Rev.  01/06 
-3.13-   
HTTP 
MIN RATE - Minimum Flow Rate Settings 
Rate Unit/Rate Interval 
A minimum volumetric flow rate setting is entered to 
establish filter software settings. Volumetric entries will 
be in the Engineering Rate Units and Interval selected 
on pages 3.10-3.11 of this manual. For uni-directional 
measurements,  set  MIN  RATE  to  zero.  For  bi-
directional measurements, set to the highest negative 
(reverse) flow rate expected in the piping system. 
MAX RATE - Maximum Flow Rate Settings 
Rate Unit/Rate Interval 
A maximum volumetric flow rate setting is entered to 
establish filter software settings. Volumetric entries will 
be in the Engineering Rate Units and Interval selected 
on pages 3.10-3.11 of this manual. For uni-directional  
or bi-directional measurements, set MAX RATE to the 
highest  (positive)  flow  rate  expected  in  the  piping 
system. 
FL C-OFF - Low Flow Cut-off 
Percent of the range between MIN RATE and 
MAX RATE 
A Low Flow Cut-off entry is provided to allow very low 
flow rates (that can be present when pumps are off 
and valves are closed) to be displayed as Zero flow.  
Typical  values that  should  be  entered  are  between 
1.0% and 5.0% of the flow range between MIN RATE 
and MAX RATE. 
DAMP PER - System Damping 
Relative Percent Entry: 0-100% 
Flow  Filter  Damping establishes a maximum 
adaptive filter value. Under the stable flow conditions 
(flow varies less than 10% of reading) this adaptive 
filter  will  increase  the  number  of  successive  flow 
PART 3 - KEYPAD CONFIGURATION 
System Damping 
Low Flow Cut-off 
Maximum 
Flow Rate 
Minimum 
Flow Rate 
Copy text from encrypted pdf - extract text content from PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Feel Free to Extract Text from PDF Page, Page Region or the Whole PDF File
extract formatted text from pdf; c# read text from pdf
Copy text from encrypted pdf - VB.NET PDF Text Extract Library: extract text content from PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
How to Extract Text from PDF with VB.NET Sample Codes in .NET Application
c# extract text from pdf; extract all text from pdf
Rev.  01/06 
-3.14-   
HTTP 
readings  that  are  averaged  together  up  to  this 
maximum value. If flow changes outside of the 10% 
window,  the  Flow  Filter  adapts  by  decreasing  and 
allows the meter to react faster. Increasing this value 
tends to provide smoother steady-state flow readings 
and outputs. If erratic flow conditions are present or 
expected, more advanced filters are available for use 
in the UltraLink
software utility.  See Part 4 for further 
information. 
ISO-MOD
Data Logger 
File Number/Location ID 1-30,000 
[16 total location IDs] 
INTERVAL 1-30,000 seconds 
DURATION  1-30,000 hours 
The Series HTTP has an optional 200,000-point data 
logger/electronic  stripchart recorder.  The logger can 
be configured in a couple of different ways to match 
user  applications.  The  logger  stores  time-stamped, 
high resolution (16-bit) data at user-selected intervals 
ranging  from  1  to  30,000  (8.33  hours)  seconds. 
Configuration and data retrieval from the logger can be 
accomplished in one of two ways: 
• 
The module is hot-swappable -- it can be installed, 
removed or replaced within the flow meter without 
disconnecting power. The module can be carried in 
a pocket and plugged into a PC serial port via the 
module's  integral  DB9  connector.  This  feature 
eliminates the need for a laptop computer to be at 
the flow meter site. 
• 
Via the Data Logger software utility and the serial 
DB9 interface cable included with the logger.  Data 
can be accessed by connecting the cable to the 
logger, which is located in the pocket on the front 
faceplate of the  instrument.   See Figure 3.2. e 
2.  DATALOG OPERATION MENU 
Figure 3.2 
PART 3 - KEYPAD CONFIGURATION 
C# PDF Password Library: add, remove, edit PDF file password in C#
passwordSetting.IsExtract = true; // Copy is allowed. In this part, you will know how to change and update password for an encrypted PDF file in C# programming
find and replace text in pdf file; copy text from scanned pdf
VB.NET PDF Password Library: add, remove, edit PDF file password
passwordSetting.IsExtract = True ' Copy is allowed In this part, you will know how to change and update password for an encrypted PDF file in VB.NET programming
copy and paste text from pdf to word; can't copy text from pdf
Rev.  01/06 
-3.15-   
HTTP 
See Section 4 for details regarding operation of the 
Data Logger and UltraLink
software utilities. 
