R&S ZVL 
Measurement Examples 
Basic Measurement Tasks 
Operating Manual 1303.6580.32-05 
105
Screen Control     
Control of the analyzer's display elements is based on the standard Windows™ functionality. The following 
examples show you how to make efficient use of the functions.   
To move an active setup window within the available space ... 
Make sure that the window is not maximized.  
Click the 
icon in the title bar to open the setup control menu.  
Click Move. 
After the pointer changes to the four-headed arrow: 
Press one of the direction keys (left, right, up, or down arrow key) to move the entire window without 
changing its size.  
Press ENTER or left-click the window when it is at the position you want.  
To size the active setup window ... 
Make sure that the window is not maximized.  
Click the 
icon in the title bar to open the setup control menu.  
Click Size. 
After the pointer changes to the four-headed arrow: 
Press one of the direction keys (left, right, up, or down arrow key) to move the pointer to the border 
you want to move.   
Press a direction key to move the border.  
Press ENTER or left-click the window when it is the size you want.  
Data Transfer     
The following examples show you how to perform basic tasks related to data transfer from and to the 
analyzer.   
To copy the active setup ... 
Connect a storage device like an USB stick or a CD-ROM drive to an USB connector. 
Click Nwa-File — Save Nwa As to open the Save As dialog.  
Open the Save in drop-down list and select the storage device and a folder.  
Select a File name and click Save to store the data on the storage device and to close the Save As 
dialog. 
The default extension for setup files (*.nwa) is appended automatically. 
Add metadata to pdf programmatically - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
view pdf metadata; embed metadata in pdf
Add metadata to pdf programmatically - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
clean pdf metadata; edit multiple pdf metadata
R&S ZVL 
Measurement Examples 
Basic Measurement Tasks 
Operating Manual 1303.6580.32-05 
106
To load a setup stored on an external device... 
Connect the storage device to an USB connector. 
Click Nwa-File — Recall Nwa to open the Open dialog.  
Open the Look in drop-down list and select the storage device. 
In the center of the dialog, select the folder containing the setup files.  
Select a file and click Open to load the setup as the active setup. 
Setting up a Sweep     
Configurations of the different sweep types require different procedures. The following examples list the 
essential settings for various sweep types.  
To Set up a Lin Frequency Sweep... 
Press the Sweep key or click Channel - Sweep to access the Sweep submenu. 
Click Sweep Type – Lin Frequency. 
Click Channel – Center or Span to access the submenu defining the sweep range. 
To define a sweep range confined by a start and a stop frequency, use Start and Stop. Alternatively, 
to define a sweep range of definite width around a known center frequency (e.g. an expected peak), 
use Center and Span.  
To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce the 
sweep time), click Channel – Sweep – Number of Points. 
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and labeled with the Start and 
Stop frequencies. 
To Set up a Log Frequency Sweep... 
Press the Sweep key or click Channel - Sweep to access the Sweep submenu. 
Click Sweep Type - Log Frequency. 
Click Channel - Center or Span to access the submenu defining the sweep range. 
To define a sweep range confined by a start and a stop frequency, use Start and Stop. Alternatively, 
to define a sweep range of definite width around a known center frequency (e.g. an expected peak), 
use Center and Span.  
To vary the number of measurement points (e.g. to improve the resolution of the sweep or reduce the 
sweep time), click Channel - Sweep - Number of Points. 
If the active diagram is a Cartesian diagram, the x-axis is logarithmically scaled and labeled with the Start 
and Stop frequencies. 
VB.NET PowerPoint: Read, Edit and Process PPTX File
How to convert PowerPoint to PDF, render PowerPoint to SVG and effective VB.NET solution to add desired watermark VB.NET PowerPoint: Read & Edit PPTX Metadata,
get pdf metadata; pdf metadata
C# Create PDF Library SDK to convert PDF from other file formats
Besides, using this PDF document metadata adding control, you can add some additional information to generated PDF file.
search pdf metadata; edit pdf metadata
R&S ZVL 
Measurement Examples 
Basic Measurement Tasks 
Operating Manual 1303.6580.32-05 
107
To Set up a Segmented Frequency Sweep... 
