R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
295
Remote control: 
SYSTem:LXI:MDEScription  
LAN Reset 
Initiates the network configuration reset mechanism (Lan Configure Initialize, LCI) for the instrument. 
Default state of the network settings 
According to the LXI standard, an LCI must place the following parameters to a default state.  
Parameter 
Value 
TCP/IP Mode 
DHCP + Auto IP Address 
Dynamic DNS 
Enabled 
ICMP Ping 
Enabled 
Password for LAN configuration 
LxiWebIfc 
The LCI for the network analyzer also resets the following parameters:  
Parameter 
Value 
Description 
Rohde & Schwarz Vector Network Analyzer / ZVL / <serial no.> 
Negotiation 
Auto Detect 
VXI-11 Discovery 
Enabled 
The LAN settings are configured using the  instrument's LXI Browser Interface. With firmware versions 
V3.00 and higher, a  user account with administrator  rights  is  necessary for the LXI  functionality (see 
Operation with and without Administrator Rights in chapter 1). 
Remote control: 
SYSTem:LXI:LANReset  
LXI Browser Interface 
LAN eXtensions for Instrumentation (LXI) is an instrumentation platform for measuring instruments and 
test  systems  that  is  based  on  standard  Ethernet  technology.  LXI  is  intended  to  be  the  LAN-based 
successor to GPIB, combining the advantages of Ethernet with the simplicity and familiarity of GPIB.  
LXI classes and LXI functionality 
LXI-compliant instruments are divided into three classes, A, B and C, with the functionality of the classes 
hierarchically based one upon the other:  
Remove metadata from pdf file - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
batch pdf metadata; acrobat pdf additional metadata
Remove metadata from pdf file - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
remove pdf metadata; online pdf metadata viewer
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
296
Class C instruments  are characterized by a common LAN implementation, including an ICMP 
ping responder for diagnostics. The  instruments  can be configured via a  web browser; a LAN 
Configuration  Initialize  (LCI)  mechanism  resets  the  LAN  configuration.  The  LXI  class  C 
instruments shall also support automatic detection in a LAN via the VXI-11 discovery protocol and 
programming by means of IVI drivers.  
Class B adds IEEE 1588 Precision Time Protocol (PTP) and peer-to-peer communication to the 
base class. IEEE 1588 allows all instruments on the same network to automatically synchronize to 
the most accurate clock available and then provide time stamps or time-based synchronization 
signals to all instruments with exceptional accuracy.  
Class A instruments are additionally equipped with the eight-channel hardware trigger bus (LVDS 
interface) defined in the LXI standard.  
Instruments  of  classes  A and B  can generate  and  receive  software  triggers  via  LAN  messages and 
communicate with each other without involving the controller.   
The network analyzer complies with LXI class C. In addition to the general class C features described 
above, it provides the following LXI-related functionality:  
Integrated LXI  menu for LXI activation  and reset of the LAN configuration  (LAN Configuration 
Initialize, LCI) 
The LXI functionality requires a Windows XP operating system that has been upgraded to service 
pack 2. Please contact your Rohde & Schwarz service representative if your network analyzer's 
operating system needs an upgrade. To check the version of your operating system, open the 
System – General dialog in the Control Panel.  
To upgrade an R&S ZVL which already supports LXI to a newer firmware version, the old LXI 
version must be de-installed first. This is done automatically by the setup file. Do not prevent de-
installation to avoid problems.    
After a firmware update, use the standby key to shut-down and re-start the instrument in order to 
enable the full LXI functionality.  
The LXI functionality including the home page is unavailable while the instrument is operated 
without administrator rights.  
For information about the LXI standard refer to the LXI website at http://www.lxistandard.org. See 
also "News from Rohde & Schwarz, article 2006/II - 190".  
LXI Browser Interface  
The  instrument's  LXI browser  interface works  correctly  with  all  W3C  compliant  browsers.  Typing  the 
instrument's host name or IP address in the address field of the browser on your PC, e.g.  
http://10.113.10.203  
C# PDF Password Library: add, remove, edit PDF file password in C#
C# Sample Code: Remove Password from PDF File in C#.NET. These C# demos will help you to delete password for an encrypted PDF file. // Define input file path.
add metadata to pdf file; view pdf metadata
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Remove bookmarks, annotations, watermark, page labels and article threads from PDF Document and metadata. NET Demo Code to Optimize An Exist PDF File in Visual
read pdf metadata online; pdf metadata editor online
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
297
opens the Instrument Home Page (welcome page).  
The instrument home page displays the device information required by the LXI standard including the 
VISA resource string in read-only format. The Device Indicator toggle button causes the LXI symbol in the 
status bar of the analyzer to blink (if active). A green LXI status symbol indicates that a LAN connection 
has been established; a red symbol indicates that no LAN cable is connected. The Device Indicator setting 
is not password-protected. 
The navigation pane of the browser interface contains the following control elements:  
Lan Configuration opens the LAN Configuration page.  
Status displays information about the LXI status of the instrument.  
