R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
411 
SCPI, 
Command 
Types 
Device-specific, no query 
Example: 
CORR:COLL:METH:DEF 'Test',RSHort,1 
Select a one-port normalization at port 1 with a short standard as calibration type. 
CORR:COLL:SEL SHOR,1 
Measure a short standard connected to port 1 and store the measurement results of 
this standard.  
CORR:COLL:SAVE; DEL  
Calculate the system error correction data and apply them to the active channel, 
then delete the data.  
[SENSe<Ch>:]CORRection:COLLect:METHod FRTRans | TOSM | FOPort1 | 
FOPort2 | FOPort12 | FOPTport | ROPTport | REFL1 | REFL2 | REFL12 | TPORt 
Selects a one-port or two-port calibration type at ports 1/2.  
Use the generalized command [SENSe<Ch>:]CORRection:COLLect:METHod:DEFine to 
select the calibration type for arbitrary analyzer ports or a multiport calibration type.  
<Ch> 
Channel number of the calibrated channel.  
Parameters 
Calibration types: TOSM, Full One Port, One Path Two Port, Normalization 
(REFL1, REFL2 and REFL12 for one-port, TPORT for two-port), T  
The numbers in the parameters denote the analyzer ports. Parameters for two-port 
calibration types contain no numbers because the command is only valid for ports 
1 and 2.  
*RST value 
– 
SCPI, 
Command 
Types 
Confirmed (with device-specific calibration types), command or query. If several 
calibration types are assigned to channel <Ch>, the query returns a list of all 
calibration types.   
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE.  
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine '<cal_name>', REFL | 
RSHort | FOPort | FRTRans | OPTPort | TOSM, <port_no>[,<port_no>] 
Selects a one-port or two-port calibration type at arbitrary analyzer ports.  
<Ch> 
Channel number of the calibrated channel.  
'<cal_name>' 
Name of the calibration (string parameter). The name serves as a reference to 
delete a particular set of system correction data 
([SENSe<Ch>:]CORRection:COLLect:DELete).  
Parameters 
Calibration types: One-port Normalization (reflection), using an open (REFL) or a 
short (RSHort) standard, Full One Port, Two-port Normalization (transmission, 
FRTRans), One Path Two Port, TOSM, .  
<port_no> 
Port numbers of the analyzer. For an n-port calibration type (n = 1 to 2), n port 
Add page to pdf reader - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add page numbers to pdf in reader; adding page numbers to pdf files
Add page to pdf reader - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add a page to pdf file; adding a page to a pdf
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
412 
numbers must be specified. If more than n numbers are defined, the spare 
numbers (the last ones in the list) are ignored. Entering less than n numbers 
causes an error message.  
*RST value 
– 
SCPI, 
Command 
Types 
Device-specific, no query. Use [SENSe<Ch>:]CORRection:COLLect:METHod? 
to obtain a list of all calibration types for channel <Ch>.  
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE:SELected.  
[SENSe<Ch>:]CORRection:COLLect:SAVE 
Calculates the system error correction data from the acquired one or two-port measurement results 
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]), stores them and applies them to the calibrated 
channel <Ch>. To avoid incompatibilities, older system error correction data is deleted unless it has been 
transferred into a cal pool (MMEMory:STORe:CORRection <Ch>, '<file_name>'). 
This 
command 
is 
the 
ZVR-compatible 
equivalent 
of 
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected. It must be used in combination 
with the ZVR-compatible commands for calibration method and standard selection; see 
example below. 
<Ch> 
Channel number of the calibrated channel.  
*RST value 
– 
SCPI, Command 
Types 
Confirmed, no query 
Example: 
CORR:COLL:METH REFL1 
Select a one-port normalization at port 1 as calibration type. 
CORR:COLL OPEN1  
Measure an open standard connected to port 1 and store the measurement 
results of this standard.  
CORR:COLL:SAVE  
Calculate the system error correction data and apply them to the active 
channel. 
[SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault 
Generates a set of default system error correction data for the selected ports and calibration type. The 
default data set corresponds to a test setup which does not introduce any systematic errors; none of the 
measurement results acquired previously ([SENSe<Ch>:]CORRection:COLLect[:ACQuire]) is 
taken into account.  
