pdf reader library c# : How to erase in pdf text control software platform web page winforms wpf web browser NReconUserGuide0-part968

SKYSCAN NV 
NRecon User Manual 
Skyscan NV 
4/20/2011 
How to erase in pdf text - delete, remove text from PDF file in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# developers to use mature APIs to delete and remove text content from PDF document
how to delete text from a pdf reader; how to delete text from a pdf document
How to erase in pdf text - VB.NET PDF delete text library: delete, remove text from PDF file in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
VB.NET Programming Guide to Delete Text from PDF File
how to edit and delete text in pdf file online; delete text pdf file
SkyScan NRecon user manual 2011  
User manual for SkyScan reconstruction program NRecon 
Index  
Introduction: 
NRecon in a nutshell
Chapter 1.   Installation, configuration and license
Chapter 2.  General layout and usage of the software 
2.1  Start the cluster reconstruction program
2.2  Layout and walk around
2.2.1 The Start page
2.2.2 The Settings and Advanced page
2.2.3 The Output page  
2.2.4 The Summary page
2.2.5 The batch manager
2.2.6 Optional functions
2.3  Customize the application 
Chapter  3.  Data format and conventions
Chapter  4.  How to import a non-SkyScan dataset
Introduction: NRecon in a nutshell 
SkyScan NRecon package is  used to reconstruct cross-section images from tomography 
projection images, mainly the cone-beam X-ray projections. The algorithm implemented is 
the  Feldkamp  algorithm.  The  native  SkyScan  data  are  recognized  automatically.  This 
program can also reconstruct non-native tomography data, if the file format and file names 
follow the conventions  described  in chapter 3
,  and a  corresponding  log  file  is  created 
following the instruction described in chapter 4
.  
The program is provided together with a help file with detailed information on how to use the 
software. This additional document serves as a manual for quick startup, and it overlaps 
largely with the online help. There are also a number of related technical notes available upon 
request.  
The reconstruction software package consists of 2 programs:  the user interface program 
NRecon and the reconstruction engine NReconServer. The two programs can run either on 
the same computer or on a cluster of computers. The communication between NRecon and 
C# WinForms Viewer: Load, View, Convert, Annotate and Edit PDF
Draw PDF markups. PDF Protection. • Sign PDF document with signature. • Erase PDF text. • Erase PDF images. • Erase PDF pages. Miscellaneous.
how to delete text from pdf with acrobat; deleting text from a pdf
C# WPF Viewer: Load, View, Convert, Annotate and Edit PDF
Draw markups to PDF document. PDF Protection. • Add signatures to PDF document. • Erase PDF text. • Erase PDF images. • Erase PDF pages. Miscellaneous.
how to delete text from pdf document; how to delete text in pdf file online
SkyScan NRecon user manual 2011  
NReconServer uses a client-server model (a common port). If windows firewall is enabled, 
you will be asked to unblock it for the server program.  Data are shared among the PCs using 
shared directory via network:  please make sure that all PCs involved have access to both 
input and output directories.  
A  typical  configuration  (see  Figure 1)  is  a  combination  of  1  host  computer  and  N 
(N=1,2,3,4,....) reconstruction nodes, as configuration 1 shown below. Or, you can run it as in 
configuration 2, with or without extra node PCs.  Ideally, the node PCs should be high-
definition: dual (or more) processors, large RAM (minimum 1GB if  8Kx8K reconstruction 
will be performed), fast FSB, minimum 40GB hard disk (excluding data storage) if 8Kx8K 
reconstruction will be performed.  
Figure 1  Typical layouts of NRecon/NReconServer 
Chapter 1    Installation, configuration and license   
It is rather straightforward to install and configure the reconstruction program NRecon: 
simply copy the programs to your destination folder. More details can be found in the 
installation guide which is part of the online help.   
The programs run under Windows operating systems (32-bit and 64-bit). It is important to 
install the 64-bit server program on a 64-bit computer to ensure maximum performance. For 
the user interface program NRecon, only 32-bit version is provided, which runs on both 32-
bit and 64-bit systems.  
There are 2 versions of NRecon/NReconServer available: the local and distributed/cluster 
version.  The local version is free.  