Glossary opens a document with a glossary of terms related to the LXI standard.  
www.rohde-schwarz.com opens the Rohde & Schwarz homepage.  
LAN Configuration 
The Lan Configuration web page displays all mandatory LAN parameters and allows their modification.  
VB.NET PDF Page Delete Library: remove PDF pages in vb.net, ASP.
Ability to remove consecutive pages from PDF file in VB.NET. Enable specified pages deleting from PDF in Visual Basic .NET class.
pdf metadata editor; edit pdf metadata
VB.NET PDF remove image library: remove, delete images from PDF in
Image: Insert Image to PDF. Image: Remove Image from PDF Page. Image Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. Form Process.
read pdf metadata java; change pdf metadata
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
298
The TCP/IP Mode configuration field controls how the IP address for the instrument gets assigned (see 
also Assigning an IP Address). For the manual configuration mode, the static IP address, subnet mask, 
and default gateway are used to configure the LAN. The automatic configuration mode uses DHCP server 
or Dynamic Link Local Addressing (Automatic IP) to obtain the instrument IP address.  
Changing the LAN configuration is password-protected. The default password reads LxiWebIfc 
(notice upper and lower case characters). This password can be changed in the LXI configuration 
menu. This password is restored by a Lan Reset (LCI).  
The subentries of the Lan Configuration menu open other pages:  
Ping Client provides the ping utility to verify the connection  between the  instrument and other 
devices.  
Advanced LAN Configuration provides LAN settings that are not declared mandatory by the LXI 
standard.  
C# PDF Page Delete Library: remove PDF pages in C#.net, ASP.NET
Ability to remove a range of pages from PDF file. Free trial package for quick integration in .NET as well as compatible with 32 bit and 64 bit windows system.
pdf keywords metadata; pdf metadata
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Remove bookmarks, annotations, watermark, page labels and article threads from PDF while compressing. Document and metadata. C#.NET DLLs: Compress PDF Document.
endnote pdf metadata; pdf metadata online
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
299
The Advanced LAN Configuration parameters are used as follows:  
The Negotiation configuration field provides different Ethernet speed and duplex mode settings. In 
general, the Auto Detect mode is sufficient.  
ICMP Ping must be enabled to use the ping utility.  
VXI-11 is the protocol that is used for discovery of the instrument in the LAN. According to the 
standard,  LXI  devices  must  use  VXI-11  to  provide  a  discovery  mechanism;  other  additional 
discovery mechanisms are permitted.  
Ping  
Ping is a utility that verifies the connection between the analyzer and another device. The ping command 
uses  the  ICMP  echo  request  and  echo  reply  packets  to  determine  whether  the  LAN  connection  is 
functional. Ping is useful for diagnosing IP network or router failures.  
The ping utility is not password-protected. To initiate a ping between the analyzer and a second connected 
device,  
Enable ICMP Ping on the Advanced LAN Configuration page (enabled after an LCI).  
Enter the IP address of the second device without the ping command and without any further 
parameters into the Destination Address field (e.g. 10.113.10.203).  
Click Submit. 
VB.NET PDF File Permission Library: add, remove, update PDF file
Image: Insert Image to PDF. Image: Remove Image from PDF Page. Image Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. Form Process.
pdf keywords metadata; acrobat pdf additional metadata
C# PDF bookmark Library: add, remove, update PDF bookmarks in C#.
Help to add or insert bookmark and outline into PDF file in .NET framework. Ability to remove and delete bookmark and outline from PDF document.
pdf xmp metadata editor; view pdf metadata
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
300
GPIB (option GPIB Interface, with option R&S FSL-B10 only) 
Opens a submenu to set the parameters of the remote control interface. 
Menu (Key)
Submenu 
Submenu
Command 
SETUP g
General Setup g
GPIB g
GPIB Address 
ID String Factory 
ID String User 
GPIB Language 
Display Update On/Off 
More g
GPIB Terminator LFEOI/EOI 
*IDN Format Leg./ZVL 
I/O Logging On/Off 
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
301
GPIB Address 
Opens an edit dialog box to enter the GPIB address. Values from 0 to 30 are allowed. The default address 
is 20.  
If, e.g. after a firmware update, the R&S ZVL does not maintain the GPIB address after reboot, the 
shutdown file needs to be recreated. Perform the following steps: 
Set the GPIB address. 
Create a shutdown file by switching the R&S FSL in the standby mode:  
Press the ON/STANDBY key on the front panel and wait until the yellow LED is on. With the battery 
pack option, use a USB keyboard and terminate the R&S ZV firmware with ALT+F4 to create the 
shutdown file. 
If the shutdown file was created once, the R&S ZVL boots with exactly those settings after a reboot with 
the main switch on the rear panel. During a firmware update or cold boot for service reasons the shutdown 
file is deleted, and the setup is necessary once again. 
Remote Control: 
SYST:COMM:GPIB:ADDR 20 
ID String Factory 
Selects the default response to the *IDN? query. 
ID String User 
Opens an edit dialog box to enter a user–defined response to the *IDN? query. Max. 36 characters are 
allowed. 
