Rev.  01/06 
- 2.6 -   
HTTP 
The  HTTP  contains  a  tactile  feedback  keypad 
interface that allows the user to configure parameters 
used by the HTTP operating system.   
The following “Soft Key” menu items will be displayed 
immediately above the two keys located in the lower 
corners of the Graphics Display. 
1.  The (soft)MENU key is pressed from RUN mode to 
enter  PROGRAM  mode.  The  (soft)EXIT  key  is 
pressed in PROGRAM mode to exit configuration 
parameters  and  menus.  If  any  configuration 
changes are made, the user will be prompted with 
a SAVE? (soft)YES or (soft)NO when returning to 
RUN mode. If no changes are made, the user will 
not be prompted to SAVE. 
2.  The  UP/DOWN ARROW keys are used to scroll 
through menus and configuration parameters.  The 
ARROW  keys  can  also  be  used  to  adjust 
parameter numerical values. In RUN mode the UP/
DOWN  ARROW  keys  are  used  to  adjust  the 
display contrast level.   
3.  The  Numerical  Keypad  is  used  for  entering 
numerical values. 
4.  The (soft)EDIT key is used to 
• 
access the configuration parameters in the 
various menus. 
• 
initiate changes in configuration parameters. 
PART 2 - TRANSDUCER POSITIONING 
Keypad Entry 
The HTTP system can be configured using UltraLink
and a laptop computer.  Please refer to page 2.13 for 
details regarding data entry in UltraLink
.
Graphics Display 
Configuration 
Copy formatted text from pdf - extract text content from PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Feel Free to Extract Text from PDF Page, Page Region or the Whole PDF File
copy text from pdf reader; copy paste text pdf file
Copy formatted text from pdf - VB.NET PDF Text Extract Library: extract text content from PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
How to Extract Text from PDF with VB.NET Sample Codes in .NET Application
copy text from pdf in preview; c# read text from pdf
Rev.  01/06 
- 2.7 -   
HTTP 
5.  The (soft)ACCEPT key is used to 
• 
accept configuration parameter changes. 
6.  The (soft)SELECT key is used to 
• 
configure  the  engineering  units  on  the 
graphics display.  Press the (soft)SELECT 
key  from  RUN  mode  to  highlight  the 
engineering  unit  presently  being  displayed 
on  the  graphics  display  (pressing  the 
SELECT key multiple times will toggle the 
highlighted unit from line to line).  Use the 
UP/DOWN ARROW keys to select display 
units of: 
• 
RATE 
• 
TOTALizer 
• 
VELocity 
• 
SIGNAL STRength 
From Menu 7, Display Menu, the number of   
graphics  display  lines  can  be  toggled 
between two and four lines. 
Menu  1,  the BASIC  menu  contains  all  of  the 
configuration  parameters  necessary  to  make  the 
transducer spacing calculation. 
UNITS 
ENGLSH 
METRIC 
Installs  a  global  measurement  standard  into  the 
operation of the instrument.  The choices are either 
English or Metric measurements. 
• 
Select  ENGLSH  if  all  configurations  (pipe 
sizes, etc.) are to be made in inches.  Select 
METRIC if the meter is to be configured in 
millimeters. 
• 
The  ENGLSH/METRIC  selection  will  also 
configure the HTTP to display sound speeds 
in pipe materials and liquids as either feet 
per  second  or  meters  per  second 
respectively. 
PART 2 - TRANSDUCER POSITIONING 
UNITS Entry 
C# PDF insert text Library: insert text into PDF content in C#.net
class program. Insert formatted text and plain text to PDF page using .NET XDoc.PDF component in C#.NET class. Supports adding text
extract text from pdf file using java; get text from pdf into excel
VB.NET PDF insert text library: insert text into PDF content in vb
VB.NET. Insert formatted text and plain text to PDF page. Add text to certain position of PDF page in Visual Basic .NET class. Add
extract text from pdf using c#; get text from pdf file c#
Rev.  01/06 
- 2.8 -   
HTTP 
XDCR MNT -- Transducer Mounting Method 
Selects the mounting orientation for the transducers. 
