GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
151 
The Marker menu contains the following functions:  
• Marker 1, 2, 3 creates the markers numbered 1, 2, and 3.  
• Ref Marker... creates the reference marker which is used to measure relative values and 
distances.  
• Delta Mode activates the display of the active marker values relative to the reference marker.  
• Ref.Marker -> Marker places the reference marker to the position of the active marker.   
• All Mkrs Off removes all markers from all traces of the active setup.   
• Mkr Format defines an output format for the (complex) marker values.  
• More Mkrs opens a submenu to create the markers numbered 4 to 10.  
• Coupled Mkrs couples the markers on different traces.  
• Discrete Mkrs turns the active marker into a discrete marker and vice versa.  
• Mkr Properties... opens a dialog with extended marker settings.  
• Export Mkrs... exports the marker values to an ASCII file. 
Toggle functions in the 
Marker 
menu 
Some of the Marker functions toggle between two alternative states when they are pressed repeatedly: 
• Delta Mode switches the delta mode for the active marker on or off.  
• Coupled Mkrs activates or deactivates marker coupling.  
• Marker 1, ..., Marker 10 and Ref. Marker create a marker or remove it from the display. A removed 
marker remembers its properties (stimulus value, format, delta mode, number) and will be 
restored with these properties when Marker <n> or Ref. Marker is selected again. The marker 
properties are definitely lost if the associated trace is deleted.  
Markers are available for all diagram types (Trace – Format).   
Add page to pdf - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add page numbers to a pdf in preview; adding page numbers to pdf documents
Add page to pdf - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add and delete pages from pdf; add page numbers pdf file
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
152 
Marker 1, 2, 3 
Creates the markers numbered 1, 2, and 3, respectively, and assigns them to the active trace (toggle 
function). Marker 1/2/3 opens the numeric entry bar to define the marker position (Stimulus Mkr 1/2/3). 
The default position is the center of the sweep range.  
On closing the Stimulus Mkr 1/2/3 numeric entry bar a marker symbol (triangle) labeled Mkr <n> is 
positioned on the trace and the marker coordinates are displayed in the Info Field.  
Activating and moving markers 
To select one of several markers as an active marker, do one of the following:   
• Click the marker symbol.  
• Click the marker line in the marker info field. 
To change the position of the active marker on the trace use one of the following methods:  
• Drag-and-drop the marker symbol to the desired position. 
• Click the Marker <n> or Ref. Marker softkey to call up the entry bar for the new stimulus value.  
• Right-click the diagram area or select Mkr. Properties to call up the Marker Properties dialog and 
select the new stimulus value.  
• Use the Search functions to place the marker to a specific point on the trace.  
If the marker position is defined explicitly by entering a numeric value, the marker position 
can be outside the sweep range. If it is just varied using the rollkey, the mouse or the cursor 
keys, it always remains within the sweep range. If the position of a marker outside the sweep 
range is varied, it is automatically moved to the start or stop value of the sweep range, 
whichever is closer.   
Remote control: 
CALCulate<Chn>:MARKer<Mk>[:STATe] 
ON 
CALCulate<Chn>:MARKer<Mk>:Y? 
Ref Marker 
Creates a reference marker and assigns it to the active trace (toggle function). Ref Marker opens the 
numeric entry bar to define the marker position (Stim. Ref.Mkr.). The default position is the center of the 
sweep range.  
On closing the Stimulus Ref Mkr numeric entry bar a marker symbol (triangle) labeled Ref is positioned on 
the trace and a line indicating Ref plus the marker coordinates is inserted in the marker info field.  
The reference marker defines the reference value for all markers that are in Delta Mode.   
Remote control: 
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] 
ON 
CALCulate<Chn>:MARKer<Mk>:REFerence:Y? 
C# PDF insert image Library: insert images into PDF in C#.net, ASP
C#.NET PDF SDK - Add Image to PDF Page in C#.NET. How to Insert & Add Image, Picture or Logo on PDF Page Using C#.NET. Add Image to PDF Page Using C#.NET.
add pages to pdf reader; add and remove pages from pdf file online
VB.NET PDF insert image library: insert images into PDF in vb.net
VB.NET PDF - Add Image to PDF Page in VB.NET. Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
add a page to a pdf in acrobat; add a page to a pdf file
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
153 
Delta Mode  
Converts the active marker to a delta marker so that its values are measured and indicated relative to the 
reference marker (toggle function). A ∆ sign placed in front of the marker line indicates that the marker is 
in Delta Mode.  