C# HTML5 Viewer: Load, View, Convert, Annotate and Edit PDF
Redact tab on viewer empower users to redact and erase PDF text, erase PDF images and erase PDF pages online. Miscellaneous. • RasterEdge XDoc.
remove text from pdf reader; how to delete text in a pdf file
C# PDF Text Redact Library: select, redact text content from PDF
Free online C# source code to erase text from adobe PDF file in Visual Studio. How to Use C# Code to Erase PDF Text in C#.NET. Add necessary references:
pdf editor online delete text; how to remove text watermark from pdf
SkyScan NRecon user manual 2011  
The local/free version:   please download and install the local NRecon/NReconServer.  There 
is no limitation in functionality. As the name suggests, both programs have to run on the 
same computer.  For each NRecon, only one server can be connected to.  
The distributed/cluster version:   you need a license to run this copy.   To apply for a license, 
please note down the product ID (found in menu Help->About NRecon ...; or in the warning 
message when no license or an invalid license is found by a distributed version), and send the 
ID together with a user name and your institution/company name to your distributor or to 
SkyScan. More information can be found on our website: http://www.skyscan.be
.    
To install license:  the NRecon.lic has to be in the same directory as NRecon.exe. By the first 
startup, you need to fill in the product ID, name and company name exactly as you received 
from SkyScan (case-sensitive; please use copy-and-paste to avoid typing   errors).    
Chapter 2    General layout and usage of the software 
If you have a local version, just start up NRecon: the server program will be started up 
automatically if it was not yet on. You may skip 2.1 which is specific for distributed/cluster 
version.   
The NReconServer(local) will only accept connection requested by NRecon running on the 
same computer.    
In case of multi-user environment, it is recommended to run only one copy of the server 
program at the same time:  the computation resources are used more optimally. 
2.1 Start the cluster reconstruction program and the client-server connections  
In general, the server program should be started first, but it is not strictly necessary.  If 
you exit the server program, turn off one of the node PCs or unplug one of the network 
cables while NRecon is on and connected,  NRecon will be informed and show the status 
correctly. However, if you start up a server program while NRecon is already on, you need to 
re-establish the connection.   
Neither NRecon nor NReconServer allows more than one copy running on the same 
computer.  In case of multi-user workstation (multiple users can log on simultaneously), only 
one copy of the server program can be started at a fixed port number: it is then recommended 
to install the server program only on one (common) account, so that you do not have to worry 
about different port number or multiple instances of the server program on a single machine.   
The server only "serves" one NRecon at one time (the active client), but it does hold a 
numbered  waiting  list  and  it  will  inform  all  connections  (the  connected  NRecon 
programs) whenever this status changes.    An active client gives up its "privilege" only when 
it disconnects from the server, and the first waiting client will be promoted to be the active 
client. In case of multiple  NRecon instances, communication among  the  users  may  be 
required to use the limited computer resource efficiently.  For a smooth operation in a multi-
user environment, it is recommended to disconnect the servers whenever they are not needed 
any more. 
C# HTML5 PDF Viewer SDK to view, annotate, create and convert PDF
setting PDF file permissions. Help C# users to erase PDF text content, images and pages online in ASP.NET. RasterEdge C#.NET HTML5
delete text pdf preview; how to delete text in pdf document
C# PDF Image Redact Library: redact selected PDF images in C#.net
Same as text redaction, you can specify custom text to appear over the image redaction area. How to Erase PDF Images in .NET Using C# Class Code.
deleting text from a pdf; how to delete text from pdf document
SkyScan NRecon user manual 2011  
2.2 Layout and walk around     
Figure 2 shows the typical appearance of NRecon and NReconServer. Little explanation is 
needed for NReconServer:  only the communication port number and the number of threads 
can be altered by the users.  The port number should be the same as that used in NRecon 
(default to 1) and usually does not require modification.  The number of threads influences 
reconstruction speed. The layout of NRecon is more complicated.  
