how to disable save option in pdf using c# : Remove metadata from pdf acrobat software application project winforms html azure UWP ZVL_Operating_008_V3_2021-part1061

R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
195
Y= Frequency/angular velocity in radians/s 
f= Frequency in Hz 
In practice, the analyzer calculates an approximation to the derivative of the phase response, taking a 
small frequency interval Qf and determining the corresponding phase change Qa. The delay is thus 
computed as: 
The aperture Qf must be adjusted to the conditions of the measurement. 
If the delay is constant over the considered frequency range (non-dispersive DUT, e.g. a cable), then `
g
and `
g,meas
are identical and  
where Qt is the propagation time of the wave across the DUT, which often can be expressed in terms of its 
mechanical length L
mech
,the permittivity ], and the velocity of light c. The product of L
mech
.
sqrt(N) is termed 
the electrical length of the DUT and is always larger or equal than the mechanical length (] > 1 for all 
dielectrics and ] = 1 for the vacuum).  
Application: Transmission measurements, especially with the purpose of investigating deviations from 
linear phase response and phase distortions. To obtain the delay a frequency sweep must be active.  
The cables connecting the analyzer test ports to the DUT introduce an unwanted delay, which often 
can be assumed to be constant. Use the Zero Delay at Marker function, define a numeric length Offset or 
use the Auto Length function to mathematically compensate for this effect in the measurement results. To 
compensate for a frequency-dependent delay in the test setup, a system error correction is required.  
Remote control:
CALCulate<Chn>:FORMat GDELay  
Aperture 
Sets a delay aperture for the delay calculation. The aperture Qf is entered as an integer number of 
Aperture Steps: 
The analyzer calculates the aperture from the sweep points of the current frequency sweep.  
Properties: The delay at each sweep point is computed as: 
where the aperture Qf is a finite frequency interval around the sweep point f
o
and the analyzer measures 
the corresponding phase change Qa.
Remove metadata from pdf acrobat - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
edit pdf metadata; modify pdf metadata
Remove metadata from pdf acrobat - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
embed metadata in pdf; pdf metadata extract
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
196
Calculation of Jf and JK 
With a given number of aperture steps n  the delay at sweep point no. m is calculated as follows: 
If n is even (n = 2k), then Qf (m) = f (m+k) – f (m–k) and Qa(m) = Qa (m+k) – Qa (m–k).   
If n is odd (n = 2k+1), then Qf (m) = f (m+k) – f (m–k–1) and (m)= Qa (m+k) – Qa (m–k–1).   
The delay calculation is based on the already measured sweep points and does not slow down the 
measurement. 
Qf is constant over the entire sweep range, if the sweep type is a Lin. Frequency sweep. For Log. 
Frequency and Segmented Frequency sweeps, it varies with the sweep point number m.   
Application The aperture must be adjusted to the conditions of the measurement. A small aperture 
increases the noise in the group delay; a large aperture tends to minimize the noise effects, but at the 
expense of frequency resolution. Phase perturbations which are narrower in frequency than the aperture 
tend to be smeared over and can not be measured.  
Remote control:
CALCulate<Chn>:GDAPerture:SCOunt  
SWR 
Calculates the Standing Wave Ratio (SWR) from the measured quantity (primarily: from a reflection S-
parameter) and displays it in a Cartesian diagram.  
Properties: The SWR (or Voltage Standing Wave  Ratio, VSWR) is a measure of the power reflected at 
the input of the DUT. It is calculated from the magnitude of the reflection coefficients S
ii
(where i denotes 
the port number of the DUT) according to: 
The superposition of the incident and the reflected wave on the transmission line connecting the analyzer 
and the DUT causes an interference pattern with variable envelope voltage. The SWR is the ratio of the 
maximum voltage to the minimum envelope voltage along the line.  
Interpretation of the SWR 
The superposition of the incident wave I and the reflected wave R on the transmission line connecting the 
analyzer and the DUT causes an interference pattern with variable envelope voltage. The SWR is the ratio 
of the maximum voltage to the minimum envelope voltage along the line: 
C# PDF Converter Library SDK to convert PDF to other file formats
manipulate & convert standard PDF documents in .NET class applications independently, without using other external third-party dependencies like Adobe Acrobat.
pdf metadata viewer online; preview edit pdf metadata
.NET PDF Document Viewing, Annotation, Conversion & Processing
Read PDF metadata. Search text content inside PDF. Insert images into PDF. Edit, remove images from PDF. Print. Support for all the print modes in Acrobat PDF.
add metadata to pdf; batch pdf metadata editor
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
197
SWR = V
Max
/V
Min
= (|V
I
|+ |V
R
|) / (|V
I
|– |V
R
|) = (1 + |S
ii
|) / (1 – |S
ii
|) 
Application: Reflection measurements with conversion of the complex S-parameter to a real SWR. 
