c# free pdf viewer : Add a page to a pdf file application control utility azure web page winforms visual studio R&S%20ZVL-MANUAL46-part770

R&S
®
ZVL 
Command Reference 
TRACe 
Operating Manual 1303.6580.32-01 
461 
mathematical trace is active, the associated data trace is copied.   
• As a trace with a name (string variable)  
The created memory trace can be specified as follows: 
• As the memory trace named Mem<n>[Trc<m>], where n = 1, ... 8 and Trc<m> is the name of the 
copied data trace (MDATA1, MDATA2, MDATA3, MDATA4, MDATA5, MDATA6, MDATA7, 
MDATA8)  
• As a memory trace with an arbitrary name (string variable)  
An existing memory trace with the same name is overwritten.  
The copied trace is the data trace which is not modified by any mathematical operations. To 
copy a mathematical trace to a memory trace, use TRACe:COPY:MATH. To copy the active 
trace to the memory using an automatic memory trace name, use 
CALCulate<Chn>:MATH:MEMorize.  
<memory_trc> 
Name of the memory trace.   
Range [def. unit] 
<memory_trace> is either a string variable (enclosed in single or double quotes) 
or one of the following reserved names (no string variables): 
MDATA1 | MDATA2 | MDATA3 | MDATA4 | MDATA5 | MDATA6 | 
MDATA7 | MDATA8 (only for memory traces Mem<n>[Trc<m>], where n = 1, ... 
8; see list of trace names). [–] 
*RST value 
– 
<data_trc> 
Name of the data trace  
Range [def. unit] 
<data_trace> is either a string variable (enclosed in single or double quotes) or 
one of the following reserved names (no string variables): 
CH1DATA | CH2DATA | CH3DATA | CH4DATA (only for the active data 
trace in channels Ch1, Ch2, Ch3, Ch4; see list of trace names). [–] 
*RST value 
– 
SCPI, Command 
Types 
Confirmed, no query 
Example: 
*RST; SWE:POIN 20  
Create a trace with 20 sweep points, making the created trace the active trace 
of channel 1 (omitted optional mnemonic SENSe1). 
TRAC:COPY "Mem_Pt20",CH1DATA  
Copy the current state of the created trace to a memory trace named 
"Mem_Pt20". The memory trace is not displayed.  
DISP:WIND:TRAC2:FEED "MEM_PT20"  
Display the created memory trace in the active diagram area (diagram area no. 
1). 
TRACe:COPY:MATH <memory_trc>,<data_trc> 
Copies a mathematical trace to a memory trace. The trace to be copied can be specified by two 
alternative methods: 
• As the active mathematical trace of channels 1 to 4 (CH1DATA, CH2DATA, CH3DATA, 
Add a page to a pdf file - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
adding page numbers pdf file; add page to pdf preview
Add a page to a pdf file - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add a page to a pdf file; add page to a pdf
R&S
®
ZVL 
Command Reference 
TRACe 
Operating Manual 1303.6580.32-01 
462 
CH4DATA)  
• As a trace with a name (string variable)  
The created memory trace can be specified as follows: 
• As the memory trace named Mem<n>[Trc<m>], where n = 1, ... 8 and Trc<m> is the name of the 
copied data trace (MDATA1, MDATA2, MDATA3, MDATA4, MDATA5, MDATA6, MDATA7, 
MDATA8)  
• As a memory trace with an arbitrary name (string variable)  
An existing memory trace with the same name is overwritten.  
To copy a data trace which is not modified by any mathematical operations, use 
TRACe:COPY.   
<memory_trc> 
Name of the memory trace.   
Range [def. unit] 
<memory_trace> is either a string variable (enclosed in single or double quotes) 
or one of the following reserved names (no string variables): 
MDATA1 | MDATA2 | MDATA3 | MDATA4 | MDATA5 | MDATA6 | 
MDATA7 | MDATA8 (only for memory traces Mem<n>[Trc<m>], where n = 1, ... 
8; see list of trace names). [–] 
*RST value 
– 
<data_trc> 
Name of the data trace  
Range [def. unit] 
<data_trace> is either a string variable (enclosed in single or double quotes) or 
one of the following reserved names (no string variables): 
CH1DATA | CH2DATA | CH3DATA | CH4DATA (only for the active data 
trace in channels Ch1, Ch2, Ch3, Ch4; see list of trace names). [–] 
*RST value 
– 
SCPI, Command 
Types 
Confirmed, no query 
Example: 
*RST; SWE:POIN 20  
Create a trace with 20 sweep points, making the created trace the active trace 
of channel 1 (omitted optional mnemonic SENSe1). 
CALC:MATH:SDEF 'Trc1 / 2'; CALC:MATH:STAT ON  
Define a mathematical trace, dividing the data trace by 2. Activate the 
mathematical mode and display the mathematical trace instead of the data 
trace.  
