Standards. Hiding the diagram leaves more space for displaying the characteristics of the 
measured standards.  
Keep Measurement Data for >Repeat Previous Cal< causes the raw measurement data of the 
standards to be stored after the calibration is completed. This enables the Repeat Prev Cal... 
command, which can be used to optimize a previous calibration without repeating the 
measurement of all standards. If Keep Measurement Data... is not active, then the raw 
measurement data of the standards is deleted and the analyzer only stores the system error 
correction data. Deleting the raw data saves disk space.  
Apply is enabled as soon as data has been acquired for all standards. The button starts the 
calculation of the system error correction data and closes the calibration wizard. The current 
instrument settings are stored with the correction data. To avoid incompatibilities, older system 
error correction data is deleted unless it has been transferred into a Cal Pool using the Calibration 
Manager.  
The Keep Measurement Data for >Repeat Previous Cal< setting is valid for the current calibration 
only. To activate this function in general, use the parameter in the User Interface tab of the Sytem 
Configuration dialog (menu Nwa-Setup – System Config). 
Checks during the calculation of correction data 
Incompatibilities between the selected calibration type, the standards and the channel settings may cause 
the calibration to be inaccurate. The analyzer auto-detects potential sources of errors and displays 
appropriate, self-explanatory notice boxes.  
Remote control:
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe  
[SENSe<Ch>:]CORRection:COLLect[:ACQUire]:SELected 
[SENSe<Ch>:]CORRection:COLLect:SAVE 
[SENSe<Ch>:]CORRection:COLLect:DELete ["<cal_name>"]
Calibration Labels  
The following labels in the trace list inform you about the status or type of the current calibration.  
Label 
Meaning 
Cal 
The system error correction is applied without interpolation. This means that a set of measured 
correction data is available at each sweep point.  
Cal 
int 
The system error correction is applied, however, the correction data for at least one sweep 
point is interpolated from the measured values. This means that the channel settings have 
been changed so that a current sweep point is different from the calibrated sweep points. It is 
not possible to disable interpolation.   
Cav 
The system error correction uses variable calibration methods to calculate a measurement 
parameter. This happens for example, if a Z-parameter is calculated from S-parameters which 
are partly factory calibrated and partly normalized. This can also happen, if the analyzer uses a 
one port calibration at port 1 and a normalization at port 2. 
Ca? 
The system error correction is applied, however, the accuracy is questionable because of of 
the following applies:  
The attenuator settings during the calibration differ from the attenuator settings during 
the measurement. 
Modify pdf metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
adding metadata to pdf files; c# read pdf metadata
Modify pdf metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
clean pdf metadata; get pdf metadata
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
226
The bandwidth settings during the calibration differ from the bandwidth settings during 
the measurement. 
The point delay settings during the calibration differ from the point delay settings during 
the measurement.  
Cal 
Off ! 
The system error correction is no longer applied (e.g. turned off by the user). See also 
Calibration Overview. 
Repeat Prev Cal     
Reopens the Measure Standards dialog of the calibration wizard to repeat and optimize the previous 
calibration without necessarily repeating the measurement of all standards.  
Repeat Prev Cal is active only if Keep Measurement Data for >Repeat Cal< in the Measure Standards 
dialog was enabled when the last calibration for the active channel was successfully completed.  
Additional conditions for repeating a calibration 
Changing the channel settings can cause incompatibilities between the current instrument state and the 
previous calibration. In particular the following conditions must be met:   
The configuration of all ports needed must be correct (see Select Connector– Physical Port 
Number #).
The Sweep Type must be unchanged.  
If incompatible settings occur the analyzer displays an error message and does not repeat the calibration.  
Correction Off 
Activates or deactivates the system error correction in the active channel. Correction Off is active only if a 
valid system error correction is assigned to the active channel; see Cal State in the Calibration Manager. 
Ared label Cal Off ! appears behind the trace list if the system error correction is switched off; see 
also Calibration Overview.  
Remote control: [:SENSe<Ch>]:CORRection:STATe ON | OFF
Port Extensions     
The Port Extensions submenu defines a length offset and a loss for the test ports. The offset parameters 
complement the system error correction, compensating for the known length and loss of a (non-dispersive 
and perfectly matched) transmission line between the calibrated reference plane and the DUT. 
