Appendix G. FX
G-5
Fig. G-3. Enabling Channel 2.
Enabled channels have a green check mark on their tabs.
You do not need to enable Channels 2–4 in sequence. For example, you can set up 
your four most common applications using the different channels, but only enable 
Channel 4 for a particular gel. Channel 1 would be scanned first, Channel 4 second.
Enabled channels are scanned sequentially. The separate scans are displayed and 
saved as separate images. The total scanning time depends on the number and type of 
enabled applications.
If you have selected Auto Save After Scan, the image created using Channel 1 will 
have the base file name, and images created using subsequent channels will have the 
base file name plus a version number. (v. 2, v. 3, v. 4). Note that the image version 
number does not necessarily correspond to the channel number. For example, if you 
scanned an image using only Channels 1 and 4, the image created using Channel 4 
will be saved as version 2 (v. 2).
G.2.b Selecting an Application
To select an application (i.e., the appropriate filter and other parameters for the type of 
object you are imaging), click on the Select button under a channel tab.
Select the checkbox
to enable the channel 
Click on the Select
button to select the
application 
Click on the tab to 
select the channel 
How to extract pictures from pdf files - Select, copy, paste PDF images in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document
extract images from pdf c#; how to extract text from pdf image file
How to extract pictures from pdf files - VB.NET PDF Image Extract Library: Select, copy, paste PDF images in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Support PDF Image Extraction from a Page, a Region on a Page, and PDF Document
pdf extract images; extract jpg from pdf
Quantity One User Guide
G-6
Fig. G-4. Example of an application tree: Ethidium Bromide gel.
Standard Applications
The standard applications and associated settings are listed in a tree that expands from 
left to right. When you select a standard application, the software automatically 
selects the appropriate filter(s) in the FX for that particular application. Some 
applications are inactive in the FX Pro and FX Pro Plus systems.
Standard FX Applications
Category
Application
Radioisotopes
CS- or BI-Screen (Bio-Rad)1
K-Screen (Kodak)2
Fuji-Screen2
Chemiluminescence
Chemi-Screen (Bio-Rad)1
Chemifluorescence
ECL-Plus
Attophos
DNA Stain Gel
Ethidium Bromide
Sybr Green I & II
Sybr Gold
VB.NET PDF Convert to Word SDK: Convert PDF to Word library in vb.
be converted to separate Word files within a short time in VB.NET class application. In addition, texts, pictures and font formatting of source PDF file are
extract image from pdf; extract images from pdf online
VB.NET Image: VB.NET Codes to Load Images from File / Stream in .
When evaluating this VB.NET imaging library with pictures of your own powerful & profession imaging controls, PDF document, image to pdf files and components
how to extract images from pdf file; extract image from pdf c#
Appendix G. FX
G-7
1Not supported in the FX Pro and FX Pro Plus systems.
2
Not supported in the FX Pro systems.
First select your general application, next select the particular stain or medium you 
are using, and finally (if appropriate) select the intensity of your samples.
Note:
Some applications require an external laser. If you choose one of these without 
having an external laser attached, you will receive a warning.
To exit the tree without selecting, press the 
ESC
key.
Your selection and settings will be displayed below the Select button.
Protein Stain Gel
Sypro Orange
Sypro Red
Nile Red
Sypro Ruby
Fluorophores
Alexa 488
Alexa 532
Alexa 546
FITC
FAM
CY3
HEX
R6G
Texas Red
Microtiter Plate
DNA (Sybr Green I)
Protein (Nano Orange)
ssDNA (Oligreen)
DNA (Picogreen)
B-Gal (FDG)
GUS (FDG)
Densitometry
Coomassie Blue Gel/Blot
Copper Stain Gel/Blot
Silver Stain Gel/Blot
X-Ray Film (Grey Type)
Standard FX Applications
VB Imaging - VB Code 93 Generator Tutorial
VB developers to create Code 93 on popular image files, including BMP developers to create and write Code 93 linear barcode pictures on PDF documents, multi
extract photo from pdf; extract images from pdf acrobat
C#: Use OCR SDK Library to Get Image and Document Text
a digital camera, scanned document or image-only PDF using C# color image recognition for scanned documents and pictures in C#. Steps to Extract Text from Image.
some pdf image extract; pdf image text extractor
Quantity One User Guide
G-8
Sample Intensity
Many FX applications require that you select a sample intensity (High, Medium, or 
Low) from the application tree. This is simply a rough estimate of how much sample 
is visible in your gel or other object.
