HW Interfaces 
Front Panel Connectors 
Operating Manual 1303.6580.32-01 
491 
9 HW Interfaces 
This chapter provides a detailed description of the hardware interfaces and connectors of the instrument. 
For a graphical overview of the front panel and rear panel connectors and their use refer to chapter 
Preparing for Use. 
Front Panel Connectors 
Test Ports 
N-connectors labelled PORT 1 and PORT 2/ RF INPUT. The test ports serve as outputs for the RF 
stimulus signal and as inputs for the measured RF signals from the DUT (response signals).   
• With a single test port, it is possible to generate a stimulus signal and measure the response 
signal in reflection.   
• With 2 test ports, it is possible to perform full two-port measurements; see S-Parameters section 
in Chapter 3.  
• The two network analyzer ports are equivalent. If the Spectrum Analysis option (R&S ZVL-K1) is 
active, test port PORT 2 serves as an AC-coupled input for the analyzed RF signal; PORT 1 is not 
used.  
Attention! 
The maximum input levels at all test ports according to the front 
panel labeling or the data sheet must not be exceeded.   
In addition, the maximum input voltages of the other input 
connectors at the front and rear panel must not be exceeded.   
USB Connectors 
Two single Universal Serial Bus connectors of type A (master USB), used to connect a keyboard 
(recommended: PSL–Z2, order number 1157.6870.03), mouse (recommended: PSL–Z10, order number 
1157.7060.03) or other pointing devices, a printer or an external storage device (USB stick, CD-ROM 
drive etc.).  
Add pages to pdf preview - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
adding pages to a pdf; add page numbers to a pdf document
Add pages to pdf preview - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
add page to pdf; add pages to pdf
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
492 
Using an adapter cable (R&S NRP–Z4), a power sensor can be connected, as an alternative to the power 
sensor connector on the rear panel that is only available with option Additional Interfaces, R&S FSL–B5.  
For maintaining the EMI conformity of the R&S ZVL only appropriate USB accessories may 
be used.  
Passive connecting USB cables should not exceed 4 m in length. Use the original USB 
connecting cable or another high–quality cable. The maximum current per USB port is 500 
mA.  
PROBE POWER 
Connector for supply voltages of +15 V to –12 V and ground for active probes and preamplifiers. A 
maximum current of 140 mA is available. This connector is suitable as power supply for high-impedance 
probes from Agilent.  
Rear Panel Connectors 
LAN 
8-pin connector RJ-45, used to connect the analyzer to a Local Area Network (LAN); see Remote Control 
in a LAN. The pin assignment of the RF-45 connector supports category 5 UTP/STP (Unshielded/Shielded 
Twisted Pair) cables. 
EXT TRIGGER 
BNC female connector, used as an input for external TTL trigger signals. Input levels <0.7 V are 
interpreted as logic low; input levels >1.4 V as logic high. The typical input impedance is 10 kΩ.  
C# WinForms Viewer: Load, View, Convert, Annotate and Edit PDF
Add text to PDF document in preview. • Add text box to PDF file in preview. Erase PDF pages. Miscellaneous. Search PDF text in preview.
add page to pdf reader; add page break to pdf
C# WPF Viewer: Load, View, Convert, Annotate and Edit PDF
PDF Protection. • Add signatures to PDF document. Erase PDF pages. Miscellaneous. • Select PDF text on viewer. • Search PDF text in preview.
add a page to a pdf online; add page number to pdf preview
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
493 
EXT REF 
BNC female connector, used as an input or output for the 10 MHz reference clock signal.   
The function of the 10 MHz REF connector depends on the Int. Reference or Ext. Reference setting in the 
SETUP menu:  
• If Int. Reference is active, 10 MHz REF is used as an output connector for the 10 MHz internal 
reference clock signal of the analyzer. This configuration requires the OCXO option (R&S FSL–
B4, OCXO Reference Frequency).  
• If Ext. Reference is active, 10 MHz REF is used as an input connector for an external 10 MHz 
reference clock signal. The external reference signal must meet the specifications of the data 
sheet. The internal reference signal is synchronized to the external signal.  
The AC power switch also interrupts the power supply of the OCXO. When you switch the 
instrument back on, be sure to comply with the extended warm–up phase specified in the 
data sheet. 
POWER SENSOR (Option R&S FSL–B5, Additional Interfaces) 
LEMOSA female connector, used for connecting power sensors of the R&S NRP–Zxy family. With an 
adapter cable R&S NRP–Z4, the USB port on the front panel can be used for this purpose.  
