Appendix E. GS-800
E-9
Fig. E-8. Entering a custom resolution (with Oversample selected).
Image File Size
The size of the scan file for the selected resolution is listed next to Image File Size. If 
you do not have enough computer memory for the specified file size, an error 
message will appear when you attempt to scan. If this happens, select a lower 
resolution or decrease the size of the area to be scanned. (Macintosh users can also 
increase the application memory partition. See your Macintosh computer 
documentation for guidance.)
E.5 Calibration
The GS-800 automatically calibrates the densities of scanned images to ensure 
accuracy and reproducibility of results. The GS-800 has built-in step tablets for both 
transmissive and reflective scanning.
The transmissive step tablet calibrates transmissive scans from 0 to 3.0 O.D.
The reflective step tablet targets reflective scans from 0 to 2.0 O.D.
The first time you use the GS-800, you must select some settings to ensure accurate 
calibration. Click on the Options button in the main acquisition window. This will 
open the Densitometer Options dialog box.
How to extract a picture from a pdf - Select, copy, paste PDF images in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document
extract images from pdf c#; extract image from pdf file
How to extract a picture from a pdf - VB.NET PDF Image Extract Library: Select, copy, paste PDF images in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document
extract images from pdf acrobat; pdf extract images
Quantity One User Guide
E-10
Fig. E-9. Densitometer Options dialog box.
Note that calibration is always on for the GS-800.
E.5.a Step Tablet Values
The built-in transmissive and reflective step tablets on your densitometer have 
specific density values. Before scanning for the first time, you must enter the values 
for your transmissive step tablet into the software. You can use the default values in 
the software for the reflective step tablet.
Transmissive Step Tablet
Before you scan in transmissive mode for the first time, you must specify the values 
for your transmissive step tablet.
First, make sure that the Transmissive checkbox is selected in the main acquisition 
window, then click on the Options button. In the Densitometer Options dialog, click 
on the Edit Step Tablet button. The Step Tablet Values for GS-800 Transmissive 
dialog box will open.
C# PDF insert image Library: insert images into PDF in C#.net, ASP
How to Insert & Add Image, Picture or Logo on PDF Page Using C#.NET. Import graphic picture, digital photo, signature and logo into PDF document.
extract images from pdf; extract images from pdf files
C# Word - Paragraph Processing in C#.NET
C# users can set paragraph properties and create content such as run, footnote, endnote and picture in a paragraph. Create Picture in Paragraph.
pdf image extractor online; how to extract images from pdf
Appendix E. GS-800
E-11
Fig. E-10. Step Tablet dialog box.
Attached to the outside of your GS-800, you will find a copy of the manufacturer’s 
printout of the diffuse density values for each segment of the transmissive step tablet, 
as well as a serial number for the tablet. These exact density values must be entered 
into the software to associate a correct density value with each step on the step tablet.
Note:
Scanning in transmissive mode with incorrect step tablet values can cause 
significant errors in the reported densities of your scans.
First, type the serial number for the tablet into the Tablet Serial Number field.
The Quantity Offset field does not apply in the GS-800 in transmissive mode. This 
value should remain at zero.
Next, enter the step tablet values into the appropriate fields under the Diffuse column. 
After the step tablet is scanned, the software will associate each density value with its 
corresponding segment on the step tablet. The density values do not need to be 
reentered each time you calibrate.
When you are finished entering the transmissive step tablet values, click on OK.
VB.NET TIFF: How to Draw Picture & Write Text on TIFF Document in
drawing As RaterEdgeDrawing = New RaterEdgeDrawing() drawing.Picture = "RasterEdge" drawing provide powerful & profession imaging controls, PDF document, tiff
extract images pdf acrobat; some pdf image extractor
VB.NET Image: Image Cropping SDK to Cut Out Image, Picture and
first! VB.NET Image & Picture Cropping Application. Do you need to save a copy of certain part of an image file in a programming way?
extract images from pdf files without using copy and paste; extract image from pdf online
Quantity One User Guide
E-12
Reflective Step Tablet
For the reflective step tablet, it is recommended that you use the default target values 
in the software. Although you can edit these values based on your own measurements 
and testing, the default values have been preset based on careful testing and review.
To view or edit the default reflective values, make sure that the Reflective checkbox 
is selected in the main acquisition window, then click on the Options button. In the 
Densitometer Options dialog, click on the Edit Step Tablet button. The Step Tablet 
Values for GS-800 Reflective dialog box will open.
When you are finished viewing/editing the reflective step tablet values, click on OK.
Diffuse Versus Specular O.D.
In the step tablet form, you enter O.D. as diffuse density, and then the software 
automatically calculates the specular density.
Specular density is a measure of the light that passes directly through a medium. 
Diffuse density includes light that is scattered as it passes through the medium. Step 
tablet values are given in diffuse density, but are measured by the scanner in specular 
density, and therefore must be converted according to the specular/diffuse O.D. ratio. 
This conversion does not affect quantitation.
Fig. E-11. Specular and diffuse density
Diffuse density values are converted to specular optical density units according to the 
following formula:
Specular Light (passes directly through medium)
Diffuse Light (is scattered by medium)
Medium (gel, film, etc.)
Specular OD
1.4 Diffuse OD
=
VB.NET Image: Image Resizer Control SDK to Resize Picture & Photo
VB.NET Method to Resize Image & Picture. Here we display the method that We are dedicated to provide powerful & profession imaging controls, PDF document, image
extract image from pdf acrobat; extract text from image pdf file
VB.NET Image: Image Scaling SDK to Scale Picture / Photo
VB.NET DLLs to Scale Image / Picture. There are two dlls that will be involved in the process of VB.NET image scaling, which are RasterEdge.Imaging.Basic.dll
extract pictures from pdf; extract photo from pdf
Appendix E. GS-800
E-13
E.5.b Other Calibration Settings
After you have entered the step tablet values, you can immediately calibrate by 
clicking on the Calibrate Now button (in the Densitometer Options dialog box).
