Standby and Ready State 
The STANDBY toggle switch is located in the bottom left corner of the front panel.  
After switching on the AC power, press the STANDBY key briefly to switch the analyzer from the 
standby to ready state or vice versa. 
In standby state, the right, amber LED is on. The standby power only supplies the power switch 
circuits and the optional oven quartz (OCXO, 10 MHz reference oscillator, option ZVAB-B4, order 
no. 1164.1757.02). In this state it is safe to switch off the AC power and disconnect the instrument 
from the power supply. 
In ready state, the left, green LED is on. The analyzer is ready for operation. All modules are 
power-supplied and the analyzer initiates its startup procedure.  
CAUTION! 
The instrument is still power-supplied while it is in standby mode.  
Replacing Fuses 
The DC inputs PORT BIAS at the rear panel are protected by four fuses IEC 127 - F 250 L 
(250 mA flink).  
To replace the fuses open the fuse holder by slightly turning the lid counterclockwise.  
Replacement fuses are provided with the instrument.  
Batch pdf metadata editor - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
add metadata to pdf programmatically; remove pdf metadata
Batch pdf metadata editor - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
edit pdf metadata acrobat; pdf metadata viewer online
Starting the Analyzer, Shutdown 
R&S
® 
ZVB 
1145.1055.62 
1.14 
E-3 
Starting the Analyzer, Shutdown 
To start the analyzer, proceed as follows: 
1.  Make sure that the instrument is connected to the AC power supply and the power switch on 
the rear panel is in position I (On).  
2.  If necessary, press the STANDBY toggle switch on the front panel to switch the instrument to 
ready state (the green LED is on).   
In ready state, the analyzer automatically performs a system check, boots the Windows XP operating 
system and then starts the network analyzer (NWA) application. If the last analyzer session was 
terminated regularly, the NWA application uses the last setup with all instrument settings.  
To shut down the analyzer, proceed as follows: 
1.  Press the STANDBY key to save the current setup, close the NWA application, shut down 
Windows XP and set the instrument to standby state. Of course you can also perform this 
procedure step by step like in any Windows session.  
2.  If desired, set the AC power switch to position 0 (Off).  
Attention! 
It  is strongly  recommended  to  switch  the  analyzer  to  standby  state  before 
disconnecting it from the AC supply. If you set the power switch to 0 while the 
NWA application is still running, you will lose the current settings. Moreover, loss of 
program data can not be excluded if the application is terminated improperly.  
C# PDF Convert to Tiff SDK: Convert PDF to tiff images in C#.net
Studio .NET project. Powerful .NET control to batch convert PDF documents to tiff format in Visual C# .NET program. Free library are
get pdf metadata; change pdf metadata
C# PDF Convert to Word SDK: Convert PDF to Word library in C#.net
Powerful components for batch converting PDF documents in C#.NET program. Convert PDF to multiple MS Word formats such as .doc and .docx.
remove metadata from pdf; pdf xmp metadata editor
R&S
®
ZVB 
Connecting External Accessories 
1145.1055.62 
1.15 
E-3 
Connecting External Accessories 
The equivalent USB ports on the front and rear panel of the analyzer can be used to connect a variety 
of accessories: 
A mouse simplifies operation of the instrument using the controls and dialogs of the Graphical User 
Interface (GUI). 
A keyboard simplifies the entry of data.  
A printer generates hard copies of the screen contents. 
In addition the analyzer provides interfaces for monitor connection and network integration:  
An external monitor shows the magnified Graphical User Interface (GUI) with all diagram areas 
and controls. 
A LAN connection can be established in order to access the hard disk or control the analyzer from 
an external PC.  
Connecting a Mouse 
A USB mouse can be connected to one of the Universal Serial Bus connectors on the front panel or on 
the rear panel (double USB connectors).   
The mouse is detected automatically when it is connected. It is safe to connect or disconnect the 
mouse during the measurement.  
Use the Start – Control Panel – Mouse menu of Windows XP to configure the mouse properties. 
To access Windows XP, press the Windows key on the external keyboard or on the front panel.  
Operating the analyzer does not require a mouse. You can access all essential functions using 
the keys on the front panel.  
Connecting a Keyboard 
A keyboard can be connected to one of the Universal Serial Bus connectors on the front panel or on the 
rear panel (double USB connectors). 
