Pre-Sample
Using TestWorks 4
Test Flow / Test Segment Configuration
61
Pre-Sample
This section contains the test segments that will be performed when the first 
specimen is run in a new sample.
Required inputs
This test segment will prompt you for the inputs that are listed in the table. 
However, in the MTS templates (methods provided with software), the are 
no inputs listed. To add an input to the table, you must edit the desired input 
and change its When setting to Pre-Sample and place a check mark beside 
the Required Input option. These inputs deal with information pertaining to 
the sample. For example, Sample ID.
To modify this test segment, either double-click on an existing Required 
Inputs Test Segment (or right-click and select Edit) or right-click and select 
Add/Remove or Create New. Complete the following steps:
1. Place a check mark inside the Enable box to activate the required 
inputs test segment. You can also select Enable/Disable by right-
clicking on the Required Inputs Test Flow parameter.
2. To add inputs within this test segment, right-click within the Required 
Inputs Test Segment dialog and select Add/Remove Required Input. 
The following dialog enables you add or remove required inputs from 
this test segment. Just select the inputs desired to be added or removed 
from choice list for the method.
3. To create an input please refer to “Inputs” on page 87.
Note
The Unattended check box is for Robot Systems only.
Note
If applicable, the Strings Cannot Be Empty checkbox when 
checked indicates that a value (text) must be entered. 
Cannot select text in pdf - search text inside PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Learn how to search text in PDF document and obtain text content and location information
how to make a pdf file text searchable; search pdf documents for text
Cannot select text in pdf - VB.NET PDF Text Search Library: search text inside PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Learn How to Search Text in PDF Document and Obtain Text Content and Location Information in VB.NET application
search pdf files for text; search text in multiple pdf
Using TestWorks 4
Pre-Sample
Test Flow / Test Segment Configuration
62
Formulas
This test segment will calculate the formulas that are listed in the table. To 
add a formula to the table, you must edit the desired formula and change its 
When setting to Pre-Sample. A typical formula used in this section would be 
the SampleName formula. This configures the test method to utilize the 
SampleID sample information input to automatically name your sample file 
when you test the first specimen in the sample.
To modify this test segment, either double-click on the Formulas or right- 
click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the Formulas test 
segment.
2. To add inputs within this test segment, right click within the Edit 
Formulas dialog and select Add/Remove Formula. The following 
dialog enables you Add or Remove formulas from this test segment. 
Just select the inputs desired to be added or removed from the choice 
list for the method.
3. To create a custom formula, please refer to “Formula Builder - 
(Creating Formulas)” on page 111.
Verify sample name
This test segment evaluates the sample name entered by the user and 
provides a warning message if the file already exists on disk or if the sample 
name contains an invalid character that will cause an error when saving the 
file. This segment is only useful if the test method has been set up so that the 
formula whose internal name is SampleName has been enabled.
This test segment is only valid in the Pre-Sample section of the test flow. It is 
typically executed after the pre-sample formulas have been calculated.
To modify this test segment, either double-click on the Verify Sample Name 
or right-click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the Verify Sample 
Name test segment.
2. Type a run-time message into the Run-Time Message text box.
C# HTML5 Viewer: Deployment on AzureCloudService
RasterEdge.XDoc.PDF.dll. RasterEdge.XDoc.PDF.HTML5Editor.dll. Or you can select x86 if you use x86 dlls. (The application cannot to work without this node.).
search a pdf file for text; how to select text in pdf reader
C# HTML5 Viewer: Deployment on ASP.NET MVC
RasterEdge.XDoc.PDF.HTML5Editor.dll. When you select x64 and directly run the application, you may get following error. (The application cannot to work without
select text in pdf reader; pdf make text searchable
Pre-Specimen
Using TestWorks 4
Test Flow / Test Segment Configuration
63
Pre-Specimen
This section contains test segments that will be performed for each specimen 
that is tested.
Play WAV file
This test segment plays the WAV file that is specified. A WAV test segment 
can be used to play a file that provides verbal instructions for a test 
procedure.
To modify this test segment, either double-click on the Play WAV File or 
right click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the WAV File test 
segment.
2. Click on the Browse button to load the specific WAV file.
3. Type a run-time message into the Run-Time Message text box.
Required inputs
This test segment will prompt you for the inputs that are listed in the table. 
These inputs are generally used to describe each specimen within a sample. 
For example, Width and Thickness would be required inputs (Pre-Specimen) 
because these inputs would vary from specimen to specimen.
To modify this test segment, either double-click on an existing Required 
Inputs Test Segment (or right-click and select Edit) or right-click and select 
Add/Remove or Create New. Complete the following steps:
1. Place a check mark inside the Enable box to activate the required 
inputs test segment. You can also select Enable/Disable by right-
clicking on the Required Inputs Test Flow parameter.
