Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
431 
[SENSe<Ch>:]SEGMent<Seg>:DEFine <Start>,<Stop>,<Points>,<Power>,<Point 
Delay>,<Unused>,<Meas. Bandwidth>[,<LO>] 
Re-defines a sweep segment with specific channel settings (Insert New Segment). The segment replaces 
an existing segment <Seg> in the segment list. The modified segment must not overlap with any of the 
existing segments.  
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings 
except <LO> can be changed for existing segments using other commands of the 
[SENSe<Ch>:]SEGMent<Seg>... subsystem.   
Note: Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use 
[SENSe<Ch>:]SEGMent<Seg>:INSert (no query) to insert a new segment into the current 
segment list.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. Segment numbers must be sequential. The specified 
segment number must be smaller or equal to the number of existing segments plus 
1.   
If segment number <Seg> already exists, it is replaced by the new segment.  
<Start>, 
<Stop> 
Start and stop frequency of the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.  
Range [def. 
unit] 
A new segment must not overlap with any of the existing segments. Besides the 
frequency range depends on the analyzer model. [Hz]  
<Points> 
Number of sweep points in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts.  
Range [def. 
unit] 
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
<Power> 
Internal source power in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:POWer. 
Range [def. 
unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to the 
data sheet [dBm]. UP and DOWN increment/decrement the source power in 1-dB 
steps.  
<Point 
Delay> 
Delay for each partial measurement in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting 
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime, this parameter is 
replaced by <Time>.   
Range [def. 
unit] 
0 s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the segment, 
which is equivalent to a meas. delay of 0 s.  
<Unused> 
Ignored parameter, should be set to the default value 0.  
<Meas. 
Bandwidth> 
Resolution bandwidth in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]. 
Range [def. 
unit] 
1.0E-6 Hz to 5 MHz [Hz].  
<LO> 
Position of the local oscillator frequency LO relative to the RF frequency (Spur 
Avoid). In remote control this parameter must be set when a sweep segment is 
created.  
Range [def. 
unit] 
POSitive: LO > RF 
SCPI, 
Device-specific, with query (used to retrieve the channel settings for a particular 
Adding a page to a pdf in reader - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
adding pages to a pdf document in preview; adding page numbers to pdf in
Adding a page to a pdf in reader - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add pages to an existing pdf; adding a page to a pdf document
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
432 
Command 
Types 
sweep segment).  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings.  
SEGM:DEF?  
Query the channel settings for the new segment.  
[SENSe<Ch>:]SEGMent<Seg>:DELete 
Deletes 
the 
specified 
(single) 
sweep 
segment 
(Del. 
Selected 
Segment)
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL deletes all segments in the channel.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. If unspecified, the value is set to 1.  
*RST value 
– 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus NORMal selectivity.  
SEGM:DEL  
Delete the created segment. 
[SENSe<Ch>:]SEGMent:DELete:ALL 
Deletes all sweep segments in the channel (Del. All Segments). [SENSe<Ch>:]SEGMent<Seg>:DELete 
deletes a single segment.  
<Ch> 
Channel number.   
*RST value 
– 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings 
and thus NORMal selectivity.  
SEGM:ALL  
Delete the created segment and all segments in the channel created before.  
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
page modifying page, you will find detailed guidance on creating, loading, merge and splitting PDF pages and Files, adding a page into PDF document, deleting
add and delete pages in pdf online; add page to pdf acrobat
VB.NET PDF Library SDK to view, edit, convert, process PDF file
Perform annotation capabilities to mark, draw, and visualize objects on PDF document page. Capable of adding PDF file navigation features to your VB.NET program
adding page numbers to a pdf in reader; add page to a pdf
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
433 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer? 
Returns the center frequency of sweep segment no. <Seg>. 
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
Response 
Center frequency of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz].   
*RST value 
– 
SCPI, Command 
Types 
Device specific, query only. 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz.   
SEGM:FREQ:CENT?  
Query the center frequency of the created segment. The response is 
1.2500000000E+006.  
Note:  The frequency range of the sweep segment can be changed via 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.  
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN? <numeric_value> 
Returns the width of the frequency range of sweep segment no. <Seg>. 
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
Response 
Frequency span of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz].   
*RST value 
– 
SCPI, Command 
Types 
Device specific, query only. 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz.   
SEGM:FREQ:SPAN?  
Query the span of the created segment. The response is 
5.0000000000E+005.  
Note:  The frequency range of the sweep segment can be changed via 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.  
C# PDF insert image Library: insert images into PDF in C#.net, ASP
digital photo, scanned signature or logo into PDF document page in C# solve this technical problem, we provide this C#.NET PDF image adding control, XDoc
add contents page to pdf; add a page to a pdf
C# PDF insert text Library: insert text into PDF content in C#.net
text and plain text to PDF page using .NET XDoc.PDF component in C#.NET class. Supports adding text to PDF in preview without adobe reader installed in ASP.NET.
add a blank page to a pdf; add and delete pages from pdf
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
434 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt <numeric_value> 
Defines the Start frequency of sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<numeric_value> 
Start frequency of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz]. The increment is 0.1 kHz.  
