R&S ZVL 
Remote Control 
Remote Control Operation 
Operating Manual 1303.6580.32-05 
315
After the GPIB Explorer is started, the interface for the connection to the instrument can be selected in a 
dialog:  
The following options are provided:  
NT named pipe (not supported at present) 
GPIB address (for connection to controllers equipped with a National Instrument GPIB interface 
using the GPIB bus connector) 
RSIB address (for LAN connection via RSIB protocol, requires an appropriate IP or local host 
address; see LAN Connection sections in Chapter 1)   
VXI-11/VISA (for LAN connections via the VXI-11 protocol; requires an installed VISA library)  
NT pipe A/B (only for local connection on the analyzer, recommended for remote test on the 
instrument)  
EB200 (not supported at present)  
Aconnection requires the default settings "Rohde & Schwarz" (for the SCPI command set) 
and "Use SRQ: On". Select Nwa-Setup – Setup Info to look up the IP address information of 
your analyzer.  
After the connection is established, the GPIB explorer displays a tree view of all commands included in the 
current firmware version of the network analyzer. The programs can be selected for execution by a single 
mouse click.   
Analyze pdf metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
read pdf metadata; remove metadata from pdf
Analyze pdf metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
search pdf metadata; batch pdf metadata editor
R&S ZVL 
Remote Control 
Remote Control Operation 
Operating Manual 1303.6580.32-05 
316
Refer to the GPIB explorer's help system for more information.  
Switchover to Remote Control 
On power-up, the instrument is always in the manual operating state and can be operated via the front 
panel controls. The instrument is switched to remote control as soon as it receives a command from the 
controller. If the instrument is controlled via RSIB or VX 11 protocol, the alternative commands @REM 
and @LOC can be used to switch from manual to remote control and back.  
VB.NET TIFF: TIFF Tag Viewer SDK, Read & Edit TIFF Tag Using VB.
Therefore, if you want to analyze or diagnose one TIFF file format, you must need a of TIFF file which is used as a tool to save metadata information), we
rename pdf files from metadata; bulk edit pdf metadata
R&S ZVL 
Remote Control 
Remote Control Operation 
Operating Manual 1303.6580.32-05 
317
While remote control is active, operation via the front panel is disabled with the exception of two softkeys. 
The instrument settings are optimized for maximum measurement speed; the display is switched off:  
The two softkeys in the remote screen are used to modify or quit the remote state:  
Display Update On/Off switches the display on or off; see background information below.  
Local switches the instrument back to local state.  
Display
 On/Off
states 
Switching on the display means that the analyzer shows the measurement screen with the current setups, 
diagram areas and traces without leaving the remote state. In this operating mode, it is possible to 
observe the screen while a remote control script is executed and the control elements on the front panel 
are still disabled:  
Switching on the display is ideal for program test purposes but tends to slow down the measurement. 
Therefore it is recommended to switch off the display in real measurement applications where a tested 
program script is to be executed repeatedly.  
R&S ZVL 
Remote Control 
Remote Control Operation 
Operating Manual 1303.6580.32-05 
318
The analyzer provides a third display option where the measurement screen is only updated 
when this is triggered by a remote control command; see SYSTem:DISPlay:UPDate ONCE.
The instrument remains in the remote state until it is reset to the manual state via the GUI or via GPIB bus 
(see section Return to Manual Operation).
You can create your own keys to replace the two default softkeys Go to Local and Display 
On/Off. See Combining Manual and Remote Control.  
Atooltip across the bottom of the remote screen indicates a remote command error, e.g. 
.You can switch off this tooltip using 
SYSTem:ERRor:DISPlay OFF.
Remote control: 
@REM 
SYSTem:DISPlay:UPDate 
SYSTem:USER:DISPlay:TITLe (define a title for the remote 
display) 
SYSTem:ERRor:DISPlay (switch tooltip on or off) 
Setting the Device Address 
The GPIB address (primary address) of the instrument is factory-set to 20. It can be changed manually in 
the System menu or via GPIB bus. For remote control, addresses 0 through 30 are permissible. The GPIB 
address is maintained after a reset of the instrument settings.  
Remote control: 
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 
Return to Manual Operation 
Return to manual operation can be initiated via the front panel or via remote control.  