If you are unsure of the level of intensity of your sample, you can always select a 
level, capture an image, then adjust the level and capture another image.
For example, if you select Low Sample Intensity and the resulting image has too 
many saturated pixels, you will receive a warning message. Simply change the setting 
to Medium Sample Intensity and rescan. If you select High Sample Intensity and the 
resulting image is too faint, select Medium or Low and rescan.
Custom Applications
If your application is not listed, if you want to use user-installed filters, or if you want 
to use an external laser, you can create and save your own custom application.
From the application tree, select Custom, then Create. This will open a dialog box in 
which you can name your application and select your settings.
Fig. G-5. Creating a custom application.
C# Imaging - Scan RM4SCC Barcode in C#.NET
& decode RM4SCC barcode from scanned documents and pictures in your Decode RM4SCC from documents (PDF, Word, Excel and PPT) and extract barcode value as
extract text from image pdf file; extract image from pdf acrobat
C# Imaging - C# Code 93 Generator Tutorial
to write and draw the best Code 93 barcode pictures in png, jpeg, gif, bmp, TIFF, PDF, Word, Excel Code93 barcode and save it to image files/object using
how to extract a picture from a pdf; extract photos from pdf
Appendix G. FX
G-9
To select a filter (including user-defined) or filter combination, click on the buttons 
for Filters A, B, and C, and make your choice from each pop-up list.
Note:
Filter wheel C is not available in the FX Pro or the FX Pro Plus systems.
Note:
The user-defined filters (User1, User2, etc.) cannot be renamed in the pop-up list, 
so be sure to remember which filter you insert into each position in the FX (i.e. 
690 nm filter, 605 nm BPfilter).
To use an external laser, click on the Laser button and select it from the pop-up list 
(i.e. 488 nm or 635 nm laser). Otherwise, use the default internal laser (532/1064nm).
Note:
The dual laser (532/1064 nm) is not available in the FX Pro or the FX Pro Plus 
systems.
Click on the PMT Voltage button to select a standard voltage for your custom 
application or create a custom PMT voltage.
To select a custom voltage, click on the Custom option. In the dialog box, adjust the 
slider to select a PMT voltage as a percentage of the maximum.
Fig. G-6. Selecting a custom PMT voltage.
Note:
For voltages above 80% of maximum, you will receive a warning message that 
the high voltage could damage the PMT. 
If you select Choose Later from the list of PMT voltages, the choices of sample 
intensity will be displayed when you select your custom application.
C# Imaging - Scan ISBN Barcode in C#.NET
which can be used to track images, pictures and documents in Load an image or a document(PDF, TIFF, Word barcodes from png image files and extract ISBN barcode
extract image from pdf java; extract photos pdf
Save, Print Images in Web Image Viewer| Online Tutorials
of single page printing and multi-page printing for pictures and documents; various file formats like PNG, JPEG, GIF, BMP, TIFF, PDF, MS Word Save Images & Files.
how to extract images from pdf in acrobat; extract images pdf
Quantity One User Guide
G-10
Finally, enter a name for your application in the Name field and click on OK to 
implement your changes.
After you have created an application, you can select it from the application tree by 
selecting Custom and the name you created. You can delete the application by 
selecting Custom, Delete, and the name of the application.
Fig. G-7. Selecting a custom application.
You can edit a custom application by selecting Custom, Edit, and the name of 
the application. You can also use this feature to create a new custom 
application from an existing one.
Fig. G-8. Application selection and settings.
VB.NET Image: Mark Photo, Image & Document with Polygon Annotation
SDK, which can be used to create the most common 7 types of annotations on various image files. What's more, if coupled with .NET PDF document imaging add-on
extract vector image from pdf; how to extract pictures from pdf files
VB.NET Image: Sharpen Images with DocImage SDK for .NET
VB.NET Coding. When you have made certain corrections in your VB.NET project photo or image files, you might want to sharpen your pictures before saving them
how to extract images from pdf; extract images pdf acrobat
Appendix G. FX
G-11
Once you select an application, the application name and settings appear 
below the Select button.
G.3 Step II. Select Scan Area
To select a scan area, drag your mouse within the scanning window. (In the scanning 
window, your cursor appearance will change to a cross.) The border of the scan area 
you are selecting is marked by a frame.