*RST value 
–  (no sweep segment defined after reset) 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz.   
SEGM:FREQ:STAR?  
Query the start frequency of the created segment. The response is 
1.0000000000E+006.  
Note:  If the start frequency entered is greater than the current stop frequency 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP), the stop frequency is set to the start 
frequency plus the minimum frequency span of 1 Hz.  
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP <numeric_value> 
Defines the Stop frequency of sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<numeric_value> 
Stop frequency of the sweep. 
Range [def. unit] 
Depending on the instrument model [Hz]. The increment is 0.1 kHz.  
*RST value 
– (no sweep segment defined after reset) 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ   
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz.   
SEGM:FREQ:STOP?  
Query the stop frequency of the created segment. The response is 
1.5000000000E+006.  
Note:  If the stop frequency entered is smaller than the current start frequency 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt), the start frequency is set to the stop 
frequency minus the minimum frequency span 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN).   
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
Provides you with examples for adding an (empty) page to a PDF and adding empty pages to a PDF from a supported file format, with customized options.
add a page to a pdf in acrobat; add pages to pdf file
VB.NET PDF insert text library: insert text into PDF content in vb
Studio .NET PDF SDK library supports adding text content to Add text to certain position of PDF page in Visual Add text to PDF in preview without adobe reader
add a page to a pdf document; add page to pdf reader
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
435 
[SENSe<Ch>:]SEGMent<Seg>:INSert <Start>,<Stop>,<Points>,<Power>,<Point 
Delay>,<Unused>,<Meas. Bandwidth> 
Inserts a new sweep segment with specific channel settings (Define Segments - Add). The new segment 
must not overlap with any of the existing segments. 
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings 
except <LO> can be changed for existing segments using other commands of the 
[SENSe<Ch>:]SEGMent<Seg>... subsystem.   
Note:  Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use 
[SENSe<Ch>:]SEGMent<Seg>:DEFine to change or query all settings of an existing segment.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. Segment numbers must be sequential. The specified 
segment number must be smaller or equal to the number of existing segments plus 1. 
Moreover, segment numbers must selected such that the corresponding frequency 
ranges are in ascending order.  
If one or more sweep segments with segment numbers <Seg> or larger exist in the 
current channel, then all these existing segment numbers are incremented by 1 and 
the new segment is inserted as segment no. <Seg>.   
<Start>, 
<Stop> 
Start and stop frequency of the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.  
Range [def. 
unit] 
A new segment must not overlap with any of the existing segments. Besides the 
frequency range depends on the analyzer model. [Hz]  
<Points> 
Number of sweep points in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts.  
Range [def. 
unit] 
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
<Power> 
Internal source power in the segment. See [SENSe<Ch>:]SEGMent<Seg>:POWer. 
Range [def. 
unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to the 
data sheet [dBm]. UP and DOWN increment/decrement the source power in 1-dB 
steps.  
<Point 
Delay> 
Delay for each partial measurement in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting 
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELec SWTime, this parameter is 
replaced by <Time>.   
Range [def. 
unit] 
0 s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the segment, 
which is equivalent to a meas. delay of 0 s.  
<Unused> 
Ignored parameter, should be set to the default value 0.  
<Meas. 
Bandwidth> 
Resolution bandwidth in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]. 
Range [def. 
unit] 
1.0E-6 Hz to 5 MHz [Hz].  
SCPI, 
Command 
Types 
Device-specific, no query.  
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.  
C# PDF Annotate Library: Draw, edit PDF annotation, markups in C#.
notes on adobe PDF file without adobe reader installed. Provide users with examples for adding text box to users to draw various annotation markups on PDF page.
add pdf pages together; add page number to pdf in preview
VB.NET TIFF: Read, Edit & Process TIFF with VB.NET Image Document
at the page level, like TIFF page adding & deleting Viewers in VB.NET, TIFF Page Processing Within VB powerful & profession imaging controls, PDF document, image
add page number pdf; add page number to pdf preview
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
436 
SEGM2:ADD  
Create a second sweep segment. The frequency range of the second segment will be 
between 1.5 MHz and the maximum frequency of the analyzer.  
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime | DWELl 
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the command 
[SENSe<Ch>:]SEGMent<Seg>:INSert, is entered as a segment sweep time or as a meas. delay.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
SWTime 
Use segment sweep time.  
DWELl 
Use meas delay. 
*RST value 
SWTime  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:INS:SEL DWEL  
Select the meas. delay to determine the sweep time in a new sweep segment. 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz 
and a meas. delay of 10 ms..  
SEGM:SWE:TIME?  
Query the sweep time in the new segment. 
[SENSe<Ch>:]SEGMent<Seg>:OVERlap <Boolean> 
Queries whether the analyzer supports overlapping sweep segments.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON | OFF: No effect.  