NOISE SOURCE CONTROL (Option R&S FSL–B5, Additional 
Interfaces) 
BNC female connector, provides the supply voltage for an external noise source, e.g. to measure the 
noise figure and gain of amplifiers and frequency converting DUTs.  
Conventional noise sources require a voltage of +28 V in order to be switched on and 0 V to be switched 
off. The output supports a maximum load of 100 mA. 
How to C#: Preview Document Content Using XDoc.Word
With the SDK, you can preview the document content according to the preview thumbnail by the ways as following. C# DLLs for Word File Preview. Add references:
add pdf pages together; add page numbers to pdf in reader
How to C#: Preview Document Content Using XDoc.PowerPoint
C# DLLs: Preview PowerPoint Document. Add necessary XDoc.PowerPoint DLL libraries into your created C# application as references. RasterEdge.Imaging.Basic.dll.
adding a page to a pdf file; add a page to a pdf
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
494 
IF/VIDEO OUT (Option R&S FSL–B5, Additional Interfaces) 
BNC female connector, used as an intermediate frequency (IF) output of approximately 20 MHz or as a 
video output at the selected video and resolution bandwidth (with spectrum analyzer option R&S ZVL-K1).  
This connector cannot be used simultaneously with the AF output connector on the front 
panel unless it is configured as a video output.  
AUX PORT (Option R&S FSL–B5, Additional Interfaces) 
9-pole SUB–D male connector, provides output signals for controlling external devices. The voltage levels 
are of TTL type (max. 5 V). The pin assignment is shown below.  
Pin No. 
Name 
Input (I) or  
Output (O) 
or  
Bidirectional 
(B) 
Voltage Range 
Function 
SUPPLY 
+ 5 V / max. 250 mA 
Supply voltage for 
external circuits 
2 to 7 
I/O 
– 
Reserved for future use 
GND 
– 
Ground 
READY FOR 
TRIGGER 
0 V 
Signal indicating that 
the instrument is ready 
to receive trigger signal. 
Attention! 
Watch the pin assignment carefully. A short-circuit may damage the 
instrument.  
C# PDF insert image Library: insert images into PDF in C#.net, ASP
viewer component supports inserting image to PDF in preview without adobe Insert images into PDF form field. How to insert and add image, picture, digital photo
add and delete pages from pdf; adding a page to a pdf in preview
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Remove bookmarks, annotations, watermark, page labels and article threads from PDF while compressing. Also a preview component enables compressing and
add page number to pdf document; add and delete pages in pdf
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
495 
GPIB Interface (Option R&S FSL–B10) 
GPIB interface in compliance with IEEE488 and SCPI. A computer for remote control can be connected 
via this interface. To set up the connection, a shielded cable is recommended. See GPIB Bus Interface 
above in this chapter. 
DC Power Supply (Option R&S FSL-B30) 
A DC power supply can be connected alternatively to the AC power supply; see DC Power Supply and 
Battery in chapter 1. The connector is supplied with the other accessories of the option and is connected 
according to the following diagram: 
Pin No. 
Input (I) or  
Output (O) 
or  
Bidirectional 
(B) 
Voltage Range 
Function 
+11 V to +28 V 
7 A to 2.7 A 
Positive voltage 
– 
– 
Ground 
– 
– 
Not used 
Caution!  
DC 
power 
supply 
The power supply (SELV) that is used must fulfill the requirements 
for reinforced/double insulation for main supply circuits in 
accordance to DIN/EN/IEC 61010 (UL 61010B–1, CSA C22.2 No. 
1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). It is 
recommended to fuse the DC power supply according to the table 
below. Before switching on the instrument check the connection for 
correct polarity.  
VB.NET PDF insert image library: insert images into PDF in vb.net
inserting image to PDF in preview without adobe provide users the most individualized PDF page image inserting function, allowing developers to add and insert
adding a page to a pdf in reader; adding page numbers to a pdf in reader
How to C#: Preview Document Content Using XDoc.excel
following. C# DLLs: Preview Excel Document without Microsoft Office Installed. Add necessary references: RasterEdge.Imaging.Basic.dll.
add page numbers to pdf reader; add page numbers to pdf document in preview
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
496 
In continuous operation the actual breaking current can differ from the rated breaking current. For fuse 
selection take the characteristics of the fuse into account. 