Length offset parameters: Definition 
The Delay is the propagation time of a wave traveling through the transmission line. The Electrical 
Length is equal to the Delay times the speed of light in the vacuum and is a measure for the length 
of transmission line between the standard and the actual calibration plane. For a line with 
permittivity N
r
and mechanical length L
mech
the delay and the electrical length are calculated as 
follows: 
How to C#: Modify Image Metadata (tag)
VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel, VB.NET PowerPoint, VB.NET Tiff, VB.NET Imaging, VB.NET OCR, VB How to C#: Modify Image Metadata (tag).
add metadata to pdf; change pdf metadata creation date
VB.NET PDF File & Page Process Library SDK for vb.net, ASP.NET
VB.NET PDF - How to Modify PDF Document Page in VB.NET. VB.NET Guide for Processing PDF Document Page and Sorting PDF Pages Order.
analyze pdf metadata; view pdf metadata in explorer
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
227
Electrical Length, Mechanical Length or Delay are coupled parameters. When one of them is 
changed, the other two follow.  
The velocity factor is 1/sqrt(]
r
)and is a measure for the velocity of light in a dielectric with 
permittivity ]
r
relative to the velocity of light in the vacuum (velocity factor < 1). Permittivity and 
velocity factor are coupled parameters. 
For a non-dispersive DUT, the delay defined above is constant over the considered frequency 
range and equal to the negative derivative of the phase response with respect to the frequency 
(see mathematical relations). The length offset parameters compensate for a constant delay, which 
is equivalent to a linear phase response.  
If a dispersive connector type (i.e. a waveguide; see Offset Model dialog) is assigned to a test port 
that is related to a particular quantity, then the phase of the quantity is calculated taking dispersion 
effects into account.  
Loss parameters: Definition 
The loss L is the attenuation of a wave traveling through the offset transmission line. In logarithmic 
representation, the loss can be modeled as the sum of a constant and a frequency-dependent part. 
The frequency dependence is essentially due to the skin effect; the total loss can be approximated 
by an expression of the following form:  
The DC loss Loss
DC
,the reference frequency f
ref
,and the loss at the reference frequency Loss(f
ref
)
are empirical parameters for the transmission lines connected to each port which can be entered 
into any of the dialogs in the Offset menu. For a lossless transmission line, both Loss
DC
and 
Loss(f
ref
)are zero. In practice, Loss(f
ref
)often represents the dominant contribution so that Loss
DC
can be set to zero.   
Offset parameters: Application and effect 
Offset parameters can be particularly useful if the reference plane of the calibration cannot be 
placed directly at the DUT ports, e.g. because the DUT has non-coaxial ports and can only be 
measured in a test fixture. Offset parameters can also help to avoid a new complete system error 
correction if a cable with known properties has to be included in the test setup.  
Apositive length offset moves the reference plane of the port towards the DUT, a negative offset 
moves the reference plane away from the DUT. The offset parameters cannot compensate for a 
possible mismatch in the test setup.  
Apositive length offset moves the reference plane of the port towards the DUT, which is 
equivalent to deembedding the DUT from a (perfectly matched) transmission line at that 
port.  
Anegative offset moves the reference plane away from the DUT, which is equivalent to 
embedding the DUT into a (perfectly matched) transmission line at that port.   
In contrast to the embedding/deembedding procedure, the offset parameters cannot compensate 
for a possible mismatch in the test setup.  
Each offset parameter is assigned to a particular port. The delay parameters affect the phase of all 
measured quantities related to this port; the loss parameters affect their magnitude. An offset at 
port 1 affects the S-parameters S
11
,S
21
,S
12
.Some quantities depend on the whole of all S-
parameters, so they are all more or less affected when one S-parameter changes due to the 
addition of an offset length.  
C# PDF File & Page Process Library SDK for C#.net, ASP.NET, MVC
Image: Copy, Paste, Cut Image in Page. Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. C# PDF - Modify PDF File and Page Using C#.
delete metadata from pdf; remove pdf metadata
C# Raster - Modify Image Palette in C#.NET
in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET C# Raster - Modify Image Palette in C#.NET.
rename pdf files from metadata; pdf metadata editor
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
228
To account for the propagation in both directions, the phase shift of a reflection parameter due 
to a given length offset is twice the phase shift of a transmission parameter. If, at a frequency of 
300 MHz, the electrical length is increased by 250 mm (d/4), then the phase of S
21
increases by 90 
deg, whereas the phase of S
11
increases by 180 deg. Equivalent relations hold for the loss.  
