R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
211 
Storing connector type settings 
Calibration kits and connector types are global resources; the parameters are stored independently and 
available irrespective of the current setup. The connector type settings are always stored together with the 
associated calibration kit parameters. It is possible to export and import the connector settings using 
Export Kit... and Import Kit... in the Calibration Kits dialog.  
The name, polarity and reference impedance of a user-defined connector can be changed in the table. 
The 
button in the Offset Model column opens the Offset Model dialog to define the propagation mode 
in detail.  
After assigning a calibration or adapter kit to a user-defined connector type, you can still 
change its name, offset model and reference impedance. Switching between sexed and 
sexless will delete all kits assigned to the connector type.  
Remote control: 
[SENSe<Ch>:]CORRection:CONNection 
[SENSe<Ch>:]CORRection:CONNection:DELete 
Offset Model Dialog  
Defines the mode of wave propagation in the lines of the standards associated with the connector type. 
This dialog is opened from the Available Connector Types dialog (
button in the table).  
The Parameters to be selected depend on the Line Type:  
• If the calibration kit standards contain lines with transverse electric propagation mode (TEM), then 
the Relative Permittivity ε
r
of the dielectric can be defined. The default permittivity is the value for 
air. TEM-type lines have not cutoff frequency.  
• If the calibration kit standards contain waveguides, then the lowest frequency where a wave 
propagation is possible (Cutoff Frequency f
c
) can be defined. The default cutoff frequency if 0 Hz 
(propagation at all frequencies). No relative permittivity is needed for waveguides.  
The impedance for waveguides is frequency-dependent. If a waveguide line type is selected, 
various dialogs (e.g. Add Standard...) will indicate varies instead of a definite impedance 
value.  
Impact of offset model parameters 
The offset model parameters are used for the calculation of the S-parameters for the calibration standards 
Adding a page to a pdf file - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add and delete pages from pdf; add page numbers to a pdf
Adding a page to a pdf file - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add pages to pdf file; add page numbers to pdf preview
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
212 
associated with the connector type, provided that they are derived from a circuit model (Add/Modify 
Standard dialog).  
• For TEM-type lines, the relative permittivity ε
r
is needed for the conversion of and ZVR-type Loss 
(in units of dB/sqrt(GHz)) into an Agilent-type Offset Loss (in units of GΩ/s) and vice versa (see 
Offset andLoad parameters). The Electrical Length and Delay values in the Modify Offset dialog 
are directly entered and therefore independent of ε
r
.  
• For waveguides, the low frequency cutoff frequency f
c
is important because no wave propagation 
is possible at frequencies below f
c
. If a standard is measured in order to acquire calibration data, 
the analyzer checks the low frequency cutoff. If the start frequency of the sweep range is below f
c
then the calibration wizard generates and error message. 
The offset model parameters are not used except in the context of calibration. The offset 
parameterdefinitions (seeMechanical Length) are based on independent ε
r
values.   
Remote control: 
[SENSe<Ch>:]CORRection:CONNection 
Add or View / Modify Standard      
Defines, displays or modifies the properties of the calibration standards in a particular calibration kit. This 
dialog is opened from the Add or View / Modify Calibration Kit dialog (Add Standard... or View / Modify 
Standard... buttons)Depending on the title, some control elements may not be active.  
For an overview of calibration standards and their properties refer to section Calibration standard 
types below. 
In its upper part the Add Standard or View / Modify Standard dialog contains several controls to do the 
following: 
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
guidance on creating, loading, merge and splitting PDF pages and Files, adding a page into PDF document, deleting unnecessary page from PDF file and changing
adding page numbers to pdf in; add page number to pdf hyperlink
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
Provides you with examples for adding an (empty) page to a PDF and adding empty pages to a PDF from a supported file format, with customized options.
adding page numbers pdf file; add page to pdf without acrobat
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
213 
• Select a standard Type and its Gender (for polarized/not sexless connector types and if the port 
assignment is not restricted) and assign a Label.  
• Restrict Port Assignment 
• 
Select S-Params From
Qualify whether the standard is described by a Circuit Model from which the analyzer can calculate the S-
parameters or by a table of measured or simulated S-parameters stored in a Touchstone file. Pressing the 
Read Data from File... button opens a file selection dialog where the appropriate file type (*.s1p for one-
port standards and *.s2p for two-port standards) is automatically selected.  
The Sliding Match and Attenuation are special standard types which must be described by a 
circuit model. The controls in the S-Params From panel are disabled. 
For two-port standards described by a *.s2p file, the implicit ports 1 and 2 (given by the order of S-
parameters Re(S11)   Im(S11)   Re(S21)   Im(S21)   Re(S12)   Im(S12)   Re(S22)   Im(S22) in the file) are 
assigned to the test ports that the analyzer actually calibrates as follows: Port 1 is always assigned to the 
lower-numbered calibrated test port, port 2 to the other (higher-numbered) calibrated test port.  
