Rev.  01/06 
-3.23-   
HTTP 
ISO-MOD
RS485 
RS485 MO  — MODE 
SLAVE 
MASTER 
RS485 BA — BAUD RATE 
1200 
2400 
9600 
19200 
ADDRESS — Device Address 
1-127 
The RS485 Module allows up to 126 HTTP systems to 
be daisy-chained on a single three-wire cable network. 
Communications are via a proprietary digital protocol, 
detailed in the Appendix. All meters are assigned a 
unique one byte serial number that allows all of the 
meters  on  the  cable network  to  be  accessed  inde-
pendently.  Baud rates up to 19.2K and cable lengths 
to 5,000 feet (1,500 meters) are supported without the 
need for repeaters. 
RS485 MO 
Select SLAVE for all of the HTTP meters. 
RS485 BA 
Select a Baud rate that is compatible with the operat-
ing system. 
ADDRESS 
Each  HTTP  connected  on  the  communications  bus 
must  have  a unique address number assigned. Ad-
dress 127 is a universal address that will result in all 
HTTP instruments on the network responding simulta-
neously—regardless of address—resulting in CRC er-
rors. Only select address location 127 if one meter is 
on the network. 
PART 3 - KEYPAD CONFIGURATION 
Optional 
RS485 Module 
Copy text from pdf - extract text content from PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Feel Free to Extract Text from PDF Page, Page Region or the Whole PDF File
copy pdf text to word document; copy and paste text from pdf to excel
Copy text from pdf - VB.NET PDF Text Extract Library: extract text content from PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
How to Extract Text from PDF with VB.NET Sample Codes in .NET Application
copy text from protected pdf; copy text from pdf reader
Rev.  01/06 
-3.24-   
HTTP 
The SEN MENU is utilized to select the type of trans-
ducer  that will be  interfaced  with  the HTTP  meter.  
Select the appropriate  transducer  from  the list  and 
save the configuration. If the transducer selection is 
modified, a system reset is required
SEC MENU allows users to make password revisions, 
reset flow totalizer and reset transmitter microproces-
sor. 
TOT RES 
NO 
YES 
Select YES to reset flow totalizer/accumulator to Zero. 
SYS RSET 
NO 
YES 
Select YES to initiate a microprocessor reset. All  con-
figurations and totalizer values will be maintained. 
CH PSWD? -- Change the Security Password 
0-9999 
By changing the Security Password from 0 to another 
value (between 1-9999), configuration parameters will 
not  be  accessible  without  first  entering  that  value 
when prompted.  If the value is left at 0, no security is 
invoked and unauthorized changes could be made. 
PART 3 - KEYPAD CONFIGURATION 
5.  
SEC MENU 
-- SECURITY MENU 
System RESET 
Totalizer RESET 
Change Password 
4.  
SEN MENU 
-- SENSOR MENU 
C# PDF Page Extract Library: copy, paste, cut PDF pages in C#.net
C#.NET PDF Library - Copy and Paste PDF Pages in C#.NET. Easy Ability to copy selected PDF pages and paste into another PDF file. The
delete text from pdf preview; copy text pdf
VB.NET PDF Page Extract Library: copy, paste, cut PDF pages in vb.
Extract, Copy, Paste PDF Pages. |. Home ›› XDoc.PDF ›› VB.NET PDF: Copy and Paste PDF Page. Ability to copy PDF pages and paste into another PDF file.
cut text pdf; copy highlighted text from pdf
Rev.  01/06 
-3.25-   
HTTP 
The  SERVICE  Menu  makes  available  two different 
system  measurements  that  are  used  for  trouble-
shooting and fine tuning of the instrument.  Actual liq-
uid sound speed and system signal strength readings 
can be accessed through this menu.  
The SERVICE Menu also has features that allow ad-
justment of  Signal Strength Cutoff,  Error-Mode  out-
puts, Zero Flow Rate Set and entry of a universal cor-
rection factor. 
SSPD MPS - Sound Speed in the Liquid Metric 
SSPD FPS - Sound Speed in the Liquid U.S. 
