be combined in 
the Background Library (function "Apply to Background Library"). Note that an 
individual Background Library is created for each installed detector. 
Image of background library (example) 
Parameter 
Range 
Description 
• Detector 
Dropdown list 
The Background Library is displayed 
for the selected detector type: 
• Faraday: the table for the 
Faraday detector is displayed. 
• SEM: the table for the SEM 
detector is displayed. 
• IonCounter (only QMG700): 
the table for the ion counter is 
displayed. 
• Mass 
Mass number 
(not editable) 
Mass number for which the 
background has been determined: 
applied from the related MID recipe 
("Mass" parameter). 
• Intensity 
Measured 
value 
Background intensity at the related 
mass number: 
corresponds to the currently 
measured value for this mass 
number. 
• Apply to 
Backgroun
d Library 
Button 
Apply the result of the current 
background determination to the 
Background Library for the detector 
that is selected in the recipe. 
NOTE: 
• New background measurements 
are added to the Background 
Library, 
• Current measuring values 
Adding page to pdf - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
adding page numbers to pdf document; add pages to an existing pdf
Adding page to pdf - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
adding page numbers in pdf; add page numbers to pdf online
QUADERA HTML Help 
142 
replace existing data sets in the 
Background Library. 
• Delete 
Backgroun
d Library 
Button 
Delete all data sets of the 
Background Library, for all 
detectors. 
NOTE: 
"Delete Background Library" 
deletes the data sets of the 
Background Library 
irretrievably. In this case, a new 
background determination is 
required for new data sets. 
VB.NET PDF Library SDK to view, edit, convert, process PDF file
Perform annotation capabilities to mark, draw, and visualize objects on PDF document page. Capable of adding PDF file navigation features to your VB.NET program
add and delete pages in pdf online; add page number to pdf reader
C# PDF insert image Library: insert images into PDF in C#.net, ASP
image adding library control for PDF document, you can easily and quickly add an image, picture or logo to any position of specified PDF document file page.
add page to pdf without acrobat; adding a page to a pdf in preview
QUADERA 
143 
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
page modifying page, you will find detailed guidance on creating, loading, merge and splitting PDF pages and Files, adding a page into PDF document, deleting
add pages to pdf preview; add page numbers pdf files
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
Provides you with examples for adding an (empty) page to a PDF and adding empty pages to a PDF from a supported file format, with customized options.
add page number to pdf document; adding page numbers pdf file
QUADERA HTML Help 
144 
Gas Specific Calibration 
Gas Specific Calibration 
NOTE: 
Perform the following preparatory measurements before the gas specific 
calibration: 
• Calibrate the mass scale, 
• Determine the offset, 
• Determine the background. 
Basics 
An MCD measurement project uses a matrix of calibration factors (analysis 
matrix) to convert ion currents into gas concentrations. The calibration factors 
are not universally valid. They have to be quantified by a separate calibration 
measurement using known gases (calibration gases). Calibration gases are 
selected according to the components in the measurement project whose 
concentrations are to be measured. 
NOTE: 
Gas mixtures that contain components with overlapping masses are not 
suited for use as calibration gas. The same mass number must not be 
assigned to several components. 
Procedure 
One MID measurement project is created per calibration gas. The recipe 
parameters include the mass numbers of the calibration gas components and can 
be modified in the recipe editor. The "GSC Matrix" editor specifies the 
concentration of each calibration gas component and its contribution on the 
related mass number. The gas specific calibration factors are required for the 
actual MCD measurement project. They are determined in the calibration project 
(GSC Result View) and saved in the MCD Calibration Factor Library. 
C# PDF insert text Library: insert text into PDF content in C#.net
C#.NET PDF SDK - Insert Text to PDF Document in C#.NET. Providing C# Demo Code for Adding and Inserting Text to PDF File Page with .NET PDF Library.
adding page numbers to pdf in reader; add and delete pages from pdf
VB.NET PDF insert text library: insert text into PDF content in vb
VB.NET PDF - Insert Text to PDF Document in VB.NET. Providing Demo Code for Adding and Inserting Text to PDF File Page in VB.NET Program.
adding a page to a pdf in reader; add page numbers to pdf in reader
QUADERA 
145 
Determining the GSC Factors 
NOTE: 
Before starting the measurement, check the correct calibration of the 
mass spectrometer components. If calibration is necessary, perform the 
following preparatory measurements: 
• Calibrate the mass scale, 
• Determine the offset. 
• Determine the background. 
NOTE: 
Gas mixtures that contain components with overlapping masses are not 
suited for use as calibration gas. The same mass number must not be 
assigned to several components. 
1. Feed a suitable calibration gas into the vacuum chamber. 
2. Open the start page: In the View menu, click Show Start Page
3. Click on the icon for the required device. 
4. Create a new calibration project: In the "Templates" list, double-click e.g. 
Gas Specific Calibration.qmt. This way you create a new calibration 
project based on the selected template. You can modify it according to 
your needs. 
5. The new measurement project window appears. The working area displays 
the calibration project. 
• Edit the recipe according to your needs (if necessary): Click the 
corresponding tab in the working area selection. 
• Display the measuring data: Click the corresponding tab in the working 
area selection. 
6. Input the parameters which define the calibration gas in the "GSC Matrix" 
editor. 
• Use the commands of the related context menu to edit the 
measurement tasks and masses. 
• Note: Use an identical name for the gas component in an MCD project. 