There are three configuration parameters to enter for 
data logger operation: 
• 
FILE NUMBER/LOCATION ID 
• 
INTERVAL 
• 
DURATION 
The HTTP and the data logger module can be used to 
monitor and store data on up to 16 different locations.  
These locations are identified by the FILE NUMBER/
LOCATION ID that is assigned. The data logger will 
not write over an existing file. Existing data files will 
need to be uploaded and then erased from the logger 
before new data can be written into the space. The 
data  logger  will  stop  logging  once  it  has  reached 
30,000 location IDs.  Uploading and storing of files is 
completed using the Data Logger software utility.  See 
Section 4 of this manual for details regarding the Data 
Logger and UltraLink
utilities. 
From the OUTPUT 1 menu, adjust the time INTERVAL 
between readings.  INTERVAL values between 1 and 
30,000 seconds are acceptable. 
For reference there are: 
60 seconds in 1 minute 
300 seconds in 5 minutes 
1,800 seconds in 30 minutes 
3,600 seconds in 1 hour 
30,000 seconds in 8.33 hours 
PART 3 - KEYPAD CONFIGURATION 
Logging Interval 
File Number or 
Location 
Data Logger 
Configuration 
NOTE:  There is a maximum of 30,000 location IDs 
(samples) per location. The logger will stop logging 
once it reaches this point, and a new data file will need 
to be entered. 
VB.NET Excel: Render and Convert Excel File to TIFF Image by Using
contained in the Excel needs to be encrypted, it will If you want to view or edit PDF, Word, Excel or Excel document to REImage, you can just copy following VB
copy pdf text to word; extract text from pdf c#
Rev.  01/06 
-3.16-   
HTTP 
Table 3.2 describes some typical configurations of the 
INTERVAL and DURATION times with what the ex-
pected data samples collected count will be. 
If the HTTP is going to be left unattended, logging flow 
data,  for  extended  periods  of  time,  the  DURATION 
time can be configured to stop logging after the DU-
RATION of time has passed.  DURATION is config-
ured in hours and values between 1 and 30,000 hours 
are acceptable. 
Datalog Maintenance permits files to be deleted from 
the data logger module.  The Menu contains three op-
tions for deleting  files:   delete the  last  file that was 
generated, delete the first file that was generated or 
delete all of the files on the logger. 
PART 3 - KEYPAD CONFIGURATION 
Example No. 
INTERVAL 
Seconds 
DURATION 
Hours 
Operated 
Samples  
Collected 
21,600 
10 
72 (3 days) 
25,920 
60 (1min) 
480 (20 days) 
28,800 
300 (5 min) 
2,016 (12 wks) 
24,192 
1,800 (30 min) 
8,760 (1 yr) 
17,520 
3,600 (1 hr) 
8,760 (1 yr) 
8,760 
18,000 (5 hr) 
26,280 (3 yr) 
17,520 
Table 3.2 
Logging Duration 
3.  DATALOG MAINTENANCE 
NOTE:  There is a maximum of 30,000 location IDs 
(samples) per location.  
Rev.  01/06 
-3.17-   
HTTP 
ISO-MOD
4-20 mA 
FL    4MA 
FL    20MA 
CAL  4MA 
CAL  20MA 
4-20  TST 
Configured via jumper selections for either a passive 
or  active  transmission  mode,  the  4-20  mA  Output 
Module interfaces with virtually all recording and log-
ging systems by transmitting an analog current signal 
that is proportional to system flow rate.  Independent  
4 mA and 20 mA span settings are established in soft-
ware using the Flow Measuring Range entries.  These 
entries can be set anywhere in the –40 to +40 FPS     
(-12 to +12 MPS) measuring range of the instrument. 
Output resolution of the module is 12-bits (4096 dis-
crete  points)  and  the  module  can  drive  up  to  800 
Ohms  of  load  with  its  internal  24V  isolated  power 
source. 
The FL  4MA and FL  20MA entries are used to set the 
span of the 4-20 mA analog output. These entries are 
volumetric rate units that are equal to the volumetric 
units configured as Engineering Rate Units and Engi-
neering Units Time Interval entered on page 3.10.  