The HTTP performs an actual speed of sound calcu-
lation for the liquid it measures.  This calculation will 
vary with temperature, pressure and fluid composi-
tion.  The value indicated should be within 2% of the 
value  entered/indicated  in  the  BASIC  menu  item 
FLUID SS.  The value cannot be modified. If the ac-
tual measured value is significantly different than the 
BASIC MENU’s FLUID SS value, it typically indicates 
a problem with the instrument setup. Any entry such 
as PIPE OD or wall thickness may be in error, the 
pipe may not be round, or the transducer spacing is 
not correct. Table 3.3 on page 3.26 lists sound speed 
values for water at varying temperatures. If the HTTP 
is measuring sound speed within 2% of the table val-
ues, then the installation and setup of the instrument 
is correct and accurate readings may be assured. 
PART 3 - KEYPAD CONFIGURATION 
Liquid Sound 
Speed 
6.  
SER MENU 
-- SERVICE MENU 
C# PDF Image Extract Library: Select, copy, paste PDF images in C#
PDF ›› C# PDF: Extract PDF Image. How to C#: Extract Image from PDF Document. Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document.
copy pdf text with formatting; extract text from pdf online
VB.NET PDF Image Extract Library: Select, copy, paste PDF images
VB.NET PDF - Extract Image from PDF Document in VB.NET. Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document in VB.NET Project.
delete text from pdf with acrobat; extract text from pdf java
Rev.  01/06 
-3.26-   
HTTP 
SIG STR - Signal Strength 
The measurement of Signal Strength assists service 
personnel  in  troubleshooting  the HTTP  system.   In 
general,  expect  the  signal  strength  readings  to  be 
greater than 4% on a full pipe with the transducers 
properly mounted.  Signal strength readings that are 
less than 4% may indicate a need to chose an alter-
native mounting method for the transducers, or that 
an improper pipe size has been entered. 
Signal Strength readings in excess of 95% may indi-
cate that a mounting method with a longer path length 
may  be  required.  For  example,  if  transducers 
mounted on a 3 inch PVC pipe in V-mode causes the 
measured  Signal  Strength  value  to  exceed  95%, 
change the mounting method to W-mode for greater 
stability in readings. 
PART 3 - KEYPAD CONFIGURATION 
Deg. C 
Deg. F 
Vs (m/s) 
Vs (f/s) 
32 
1402 
4600 
10 
50 
1447 
4747 
20 
68 
1482 
4862 
30 
86 
1509 
4951 
40 
104 
1529 
5016 
50 
122 
1543 
5062 
60 
140 
1551 
5089 
70 
158 
1555 
5102 
80 
176 
1554 
5098 
90 
194 
1550 
5085 
100 
212 
1543 
5062 
110 
230 
1532 
5026 
120 
248 
1519 
4984 
130 
266 
1503 
4931 
140 
284 
1485 
4872 
150 
302 
1466 
4810 
160 
320 
1440 
4724 
170 
338 
1412 
4633 
180 
356 
1390 
4560 
190 
374 
1360 
4462 
200 
392 
1333 
4373 
220 
428 
1268 
4160 
240 
464 
1192 
3911 
260 
500 
1110 
3642 
Table 3.3 
Sound Speed in 
Liquid Water vs. 
Temperature 
Signal Strength 
VB.NET PDF copy, paste image library: copy, paste, cut PDF images
XDoc.PDF ›› VB.NET PDF: Copy, Paste, Cut Image in Page. VB.NET PDF - Copy, Paste, Cut PDF Image in VB.NET. VB.NET: Copy and Paste Image in PDF Page.
export text from pdf; edit pdf replace text
C# PDF copy, paste image Library: copy, paste, cut PDF images in
|. Home ›› XDoc.PDF ›› C# PDF: Copy, Paste, Cut Image in Page. C#.NET PDF SDK - Copy, Paste, Cut PDF Image in C#.NET. C# Guide
acrobat remove text from pdf; extract text from pdf open source
Rev.  01/06 
-3.27-   
HTTP 
Signal Strength Cutoff SIG C-OF is used to drive the 
flow  meter  and  its  outputs  to  the  SUB  FLOW 
(Substitute Flow– described below) state if conditions 
occur that cause low signal strength. A signal strength 
indication of between 0.5 and 0.8 is inadequate for 
measuring flow reliably, so minimum settings for SIG 
C-OF are in the range of 1.0 to 2.0. A good practice is 
to set the SIG C-OF at approximately 60-70% of ac-
tual measured signal strength (described above). 