Otherwise, QUADERA® cannot refer to this component. As an 
example, QUADERA® does not equate "O2" and "Oxygen". 
NOTE: 
You can start the measurement regardless whether the recipe editor or 
the working area has been selected. 
7. Start the measurement: In the device status view, click on the "Start" 
icon. The connection to the device is automatically established. The device 
now receives the parameters for the current measurement task. The 
measurement is running automatically. 
C# PDF Annotate Library: Draw, edit PDF annotation, markups in C#.
text comments on PDF page using C# demo code in Visual Stuodio .NET class. C#.NET: Add Text Box to PDF Document. Provide users with examples for adding text box
adding page numbers to pdf in preview; adding page to pdf
C# PDF Library SDK to view, edit, convert, process PDF file for C#
Capable of adding PDF file navigation features to your C# program. Perform annotation capabilities to mark, draw, and visualize objects on PDF document page.
add pages to pdf online; adding page numbers to pdf files
QUADERA HTML Help 
146 
8. In the GSC Result View, check the measured values and the calculated 
calibration factors. 
9. Click Apply to MCD Calibration Factor Library to save the calibration 
factors in the MCD calibration factor library. Note: 
• New calibration factors are added to the MCD Calibration Factor 
Library, 
• Current values replace existing data sets in the MCD Calibration Factor 
Library. 
10. When required for future applications: save the calibration project as 
template (menu File > Save as Template…). 
11. Close and save the measurement project. In the File menu, click Close
You are requested if you want to save the measurement project. 
QUADERA® suggests an automatically generated file name that is based 
on date/time of the measurement and the used template. 
• Yes: the "Save As" window used for saving measuring data opens up. 
Select a suitable name and path and then click Save. The 
measurement project will be saved and closed. 
Please note: The file extension for measurement projects is *.qmp 
• No (recommended option): the measurement project will be closed 
without saving the data. 
• Cancel: the action is cancelled. You will return to the measurement 
project without saving the data. 
QUADERA 
147 
GSC Matrix Editor 
The parameters that define the calibration gas are displayed in the "GSC Matrix" 
editor of the calibration project. Each calibration gas component is defined by its 
name, concentration, and the measurement task(s). Use the "Add MCD Task" tool 
to add a new component (measurement task). 
Image of "GSC Matrix" editor (example) 
Parameter 
Range 
Description 
• Component 
Letters and 
digits 
Name of the calibration gas 
component. 
Note: 
In an MCD project, use an 
identical name for the gas 
component. Otherwise, 
QUADERA® cannot refer to this 
component. As an example, 
QUADERA® does not equate "O2" 
and "Oxygen". 
• Concentration 
Number > 0 
Concentration of the calibration 
gas component. 
• Mass 
(not 
editable) 
The gas specific calibration factors 
are determined for these 
measurement tasks: 
accepted from the MID recipe 
("Task Name" parameter). 
• Mass # 
Checkbox 
Defines whether a calibration 
factor for the component is 
determined in the corresponding 
measurement task, or not. 
Note: 
• Only one calibration factor can 
be calculated per 
measurement task. 
• Several calibration factors per 
component can be determined 
at the same time. 
• A red colored cell indicates the 
internal standard. Use the 
context menu of the "GSC 
Matrix" editor to set the 
internal standard. 
QUADERA HTML Help 
148 
Context menu of the "GSC Matrix" editor 
In this context menu the following commands are available: 
• Delete selected Component: delete the marked measurement task 
(table row) from the "GSC Matrix" editor. 
• Delete selected Mass: delete the marked mass (table column) from the 
"GSC Matrix" editor. 
• Set as internal standard: set the internal standard for the calibration 
factors. 
Proceed as follows to set the internal standard for the calibration factors: 
1. Open the context menu in the table element related to the 
mass/component pair you want to set as internal standard. 
2. Click Set as internal standard. The table element for the internal 
standard appears with a red background color. 
QUADERA 
149 
QUADERA HTML Help 
150 
GSC Result View 
The parameters required to determine the calibration factors are displayed in the 
"GSC Result View" of the calibration project. 
The "GSC Result View" displays the calibration factors in a tabular form. They are 
calculated for all measurement tasks of the MID recipe, using the corresponding 
measured data. In the course of time you can extend and complete the MCD 
Calibration Factor Library (function "Apply to MCD Calibration Factor Library") by 
using various calibration gases to determine the calibration factors. 
Image of "GSC Result View" (example) 
Parameter 
Range 
Description 
• Component 
(not editable) 
Name of the calibration gas 
component: 
applied from the "GSC Matrix" 
editor ("Component" parameter). 
• Mass 
(not editable) 
Measurement tasks for which the 
gas specific calibration factors are 
determined: 
applied from the related MID recipe 
("Task Name" parameter). 
• Intensity 
(not editable) 
Intensity at the related mass 
number: 
corresponds to the currently 
measured value for this mass 
number. 
• Factor 
(not editable) 
The calibration factor is 
automatically calculated from the 
measured intensity and the 
specified concentration. 
• Apply to 
MCD 
Calibration 
Factor 
Library 
Button 
Apply the current gas specific 
calibration factors to the MCD 
Calibration Factor Library. 
NOTE: 
• New calibration factors are 
added to the MCD Calibration 
Factor Library, 
• Current values replace existing 
data sets in the MCD Calibration 
Factor Library. 