R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
175
Remote control:
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON
CALCulate<Chn>:MARKer<Mk>:REFerence:Y? 
Delta Mode  
Converts the active marker to a delta marker so that its values are measured and indicated relative to the 
reference marker (toggle function). A Q sign placed in front of the marker line indicates that the marker is 
in Delta Mode.  
The reference marker itself can not be set to delta mode but must be present when another marker is set 
to delta mode. The analyzer takes into account these conditions when Delta Mode is selected:  
If Delta Mode is selected while the reference marker is active, the marker in the info list after the 
reference marker is activated and set to delta mode. If the current trace contains the reference 
marker only, a new Mkr 1 is created and set to delta mode.  
If Delta Mode is selected for a normal marker while the current trace contains no reference 
marker, a reference marker is created.  
Remote control: CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe] ON
Ref.Marker -> Marker  
Places the reference marker to the position of the active marker. As a consequence, the active marker 
takes the role of a reference marker.   
Ref.Marker -> Marker is not active if the active marker is a reference marker.   
Remote control:
All Mkrs Off 
Removes all markers from all traces of the active setup. The removed markers remember their properties 
(stimulus value, format, delta mode, number) when they are restored. The marker properties are definitely 
lost if the associated trace is deleted.   
Remote control:
CALCulate<Chn>:MARKer<Mk>:AOFF
Mkr Format  
Opens a submenu to select an output format for the (complex) active marker value in the marker info field. 
The default marker format is the format of the associated trace. The current format is indicated with a .
Metadata in pdf documents - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
pdf xmp metadata; pdf metadata
Metadata in pdf documents - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
remove metadata from pdf; batch pdf metadata
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
176
All marker formats are available irrespective of the measured quantity. The output values are calculated 
by a simple conversion of a complex measurement result, where the marker format defines the conversion 
rules. This flexibility in the calculation must be kept in mind when interpreting the results and physical units 
displayed; see also Measured Quantities and Display Formats.   
Short description of marker formats 
The formats of the markers assigned to a trace are independent of each other and of the trace format 
settings. The following table gives an overview on how a complex marker value z = x + jy is converted.  
Marker Format 
Description 
Formula 
dB Mag 
Magnitude of z in dB 
|z| = sqrt ( x
2
+y
2
)
dB Mag(z) = 20 * log|z| dB  
Lin Mag 
Magnitude of z, unconverted 
|z| = sqrt ( x
2
+y
2
)
Phase 
Phase of z 
X(z) = arctan (y/x) 
Real 
Real part of z 
Re(z) = x 
Imag 
Imaginary part of z 
Im(z) = y 
SWR 
(Voltage) Standing Wave Ratio 
SWR = (1 + |z|) / (1 – |z|) 
Delay 
Group delay, neg. derivative of the phase response*
)
–d X (z) / d Y
dB Mag and Phase 
Magnitude of z in dB and phase in two lines 
20 * log|z| dB 
arctan ( Im(z) / Re(z) ) 
Lin Mag and Phase 
Magnitude of z (unconverted) and phase in two lines 
|z| 
arctan ( Im(z) / Re(z) ) 
Real and Imag 
Real and imaginary part of z in two lines 
x
y
Default (Trace) 
Marker format identical with trace format 
– 
R+jX
Unnormalized resistance and reactance; L or C in three lines 
(Smith diagram) 
R
X
Lor C**
)
G+jB
Unnormalized conductance and susceptance; L or C in three lines
G
C# PDF Library SDK to view, edit, convert, process PDF file for C#
NET empowers C# developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. C#.NET: Edit PDF Metadata.
view pdf metadata in explorer; analyze pdf metadata
C# PDF Print Library: Print PDF documents in C#.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
pdf metadata editor; read pdf metadata online
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
177
(Inverted Smith diagram) 
Lor C**
)
*
)
The delay aperture is defined in the Trace – Format menu.  