GPIB Language 
Only the remote control language SCPI is available for the R&S ZVL and is set by default.  
Display Update On/Off 
Defines whether the instrument display is switched off when changing from manual operation to remote 
control. In remote control mode, this softkey is displayed in the local menu. 
Remote Control: 
SYST:DISP:UPD 
GPIB Terminator LFEOI/EOI 
Changes the GPIB receive terminator.  
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary data transfers (e.g. 
trace data) from the control computer to the instrument, the binary code used for <LF> might be included 
in the binary data block, and therefore should not be interpreted as a terminator in this particular case. 
This can be avoided by changing the receive terminator to EOI. 
Remote Control: 
SYST:COMM:GPIB:RTER 
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
302
*IDN Format Leg./ZVL 
If set to Leg., provides the response to the *IDN? remote command in a format that is compatible to the 
R&S FSP/FSU/FSQ family (e.g. "Rohde&Schwarz,ZVL-6,100007/006,1.10"). This function is intended for 
re–use of existing control programs together with the R&S ZVL. The ZVL-compliant ID string contains the 
order number of the instrument; e.g. "Rohde&Schwarz,ZVL-6,1303.6509K06/100007,1.10".  
The *IDN format is not changed upon a reset/preset of the analyzer settings.  
I/O Logging On/Off 
Activates logging all remote control commands received by the ZVL in the log file ScpiLog.txt for debug 
purposes. Logging the commands may be useful in order to find misspelled keywords in control programs. 
The corresponding log file will be stored in folder C:\R_S\instr\ScpiLogging. 
System Info 
Opens a submenu to display detailed information on module data, device statistics and system messages. 
Menu (Key)
Submenu 
Command 
SETUP g
System Info g
Hardware Info 
Versions+Options 
System Messages 
Clear All Messages
Hardware Info 
Opens a dialog box in which all modules (installed hardware components) are listed together with the 
serial numbers, order numbers, model information, hardware codes, and hardware revisions.  
Versions+Options 
Opens a dialog box that displays hardware and firmware information, e.g. the firmware version, the image 
version, BIOS version, data sheet version of the basic device, installed options (hardware and firmware 
options).   
Remote Control: 
*IDN? (instrument identification) 
*OPT? (installed options) 
System Messages 
Opens the System Messages dialog box that displays the generated system messages in the order of 
their occurrence. The most recent messages are placed at the top of the list. Messages that have 
occurred since the last display of system messages menu are marked with an asterisk '*'. The following 
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
303
information is available: 
No 
device–specific error code 
Message 
brief description of the message 
Component 
hardware messages: name of the affected module 
software messages: name of the affected software 
Date/Time 
date and time of the occurrence of the message 
If the number of error messages exceeds the capacity of the error buffer, Message buffer overflow is 
displayed. To delete messages see Clear All Messages softkey. 
Remote Control: 
SYST:ERR:LIST? 
Clear All Messages 
Deletes all system messages. The softkey is only available if the System Messages dialog box is 
displayed. 
Remote Control: 
SYST:ERR:CLE:ALL? 
Service 
Opens a submenu that contains additional functions for maintenance and/or trouble shooting.  
Menu (Key)
Submenu 
Command 
SETUP g
Service g
Input RF/Cal/TG 
Comb Frequency 
Reset Password 
Selftest 
Selftest Results 
Password 
Service Function 
R&S ZVL 
GUI Reference 
SETUP Key 
Operating Manual 1303.6580.32-06 
304
Attention! 
The service functions are not necessary for normal measurement operation. However, 
incorrect use can affect correct operation and/or data integrity of the R&S ZVL.  
Therefore, many of the functions can only be used after entering a password. They are 
described in the instrument service manual. 
Input RF/Cal/TG 
Selects the source signal for the RF input path in spectrum analysis mode (with option R&S ZVL-K1). 
RF switches the RF path to the input connector for spectrum analyzer mode (PORT 2). The R&S 
ZVL measures the RF input signal fed in via PORT 2.  
Cal switches the RF path to the internal calibration source (65.8333 MHz; see Comb Frequency)
and activates the data entry of the output level of the calibration source. Possible values are 0 dB 
and –30 dB.  
TG is for future extensions.  
The Input RF/Cal/TG softkey is available in spectrum analyzer mode only.  
Comb Frequency 
Opens a dialog box to set the comb generator frequency for the internal calibration (see Input Cal):
Comb/1: 65.8333 MHz / 1  
Comb/64: 65.8333 MHz / 64  
Comb/65: 65.8333 MHz / 65  
The Comb Frequency is available in spectrum analyzer mode only, provided that Input Cal is active.  
Reset Password 
Deactivates all set passwords. 
Remote Control: 
SYST:PASS:RES 
Selftest 
Initiates the self test of the instrument modules to identify a defective module in case of failure. All 
modules are checked consecutively and the test result is displayed.  
Remote Control: 
*TST? 
Selftest Results 
Opens the Selftest Result dialog box that contains the test results. In case of failure a short description of 
the failed test, the defective module, the associated value range and the corresponding test results are 