R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
375
Query the results of the bandfilter search. An error message is generated if 
the bandfilter search fails so that no valid results are available.  
CALCulate<Chn>:MARKer<Mk>:COUPled[:STATe] <Boolean> 
Couples the markers of all traces in the active setup to the markers of trace no. <Chn>, provided that they 
have the same sweep type (SENSe<Chn>:FUNCtion).  
<Chn> 
Channel number used to identify the active trace. The effects of marker coupling 
depend on the active trace number; see Coupled Markers in the GUI Reference 
chapter.  
<Mk> 
Marker number in the range 1 to 10. This suffix is ignored because the command 
affects all markers.   
<Boolean> 
ON | OFF - Enables or disables marker coupling.   
*RST value 
OFF   
SCPI, 
Command 
Types 
Device-specific, command or query.   
Example: 
Suppose that the active setup contains two traces Trc1 and Trc2, assigned to 
channels no. 1 and 2, respectively.  
:CALC1:PAR:SEL 'TRC1'; CALC1:MARK1 ON; MARK2 ON  
Select Trc1 as the active trace and create the two markers no. 1 and 2. The 
default position for both markers is the center of the sweep range.  
CALC1:MARK:COUP ON  
Create two markers no. 1 and 2 on Trc 2 and couple them to the markers of Trc 1. 
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe] <Boolean> 
Switches the delta mode for marker <Mk> on trace no. <Chn> on or off. The marker must be created 
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON. If the active trace contains no reference 
marker, the command also creates a reference marker.   
<Chn> 
Channel number used to identify the active trace.   
<Mk> 
Marker number in the range 1 to 10.   
<Boolean> 
ON | OFF - Enables or disables the delta mode.  
*RST value 
OFF   
SCPI, Command 
Types 
Device-specific, command or query.   
Example: 
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK ON  
Create marker no. 1 and set it to the center of the sweep range.  
CALC:MARK:DELT ON  
Add metadata to pdf file - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
pdf xmp metadata editor; extract pdf metadata
Add metadata to pdf file - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
add metadata to pdf; add metadata to pdf programmatically
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
376
Create a reference marker at the center of the sweep range and set marker 1 
to delta mode.   
CALCulate<Chn>:MARKer<Mk>:FORMat DEFault | MLINear | MLOGarithmic | 
PHASe | POLar | COMPlex | GDELay | REAL | IMAGinary | SWR | LINPhase | 
LOGPhase | IMPedance | ADMittance | MDB | MLPHase | MDPHase 
Defines the output format for the (complex) value of marker <Mk> on trace no. <Chn>. 
Note: The formats of the markers assigned to a trace are independent of each other and of the trace 
format settings; see CALCulate<Chn>:FORMat.
<Chn> 
Channel number used to identify the active trace.  
<Mk> 
Marker number in the range 1 to 10.   
Parameters 
See list of parameters below. 
*RST value 
DEFault  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK ON; CALC:MARK:FORM?  
Create marker 1, assign it to the trace no. 1 and query its format. The analyzer 
returns the format of the active trace.  
Assume that the marker result is given by the complex quantity z = x + jy. The meaning of the parameters 
is as follows:  
DEFault 
The format of the trace no. <Chn> (MLOG after *RST); see CALCulate<Chn>:FORMat.
MLINear  
|z| = sqrt ( x2 + y2 )   
MLOGarithmic
MDB 
|z| = sqrt ( x2 + y2 )  
dB Mag(z) = 20 * log|z| dB  
PHASe  
W(z) = arctan ( Im(z) / Re(z) )  
POLar  
COMPlex
x, y (Real and Imag)  
GDELay 
Group Delay, – d (z)/dX
REAL 
x
IMAGinary  
y
SWR 
Standing Wave Ratio, SWR = (1 + |z|) / (1 – |z|) 
LINPhase 
MLPhase
Lin Mag and Phase, |z|, arctan ( Im(z) / Re(z) )  
LOGPhase 
MDPhase
dB Mag and Phase, 20 * log|z| dB, arctan ( Im(z) / Re(z) )  
IMPedance 
R, X, L or C (depending on sign(X))  
ADMittance 
G, B, L/C (depending on sign(X)) 
C# PDF Password Library: add, remove, edit PDF file password in C#
This example shows how to add PDF file password with access permission setting. passwordSetting.IsAssemble = true; // Add password to PDF file.
view pdf metadata; view pdf metadata in explorer
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Document and metadata. All object data. File attachment. Flatten visible layers. VB.NET Demo Code to Optimize An Exist PDF File in Visual C#.NET Project.
preview edit pdf metadata; analyze pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
377
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth <x dB Bandwidth> 
Defines the bandfilter level, i.e. the minimum excursion for the bandpass and bandstop peaks.   
Use CALCulate<Chn>:MARKer<Mk>:BWIDth to set the x dB Bandwidth and query the results of a 
bandfilter search. Note the sign convention for input values.  
<Chn> 
Channel number used to identify the active trace.  
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may be set 
to any value because the bandfilter search functions always use markers Mkr1 to 
Mkr4. 
<x dB 
Bandwidth> 
xdB Bandwidth parameter.  
Range [def. unit] 
0.01 dB to 100.00 dB, the increment (UP. DOWN) is 0.3 dB. [dB]   
*RST value 
3dB. 
SCPI, 
Command 
Types 
Device-specific, command or query.  
Example: 
See CALCulate<Chn>:MARKer<Mk>:BWIDth.
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE BPASs | BSTop | 
BPRMarker | BSRMarker | BPAB | BSAB  
Selects the bandfilter search mode.  