**
)
The equivalent inductances or capacitances
Lor C are calculated from the imaginary part of the 
impedance according to 
Remote control:
CALCulate<Chn>:MARKer<Mk>:FORMat ...
More Mkrs  
Opens a submenu to create the markers numbered 4 to 10. The markers are analogous to marker no. 1 to 
3.  
Remote control: CALCulate<Chn>:MARKer<Mk>[:STATe] ON
Coupled Mkrs  
Couples the markers of all traces in the active setup to the markers of the active trace (toggle function). 
While marker coupling is active, the active trace markers assume the role of master markers; the other 
markers behave as slave markers, following any change of position of the master marker.  
Effects of marker coupling 
The concept of marker coupling means that corresponding markers on different traces (i.e. markers with 
the same number or reference markers) are positioned to the same stimulus values but keep their 
independent format and type settings. When a trace with markers is selected as the active trace and 
marker coupling is switched on, the following happens: 
The active trace and all associated markers are left unchanged. The active trace markers become 
the master markers of the setup.  
Markers on the other traces which have no corresponding master marker are removed but 
remember their properties and can be re-activated after the coupling is released.  
The remaining markers on the other traces become slave markers and are moved to the position 
of the corresponding master markers. "Missing" slave markers are created so that each trace has 
the same number of markers placed at the same position.  
If the position of a master marker is outside the sweep range of the slave trace, the slave marker 
is displayed at the edge of the diagram. The marker info field indicates an invalid measurement 
result: 
While marker coupling is active, it is possible to: 
Move a master marker and thus change the position of all corresponding slave markers.  
Activate another trace in order to make the associated markers the new master markers.  
Marker coupling makes sense only if the master and the slave traces use the same stimulus variable. 
Channels with a different stimulus variable (sweep type) are not coupled.  
VB.NET PDF Print Library: Print PDF documents in vb.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
c# read pdf metadata; read pdf metadata java
C# PDF File Compress Library: Compress reduce PDF size in C#.net
compressing control is designed to offer C# developers to compress existing PDF documents in .NET Document and metadata. C#.NET DLLs: Compress PDF Document.
rename pdf files from metadata; remove metadata from pdf acrobat
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
178
Remote control:
CALCulate<Chn>:MARKer<Mk>:COUPled[:STATe] ON
Discrete Mkrs  
Turns the active marker into a discrete marker and vice versa.  
The stimulus value of a discrete marker always coincides with a sweep point. Use discrete markers in 
order to avoid that the marker indicates an interpolated measurement value. 
Remote control:
CALCulate<Chn>:MARKer<Mk>:MODE  
CALCulate<Chn>:MARKer<Mk>:REFerence:MODE
Mkr Properties...  
Opens a dialog to define the properties of all markers of the active trace.  
In the left part the dialog contains four input fields and drop-down lists to select a marker and define the 
basic properties of a selected marker which is switched on in the Mode panel.   
Select Mkr. opens a drop-down list to select the Ref. Marker or one of the Marker 1, ... 10 that 
may be associated with the active trace.  
Name defines a marker name which may exceed the length of the input box and contain letters, 
numbers, blanks and special characters.  
Stimulus defines the stimulus value (in Cartesian diagrams: the x-axis position) of the marker.  
Format defines the output format of the marker value.  
The Mode panel contains check boxes to select properties that are related to the marker positions. All 
properties can be combined. 
Marker On displays the selected marker in the diagram and in the marker info field or removes it. 
Marker configurations are only available while the marker is switched On. A removed marker 
remembers its properties (stimulus value, format, delta mode, number ...) when it is switched on 
again. The marker properties are definitely lost if the associated trace is deleted.    
Fixed Marker freezes thecurrent response value of the selected marker. The marker can still be 
shifted horizontally but the vertical position remains fixed if the other marker settings are changed. 