The keyboard is detected automatically when it is connected. The default input language is English – 
US. It is safe to connect or disconnect the external keyboard during the measurement.  
Use the Start – Control Panel – Keyboard or Regional and Language Options menu of Windows 
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
NET components for batch combining PDF documents in C#.NET class. Powerful library dlls for mering PDF in both C#.NET WinForms and ASP.NET WebForms.
pdf remove metadata; batch pdf metadata
C# PDF Print Library: Print PDF documents in C#.net, ASP.NET
NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET Quicken PDF printer library allows C# users to batch print PDF file in
pdf keywords metadata; batch pdf metadata
Connecting External Accessories 
R&S
® 
ZVB 
1145.1055.62 
1.16 
E-3 
XP to configure the keyboard properties. To access Windows XP, press the Windows key on the 
external keyboard or on the front panel.  
Operating the analyzer does not require a keyboard. You can access all essential functions 
using the keys on the front panel. In combination with a mouse, the front panel keys provide 
access to all instrument functions.  
Connecting a Printer 
A printer can be connected to one of the Universal Serial Bus connectors on the front panel or on the 
rear panel (double USB connectors). 
It is safe to connect or disconnect the printer during the measurement. When printing a copy (File – 
Print), the analyzer checks whether a printer is connected and turned on and whether the appropriate 
printer driver is installed. If required, printer driver installation is initiated using Windows XP's Add 
Printer Wizard. The wizard is self-explanatory. A printer driver needs to be installed only once, even 
though the printer may be temporarily removed from the analyzer.  
Printer driver installation 
A great variety of printer drivers is available on the analyzer. To obtain the complete list, access 
Windows XP (press the Windows key) and open the Add Printer Wizard in the Start  Control Panel – 
Printer and Faxes menu. 
You can load updated and improved driver versions or new drivers from an installation disk, USB 
memory stick or another external storage medium. Alternatively, if the analyzer is integrated in a 
network, you can install driver data stored in a network directory. In either case, use the Add Printer 
Wizard to complete the installation.   
Use the Page Setup dialog or the Start – Control Panel – Printers and Faxes menu of Windows 
XP to configure the printer properties. To access Windows XP, press the Windows key on the external 
keyboard or on the front panel.  
C# Create PDF from Tiff Library to convert tif images to PDF in C#
NET component for batch converting tiff images to PDF documents in C# class. Create PDF from Tiff in both .NET WinForms and ASP.NET application.
clean pdf metadata; batch pdf metadata editor
C# Create PDF from PowerPoint Library to convert pptx, ppt to PDF
PowerPoint. Professional .NET PDF converter control for batch conversion. Export PowerPoint hyperlink to PDF in .NET console application.
pdf remove metadata; google search pdf metadata
R&S
®
ZVB 
Connecting External Accessories 
1145.1055.62 
1.17 
E-3 
Connecting a Monitor 
A standard VGA monitor can be connected to the 15-pole Sub-Min-D MONITOR connector on the rear 
panel of the analyzer.  
Attention! 
The monitor must be connected while the instrument is switched off (in standby 
mode). Otherwise correct operation can not be guaranteed.  
The monitor displays the magnified analyzer screen with all diagram areas, measurement results and 
control elements. No extra configuration is required.  
With  an  additional  mouse  or  keyboard  connected  to  the  analyzer,  you  can control the 
measurement from the external monitor. If desired, click Display – Config/View – Hardkey Bar to add 
the Hardkey Bar (front panel key bar) to the analyzer screen. 
Connecting a LAN Cable 
A LAN cable can be connected to one of the LAN connectors on the rear panel of the analyzer. To 
establish a LAN connection proceed as follows:  
1.  Refer to section Assigning an IP Address and learn how to avoid connection errors. 
2.  Connect an appropriate LAN cable to one of the LAN ports. Use a commercial RJ-45 cable to 
establish a non-dedicated network connection, or a cross-over RJ-45 cable to establish a 
dedicated connection between the analyzer and a single PC.    
Dedicated vs. non-dedicated network connections 
There are two methods to establish a LAN connection of the analyzer:  
A non-dedicated network (Ethernet) connection from the analyzer to an existing network made with 
an ordinary RJ-45 network cable. The analyzer is assigned an IP address and can coexist with a 
computer and with other hosts on the same network.  