TRAC:COPY:MATH 'Mem_Pt20',CH1DATA; CALC:MATH:STAT OFF  
Copy the current state of the mathematical trace to a memory trace named 
"Mem_Pt20". The memory trace is not displayed. Switch the display back to the 
data trace.  
DISP:WIND:TRAC2:FEED 'MEM_PT20'  
Display the created memory trace together with the data trace.  
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
C# File: Split PDF; C# Page: Insert PDF pages; C# Page: Delete PDF pages; C# C# Read: PDF Image Extract; C# Write: Insert text into PDF; C# Write: Add Image to
add page numbers to pdf; adding page to pdf in preview
VB.NET PDF Password Library: add, remove, edit PDF file password
On this page, we will illustrate how to protect PDF document via Add necessary references Visual Basic coding example will help you open an encrypted PDF file.
add pdf pages to word document; adding page numbers to a pdf in preview
R&S
®
ZVL 
Command Reference 
TRACe 
Operating Manual 1303.6580.32-01 
463 
TRACe[:DATA][:RESPonse][:ALL]? CH1DATA | CH2DATA | CH3DATA | CH4DATA | 
CH1MEM | CH2MEM | CH3MEM | CH4MEM | MDATA1 | MDATA2 | MDATA3 | 
MDATA4 | MDATA5 | MDATA6 | MDATA7 | MDATA8  
Returns the response values of the active data trace or memory trace (see trace names).  
To read the response values of an arbitrary data or memory trace, use 
CALCulate<Chn>:DATA? To read the response values of a trace acquired in single sweep 
mode (INITiate<Ch>:CONTinuous OFF), use CALCulate<Chn>:DATA:NSWeep?   
<Response> 
Response data of the selected trace, see list of trace names.  
Range [def. 
unit] 
The data is transferred in the data format defined via FORMat[:DATA] and 
FORMat:DEXPort:SOURce. The unit is the default unit of the measured 
parameter; see CALCulate<Ch>:PARameter:SDEFine.   
*RST value 
– 
SCPI, 
Command 
Types 
Device-specific, query only 
Example: 
SWE:POIN 20  
Create a trace with 20 sweep points, making the created trace the active trace of 
channel 1 (omitted optional mnemonic SENSe1).   
CALC:FORM MLIN; :FORM ASCII; FORM:DEXP:SOUR FDAT  
Select the trace data format: linear magnitude values, ASCII format and formatted 
trace data (1 value per sweep point).   
TRAC? CH1DATA  
Query the 20 response values of the created trace according to the previous 
format settings.    
TRACe[:DATA]:STIMulus[:ALL]? CH1DATA | CH2DATA | CH3DATA | CH4DATA | 
CH1MEM | CH2MEM | CH3MEM | CH4MEM | MDATA1 | MDATA2 | MDATA3 | 
MDATA4 | MDATA5 | MDATA6 | MDATA7 | MDATA8  
Returns the stimulus values of the active data trace or memory trace (see trace names).  
To read the stimulus values of an arbitrary data or memory trace, use 
CALCulate<Chn>:DATA:STIMulus?  
<Response> 
Response data of the selected trace, see list of trace names.  
Range [def. unit] 
The data is transferred in the data format defined via FORMat[:DATA]. The 
numeric values are expressed in the default unit of the stimulus variable.  [Hz, 
dBm or s, depending on the sweep type]  
*RST value 
– 
SCPI, Command 
Types 
Device-specific, query only 
Example: 
SWE:POIN 20  
C# PDF Password Library: add, remove, edit PDF file password in C#
your PDF document in C# project, XDoc.PDF provides some PDF security settings. On this page, we will talk about how to achieve this via Add necessary references
add page number to pdf hyperlink; add page to pdf acrobat
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
there is a newly created page, you may need to add it into Files, adding a page into PDF document, deleting unnecessary page from PDF file and changing
adding page numbers pdf; add page numbers to a pdf in preview
R&S
®
ZVL 
Command Reference 
TRACe 
Operating Manual 1303.6580.32-01 
464 
Create a trace with 20 sweep points, making the created trace the active trace of 
channel 1 (omitted optional mnemonic SENSe1).   
TRAC:STIM? CH1DATA  
Query the 20 stimulus values of the created trace. In the default format setting, 
the data is returned as a comma-separated list of 10-digit ASCII values.   
