R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
301 
the limit line.  
• TLIN and TLOG select time units [default: s]. 
• PLIN, PLOG and PSINgle select power units [default: dBm].  
*RST value 
FLIN 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
SWE:TYPE POW  
Select a power sweep. 
CALC:LIM:CONT:DOM PLIN  
Delete all existing limit line segments and select level units for the limit line of the 
active trace.  
CALC:LIM:CONT -20 DBM, -10 DBM  
Define a limit line segment in the stimulus range between –20 dBm and –10 dBm. 
CALCulate<Chn>:LIMit:CONTrol:SHIFt <numeric_value> 
Shifts the limit line in horizontal direction.  
<Chn> 
Channel number used to identify the active trace.  
<numeric_value> 
Amount that the limit line is shifted. 
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. [Hz 
for frequency sweeps, dBm for power sweeps, s for time sweeps, see 
CALCulate<Chn>:LIMit:CONTrol:DOMain]  
*RST value 
–  
SCPI, Command 
Types 
Device-specific, no query.   
Example: 
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ  
Define a limit line segment in the stimulus range between 1 GHz and 2 GHz, 
using default response values.  
CALC:LIM:CONT:SHIF 1; CALC:LIM:CONT?  
Shift the segment by 1 Hz. The modified limit line segment extends from 
1000000001 (Hz) to 2000000001 (Hz).  
CALCulate<Chn>:LIMit:DATA <type>, <start_stim>, <stop_stim>, <start_resp>, 
<stop_resp>{,<type>, <start_stim>, <stop_stim>, <start_resp>, <stop_resp>} 
Defines the limit line type, the stimulus and response values for a limit line with an arbitrary number of limit 
line segments.  
Add page to pdf reader - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add page number to pdf file; adding page numbers to a pdf in reader
Add page to pdf reader - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add page numbers to a pdf document; add a page to a pdf in reader
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
302 
In contrast to CALCulate<Chn>:LIMit:CONTrol[:DATA], this command does not overwrite 
existing limit line segments. The defined segments are appended to the segment list as new 
segments.  
<Chn> 
Channel number used to identify the active trace.  
<type> 
Identifier for the type of the limit line segment. 
Range [def. unit] 
0 – Limit line segment off, segment defined but no limit check performed  
1 – Upper limit line segment 
2 – Lower limit line segment 
<start_stim>, 
<stop_stim>, 
<start_resp>, 
<stop_resp> 
Stimulus and response values of the first and last points of the limit line segment. 
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. [Hz for 
frequency sweeps, dBm for power sweeps, s for time sweeps, see 
CALCulate<Chn>:LIMit:CONTrol:DOMain]   
*RST values 
– (no limit line defined after *RST)   
SCPI, 
Command 
Types 
Device-specific, with query.  
Example: 
*RST; CALC:LIM:CONT 1 GHZ, 1.5 GHZ  
Define an upper limit line segment in the stimulus range between 1 GHz and 1.5 
GHz, using default response values.  
CALC:LIM:DATA 1,1500000000, 2000000000,2,3  
Define an upper limit line segment in the stimulus range between 1.5 GHz and 2 
GHz, assigning response values of +2 dBm and +3 dBm. 
CALC:LIM:DISP ON  
Show the limit line segment in the active diagram.  
CALCulate<Chn>:LIMit:DELete:ALL  
Deletes all limit line segments.   
<Chn> 
Channel number used to identify the active trace.  
*RST value 
– 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
*RST; CALC:LIM:CONT 1 GHZ, 1.5 GHZ  
Define an upper limit line segment in the stimulus range between 1 GHz and 1.5 
GHz, using default response values.  
CALC:LIM:DATA 1,1500000000, 2000000000,2,3  
Define an upper limit line segment in the stimulus range between 1.5 GHz and 2 
GHz, assigning response values of +2 dBm and +3 dBm. 
CALC:LIM:DEL:ALL 
C# PDF insert image Library: insert images into PDF in C#.net, ASP
How to insert and add image, picture, digital photo, scanned signature or logo into PDF document page in C#.NET class application?
add page numbers to pdf document in preview; add a page to a pdf file
VB.NET PDF insert image library: insert images into PDF in vb.net
VB.NET PDF - Add Image to PDF Page in VB.NET. Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
add and delete pages in pdf online; adding page numbers to pdf
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
303 
Delete both created limit line segments.  