Remote control: CALCulate<Chn>:FORMat SWR 
Lin Mag 
Selects a Cartesian diagram with a linear vertical axis scale to display the magnitude of the measured 
quantity.  
Properties: The stimulus variable appears on the horizontal axis, scaled linearly. The magnitude of the 
complex quantity C, i.e. |C| = sqrt ( Re(C)
2
+Im(C)
2
), appears on the vertical axis, also scaled linearly.  
Application: Real measurement data (i.e. the Stability Factors, DC Input 1/2, and the PAE) are always 
displayed in a Lin Mag diagram. 
Alternative Formats 
The magnitude of each complex quantity can be displayed on a logarithmic scale. It is possible to view the 
real and imaginary parts instead of the magnitude and phase. 
Remote control: CALCulate<Chn>:FORMat MLINear  
Real 
Selects a Cartesian diagram to display the real part of a complex measured quantity.  
Properties: The stimulus variable appears on the horizontal axis, scaled linearly. The real part Re(C) of 
the complex quantity C = Re(C) + j Im(C), appears on the vertical axis, also scaled linearly.  
Application: The real part of an impedance corresponds to its resistive portion.   
Alternative Formats 
It is possible to view the magnitude and phase of a complex quantity instead of the real and imaginary 
part. The magnitude can be displayed on a linear scale or on a logarithmic scale. Both the real and 
imaginary parts are displayed in the polar diagram. 
Remote control:
CALCulate<Chn>:FORMat REAL  
Imag 
Selects a Cartesian diagram to display the imaginary part of a complex measured quantity. 
Properties: The stimulus variable appears on the horizontal axis, scaled linearly. The imaginary part 
Im(C) of the complex quantity C = Re(C) + j Im(C), appears on the vertical axis, also scaled linearly.  
Application: The imaginary part of an impedance corresponds to its reactive portion. Positive (negative) 
values represent inductive (capacitive) reactance.   
Alternative Formats 
It is possible to view the magnitude and phase of a complex quantity instead of the real and imaginary 
part. The magnitude can be displayed on a linear scale or on a logarithmic scale. Both the real and 
C# Windows Viewer - Image and Document Conversion & Rendering in
standard image and document in .NET class applications independently, without using other external third-party dependencies like Adobe Acrobat. Convert to PDF.
delete metadata from pdf; add metadata to pdf programmatically
C# Word - Word Conversion in C#.NET
Word documents in .NET class applications independently, without using other external third-party dependencies like Adobe Acrobat. Word to PDF Conversion.
pdf remove metadata; pdf keywords metadata
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
198
imaginary parts are displayed in the polar diagram. 
Remote control:
CALCulate<Chn>:FORMat IMAGinary  
Inv Smith 
Selects an inverted Smith chart to display a complex quantity, primarily a reflection S-parameter. 
Properties: The Inverted Smith chart is a circular diagram obtained by mapping the positive complex 
semi-plane into a unit circle. If the measured quantity is a complex reflection coefficient (S
11
,S
22
etc.), then 
the unit Inverted Smith chart represents the normalized admittance. In contrast to the polar diagram, the 
scaling of the diagram is not linear.   
Application: Reflection measurements, see application example.  
The axis for the sweep variable is lost in Smith charts but the marker functions easily provide the 
stimulus value of any measurement point. dB values for the magnitude and other conversions can be 
obtained by means of the Marker Format functions. 
Remote control:
CALCulate<Chn>:FORMat ISMith 
Unwrapped Phase 
Selects a Cartesian diagram with an arbitrarily scaled linear vertical axis to display the phase of the 
measured quantity. 
Properties: The stimulus variable appears on the horizontal axis, scaled linearly. The phase of the 
complex quantity C, i.e. X (C) = arctan ( Im(C) / Re(C) ), appears on the vertical axis. X (C) is measured 
relative to the phase at the start of the sweep (reference phase = 0°). In contrast to the normal Phase 
format, the display range is not limited to values between –180° and +180°. This avoids artificial jumps of 
the trace but can entail  a relatively wide phase range if the sweep span is large.  
Application: Phase measurements, e.g. phase distortion, deviation from linearity. 
After changing to the Unwrapped Phase format, use Trace – Scale – Autoscale to re-scale the vertical 
axis and view the entire trace.  
Remote control:
CALCulate<Chn>:FORMat UPHase   
Scale     
The Scale settings define how the current trace is presented in the diagram selected in the Format 
submenu.  