Fig. G-9. Selecting a scan area.
If you are in quadrant mode, note that the frame “locks” onto the next quadrant as you 
drag. When you release the mouse button, the border changes to a dashed blue line, 
indicating a selected area.
To reposition the scanning box you have selected, position your cursor inside the 
box and drag. The entire box will move.
To resize the box, position your cursor on a box side and drag. The side you have 
selected will move.
To redo the box entirely, position your cursor outside the box and drag. The old 
box will disappear and a new box will be created.
Drag cursor to
define scan
area
Quantity One User Guide
G-12
You can also select the scanning area by entering coordinates in the appropriate fields 
(Top, Bottom, Left, Right). After you enter a coordinate, the position of the scanning 
area box will change accordingly.
When selecting, be sure to include the entire area of interest, and be generous with 
borders. You can always crop the image later.
G.4 Step III. Select Resolution
The FX acquisition window allows you to scan at 50, 100, 200, or 800 micrometers. 
These resolutions are listed as option buttons in the control panel.
Fig. G-10. Resolution option buttons.
The resolution you select should be based on the size of the objects (e.g., bands, 
spots) you are interested in. For example:
50 micrometer resolution should be reserved for images requiring the highest 
level of detail, e.g., high density in situ samples, 1,536-well microplates, high 
density arrays, samples with very closely spaced bands. Files scanned at 50 
micrometers can be very large. 
100 micrometer resolution is useful for typical gels and arrays.
200 micrometer resolution is useful for gels with large bands and dot blots.
800 micrometer resolution should be reserved for very large objects, such as CAT 
assays.
Appendix G. FX
G-13
File Size of Images
Image File Size (below Select Resolution) shows the size of the scan file you are 
about to create. If you do not have enough computer memory for the specified file 
size, an error message will appear when you attempt to scan. If this happens, select a 
lower resolution or decrease the size of the area to be scanned. (Macintosh users can 
also increase the application memory partition. See your Macintosh computer 
documentation for guidance.)
G.5 Acquire the Image
Once you have selected your application, scan area, and resolution, you are ready to 
acquire an image.
Click on the Acquire button. There may be a short delay while the image laser warms 
up; then the scanned image will begin to appear in the scanning window, line by line.
To interrupt a scan, click on the Stop button. A message will ask you to confirm the 
interrupt, and then you will be asked if you want to keep the partial scan. This feature 
is useful if you overestimated the size of the area you selected.
Note:
If the image you are scanning has more than 10 saturated pixels, you will receive 
a warning message. If this happens, you can go back and select a higher sample 
intensity in the application tree.
Saving the Image
After the scan is complete, a message will appear asking you if you want to keep the 
scan. If you select Yes, a separate window will pop up containing the new image.
You can then save and analyze the image using the standard menu and toolbar 
functions.
Quantity One User Guide
G-14
G.6 Options
Auto Save After Scan
To automatically save any scan you create, click on the Auto Save After Scan 
checkbox.
With this checkbox selected, when you click on Acquire, a Save As dialog box will 
open asking you to specify a file name and location for the image you are about to 
create. The scan will begin when you click on the Save button.
If you are scanning using multiple channels, the image created using Channel 1 will 
be saved using the base file name, and images created using subsequent channels will 
have the base file name plus a version number. (v. 2, v. 3, v. 4). Note that the image 
version number does not necessarily correspond to the channel number. For example, 
if you scanned an image using only Channels 1 and 4, the image created using 
Channel 4 will still be saved as version 2 (v. 2).
Make Backup Copy
You can automatically create a backup copy of any scan you create. To do so, first 
select Auto Save After Scan (see above), then select the Make Backup Copy 
checkbox.
With this checkbox selected, when you save a scan, a backup copy will be placed in 
the same directory as the scanned image. Backup images are identified by "Backup 
of" followed by the original image file name. 
Note:
Backup files are read only files. If you attempt to save a backup file, you will be 
promted to give it a new name. This protects the backup file and preserves it from 
any changes.
Highlight Saturated Pixels
When this box is checked, any saturated pixels in the image will appear highlighted in 
red in the scan window and in the pop-up image window. To view/hide saturated 
pixels in the pop-up image window, use the Image > Transform command.
Documents you may be interested