CALCulate<Chn>:LIMit:DISPlay[:STATe] <Boolean> 
Displays or hides the entire limit line (including all segments) associated to the active trace.  
<Chn> 
Channel number used to identify the active trace.   
<Boolean> 
ON | OFF - Limit line on or off.   
*RST value 
OFF   
SCPI, Command 
Types 
Device-specific, command or query.   
Example: 
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ  
Define an upper limit line segment in the stimulus range between 1 GHz and 2 
GHz, using default response values.  
CALC:LIM:DISP ON  
Show the limit line segment in the active diagram.  
CALCulate<Chn>:LIMit:FAIL? 
Returns a 0 or 1, to indicate whether or not the limit check has failed.  
Use CALCulate<Chn>:CLIMits:FAIL? to perform a composite (global) limit check.  
<Chn> 
Channel number used to identify the active trace.   
Response 
0 | 1 - 0 represents pass, 1 represents fail.  
*RST value 
 
SCPI, Command 
Types 
Confirmed, query only.   
Example: 
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ  
Define an upper limit line segment in the stimulus range between 1 GHz and 2 
GHz, using default response values.  
CALC:LIM:STAT ON; FAIL?  
Switch the limit check on and query the result.   
CALCulate<Chn>:LIMit:LOWer[:DATA] 
<numeric_value>,<numeric_value>{,<numeric_value>,<numeric_value>}  
Defines the response (y-axis) values of the lower limit line and/or creates new limit line segments.  
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
VB.NET Page: Insert PDF pages; VB.NET Page: Delete PDF Add Image to PDF; VB.NET Protect: Add Password to VB.NET Annotate: PDF Markup & Drawing. XDoc.Word for C#;
add page pdf reader; adding pages to a pdf document
VB.NET PDF Password Library: add, remove, edit PDF file password
On this page, we will illustrate how to protect PDF document via password by using simple VB.NET demo code. Open password protected PDF. Add password to PDF.
add page numbers pdf files; adding a page to a pdf
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
304 
The 
commands 
CALCulate<Chn>:LIMit:LOWer[:DATA] 
and 
CALCulate<Chn>:LIMit:UPPer[:DATA] use a fixed numbering scheme for limit line 
segments: Upper limit line segments are assigned odd numbers (1, 3, 5,...), lower limit line 
segments are assigned even numbers (2, 4, 6,...).   
Rules for creating segments 
The following rules apply to an active trace with n existing upper and n existing lower limit line segments: 
• An odd number of values is rejected; an error message –109,"Missing parameter..." is generated.  
• An even number of 2*k values updates or generates k lower limit line segments.  
• For n > k the response values of all existing lower limit line segments no. 2, 4, ...,2*k are updated, 
the existing upper and lower limit line segments no. 2*k+1, ..., 2*n are deleted. The existing upper 
limit line segments no. 1, 3, 2*k–1 are not affected.  
• For n < k the response values of the lower limit line segments no. 2, 4 to 2*n are updated, the 
lower limit line segments 2*n+2, 2*n+4,..., 2*k are generated with default stimulus values (see 
CALCulate<Chn>:LIMit:CONTrol[:DATA]. In addition, the missing upper limit line segments 
2*n+1, 2*n+3,..., 2*k–1 are generated with default stimulus and response values  
<Chn> 
Channel number used to identify the active trace.  
<numeric_value> 
Pair(s) of response values.  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. [dB] 
*RST value 
– 
The response value of a segment that is created implicitly, e.g. an upper limit 
line segment, is –20 dB.   
SCPI, Command 
Types 
Confirmed, with query.  
Example: 
CALC:LIM:LOW -10, 0, 0, -10  
Define the following lower and (default) upper limit line segments: 
CALC:LIM:DISP ON  
Show the limit line segments in the active diagram.  
CALCulate<Chn>:LIMit:LOWer:FEED 
<stimulus_offset>,<response_offset>[,<trace_name>] 
Generates a lower limit line using the stimulus values of a data or memory trace and specified offset 
values.   