If the trace is displayed in Delay format, changing the offset parameters simply shifts the whole 
trace in vertical direction. The sign of the phase shift is determined as follows: 
Apositive offset parameter causes a positive phase shift of the measured parameter and 
therefore reduces the calculated group delay.   
Anegative offset parameter causes a negative phase shift of the measured parameter and 
therefore increases the calculated group delay.   
The offset parameters can be defined separately for each port.   
Reset Lengths restores the default values for all length offsets, i.e. it resets all values to zero. The 
(de-)embedding parameters are left unchanged. 
Electrical Length defines electrical length offsets and loss parameters at all test ports.  
Mechanical Length defines mechanical length offsets and loss parameters at all test ports.  
Delay defines delay times and loss parameters at all test ports.  
Auto Length determines a length offset for the receiving port of the active measured quantity with 
the condition that the residual group delay of the active trace is minimized.  
Auto Offset determines a length offset an loss for the receiving port of the active measured 
quantity with the condition that the residual group delay of the active trace is minimized and the 
measured loss is minimized as far as possible across the entire sweep range.  
De-Embedding and Embedding open configuration dialogs for (de-)embedding transformation 
networks.  
The Zero Delay at Marker function overwrites the Offset parameters.  
Remote control 
(for Reset 
Offsets): 
[SENSe<Ch>:]CORRection:OFFSet<port_no>[:STATe] <numeric 
_value> 
Electrical Length 
Opens a dialog to define the length offset parameters for the physical test ports as electrical lengths. The 
dialog also contains the DC loss, the loss at the reference frequency Loss(f
ref
); and the reference 
frequency f
ref
;see Loss parameters: Definition above.  
How to C#: Modify Image Bit Depth
text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET VB.NET How-to, VB.NET PDF, VB.NET Word, VB.NET Excel How to C#: Modify Image Bit Depth.
add metadata to pdf file; adding metadata to pdf
How to C#: Overview of Using XImage.Raster
See this C# guide to learn how to use RasterEdge XImage SDK for .NET to edit the image file Metadata, Color Palette and modify the compression method.
pdf xmp metadata viewer; pdf metadata extract
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
229
The dialog provides the following additional control elements:  
Select Cable Type opens a dialog to use the parameters of a standard (predefined) cable type or 
add a cable type with arbitrary parameters.  
If Same Offset Line Characteristics at All Ports is selected, the selected cable type is applied to all 
ports.   
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:LOSS<port_no> 
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet 
Mechanical Length 
Opens a dialog to define length offset parameters for the physical test ports as mechanical lengths and 
permittivities. The dialog also contains the DC loss, the loss at the reference frequency Loss(f
ref
); and the 
reference frequency f
ref
;see Loss parameters: Definition above.  
The Mechanical Length dialog contains the three editable columns Mech. Length, Permittivity (]
r
)and 
Velocity Fact 1/sqrt(]
r
); see Offset parameters: Definition above. Permittivity and velocity factor are 
coupled parameters. The dialog provides the following additional control elements:  
Select Cable Type opens a dialog to use the parameters of a standard (predefined) cable type or 
add a cable type with arbitrary parameters.  
If Same Offset Line Characteristics at All Ports is selected, the selected cable type is applied to all 
ports.   
Select Same Offset Line Characteristics at All Ports to change the permittivity or velocity factor for all 
ports by entering a single value. 
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:DISTance 
[SENSe<Ch>:]CORRection:EDELay<port_no>:DIELectric
[SENSe<Ch>:]CORRection:LOSS<port_no> 
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency 
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet 
How to C#: Modify Color and Contrast
How to C#: Modify Color and Contrast. Overview for How to Modify Color and Contrast. Overview. By two. Steps to Modify Color and Contrast.
remove pdf metadata online; google search pdf metadata
VB.NET PDF Library SDK to view, edit, convert, process PDF file
NET developers can redact, delete, view and save PDF metadata. added to a specific location on PDF file page addition, you can easily create, modify, and delete
add metadata to pdf programmatically; pdf metadata online
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
230
Delay 
Opens a dialog to define the length offset parameters for the physical test ports as delays. The dialog also 
contains the DC loss, the loss at the reference frequency Loss(f
ref
); and the reference frequency f
ref
;see 
Loss parameters: Definition above.  
The dialog provides the following additional control elements:  
Select Cable Type opens a dialog to use the parameters of a standard (predefined) cable type or 
add a cable type with arbitrary parameters.  