The main purpose of the default correction data set is to provide a dummy system error 
correction which you can replace with your own, external correction data. You may have 
acquired the external data in a previous session or even on an other instrument. If you want 
C# PDF insert image Library: insert images into PDF in C#.net, ASP
How to insert and add image, picture, digital photo, scanned signature or logo into PDF document page in C#.NET class application?
add page numbers to pdf; add pages to pdf acrobat
VB.NET PDF insert image library: insert images into PDF in vb.net
VB.NET PDF - Add Image to PDF Page in VB.NET. Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
add pages to pdf; adding page to pdf in preview
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
413 
to use the external correction data on the analyzer, simply generate the default data set 
corresponding to your port configuration and calibration type and overwrite the default data. 
For details refer to the program example below.  
This command must be used in combination with the ZVR-compatible commands 
[SENSe<Ch>:]CORRection:COLLect:METHod and [SENSe<Ch>:]CORRection:DATa. Use 
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault if you want to use ZVL-
specific calibration commands or if you want to calibrate more than 2 ports.  
<Ch> 
Channel number of the calibrated channel.  
*RST value 
– 
SCPI, 
Command 
Types 
Device-specific, no query 
Example: 
CORR:COLL:METH REFL1 
Select a one-port normalization at port 1 with an open standard as calibration type. 
CORR:COLL:SAVE:DEF  
Calculate a dummy system error correction for the normalization at port 1. The 
dummy system error correction provides the reflection tracking error term 
'SCORR3'.  
CORR:DATA? 'SCORR3'  
Query the dummy system error correction term. The response is a 1 (written as 1,0 
for the real and imaginary part) for each sweep point (no attenuation and no phase 
shift between the analyzer and the calibration plane).  
CORR:DATA 'SCORR3',<ASCII_data> 
Replace the dummy system error correction term with your own correction data, 
transferred in ASCII format.  
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected 
Calculates the system error correction data from the acquired measurement results 
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected), stores them and applies them to 
the calibrated channel <Ch>. To avoid incompatibilities, older system error correction data is deleted 
unless it has been transferred into a cal pool (MMEMory:STORe:CORRection <Ch>, 
'<file_name>'). 
<Ch> 
Channel number of the calibrated channel.  
*RST value 
– 
SCPI, Command 
Types 
Confirmed, no query 
Example: 
CORR:COLL:METH:DEF 'Test',RSHort,1 
Select a one-port normalization at port 1 with a short standard as calibration 
type. 
CORR:COLL:SEL SHOR,1 
Measure a short standard connected to port 1 and store the measurement 
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
VB.NET Page: Insert PDF pages; VB.NET Page: Delete PDF Add Image to PDF; VB.NET Protect: Add Password to VB.NET Annotate: PDF Markup & Drawing. XDoc.Word for C#;
add pages to pdf without acrobat; add page numbers to pdf files
VB.NET PDF Password Library: add, remove, edit PDF file password
On this page, we will illustrate how to protect PDF document via password by using simple VB.NET demo code. Open password protected PDF. Add password to PDF.
add page numbers to pdf using preview; add page to pdf
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
414 
results of this standard.  
CORR:COLL:SAVE:SEL  
Calculate the system error correction data and apply them to the active 
channel. 
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault 
Generates a set of default system error correction data for the selected ports and calibration type. The 
default data set corresponds to a test setup which does not introduce any systematic errors; none of the 
measurement 
results 
acquired 
previously 
([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected) is taken into account.  
The main purpose of the default correction data set is to provide a dummy system error 
correction which you can replace with your own, external correction data. You may have 
acquired the external data in a previous session or even on an other instrument. If you want 
to use the external correction data on the analyzer, simply generate the default data set 
corresponding to your port configuration and calibration type and overwrite the default data. 
For details refer to the program example below.  
<Ch> 
Channel number of the calibrated channel.  
*RST value 
– 
SCPI, 
Command 
Types 
Device-specific, no query 
Example: 
CORR:COLL:METH:DEF 'Test',RSHort,1 
Select a one-port normalization at port 1 with a short standard as calibration type. 
CORR:COLL:SAVE:SEL:DEF  
Calculate a dummy system error correction for the normalization at port 1. The 
dummy system error correction provides the reflection tracking error term 
'REFLTRACK'.  