<Chn> 
Channel number used to identify the active trace. This trace provides the 
stimulus data for the limit line unless another trace <trace_name> is 
C# PDF Password Library: add, remove, edit PDF file password in C#
your PDF document in C# project, XDoc.PDF provides some PDF security settings. On this page, we will talk about how to achieve this via Add necessary references
add page to pdf online; add a page to a pdf
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
DLLs for Deleting Page from PDF Document in VB.NET Class. In order to run the sample code, the following steps would be necessary. Add necessary references:
adding a page to a pdf in reader; add page number to pdf preview
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
305 
specified.  
<stimulus_offset> 
Stimulus offset value, used to shift all imported limit line segments in 
horizontal direction.  
Range [def. unit] 
-1000 GHz to +1000 GHz [Hz]  
*RST value 
0 Hz 
<response_offset> 
Response offset value, used to shift all imported limit line segments in vertical 
direction.  
Range [def. unit] 
-10
12
dB to +10
12
dB [dB]  
*RST value 
0 dB 
<trace_name> 
Name of the selected trace as used e.g. in 
CALCulate<Ch>:PARameter:SDEFine. If no trace name is specified the 
analyzer uses the active trace no. <Chn>.  
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
CALC:LIM:LOW:FEED 1 GHZ, -10 
Use the stimulus values of the active trace, shifted by 1 GHz to the right and 
decreased by 10 dB, to create a lower limit line.  
CALC:LIM:LOW:SHIF -3; CALC:LIM:CONT:SHIF 1 GHz 
Shift the limit line by an additional –3 dB in vertical and by 1 GHz in horizontal 
direction. If an upper limit line exists it is also shifted. 
CALCulate<Chn>:LIMit:LOWer:SHIFt <numeric_value>  
Shifts all lower and upper limit line segments assigned to the active trace in vertical direction.   
This command is identical with CALCulate<Chn>:LIMit:UPPer:SHIFt.  
<Chn> 
Channel number used to identify the active trace.   
<numeric_value> 
Response offset value for all limit line segments.  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. 
[dB]  
*RST value 
–  
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
See CALCulate<Chn>:LIMit:LOWer:FEED. 
CALCulate<Chn>:LIMit:LOWer:STATe <Boolean> 
Switches the lower limit check on or off. Lower limit line segments are assigned even numbers; see 
CALCulate<Chn>:LIMit:LOWer[:DATA]. The command does not affect segments with odd numbers.  
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
C#.NET Project DLLs for Deleting PDF Document Page. In order to run the sample code, the following steps would be necessary. Add necessary references:
adding page numbers to pdf in preview; add page numbers to pdf
C# PDF Sticky Note Library: add, delete, update PDF note in C#.net
C#.NET Tutorial for How to Add a Sticky Note Annotation to PDF Page with Visual enable users to annotate PDF without adobe PDF reader control installed.
add page numbers to pdf preview; add page numbers pdf file
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
306 
Use CALCulate<Chn>:LIMit:STATe to switch on or off the entire limit check, including 
upper and lower limit lines.  
<Chn> 
Channel number used to identify the active trace.   
<Boolean> 
ON | OFF - Limit check on or off.  
*RST value 
OFF  
SCPI, Command 
Types 
Confirmed, command or query.   
Example: 
CALC:LIM:LOW -10, 0, 0, -10  
Define the following lower and (default) upper limit line segments: 
CALC:LIM:LOW:STAT ON; CALC:LIM:FAIL?  
Switch the limit check on and query the result.   
CALCulate<Chn>:LIMit:RDOMain:COMPlex S | SINV | Y | Z | YREL | ZREL  
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line. 
The units of the stimulus values are defined via CALCulate<Chn>:LIMit:CONTrol:DOMain.  
This command is complemented by CALCulate<Chn>:LIMit:RDOMain:FORMat and 
CALCulate<Chn>:LIMit:RDOMain:SPACing.  
<Chn> 
Channel number used to identify the active trace.  
Parameters 
Keyword for the physical unit of the response values.  