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
VB.NET Guide and Sample Codes to Merge PDF Documents in VB.NET Project. Batch merge PDF documents in Visual Basic .NET class program.
batch edit pdf metadata; adding metadata to pdf files
VB.NET PDF Library SDK to view, edit, convert, process PDF file
for .NET empowers VB.NET developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. PDF Metadata Edit.
remove metadata from pdf; pdf metadata editor online
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
179
Markers must be inside the sweep range and have a valid response value when they are fixed.     
Delta Mode sets the selected marker to delta mode and displays its values relative to the 
reference marker.     
Discrete Mode means that a marker can be set to discrete sweep points only. If discrete mode is 
switched off, the marker can be positioned on any point of the trace, and its response values are 
obtained by interpolation.     
All Mkrs Coupled couples the markers of all traces in the active setup to the markers of the active 
trace; see Coupled Markers.  
The Style settings qualify how the selected marker is displayed on the screen. 
The Show Info settings select the marker info to be displayed at the Marker Position, in the marker info 
field or in a separate Table below the diagram area. The display options may be selected simultaneously 
or all switched off. The table provides more information than the marker info field:  
The Reduced table provides a condensed overview with less entries.  
Remote control:
CALCulate<Chn>:MARKer<Mk>[:STATe] ON  
CALCulate<Chn>:MARKer<Mk>:TYPE NORMal | FIXed  
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE NORMal | FIXed 
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe] ON  
CALCulate<Chn>:MARKer<Mk>:MODE CONTinuous | DISCrete  
CALCulate<Chn>:MARKer<Mk>:COUPled[:STATe] ON  
Export Mkrs...  
Calls up a Save As... dialog to store the current marker values to a marker file.   
The analyzer uses a simple ASCII format to export marker values. By default, the marker file extension is 
*.txt. The file contains all traces in the active setup together with their names and measured quantities. 
Below each trace, the file shows a list of all markers with their names, stimulus and response values.   
The following example of a marker file describes a setup with two traces Trc1 and its memory trace. Trc1 
has no markers assigned, the memory trace has four markers named Mkr 1, ..., Mkr 4.  
Remote control:
MMEMory:STORe:MARKer "file_name"
Meas     
The Measure submenu selects the quantity to be measured and displayed.  
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
C#.NET PDF Library - Merge PDF Documents in C#.NET. Provide NET components for batch combining PDF documents in C#.NET class. Powerful
remove pdf metadata online; remove pdf metadata
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
Our .NET SDKs own the most advanced & comprehensive documents and images reading and has a mature imaging utility which allows Tiff image file metadata to be
read pdf metadata; google search pdf metadata
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
180
S11, S12, S21, S22 select the four elements of the standard 2-port scattering matrix (S-
parameters).  
S-Param Wizard opens a series of dialogs providing the settings for a standard S-parameter 
measurement. 
Distance-to-Fault enables or disables the location of a discontinuity on a transmission line (with 
option R&S ZVL-K2).  
Define Distance-to-Fault opens a submenu to configure the Distance-to-Fault measurement.  
Impedance opens a submenu to convert reflection S-parameters into matched-circuit impedances 
(converted Z-parameters).  
Admittance opens a submenu to convert reflection S-parameters into matched-circuit admittances 
(converted Y-parameters).  
Stability Factor selects one of the three factors K, µ
1
or µ
2
to assess the stability of linear circuits 
(e.g. amplifiers). 
Port assignments of the DUT and the analyzer 
S-parameters S
<out>< in> 
characterize the Device Under Test, so the indices <out> and < in> denote the 
output and input ports of the DUT. Analogously the waves a
1
and a
2
are referred to as incident waves, b
1
and b
2
to as outgoing waves. 
The analyzer measures the waves a
1
,b
1
,... at its own test ports in order to obtain S-parameters, ratios 
and other derived quantities. The test ports of the analyzer are numbered, so it is convenient to use them 
as a reference, defining port number n of the DUT as the port connected to test port n of the analyzer. 