*RST value 
OFF. If used as a query, the command returns the information that overlapping 
sweep segments are not supported (OFF)  
SCPI, Command 
Types 
Device-specific, command (no effect) or query 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
437 
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel] <numeric_value> 
Defines the Power of the internal signal source in sweep segment no. <Seg>. At the same time, the 
command 
activates 
separate 
power 
control 
in 
all 
sweep 
segments 
([SENSe<Ch>:]SEGMent<Seg>:POWer[:Level]:CONTrol ON).  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<numeric_value> 
Internal source power.  
Range [def. unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; 
refer to the data sheet [dBm].  
*RST value 
–10 dBm. 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings 
and thus –10 dBm internal source power.  
SEGM:POW -20 
Decrease the power to –20 dBm.  
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol <Boolean> 
Qualifies whether or not the Power of the internal signal source can be set independently for each sweep 
segment.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON: The power can be set independently for each sweep segment.  
OFF: The power in all sweep segments is equal to the internal source power for 
unsegmented sweeps set via 
SOURce<Ch>:POWer[:LEVel][:IMMediate][:AMPlitude].  
*RST value 
OFF. The parameter is automatically switched to ON when a segment level is 
entered using [SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].   
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –
10 dBm internal source power.  
SEGM:POW -20 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
438 
Decrease the power to –20 dBm.  
SEGM:POW:CONT OFF  
Couple the powers in all segments and reset the power in segment no. 1 to the initial 
value.   
[SENSe<Ch>:]SEGMent<Seg>[:STATe] <Boolean> 
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inactive segments only 
are not measured    
<Ch> 
Channel number.  
<Seg> 
Sweep segment number.  
<Boolean> 
ON | OFF - Activates or deactivates the measurement in sweep segment 
<Seg>.  
*RST value 
ON 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings. 
SEGM OFF 
Disable the measurement in the created sweep segment.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts <numeric_value> 
Defines the total number of measurement Points in sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<numeric_value> 
Number of points in the segment.  
Range [def. unit] 
1 to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
*RST value 
51 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings 
and thus 51 sweep points.  
SEGM:SWE:POIN 401 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
439 
Increase the number of points to 401.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME <numeric_value> 
Sets the duration of the sweep in sweep segment no. <Seg> (Segment Sweep Time). At the same time, 
the command activates separate sweep time setting in all sweep segments 
([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol ON).  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<numeric_value> 
Internal source power.  
Range [def. unit] 
The minimum duration depends on the other channel settings, in particular on 
the number of points ([SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts), the IF 
bandwidth (BWID) and the delay for each partial measurement 
([SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl). The maximum is 1000 s. [s] 
Changing the duration leaves the number of points unchanged but directly 
affects the delay.  
*RST value 
MIN, depending on the channel settings. This default value corresponds to 
automatic sweep time setting in manual control. 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus –10 dBm internal source power.  
SEGM:SWE:TIME 0.1 
Increase the segment sweep time to 0.1 s.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME:CONTrol <Boolean> 
Qualifies whether or not the Segment Sweep Time can be set independently for each sweep segment.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON: The time can be set independently for each sweep segment.  
OFF: The time in all sweep segments is equal to the sweep time for unsegmented 
sweeps set via [SENSe<Ch>:]SWEep:TIME.    
*RST value 
OFF. The parameter is automatically switched to ON when a segment sweep time is 
entered using [SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME or if the channel 
settings in a sweep segment require a sweep time larger than the unsegmented 
R&S
®
ZVL 
Command Reference 
[SENSe...] 
Operating Manual 1303.6580.32-01 
440 
sweep time.  
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –
10 dBm internal source power.  
SEGM:SWE:TIME 0.1  
Increase the segment sweep time to 0.1 s.  
SEGM:SWE:TIME:CONT OFF  
Couple the sweep times in all segments and reset the sweep time in segment no. 1 
to the initial value.   
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl <numeric_value> 
Defines the delay time for each partial measurement in sweep segment no. <Seg> (Meas. Delay). If 
coupling of the segments is switched on ([SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol ON) 
the delay is valid for all sweep segments in the current channel.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number.  
<numeric_value> 
Meas. delay before each partial measurement.  
Range [def. unit] 
0 s to 2.5E+003 s. [s] Changing the delay leaves the number of points 
unchanged but has an impact on the duration of the sweep 
([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME).   
*RST value 
0 s 
SCPI, Command 
Types 
Device-specific, command or query. This default value corresponds to 
automatic sweep time setting in manual control. 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings.  
SEGM:SWE:DWEL 1 MS 
Set the meas. delay in segment no. 1 to 1 ms. 
SEGM:DEF? Response: 300000,8000000000,51,-300,0.056559,0,10000,POS,NORM 
Query the channel parameters for sweep segment 1. The response value for 
the segment sweep time (olive) implicitly contains the defined meas. delay.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol <Boolean> 
Qualifies whether or not the Meas. Delay defined via [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl can 
be set independently for each sweep segment.  