Range [def. unit] 
The parameters form four groups:  
• S and SINV select relative units (dB) for the limit line. 
• Y selects admittance units (S/Siemens).  
• Z selects impedance units (Ω).  
• YREL and ZREL select dimensionless numbers (U).  
[–]  
*RST value 
–   
SCPI, Command Types 
Device-specific, no query.   
CALCulate<Chn>:LIMit:RDOMain:FORMat COMPlex | MAGNitude | PHASe | REAL | 
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
307 
IMAGinary | SWR | GDELay | L | C 
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line. 
The units of the stimulus values are defined via CALCulate<Chn>:LIMit:CONTrol:DOMain.  
This command is complemented by CALCulate<Chn>:LIMit:RDOMain:COMPlex and 
CALCulate<Chn>:LIMit:RDOMain:SPACing.  
<Chn> 
Channel number used to identify the active trace.  
Parameters 
Keyword for the physical unit of the response values.  
Range [def. unit] 
The parameters form four groups:  
• COMPlex, REAL, IMAGinary, and SWR select dimensionless numbers 
(U) for the limit line.  
• MAGNitude selects relative units (dB).  
• PHASe selects phase units (deg).  
• GDELay selects time units (s).  
• L selects inductance units (H/Henry).  
• C selects capacitance units (F/Farad). 
[–]  
*RST value 
– 
SCPI, Command 
Types 
Device-specific, no query.  
CALCulate<Chn>:LIMit:RDOMain:SPACing LINear | LOGarithmic | dB | SIC 
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line. 
The units of the stimulus values are defined via CALCulate<Chn>:LIMit:CONTrol:DOMain.  
This command is complemented by CALCulate<Chn>:LIMit:RDOMain:COMPlex and 
CALCulate<Chn>:LIMit:RDOMain:FORMat.  
<Chn> 
Channel number used to identify the active trace.  
Parameters 
Keyword for the physical unit of the response values.  
Range [def. unit] 
The analyzer uses dB units, irrespective of the parameter selected.
[–]  
*RST value 
– 
SCPI, Command Types 
Device-specific, no query.  
CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPlitude:STARt <numeric_value> 
Changes the start response value (i.e. the response value assigned to the start stimulus value) of a limit 
line segment. A segment must be created first to enable this command (e.g CALC:LIM:DATA).  
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
308 
To define the response values of several limit line segments with a single command, use 
CALCulate<Chn>:LIMit:LOWer[:DATA] 
or 
CALCulate<Chn>:LIMit:UPPer[:DATA].  
<Chn> 
Channel number used to identify the active trace.  
<Seg> 
Segment number  
<numeric_value> 
Response value  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition.  
[dB]  
*RST value 
–  
The default response values of a segment that is created by defining its 
stimulus values only (e.g. by means of 
CALCulate<Chn>:LIMit:CONTrol[:DATA]), is –20 dB.  
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
CALC:LIM:DATA 1,1500000000, 2000000000,2,3  
Define an upper limit line segment (segment no. 1) in the stimulus range 
between 1.5 GHz and 2 GHz, assigning response values of +2 dBm and +3 
dBm. 
:CALC:LIM:SEGM:AMPL:STAR 5; STOP 5; CALC:LIM:SEGM:TYPE LMIN 
Change the segment to a lower limit line segment with a constant response 
value of +5 dB.  
CALC:LIM:DATA?  
Query the type, the stimulus and response values of the created segment with 
a single command. The response is 2,1000000,2000000,5,5. 
CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPlitude:STOP <numeric_value> 
Changes the stop response value (i.e. the response value assigned to the stop stimulus value) of a limit 
line segment. A segment must be created first to enable this command (e.g CALC:LIM:DATA).  
To define the response values of several limit line segments with a single command, use 
CALCulate<Chn>:LIMit:LOWer[:DATA] 
or 
CALCulate<Chn>:LIMit:UPPer[:DATA].  
<Chn> 
Channel number used to identify the active trace.  
<Seg> 
Segment number  
<numeric_value> 
Response value  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition.  