Press the Sweep key or click Channel - Sweep to access the Sweep submenu. 
Click Sweep Type - Define Segments to call up the dialog defining the individual sweep segments. 
Proceed as described in Define Segments to customize your sweep range. Make sure to define at 
least 2 distinct sweep frequencies. 
Use the controls in the Individual Settings panel to define the internal generator power, measurement 
(IF) bandwidth and other measurement settings for each sweep segment.  
Close the Define Segments dialog. 
Click Sweep Type - Segmented Frequency.
If the active diagram is a Cartesian diagram, the x-axis is linearly scaled and ranges from the lowest to the 
highest frequency in all segments. 
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Document and metadata. All object data. Flatten visible layers. C#.NET DLLs: Compress PDF Document. Add necessary references: RasterEdge.Imaging.Basic.dll.
edit pdf metadata acrobat; read pdf metadata online
C# PDF Text Search Library: search text inside PDF file in C#.net
Ability to search and replace PDF text in ASP.NET programmatically. C#.NET PDF DLLs for Finding Text in PDF Document. Add necessary references:
pdf metadata editor; pdf metadata viewer online
R&S ZVL 
Measurement Examples 
Optimization 
Operating Manual 1303.6580.32-05 
108
Optimization 
Optimizing the Measurement Speed     
Minimizing the measurement time per sweep contributes to a high measurement throughput. The following 
parameters have an influence on the sweep time:  
To find the best set of sweep points... 
Limit the measurement to the stimulus range that you need for your device:  
Click Channel - Center or Span to access the submenu defining the sweep range. 
To define a sweep range confined by a start and a stop frequency, use Start and Stop. Alternatively, 
to define a sweep range of definite width around a known center frequency (e.g. an expected peak), 
use Center and Span.
Alternatively, use the marker functions to define the sweep range.  
Use the minimum number of points that will produce an acceptable resolution:  
Click Channel – Sweep –  Number of Points and reduce the number of sweep points. 
Use the segmented sweep to focus on different subranges of the sweep, using optimized channel settings 
for each subrange:  
Set up a segmented frequency sweep as described in Segmented Frequency.  
Optimize the number of points and the noise reduction settings for each individual sweep segment.  
The sweep time increases with the Frequency Step Size. 
To find the appropriate noise reduction settings... 
Use the fastest IF filter that will produce acceptable dynamic range:  
Press the PWR BW key or click Channel – Pwr Bw to access the corresponding submenu.  
Select the widest bandwidth compatible with your measurement task.
If possible, click Fine Adjust to further increase the bandwidth. Select the Fast filter type unless a 
Normal filter is required for your measurement. 
Reduce the number of repeated measurements made to obtain an averaged result. Averaging does not 
slow down the measurement, however, several consecutive sweeps must be performed until an average 
result is available. 
If averaging is not necessary, open the Channel – Power Bandwidth Average menu and make sure 
that Average On is not active.  
Alternatively, click Average Factor and reduce the factor to a minimum.  
.NET PDF Document Viewing, Annotation, Conversion & Processing
Extract images from PDF file. Read PDF metadata. Search text content inside PDF. Edit, remove images from PDF. Add, edit, delete links. Form Process.
adding metadata to pdf files; add metadata to pdf programmatically
VB.NET PDF - How to Add Barcode on PDF Page
text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET VB.NET PDF barcode creator add-on, which combines the PDF reading add-on with
acrobat pdf additional metadata; pdf metadata editor online
R&S ZVL 
Measurement Examples 
Optimization 
Operating Manual 1303.6580.32-05 
109
Calibrating a Measurement Channel     
Calibration is the process of eliminating systematic, reproducible errors from the measurement results 
(system error correction). A measurement channel can be calibrated either manually or automatically 
using a calibration unit (accessory R&S ZV-Z5x).  
Select and perform a manual calibration 
In the following example a 2-port TOSM calibration is performed using the Calibration Wizard. Other 
calibrations can be performed in an analogous way.  
Perform all channel settings for your measurement.  
Click Channel – Cal – Start Cal – Two-Port P1 P2 – TOSM to open the calibration wizard for the 
desired calibration type. 