Manually: Click the Local softkey in the remote screen.  
Via GPIB bus: CALL IBLOC(device%)  
Via RSIB or VX 11 protocol: @LOC and @REM can be used to switch from remote to manual 
control and back.  
Local lockout 
Before returning to manual control, command processing must be completed. If this is not the case, the 
analyzer switches back to remote control immediately. 
Returning to manual control by pressing the Local softkey can be disabled by the GPIB Local Lockout 
Message (LLO; see GPIB Bus Interface, Universal Commands) which is also included in the NI 
commands SetRWLS (Set Remote With Lockout State) or SendLLO. This prevents unintentional switch-
over, i.e. return to manual control is possible via the GPIB bus only.  
Returning to manual control via the front panel keys can be enabled again by deactivating the REN control 
line of the GPIB bus.  
R&S ZVL 
Remote Control 
Remote Control Operation 
Operating Manual 1303.6580.32-05 
319
Combining Manual and Remote Control 
Using a remote control script is the quickest and easiest way of performing complicated tasks which need 
to be repeated many times. On the other hand, it is often preferable to control a previously configured 
measurement manually in order to observe the result on the screen.   
The analyzer provides a number of tools for combining manual and remote control:  
User Keys 
The remote control commands SYSTem:USER:KEY... place up to 8 softkeys with arbitrary 
functionality on the remote screen. The softkeys replace the default softkeys Go to Local and 
Display On/Off. 
Pressing a user softkey or clicking the corresponding command across the top of the remote 
screen executes the assigned functionality.   
If you want to use more than 8 user keys, you can easily introduce additional levels: Simply 
reserve one key for re-defining the entire user key bar. If you press or click the key, your 
remote script should re-assign the labels and functions of the existing keys. If you use this 
procedure repeatedly, you can emulate an arbitrary number of manual control features on 
your remote screen.  
Menu key commands 
The commands in the DISPlay:MENU:KEY... subsystem open the manual control screen, 
activate menus or submenus for manual control, or execute a specified menu command.  
DISPlay:MENU:KEY:EXECute '<menu_key>' can be used in different ways:  
If the menu key causes an event (e.g. selection of the measured quantity: S11, S12 etc.), 
the function is executed immediately. No further action is required; the instrument remains 
in remote control mode.  
If the menu key requires a numeric entry (e.g. entry of the sweep range: Start, Stop etc.) 
the command opens the numeric entry bar. The instrument is switched back to remote 
control mode as soon as the entry has been made manually.  
R&S ZVL 
Remote Control 
Messages 
Operating Manual 1303.6580.32-05 
320
If the menu key requires several entries to be made in a dialog or in a wizard (e.g. the S-
Param Wizard), the command opens this dialog/wizard. All entries can be made manually. 
The instrument is switched back to remote control mode as soon as the dialog or wizard is 
closed.   
DISPlay:MENU:KEY:EXECute '<menu_key>' works for all menu keys, even though the 
analyzer may not provide an equivalent remote control command.  
Remote control: 
SYSTem:USER:KEY 
SYSTem:USER:KEY:FUNCtion 
DISPlay:MENU:KEY:EXECute 
DISPlay:MENU:KEY:SELect 
Messages 
The messages transferred on the data lines of the GPIB bus or via the RSIB / VXI-11 interface can be 
either interface messages or device messages. 
GPIB Interface Messages 
GPIB interface messages are transferred on the data lines of the GPIB bus, the ATN control line being 
active. They are used for communication between controller and instrument and can only be sent by a 
computer which has the function of an GPIB bus controller.  
GPIB interface messages can be further subdivided into 
Universal commands act on all devices connected to the GPIB bus without previous addressing.  
Addressed commands only act on devices previously addressed as listeners.  
The interface messages relevant to the instrument are listed in the GPIB bus section. 
RSIB and VXI-11 Interface Messages 
The RSIB and VXI-11 interfaces allow the instrument to be controlled in a Local Area Network. For a short 
introduction and a list of interface functions refer to the following sections:   
RSIB Interface  
VXI-11 Interface  
Device Messages (Commands and Device Responses) 
Device messages are transferred via the data lines of the GPIB bus, the "ATN" control line not being 
active. The ASCII character set is used. A distinction is made according to the direction in which device 
messages are transferred: 
Commands are messages the controller sends to the instrument. They operate the device 
functions and request information.  