Assigning a label to standards is optional. However, the label is displayed in many dialogs 
and can provide useful information about the standard, e.g. its serial number.  
If Circuit Model is selected in the S-params From panel, then the controls in the central panel of the dialog 
are enabled. The circuit diagram is adjusted to the selected standard type. The following parameters can 
be set: 
• Frequency range (Min. Freq. to Max. Freq) for which the circuit model is valid. During calibration, 
the analyzer checks whether the sweep range is contained in the validity range of all measured 
standards and possibly generates a warning (see Measure Standards dialog).  
• Offset and Load parameters of the circuit model. 
The impedance for waveguides is frequency-dependent. If a waveguide line type is selected 
in the Offset Model dialog, the circuit model indicates varies instead of a definite impedance 
value.  
Remote 
control: 
[SENSe<Ch>:]CORRection:CKIT:<std_type>  
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> 
MMEMory:LOAD:CKIT:SDATa 
Restrict Port Assignment 
Opens a dialog to define whether the standard can be connected to any port of the analyzer or to just one 
port (for one-port standards) or a pair of ports (for two-port standards).  
VB.NET PDF Library SDK to view, edit, convert, process PDF file
Perform annotation capabilities to mark, draw, and visualize objects on PDF document page. Capable of adding PDF file navigation features to your VB.NET program
add pages to an existing pdf; adding page numbers to a pdf in reader
C# PDF insert image Library: insert images into PDF in C#.net, ASP
image adding library control for PDF document, you can easily and quickly add an image, picture or logo to any position of specified PDF document file page.
add page numbers to pdf document in preview; add page number to pdf reader
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
214 
The port assignment is displayed in the Add or View / Modify Calibration Kit dialogs.  
Port assignment and gender 
The standards are handled differently, depending on their port assignment:  
• If the port assignment is not restricted, the gender belongs to the definition of polarized standards. 
When the connector type and calibration kit is selected in the calibration wizard, the analyzer 
checks whether the kit contains the necessary standard types and whether the standards have 
the right gender.  
• Standards with restricted port assignment are assumed to have the right gender (the one required 
for this port). In the View Modify Standard dialog, the Gender: input field is disabled. In the 
calibration wizard, the analyzer checks whether the kit contains the necessary standard types for 
the required ports. Instead of the gender, the port assignment is stored in the calibration kit file.  
This approach simplifies the definition of standards and helps to avoid inconsistencies.  
Remote control: 
[SENSe<Ch>:]CORRection:CKIT:<std_type>  
Modify Offset 
Specifies the offset parameters for the transmission lines of a particular calibration standard. This dialog is 
opened from the Add or View / Modify Standard... dialog (Modify Offset... button).  
The offset parameters depend on whether or not the circuit model is defined in Agilent Mode (see 
Add/Modify Calibration Kit dialog):  
• If Agilent Mode is active, then the standard is characterized by its Delay (in s), its characteristic 
impedance Z
0
(in Ω) and its Offset Loss (in GΩ/s).  
• If Agilent Mode is switched off, then the standard is characterized by the ZVR-compatible 
parameters Electrical Length (in m), its Char. Impedance (in Ω) and its Loss (in dB/sqrt(GHz)). 
The loss is zero and not editable as long as the electrical length is zero.  
Both parameter sets are closely related. The Electrical Length is proportional to the DelayZ
0
corresponds 
to the Char. Impedance. Moreover the analyzer converts an Agilent-type Offset Loss into a ZVR-type Loss 
and vice versa using the Relative Permittivity 
ε
r
for the connector type defined in the Offset Model... dialog. 
Offset parameters 
C# PDF insert text Library: insert text into PDF content in C#.net
C#.NET PDF SDK - Insert Text to PDF Document in C#.NET. Providing C# Demo Code for Adding and Inserting Text to PDF File Page with .NET PDF Library.
add page to pdf online; add page to pdf acrobat
C# PDF Library SDK to view, edit, convert, process PDF file for C#
Capable of adding PDF file navigation features to your C# program. Perform annotation capabilities to mark, draw, and visualize objects on PDF document page.
add page numbers to pdf document; add blank page to pdf
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
215 
The offset parameters have the following physical meaning:  
• The Delay is the propagation time of a wave traveling through the standard. The Electrical Length 
is equal to the Delay times the speed of light in the vacuum and is a measure for the length of 
transmission line between the standard and the actual calibration plane. For a waveguide with 
permittivity 
ε
r
and mechanical length L
mech
the following relations hold: 
The default delay is 0 s, the default step width is 1 ns, corresponding to a step width of 299.792 mm for 
the electrical length. The relations hold for one-port and 2-port standards.  