2. To add inputs within this test segment, right-click within the Required 
Inputs Test Segment dialog and select Add/Remove Required Input. 
The resultant dialog allows you to add or remove required inputs from 
this test segment. Just select the inputs desired to be added or removed 
from choice list for the method.
3. To create a custom input please refer to “Inputs” on page 87.
Note
The Unattended check box is for Robot Systems only.
Note
If applicable, the Strings Cannot Be Empty checkbox when 
checked indicates that a value (text) must be entered. 
C# PDF: PDF Document Viewer & Reader SDK for Windows Forms
Choose Items", and browse to locate and select "RasterEdge.Imaging open a file dialog and load your PDF document in will be a pop-up window "cannot open your
find text in pdf files; pdf search and replace text
C# Image: How to Deploy .NET Imaging SDK in Visual C# Applications
RasterEdge.Imaging.MSWordDocx.dll; RasterEdge.Imaging.PDF.dll; in C# Application. Q: Error: Cannot find RasterEdge Right click on projects, and select properties.
pdf find and replace text; how to make pdf text searchable
Using TestWorks 4
Pre-Specimen
Test Flow / Test Segment Configuration
64
Test message
This test segment displays a message dialog that can be customized to 
display until you respond to it or to display for a fixed amount of time. 
Typically, this segment is used to convey instructions to the user who is 
running a test. For instance, you might create a message that says “Have you 
calibrated your load cell? Click Cancel if you have not.”
To modify this test segment, complete the following steps:
1. Place a check mark inside the Enable box to activate the Test Message 
Test Segment.
2. Type the desired message into the Message Text box.
3. Click on the Wait for user response option if you want TestWorks 4 to 
display your message until you respond to it.
Click on the OK dialog option if you want merely an OK response 
available within the test message.
Click on the OK/Cancel dialog option if you want OK and Cancel 
responses available within the test message.
4. Click on the Pause for specified time option if you want TestWorks 4 
to display the message for a specified amount of time.
Type the desired amount of display time into the Pause Time text 
box.
Place a check mark inside the Include Resume button box if you 
want a Resume button to appear within the message box. This 
button allows you to force the test to continue before the display 
time has expired.
Tare (channel)
This test segment resets your selected data channel to 0. It is commonly 
used to zero the crosshead or actuator rod position, or the time before a test 
begins.
To modify this test segment, complete the following steps:
1. Place a check mark inside the Enable box to activate the Tare Test 
Segment.
2. Click on the Channel combo-box to select the data channel that will be 
reset to 0.
3. Type a run-time message into the Run-Time Message text box.
GIF to PNG Converter | Convert GIF to PNG, Convert PNG to GIF
Imaging SDK; Save the converted list in memory if you cannot convert at Select "Convert to PNG"; Select "Start" to start conversion procedure; Select "Save" to
how to make a pdf document text searchable; search text in pdf using java
C# PowerPoint: Document Viewer Creating in Windows Forms Project
You can select a PowerPoint file to be loaded into the WinViewer control. is not supported by WinViewer control, there will prompt a window "cannot open your
searching pdf files for text; pdf find highlighted text
Pre-Specimen
Using TestWorks 4
Test Flow / Test Segment Configuration
65
Formulas
This test segment will calculate the formulas that are listed in the Edit 
Formula table. To add a formula to the table, you must edit the desired 
formula and change its When setting to Pre-Specimen. Some typical 
formulas used in this section would be Area, Load Cell Capacity, and 
Adjusted Gage Length.
To modify this test segment, either double-click on the Formulas Display 
Name or right-click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the Formulas 
Test Segment.
2. To add formulas within this test segment, right-click within the 
Formulas dialog and select Add/Remove Formula. The resultant 
dialog allows you to add or remove formulas from this test segment. 
Just select the inputs desired to be added or removed from choice list 
for the method.
3. To create a custom formula, please refer to“Formula Builder - (Creating 
Formulas)” on page 111.
C# Image: Create C#.NET Windows Document Image Viewer | Online
DeleteAnnotation: Delete all selected text or graphical annotations. You can select a file to be loaded into the there will prompt a window "cannot open your
pdf text search tool; how to select all text in pdf file
C# Image: How to Use C# Code to Capture Document from Scanning
installed on the client as browsers cannot interface directly a multi-page document (including PDF, TIFF, Word Select Fill from the Dock property located in
pdf editor with search and replace text; pdf text search
Using TestWorks 4
Specimen
Test Flow / Test Segment Configuration
66
Specimen
This section contains the test segments that perform the physical testing of 
the specimen. Data can only be collected within this section.