CORR:CDAT? 'REFLTRACK',1,0  
Query the dummy system error correction term. The response is a 1 (written as 1,0 
for the real and imaginary part) for each sweep point (no attenuation and no phase 
shift between the analyzer and the calibration plane).  
CORR:CDAT 'REFLTRACK',1,0,<ASCII_data> 
Replace the dummy system error correction term with your own correction data, 
transferred in ASCII format.  
[SENSe<Ch>:]CORRection:COLLect:SCONnection<port_no> '<conn_name>', 
MALE | FEMale  
Selects a connector type at a specified port <port_no> and its gender. In contrast to 
[SENSe<Ch>:]CORRection:COLLect:CONnection<port_no>, this command uses a string variable 
to identify the connector type.  
C# PDF Password Library: add, remove, edit PDF file password in C#
your PDF document in C# project, XDoc.PDF provides some PDF security settings. On this page, we will talk about how to achieve this via Add necessary references
adding page numbers pdf; adding page numbers to a pdf document
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
DLLs for Deleting Page from PDF Document in VB.NET Class. In order to run the sample code, the following steps would be necessary. Add necessary references:
add pages to pdf in preview; add page to pdf without acrobat
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
415 
<Ch> 
Channel number of the calibrated channel.  
<port_no> 
Port numbers of the analyzer.  
Note: If the analyzer is set to use the same connectors at all ports 
([SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>:PORTs 
ALL), then a change of a connector type is valid for all ports. The gender of the 
connectors can still be different.  
Parameters 
Connector type (string parameter) and gender of the connectors. The gender 
designation MALE | FEMale is not necessary (and ignored) for sexless connector 
types.  
*RST value 
'N 50 Ohm',FEM for all ports.  
SCPI, 
Command 
Types 
Device-specific, command or query. 
Example: 
*RST; CORR:COLL:SCON1 'N 75 Ohm', MALE; SCON4?  
Change the connector type at port 1 from 'N 50 Ohm', FEM to 'N 75 Ohm', MALE. 
The connector type at the other ports is also changed to N 75 Ohm, however, the 
gender (female) is maintained. CORR:COLL:SCON4? returns 'N 75 Ohm', FEM.  
[SENSe<Ch>:]CORRection:CONNection    '<conn_name>', TEM | WGUide, GENDer 
| NGENder, <perm_rel>, <imped> 
Configures the user-defined connector types. 
<Ch> 
Channel number  
'<conn_name>' 
Name of the user-defined connectors, string parameter  
TEM | WGUide 
Transverse electric or waveguide type propagation mode  
GENDer | NGENder 
Polar connector type (m/f) 
Sexless connector type  
<perm_rel> 
Relative permittivity  
Range [def. unit] 
0.0000000001 to 1000. [–] UP/DOWN/MIN/MAX parameters are not 
available for this command  
*RST value 
–  
<imped> 
For TEM type connectors: reference impedance in Ω (without unit), 
For WGUide type connectors: cutoff frequency in Hz (without unit) 
Range [def. unit] 
Ref. impedance: 1µΩ to 1000 MΩ 
Cutoff frequency: 0 Hz to 1000 GHz  
*RST value 
–  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
CORR:CONN 'USERCON',TEM,GEND,1.00000,50 
Define a TEM type connector type names USERCON.  
CORR:CONN? 'USERCON' 
Query the properties of the configured connector type.   
CORR:CONN:DEL 'USERCON' 
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
C#.NET Project DLLs for Deleting PDF Document Page. In order to run the sample code, the following steps would be necessary. Add necessary references:
add page pdf; add multi page pdf to word document
C# PDF Sticky Note Library: add, delete, update PDF note in C#.net
C#.NET Tutorial for How to Add a Sticky Note Annotation to PDF Page with Visual enable users to annotate PDF without adobe PDF reader control installed.
adding page numbers to pdf; adding page numbers to a pdf file
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
416 
Delete the configured connector type.    
[SENSe<Ch>:]CORRection:CONNection:DELete '<conn_name>' 
Deletes a user-defined connector type named '<conn_name>'. 
