GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
131 
Trace File Formats 
Trace files are ASCII files with selectable file format. The analyzer provides two classes of trace files:  
• Touchstone (*.s<n>p) files  
• ASCII (*.csv) files 
• Matlab (*.dat) files
are ASCII files which can be imported and processed in Matlab.  
The trace file formats complement each other; see Selecting an appropriate file format.   
Touchstone files 
All Touchstone files contain a header, a comment section, and the actual trace data: 
 HZ   S   RI   R     50.0000  
! Rohde & Schwarz ZVL  
! Measurement: S11 
! 2003-07-07 
!  
60297750.000000          0.498113   -0.054290    
80297000.000000          0.504888   -0.081229    
... 
The header consists of the following data elements:  
#
specifies beginning of header line (required at top of file).  <Frequency unit>
HZ / KHZ / 
MHZ / GHZ allowed for imported files. The analyzer always uses HZ for exported data.    
<Data file type>
at present: S for S-parameter files.    
<Data format>
RI for Re/Im, MA for lin. Mag-Phase, DB for dB Mag-Phase. The data 
format for export files can be selected in the Export Data dialog.    
Add pages to pdf online - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add page number to pdf reader; add page to pdf reader
Add pages to pdf online - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add page number to pdf document; add a blank page to a pdf
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
132 
<Normalizing impedance>
Impedance system in which the data was defined. The analyzer 
uses 50.0000 Ω.   
Comment lines start with the exclamation mark (!) and may contain any text used for documentation of the 
trace data file. Any number of comment lines may be inserted before or after the header line.  
The trace data depend on the number of ports <n> and the data format. For real and imaginary values 
(data format = RI) the trace data for each stimulus frequency are arranged as follows: 
• 1-port files (*.s1p) 
• Freq    Re(S11)   Im(S11) 
• S11 can be replaced by an any S-parameter, so the *s1p format is suitable for exporting an 
arbitrary data trace representing an S-parameter.  
• 2-port files (*.s2p) 
• Freq    Re(S11)   Im(S11)    Re(S21)   Im(S21)    Re(S12)   Im(S12)    Re(S22)   Im(S22) 
• (all values arranged in 1 line) 
• 3-port files (*.s3p) 
• Freq    Re(S11)   Im(S11)    Re(S12)   Im(S12)    Re(S13)   Im(S13) 
• Re(S21)   Im(S21)    Re(S22)   Im(S22)    Re(S23)   Im(S23) 
• Re(S31)   Im(S31)    Re(S32)   Im(S32)    Re(S33)   Im(S33) 
• (values arranged in 3 lines) 
• 4-port files (*.s4p) 
• Freq    Re(S11)   Im(S11)    Re(S12)   Im(S12)    Re(S13)   Im(S13)    Re(S14)   Im(S14) 
• Re(S21)   Im(S21)    Re(S22)   Im(S22)    Re(S23)   Im(S23)    Re(S24)   Im(S24) 
• Re(S31)   Im(S31)    Re(S32)   Im(S32)    Re(S33)   Im(S33)    Re(S34)   Im(S34) 
• Re(S41)   Im(S41)    Re(S42)   Im(S42)    Re(S43)   Im(S43)    Re(S44)   Im(S44) 
• (values arranged in 4 lines) 
The stimulus frequencies are arranged in ascending order. If a lin. Mag-Phase (MA) or dB Mag-Phase 
(DB) data format is selected the real and imaginary S-parameter values Re(S
ij
), Im(S
ij
) are replaced by lin 
Mag(S
ij
), phase(S
ij
) or dB Mag(S
ij
), phase(S
ij
), respectively.  
ASCII (*.csv) files 
An ASCII file contains a header and the actual trace data: 
freq;reTrc1_S21;imTrc1_S21;reMem2[Trc1]_S21;imMem2[Trc1]_S21; 
300000.000000;0.000000;0.000000;0.000000;0.000000; 
40499497.487437;0.000000;0.000000;0.000000;0.000000; 
80698994.974874;0.494927;-0.065174;0.500833;-0.074866; 
120898492.462312;0.497959;-0.111724;0.488029;-0.107375; 
... 
The header consists of the following data elements:  
<Stimulus>
stimulus variable: freq for Frequency sweep, power for Power sweep, time for 
Time sweep, trigger for CW Mode sweep.    
<reTrace1>
first response value of first trace: re<Trace_Name>, mag<Trace_Name> or 
db<Trace_Name> for output format Real-Imag, Lin. Mag-Phase or dB Mag-Phase, 
respectively. The data format for export files can be selected in the Export Data dialog.    