With this convention the waves a
n
and b
n
have the following meaning: 
a
n
is the wave transmitted at test port no. n of the analyzer (stimulus signal) and fed to port 
number n of the DUT (incident wave). 
b
n
is the wave transmitted at port number n of the DUT (response signal) and received at test port 
no. n of the analyzer (received wave).    
The analyzer offers various diagrams and display formats to visualize the results. The analyzer places 
no restriction on the formats that are available for each measured quantity. However, to keep the 
interpretation simple, it is recommended to select an appropriate display format; see Trace - 
Format.  
S11, S12, S21, S22  
Select one of the four elements of the standard 2-port scattering matrix (S-parameters).  
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
181
The S-parameters are the basic measured quantities of a network analyzer. They describe how the DUT 
modifies a signal that is transmitted or reflected in forward or reverse direction. S-parameters are 
expressed as S
<out>< in>
,where <out> and <in> denote the output and input port numbers of the DUT.  
Afull 2-port S-parameter measurement involves 2 stages (referred to as partial measurements) with 
interchanged drive and receive ports. The analyzer automatically switches the internal power sources and 
the receivers to obtain the desired S-parameters.
Remote 
control: 
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "S11" | "S12" | 
"S21" | "S22"  
[SENSe<Chn>:]FUNCtion[:ON] "...:POWer:S<11 | 12 | 21 | 22>"  
Create new trace and select name and measurement parameter: 
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "S11" | "S12" | 
"S21" | "S22"  
S-Parameter Wizard     
The S-Parameter Wizard consists of a series of dialogs providing the settings for a standard S-parameter 
measurement.  
The measurement comprises the following stages:  
Select the measurement parameters and the diagram areas 
Choose an S-parameter or a group of S-parameters to be measured and displayed. The 
measurement parameters are displayed in separate diagram areas: Cartesian diagram areas are 
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
182
used for transmission S-parameters, Cartesian or Smith diagrams for reflection parameters.  
Select the sweep settings 
Choose the frequency range and the number of measurement points per sweep. The sweep 
range is defined by two values (start and stop frequency or center frequency and span). The 
measurement points are equidistant across the sweep range. Increasing the number of points 
also increases the measurement time per single sweep.  
Select the measurement bandwidth and source power 
Choose a typical measurement bandwidth and one of two typical source power values. A smaller 
measurement bandwidth increases the dynamic range but slows down the measurement. A 
smaller source power protects the input port of the analyzer from being overdriven, if an active 
DUT with high gain is measured.   
Perform a calibration (optional) 
Call up the calibration wizard and perform a calibration. The calibration type is a full 2-port 
(TOSM) calibration. This calibration can be performed as well using automatic calibration (if 
accessory R&S ZV-Z5x, Calibration Unit, is available).  
You can skip the automatic calibration stage (select Finish without Calibration) if  
Avalid calibration is already assigned to the active channel 
You want to apply a valid calibration stored in the cal pool 
You want to perform the calibration manually, e.g. because you wish to select another 
calibration type.  
You don't want to use a calibration, e.g. because the factory calibration is accurate 
enough for your measurement.  
In order to obtain a predictable result the measurement wizard has to reset all settings except the 
current calibration data. Store your setup if you don't want to lose the current configuration. 
Distance-to-Fault     
Opens a submenu to locate faults and discontinuities on cables. The measurement requires option ZVL-
K2 , Distance-to-Fault.  
Distance-to-fault measurement 
The network analyzer measures and displays complex S-parameters and other quantities as a function of 
the frequency. The measurement results can be filtered and mathematically transformed in order to obtain 
the impulse response, which often gives a clearer insight into the characteristics of the DUT.  