The selection of an appropriate mounting orientation is 
based  on  pipe  and  liquid  characteristics.    Refer  to 
Table 2.2 in this manual. 
V -- Mount.  A reflective type (transducers mounted on 
one side of the pipe) of installation used primarily on 
pipe  sizes  in  the  3-10  inch  (75-250  mm)  internal 
diameter range. 
W -- Mount.  A reflective type (transducers mounted 
on one side of the pipe) of installation used primarily 
on  pipe  sizes  in  the  2-3  inch  (50-75  mm)  internal 
diameter range. 
Z --  Mount.  A direct type (transducers mounted on 
opposite  sides  of  the  pipe)  of  installation  used 
primarily on pipe sizes in the 10-100 inch (250-2540 
mm) internal diameter range. 
PIPE OD -- Pipe Outside Diameter Entry 
ENGLSH (Inches) 
METRIC (Millimeters) 
Enter the pipe outside diameter
in inches if ENGLSH 
was selected as UNITS; in millimeters if METRIC was 
selected. 
PIPE WT -- Pipe Wall Thickness Entry 
ENGLSH (Inches) 
METRIC (Millimeters) 
Enter the pipe wall thickness
in inches if ENGLSH was 
selected  as  UNITS;  in  millimeters  if  METRIC  was 
selected. 
PART 2 - TRANSDUCER POSITIONING 
Transducer Mount 
Configuration 
Pipe O.D. Entry 
Pipe Wall Entry 
Rev.  01/06 
- 2.9 -   
HTTP 
PIPE MAT -- Pipe Material Selection 
CARBON S - Carbon Steel 
STAINLES - Stainless Steel 
CAST IRO - Cast Iron 
DUCTILE - Ductile Iron 
COPPER - Copper 
PVC - Polyvinylchloride 
PVDF LOW - 
Low Density Polyvinylidene Flouride
PVDF HI - 
High Density Polyvinylidene Flouride
ALUMINUM - Aluminum 
ASBESTOS - Asbestos Cement 
FIBERGLA - Fiberglass 
OTHER 
This  list  is  provided  as  an  example.  Additional 
materials  are  being  added  continuously.  Select  the 
appropriate  pipe  material  from  the  list  or  select 
OTHER if the material is not listed. 
PIPE SS -- Speed of Sound in the Pipe Material 
ENGLSH (Feet per Second) 
METRIC (Meters per Second) 
Allows adjustments to be made to the speed of sound 
in  the  pipe  wall.  If  the  UNITS  value  was  set  to 
ENGLSH,  the  entry  is  in  FPS  (feet  per  second).  
METRIC  entries  are  made  in  MPS  (meters  per 
second). 
If a pipe material was chosen from the PIPE MAT list, 
a nominal value for speed of sound in that material will 
be automatically loaded.  If the actual sound speed 
rate  is known for  the application  piping  system  and 
that value varies from the automatically loaded value, 
the value can be revised. 
If OTHER was chosen as PIPE MAT, a PIPE SS will 
need to be entered. 
PART 2 - TRANSDUCER POSITIONING 
Pipe Material Entry 
Pipe Sound Speed 
Entry 
Rev.  01/06 
- 2.10 -  
HTTP 
PIPE R -- Pipe Material Relative Roughness 
UNITLESS VALUE 
The HTTP provides Reynolds Number compensation 
in  its  flow  measurement  calculation.    The  ratio  of 
average surface imperfection as it relates to the pipe 
internal diameter is used in this compensation. 
Linear RMS measurement of the pipe 
PIPE R = 
internal wall surface  
Internal Diameter of the pipe 
If a pipe material was chosen from the PIPE MAT list, 
a nominal value relative roughness in that material will 
be automatically loaded.  If the actual roughness  is 
known for the application piping system and that value 
varies from the automatically loaded value, the value 
can be revised. 
If OTHER was chosen as PIPE MAT, a PIPE R may 
need to be entered. 