Note
Please note that some test segments can appear within your 
method. This manual covers the typical method settings. If you need 
help configuring a test segment not covered, please consult MTS.
Set clutch
This test segment changes the machine clutch to the setting that you have 
selected.
Note
Some EM machines and all servohydraulic and static hydraulic 
machines will not have a clutch. To verify a clutch machine, check 
for the lever icon in the lower right-hand corner of the TestWorks 
software.
To modify this test segment, complete the following steps:
1. Place a check mark inside the Enable box to activate the Set Clutch 
Test Segment.
2. Click on the Clutch Value combo-box to select the clutch to be set.
3. Type a run-time message into the Run-Time Message text box.
Tare (channel)
To configure refer to “Tare (channel)” on page 64 in the Pre-Specimen 
section. For MTS method templates, the Tare is defaulted to zero the Time 
channel. This should not be altered – so each specimen run will have 
accurate time values.
GoTo
These are the instructions for Editor/Definer/Creator Level Users using the 
Edit Method Menu Option.
Test Segment
This test segment is designed to move the crosshead or actuator rod at a 
specified rate. The segment can stop at a specified value or go forever until 
break (or some other limit) is detected. The dialog is divided into a few 
portions:
The Rate Value will always appear.
The Stop Value will only appear if the test segment has been 
configured by a Definer/Creator-Level user to use a Stop Value.
C# Word: How to Create C# Word Windows Viewer with .NET DLLs
and browse to find and select RasterEdge.XDoc control, there will prompt a window "cannot open your powerful & profession imaging controls, PDF document, tiff
can't select text in pdf file; convert a scanned pdf to searchable text
C# Excel: View Excel File in Window Document Viewer Control
Items", and browse to find & select WinViewer DLL; there will prompt a window "cannot open your powerful & profession imaging controls, PDF document, image
pdf searchable text converter; how to select text in pdf and copy
Specimen
Using TestWorks 4
Test Flow / Test Segment Configuration
67
To modify this test segment, double-click on the GoTo segment or right-
click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the GoTo tests    
segment (or right-click on GoTo and select Enable).
2. Type in the desired speed of the crosshead or actuator rod for this GoTo 
test segment in the Rate Value edit box. Use the Units combo box to 
assign the correct units.
Important
The GoTo command moves the crosshead or actuator rod to a 
position that is an absolute distance from the user defined zero 
position; not a distance relative to the current crosshead or 
actuator rod position.
3. If the Stop Value section is displayed, then type in the desired stopping 
condition for the GoTo test segment in the edit box and select the 
appropriate units using the Units combo box. Keep in mind that Stop 
Value is an absolute value relative to the user defined zero position; not 
a distance value relative to the current crosshead or actuator rod 
position.
Using TestWorks 4
Specimen
Test Flow / Test Segment Configuration
68
4. If the GoTo test segment has been configured to use a rate control 
channel other than Crosshead/Actuator rod, the segment might 
require PID parameters to be tuned. PID parameters are parameters 
which govern how the testing equipment moves to control based on an 
Advanced Rate Control channel. If PID tuning is required, a Gain edit 
box and an edit PID Parameters button are enabled. To tune the kP 
factor of the PID parameters, enter a value into the Gain edit box. To 
adjust the rest of the PID parameters, click the edit PID Parameters 
button. This is for advanced testing and should be altered by advanced 
users.
Data collection
By clicking on the Data Collection button, a dialog is displayed that allows 
you to configure how TestWorks 4 collects data. You can not change how 
TestWorks 4 collects data (e.g. Data Rate vs. Level Crossing), but you can 
change the particulars of the collection method (e.g. the data rate or the 
magnitude of level change). You can have up to five tabs to collect data for a 
method. Each tab can collect data in a different manner (e.g. one tab for Data 
Rate and one tab for Level Crossing).
*Quick Tip: In Data Rate mode, to decrease the amount of data TestWorks 4 
collects during a test segment, decrease the Data Acquisition Rate.
There are three types of data collection in TestWorks 4. They are as follows:
Data Rate - TestWorks 4 will continuously collect data at a rate you specify. 
The rate is in units of Hertz, or points per second. For example, if your Data 
Rate Value is set 10 Hertz, then TestWorks 4 will collect 10 data points per 
second while the test segment is executed.
Peak/Valley - TestWorks 4 will collect any data point that is a minimum or a 
maximum. A change in data direction determines if the point is a minimum 
or a maximum value. Consider the following set of points: 
S1={10,11,10,9,10}. Point 2 (value of 11) is considered a maximum, 
because it is greater than the points on either side of it. Point 4 (value of 9) is 
considered a minimum, because it is less than the points on either side of it. 
