A: Calibrating the Wavelength of the QE Pro 
891-00000-000-02-201603 
23 
Calibrating the Spectrometer  
Preparing for Calibration  
To recalibrate the wavelength of your spectrometer, you need the following components: 
A light source capable of producing spectral lines  
Note 
Ocean Optics’ HG
-1 Mercury-Argon lamp is ideal for recalibration. If you do not have an 
HG-1, you need a light source that produces several (at least 4-6) spectral lines in the 
wavelength region of your spectrometer. 
A QE Pro  spectrometer 
An optical fiber (for spectrometers without a built-in slit, a 50-
m fiber works best)  
A spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-
order linear regressions 
Note 
If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak 
and AnalysisTookPak-VBA
Calibrating the Wavelength of the Spectrometer 
Procedure 
Perform the steps below to calibrate the wavelength of the spectrometer: 
1.  Place the spectrometer operating software into Quick View mode in OceanView and take a 
spectrum of your light source. Adjust the integration time (or the A/D conversion frequency) until 
there are several peaks on the screen that are not off-scale. 
2.  Move the cursor to one of the peaks and position the cursor so that it is at the point of maximum 
intensity.  
3.  Record the pixel number that is displayed in the status bar or legend (located beneath the graph). 
Repeat this step for all of the peaks in your spectrum. 
4.  Use the spreadsheet program or calculator to create a table like the one shown in the following 
figure. In the first column, place the exact or true wavelength of the spectral lines that you used.  
In the second column of this worksheet, place the observed pixel number. In the third column, 
calculate the pixel number squared, and in the fourth column, calculate the pixel number cubed. 
Find and replace text in pdf - search text inside PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Learn how to search text in PDF document and obtain text content and location information
search pdf documents for text; make pdf text searchable
Find and replace text in pdf - VB.NET PDF Text Search Library: search text inside PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Learn How to Search Text in PDF Document and Obtain Text Content and Location Information in VB.NET application
pdf find and replace text; pdf select text
A: Calibrating the Wavelength of the QE Pro 
24 
891-00000-000-02-201603 
True Wavelength (nm) 
Pixel # 
Pixel # 
2
Pixel # 
3
Predicted 
Wavelength 
Difference 
253.65 
296.73 
302.15 
313.16 
334.15 
365.02 
404.66 
407.78 
435.84 
546.07 
576.96 
579.07 
696.54 
706.72 
727.29 
738.40 
751.47 
175 
296 
312 
342 
402 
490 
604 
613 
694 
1022 
1116 
1122 
1491 
1523 
1590 
1627 
1669 
30625 
87616 
97344 
116964 
161604 
240100 
364816 
375769 
481636 
1044484 
1245456 
1258884 
2223081 
2319529 
2528100 
2647129 
2785561 
5359375 
25934336 
30371328 
40001688 
64964808 
117649000 
220348864 
230346397 
334255384 
1067462648 
1389928896 
1412467848 
3314613771 
3532642667 
4019679000 
4306878883 
4649101309 
253.56 
296.72 
302.40 
313.02 
334.19 
365.05 
404.67 
407.78 
435.65 
546.13 
577.05 
579.01 
696.70 
706.62 
727.24 
738.53 
751.27 
0.09 
0.01 
-0.25 
0.13 
-0.05 
-0.04 
-0.01 
0.00 
0.19 
-0.06 
-0.09 
0.06 
-0.15 
0.10 
0.06 
-0.13 
0.19 
5.  Use the spreadsheet or calculator to calculate the wavelength calibration coefficients. In the 
spreadsheet program, find the functions to perform linear regressions. 
If using Quattro Pro, look under Tools | Advanced Math 
If using Excel, look under Analysis ToolPak 
6.  Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number 
squared, and the pixel number cubed as the independent variables (X). After executing the 
regression, you will obtain an output similar to the one shown below. Numbers of importance are 
noted. 