C# PDF insert image Library: insert images into PDF in C#.net, ASP
using RasterEdge.Imaging.Basic; using RasterEdge.XDoc.PDF; Have a try with this sample C#.NET code to add an image to the first page of PDF file.
add a page to a pdf in acrobat; adding a page to a pdf
VB.NET PDF insert image library: insert images into PDF in vb.net
using RasterEdge.XDoc.PDF; Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
adding page numbers to pdf in; add page number to pdf online
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
465 
8 Programming Examples.....................................................................................466 
Basic Tasks......................................................................................................................................466
Typical Stages of a Remote Control Program................................................................................466
Basic Instrument Settings.............................................................................................................466
Adjusting the Test Setup..............................................................................................................467
Start of the Measurement and Command Synchronization.........................................................467
Retrieving Measurement Results.................................................................................................468
Handling of Channels, Traces and Diagram Areas.........................................................................469
Several Traces with Equal Channel Settings...............................................................................469
Several Traces with Different Channel Settings...........................................................................474
Markers and Limit Lines...............................................................................................................478
Condensed Programming Examples.............................................................................................483
Calibration.......................................................................................................................................483
Modeling a Max Hold Function........................................................................................................485
Retrieving the Results of Previous Sweeps....................................................................................486
C# PDF File Split Library: Split, seperate PDF into multiple files
If your page number is set as 1, then the two output PDF files will contains the first page and the later C# DLLs: Split PDF Document. Add necessary references:
adding page numbers in pdf; add page to existing pdf file
VB.NET PDF File Split Library: Split, seperate PDF into multiple
Length outputFilePaths.Add(Program.RootPath + "\\" outputFileName + "_" + i.ToString() + ".pdf") Next ' Split input PDF file to 4 files: ' File 0: page 0
add remove pages from pdf; add pages to pdf file
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
466 
8 Programming Examples 
This chapter contains detailed program examples.  
The syntax and use of all SCPI commands is described in the SCPI Reference chapter. For a general 
introduction to remote control of the analyzer refer to chapter Remote Control. For an overview of special 
remote control features of the network analyzers refer to section NWA Remote Control Features in 
Chapter 5.  
Basic Tasks 
Typical Stages of a Remote Control Program 
A typical remote control program comprises the following stages:   
Very often, steps 3 and 4 (or steps 2 to 4) must be repeated several times.    
All example programs in this section have been developed and tested by means of the GPIB 
Explorer provided with the network analyzer. No extra programming environment is needed.  
Basic Instrument Settings  
Programming task: Adjust the basic network analyzer settings to your measurement tasks, optimizing 
the instrument for fast measurements.   
Considerations for high measurement speed  
The measurement speed depends on the sweep time but also on an efficient preparation of the instrument 
and on proper command synchronization. The following items should be kept in mind:  
! For maximum speed the basic channel settings should be set while the sweep is stopped and with 
a minimum of sweep points. It is advisable to increase the number of points after all instrument 
settings have been performed, and to initiate the sweep after the test setup has been completed.  
! Execution of the INITiate[:IMMediate] command is fastest in synchronized mode. Insertion 
of fixed waiting periods into the command sequence is possible but generally less efficient.  
! The sweep time depends on several parameters discussed in section Optimizing the 
Measurement Speed. In particular it is recommended to select the best set of sweep points, e.g. 
using the segmented sweep.  
// Reset the instrument, switch off the measurement (after one sweep),  
// reduce the number of sweep points. 
*RST 
INITiate1:CONTinuous OFF  
SENSe1:SWEep:POINts 2  
// 
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
467 
// Avoid a delay time between different partial measurements and before the start of the sweeps (is default 
setting). 
SENSe1:SWEep:TIME:AUTO ON  
TRIGger1:SEQuence:SOURce IMMediate  
// 
// Select the widest bandwidth compatible with your measurement. 
SENSe1:BANDwidth:RESolution 10   
// 
// Adjust your sweep points to your measurement task, e.g. using a segmented sweep. 
SENSe1:SEGMent...   
Adjusting the Test Setup 
In general the preparatives described above can be used for a series of measurements. In-between the 
measurements it is often necessary to change the test setup, e.g. in order to replace the DUT, change the 
connected ports, connect external devices etc.   
Start of the Measurement and Command Synchronization 
Programming task: Start a measurement in single sweep mode. Wait until all single sweep data has 
been acquired before you proceed to the next stage of the measurement.   
INITiate<Ch>[:IMMediate] is used to start a single sweep or a group of single sweeps. This 
command has been implemented for overlapped execution. The advantage of an overlapped 
command is that they allow the program to do other tasks while being executed.  
In the present example the sweep must be completed before measurement results can be 
retrieved. To prevent wrong results (e.g. a mix-up of results from consecutive sweeps) the 
controller must synchronize its operation to the execution of INITiate<Ch>[:IMMediate]. 
IEEE 488.2 defines three common commands (*WAI, *OPC?, *OPC) for synchronization.  
// 
// 1. Start single sweep, use *WAI  
// *WAI is the easiest method of synchronization. It has no effect when sent after sequential commands.  