If Same Offset Line Characteristics at All Ports is selected, the selected cable type is applied to all 
ports.   
Remote control: [SENSe<Ch>:]CORRection:EDELay<port_no>[:TIME] 
[SENSe<Ch>:]CORRection:LOSS<port_no> 
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet 
Auto Length 
Adds an electrical length offset to the active test port with the condition that the residual delay of the active 
trace (defined as the negative derivative of the phase response) is minimized across the entire sweep 
range. If Delay is the selected trace format, the entire trace is shifted in vertical direction and centered 
around zero. In phase format, the Auto Length corrected trace shows the deviation from linear phase. The 
effect of a dispersive connector type (i.e. a waveguide; see Offset Model dialog) assigned to the receiving 
port of the measured quantitiy is taken into account.   
If the active trace shows an S-parameter S
ij
,then Auto Length adds a length offset at port i.   
Preconditions for Auto Length, effect on measured quantities and exceptions 
Auto Length is enabled if the measured quantity contains the necessary phase information as a 
function of frequency, and if the interpretation of the results is unambiguous:  
Afrequency sweep must be active. 
The measured quantity must be an S-parameter, a converted impedance or a converted 
admittance.  
The effect of Auto Length on S-parameters is to eliminate a linear phase response as described 
above. The magnitude of the measured quantity is not affected. Converted admittances or 
impedances are calculated from the corresponding Auto Length corrected S-parameters. Stability 
factors are not derived from a single S-parameter, therefore Auto Length is disabled.   
Use Zero Delay at Marker to set the delay at a special trace point to zero. 
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:AUTO ONCE
C# PDF Library SDK to view, edit, convert, process PDF file for C#
NET allows you to read, add, edit, update, and delete PDF file metadata, like Title In addition, you can easily create, modify, and delete PDF annotations.
pdf metadata editor online; embed metadata in pdf
How to C#: Modify Alpha Channel
Image Access and Modify. Image Information. Metadata(tag) Edit. |. Home ›› XImage.Raster ›› C# Raster: Modify Alpha Channel. PDF in C#, C# convert PDF to HTML
batch pdf metadata; pdf remove metadata
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
231
Auto Offset 
Determines all offset parameters such that the residual group delay of the active trace (defined as the 
negative derivative of the phase response) is minimized and the measured loss is minimized as far as 
possible across the entire sweep range. Auto Offset involves a two-step procedure:  
An Auto Length correction modifies the phase of the measured quantity, minimizing the residual 
group delay. The magnitude of the measured quantity is not affected.   
The auto loss correction modifies the magnitude of the measured quantity, leaving the (auto 
length-corrected) phase unchanged.   
Preconditions for Auto Offset, effect on measured quantities and exceptions 
Auto Offset is enabled if the measured quantity contains the necessary phase information as a 
function of the frequency, and if the interpretation of the results is unambiguous:  
Afrequency sweep must be active. 
The measured quantity must be an S-parameter, a converted impedance or a converted 
admittance.  
The effect of Auto Offset on S-parameters is to eliminate a linear phase response and account for a 
loss as described above. Converted admittances or impedances are calculated from the 
corresponding Auto Offset corrected S-parameters. Stability factors are not derived from a single S-
parameter, therefore Auto Offset is disabled.   
Calculation of loss parameters 
The loss is assumed to be given in terms of the DC loss Loss
DC
,the reference frequency f
ref
,and 
the loss at the reference frequency Loss(f
ref
). The formula used in the Auto Loss algorithm is similar 
to the formula for manual entry of the loss parameters (see Loss parameters: Definition). The result 
is calculated according to the following rules:  
The reference frequency f
ref
is kept at its previously defined value (default: 1 GHz).   
The DC loss c is zero except for S-parameters with maximum dB magnitude larger than –
0.01 dB.  
Auto Offset for S-parameters centers the corrected dB magnitude as close as possible 
around 0 dB.  
The resulting offset parameters are displayed in the Electrical Length, Mechanical Length, and 
Delay dialogs.   
Remote control:
[SENSe<Ch>:]CORRection:LOSS<port_no>:AUTO ONCE
Deembedding and Embedding     
The Embedding... and Deembedding... dialogs define virtual networks (transformation networks) to be 
added to/removed from the measurement circuit for a DUT.  
