R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
71 
Mkr 1 indicates the maximum (minimum) of the peak.  
Mkr 2 and Mkr 3 indicate the lower and upper band edge where the trace value has 
decreased (increased) by a definite Level value.  
Mkr 4 indicates the center of the peak, calculated as the arithmetic mean value of 
the LBE and UBE positions 
3.1.5.6 Marker Info Field 
The coordinates of all markers defined in a diagram area are displayed in the info field, 
which by default is located in the upper right corner.  
The list contains the following information:  
Mkr 1, Mkr2, ... denote the marker numbers.  
Markers are displayed with the same color as the associated trace.  
The marker coordinates are expressed in one of the marker formats selected via 
Marker Format. The formats of the markers assigned to a trace are independent of 
each other and of the trace format settings.  
The active marker has a dot placed in front of the marker line.  
A ∆ sign placed in front of the marker line indicates that the marker is in Delta 
Mode. 
Customizing the marker info field 
To change the position, appearance or contents of the marker info field use one of the 
following methods:  
Double-click the info field to open the Marker Properties dialog with extended 
settings for all markers of the active trace. 
Right-click the info field to open a context menu providing frequently used marker 
settings.  
Adding page numbers to a pdf document - insert pages into PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Guide C# Users to Insert (Empty) PDF Page or Pages from a Supported File Format
add pages to pdf; adding page numbers to a pdf in preview
Adding page numbers to a pdf document - VB.NET PDF Page Insert Library: insert pages into PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Easy to Use VB.NET APIs to Add a New Blank Page to PDF Document
adding a page to a pdf; add pages to pdf document
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
72 
To change the position of the marker info field, select Movable Marker Info from the 
context menu. Drag-and-drop the info field to any position in the active diagram 
area. 
To change the format of the active marker, select Mkr Format.  
To express the coordinates of the active marker relative to the reference marker, 
activate the Delta Mode.  
For more information: Show Info Table 
In addition to the marker info field, the analyzer provides an info table with extended 
marker information.  
The table is hidden by default. To display the table double-click the marker info field to 
open the Marker Properties dialog.  
Context menu of the marker info field 
A right mouse click on the marker info field opens a context menu:  
Movable Marker Info allows the marker info field to be placed to any position in the 
diagram area. The remaining settings correspond to the most common commands in 
the Trace – Marker and Trace – Search menus.  
3.1.5.7 Channel Settings     
The main properties of all channels assigned to the traces in the diagram area are 
displayed in the channel list below the diagram.   
Each line in the channel list describes a single channel. The channel of the active trace 
is highlighted. The lines are divided into several sections with the following contents 
(from left to right):  
Channel name indicates the current channel name. The default names for new 
C# Create PDF Library SDK to convert PDF from other file formats
PDF document file but also offer them the ability to count the page numbers of generated Besides, using this PDF document metadata adding control, you
adding page numbers to pdf in preview; add page numbers pdf files
C# Word - Word Create or Build in C#.NET
but also offer them the ability to count the page numbers of generated Besides, using this Word document adding control, you can add some Create Word From PDF.
add and remove pages from a pdf; add pages to pdf without acrobat
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
73 
channels are Ch<n> where <n> is a current number. Right-click the section and 
call the Channel Manager from the context menu to change the channel name.  
Start value of the sweep indicates the lowest value of the sweep variable (e.g. the 
lowest frequency measured), corresponding to the left edge of the Cartesian 
diagram.  
Color legend shows the display color of all traces assigned to the channel. The 
colors are different, so the number of colors is equal to the numbers of traces 
assigned to the channel.  
Additional stimulus parameter shows either the power of the internal signal source 
(for frequency sweeps and time sweeps) or the CW frequency (for power sweeps).  
Stop value of the sweep indicates the highest value of the sweep variable (e.g. the 
highest frequency measured), corresponding to the right edge of the Cartesian 
diagram. Right-click any of the sections in the trace list (except Color legend) to 
open a context menu and access the most common tasks related to the section. 
Context menus of the channel list 
A right mouse click on the channel name, the sweep range, and the additional 
parameter section of the channel list opens the following context menus, respectively:  
The settings correspond to the most common commands in the Channel  Channel 
Select, Channel – Center, Span and Channel – Pwr Bw menus.  
3.1.5.8 Context Menus 
To provide access to the most common tasks and speed up the operation, the analyzer 
offers context menus (right-click menus) for the following display elements:  
Diagram area  
Marker info field  
Trace list (separate context menus for trace name section, measured quantity 
section, format section, scale section, and channel section)  
Channel list (separate context menus for channel name section, sweep range 
section, additional parameter section)  
Working with context menus requires a mouse. Click inside the display element that 
you want to work with using the right mouse button.  
Except from some particular screen configurations, anything you can do from a context 
menu you can also do from the menu bar using front panel keys and softkeys. Use 
whatever method is most convenient for you.  