The impulse response is special type of a time-domain representation. A fault (discontinuity) on a 
transmission line causing reflection results in a spike on the impulse response trace of the reflection 
coefficient. The distance between the reference plane and the fault can be calculated from the 
propagation time, taking into account the electrical properties of the transmission line (Cable Type). The 
following example shows the reflection at the end of an open cable with a length of approx. 959 mm (Mkr 
2position). The cable is connected to the test port no. 2 of the analyzer. A full one-port calibration ensures 
that the distance is measured from the test port position (reference plane = distance zero). The additional 
peak at the position of Mkr 1 is due to the connection between the cable and the test port.  
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
183
The parameters of the impulse response and frequency-domain representation are related due to the 
mathematical properties of the time-domain transform. In practice, this means that the sweep points may 
have to be adjusted to the length of the transmission line and the expected distance-to-fault:  
The impulse-response resolution is proportional to the measured sweep range: The larger the 
sweep range (span), the better the resolution.   
At fixed frequency span, the maximum distance that can be measured is proportional to the 
number of sweep points (i.e. inversely proportional to the Frequency Step Size). The larger the 
Number of Points, the longer the maximum distance between the calibrated reference plane (most 
conveniently, PORT 2 of the analyzer) and the located fault.  
The R&S ZVL can automatically determine a suitable number of sweep points (Auto Number of 
Points). Moreover it is possible to define which of the peaks will be considered as being due to a 
fault, and to draw up and export a list of the detected peaks.  
Start Distance and Stop Distance define the distance window for the distance-to-fault 
measurement. 
Stimulus Frequency opens a submenu to define the frequency stimulus axis.  
Auto Number of Points adjusts the number of sweep points in order to provide an unambiguous 
display of the fault locations.  
Cable Type opens a dialog to select a standard (predefined) cable type or add a cable with 
arbitrary permittivity.  
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
184
Full One-Port P2 Cal opens the calibration wizard to perform a full one-port calibration at port no. 
2and re-adjust the reference plane (distance zero).  
Fault Limit defines a minimum response value in dB for a trace maximum to be considered as 
being due to a fault.  
Show Fault List displays a list of all peaks that fulfil the Fault Limit condition.  
Export Fault List opens a Save As... dialog to write the fault list data to an ASCII file.   
Distance-to-Fault 
Enables or disables the location of a discontinuity on a transmission line (with option R&S ZVL-K2). The 
distance-to-fault measurement requires a linear frequency sweep (Channel – Sweep – Sweep Type – Lin. 
Frequency). The analyzer automatically disables the distance-to-fault measurement as soon as a different 
sweep type is selected.  
Default settings for distance-to-fault 
The following settings are performed automatically when a distance-to-fault measurement is enabled 
(Trace – Meas– Distance-to-Fault: On): 
The reflection coefficient S22 is selected as measured quantity.  
The stimulus axis shows a distance window. The start and stop distance is displayed below the 
channel settings.  
The cable type (e.g. Ideal Air Line) is displayed in the trace list. A context menu opens the Cable 
Type dialog.  
The number of points is increased, if necessary. See Auto Number of Points. 
Remote control:
CALCulate<Chn>:TRANsform:DTFault:STATe
Start Distance..., Stop Distance... 
Defines the distance range (x-axis range) for the distance-to-fault measurement. The Start Distance and 
the Stop Distance correspond to the left and right edge of the diagram area; they are shown below the 
channel settings:  
Distance trace settings 
While a distance-to-fault measurement is active the trace settings behave as follows:  
Start Distance... and Stop Distance... configure the distance axis.  
dB Mag is the recommended Trace Format, although other formats are available.  
Limit lines can be defined like the limit lines for time sweeps.  
The marker and bandfilter search functions are available for the transformed trace.  
Note that only the reflection coefficient S
22
(and the corresponding impedance and admittance 
parameters) are available as measured quantities.  
Remote control:
CALCulate<Chn>:TRANsform:DTFault:STARt 
CALCulate<Chn>:TRANsform:DTFault:STOP 
CALCulate<Chn>:TRANsform:DTFault:CENTer 
CALCulate<Chn>:TRANsform:DTFault:SPAN 