<imTrace1>   
second response value of first trace: im<Trace_Name> for output format 
Real-Imag, ang<Trace_Name> for output formats Lin. Mag-Phase or dB Mag-Phase. The 
data format for export files can be selected in the Export Data dialog.  
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
File: Split PDF; C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# Read C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image to
add page numbers to pdf preview; adding page numbers to a pdf document
VB.NET PDF insert image library: insert images into PDF in vb.net
Access to freeware download and online VB.NET to provide users the most individualized PDF page image inserting function, allowing developers to add and insert
add page to pdf; adding page numbers to pdf document
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
133 
<reTrace2>   
first response value of second trace: re<Trace_Name>, mag<Trace_Name> 
or db<Trace_Name> for output format Real-Imag, Lin. Mag-Phase or dB Mag-Phase, 
respectively. The data format for export files can be selected in the Export Data dialog.  
<imTrace2>...   second response value of first trace: im<Trace_Name> for output format 
Real-Imag, ang<Trace_Name> for output formats Lin. Mag-Phase or dB Mag-Phase. The 
data format for export files can be selected in the Export Data dialog. first response value 
of second trace. 
HZ / KHZ / MHZ / GHZ allowed for imported files. The analyzer alway 
uses HZ for exported data. second response value of first trace: im<Trace_Name> for 
output format Re/Im, ang<Trace_Name> for output formats lin. Mag-Phase or dB Mag-
Phase. The data format for export files can be selected in the Export Data dialog.  
The trace data is arranged as described in the header. Different values are separated by semicolons. A 
semicolon is also inserted before the end of each line.  
The stimulus values are arranged in ascending order.  
Shift Response Value... 
Modifies all points of the active trace by means of an added and/or a multiplied complex constant. The 
constant complex values are entered in the four input fields of the Shift Response Value dialog. 
The units of the constants are adjusted to the format of the active trace. Setting all values to zero (Clear 
All Values) restores the original trace.    
Effect of the constants 
The trace points are modified according to the formula displayed in the lower part of the dialog: 
The formula and the different constants are adjusted to the different display formats of a trace:  
• The Magnitude factor shifts a dB Mag trace in vertical direction, leaving the phase of a complex 
parameter unchanged.  
• The Phase factor rotates a trace that is displayed in a polar diagram around the origin, leaving the 
magnitude unchanged.   
• The Real added constant shifts a real trace in vertical direction, leaving the imaginary part 
unchanged.  
• The Imaginary added constant shifts a imaginary trace in vertical direction, leaving the real part 
unchanged.  
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
In order to run the sample code, the following steps would be necessary. Add necessary references: How to VB.NET: Delete Consecutive Pages from PDF.
add and remove pages from pdf file online; adding page numbers in pdf file
VB.NET PDF Password Library: add, remove, edit PDF file password
VB: Add Password to PDF with Permission Settings Applied. This VB.NET example shows how to add PDF file password with access permission setting.
add a page to a pdf in reader; add page to existing pdf file
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
134 
Shifting the trace by means of constant values is a simple case of trace mathematics. Use the 
Define Math dialog to define more complicated mathematical operations.   
Remote control: 
DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y:OFFset <Magnitude>[, 
<Phase>, <Real>, <Imag>]  
Remote control: 
CALCulate<Chn>:MATH:WUNit:STATe ON | OFF 
Shift Stimulus Value...  
Shifts the active trace in horizontal direction, leaving the positions of all markers unchanged. 
The positive or negative offset value for the stimulus variable is entered into an input field. 
The unit depends on the sweep type. 
Shift Stimulus Value can be used in Cartesian as well as in polar diagrams. The visible effect 
depends on the diagram type: 
• In Cartesian diagrams, the trace is shifted relative to the markers and the x-axis.  
• In polar diagrams, the trace is not affected, however, markers change their position.  
Use a negative offset value to reset a shifted trace to its original position.   
Remote control: 
DISPlay:WINDow<Wnd>:TRACe<WndTr>:X:OFFset <numeric_value> 
Max Hold On 
Enables (if selected) or disables the max hold (peak hold) function for the active trace. With enabled max 
hold function, the displayed trace shows the maximum values that the analyzer acquired since the start of 
the measurement.  
The max hold process can be restarted any time using Restart Hold. It is also restarted automatically 
when the channel or trace settings are changed so that the previous measurement results are no longer 
valid.  
Remote control: 
CALCulate<Chn>:PHOLd MAX | OFF  
Restart Hold 
Restarts the max hold (peak hold) function for the active trace, discarding the old maximum measurement 
results. This softkey has no effect unless Max Hold On is active.  