• Z
0
is the Characteristic Impedance of thestandard. If the standard is terminated with Z
0
, then its 
input impedance is also equal to Z
0
Z
0
is not necessarily equal to the reference impedance of the 
system (depending on the Connector Type) or the terminal impedance of the standard. The 
characteristic impedance of the standard is only used in the context of calibration.  
The default characteristic impedance is equal to the reference impedance of the system.  
• The Loss is the energy loss along the transmission line due to the skin effect. For resistive lines 
and at RF frequencies the loss is approximately proportional to the square root of the frequency.   
In Agilent mode the Offset Loss is expressed in units of Ω/s at a frequency of 1 GHz. The following 
formula holds: 
To determine an offset loss value experimentally, measure the delay in seconds and the loss in dB at 1 
GHz and use the formula above.  
The default Loss or Offset Loss is zero.  
The impedance for waveguides is frequency-dependent. If a waveguide line type is selected 
in the Offset Model dialog, the Char. Impedance field is disabled and indicates "varies" 
instead of a definite impedance value. Moreover no Loss or Offset Loss can be set.  
Offset parameters and standard types 
Offset parameters are used to describe all types of standards except the Sliding Match and the 
Attenuation. 
• The Sliding Match is a one-port standard with variable load parameters (sliding load) and 
unspecified length. The reference impedance is fixed and equal to the characteristic impedance of 
the connector type. No load and offset parameters need to be set.  
• The Attenuation is a two-port standard which is fully matched in both directions (the reflection 
factor at both ports is zero). No load and offset parameters need to be set.   
Remote 
control: 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> 
Modify Load 
Specifies the load parameters for a particular calibration standard describing its terminal impedance. This 
C# PDF Password Library: add, remove, edit PDF file password in C#
Able to perform PDF file password adding, deleting and changing C# project, XDoc.PDF provides some PDF security settings On this page, we will talk about how
add multi page pdf to word document; add a page to pdf file
C# PDF Annotate Library: Draw, edit PDF annotation, markups in C#.
Provide users with examples for adding text box to PDF and edit C#.NET: Draw Markups on PDF File. Enable users to draw various annotation markups on PDF page.
add a page to a pdf in acrobat; adding page numbers to a pdf document
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
216 
dialog is opened from the Add or View / Modify Standard... dialog (Modify Load... button).  
The circuit model for the load consists of capacitance C which is connected in parallel to an inductance L 
and a resistance R, both connected in series.  
• R is the constant resistive contribution. It is possible to select a special value (Open for 
Ω so 
that the inductance coefficients are irrelevant, Short for 0 Ω, Match for the reference impedance of 
the current connector type) or set any resistance R.  
• The fringing capacitance C and the residual inductance L are both assumed to be frequency-
dependent and approximated by the first four terms of the Taylor series around f = 0 Hz.   
Load parameters and standard types 
Load parameters are used to describe all types of standards except a Through, Sliding Match, aLineand 
an Attenuation. 
• The Through standard is a through-connection between two ports with minimum loss which is 
taken into account by the Offset Parameters. 
• The Sliding Match is a one-port standard with variable load parameters (sliding load), so there is 
no fixed load model.  
• The Line standard is a line of variable length with minimum loss which is taken into account by the 
Offset Parameters. 
• The Attenuation is a two-port standard which is fully matched in both directions (the reflection 
factor at both ports is zero). No load and offset parameters need to be set.   
Remote 
control: 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> 
Calibration Standard Types  
The following table gives an overview of the different standards and their offset and load models:  
Standard 
Type 
Characteristics 
Ideal Standard 
Offset 
Model 
Load 
Model 
Open 
Open circuit (one-port) 
Ω 
Short 
Short circuit (one-port) 
0 Ω 
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
217 
Offset short 
Short circuit with added electrical 
length offset, for waveguide 
calibration (one-port)  
0 Ω 
Match 
Matched broadband termination 
(one-port) 
Z
0
(reference 
impedance of the 
connector type) 
Sliding 
match 
One-port standard consisting of an 
air line with a movable, low-
reflection load element (sliding 
load). .  
– 
–  
– 
Reflect 
Unknown mismatched standard 
(one-port) 
Ω 
Through 
Through-connection with minimum 
loss (two-port) 
– 
– 
Line1, Line 
2  
Line(s) for TRL calibration with 
minimum loss (two-port) 
– 
– 
Attenuation 
Fully matched standard in both 
directions (two-port; the reflection 
factor at both ports is zero). 