LINER T -- Pipe Liner Thickness Entry 
ENGLSH (Inches) 
METRIC (Millimeters) 
Enter  the  pipe  liner  thickness
   Enter  this  value  in 
inches  if  ENGLSH  was  selected  as  UNITS;  in 
millimeters if METRIC was selected. 
[If a LINER Thickness was selected] 
LINER MAT - Liner Material 
TAR EPOXY 
RUBBER 
MORTAR 
POLYPROPYLENE 
POLYSTYROL 
POLYSTYRENE 
POLYESTER 
POLYETHYLENE 
EBONITE 
TEFLON 
Other 
PART 2 - TRANSDUCER POSITIONING 
Pipe Roughness 
Entry 
Liner Material Entry 
Liner Thickness 
Entry 
Rev.  01/06 
- 2.11 -  
HTTP 
This  list  is  provided  as  an  example.    Additional 
materials are being added continuously.  Select the 
appropriate material from the list or select OTHER if 
the liner material is not listed. 
LINER SS -- Speed of Sound in the Liner 
ENGLSH (Feet per Second) 
METRIC (Meters per Second) 
Allows adjustments to be made to the speed of sound 
in the liner.  If the UNITS value was set to ENGLSH, 
the entry is in FPS (feet per second).  METRIC entries 
are made in MPS (meters per second).  If a liner was 
chosen from the LINER MAT list, a nominal value for 
speed  of  sound  in  that  media  will  be  automatically 
loaded.  If the actual sound speed rate is known for 
the  pipe  liner  and  that  value  varies  from  the 
automatically loaded value, the value can be revised. 
FL TYPE - Fluid/Media Type 
TAP WATER 
SEWAGE-RAW  
ALCOHOL 
SEA WATER 
KEROSENE 
GASOLINE 
FUEL OIL 
CRUDE OI 
PROPANE 
BUTANE 
OTHER 
This list is provided as an example.  Additional liquids 
are being added continuously.  Select the appropriate 
liquid from the list or select OTHER if the liquid is not 
listed. 
FLUID SS -- Speed of Sound in the Fluid 
ENGLSH (Feet per Second) 
METRIC (Meters per Second) 
PART 2 - TRANSDUCER POSITIONING 
Liner Sound Speed 
Entry 
Fluid Type Entry 
Fluid Sound Speed 
Entry 
Rev.  01/06 
- 2.12 -  
HTTP 
Allows adjustments to be made to the speed of sound 
in the liquid.  If the UNITS value was set to ENGLSH, 
the entry is in FPS (feet per second).  METRIC entries 
are made in MPS (meters per second).  If a fluid was 
chosen  from the FL TYPE list,  a nominal  value for 
speed  of  sound  in  that  media  will  be  automatically 
loaded.  If the actual sound speed rate is known for 
the  application  fluid  and  that  value  varies  from  the 
automatically loaded value, the value can be revised. 
If OTHER was chosen as FL TYPE, a FLUID SS will 
need to be entered.  A list of alternate fluids and their 
associated sound speeds are located in the Appendix 
located at the back of this manual. 
FLUID VI -- Absolute Viscosity of the Fluid 
cps 
Allows  adjustments  to  be  made  to  the  absolute 
viscosity of the liquid. If a fluid was chosen from the FL 
TYPE list, a nominal value for viscosity in that media 
will be automatically loaded.  If the actual viscosity is 
known for the application fluid and that value varies 
from the automatically loaded value, the value can be 
revised.   
If OTHER was chosen as FL TYPE, a FLUID VI will 
need to be entered.  A list of alternate fluids and their 
associated  viscosities  are  located  in  the  Appendix 
located at the back of this manual. 
SP GRVTY -- Fluid Specific Gravity Entry 
unitless 
Allows adjustments to be made to the specific gravity 
(density) of the liquid.  If a fluid was chosen from the 
FL TYPE list, a nominal value for specific gravity in 
that media will be automatically loaded.  If the actual 
specific gravity is known for the application fluid and 
that value varies from the automatically loaded value, 
the value can be revised. 