<Ch> 
Channel number  
'<conn_name>' 
Name of the user-defined connectors, string parameter  
SCPI, Command Types 
Device-specific, no query 
Example: 
See [SENSe<Ch>:]CORRection:CONNection 
[SENSe<Ch>:]CORRection:DATA   'SCORR1' | ... | 'SCORR12' 
Writes or reads system error correction data for a specific channel <Ch> and calibration method 
([SENSe<Ch>:]CORRection:COLLect:METHod). The analyzer test ports 1 or 2 are implicitly specified with 
the correction terms. The setting command can be used to transfer user-defined correction data to the 
analyzer; the query returns the current correction data set. ASCII or block data can be transferred, 
depending on the selected data transfer format (FORMat[:DATA])  
Use the generalized command [SENSe<Ch>:]CORRection:COLLect:CDATa to transfer 
calibration data for arbitrary analyzer ports.  
<Ch> 
Channel number of the calibrated channel. If unspecified the numeric suffix is set to 
1. 
Error term 
parameters 
String parameters describing the different error terms, depending on the current 
calibration method; see table below. Each term contains one complex value (real and 
imaginary part) for each sweep point.  
Range [def. 
unit] 
The error terms consist of dimensionless complex numbers. The parameters must be 
transferred in full length and have the following meaning: 
'SCORR1' – Directivity at port 1  
'SCORR2' – Source match at port 1 
'SCORR3' – Reflection tracking at port 1 
'SCORR4' – Forward isolation between port 1 and port 2 
'SCORR5' – Load match at port 2 
'SCORR6' – Forward transmission tracking between port 1 and port 2 
'SCORR7' – Directivity at port 2  
'SCORR8' – Source match at port 2 
'SCORR9' – Reflection tracking at port 2 
'SCORR10' – Reverse isolation between port 2 and port 1 
'SCORR11' – Load match at port 1 
'SCORR12' – Reverse transmission tracking between port 2 and port 1 
*RST value 
–  
The analyzer provides a default calibration corresponding to a test setup which does 
not introduce any systematic errors; see 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
417 
[SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault.  
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault.   
The different calibration types of the analyzer provide the following error terms:  
Calibration type 
Parameter in 
[SENSe<Ch>:]CORRection:COLLect:METHod 
Available error terms 
(depending on port numbers) 
One-port normalization 
(reflection) using an open 
standard 
REFL1 
REFL2 
REFL12 
'SCORR3' 
'SCORR9' 
'SCORR3' and 'SCORR9'  
Full one port 
FOPort1  
FOPort2  
FOPort12 
'SCORR1' to 'SCORR3'  
'SCORR7' to 'SCORR9' 
'SCORR1' to 'SCORR3' and 
'SCORR7' to 'SCORR9' 
Two-port normalization 
FTRans 
RTRans 
FRTRans 
'SCORR6' 
'SCORR12' 
'SCORR6' and 'SCORR12' 
One path two port 
FOPTport 
ROPTport 
'SCORR1' to 'SCORR3', 
'SCORR6'  
'SCORR7' to 'SCORR9', 
'SCORR12'  
TOSM 
TOSM 
'SCORR1' to 'SCORR12' 
(at present the isolation terms 
'SCORR4' and 'SCORR10' are 
not included)  
[SENSe<Ch>:]CORRection:DATA:PARameter? 
Returns the sweep settings of the active system error correction for channel <Ch>.  
<Ch> 
Channel number of the calibrated channel.  
Response 
Five values, corresponding to the start and stop frequency of the calibration 
sweep, the number of points, the source power, and the sweep type (LIN | LOG | 
SEGM).   
*RST value 
– (the command generates an execution error if no system error correction is 
active) 
SCPI, 
Command 
Types 
Device-specific, query only 
Example: 
See [SENSe<Ch>:]CORRection:DATE?  
[SENSe<Ch>:]CORRection:DATE? 
Returns the date and time when the active system error correction data for channel <Ch> was acquired.  
<Ch> 
Channel number of the calibrated channel.  
*RST value 
– (the command generates an execution error if no system error correction is active)
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
418 
SCPI, 
Command 
Types 
Device-specific, query only 
Example: 
CORR:COLL:METH REFL1 
Select a one-port normalization at port 1 as calibration type. 
CORR:COLL OPEN1  
Measure an open standard connected to port 1 and store the measurement results 
of this standard.  
CORR:COLL:SAVE  
Calculate the system error correction data and apply them to the active channel. 
CORR:DATE?  
Query the time when the system error correction became active. The analyzer 
returns the data and time, e.g. '03/20/06,18:30:39'.   