If the measured signal strength is lower than the SIG 
C-OF setting, an ERROR 0010 will be displayed on 
the HTTP display until the measured signal strength 
becomes greater than the cutoff value. 
Substitute Flow or SUB FLOW is a value that the ana-
log  outputs  and  the  flow  rate  display  will  indicate 
when an error condition in the flow meter occurs. The 
typical setting for this entry is a value that will make 
the instrument display zero flow during an error condi-
tion.  
Table 3.4 below lists some typical settings to achieve 
“Zero” with respect to MIN and MAX FLOW settings. 
*UltraLink
is required to set values outside of 0.0-
100.0. 
PART 3 - KEYPAD CONFIGURATION 
Substitute Flow 
Entry 
MIN RATE 
SETTING 
MAX RATE 
SETTING 
SUB FLOW 
SETTING 
DISPLAY 
READING DURING 
ERRORS 
0.0 
1,000.0 
0.0 
0.000 
-500.0 
500.00 
50.0 
0.000 
-100.0 
200.0 
33.3 
0.000 
0.0 
1,000.0 
-5.0* 
-50.00 
TABLE 3.4—Substitute Flow Entry 
Signal Strength 
Cutoff 
C# PDF insert text Library: insert text into PDF content in C#.net
Text to PDF. C#.NET PDF SDK - Insert Text to PDF Document in C#.NET. Providing C# Demo Code for Adding and Inserting Text to PDF File Page with .NET PDF Library.
delete text from pdf acrobat; find and replace text in pdf
C# PDF Text Search Library: search text inside PDF file in C#.net
Text: Search Text in PDF. C# Guide about How to Search Text in PDF Document and Obtain Text Content and Location Information with .NET PDF Control.
.net extract pdf text; copy paste pdf text
Rev.  01/06 
-3.28-   
HTTP 
Because every flow meter installation is slightly differ-
ent and sound waves can travel in slightly different 
ways through these various installations, a provision 
is made in this entry to establish “Zero” flow—SET 
ZERO. 
To zero the meter: 
1.  The pipe must be full of liquid. 
2.  Flow must be absolute zero—securely close any 
valves and allow time for any settling to occur. 
3.  Press ENTER, use the arrow keys to make the 
display read YES. 
4.  Press ENTER. 
5.  The procedure is complete. 
If the flow in a piping system cannot be shut off, allow-
ing the SET ZERO procedure described above to be 
performed, then the  factory  default  zero  should be 
utilized. To utilize the D-FLT 0 function, simply press 
ENTER, then press an ARROW key to display YES 
on the display and then press ENTER. This function 
can also be utilized to correct an inadvertently entered 
or erroneous SET ZERO entry. 
This function can be used to make the HTTP system 
agree with a different or reference flow meter, by ap-
plying  a  correction  factor/multiplier  to  the  readings 
and outputs. A factory calibrated system should be 
set to 1.000. The range of settings for this entry is 
0.500 to 1.500. The following examples describe two 
uses for the COR FTR entry. 
• 
The  HTTP meter is indicating a flow rate that is 
4% higher than another flow meter located in the 
same pipe line. To make the HTTP indicate the 
same flow rate as the other meter, enter a COR 
FTR of 0.960, to lower the readings by 4%. 
PART 3 - KEYPAD CONFIGURATION 
Factory Default 
Zero Calibration 
Correction Factor 
Setting/Calibrating 
Zero Flow 
C# PDF Convert to Text SDK: Convert PDF to txt files in C#.net
C#.NET PDF SDK - Convert PDF to Text in C#.NET. Integrate following RasterEdge C#.NET text to PDF converter SDK dlls into your C#.NET project assemblies;
get text from pdf file c#; copy pdf text to word with formatting
Rev.  01/06 
-3.29-   
HTTP 
• 
An out-of-round pipe, carrying water, causes the 
HTTP to indicate a measured sound speed that is 
7.4% lower than the TABLE 3.3 value. This pipe 
condition will cause the flow meter to indicate flow 
rates that are 7.4% lower than actual flow. To cor-
rect the flow readings, enter 1.074. 
Allows the selection of a two line or four line display 
format on the graphics display module. 
In 2 Line mode, the display will display flow measure-
ments with larger characters on the top half of the 
window and smaller standard sized characters on the 
lower half of the window.  In 4 Line mode, the display 
will display flow measurements with standard sized 
characters on four lines in the window. 