PART 2 - TRANSDUCER POSITIONING 
Fluid Viscosity Entry 
Fluid Specific 
Gravity Entry 
Rev.  01/06 
- 2.13 -  
HTTP 
If OTHER was chosen as FL TYPE, a SP GRVTY may 
need to be entered if mass flows are to be calculated.  
A list of alternate fluids and their associated specific 
gravities  are located  in the Appendix  located at the 
back of this manual. 
XDCR SPAC -- Transducer Spacing Calculation 
ENGLSH (Inches) 
METRIC (Millimeters) 
This  value  represents  the  one-dimensional  linear 
measurement between the transducers (the upstream/
downstream  measurement  that  runs  parallel  to  the 
pipe).  This value is in inches if ENGLSH was selected 
as  UNITS,  in  millimeters  if  METRIC  was  selected.  
This  measurement  is  taken  from  the  line  which  is 
scribed into the side of the transducer block. 
UltraLink
Data Entry 
The UltraLink
™ 
Windows
®
-based  software  utility 
provides  an  efficient means  for  entering  piping  and 
liquid parameters through the use of pop-up window/
pull-down menu structures.  Data can be entered into 
UltraLink
, stored, later retrieved and downloaded  at 
the HTTP installation site (provided that UltraLink
™ 
and 
HTTP communications are not enabled at the time of 
data entry) or it can be downloaded immediately to the 
HTTP  meter  (provided  that UltraLink
™ 
and  HTTP 
communications are enabled during data entry). 
PART 2 - TRANSDUCER POSITIONING 
Transducer Spacing 
Calculation 
IMPORTANT: Pipe sizes under 2 inches (50 mm) 
require optional transducers and a modified HTTP 
system that can be purchased separately.  See Page 
2.22 for additional details or consult the Hedland 
factory for information.
UltraLink
Entry 
Rev.  01/06 
- 2.14 -  
HTTP 
To  install UltraLink
™ 
and  establish  communications 
with a PC, please follow the instructions enclosed with 
the UltraLink
™ 
software package or in the Appendix of 
this manual. 
The  system  information  required  for  entry  into  the 
UltraLink
™ 
package  is  identical  to  that  required  for 
Keypad Entry covered in the previous section.  See 
pages 2.3-2.5. 
After initializing UltraLink
, click on the button labeled 
Config The window shown in Figure 2.1 will appear.  
Enter  the  pipe  and  liquid  parameters  into  the 
appropriate  data  fields  in  the Basic  window.    The 
correct  transducer  spacing  will  appear  in  the 
Transducer - Spacing data field. 
PART 2 - TRANSDUCER POSITIONING 
Figure 2.1   UltraLink
™ 
Windows
®
-based software utility configuration 
screen. 
Transducer 
spacing 
appears here. 
Rev.  01/06 
- 2.15 -  
HTTP 
After all data fields have been entered, Download to 
the HTTP or File Save to a disk by clicking on the 
appropriate button in the Config window.  Download 
is  not  possible  unless  communications  are  enabled 
between the HTTP and UltraLink
 Communications 
are enabled when a green OK is indicated in the lower 
right-hand COMM: status box.  If communications are 
not  enabled,  please  review  the  documentation  that 
details the installation and initialization of UltraLink
 
Refer to Part 4, page 4.1. 
After selecting an optimal mounting location, Step 1, 
and  successfully  determining  the  proper  transducer 
spacing, Step 2, the transducers can now be mounted 
onto the pipe. 
The HTT transducers need to be properly oriented on 
the  pipe  to  provide  optimum  reliability  and 
performance.  On  horizontal  pipes,  the  transducers 
should  be  mounted  180  radial  degrees  from  one 
another and at least 45 degrees from the top-dead-
center  and  bottom-dead-center  of  the  pipe.  See 
Figure  2.2. Figure 2.2 does not apply to vertically 
oriented pipes. 
PART 2 - TRANSDUCER POSITIONING 
3.  Transducer Mounting 
Figure 2.2  Horizontal pipe transducer mounting 