Input Voltage 
Max. Current or Power 
11 V to 12.5 V 
125 VA 
12.5 V to 18.7 V 
10 A 
18.7 V to 28 V 
200 VA 
The instrument is switched on or off using the standby key on the front panel; see sections Power Supply 
Options and DC Power Supply and Battery in chapter 1.  
Battery Pack (Option R&S FSL-B31) 
The battery pack can be used alternatively as power supply. If the battery runs low during operation, a 
message is displayed. In this case, use another power supply or switch off the instrument. For an 
overview on available power supplies refer to section Power Supply Options in chapter 1.The power 
supply can be changed during operation.  
The instrument is switched on or off using the ON/STANDBY function key on the front panel. For details 
refer to section DC Power Supply and Battery in chapter 1. 
The battery pack can be reloaded via the AC or DC power supply. For details on charging refer to section 
Charging the Battery in chapter 1.  
If the battery is not to be used for a longer time, it is recommended to remove it and store it 
separately. 
IEC/IEEE Bus Interface  
The standard instrument is equipped with a GPIB bus (IEC/IEEE bus) connection. The two interface 
connectors labeled IEC BUS and IEC SYSTEM BUS are located on the rear panel of the instrument.  
• The IEC BUS connector is intended for remote control of the analyzer from a controller.  
• The IEC SYSTEM BUS can be used to control further devices from the analyzer.  
Always use a shielded cable to connect the GPIB bus interfaces.  
Characteristics of the interface 
• 8-it parallel data transfer 
• Bidirectional data transfer 
• Three-line handshake 
• High data transfer rate of max. 1 MByte/s 
• Up to 15 devices can be connected 
• Maximum length of the connecting cables 15 m. The length of a 
single connecting cable should not exceed 2 m, if many devices are 
used, it should not exceed 1 m. 
C# HTML5 Viewer: Load, View, Convert, Annotate and Edit Raster
zoom in or zoom out images while in preview. to Tiff (.tif, .tiff) online, create PDF document from Supported annotation features includes add text comments to
add page numbers to pdf preview; add pages to pdf reader
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
497 
• Wired OR if several instruments are connected in parallel 
Pin assignment 
Bus lines  
1. Data bus with 8 lines D0 to D7
The transmission is bit-parallel and 
byte-serial in the ASCII/ISO code. D0 is the least significant bit, D7 
the most significant bit.  
2. Control bus with five lines  
IFC (Interface Clear):
active LOW resets the interfaces of the 
instruments connected to the default setting. 
ATN (Attention):
active LOW signals the transmission of interface 
messages, inactive HIGH signals the transmission of device 
messages. 
SRQ (Service Request):
active LOW enables the connected device to 
send a service request to the controller. 
REN (Remote Enable):
active LOW permits switchover to remote 
control. 
EOI (End or Identify):
has two functions in connection with ATN: 
ATN=HIGH active LOW marks the end of data transmission. 
ATN=LOW active LOW triggers a parallel poll. 
3.Handshake bus with three lines 
DAV (Data Valid):
active LOW signals a valid data byte on the data 
bus. 
NRFD (Not Ready For Data):
active LOW signals that one of the 
connected devices is not ready for data transfer. 
NDAC (Not Data Accepted):
active LOW signals that the instrument 
connected is accepting the data on the data bus. 
The analyzer provides the following functions to communicate via GPIB bus: 
• Interface functions 
• Interface messages 
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
498 
• Instrument messages 
Interface Functions 
Instruments which can be controlled via GPIB bus can be equipped with different interface functions. The 
interface function for the network analyzer are listed in the following table.  
Control character 
Interface function 
SH1 
Handshake source function (source handshake), full capability 
AH1 
Handshake sink function (acceptor handshake), full capability 
L4 
Listener function, full capability, de-addressed by MTA. 
T6 
Talker function, full capability, ability to respond to serial poll, deaddressed by MLA 
SR1 
Service request function (Service Request), full capability 
PP1 
Parallel poll function, full capability 
RL1 
Remote/Local switch over function, full capability 
DC1 
Reset function (Device Clear), full capability 
DT1 
Trigger function (Device Trigger), full capability 
Interface Messages 
Interface messages are transmitted to the instrument on the data lines, with the attention line being active 
(LOW). They serve to communicate between controller and instrument. 
Universal Commands 
Universal commands are encoded in the range 10 through 1F hex. They are effective for all instruments 
connected to the bus without pervious addressing. 