The reference marker itself can not be set to delta mode but must be present when another marker is set 
to delta mode. The analyzer takes into account these conditions when Delta Mode is selected:  
• If Delta Mode is selected while the reference marker is active, the marker in the info list after the 
reference marker is activated and set to delta mode. If the current trace contains the reference 
marker only, a new Mkr 1 is created and set to delta mode.  
• If Delta Mode is selected for a normal marker while the current trace contains no reference 
marker, a reference marker is created.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe] ON 
Ref.Marker -> Marker  
Places the reference marker to the position of the active marker. As a consequence, the active marker 
takes the role of a reference marker.   
Ref.Marker -Marker is not active if the active marker is a reference marker.   
Remote control: 
– 
All Mkrs Off 
Removes all markers from all traces of the active setup. The removed markers remember their properties 
(stimulus value, format, delta mode, number) when they are restored. The marker properties are definitely 
lost if the associated trace is deleted.   
Remote control: 
CALCulate<Chn>:MARKer<Mk>:AOFF 
Mkr Format  
Opens a submenu to select an output format for the (complex) active marker value in the 
marker info field. The default marker format is the format of the associated trace. The 
current format is indicated with a !.  
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# Read: PDF Text Extract; C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image
add page numbers to pdf in preview; add page number to pdf reader
VB.NET PDF Password Library: add, remove, edit PDF file password
On this page, we will illustrate how to protect PDF document via password by using simple VB.NET demo code. Open password protected PDF. Add password to PDF.
adding page to pdf in preview; adding page numbers pdf file
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
154 
All marker formats are available irrespective of the measured quantity. The output values are calculated 
by a simple conversion of a complex measurement result, where the marker format defines the conversion 
rules. This flexibility in the calculation must be kept in mind when interpreting the results and physical units 
displayed; see also Measured Quantities and Display Formats.   
Short description of marker formats  
The formats of the markers assigned to a trace are independent of each other and of the trace format 
settings. The following table gives an overview on how a complex marker value z = x + jy is converted.  
Marker Format 
Description 
Formula 
dB Mag 
Magnitude of z in dB 
|z| = sqrt ( x
2
+ y
2
)  
dB Mag(z) = 20 * log|z| dB  
Lin Mag 
Magnitude of z, unconverted 
|z| = sqrt ( x
2
+ y
2
Phase 
Phase of z 
φ (z) = arctan (y/x) 
Real 
Real part of z 
Re(z) = x 
Imag 
Imaginary part of z 
Im(z) = y 
SWR 
(Voltage) Standing Wave Ratio 
SWR = (1 + |z|) / (1 – |z|) 
Delay 
Group delay, neg. derivative of the 
phase response*
)
– d φ (z) / d ω 
dB Mag and Phase 
Magnitude of z in dB and phase in two 
20 * log|z| dB 
C# PDF Password Library: add, remove, edit PDF file password in C#
your PDF document in C# project, XDoc.PDF provides some PDF security settings. On this page, we will talk about how to achieve this via Add necessary references
adding page numbers to a pdf file; add a page to a pdf in reader
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
DLLs for Deleting Page from PDF Document in VB.NET Class. In order to run the sample code, the following steps would be necessary. Add necessary references:
add a page to pdf file; add page numbers to pdf files
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
155 
lines 
arctan ( Im(z) / Re(z) ) 
Lin Mag and Phase 
Magnitude of z (unconverted) and 
phase in two lines 
|z| 
arctan ( Im(z) / Re(z) ) 
Real and Imag 
Real and imaginary part of z in two 
lines 
Default (Trace) 
Marker format identical with trace 
format 
– 
R + j X 
Unnormalized resistance and 
reactance; L or C in three lines 
(Smith diagram) 
L or C**
)
G + j B 
Unnormalized conductance and 
susceptance; L or C in three lines 
(Inverted Smith diagram) 
L or C**
)
*
The delay aperture is defined in the Trace – Format menu.  
**
The equivalent inductances or capacitances
L or C are calculated from the imaginary part of the 
impedance according to 
Remote control: 
CALCulate<Chn>:MARKer<Mk>:FORMat ... 
More Mkrs  
Opens a submenu to create the markers numbered 4 to 10. The markers are analogous to marker no. 1 to 
3.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>[:STATe] ON 
Coupled Mkrs  
Couples the markers of all traces in the active setup to the markers of the active trace (toggle function). 