For example, to span the 4-20 mA output from –100 
GPM to +100 GPM, with 12 mA being 0 GPM, set the 
FL  4MA and FL  20MA inputs as follows: 
FL  4MA  =  -100.0 
FL  20MA = 100.0 
For example, to span the 4-20 mA output from 0 GPM 
to +100 GPM, with 12 mA being 50 GPM, set the FL  
4MA and FL  20MA inputs as follows: 
FL  4MA  =  0.0 
FL  20MA = 100.0 
PART 3 - KEYPAD CONFIGURATION 
4.  OUTPUT 2 MENU 
4-20 mA Span 
Standard 
4-20 mA 
Rev.  01/06 
-3.18-   
HTTP 
The  4-20  mA  ISO-MOD  is  factory  calibrated  and 
should not require adjustment unless it is replaced.   
The  CAL4MA  entry  allows  fine  adjustments  to  be 
made to the “zero” of the 4-20 mA output.  To adjust 
the  4  mA  output,  an  ammeter or reliable  reference 
connection to the 4-20 mA output must be present. 
Procedure: 
1.  Disconnect  one  side  of  the  current  loop  and 
connect the ammeter in series (disconnect either 
wire at the terminals labeled +/- on the 
ISO-MOD
4-
20 mA module).  
2.  Using the arrow keys, increase the numerical value 
to increase the current in the loop to 4 mA.  De-
crease  the  value  to  decrease  the  current  in  the 
loop to 4 mA.  Typical values range between 40-80 
counts. 
Re-connect the 4-20 mA output circuitry as required. 
Calibration of the 20 mA setting is conducted much the 
same way as the 4 mA adjustments. 
Procedure: 
1.  Disconnect  one  side  of  the  current  loop  and 
connect the ammeter in series (disconnect either 
wire at the terminals labeled +/- on the 
ISO-MOD
4-
20 mA module).  
2.  Using the arrow keys, increase the numerical value 
to increase the current in the loop to 20 mA.  De-
crease  the  value  to  decrease  the  current  in  the 
loop  to  20  mA.    Typical  values  range  between 
3700-3900 counts. 
Re-connect the 4-20 mA output circuitry as required. 
PART 3 - KEYPAD CONFIGURATION 
4-20 mA Calibration 
NOTE:  The CAL 4MA and CAL 20MA entries should 
not be used in an attempt to set the 4-20 mA range.  
Utilize FL 4MA and FL 20MA, detailed above, for this 
purpose. 
Rev.  01/06 
-3.19-   
HTTP 
4-20 TST - 4-20 mA Output Test 
4-20 
Allows a simulated value to be output on from the 4-20 
mA output.  By incrementing this value, the 4-20 mA 
output will transmit the indicated current value. 
The Series HTTP allows other input/output modules to 
be used in place of the 4-20 mA output.  Please con-
tact the Hedland factory regarding these options:  
ISO-MOD
RATE PULSE 
FL 0H 
FL 25KH 
The Rate Pulse Output Module is utilized to transmit 
information to external counters and PID systems via a 
frequency output that is proportional to  system  flow 
rate. Independent Zero and Span settings are estab-
lished in  software using  the Flow Measuring Range 
entries.  Output  resolution  of  the  module  is  12-bits 
(4096 discrete points) and the maximum  output fre-
quency setting is 2,500Hz. The module has two output 
modes, turbine meter simulation and “open collector”.  
The  turbine  meter simulation  sources  a non-ground 
referenced saw-tooth waveform with a maximum peak 
amplitude  of  approximately  500mVpp.  The  open-
collector output utilizes a 0.21-Ohm FET output that is 
rated to operate at 100 V and 1 A maximum. If the 
open-collector output type is utilized, an external volt-
age source and limit resistor must be present. 
The FL 0H and FL 25KH entries are used to set the 
span of the 0-2.5KHz frequency output. These entries 
are volumetric rate units that are equal to the volumet-
ric units configured as Engineering Rate Units and En-
gineering Units Time Interval entered on page 3.11.   