R&S ZVL 
Remote Control 
Messages 
Operating Manual 1303.6580.32-05 
321
Device responses are messages the instrument sends to the controller after a query. They can 
contain measurement results, instrument settings and information on the instrument status.  
Commands are subdivided according to two criteria:  
1. According to the effect they have on the instrument:  
Setting commands cause instrument settings such as a reset of the instrument or setting the 
output level to some value.  
Queries cause data to be provided for output on the GPIB bus, e.g. for identification of the device 
or polling the active input.  
2. According to their definition in standard IEEE 488.2: 
Common commands have a function and syntax that is exactly defined in standard IEEE 488.2. 
Typical tasks are the management of the standardized status registers, reset and selftest.  
Instrument-control commands are functions that depend on the features of the instrument such as 
frequency settings. A majority of these commands has also been standardized by the SCPI 
consortium.  
The device messages have a characteristic structure and syntax. In the SCPI reference chapter all 
commands are listed and explained in detail. 
SCPI Command Structure and Syntax     
SCPI commands consist of a so-called header and, in most cases, one or more parameters. The header 
and the parameters are separated by a "white space" (ASCII code 0 to 9, 11 to 32 decimal, e.g. blank). 
The headers may consist of several key words. Queries are formed by directly appending a question mark 
to the header. 
Common commands and device-specific commands differ in their syntax. 
Common Commands 
Common (=device-independent) commands consist of a header preceded by an asterisk "*" and possibly 
one or more parameters. 
Examples: 
*RST 
RESET, resets the instrument. 
*ESE 253 
EVENT STATUS ENABLE, sets the bits of the event status enable registers. 
*ESR? 
EVENT STATUS QUERY, queries the contents of the event status register. 
Instrument-Control Commands 
Instrument-control commands are based on a hierarchical structure and can be represented in a 
command tree. The command headers are built with one or several mnemonics (keywords). The first level 
(root level) mnemonic identifies a complete command system. 
R&S ZVL 
Remote Control 
Messages 
Operating Manual 1303.6580.32-05 
322
Example: 
SENSe      This mnemonic identifies the command system SENSe.
For commands of lower levels, the complete path has to be specified, starting on the left 
with the highest level, the individual key words being separated by a colon ":". 
Example: 
SENSe:FREQuency:STARt 1GHZ 
This command is located on the third level of the SENSe system. It defines the start 
frequency of the sweep. 
The following rules simplify and abbreviate the command syntax: 
Multiple mnemonics 
Some mnemonics occur on several levels within one command system. Their effect depends on 
the structure of the command, i. e. on the position in the command header they are inserted in. 
Example: 
SOURce:FREQuency:CW lGHZ 
This command contains the key word SOURce in the first command level. It defines the frequency 
for sweep types operating at fixed frequency. 
TRIGger:SOURce EXTernal 
This command contains the key word SOURce in the second command level. It defines the trigger 
source “external trigger”. 
Optional mnemonics 
Some command systems permit certain mnemonics to be optionally inserted into the header or 
omitted. These mnemonics are marked by square brackets in this manual. The full command 
length must be recognized by the instrument for reasons of compatibility with the SCPI standard. 
Some commands are considerably shortened by omitting optional mnemonics. 
Example: 
TRIGger[:SEQuence]:SOURce EXTernal 
This command defines the trigger source “external trigger”. The following command has the same 
effect: 
TRIGger:SOURce EXTernal 
R&S ZVL 
Remote Control 
Messages 
Operating Manual 1303.6580.32-05 
323
An optional mnemonic must not be omitted if its effect is additionally specified by a numeric 
suffix. 
Long and short form 
The key words feature a long form and a short form. Either the short form or the long form can be 
entered; other abbreviations are not permitted. 
Example:  
TRIGger:SOURce EXTernal 
TRIG:SOUR EXT 
The short form is marked by upper case letters, the long form corresponds to the complete 
word. Upper case and lower case notation only serves to distinguish the two forms in the 
manual, the instrument itself is case-insensitive. 