[dB]  
*RST value 
–  
The default response values of a segment that is created by defining its 
stimulus values only (e.g. by means of 
CALCulate<Chn>:LIMit:CONTrol[:DATA]), is –20 dB.  
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
309 
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
See CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPlitude:STARt.   
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt <numeric_value> 
Changes the start stimulus value (i.e. the largest or smallest stimulus value) of a limit line segment. A 
segment must be created first to enable this command (e.g CALC:LIM:DATA).  
To define the stimulus values of several limit line segments with a single command, use 
CALCulate<Chn>:LIMit:CONTrol[:DATA].  
<Chn> 
Channel number used to identify the active trace.  
<Seg> 
Segment number  
<numeric_value> 
Frequency, power or time value, to be defined in accordance with the domain 
setting (CALCulate<Chn>:LIMit:CONTrol:DOMain).  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. In 
particular, the start value can be larger than the stop value 
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP.  
[Hz for frequency sweeps, dBm for power sweeps, s for time sweeps]  
*RST value 
–  
A segment that is created implicitly, e.g. by means of 
CALCulate<Chn>:LIMit:UPPer[:DATA] or 
CALCulate<Chn>:LIMit:LOWer[:DATA,], covers the maximum sweep 
range of the analyzer.   
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
CALC:LIM:DATA 1,1500000000, 2000000000,2,3  
Define an upper limit line segment (segment no. 1) in the stimulus range 
between 1.5 GHz and 2 GHz, assigning response values of +2 dBm and +3 
dBm. 
CALC:LIM:SEGM:STIM:STAR 1GHZ; STOP 2 GHZ; CALC:LIM:SEGM:TYPE LMIN  
Change the segment to a lower limit line segment with a stimulus range 
between 1 GHz and 2 GHz.   
CALC:LIM:DATA?  
Query the type, the stimulus and response values of the created segment with a 
single command. The response is 2,1000000,2000000,2,3.  
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP <numeric_value> 
Changes the stop stimulus value (i.e. the largest or smallest stimulus value) of a limit line segment. A 
segment must be created first to enable this command (e.g CALC:LIM:DATA).  
R&S
®
ZVL 
Command Reference 
CALCulate 
Operating Manual 1303.6580.32-01 
310 
To define the stimulus values of several limit line segments with a single command, use 
CALCulate<Chn>:LIMit:CONTrol[:DATA].  
<Chn> 
Channel number used to identify the active trace.  
<Seg> 
Segment number  
<numeric_value> 
Frequency, power or time value, to be defined in accordance with the domain 
setting (CALCulate<Chn>:LIMit:CONTrol:DOMain).  
Range [def. unit] 
Almost no restriction for limit segments; see Rules for Limit Line Definition. In 
particular, the stop value can be smaller than the start value 
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:START.  
[Hz for frequency sweeps, dBm for power sweeps, s for time sweeps]  
*RST value 
–  
A segment that is created implicitly, e.g. by means of 
CALCulate<Chn>:LIMit:UPPer[:DATA] or 
CALCulate<Chn>:LIMit:LOWer[:DATA,], covers the maximum sweep 
range of the analyzer.   
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
See CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt.    
CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE LMIN | LMAX | OFF 
Selects the limit line type for a limit line segment. This can be done before or after defining the stimulus 
and response values of the segment, however, a segment must be created first to enable this command 
(e.g CALC:LIM:DATA).  
The type command overwrites the CALCulate<Chn>:LIMit:DATA settings and is overwritten 
by them. It is not affected by the other commands in the LIMit subsystem defining stimulus 
and response values of limit lines.  
<Chn> 
Channel number used to identify the active trace.  
<Seg> 
Segment number 
Parameters 
Limit line type  
Range [def. unit] 
LMAX (upper limit line segment), LMIN (lower limit line segment), OFF (limit 
check switched off, limit line segment not deleted) [–]  
*RST value 
LMAX 
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
*RST; CALC:LIM:UPP 1 GHZ, 2 GHZ  
Define an upper limit line segment in the stimulus range between 1 GHz and 2 
GHz, using default response values.  
CALC:LIM:SEG:TYPE LMIN  
Turn the defined limit line segment into a lower limit line segment.  
Documents you may be interested
Documents you may be interested