Embedding a DUT into a matching network 
To be integrated in application circuits, high-impedance components like Surface Acoustic Wave (SAW) 
filters are often combined with a matching network. To obtain the characteristics of a component with an 
added matching network, both must be integrated in the measurement circuit of the network analyzer. The 
figure below shows a 2-port DUT that is combined with a real matching circuit consisting of a serial 
capacitor and a shunt inductor.  
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
232
The idea of virtual embedding is to simulate the matching network and avoid using physical circuitry so 
that the analyzer ports can be directly connected to the input and output ports of the DUT. The matching 
circuit is taken into account numerically. The analyzer measures the DUT alone but provides the 
characteristics of the DUT, including the desired matching circuit. This method provides a number of 
advantages:  
The measurement uncertainty is not impaired by the tolerances of real test fixtures. 
There is no need to fabricate test fixtures with integrated matching circuits for each type of DUT.  
Calibration can be performed at the DUT's ports. If necessary (e.g. for compensating for the effect 
of a test fixture) it is possible to shift the calibration plane using length offset parameters.  
Deembedding a DUT 
Deembedding and embedding are inverse operations: A deembedding problem is given if an arbitrary real 
network connected to the DUT is to be virtually removed in order to obtain the characteristics of the DUT 
alone. Deembedding is typically used for DUTs which are not directly accessible because they are 
inseparably connected to other components, e.g. for MMICs in a package or connectors soldered to an 
adapter board.  
To be numerically removed, the real network must be described by a set of S-parameters or by an 
equivalent circuit of lumped elements. Deembedding the DUT effectively extends the calibration plane 
towards the DUT ports, enabling a realistic evaluation of the DUT without the distorting network. 
Deembedding can be combined with length offset parameters.  
The simplest case of one-port deembedding can be depicted as follows:  
The embedding/deembedding function in the Virtual Transform menu has the following characteristics:  
Two-port networks can be applied to any of the DUT ports 1 or 2.  
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
233
Transformation networks can be defined by a set of S-parameters stored in a Touchstone file or 
by an equivalent circuit with lumped elements.  
The same networks are available for embedding and deembedding.   
De-embedding opens the configuration dialog and activates or deactivates deembedding.   
Embedding opens the configuration dialog and activates or deactivates embedding.   
Embedding/Deembedding  
Selects a 2-port transformation network for (de-)embedding, defines its parameters, assigns it to a 
physical port and enables (de-)embedding. The two dialogs for deembedding and embedding are identical 
except for their inverse effect.  
The dialogs contain the following control elements:  
Port is the analyzer port for the added or removed circuit. The transformation networks are 
defined such that the analyzer test port is connected to the left of the circuit; the DUT port is on 
the right side.  
(De-)embed DUT enables or disables the (de-)embedding function.  
R&S ZVL 
GUI Reference 
Channel Menu 
Operating Manual 1303.6580.32-06 
234
Transformation Network contains all available 2-port networks (see below). Networks are either 
defined by lumped elements or by means of imported S-parameter data. The active network 
appears in inverse colors. The element parameters (C, R, L) for the selected network are 
displayed on the right side.  
Read Data From File... is enabled as long as the 2-Port Data network is active. This network is 
defined by its S-parameters stored in a two-port Touchstone file (*.s2p). No additional parameters 
are required.  
Set to Ideal Through is enabled as long as the 2-Port Data network is active. An imported S-
parameter set is replaced by the S-parameters of an ideal through connection, which eliminates 
the transformation network.   
Circuit models for 2-port networks 
The lumped element 2-port transformation networks for (de-)embedding consist of the following two basic 
circuit blocks:  
Acapacitor C connected in parallel with a resistor.  
An inductor L connected in series with a resistor.  
The 2-port transformation networks comprise all possible combinations of 2 basic blocks, where one block 
represents a serial, the other a shunt element. In the default setting the resistors are not effective, since 
the serial Rs are set to 0 V, the shunt Rs are set to 10 MV.
The first network is defined by its S-parameters stored in an imported two-port Touchstone file 
(*.s2p). No additional parameters are required.  
The following networks are composed of a serial C or L (as seen from the test port), followed by a shunt C 
or L. They are named Serial C, Shunt C / Serial C, Shunt L / Serial L, Shunt C / Serial L, Shunt L. 
The following networks are composed of a shunt C or L (as seen from the analyzer port), followed 
by a serial C or L. They are named Shunt C, Serial C / Shunt C, Serial L / Shunt L, Serial C / 
Shunt L, Serial L. 