CORR:DATA:PAR?  
Query the sweep settings for the calibration sweep. The analyzer returns the start 
and stop frequency, the number of points, source power, and the sweep type, e.g. 
300000,8000000000,201,0,LIN  
CORR:SST?  
Query the calibration status. The analyzer returns 'CAL OFF' (because the 
performed one-port calibration is not sufficient for the measured transmission S-
paramter S
21
).   
[SENSe<Ch>:]CORRection:EDELay<port_no>:AUTO ONCE 
Defines the offset parameter for the active test port such that the residual delay of the active trace (defined 
as the negative derivative of the phase response) is minimized across the entire sweep range. 
<Ch> 
Channel number of the offset-corrected channel.  
<port_no> 
Port number of the analyzer. This numeric suffix is ignored; the active port is 
determined by the active trace.  
ONCE 
Applies the Auto Length function.  
*RST value 
–   
SCPI, Command 
Types 
Device-specific, no query 
Example: 
*RST; CORR:EDEL:AUTO ONCE  
Reset the instrument and apply the auto length function to the default trace 
(Trc1 in channel 1). 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
419 
[SENSe<Ch>:]CORRection:EDELay<port_no>:DIELectric <permittivity> 
Defines the permittivity for the offset correction at test port <port_no>. 
<Ch> 
Channel number of the offset-corrected channel 
<port_no> 
Port number of the analyzer  
<permittivity> 
Permittivity  
Range [def. unit] 
1 to +3.4028234664E+038 m 
*RST value 
1.00062 [–]  
SCPI, Command Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:EDELay<port_no>:DISTance <length> 
Defines the offset parameter for test port <port_no> as a mechanical length. 
<Ch> 
Channel number of the offset-corrected channel  
<port_no> 
Port number of the analyzer  
<length> 
Mechanical length  
Range [def. 
unit] 
–3.402823466E+038 m to +3.4028234664E+038 m. In contrast to the electrical 
length ([SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth), the distance can 
not be incremented (the UP/DOWN parameters do not work).  
*RST value 
0 m [m]  
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth 
[SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth <length> 
Defines the offset parameter for test port <port_no> as an electrical length. 
<Ch> 
Channel number of the offset-corrected channel  
<port_no> 
Port number of the analyzer  
<length> 
Electrical length  
Range [def. 
unit] 
–3.4028234664E+038 m to +3.4028234664E+038 m [m]. The increment (UP, 
DOWN) is 1 cm. 
*RST value 
0 m   
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
CORR:EDEL2:ELEN 0.3 
Define an electrical length of 30 cm for channel 1 and port no. 2.  
CORR:EDEL2:DIST?; DIEL? 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
420 
Query the values of the mechanical length and the permittivity at port 2. The 
mechanical length is equal to the electrical length divided by the square root of the 
permittivity; the latter is set to its default value. The response is 
0.29990704322;1.00062.  
CORR:EDEL2? 
Query the value of the delay at port 2. The delay is equal to the electrical length 
divided by the speed of light in the vacuum, so the response is 1.0006922856E-009
CORR:LOSS2 2; LOSS2:FREQ 1.5 GHz; OFFS 3 dB 
Define the offset loss parameters at port 2.  
[SENSe<Ch>:]CORRection:EDELay<port_no>[:TIME] <delay> 
Defines the offset parameter for test port <port_no> as a delay time. 
<Ch> 
Channel number of the offset-corrected channel  
<port_no> 
Port number of the analyzer  
<delay> 
Delay  
Range [def. 
unit] 
-3.40282346638529E+038 s to +3.40282346638529E+038 s [s]. In contrast to the 
electrical length ([SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth), the delay 
time can not be incremented (the UP/DOWN parameters do not work).   
*RST value 
0 s   
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth 
[SENSe<Ch>:]CORRection:FACTory[:STATe] <Boolean> 
Enables or disables the factory calibration for a particular channel.  
<Ch> 
Channel number of the calibrated channel  
<Boolean> 
Enable or disable factory calibration  
*RST value 
ON   
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
*RST; CORR:FACT? 
Reset the instrument and query whether the factory calibration for channel 1 is 
enabled. The response is 1.  
[SENSe<Ch>:]CORRection:LOSS<port_no>  <DC_loss> 
Defines the frequency-independent part (DC value) of the offset loss.  