A dedicated network connection between the analyzer and a single computer made with a cross-
over RJ-45 network cable. The computer must be equipped with a network adapter and is directly 
connected to the analyzer. The use of hubs, switches, or gateways is not needed, however, data 
C# Create PDF from Excel Library to convert xlsx, xls to PDF in C#
Professional .NET PDF converter control for batch conversion. Merge all Excel sheets to one PDF file. Export PDF from Excel with cell border or no border.
batch edit pdf metadata; add metadata to pdf programmatically
C# Create PDF from Word Library to convert docx, doc to PDF in C#.
and ASP.NET. Convert both DOC and DOCX formats to PDF files. Professional .NET control for batch conversion in C#.NET class. Easy to
remove metadata from pdf acrobat; online pdf metadata viewer
Connecting External Accessories 
R&S
® 
ZVB 
1145.1055.62 
1.18 
E-3 
transfer is still made using the TCP/IP protocol.  
The IP address information is displayed in the Info – Setup Info dialog. 
Test Setups with Two LAN Connections 
The two LAN connectors on the rear panel of the analyzer are equivalent. With one LAN connector 
used to establish a connection to a home/company network, the other one can be used to connect an 
additional instrument, e.g. an additional analyzer or signal generator:  
Defining the network topology: Router vs. network client 
With two LAN connections, it is possible to use the analyzer in two alternative ways:   
As a client participating in two independent networks, one comprising the home network including 
the analyzer, the second consisting of the additional test instrument plus the analyzer.  
As a data router between the additional test instrument and the home network. This configuration 
means that the analyzer and the additional test instrument are integrated into a single network. 
The network topology is defined in Windows XP's Control Panel – Network Connections – Local Area 
Connection Status – Local Area Connection Properties – Internet Protocol (TCP/IP) Properties – 
Advanced TCP/IP Settings dialog. Both instruments must have independent IP addresses; see section 
Assigning an IP Address below. Contact your LAN administrator for details.  
Attention! 
Never use both LAN connectors to connect the analyzer in parallel to the same 
network as this will result in connection errors.  
R&S
®
ZVB 
Remote Control in a LAN 
1145.1055.62 
1.19 
E-3 
Remote Control in a LAN 
A LAN connection is used to integrate the analyzer into a home/company network. This offers several 
applications:  
Transfer data between a controller and the analyzer, e.g. in order run a remote control program.  
Control the measurement from a remote computer using the Remote Desktop application.   
Use external network devices (e.g. printers).  
Attention!  
An efficient virus protection is a prerequisite for secure operation in the network. 
Never connect your analyzer to an unprotected network because this may cause 
damage to the instrument software.   
To establish the connection proceed as follows: 
1.  Press the Windows button and open Windows XP's control panel. 
2.  Select System and open the Remote tab in the System Properties dialog. Enable Allow users to 
connect remotely to this computer.   
3.  Assign an IP address to the analyzer following the directions below and connect the analyzer to 
the network as described in Connecting a LAN Cable.  
4.  Create a Remote Desktop Connection using the analyzer's IP address.  
The analyzer uses a user name and password as credentials for remote access. In the factory 
configuration, instrument is preset for both the user name and the password. To protect the 
analyzer from unauthorized access, it is recommended to change the factory setting.   
Assigning an IP Address 
Depending on the network capacities, the TCP/IP address information for the analyzer can be obtained 
in different ways.   
If the network supports dynamic TCP/IP configuration using the Dynamic Host Configuration 
Protocol (DHCP), all address information can be assigned automatically.   
If  the  network  does not support  DHCP,  or if the analyzer is set  to use alternate TCP/IP 
configuration, the addresses must be set manually.   
By default, the analyzer is configured to use dynamic TCP/IP configuration and obtain all address 
information automatically. This means that it is safe to establish a physical connection to the LAN 
without any previous analyzer configuration.  
Attention! 
If your network does not support DHCP, or if you choose to disable dynamic 
TCP/IP  configuration,  you  must  assign  valid  address  information before 
connecting the analyzer to the LAN. Contact your network administrator to obtain 
a valid IP address, because connection errors can affect the entire network.  