PART 3 - KEYPAD CONFIGURATION 
4-20 mA Test 
Optional Modules 
Optional 
Rate Pulse 
Rate Pulse Span 
Rev.  01/06 
-3.20-   
HTTP 
In a bi-directional system, to span the 0-2.5kHz output 
from -100 GPM to +100 GPM, with 1.25kHz being 0 
GPM, set the FL 100H and FL 10KH inputs as follows: 
FL  0H  =  100.0 
FL  25KH  =  100.0 
To span the 0-2.5 kHz output from 0 GPM to +100 
GPM, with 1.25 kHz being 50 GPM, set the FL 0H and 
FL 25KH inputs as follows: 
FL  0H  =  0  
FL  25KH  =  100.0 
ISO-MOD
Dual Relay 
RELAY 1 AND RELAY 2 
NONE 
TOTALIZE 
TOT MULT 
FLOW 
ON 
OFF 
SIG STR 
ERRORS 
Two  independent  SPDT  (single-pole,  double-throw, 
Form C) relays are contained in this module. The relay 
operations are user configured via software to act in 
either a flow rate alarm, signal strength alarm, error 
alarm or totalizer/batching mode.  The relays are rated 
for 200 VAC max. and a have current rating of 0.5A 
resistive load (175 VDC @ 0.25A resistive).  It is highly 
recommended  that  a  secondary  relay  be  utilized 
whenever the Control Relay ISO-MOD is used to con-
trol inductive loads such as solenoids and motors. 
TOTALIZE mode configures the relay to output a 50 
mSec pulse (contact changeover) each time the dis-
play totalizer increments—divided by the TOT MULT. 
The TOT MULT value must be a whole, positive, nu-
merical value. 
PART 3 - KEYPAD CONFIGURATION 
Optional 
Dual Relay 
Batch/Totalizer 
Relay 
Rev.  01/06 
-3.21-   
HTTP 
• 
If the Totalizer Exponent is set to E0 (x1) and the 
Relay Multiplier is set to 1, then the relay will pulse 
each time the totalizer increments one count, or 
each single, whole measurement unit totalized. 
• 
If the Totalizer Exponent is set to E2 (x100) and 
the Relay Multiplier is set to 1, then the relay will 
pulse each time the display totalizer increments or 
once per 100 measurement units totalized. 
• 
If the Totalizer Exponent is set to E0 (x1) and the 
Relay Multiplier is set to 2, the relay will pulse once 
for every two counts that the totalizer increments. 
Flow rate relay configuration permits relay changeover 
at two separate flow rates, allowing operation with an 
adjustable switch deadband. Figure 3.3 illustrates how 
the setting of the two set points influences Rate Alarm 
operation.  
A single-point flow rate alarm would place  the ON> 
setting slightly higher than the OFF< setting -allowing 
a switch deadband to be established. If a deadband is 
not established, switch chatter (rapid switching) may 
result when flow rate is too close to the switch point. 
The SIG STR alarm will provide an indication that the 
flow  meter  signals  between  the  transducers  have 
fallen to a point where flow measurements may not be 
possible. It can also be used to indicate the pipe has 
PART 3 - KEYPAD CONFIGURATION 
Flow Rate Relay 
Minimum flow 
Maximum flow 
Set  ON > 
Set  OFF < 
Relay OFF 
Relay ON 
Figure 3.3 
Single Point Alarm Operation 
Deadband 
Signal Strength 
Alarm 
Rev.  01/06 
-3.22-   
HTTP 
emptied.  Like the  Flow Rate alarm  described previ-
ously,  the  Signal  Strength  alarm  requires  that  two 
points be entered, establishing an alarm deadband. A 
valid switch point exists when the ON> is a value lower 
than OFF<. If a deadband is not established and the 
signal strength decreases to approximately the value 
of the switch point, the relay may chatter. 
If a relay is set to ERROR mode, the relay will activate 
when errors occur in the flow meter that has caused 
the meter to stop measuring reliably. See  Appendix  
for a list of potential error codes. 
ISO-MOD
RS232C 
RS232 MO  — MODE 
HOST 
UIF 
RS232 BA — BAUD RATE 
1200 
2400 
9600 
19200 
The RS232 Module can be interfaced with serial com-
munication ports of PCs, PLCs and SCADA systems. 
The module runs a proprietary digital protocol, detailed 
in the Appendix, that is used to monitor flow rate infor-
mation  in  piping  systems.  The  RS232  Module  may 
also be used to form a hardwire connection to a PC 
that  is  running  the UltraLink
software utility. Baud 
rates up to 19.2 K are supported. 
PART 3 - KEYPAD CONFIGURATION 
Error Alarm Relay 
Optional 
RS232C Module 