While marker coupling is active, the active trace markers assume the role of master markers; the other 
markers behave as slave markers, following any change of position of the master marker.  
Effects of marker coupling  
The concept of marker coupling means that corresponding markers on different traces (i.e. markers with 
the same number or reference markers) are positioned to the same stimulus values but keep their 
independent format and type settings. When a trace with markers is selected as the active trace and 
marker coupling is switched on, the following happens: 
• The active trace and all associated markers are left unchanged. The active trace markers become 
the master markers of the setup.  
• Markers on the other traces which have no corresponding master marker are removed but 
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
C#.NET Project DLLs for Deleting PDF Document Page. In order to run the sample code, the following steps would be necessary. Add necessary references:
add a page to a pdf online; add pdf pages together
C# PDF Sticky Note Library: add, delete, update PDF note in C#.net
C#.NET PDF SDK - Add Sticky Note to PDF Page in C#.NET. C#.NET Demo Code: Add Sticky Note on PDF Page in C#.NET. Add necessary references:
add page number pdf file; add document to pdf pages
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
156 
remember their properties and can be re-activated after the coupling is released.  
• The remaining markers on the other traces become slave markers and are moved to the position 
of the corresponding master markers. "Missing" slave markers are created so that each trace has 
the same number of markers placed at the same position.  
• If the position of a master marker is outside the sweep range of the slave trace, the slave marker 
is displayed at the edge of the diagram. The marker info field indicates an invalid measurement 
result: 
While marker coupling is active, it is possible to: 
• Move a master marker and thus change the position of all corresponding slave markers.  
• Activate another trace in order to make the associated markers the new master markers.  
Marker coupling makes sense only if the master and the slave traces use the same stimulus 
variable. Channels with a different stimulus variable (sweep type) are not coupled.  
Remote control: 
CALCulate<Chn>:MARKer<Mk>:COUPled[:STATe] ON 
Discrete Mkrs  
Turns the active marker into a discrete marker and vice versa.  
The stimulus value of a discrete marker always coincides with a sweep point. Use discrete 
markers in order to avoid that the marker indicates an interpolated measurement value. 
Remote control: 
CALCulate<Chn>:MARKer<Mk>:MODE 
Mkr Properties...  
Opens a dialog to define the properties of all markers of the active trace.  
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
157 
In the left part the dialog contains four input fields and drop-down lists to select a marker and define the 
basic properties of a selected marker which is switched on in the Mode panel.   
• Select Mkr. 
opens a drop-down list to select the Ref. Marker or one of the Marker 1, ... 
10 that may be associated to the active trace.  
• Name
defines a marker name which may exceed the length of the input box and 
contain letters, numbers, blanks and special characters.  
• Stimulus
defines the stimulus value (in Cartesian diagrams: the x-axis position) of the 
marker.  
• Format
defines the output format of the marker value.  
The Mode panel contains check boxes to select properties that are related to the marker positions. All 
properties can be combined.  
• Marker On
displays the selected marker in the diagram and in the marker info field or 
removes it. Marker configurations are only available while the marker is switched On. 
A removed marker remembers its properties (stimulus value, format, delta mode, 
number ...) when it is switched on again. The marker properties are definitely lost if 
the associated trace is deleted.    
• Fixed Marker
freezes thecurrent response value of the selected marker. The marker 
can still be shifted horizontally but the vertical position remains fixed if the other 
marker settings are changed. Markers must be inside the sweep range and have a 
valid response value when they are fixed.     
• Delta Mode
sets the selected marker to delta mode and displays its values relative to 
the reference marker.     
• Discrete Mode
means that a marker can be set to discrete sweep points only. If 
discrete mode is switched off, the marker can be positioned on any point of the trace, 
and its response values are obtained by interpolation.     
• All Mkrs Coupled
couples the markers of all traces in the active setup to the markers of 
the active trace; see Coupled Markers.  
Style qualifies how the selected marker is displayed on the screen.  
Show Info selects the marker info to be displayed at the Marker Position, in the marker info field or in a 
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
158 
separate Table below the diagram area. The display options may be selected simultaneously or all 
switched off. The table provides more information than the marker info field:  
Remote control: 
CALCulate<Chn>:MARKer<Mk>[:STATe] 
ON 
CALCulate<Chn>:MARKer<Mk>:TYPE 
NORMal 
FIXed 
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe] 
ON 
CALCulate<Chn>:MARKer<Mk>:MODE CONTinuous | DISCrete 
CALCulate<Chn>:MARKer<Mk>:COUPled[:STATe] ON 
Export Mkrs...  