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
74 
3.1.6 Dialogs     
Dialogs provide groups of related settings and allow to make selections and enter data 
in an organized way. All softkeys with three dots behind their labeling (as in Mkr 
Properties...) call up a dialog. The dialogs of the analyzer have an analogous structure 
and a number of common control elements.  
TIP 
Dialog Transparency 
The Dialog Transparency function (Nwa-Setup – System Config menu) varies the 
transparency of all dialogs. With an appropriate setting, you can control the dialogs 
and at the same time view the underlying traces and display elements. 
We assume that you are familiar with standard Windows dialogs and mouse operation. 
Refer to Using Front Panel Keys to learn how to control dialogs without a mouse and 
keyboard.  
3.1.6.1 Immediate vs. Confirmed Settings 
In some dialogs, the settings take effect immediately so that the effect on the 
measurement is observable while the dialog is still open. This is especially convenient 
when a numeric value is incremented or decremented, e.g. via the rotary knob.  
In most dialogs, however, it is possible to cancel an erroneous input before it takes 
effect. The settings in such dialogs must be confirmed explicitly.  
The two types of dialogs are easy to distinguish:  
Dialogs with immediate settings provide a Close button but no OK button. Example: 
Step Size dialog.  
Dialogs with confirmed settings provide both an OK button and a Cancel button. 
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
75 
Example: On-screen keyboard.  
You can also cancel an immediate setting using Setup – Undo
3.1.6.2 On-Screen Keyboard 
A keyboard 
symbol next to a character input field opens the analyzer's on-screen 
keyboard.  
The on-screen keyboard contains two sets of characters plus the following additional 
controls: 
Shift changes between the two character sets containing lower case 
letters/numbers and upper case letters/special characters, respectively. 
<= BS deletes the current string in the alphanumeric input field.  
OK applies the current selection and closes the keyboard. The current string is 
written into the input field of the calling dialog. See also Immediate vs. Confirmed 
Settings.  
Cancel discards the current selection and closes the keyboard. The input field of 
the calling dialog is left unchanged.   
The on-screen keyboard allows you to enter characters, in particular letters, without an 
external keyboard; see Data Entry. To enter numbers and units, you can also use the 
DATA ENTRY keys on the front panel of the instrument. 
3.1.6.3 Paste Marker List 
A pull-down list symbol next to a numeric input field opens a list of all current stimulus 
and response marker values of the active trace. Any of the marker values can be 
selected as a numeric entry. If the physical unit of the selected marker value is 
inconsistent (mismatch of stimulus and response values) then the numeric value is 
used without the unit.  
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
76 
The response values in the paste marker list are not updated as the analyzer continues 
measuring, so they may differ from the values in the marker info field.  
To open the paste marker list you can also click on the input field and use the space 
bar of your keyboard or the checkmark key in the navigation area at the front panel of 
the analyzer.   
3.1.6.4 Numeric Entry Bar 
Single numeric values can be entered using the input field of the numeric entry bar. 
The numeric entry bar appears just below the menu bar as soon as a function implying 
a single numeric entry is activated. In contrast to dialogs, it does not hide any of the 
display elements in the diagram area.  
The numeric entry bar contains the name of the calling function, a numeric input field 
including the Cursor Up/Down 
buttons for data variation and a Close button. Besides 
it is closed automatically as soon as an active display element in the diagram area is 
clicked or a new menu command is activated.  
3.1.7 Display Formats and Diagram Types     
A display format defines how the set of (complex) measurement points is converted 
and displayed in a diagram. The display formats in the Trace – Format menu use the 
following basic diagram types:   
Cartesian (rectangular) diagrams are used for all display formats involving a 
conversion of the measurement data into a real (scalar) quantity, i.e. for dB Mag, 
Phase, Group Delay, SWR, Lin Mag, Real, Imag and Unwrapped Phase.  
Polar diagrams are used for the display format Polar and show a complex quantity 
as a vector in a single trace.  
Smith charts are used for the display format Smith and show vector like polar 
diagrams but with grid lines of constant real and imaginary part of the impedance.   
Inverted Smith charts are used for the display format Inverted Smith and show 
vector like polar diagrams but with grid lines of constant real and imaginary part of 
the admittance.   
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
77 
NOTE 
Trace formats and measured quantities 
The analyzer allows arbitrary combinations of display formats and measured 
quantities (Trace – Meas). Nevertheless, in order to extract useful information from 
the data, it is important to select a display format which is appropriate to the analysis 
of a particular measured quantity; see Measured Quantities and Display Formats.  
3.1.7.1 Cartesian Diagrams 
Cartesian diagrams are rectangular diagrams used to display a scalar quantity as a 
function of the stimulus variable (frequency).  
The stimulus variable appears on the horizontal axis (x-axis), scaled linearly 
(sweep type Lin Frequency) or logarithmically (sweep type Log Frequency).  