Figure 2 Typical appearance of NRecon + NReconServer: the local version 
There are a few windows on the application window of NRecon: The main display where 
the images  are  shown, the  reconstruction  window  which hosts  a  few  pages/tabs  (start, 
settings, advanced, output, summary) for reconstruction purposes, the connection window 
which shows the status and progress of all servers, the color window which allows you to 
view the images in the main display in different color and scale.  Except the main display, all 
other windows are so-called dockable windows: you can drag them out of the main window, 
How to C#: Special Effects
Erase. Set the image to current background color, the background color can be set by:ImageProcess.BackgroundColor = Color.Red. Encipher.
erase text from pdf file; delete text from pdf with acrobat
Customize, Process Image in .NET Winforms| Online Tutorials
Include crop, merge, paste images; Support for image & documents rotation; Edit images & documents using Erase Rectangle & Merge Block function;
how to delete text from a pdf; delete text in pdf file online
SkyScan NRecon user manual 2011  
or dock them differently than they are shown here. You may customize the appearance of 
NRecon to your taste and convenience like most windows programs: the last used layout will 
be saved and loaded automatically at next startup.  You can also show/hide these dockable 
windows by closing the individual windows or via the View menu.  
Since there is only one display area, a few display modes are used:  navigation mode, 
alignment mode, preview mode, image mode and idle mode.  These display modes are 
coupled to the different pages/tabs on the reconstruction window.  In other words, switching 
from one display mode to another is done by switching from one page to another on the 
reconstruction window.   
The color window is also coupled to the display modes, i.e., each display mode has got its 
own settings for display. 
A general scenario of reconstruction is the following:  after loading a dataset, NRecon will be 
set at the Start page
and the display will be in the navigation mode, where the projections can 
be viewed; go to the Settings page
to check the parameters, usually the display will be in 
navigation mode to allow you to check the alignment; go back to the Start page to do a 
preview; NRecon goes automatically to the Output page
after reconstructing a slice for 
previewing, the display will be set at previewing mode, and you can modify the parameters 
concerning final  output image;  You  may go the Advanced page
for further parameters 
adjustment,  or  to  the Summary page
to  view  the  scan  parameters  and  reconstruction 
parameters; once all parameters are set, go back to Start page to start the full reconstruction 
or send the reconstruction task to the batch manager.  
The sections below will give more details on the different pages on the reconstruction 
window and the batch manager.  
2.2.1 The Start page  
This is the default page when a dataset is loaded, if no post-alignment is set automatically at 
loading time.   
This  window/page  offers  possibility  to  select  axial  reconstruction  range,  to  start  a 
reconstruction or to submit a batch job. It is coupled to the main display: setting this page 
active will set the main display to navigation mode in which a shadow image will be shown.   
Figure 2 is a snapshot taken when this page is set active.  To view other shadow images, you 
can click on the up/down arrows 
on the toolbar; or invoke a popup 
window by clicking on the second button.   
The green line on the main display in the navigation mode indicates the slice for previewing. 
The dark-red lines indicate the slice range for volume reconstruction.  These lines can be 
dragged to a desired location (left mouse button). The corresponding slice numbers are also 
shown on the start page, and can be altered from there as well, by either editing or scrolling.  
Please note, this range indication is only a rough guide due to the cone-beam geometry.  
Therefore, one should foresee a safe margin or do a preview to confirm when selecting the 
slices. 
.NET Imaging Processing SDK | Process, Manipulate Images
Provide basic transformation functions, like Crop, Rotate, Resize, Flip and more; Basic image edit function support, such as Erase Rectangle, Merge Block, etc.
erase pdf text online; how to erase in pdf text
SkyScan NRecon user manual 2011  
The functions provided on the Start page are described here.     
Preview:  
This function is provided to adjust the various parameters. It basically reconstructs one slice 
of choice. To change the slice on which a preview reconstruction will be performed, you can 
drag on the green line with mouse move, use the spin button, or simply type in a slice 
number.    
In preview mode, only the parameters shown on settings/advanced page will be taken into 
account, while the parameters shown on Output option page will be ignored silently.  For 
example, the region-of-interest option is ignored so that you can draw ROI on the full 
preview  image.   When preview reconstruction finishes, the Output option page will be 
automatically set active, offering you the chance to modify a few output parameters for the 
final results, such as the dynamic image range, output format, scales, color and etc.   
The major aim of previewing is to help you adjust parameters.  You can always switch back 
and forth between the pages to adjust them. To help you associate the parameters with result, 
the slice number (e.g.,  Prv 0514) is shown on the status bar whenever the preview image is 
in display.  If any change occurs which might result in different image quality, a * will be 
shown (e.g., *Prv 0514).   