Parameters 
Parameters must be separated from the header by a "white space". If several parameters are 
specified in a command, they are separated by a comma ",". For a description of the parameter 
types, refer to section Parameters. 
Example: 
SOURce:GROup 1,1 
This command defines a group of measured ports. 
Numeric suffix 
If a device features several functions or features of the same kind, e.g. several channels or test 
ports, the desired function can be selected by a suffix added to the command. Entries without 
suffix are interpreted like entries with the suffix 1. 
Example: 
SOURce:GROup2 1,1 
This command defines a second group (group no 2) of measured ports. 
Structure of a Command Line 
Acommand line may consist of one or several commands. It is terminated by a <New Line>, a <New 
Line> with EOI or an EOI together with the last data byte. Visual BASIC automatically produces an EOI 
together with the last data byte. 
Several commands in a command line must be separated by a semicolon ";". If the next command 
belongs to a different command system, the semicolon is followed by a colon. 
Example:
CALL IBWRT(device%, "TRIGger:SOURce EXTernal;:SENSe:FREQuency:STARt 
1GHZ") 
This command line contains two commands. The first command belongs to the TRIGger system and 
defines the trigger source (external trigger). The second command belongs to the SENSe system and 
defines the start frequency of the sweep. 
If the successive commands belong to the same system, having one or several levels in common, the 
command line can be abbreviated. To this end, the second command after the semicolon starts with the 
level that lies below the common levels. The colon following the semicolon must be omitted in this case. 
Example:
CALL IBWRT(device%, "TRIG:
SOUR EXT;:TRIG:
TIM 0.1")
R&S ZVL 
Remote Control 
Messages 
Operating Manual 1303.6580.32-05 
324
This command line is represented in its full length and contains two commands separated from each other 
by the semicolon. Both commands are part of the TRIGger command system, i.e. they have one level in 
common. 
When abbreviating the command line, the second command begins with the level below TRIG. The colon 
after the semicolon is omitted. The abbreviated form of the command line reads as follows: 
CALL IBWRT(device%, "TRIG:
SOUR EXT; TIM 0.1") 
However, a new command line always begins with the complete path. 
Example: 
CALL IBWRT(device%, "TRIG:
SOUR EXT ")  
CALL IBWRT(device%, "TRIG:
THR LOW ") 
Responses to Queries 
Aquery is defined for each setting command unless explicitly specified otherwise. It is formed by adding a 
question mark to the associated setting command. According to SCPI, the responses to queries are partly 
subject to stricter rules than in standard IEEE 488.2. 
1. The requested parameter is transmitted without header. 
Example: TRIGger:SOURce? Response: IMM 
2. Maximum values, minimum values and all further quantities which are requested via a special text 
parameter are returned as numerical values.  
Example: SENSe:FREQuency:STOP? MAX Response: 8000000000 
3. Numerical values are output without their unit. The default unit for each command is reported in the 
SCPI command description.  
Example: SENSe:FREQuency:STOP? MAX Response: 8000000000 for 8 GHz 
4. Boolean values are returned as 0 (for OFF) and 1 (for ON). 
Example: SWEep:TIME:AUTO? Response: 1
5. Text (character data) is returned in short form (see also next section).     
Example: TRIGger:SOURce? Response: IMM 
SCPI Parameters 
Most commands require a parameter to be specified. The parameters must be separated from the header 
by a "white space". Permissible parameters are numerical values, Boolean parameters, text, character 
strings and block data. The type of parameter required for the respective command and the permissible 
range of values are specified in the command description. 
Numeric Values 
Numeric values can be entered in any form, i.e. with sign, decimal point and exponent. Values exceeding 
the resolution of the instrument are rounded up or down. The mantissa may comprise up to 255 
characters, the values must be in the value range –9.9E37 to 9.9E37. The exponent is introduced by an 
"E" or "e". Entry of the exponent alone is not allowed. In the case of physical quantities, the unit can be 
entered. Permissible unit prefixes are G (giga), MA (mega), MOHM and MHZ are also permissible), K 
(kilo), M (milli), U (micro) and N (nano). If the unit is missing, the default unit is used. 