Any change in direction denotes a minimum or maximum, regardless of how 
big the change.
Level Crossing - TestWorks 4 will collect a data point when a channel 
increases or decreases by a certain amount. For example, if the Level 
Crossing channel is Load and the Delta Value is 10 lbf, then TestWorks 4 
will collect a data point every time the Load channel increases or decreases 
by 10 lbf.
Specimen
Using TestWorks 4
Test Flow / Test Segment Configuration
69
Also, at the bottom, you can enable the Take Data on Pause check box to 
continue collecting data when you have manually paused the crosshead or 
actuator rod with the Pause button.
Break detection
By clicking on Break Detection, you are allowed to edit the break detection 
task which allows TestWorks 4 to determine when a specimen has broken. It 
looks for a drop in load that is a certain percentage of peak load. When the 
load drops below that level, TestWorks 4 assumes the specimen has failed 
and performs the commanded action.
Drop Value - Allows you to configure the magnitude of drop from peak that 
TestWorks 4 looks for to determine break. The higher the number, the more 
the load has to drop from peak before TestWorks 4 recognizes the break. The 
display name of the variable used to determine the Drop Value is displayed 
in the Drop Value section. To change the value of the Drop Value variable, 
simply type in the value in the edit box. You can also change the units of the 
Drop Value variable by using the Units combo box.
Note
If TestWorks 4 recognizes specimen failure before the specimen is 
physically broken, then increase the Drop Value.
Action - Determines what action TestWorks 4 will take when it detects 
specimen failure. There are three choices:
Abort Test - will end the current test with no more test segments 
being executed.
Next Test Segment - executes the next test segment in the Test 
Flow.
Post-Specimen - will cause TestWorks 4 to skip all remaining test 
segments in the Specimen section of the Test Flow and to then 
execute the test segments in the Post-Specimen section of the Test 
Flow.
Threshold Value - Allows you to configure what load value must be reached 
before TestWorks 4 will start looking for break. This allows the “noisy” 
region at the beginning of the test to be excluded from break detection. For 
instance, if the Threshold Value is set to 5 lbf, then the load must become 
greater than 5 lbf before TestWorks 4 will begin looking for specimen 
failure. The display name of the Threshold Value variable appears in the 
Threshold Value section. To change the value of the Threshold Value 
variable, simply type the value into the edit box. You can also change the 
units of the Threshold Value variable by using the Units combo box.
Note
If TestWorks 4 detects specimen failure before there is significant 
load on the specimen, then increase the Threshold Value.
Using TestWorks 4
Specimen
Test Flow / Test Segment Configuration
70
Hold
This test segment is designed to move the crosshead or actuator rod to 
maintain a specified value on a channel. The test segment can be configured 
to hold a channel value for a specific time limit or until break (or some other 
limit) is detected.
To modify this test segment, double-click on the Hold test segment or right-
click and select Edit. Complete the following steps:
1. Place a check mark inside the Enable box to activate the Hold test 
segment.
2. The Hold Value group will always appear. The display name of the 
control channel will be displayed in this group. If this Hold test segment 
has been configured to hold the current value of the control channel, 
then no edit box will appear. In this case, the text “Hold Current Value” 
will appear in the Hold Value group. If an edit box is visible, then the 
value of the control channel that TestWorks 4 will hold, can be changed 
by typing the value into the edit box. The units of the Hold Value 
variable can be changed by using the Units combo box.
*Quick Tip: To change the value that TestWorks 4 will hold while this test 
segment is executed, change the number in the Hold Value group's edit box.
3. If the hold test segment has been configured, by a Define-Level User, to 
stop holding the channel value at a certain limit, then the Stop Value 
group will appear. The display name of the Stop Value variable will 
appear in this group. To change the value of the Stop Value variable, 
simply type the value into the edit box. You can change the units of the 
Stop Value variable by using the Units combo box.
4. If the control channel is one other than the primary control channel for 
the system (usually the crosshead/actuator channel) then PID Parameter 
tuning is required. The first level of PID tuning is to adjust the Gain 
Factor (kP) by typing a number into the Gain edit box. The higher the 
Gain Factor, TestWorks 4 will react more quickly to changes in the 
control channel. If adequate control can not be attained by adjusting the 
Gain Factor, then the other PID parameters must be tuned.
5. To change how TestWorks 4 collects data while this hold test segment is 
executed, click the Data Collection button. (Refer to the GoTo “Test 
Segment” on page 66 for these instructions.)
6. To change how TestWorks 4 detects break (specimen failure) while this 
hold test segment is executed, click the Break Detection button. (Refer 
to the GoTo “Test Segment” on page 66 for these instructions.)
7. Type a run-time message into the Run-Time Message text box.