VB.NET PDF: How to Create Watermark on PDF Document within
logo) on any desired PDF page. And with our PDF Watermark Creator, users need no external application plugin, like Adobe Acrobat.
remove pdf metadata online; edit pdf metadata online
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
199
Autoscale adjusts the diagram scale in order to display the entire active trace in the diagram area.  
Autoscale All adjusts the diagram scale in order to display all traces in the diagram area.  
Scale/Div sets the value of the vertical diagram divisions.  
Ref Value sets the reference line of a Cartesian diagram or the outer circumference of a polar 
diagram. 
Ref Position defines the position of the reference line in a Cartesian diagram.  
Maximize maximizes the active diagram area. 
Overlay All displays all traces in one diagram area. 
Split All displays every trace in a separate diagram area. 
The Scale settings are closely related to the settings in the Format submenu and in the Display menu. All 
of them have an influence on the way the analyzer presents data on the screen. 
The Scale settings depend on the diagram type (Trace – Format) because not all diagrams can be scaled 
in the same way:  
In Cartesian diagrams, all scale settings are available.  
In circular diagrams, no Scale/Div., no Ref. Position, and no Max and Min values can be defined. 
The default scale is activated automatically when a display format (diagram type) is selected. Scale 
settings that are not compatible with the current display format are disabled (grayed).  
Relations between the scaling parameters 
The scaling parameters Scale / Div, Ref Value, Ref Position, Max, Min are coupled together in the 
following manner:  
Max – Min = Scale / Div * <Number of graticule divisions>  
Max = Ref Value when Ref Position is 10  
Min = Ref Value when Ref Position is 0  
The Marker -> provide a convenient alternative to manual diagram scaling.  
Autoscale 
Adjusts the Scale Divisions and the Ref. Value in order to display the entire active trace in the diagram 
area, leaving an appropriate display margin. 
In Cartesian diagrams, the analyzer re-calculates the values of the vertical divisions so that the 
trace fits onto 80% of the vertical grid. The reference value is chosen to center the trace in the 
diagram.     
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
200
In circular diagrams (Polar, Smith, Inverted Smith), the analyzer re-calculates the values of the 
radial divisions so that the diagram is confined to approx. 80% of the outer circumference. The 
reference value is set to the value of the outer circumference. 
Autoscale does not affect the stimulus values and the horizontal axis.     
Remote control:
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y[:SCALe]:AUTO ONCE
Autoscale All 
Adjusts the Scale Divisions and the Ref. Value in order to display all traces in the diagram area, leaving an 
appropriate display margin. This scale settings are analogous to the Autoscale function. The traces in the 
active diagram area are taken into account irrespective of their channel assignment.   
Scale / Div 
Sets the value of the vertical diagram divisions in Cartesian diagrams. 
Scale /Div corresponds to the increment between two consecutive grid lines. The unit depends on the 
display format: dB for display format dB Mag, degrees for Phase and Unwrapped Phase, ns for Delay, U
(units) for all other (dimensionless) formats. 
Scale /Div is not available (grayed) for circular diagrams (Polar, Smith, Inverted Smith).
Remote control: DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:PDIVision
Ref Value  
Sets the reference line of a Cartesian diagram or the outer circumference of a circular diagram. 
In Cartesian diagrams Ref. Value defines the value of the reference line, indicated by a 
symbol at the right edge of the diagram area. The color of the symbol corresponds to the trace 
color. As the Ref. Value is varied, the position of the reference line (Ref. Position) is left 
unchanged, so that the current trace is shifted in vertical direction. The unit of the Ref. Value 
depends on the display format: dB for display format dB Mag, degrees for Phase and Unwrapped 
Phase, ns for Delay, U (units) for all other (dimensionless) formats.     
In circular diagrams (Polar, Smith, Inverted Smith), Ref. Value defines the value of the outer 
circumference. Changing Ref. Value enlarges or scales down the diagram, leaving the center 
unchanged. The unit is U (units) for all circular diagrams. 
Use the paste marker list for convenient entry of the reference value.  
Remote control:
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:RLEVel  
Ref Position  
Defines the position of the reference line in a Cartesian diagram. 
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
201
The reference line is indicated by a 
symbol at the right edge of the diagram area. The color of the 
symbol corresponds to the trace color. Ref. Position is defined on a linear scale between 0 (bottom line of 
the diagram) and 10 (top line of the diagram). As the Ref. Position is varied, the value of the reference line 
(Ref. Value) is left unchanged, so the current trace is shifted together with the Ref. Position.
Ref. Position is not available (grayed) for polar diagrams (Polar, Smith, Inverted Smith).
Use the analyzer's drag-and-drop functionality to move the reference line symbol to the desired 
position. 
Remote control:
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:RPOSition  
Maximize 
Maximizes all diagram areas of the active setup to occupy the whole window, placing the active diagram 
area on top. Clicking Maximize again restores the previous display configuration. 