You can also specify how often you want the GS-800 to automatically recalibrate. 
Either click on the Calibrate Before Every Scan checkbox or specify a period 
between automatic recalibrations (in minutes) in the Recalibration Interval field.
Note:
The scanner will automatically recalibrate each time you change your filter or 
your reflective/transmissive setting. (If you select a different application with the 
same filter and light settings, it will not auto recalibrate.)
To print out a calibration report each time the densitometer calibrates, click on the 
Calibration Report checkbox.
E.6 Acquire the Image
Note:
Before scanning in transmissive mode, make sure the white balance region of the 
scanning area is not covered or obstructed in any way.
To begin to scan, click on the Acquire button. The scanned image will begin to 
appear in the scanning window, line by line.
To interrupt a scan, click on the Stop button. A message will ask you to confirm the 
interrupt, and then you will be asked if you want to keep the partial scan. This feature 
is useful if you overestimated the size of the area you selected.
After the scan is complete, a window will open displaying the scan image, at which 
point you can analyze and save it.
Note:
The image will open with a default file name that includes the date, time, and (if 
applicable) user name. However, unless you have selected Auto Save After 
Scan, the file will not be saved until you perform a Save or Save As operation.
C# TIFF: How to Insert & Burn Picture/Image into TIFF Document
Support adding image or picture to an existing or new creating blank TIFF are dedicated to provide powerful & profession imaging controls, PDF document, tiff
extract pdf images; how to extract a picture from a pdf
VB.NET Image: VB.NET Planet Barcode Generator for Image, Picture &
png, gif, jpeg, bmp and tiff) and a document file (supported files are PDF, Word & Generate Planet Barcode on Picture & Image in VB.NET. In this part, we will
extract jpg pdf; pdf image extractor c#
Quantity One User Guide
E-14
E.7 Other Options
Oversample
This feature allows you to scan at the maximum resolution of the scanner and then use 
spatial averaging to create an image with lower resolution. This can result in better 
images at lower resolution—however, it takes longer to scan.
To turn on oversampling, click on the Options button in the acquisition window and 
select the Oversample checkbox in the options dialog.
With oversampling on, you can specify your own resolution within the range of the 
densitometer by entering values directly in the fields next to X resolution and Y 
resolution in the main acquisition window.
Auto Save After Scan
To automatically save any scan you create, select the Auto Save After Scan 
checkbox.
With this checkbox selected, when you click on Acquire, a Save As dialog box will 
open asking you to specify a file name and location for the image you are about to 
create. The scan will begin when you click on the Save button.
Make Backup Copy
If you have selected Auto Save After Scan, you can also automatically create a 
backup copy of any scan you create.
Click on the Make Backup Copy checkbox. With this checkbox selected, when you 
save a scan, a backup copy will be placed in the same directory as the scanned image. 
Backup images are identified by "Backup of" followed by the original image file 
name. 
Note:
Backup files are read only files. If you attempt to save a backup file, you will be 
promted to give it a new name. This protects the backup file and preserves it from 
any changes.
VB.NET Image: Create Code 11 Barcode on Picture & Document Using
Write Code 11 barcode image on single-page or multi-page PDF, TIFF or Word document using VB.NET code. Code 11 Barcode Generation on PDF Using VB.NET Code.
extract vector image from pdf; how to extract images from pdf files
VB.NET Image: How to Save Image & Print Image Using VB.NET
tutorial page will tell you how to save and print image / picture file using Capable of saving and printing multi-page document files, like PDF and Word, in VB
extract images from pdf online; pdf image extractor c#
Appendix E. GS-800
E-15
Highlight Saturated Pixels
When this box is checked, any saturated pixels in the image will appear highlighted in 
red in the scan window and in the pop-up image window. To view/hide saturated 
pixels in the pop-up image window, use the Transform command on the main toolbar 
or Image menu.
Hide Grid
To hide the gridlines in the scanning area window, click on the Hide Grid checkbox.
Quantity One User Guide
E-16
F-1
Appendix F
Personal Molecular Imager FX
Fig.F-1. Personal Molecular Imager FX
Before you can begin acquiring images using the Personal Molecular Imager® FX, 
the instrument must be properly installed and connected with the host computer. See 
the Personal FX hardware manual for installation, startup, and operating instructions.
Note:
The Personal FX should be turned on and the initialization sequence completed 
before the host computer is turned on (except in the case of certain Power 
Macintosh configurations). See the hardware manual for more details.
Quantity One User Guide
F-2
PC Only: A Note About SCSI Cards
The Personal FX is connected to your computer by a Small Computer System 
Interface (SCSI) cable. To use the Personal FX, you must have a SCSI card installed 
in your PC. If you have a PC with a Windows 98 or Windows ME operating system, 
you may also need to load the SCSI and WinASPI drivers that came with the card.
Simulation Mode
Any of the imaging device acquisition windows can be opened in “simulation mode.” 
In this mode, an acquisition window will open and the controls will appear active, but 
instead of capturing real images, the window will create “dummy” images of 
manufactured data.
You do not need to be connected to an imaging device to open a simulated acquisition 
window. This is useful for demonstration purposes or practice scans.
To enter simulation mode, hold down the 
CTRL
key and select the name of the device 
from the File menu. The title of the acquisition window will indicate that it is 
simulated.
F.1 Personal FX Acquisition Window
To acquire images using the Personal FX, go to the File menu and select Personal 
FX.... The acquisition window for the imager will open, displaying a control panel 
and the scanning area window.
Documents you may be interested
Documents you may be interested