Command 
QuickBASIC 
command 
Effect on the instrument 
DCL 
(Device Clear) 
IBCMD (controller%, 
CHR$(20)) 
Aborts the processing of the commands just received and sets the 
command processing software to a defined initial state. Does not 
change the instrument settings. 
IFC 
(Interface Clear) 
IBSIC (controller%) 
Resets the interfaces to the default setting. 
LLO 
(Local Lockout) 
IBCMD (controller%, 
CHR$(17)) 
The LOC/IEC ADDR key is disabled. 
SPE 
(Serial Poll 
Enable) 
IBCMD (controller%, 
CHR$(24)) 
Ready for serial poll. 
SPD 
(Serial Poll 
Disable) 
IBCMD (controller%, 
CHR$(25)) 
End of serial poll. 
PPU  
(Parallel Poll 
Unconfigure) 
IBCMD 
(controller%, 
CHR$(21)) 
End of the parallel-poll state.  
Addressed Commands 
Addressed commands are encoded in the range 00 through 0F hex. They are only effective for 
instruments addressed as listeners. 
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
499 
Command 
QuickBASIC 
command 
Effect on the instrument 
GET 
(Group 
Execute 
Trigger) 
IBTRG (device%) 
Triggers a previously active device function (e.g. a sweep). The effect 
of the command is the same as with that of a pulse at the external 
trigger signal input. 
GTL 
(Go to Local) 
IBLOC (device%) 
Transition to the "Local" state (manual control). 
PPC 
(Parallel Poll 
Configure) 
IBPPC (device%, 
data%) 
Configures the instrument for parallel poll. Additionally, the QuickBASIC 
command executes PPE/PPD. 
SDC 
(Selected 
Device Clear) 
IBCLR (device%) 
Aborts the processing of the commands just received and sets the 
command processing software to a defined initial state. Does not 
change the instrument setting. 
Instrument Messages 
Instrument messages (commands) are transferred on the data lines of the GPIB bus while the ATN line is 
not active. ASCII code is used. 
Structure and syntax of the instrument messages are described the SCPI Reference chapter. The chapter 
also provides a detailed description of all messages implemented by the analyzers.   
VXI-11 Interface 
The VXI–11 standard is based on the RPC protocol which in turn relies on TCP/IP as the 
network/transport layer. The TCP/IP network protocol and the associated network services are pre-
configured. TCP/IP ensures connection–oriented communication, where the order of the exchanged 
messages is adhered to and interrupted links are identified. With this protocol, messages cannot be lost.  
Remote control of an instrument via a network is based on standardized protocols which follow the OSI 
reference model (see Fig. below).  
Based on TCP/UDP, messages between the controller and the instrument are exchanged via open 
network computing (ONC) – remote procedure calls (RPC). With XDR (VXI–11), legal RPC messages are 
known as VXI–11 standard. Based on this standard, messages are exchanged between the controller and 
the instrument. The messages are identical with SCPI commands. They can be organized in four groups: 
R&S
®
ZVL 
HW Interfaces 
Rear Panel Connectors 
Operating Manual 1303.6580.32-01 
500 
• Program messages (control command to the instrument) 
• Response messages (values returned by the instrument) 
• Service request (spontaneous queries of the instrument) 
• Low–level control messages (interface messages)  
A VXI–11 link between a controller and an instrument uses three channels: core, abort and interrupt 
channel. Instrument control is mainly performed on the core channel (program, response and low–level 
control messages). The abort channel is used for immediate abort of the core channel; the interrupt 
channel transmits spontaneous service requests of the instrument. Link setup itself is very complex. For 
more details refer to the VXI–11 specification. 
The number of controllers that can address an instrument is practically unlimited in the network. In the 
instrument, the individual controllers are clearly distinguished. This distinction continues up to the 
application level in the controller, i.e. two applications on a computer are identified by the instrument as 
two different controllers.  
The controllers can lock and unlock the instrument for exclusive access. This regulates access to the 
instrument of several controllers.  
VXI-11 Interface Messages 
On the Ethernet link, the interface messages are called low–level control messages. These messages can 
be used to emulate interface messages of the IEC/IEEE bus.  
Command 
Effect on the instrument 
&ABO 
(Abort) 
Aborts processing of the commands just received. 
&DCL 
(Device Clear) 
Aborts processing of the commands just received and sets the 
command processing software to a defined initial state. Does not 
change the instrument setting. 