Remote Control in a LAN 
R&S
® 
ZVB 
1145.1055.62 
1.20 
E-3 
Manual TCP/IP configuration 
To disable dynamic TCP/IP configuration and enter the TCP/IP address information manually proceed 
as follows:  
1.  Obtain the IP address and subnet mask for the analyzer and the IP address for the local default 
gateway from your network administrator. If needed, also obtain the name of your DNS domain 
and the IP addresses of the DNS and WINS servers on your network. If you use both LAN 
connectors, you need two different sets of address information.  
2.  Press the Windows key to access Windows XP.  
3.  Open the Control Panel – Network Connections – Local Area Connection Status – Local Area 
Connection Properties – Internet Protocol (TCP/IP) Properties dialog and enter the address 
information, e.g.: 
For more information refer to the Windows XP Help.  
Remote Desktop Connection 
Remote Desktop is a Windows application which can be used to access and control the analyzer from a 
remote computer through a LAN connection. While the measurement is running, the analyzer screen 
contents are displayed on the remote computer, and Remote Desktop provides access to all of the 
applications, files, and network resources of the analyzer.  
To set up a Remote Desktop connection  
1.  Connect the analyzer to a LAN and determine its IP address; see Remote Control in a LAN.  
2.  Set up your remote computer (integrated in the LAN) to use Remote Desktop and create the 
Remote Desktop connection to the analyzer.  
For detailed information about Remote Desktop and the connection refer to the Windows XP Help.  
R&S
®
ZVB 
Windows XP 
1145.1055.62 
1.21 
E-3 
Windows XP 
The analyzer is equipped with a Windows XP operating system which has been configured according to 
the instrument's features and needs. Changes in the system configuration can be necessary in order to  
Establish a LAN connection  
Customize the properties of the external accessories connected to the analyzer  
Call up additional software tools  
Attention! 
The operating system is adapted to the network analyzer. To avoid impairment of 
instrument functions, only change the settings described in this manual. Existing 
software must  be  modified  only with update software released  by Rohde & 
Schwarz. Likewise, only programs authorized by Rohde & Schwarz for use on the 
instrument must be executed.  
The Start menu of the operating system is accessed by pressing the Windows key in the SUPPORT 
keypad. All necessary settings can be accessed from the Start menu, in particular from the Control 
Panel.   
Firmware Update 
Upgrade versions of the analyzer firmware are supplied as single setup files *.msi. To perform a 
firmware update 
1.  Copy the setup file to any storage medium accessible from the analyzer. This may be the 
internal hard disk, an external storage medium (USB memory stick, CD-ROM with external 
drive) or a network connection (LAN, GPIB bus).  
2.  Double-click the setup file (or use the front panel keys to select and start the setup file; see 
Using Front Panel Keys in Chapter 2) and follow the instructions of the setup wizard.  
Setup files can be stored and installed again. The default drive name of the internal hard disk is C:. 
External storage devices are automatically mapped to the next free drive, i.e. D:, E: etc.  
Attention! 
The Calibration Unit (accessory R&S ZV-Z51/-Z52 or R&S ZV-Z58) must be 
disconnected during a firmware update.  
R&S
®
ZVB 
Getting Started 
1145.1055.62 
I-2.1 
E-2 
Contents of Chapter 2 
2  Getting Started.................................................................................................... 2.1 
Reflection Measurements................................................................................................................2.2
Instrument Setup for Reflection Measurements.......................................................................2.2
Parameter and Sweep Range Selection..................................................................................2.2
Instrument Calibration..............................................................................................................2.3
Evaluation of Data....................................................................................................................2.5
Saving and Printing Data.........................................................................................................2.7
Basic Tasks.......................................................................................................................................2.9
Control via Front Panel Keys...................................................................................................2.9
To Access a Particular Menu Command ......................................................................2.9
To Make a Selection in a Dialog..................................................................................2.10
Data Entry..............................................................................................................................2.11
Using Front Panel Keys...............................................................................................2.11
Using the Analyzer's On-Screen Keyboard.................................................................2.12
Using Windows XP's On-Screen Keyboard.................................................................2.12
Scaling Diagrams...................................................................................................................2.13
Setting the Sweep Range............................................................................................2.13
Reference Value and Position.....................................................................................2.13
Autoscale.....................................................................................................................2.14
Circular Diagrams........................................................................................................2.14
Using Marker Functions...............................................................................................2.14
Enlarging the Diagram Area........................................................................................2.15
Documents you may be interested
Documents you may be interested