C# PDF insert image Library: insert images into PDF in C#.net, ASP
Insert images into PDF form field. Access to freeware download and online C#.NET class source code. How to insert and add image, picture, digital photo, scanned
adding page numbers to a pdf in reader; add pages to pdf document
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
options, including setting a single page, a series of pages, and random pages to be C#.NET Project DLLs for Deleting PDF Document Page. Add necessary references
adding page numbers to a pdf file; add page pdf reader
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
135 
Remote control: 
CALCulate<Chn>:PHOLd MAX | OFF 
Marker ->     
The Marker -> uses the active marker to define the sweep range, scale the diagram and introduce an 
electrical length offset.   
The functions in the upper section provide search functions to locate specific points on the trace.  
• Marker Search opens a submenu with all marker search functions.   
• Max Search sets the active marker to the absolute maximum in the search range.   
• Next Peak sets the active marker to the next maximum or minimum in the search range, 
depending on the current search criterion.  
The following functions use the stimulus value of the active marker to define the sweep range:   
• Start = Marker sets the beginning of the sweep range equal to the marker value.  
• Stop = Marker sets the end of the sweep range equal to the marker value.   
• Center = Marker sets the center of the sweep range equal to the marker value.   
The following functions use the response value of the active marker to scale the y-axis of the diagram:   
• Ref. Value = Marker sets the reference value equal to the marker value.   
• Max = Marker sets the upper edge of the diagram equal to the marker value.   
• Min = Marker sets the lower edge of the diagram equal to the marker value.   
• Zero Delay at Marker corrects the measurement result by adding or subtracting a constant group 
delay.   
Next Peak 
Sets the active marker to the next maximum or minimum in the search range, depending on the current 
search criterion.  
If a Max Search or a Bandpass Search is active, then the marker is set to the next maximum. The next 
C# HTML5 PDF Viewer SDK to view PDF document online in C#.NET
File: Split PDF; C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# Read C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image to
add and delete pages from pdf; add pages to an existing pdf
VB.NET PDF- View PDF Online with VB.NET HTML5 PDF Viewer
File: Split PDF; C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# Read C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image to
add pages to pdf; add a page to a pdf in acrobat
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
136 
maximum is the maximum with the largest response value that is below the current marker response 
value.   
• If a Min Search or a Bandstop Search is active, then the marker is set to the next minimum. The 
next minimum is the minimum with the smallest response value that is above the current marker 
response value.   
By default the search range coincides with the sweep range. If the active trace contains no markers, a 
marker Mkr 1 is created to indicate the search result. Next Peak is disabled while a Target Search is 
active.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
NPEak 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Start = Marker 
Sets the beginning (start) of the sweep range equal to the stimulus value of the active marker, leaving the 
end (stop) value unchanged. The active marker appears at the left edge of the diagram.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt 
Stop = Marker 
Sets the end (stop) of the sweep range equal to the stimulus value of the active marker, leaving the 
beginning (start) unchanged. The active marker appears at the right edge of the diagram.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP 
Center = Marker 
Sets the center of the sweep range equal to the stimulus value of the active marker, leaving the span 
unchanged. The active marker appears in the center of the diagram.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:CENTer 
Ref Value = Marker 
Sets the reference value equal to the response value of the active marker, leaving the values of the 
vertical divisions (Scale Div.) unchanged.   
Max = Marker 
Sets the upper edge of the diagram equal to the response value of the active marker, leaving the values of 
the vertical divisions (Scale Div.) unchanged.  
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
137 
Min = Marker 
Sets the lower edge of the diagram equal to the response value of the active marker, leaving the values of 
the vertical divisions (Scale Div.) unchanged.  
Zero Delay at Marker 
Corrects the measurement result by adding or subtracting a constant group delay. This function must be 
applied to a trace which is displayed in group delay format. The trace is shifted in vertical direction so that 
the delay at the marker position vanishes.  
The delay represents the propagation time of the wave across the DUT, so this operation corresponds to a 
electrical length compensation, i.e. to a shift of the reference plane by adding to or subtracting from the 
test port a simulated lossless transmission line of variable length. The correction must be carried out on 
the Delay trace but has an impact on all trace formats.  
A standard application of Zero Delay at Marker is correction of the constant delay caused by 
the interconnecting cables between the analyzer test ports and the DUT (line stretch).  
The Zero Delay at Marker function modifies the Offset parameters and therefore influences 
the entire channel.  
Marker Search 
Marker Search     
The Marker Search menu uses markers to locate specific points on the trace.   
Search functions 
The search functions are tools for searching measurement data according to specific criteria. A search 
consists of analyzing the measurement points of the current trace (or of a user-defined subrange termed 
the Search Range) in order to find one of the following:   
• Absolute or relative (local) maxima and minima (minimum/maximum search).  