In the first dialog of the calibration wizard, select the connector type that you use at port 1 and 2 (both 
connector types must be equal) and the calibration kit. Click Next > to access the next dialog of the 
wizard.
Connect an open standard to port 1 and click the corresponding box in the Measured Standards list to 
initiate the calibration sweep and acquire the calibration data for this standard.  
Repeat the last step for a short and a match connected to port 1, an open, short and match connected 
to port 2, and for a through between both ports. 
Click Apply to start the calculation of the system error correction data, apply them to the current 
channel and close the calibration wizard. 
Replace the last measured standard with your DUT and perform calibrated measurements without 
changing the channel settings used for calibration.  
You can also access manual calibration from the S-Parameter Wizard.  
Perform an automatic calibration  
In the following example a 2-port TOSM calibration is performed using the calibration unit (accessory R&S 
ZV-Z53). The calibration of a single port is analogous; see Performing an Automatic Calibration.  
Connect the calibration unit:  
Switch on and start-up your network analyzer and establish a USB connection to the calibration unit; 
see Connecting the calibration unit.  
Connect port 1 of the unit to port 1 of the analyzer; repeat this for port 2.  
Perform the calibration:  
Click Channel – Cal – Start Cal – Two-Port P1 P2 – Calibration Unit and wait until the calibration is 
completed.   
Remove the connecting cables from the unit, connect your DUT instead and perform calibrated 
measurements.  
You can also access automatic calibration from the S-Parameter Wizard. 
VB.NET PDF - Convert CSV to PDF
pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET VB.NET Demo Code for Converting RTF to PDF. Add necessary references
remove metadata from pdf acrobat; batch pdf metadata editor
VB.NET PDF Text Search Library: search text inside PDF file in vb.
Ability to search and replace PDF text programmatically in VB.NET. Our VB.NET PDF Document Add-On enables you to search for text in target PDF document
pdf metadata; pdf metadata reader
R&S ZVL 
Measurement Examples 
Advanced Tasks 
Operating Manual 1303.6580.32-05 
110
Advanced Tasks 
Time Domain Measurements     
With option ZVL-K3, Time Domain, you can view the measurement results as a function of time.  
To set up a time domain measurement... 
Reset the analyzer to ensure that the following operating mode is set: Channel - Sweep - Sweep 
Type: Lin. Frequency, Trace - Meas.: S21, Trace - Format: dB Mag.  
Open the Trace - Meas menu and select the measured quantity to be transformed (preferably: an S-
parameter or ratio).  
Click Trace - Traces - Time Domain Transform - Time Domain to transform the active trace to the time 
domain.  
The transformed trace is displayed in the active diagram area. To refine the analysis, you can modify the 
number and position of the sweep points (Channel - Sweep - Number of Points, Channel - Stimulus), 
select a different transform type and window for profiling (Trace - Traces - Time Domain Transform - 
Define Transform...) and vary the time range (Trace - Traces - Time Domain Transform - Stimulus 
Transform). 
To eliminate an unwanted response from the measured trace... 
Calculate a time domain transform as described above, Setting up a time domain measurement.  
Click Trace - Traces - Time Domain Transform - Define Time Gate and place a time gate on the 
transformed trace. To remove a single response use a Notch gate type. To remove everything outside 
aspecified area, use a Bandpass gate type.  
Click Trace - Traces - Time Domain Transform - Time Gate to enable the time gate. If necessary, go 
back to the Define Time Gate dialog to correct your time gate settings.   
Click Trace - Trace Func - Transform - Frequency Domain.  
The frequency domain trace now corresponds to the modified time domain trace which is transformed 
back into the frequency domain.
Distance-to-Fault Measurements     
With option ZVL-K2, Distance-to-Fault, the analyzer can locate faults and discontinuities on cables and 
transmission lines. The faults produce spikes on the trace; the x-axis shows the distance from the 
calibrated reference plane. Most conveniently the reference plane is set to the analyzer test port PORT 2.
Preparing a Distance-to-Fault Measurement 
Consider two RG141A cables of different length. The first cable can be connected to PORT 2 of the 
analyzer using a suitable adapter. The second cable is connected to the first cable using a second adapter 
and either left open or terminated with a matched load. The first cable contains a fault which is to be 
located by the network analyzer. The lengths of the cables and adapters are shown below.  