DISPLAY UNITS SELECTION 
The  (soft)SELECT  key  is  used  to  configure  the 
engineering units on the graphics display—Press the 
(soft)SELECT key from  RUN  mode to  highlight  the 
engineering  unit  presently  being  displayed  on  the 
graphics display (pressing the SELECT key multiple 
times will toggle the highlighted unit from line to line).  
Use the UP/DOWN  ARROW keys to  select display 
units of 
• 
RATE 
• 
TOTALizer 
• 
VELocity 
• 
SIGNAL STRength 
PART 3 - KEYPAD CONFIGURATION 
7.  
DSP MENU 
-- DISPLAY MENU 
Display Units 
Graphics Display 
Mode 
Rev.  01/06 
-3.30-   
HTTP 
The LED backlighting on the HTTP is used to assist 
the operator in viewing the display in poorly lit areas—
the backlighting, when activated, doubles the power 
consumption of the flow meter.  If left on continuously, 
the charge in the battery will be depleted much more 
rapidly  than if  the backlighting is only activated for 
short  periods  of  time.    If  the  instrument  is  being 
operated  while  powered  from  an  external  power 
source, the back light may be left on permanently. 
Adjust  the  Back  Light  Timeout  to  approximate  the 
amount  of  seconds  that  the  backlighting  should 
remain active.  The time  out can be set anywhere 
between  10  and  30,000  seconds.    If  continuous 
backlighting is desired, set the Back Light Timeout to 
0 seconds. 
Back Light Timeout 
PART 3 - KEYPAD CONFIGURATION 
Rev.  01/06 
-4.1- 
HTTP 
PART 4 - SOFTWARE UTILITIES 
The  HTTP  flow  meter  can  be  used  with  two  software  utilities,       
UltraLink
and Data Logger. The UltraLink utility is used for con-
figuration, calibration and communication with the HTTP flow meter. 
The Data Logger utility is used for uploading and translating data  
accumulated in the data logger module located in the pocket on the 
front faceplate of the flow meter. 
UltraLink has been designed to provide a HTT user a powerful and 
convenient  way  to  configure  and  calibrate  HTTP  flow  meters.      
UltraLink can be used in conjunction with the infrared communica-
tions adapter included in the HTTP case or the optional ISO-MOD 
RS232 and ISO-MOD RS485. 
System Requirements
Computer type - PC, operating system - Windows
®
95/98/2000/NT, 
 communications  port  for  the  infrared  adapter,  access  to  the     
Hedland web site. 
Installation
1.  Go to www.hedland.com  
2.  Go to Products/Transit Time Ultrasonic/Software 
3.  Click the UltraLink
icon. 
4.  Follow downloading instructions. 
5. UlSetup will automatically extract and install on the hard disk 
and place a short-cut icon on the desktop. 
6.  Some PCs may require a restart after a successful installation. 
A CD of the UltraLink software can also be purchased by contacting 
Hedland sales at 800-433-5263 or 262-639-6770. Please refer to 
part number HTTP-086. 
1.  Connect the infrared communications adapter to a PC commu-
nication port and point the communicator at the HTTP infrared 
window, located in the lower right hand corner of the keypad. If 
meter is ordered with a ISO-MOD RS232 or ISO-MOD RS485, 
Important 
Notice! 
Initialization 
Rev.  01/06 
-4.2- 
HTTP 
connect the PC Communications port directly to the optionally 
installed RS232C or RS485 port located on side of HTTP meter. 
2.  The first screen is the “RUN-mode” screen, See Figure 4.1, 
which contains real-time information regarding flow rate, total-
izer accumulation, system signal strength, diagnostic data and 
the flow meter’s serial number. The indicator in the lower right-
hand  corner  will  indicate  communications  status.  If  a  red       
ERROR is indicated, click on the Communications button on the 
top bar. Click on Initialize. Choose the appropriate COM port 
and  interface  type.  Proper  communications  are  established 
when a green OK is indicated in the lower right-hand corner of 
the PC display. 
Figure 4.1 
UltraLink
Data Screen 
Data Trend Minutes 
Data Trend Flow Rate 
PART 4 - SOFTWARE UTILITIES 
Notes:  The range of the infrared communications adapter 
is roughly 3 meters. Some high-intensity lighting systems 
will reduce the infrared communications range. 
Documents you may be interested
Documents you may be interested