The measured data (response values) appears on the vertical axis (y-axis). The 
scale of the y-axis is linear with equidistant grid lines although the y-axis values 
may be obtained from the measured data by non-linear conversions 
The following examples show the same trace in Cartesian diagrams with linear and 
logarithmic x-axis scaling.  
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
78 
3.1.7.2 Conversion of Complex into Real Quantities 
The results to be selected in the Trace – Meas menu can be divided into two groups: 
S-Parameters, Impedances, and Admittances are complex.  
Stability Factors are real.  
The following table shows how the response values in the different Cartesian diagrams 
are calculated from the complex measurement values z = x + jy (where x, y, z are 
functions of the sweep variable). The formulas also hold for real results, which are 
treated as complex values with zero imaginary part (y = 0).  
Trace Format 
Description 
Formula 
dB Mag 
Magnitude of z in dB 
|z| = sqrt ( x
2
+ y
2
)  
dB Mag(z) = 20 * log|z| dB  
Lin Mag 
Magnitude of z, unconverted 
|z| = sqrt ( x
2
+ y
2
Phase 
Phase of z 
φ (z) = arctan (y/x) 
Real 
Real part of z 
Re(z) = x 
Imag 
Imaginary part of z 
Im(z) = y 
SWR 
(Voltage) Standing Wave Ratio 
SWR = (1 + |z|) / (1 – |z|) 
Group Delay 
Group delay, neg. derivative of the phase response 
– d φ (z) / dω  (ω = 2π * f) 
An extended range of formats and conversion formulas is available for markers. To 
convert any point on a trace, create a marker and select the appropriate marker format. 
Marker and trace formats can be selected independently.   
3.1.7.3 Polar Diagrams 
Polar diagrams show the measured data (response values) in the complex plane with a 
horizontal real axis and a vertical imaginary axis. The grid lines correspond to points of 
equal magnitude and phase.  
The magnitude of the response values corresponds to their distance from the 
center. Values with the same magnitude are located on circles.  
The phase of the response values is given by the angle from the positive horizontal 
axisValues with the same phase on straight lines originating at the center.  
The following example shows a polar diagram with a marker used to display a pair of 
stimulus and response values.  
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
79 
Example: Reflection coefficients in polar diagrams 
If the measured quantity is a complex reflection coefficient (S
11
, S
22
etc.), then the 
center of the polar diagram corresponds to a perfect load Z
0
at the input test port of the 
DUT (no reflection, matched input), whereas the outer circumference (|S
ii
| = 1) 
represents a totally reflected signal.   
0
180 
0
90 
0
-90 
0
Circles of equal
magnitude
Radial lines of
equal phase angle
Voltage reflection:
Open-circuited 
load (Z = infinity)
Voltage reflection:
Short-circuited 
load (Z = 0)
Matching
impedance (Z = Z
0
)
Examples for definite magnitudes and phase angles:  
The magnitude of the reflection coefficient of an open circuit (Z = infinity, I = 0) is 
one, its phase is zero.  
The magnitude of the reflection coefficient of a short circuit (Z = 0, U = 0) is one, its 
phase is –180
0
 
3.1.7.4 Smith Chart 
The Smith chart is a circular diagram that maps the complex reflection coefficients S
ii
to normalized impedance values. In contrast to the polar diagram, the scaling of the 
diagram is not linear. The grid lines correspond to points of constant resistance and 
reactance.  
Points with the same resistance are located on circles.  
Points with the same reactance produce arcs.  
The following example shows a Smith chart with a marker used to display the stimulus 
value, the complex impedance Z = R + j X and the equivalent inductance L (see 
marker format description in the help system).  
R&S
®
ZVL 
System Overview 
Basic Concepts 
Quick Start Guide 1303.6538.62-01 
80 
TIP 
Polar diagram types  
A comparison of the Smith chart, the inverted Smith chart and the polar diagram 
reveals many similarities between the two representations. In fact the shape of a trace 
does not change at all if the display format is switched from Polar to Smith or Inverted 
Smith – the analyzer simply replaces the underlying grid and the default marker 
format.  
Smith chart construction 
In a Smith chart, the impedance plane is reshaped so that the area with positive 
resistance is mapped into a unit circle.  
The basic properties of the Smith chart follow from this construction: 
The central horizontal axis corresponds to zero reactance (real impedance). The 
center of the diagram represents Z/Z
0
= 1 which is the reference impedance of the 
system (zero reflection). At the left and right intersection points between the 
horizontal axis and the outer circle, the impedance is zero (short) and infinity 
(open).  
The outer circle corresponds to zero resistance (purely imaginary impedance). 
Points outside the outer circle indicate an active component.  
The upper and lower half of the diagram correspond to positive (inductive) and 
negative (capacitive) reactive components of the impedance, respectively.  