Regression Statistics 
Multiple R 
0.999999831  
R Square 
0.999999663 
R Squared 
Adjusted R Square  0.999999607 
Standard Error  0.125540214 
Observations  22 
Intercept 
Coefficients  Standard Error 
Intercept 
190.473993 
0.369047536 
First coefficient 
X Variable 1 0.36263983 
0.001684745 
X Variable 2-1.174416E-05 
8.35279E-07 
X Variable 3-2.523787E-09  2.656608E-10 
Second coefficient 
Third coefficient 
Independent 
Variable 
Dependent 
Variables 
Values Computed 
from the Regression 
Output 
VB.NET PDF replace text library: replace text in PDF content in vb
and ASP.NET webpage. Find and replace text in PDF file in preview without adobe PDF reader component installed. Able to pull text
can't select text in pdf file; how to make a pdf document text searchable
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
When you have downloaded the RasterEdge Image SDK for .NET, you can unzip the package to find the RasterEdge.Imaging.PDF.dll in the bin folder under the root
make pdf text searchable; search pdf files for text
A: Calibrating the Wavelength of the QE Pro 
891-00000-000-02-201603 
25 
7.  Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at 
the value for R squared. It should be very close to 1. If not, you have most likely assigned one of 
your wavelengths incorrectly.  
Keep these values at hand. 
Saving the New Calibration Coefficients: USB 
Mode 
Ocean Optics programs wavelength calibration coefficients unique to each QE Pro  onto an EEPROM 
memory chip in the QE Pro .  
You can overwrite old calibration coefficients on the EEPROM if you are using the QE Pro  via the USB 
port.  
Procedure 
To save wavelength calibration coefficients using the USB mode, perform the following steps: 
1.  Ensure that the QE Pro  is connected to the PC and that you have closed all other applications. 
2.  Point your browser to http://oceanoptics.com/support/software-downloads/ and scroll down to 
Utility Programs. Select the QE Pro Update Tool for your operating system. 
3.  Save the files to your computer and extract them from the zip file.  
4.  Double-click the UpdateQE-PRO.jar file to run the software. 
5.  Click on the desired QE Pro  device displayed in the Devices pane. 
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
document. If you find certain page in your PDF document is unnecessary, you may want to delete this page directly. Moreover, when
pdf search and replace text; pdf editor with search and replace text
VB.NET Create PDF Library SDK to convert PDF from other file
Create writable PDF file from text (.txt) file in VB.NET To get this RasterEdge PDF document file creator library the RasterEdge .NET Image SDK and find the PDF
searching pdf files for text; how to search text in pdf document
A: Calibrating the Wavelength of the QE Pro 
26 
891-00000-000-02-201603 
6.  Double-click on each of the calibration coefficients displayed in the Wavelength Coefficients 
pane of the USB Programmer screen and enter the new values acquired in Step 5 of the 
Calibrating the Wavelength of the Spectrometer section in this appendix. 
7.  Repeat Step 6 for all of the new values. 
8.  Click on the Write Coefficients to Device button to save the information, and then select File| 
Exit to exit the QE Pro Reflash Utility software. 
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the QE 
Pro
C# HTML5 PDF Viewer SDK deployment on Visual Studio .NET
C#.NET rotate PDF pages, C#.NET search text in PDF, C# Unzip the download package and you can find a project Once done debugging with x86 dlls, replace the x86
text searchable pdf; search text in pdf using java
VB.NET PDF - Deploy VB.NET HTML5 PDF Viewer on Visual Studio.NET
C#.NET rotate PDF pages, C#.NET search text in PDF, C# Unzip the download package and you can find a project Once done debugging with x86 dlls, replace the x86
how to select all text in pdf; search pdf documents for text
891-00000-000-02-201603 
27 
Appendix B 
Specifications 
Overview 
This appendix contains information on spectrometer operation, specifications, and system compatibility. 
It also includes accessory connector pinout diagrams and pin-specific information. 
How the QE Pro Works 
Below is a diagram of how light moves through the optical bench of a QE Pro  Spectrometer. The optical 
bench has no moving parts that can wear or break; all the components are fixed in place at the time of 
manufacture.  
QE Pro Spectrometer with Components 
C# PDF File Permission Library: add, remove, update PDF file
Text: Replace Text in PDF. Image: Insert Image to PDF. Image: Remove read PDF, VB.NET convert PDF to text, VB.NET the following code table, you will find a piece
convert pdf to searchable text online; pdf text select tool
VB.NET PDF File Permission Library: add, remove, update PDF file
Text to PDF. Text: Delete Text from PDF. Text: Replace Text in PDF. NET rotate PDF pages, C#.NET search text in PDF In the following code table, you will find a VB
convert pdf to searchable text; how to select all text in pdf file
B: Specifications 
28 
891-00000-000-02-201603 
QE Pro Components Table 
Ocean Optics permanently secures all components in the QE Pro  at the time of manufacture. Only Ocean 
Optics technicians can replace interchangeable components, where noted. 