VB.NET Create PDF Library SDK to convert PDF from other file
Load PDF from stream programmatically in VB.NET. VB.NET: DLLs for Creating PDF. Add necessary references: RasterEdge.Imaging.Basic.dll.
edit pdf metadata online; pdf xmp metadata
C# PDF: PDF Document Viewer & Reader SDK for Windows Forms
without any dependency on Adobe products; Add PDF viewing and Support navigating, zooming, annotating and saving PDF in C# WinForms project programmatically;
adding metadata to pdf files; change pdf metadata
R&S ZVL 
Measurement Examples 
Advanced Tasks 
Operating Manual 1303.6580.32-05 
111
Mkr 2 shows the position of the fault in the first cable.  
Mkr 3 shows the position of the second adapter between the first and the second cable, causing a 
reflection at either side. The distance between the two maxima of this spike corresponds to the 
length of the adapter.  
Mkr 4 shows the reflection at the end of the second cable due to a small residual mismatch.  
With an open second cable, the result looks a little more complex: 
Compared to the situation with matched termination, the spikes at the positions of Mkr 1, Mkr 2, 
and Mkr 3 are virtually unchanged.   
Mkr 4 shows the reflection at the end of the second cable. The open cable reflects almost the 
entire signal power; full one-port calibration normalizes the trace so that the maximum of the spike 
is close to 0 dB.  
The spikes beyond Mkr 4 indicate multiple reflections. E.g. the double spike at the position of 
Mkr5 corresponds to a signal that has once traveled through the first cable (in both directions) and 
twice traveled through the second line due to bouncing between open end and adapter.   
Network (De-)embedding     
The Virtual Transform menu provides the functions for network embedding and deembedding. The 
following examples show how to use the functions.  
To add a virtual transformation network to a port... 
Suppose that your DUT is a 2-port DUT.  
Connect the DUT to your analyzer: For a two-port measurement, connect the DUT between test ports 
1and 2 of the analyzer.  
Establish the necessary Channel settings (port configuration, sweep range and type etc.) and select 
the measured quantities (Trace – Meas).  
Click Channel – Cal – Port Extension – Embedding to open the Embedding... dialog.  
In the dialog, select the Port to which you want to add a virtual transformation network (e.g. Port 1) 
and do one of the following:  
R&S ZVL 
Measurement Examples 
Advanced Tasks 
Operating Manual 1303.6580.32-05 
113
If you wish to define the added network by an imported set of 2-port S-parameters, select the 2-Port 
transformation network, click Read Data From File, and load the parameters from a 2-port (*.s2p) 
Touchstone file.
If you wish to define the added network by an equivalent circuit, select the circuit type and adjust the 
parameters R, C, and L displayed in the right half of the dialog.  
Click Embed DUT and Close the dialog.  
The traces of the active channel show the characteristics of the DUT including the virtual transformation 
network inserted between port 1 of the analyzer and the DUT.  
To remove a real transformation network from a port... 
Suppose that your 2-port DUT has a real, known 2-port transformation network at port 1.  
Connect the DUT (including the transformation network) between test ports 1 and 2 of the analyzer so 
that the transformation network is placed between port 1 of the analyzer and the DUT.  
Establish the necessary Channel settings (port configuration, sweep range and type etc.) and select 
the measured quantities (Trace – Measure).   
Click Channel – Cal – Port Extension – Deembedding to open the Deembedding... dialog.  
In the dialog, select the Port connected to the physical transformation network (e.g. Port 1) and do one 
of the following:  
If your transformation network is described by an imported set of 2-port S-parameters, select the 2-
Port transformation network, click Read Data From File, and load the data from a 2-port (*.s2p) 
Touchstone file.
If your transformation network can be described by an equivalent circuit, select the circuit type and 
adjust the parameters R, C, and L displayed in the right half of the dialog.  
Click Deembed DUT and Close the dialog.  
The traces of the active channel show the characteristics of the DUT without the effects of the removed 
physical transformation network.  