Calls up a Save As... dialog to store the current marker values to a marker file.   
The analyzer uses a simple ASCII format to export marker values. By default, the marker file extension is 
*.txt. The file contains all traces in the active setup together with their names and measured quantities. 
Below each trace, the file shows a list of all markers with their names, stimulus and response values.   
The following example of a marker file describes a setup with two traces Trc1 and its memory trace. Trc1 
has no markers assigned, the memory trace has four markers named Mkr 1, ..., Mkr 4.   
Remote control: 
MMEMory:STORe:MARKer "file_name" 
Meas     
The Measure submenu selects the quantity to be measured and displayed.  
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
159 
• S11, S12, S21, S22 select the four elements of the standard 2-port scattering matrix (S-
parameters).  
• Impedance opens a submenu to convert reflection S-parameters into matched-circuit impedances 
(converted Z-parameters).  
• Admittance opens a submenu to convert reflection S-parameters into matched-circuit admittances 
(converted Y-parameters).  
• Stability Factor selects one of the three factors K, µ
1
or µ
to assess the stability of linear circuits 
(e.g. amplifiers). 
Port assignments of the DUT and the analyzer 
S-parameters S
<out>< in> 
characterize the Device Under Test, so the indices <out> and < in> denote the 
output and input ports of the DUT. Analogously the waves a
1
and a
2
are referred to as incident waves, b
1
and b
2
to as outgoing waves. 
The analyzer measures the waves a
1
, b
1
, ... at its own test ports in order to obtain S-parameters, ratios 
and other derived quantities. The test ports of the analyzer are numbered, so it is convenient to use them 
as a reference, defining port number n of the DUT as the port connected to test port n of the analyzer. 
With this convention the waves a
n
and b
n
have the following meaning: 
• a
n
is the wave transmitted at test port no. n of the analyzer (stimulus signal) and fed to port 
number n of the DUT (incident wave). 
• b
n
is the wave transmitted at port number n of the DUT (response signal) and received at test port 
no. n of the analyzer (received wave).    
The analyzer offers various diagrams and display formats to visualize the results. The 
analyzer places no restriction on the formats that are available for each measured quantity. 
However, to keep the interpretation simple, it is recommended to select an appropriate 
display format; see Trace - Format.  
S11, S12, S21, S22     
Select one of the four elements of the standard 2-port scattering matrix (S-parameters).  
The S-parameters are the basic measured quantities of a network analyzer. They describe how the DUT 
modifies a signal that is transmitted or reflected in forward or reverse direction. S-parameters are 
expressed as S
<out>< in>
, where <out> and <in> denote the output and input port numbers of the DUT.  
A full 2-port S-parameter measurement involves 2 stages (referred to as partial measurements) with 
R&S
®
ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-01 
160 
interchanged drive and receive ports. The analyzer automatically switches the internal power sources and 
the receivers to obtain the desired S-parameters. 
Remote control: 
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "S11" | 
"S12" 
"S21" 
"S22" 
[SENSe<Chn>:]FUNCtion[:ON] "...:POWer:S<11 | 12 | 21 | 22>"  
Create new trace and select name and measurement
parameter:  
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "S11" | "S12" 
| "S21" | "S22"  
Impedance     
The Impedance submenu contains the functions to convert S-parameters into matched-circuit 
impedances.  The matched-circuit impedances describe the impedances of a DUT that is terminated at its 
outputs with the reference impedance Z
0
Z <– S11, Z <– S12, Z <– S21,  Z<– S22 select the forward and reverse matched-circuit impedances of a 
2-port DUT.  
Z <– S11, Z <– S12, Z <– S21, Z <– S22 
Selects the 2-port matched-circuit, converted impedances. The parameters describe the impedances of a 
2-port DUT, obtained in forward and reverse transmission and reflection measurements:   
• Z
11
is the input impedance of a 2-port DUT that is terminated at its output with the reference 
impedance Z
(matched-circuit impedance measured in a forward reflection measurement).   
• Z
22
is the output impedance of a 2-port DUT that is terminated at its input with the reference 
impedance Z
(matched-circuit impedance measured in a reverse reflection measurement).   
• Z
12
and Z
21
denote the forward and reverse transfer impedances, respectively.  
Use the Smith chart to obtain an alternative, graphical representation of the converted 
impedances in a reflection measurement.  
Remote control: 
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "Z-S11" | "Z-
S12" 
"Z-S21" 
"Z-S22" 