Item  Name 
Description 
SMA Connector 
Secures the input fiber to the spectrometer. Light from the input fiber enters the 
optical bench through this connector. 
Interchangeable 
Slit 
A dark piece of material containing a rectangular aperture, which is mounted 
directly behind the SMA Connector. The size of the aperture regulates the 
amount of light that enters the optical bench and controls spectral resolution. 
You can also use the QE Pro without a Slit. In this configuration, the diameter 
of the fiber connected to the QE Pro determines the size of the entrance 
aperture. 
The user can change the slit. Ocean Optics also offers a range of FC 
connector slits in the same wavelengths, with the product code INTFC-XXX. 
An INTFC-KIT is also available. 
Filter 
Restricts optical radiation to pre-determined wavelength regions. Light passes 
through the Filter before entering the optical bench. Both bandpass and 
longpass filters are available to restrict radiation to certain wavelength regions.  
Only Ocean Optics technicians can change the Filter. 
Collimating 
Mirror 
Focuses light entering the optical bench towards the Grating of the 
spectrometer. 
Light enters the spectrometer, passes through the SMA Connector, Slit, and 
Filter, and then reflects off the Collimating Mirror onto the Grating. 
Grating 
Diffracts light from the Collimating Mirror and directs the diffracted light onto 
the Focusing Mirror. Gratings are available in different groove densities, 
allowing you to specify wavelength coverage and resolution in the 
spectrometer.  
Only Ocean Optics technicians can change the Grating. 
Focusing Mirror 
Receives light reflected from the Grating and focuses the light onto the CCD 
Detector or L2 Detector Collection Lens (depending on the spectrometer 
configuration). 
Back-thinned 
Area Detector 
with TE Cooling 
Provides 90% quantum efficiency and bins pixels in a vertical column to 
acquire light form the entire height of the spectrometer’s slit image. This 
improves light collection and signal-to-noise significantly. This 2D area detector 
is back-thinned (back-illuminated) and does not require the detector upgrade 
that is normally applied to other detectors. 
Only Ocean Optics technicians can add or remove the Detector. 
Detector  
with OFLV Filter 
Eliminates second-order effects and is used with an HC-1 Grating in a 200-950 
nm wavelength system in a QE Pro
B: Specifications 
891-00000-000-02-201603 
29 
QE Pro Specifications 
The following sections provide specification information for the CCD detector in the QE Pro , as well as 
the QE Pro  Spectrometer itself. 
CCD Detector Specifications 
Specification 
Value 
Detector 
Hamamatsu S7031-1006 Low Etaloning 
Detector range 
185-1100 nm 
Pixels 
1024 active 
Pixel size 
24 μm
2
Quantum efficiency 
90% peak; 65% at 250 nm. See Quantum Efficiency of Detector graph 
below. 
QE Pro Spectrometer Specifications 
Specification 
Value 
Integration time 
8 ms to 60 minutes 
Typical dynamic range 
~85,000:1 
Signal-to-noise ratio 
1000:1 (single acquisition) 
Readout Noise 
2.5 counts RMS (typical) 
Linearity 
Corrected 
0.5% nonlinearity (max) 
f number 
f/4 
Activation time of optional internal 
shutter 
11 ms 
Trigger jitter 
10 ns 
Fiber optic connector 
SMA 905 and Ocean Optics FC 
Gratings 
14 gratings available (H1--H14), 
HC1 grating
B: Specifications 
30 
891-00000-000-02-201603 
Specification 
Value 
Entrance aperture 
5, 10, 25, 50, 100 or 200 µm wide slits 
(or SMA/FC bulkhead with no 
slit)
Optical resolution (FWHM) 
Depends on grating and size of entrance aperture 
OFLV filters 
OFLV-QE (200-950 nm); OFLV-QE-250 (250-1000 nm); OFLV-
QE-300 (300-1050 nm); OFLV-QE-350 (350-1100 nm); OFLV-QE-
400 (400-1150 nm) 
Temperature 
Operation 
TE Cooler can only cool 15 to 20 °C below ambient temperature 
0 °C to 50 °C 
Humidity 
≤ 
90% noncondensing 
Power consumption 
TEC On 
@ 40° below ambient
TEC Off  
15W (Max) 
2.5W (Max, Typical) 
Supply Voltage 
4.5 
5.5 V 
Power-up Time 
7 seconds 
Interfaces 
USB 2.0, 480 Mbps (USB 1.1 compatible); RS-232 (5-wire) 
Dimensions (LxWxH) 
182 mm (7.16 in.) x 110 mm (4.33 in.) x 47 mm (1.85 in.) 