• Trace points with a specific response value (target search).  
• Trace segments with a shape that is characteristic for bandpass or bandstop filters (bandfilter 
search).  
When the search is activated, the active marker is moved to the (next) point that meets the search criteria. 
If the trace contains no markers, a Mkr 1 is created and used for the search. The search result is 
displayed in the marker info field. If no search result can be found, the marker remains at its original 
position.   
Some search functions can be activated repeatedly in order to find all possible search results. Moreover 
the analyzer provides a Tracking mode where the search is repeated after each sweep.  
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
138 
• Max Search sets the active marker to the absolute maximum in the search range.   
• Min Search sets the active marker to the absolute minimum in the search range.   
• Peak Search > sets the active marker to the next peak with higher stimulus value.   
• < Peak Search sets the active marker to the next peak with lower stimulus value.   
• Target opens a submenu to search for a specific value on the trace.   
• Bandfilter opens a submenu to search for trace segments with characteristic bandfilter shape and 
calculate the filter parameters.  
• Marker Tracking causes the search to be repeated after each sweep.  
• Def Peak defines the criteria for the peak search.   
• Search Range... assigns a search range to each marker of the current trace.   
The search functions are available in all Cartesian and polar diagram types (see Trace – Format). In 
polar diagrams (Polar, Smith, Inverted Smith), where complex values are displayed, the magnitude of the 
response values provides the search criterion.  
Max Search 
Sets the active marker to the absolute maximum in the search range, i.e. to the largest of all response 
values. If a complex trace format (e.g. a polar diagram) is active, the marker is set to the measurement 
point with the maximum magnitude.  
By default the search range coincides with the sweep range. If the active trace contains no markers, a 
marker Mkr 1 is created to indicate the search result.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
MAXimum 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Min Search 
Sets the active marker to the absolute minimum in the search range, i.e. to the smallest of all response 
values. If a complex trace format (e.g. a polar diagram) is active, the marker is set to the measurement 
point with the minimum magnitude. 
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
139 
By default the search range coincides with the sweep range. If the active trace contains no markers, a 
marker Mkr 1 is created to indicate the search result.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
MINimum 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Peak Search >  
Sets the active marker to the next peak with higher stimulus value. If the active trace contains no markers, 
a marker Mkr 1 created to indicate the search result. The peak criteria are defined via Define Peak.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
RPEak 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
< Peak Search 
Sets the active marker to the next peak with lower stimulus value. If the active trace contains no markers, 
a marker Mkr 1 created to indicate the search result. The peak criteria are defined via Define Peak.   
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
LPEak 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Target     
Opens a submenu to search for a specific value on the trace.  
• Target Search activates the search for the target value.  
• Target Search > activates the search for the target value to the right of the active marker.  
• < Target Search activates the search for the target value to the left of the active marker.  
• Def Value specifies the target value.  
• Search Range confines the search to a subrange of the sweep.  
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
140 
Target Search 
Activates the search and sets the active marker to the defined target value. If the active trace contains no 
markers, a marker Mkr 1 is created to indicate the search result.  
If the target value occurs at several stimulus values, the marker is placed to the search result with the 
smallest stimulus value. The other measurement points with the same target value can be located using 
the Target Search > function.  
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active 
marker is not moved away from its original position.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
TARGet 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Target Search > 
Activates the search to the right of the active marker position and sets the active marker to the defined 
target value. The target search range is between the active marker position and the end (Stop) of the 
sweep range. If the active trace contains no markers, a marker Mkr 1 is created to indicate the search 
result, and the target search range starts at the beginning (Start) of the sweep range.  
If the target value occurs at several stimulus values, the marker is placed to the search result with the 
smallest stimulus value. The other measurement points with the same target value can be located using 
Target Search > repeatedly.  
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active 
marker is not moved away from its original position.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
RTARget 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
< Target Search 
Activates the search to the left of the active marker position and sets the active marker to the defined 
target value. The target search range is between the (Start) of the sweep range and the active marker 
position. If the active trace contains no markers, a marker Mkr 1 is created to indicate the search result, 
and the target search range starts at the end (Stop) of the sweep range.  
If the target value occurs at several stimulus values, the marker is placed to the search result with the 
smallest stimulus value. The other measurement points with the same target value can be located using 
Target Search > repeatedly.  
If the target is not found (e.g. because the active trace doesn't contain the target value), then the active 
marker is not moved away from its original position.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute 
LTARget 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Def Value 
Opens an dialog to select a marker for the target search and define the target value.  