– 
– 
– 
Symm. 
network 
Unknown mismatched reflection-
symmetric standard (two-port) 
– 
Remote 
control: 
For 
an 
overview 
of 
standard 
parameters 
see 
also 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> 
Sweep     
The Sweep submenu defines the scope of measurement in the current channel. This includes the sweep 
type with various parameters, the trigger conditions the periodicity of the measurement, and the sweep 
average.  
Sweeps 
A sweep is a series of consecutive measurements taken over a specified sequence of stimulus values. It 
represents the basic measurement cycle of the analyzer.  
The analyzer can perform sweeps at constant power but variable frequency (frequency sweeps); see 
Sweep Type.  
The sweeps are further specified by the number of measurement points, the total measurement time and 
the trigger mode. A measurement may consist of a single sweep or a series of sweeps repeated 
continuously. 
On the other hand, depending on the measurement task and the measured quantity, the measurement at 
each point can consist of several partial measurements with definite hardware settings. 
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
218 
The Sweep menu provides the following settings: 
• Sweep Type defines the sweep variable (frequency/power/time) and the position of sweep points 
in the sweep range. 
• Number of Points sets the total number of measurement points over the sweep range.  
• Frequency Step Size sets the distance between two consecutive frequency sweep points.  
• Meas Delay delays the start of each sweep.  
• Restart aborts the current measurement and restarts a new sweep sequence.  
• If Single (All Chans) is selected the measurement is terminated after a single sweep or a group of 
single sweeps defined in the Define Restart dialog.  
• Define Restart opens a dialog to specify how many sweeps are repeated.  
• Average On activates or de-activates the sweep average. With average on the measurement 
results are averaged over a selected number of consecutive sweeps (Average Factor). 
• Average Factor defines the number of consecutive sweeps to be averaged.  
• Restart Average starts a new average cycle, clearing all previous results and thus eliminating their 
effect on the new cycle. The new cycle is started as fast as possible; an ongoing sweep is 
terminated immediately.    
Sweep Type     
The Sweep Type submenu defines the frequency sweep type and the position of the sweep points across 
the sweep range. 
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
219 
• Lin Frequency is the default sweep type. The stimulus frequency is swept in equidistant steps 
over the continuous frequency range. In a Cartesian diagram, the x-axis is a linear frequency axis. 
• Log Frequency is analogous to Lin. Frequency, however, the frequency is swept in equidistant 
steps on a logarithmic scale. In a Cartesian diagram, the x-axis is a logarithmic frequency axis. 
• Segmented Frequency is analogous to Lin. Frequency but uses a sweep range that can be 
composed of several continuous frequency ranges or single frequency points defined via Define 
Segments. A valid segment list must be defined before activating Segment Frequency.  
Lin Frequency  
In a Lin. Frequency sweep the stimulus frequency is swept in equidistant steps over the continuous 
frequency range. The frequency range (sweep range) is defined with the Stimulus settings. The step width 
between two consecutive sweep points is constant and given by <Span>/(n - 1) where n is the specified 
Number of Points (n > 1). The internal generator power can be set, if so desired, in the Power Bandwidth 
Average submenu.   
Lin. Frequency sweep corresponds to the analysis of a signal over the frequency, as obtained e.g. by 
means of a spectrum analyzer. This is the default sweep type. In a Cartesian diagram the measurement 
result is displayed as a trace over a linear frequency scale (spectral representation). The following 
example shows a Lin. Frequency sweep with a stimulus range between 4.5 GHz and 6 GHz, the forward 
transmission parameter S12 as measured quantity, and a dB Mag scaled y-axis.  
R&S
®
ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-01 
220 
Remote control: 
[SENSe<Ch>:]SWEep:TPYE 
LINear 
[SENSe<Chn>:]FUNCtion[:ON] "XFRequency:..."  
Log Frequency  
In a Log. Frequency sweep the stimulus frequency is swept on a logarithmic scale over the continuous 
frequency range. The frequency range (sweep range) is defined with the Stimulus settings. The sweep 
points are calculated from the Span and the specified Number of Points (n > 1) with the condition that the 
step width is constant on the logarithmic scale. The internal generator power can be set, if so desired, in 
the Power Bandwidth Average submenu.     
Log Frequency sweeps are suitable for the analysis of a DUT over a large frequency range, e.g. over 
several octaves. In a Cartesian diagram the measurement result is displayed as a trace over a logarithmic 
frequency scale. The following example shows a Log. Frequency sweep with a stimulus range between 50 
MHz and 6 GHz, the forward transmission parameter S12 as measured quantity, and a dB Mag scaled y-
axis 
Remote control: 
[SENSe<Ch>:]SWEep:TPYE 
LOGarithmic 
[SENSe<Chn>:]FUNCtion[:ON] "XFRequency:..."  
Segmented Frequency  
In a Segmented Frequency sweep the sweep range can be composed of several continuous, non-