Weight 
QE Pro 
Power Supply 
1.15 kg (2.6 lbs.) 
0.45 kg (1 lb.) 
System Compatibility 
You can use the QE Pro
’s USB connectivity with any computer that meets the 
requirements for the 
spectrometer operating software being used (Windows XP/Vista/Windows 7/8, Mac OS X and Linux). 
See About Ocean Optics Software
Alternately, the QE Pro  has a serial port for connecting to PCs, PLCs, and other devices with a Windows 
32-bit operating system that support the RS-232 communication protocol. However, this connection 
method requires an external power supply to power the QE Pro , the HR4000 Breakout Box, and a serial 
cable. 
B: Specifications 
891-00000-000-02-201603 
31 
30-Pin Accessory Connector Pinout 
The QE Pro  features a 30-pin Accessory Connector, located on the side of the unit as shown: 
Location of QE Pro 30-Pin Accessory Connector 
30-Pin Accessory Connector Pinout Diagram 
When facing the 30-pin Accessory Connector on the front of the vertical wall of the QE Pro , pin 
numbering is as follows: 
USB 
Port 
 10  12  14  16  18  20  22  24  26  28  30 
30 
 11  13  15  17  19  21  23  25  27  29 
29 
30-Pin Accessory Connector Pinout Diagram 
30-Pin Accessory Connector – Pin Definitions and 
Descriptions 
The following table contains information regarding the function of each pin in the QE Pro
’s 30
-Pin 
Accessory Connector: 
Pin #  Function 
Input/Output  Description 
RS232 Rx 
Input 
RS232 receive signal 
Communicates with a PC over DB9 
Pin 3 
B: Specifications 
32 
891-00000-000-02-201603 
Pin #  Function 
Input/Output  Description 
RS232 Tx 
Output 
RS232 transmit signal 
Communicates with a PC over DB9 
Pin 2 
GPIO (2) 
Input/Output  General Purpose Input Output 
Unused 
Ground  
Input/Output  Ground 
I2C SCL 
Input/Output  I2C clock signal for communication to other I2C peripherals 
GPIO (0) 
Input/Output  General Purpose Input Output 
I2C SDA 
Input/Output  I2C data signal for communication to other I2C peripherals 
GPIO (1) 
Input/Output  General Purpose Input Output 
10 
Ext. Trigger 
In 
Input 
CMOS input trigger tolerant from 3-5V 
11 
GPIO (3) 
Input/Output  General Purpose Input Output 
12 
VOUT  
Output 
Output power pin for QE Pro 
13 
SPI_MOSI  
Output 
SPI Master Out Slave In (MOSI) signal for communication to 
other SPI peripherals 
14 
VOUT 
Output  
Output power pin for QE Pro 
15 
SPI MISO  
Input 
SPI Master In Slave Out (MISO) signal for communication to 
the other SPI peripherals 
16 
GPIO (4)* 
Input /Output  General Purpose Input Output 
17 
Single Strobe   Output 
CMOS (3.3V) output pulse used as a strobe signal 
Has a 
programmable delay relative to the beginning of the 
spectrometer integration period  
18 
GPIO (5) 
Input/Output   General Purpose Input Output 
19 
SPI Clock 
Output 
SPI clock signal for communication to other SPI peripherals 
20 
Continuous 
Strobe 
Output 
CMOS output signal used to pulse a strobe 
Divided down 
from the master clock signal 
21 
SPI CS 
Output 
External SPI chip select (active low) 
22 
GPIO (6) 
Input/Output   General Purpose Input Output 
Documents you may be interested
Documents you may be interested