Fine tuning:   
A few reconstruction parameters have to be adjusted manually in a try-and-error fashion 
using the preview function. It requires quite some effort to save the preview images,  view 
them  and  then  make  a  decision.  The  "fine-tuning"  function  is  designed  to  make  this 
adjustment  somewhat  easier  by  launching  a  series  of  previews.   You  may  adjust  one 
parameter at a time, while keeping all other parameters fixed.  In this way, you may tune 4 
parameters: post-alignment, beam-hardening correction, ring artifacts correction level and 
smoothing level.   
Figure 3 shows the popup window appearing upon clicking on the fine-tune button. You may 
choose the parameter to tune, and the parameter values are given such that they are centered 
on the current value (the valid range of the parameters are taken into account as well) with 
given step size. Once you have set this, you can click on Start to start a series of preview 
reconstructions.    
SkyScan NRecon user manual 2011  
Figure 3  Fine-tune window 
Once the preview reconstructions are done, you may use the 
buttons on the toolbar 
to scroll. The parameter is adjusted automatically when you scroll.  In other words, simply 
concentrating on the images until you find a satisfied one, and the fine-tuning function is 
done.   In case of doubt on the consistency, there is text information at the top-right of the 
image, indicating which preview you are viewing; or go to the Settings page, switch the 
alignment image off, and you will be able to view the preview image and observe how the 
parameter updates while scrolling.   If, accidentally or not, you have changed any parameter 
which does not correspond to any of the preview images, a * will be shown (e.g., *Prv 0514) 
on the toolbar, just as in the normal preview case.  
Start 
Obviously, this is the button to start a reconstruction.   It checks for disk sufficiency and 
whether you want to overwrite files if there are already there.    
When the first resulting image is available on the disk, it will be loaded automatically and 
the Overview page will be set active, showing the first reconstructed slice in image mode.  
Same as in navigation mode, you may walk through the resulting images by using the   
buttons. Please note that the images may appear non-sequentially due to work division 
among multi-threads and multi-computers.  
Add to batch 
Instead of reconstructing the volume immediately, you can send the reconstruction task to 
the batch manager. You can start the batch jobs later. Section 2.3.5
has more details about the 
batch processing.  
2.2.2 The Settings and Advanced page   
SkyScan NRecon user manual 2011  
There  are  two  windows/pages for reconstruction  parameters.  These two  windows offer 
possibility to change reconstruction parameters for the current dataset.  The most frequently 
used parameters are put in page Settings, and other more advanced parameters are grouped 
together in page Advanced.  
The Settings page can be coupled to the main display in 3 ways:  alignment mode, preview 
mode or  navigation mode.  The  alignment  mode will be switched on if both the post-
alignment  option  and  the  show  option  are  checked;  otherwise  the preview mode  (if 
previewing image is available) or the navigation mode will be switched on.   You may switch 
between two modes by check/uncheck the show option: convenient when you are adjusting 
parameters by looking at the existing preview image.    
The Settings page is set automatically active after loading if the post-alignment option is 
checked.  Figure 4 shows the settings page with alignment display mode.   This display mode 
is switched on when both the post-alignment option and the show option are checked. The 
image shown in display is the overlaid  image (alpha-blending) of 2  shadow images at 
opposite positions (180 degrees apart).  One of the 2 images is flipped. The intention of this 
option is to help you choose the correct alignment value at the first stage. However, the 2 
images may differ a lot due to the cone-beam geometry and different magnification factors. 
Therefore, it is always recommended to use the previewing choose the post-alignment value.   
The initial/default post-alignment value is estimated at loading time. Instead of comparing the 
original cone-beam projections, we compare pseudo-parallel projections obtained from the 
cone-beam projections at 0 and 180 degrees. Only the central portion of the projections are 
used (you may change the portion of data used for post-alignment in Preferences.).  If you 
interrupt/cancel the rebinning, the original version as described above will be used instead.  
The profiling function is partially enabled: clicking on 
will show you 2 curves (profiles 
along the central horizontal line, average over the full vertical length).  Together with the 
profiles, the overlaid image provides a rather interactive way of adjusting the post-alignment 
value.  
SkyScan NRecon user manual 2011  
Figure 4 The Settings page with alignment display mode.  The profile window is also visible
Figure 5 shows the Advanced page coupled with navigation display mode.  
Documents you may be interested