// If *WAI follows INITiate<Ch>[:IMMediate] (overlapped command),  
// the analyzer executes no further commands or queries until the sweep is terminated.  
// *WAI does prevent the controller from sending other commands to the analyzer or other devices  
// on the GPIB bus  
INITiate1:SCOPe SINGle  //  Single sweep will be started in the referenced channel only  
INITiate1:IMMediate; *WAI  //  Start single sweep in channel no. 1, wait until the end of the 
sweep  
<Continue program sequence>  
// 
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
468 
// 2. Start single sweep, use *OPC?  
// If *OPC follows INITiate<Ch>[:IMMediate], it places a 1 into the output queue when the sweep is 
terminated.  
// An appropriate condition in the remote control program must cause the controller to wait until *OPC? 
returns one.  
// The controller is stopped from the moment when the condition is set.  
INITiate1:IMMediate; *OPC?  //  Start single sweep in channel no. 1, indicate the end of the 
sweep by a 1 in the output queue.  
// So far the controller may still send messages to other devices on the GPIB bus.  
<Condition OPC=1>  //   Stop the controller until *OPC? returns one (program syntax depends 
on your programming environment).  
<Continue program sequence>  
// 
// 3. Start single sweep, use *OPC  
// If *OPC follows INITiate<Ch>[:IMMediate], it sets the OPC bit in the ESR after the sweep is 
terminated.   
// This event can be polled or used to trigger a service request of the analyzer.  
// The advantage of *OPC synchronization is that both the controller and the analyzer can continue  
// processing commands while the sweep is in progress.  
*SRE 32  / Enable a service request for the ESR  
*ESE 1  / Set event enable bit for operation complete bit  
INITiate1::IMMediate; *OPC  //  Start single sweep in channel no. 1, set the OPC bit in the ESR 
after the sweep is terminated.  
// The controller may still send messages, the analyzer continues to parse and execute commands.  
<Wait for service request>  //   Controller waits for service request from the analyzer (program 
syntax depends on your programming environment).  
<Continue program sequence>  
Retrieving Measurement Results 
Programming task: Read the results acquired in a single sweep.  
// 
// 1. Read single values (-> Markers)   
// Markers are the most convenient tool for determining and retrieving single values on traces.  
// The analyzer provides up to ten markers; see Markers and Limit Lines.  
// 
// 2. Read complete trace   
// Select a trace format and read formatted trace data.  
CALCulate1:FORMat MLINear  / Calculate the linear magnitude of z  
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
469 
CALCulate1:DATA? FDATa  / Read the formatted trace data  
//  
Use CALCulate<Chn>:DATA:NSWeep to retrieve a particular trace within a group of 
sweeps.  
Handling of Channels, Traces and Diagram Areas     
The following examples show you how to perform basic tasks related to channel and trace definition and 
to the display of traces in diagram areas.   
All example programs in this section have been developed and tested by means of the GPIB 
Explorer provided with the network analyzer. No extra programming environment is needed.  
Several Traces with Equal Channel Settings 
Programming task: Create up to four different traces with equal channel settings, assign the four 2-port 
standard S-parameters to the traces and display them in up to four diagram areas.   
Important remote control features for this program example  
The following command sequence illustrates the structure of the remote commands discussed in section 
Basic Remote Control Concepts. In particular it shows that:  
! A trace can be created and handled without being displayed.  
! Traces are referenced by trace names. The active trace of a channel is often referenced by the 
R&S
®
ZVL 
Programming Examples 
Basic Tasks 
Operating Manual 1303.6580.32-01 
470 
channel suffix.  
! Diagram areas are referenced by a window suffix <Wnd>. An additional suffix <WndTr> in the 
DISPlay:WINDow<Wnd>:TRACe<WndTr>... commands numbers the different traces in a 
diagram area.  
! In remote control, it is possible to display the same trace in several diagram areas.  
! The analyzer provides several commands allowing a smooth transition between remote and 
manual control.  
// 
// 1. One channel, two traces, one diagram area  
// Reset the instrument, creating the default trace Trc1 in channel 1.  
// The default measured quantity is the forward transmission S-parameter S21.  
// The default format is dB Mag. 
*RST 
// 
// 
Create 
second 
trace 
in 
channel 
1, 
assign 
the 
format 
Phase,  
// and display the new trace in the same diagram area.  
CALCulate1:PARameter:SDEFine 'Trc2', 'S21'  // the trace becomes the active trace but is not 
displayed  
CALCulate1:FORMat PHASe  // the trace is referenced by the channel suffix 1  
DISPlay:WINDow1:TRACe2:FEED 'Trc2'  // display the second trace, numbering it the second 
trace in diagram area no. 1   
// 
// 
Check the result on the local screen  
// Go to local  
SYSTem:DISPlay:UPDate ONCE 
Documents you may be interested
Documents you may be interested