Adouble-click into any point in the diagram area is equivalent to the Maximize function. To view all 
traces in a common, maximized diagram area, select Overlay all.
Remote control:
DISPlay:WINDow<Wnd>:MAXimize ON | OFF 
Overlay All 
Places all traces in a single diagram area which is maximized to occupy the whole window. This function 
is available irrespective of the trace Format and the Channel settings; it is even possible to overlay 
Cartesian and polar diagrams.   
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
202
The active trace and active channel is highlighted. The scaling of the axes corresponds to the active trace.  
To hide all traces except the active one, select Split All and Maximize.
Remote control: No command, display configuration only. 
Split All 
Splits the active window into as many diagram areas as there are traces and assigns a single trace to 
each area.   
Remote control: No command, display configuration only. 
Lines     
The commands in the Lines submenu define limits for the measurement results, visualize them in the 
diagrams and activate/deactivate the limit check. Besides the menu provides a horizontal line for each 
trace.  
Limit Lines 
Alimit line is a set of data to specify the allowed range for some or all points of a trace. Typically, limit 
lines are used to check whether a DUT conforms to the rated specifications (conformance testing).  
The upper limit line defines the maximum value for the trace points.  
The lower limit line defines the minimum value for the trace points.  
Alimit check consists of comparing the measurement results to the limit lines and display a pass/fail 
indication.  
Upper and lower limit lines are both defined as a combination of segments with a linear dependence 
between the measured quantity and the sweep variable (stimulus variable). The limit lines can be stored to 
afile and recalled. Data or memory traces can be used to define the segments of a limit line. Moreover it 
is possible to modify the limit lines globally by adding an offset to the stimulus or response values.  
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
203
(No direct access via front panel keys)
Show Limit Line displays or hides the limit line associated with the active trace.  
Limit Check On activates or deactivates the limit check.  
Define Limit Line opens a dialog to define, save or recall limit lines.
Horizontal Line displays or hides the horizontal line of the active trace and changes its position.  
Global Limit Check activates a global (composite) limit check on all traces in the active setup.  
Limit lines are available for all Cartesian diagram types (Trace – Format). For polar diagrams, the 
functions of the Lines submenu are grayed. The limit lines are hidden and the limit check is disabled when 
aCartesian trace format is replaced by a polar diagram.   
Show Limit Line 
Shows or hides the limit line associated with the active trace in a Cartesian diagram area. A checkmark 
appears next to the menu item when the limit line is shown.  
In the diagram, upper and lower limit lines can be displayed with different colors. Limit line segments with 
disabled limit check (see Define Limit Line) can also be colored differently. The limit line colors are defined 
in the Define User Color Scheme dialog (Display – Display Config. – Color Scheme...). 
Display of the limit line and limit check are independent of each other: Hiding the limit line does not 
switch off the limit check.  
Remote control:
CALCulate<Chn>:LIMit:DISPlay[:STATe] ON | OFF
R&S ZVL 
GUI Reference 
Trace Menu 
Operating Manual 1303.6580.32-06 
204
Limit Check On 
Switches the limit check of the active trace on or off. A checkmark appears next to the menu item when 
the limit check is enabled.  
When the limit check is switched on, a PASS or FAIL message is displayed in the center of the diagram. If 
the limit check fails at a measurement point, the two trace segments to the left and right of the point can 
change their color. The Limit Fail Trace color is defined in the Define User Color Scheme dialog (Display – 
Display Config. – Color Scheme...). 
Limit check and display of the limit lines are independent of each other: With disabled limit check, the 
limit line can still be displayed, however, no display elements indicating a limit failure will appear.  
If no limit lines are defined for the active trace, the limit check can be switched on but will always 
PASS the trace.  
Limits are checked at the actual measurement points, whereas a limit failure is indicated for the trace 
segments on both sides of a failed point. A small number of points causes wide trace segments so that the 
out-of tolerance regions can appear wider as they are.  
Markers show interpolated values. As a consequence, a trace segment can be failed, whereas a marker 
placed on the segment may show a response value within the allowed range. This is shown in the 
example below, with an upper limit line at 26 dB, and a marker response value of 25.968 dB. The small 
circles correspond to the sweep points; the orange part of the trace is failed.   
Remote control: CALCulate<Chn>:LIMit:STATe ON | OFF 
CALCulate<Chn>:LIMit:LOWer:STATe ON | OFF
CALCulate<Chn>:LIMit:UPPer:STATe ON | OFF
CALCulate<Chn>:LIMit:FAIL?  
Define Limit Line  
Opens a dialog to define the limit line for the active trace on a segment-by-segment basis. In each 
Documents you may be interested
Documents you may be interested