<Chn> 
Channel number used to identify the active trace.  
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may be set 
to any value because the bandfilter search functions always use markers Mkr1 to 
Mkr4. 
Parameters 
Bandfilter search type:  
BPASs: Bandpass Search Ref to Max 
BSTop: Bandstop Search Ref to Max  
BPRMarker: Bandpass Search Ref to Marker 
BSRMarker: Bandstop Search Ref to Marker  
BPAB: Bandpass Absolute Level 
BSAB: Bandstop Absolute Level 
*RST value 
BPASs 
SCPI, 
Command 
Types 
Device-specific, command or query.  
Example: 
See CALCulate<Chn>:MARKer<Mk>:BWIDth.
CALCulate<Chn>:MARKer<Mk>:FUNCtion:CENTer 
Sets the center of the sweep range equal to the stimulus value of the marker <Mk> on trace no. <Chn>.  
<Chn> 
Channel number used to identify the active trace. 
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Document and metadata. All object data. Flatten visible layers. C#.NET DLLs: Compress PDF Document. Add necessary references: RasterEdge.Imaging.Basic.dll.
search pdf metadata; change pdf metadata
VB.NET PDF insert image library: insert images into PDF in vb.net
using RasterEdge.XDoc.PDF; Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
pdf metadata viewer; add metadata to pdf file
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
378
<Mk> 
Marker number in the range 1 to 10.  
Range [def. unit] 
–[–]  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
*RST; CALC:MARK ON   
Create marker 1 in the center of the current sweep range and assign it to 
trace no. 1.   
CALC:MARK:FUNC:CENT  
Leave the sweep range unchanged. 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DELTa[:STATe] <Boolean> 
Switches the delta mode for marker <Mk> on trace no. <Chn> on or off.  
Note: This command is the ZVR-compatible equivalent of 
CALCulate<Chn>:MARKer<Mk>:DELTa[:STATe].
<Chn> 
Channel number used to identify the active trace.  
<Mk> 
Marker number in the range 1 to 10.   
<Boolean> 
ON | OFF - Enables or disables the delta mode.  
*RST value 
OFF   
SCPI, Command Types 
Device-specific, command or query.   
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value> 
Assigns a search range no. <numeric_value> to marker no <Mk> and selects the search range, e.g. in 
order to define the start and stop values.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
<numeric_value> 
Number of the search range.  
Range [def. unit] 
0– fixed full span search range (equal to the sweep range)  
1to 10 – user-definable search ranges; see example   
*RST value 
0(reserved for full span search range) 
SCPI, Command Types 
Device-specific, command or query.  
Example: 
CALC1:MARK1:FUNC:DOM:USER 2  
Select the search range no. 2, assigned to marker no. 1 and trace no. 1. 
CALC:MARK:FUNC:DOM:USER:STARt 1GHz  
Set the start frequency of the search range to 1 GHz. 
CALC:MARK:FUNC:DOM:USER:STOP 1.2GHz 
Set the stop frequency of the search range to 1.2 GHz. 
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
Add necessary references: using RasterEdge.XDoc.PDF; Note: When you get the error "Could not load file or assembly 'RasterEdge.Imaging.Basic' or any other
change pdf metadata creation date; batch pdf metadata editor
VB.NET PDF File Split Library: Split, seperate PDF into multiple
1). ' Create output PDF file path list Dim outputFilePaths As New List(Of String) Dim i As Integer For i = 0 To splitIndex.Length outputFilePaths.Add(Program
embed metadata in pdf; google search pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
379
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt <numeric_value> 
Defines the start value of the search range selected with 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
<numeric_value> 
Beginning of the search range.  
Range [def. unit] 
Maximum allowed sweep range, depending on the instrument model and on 
the sweep type. [Hz, dBm or s, depending on the sweep type]  
*RST value 
0Hz  
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value> 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP <numeric_value> 
Defines the stop value of the search range selected with 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>.
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
<numeric_value> 
End of the search range.  
Range [def. unit] 
Maximum allowed sweep range, depending on the instrument model and on 
the sweep type. [Hz, dBm or s, depending on the sweep type]  
*RST value 
0Hz  
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value> 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum | MINimum | RPEak 
|LPEak | NPEak | TARGet | LTARget | RTARget | BFILter 
Selects a search mode for marker no. <Mk> and initiates the search. The marker must be created before 
using CALCulate<Chn>:MARKer<Mk>[:STATe] ON (exception: bandfilter search).  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. For a bandfilter search (BFILter) this 
numeric suffix is ignored and may be set to any value because the bandfilter 
search functions always use markers Mkr1 to Mkr4. 
Parameters 
See list of parameters below.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.  
Example: 
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
by directly tagging the second PDF file to the target one, this PDF file merge function VB.NET Project: DLLs for Merging PDF Documents. Add necessary references
delete metadata from pdf; get pdf metadata
C# PDF Library SDK to view, edit, convert, process PDF file for C#
PDF SDK for .NET allows you to read, add, edit, update, and delete PDF file metadata, like Title, Subject, Author, Creator, Producer, Keywords, etc.
bulk edit pdf metadata; pdf xmp metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
380
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK ON   
Create marker 1 and assign it to trace no. 1.  
CALC:MARK:FUNC:EXEC MAX; RES?  
Move the created marker to the absolute maximum of the trace and query the 
stimulus and response value of the search result.  
The analyzer provides the following search modes:  
Mode 
Find... 
MAXimum 
Absolute maximum in the search range (see CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER 
<numeric_value>) 
MINimum  Absolute maximum in the search range  
RPEak   Next valid peak to the right of the current marker position  
LPEak  
Next valid peak to the left  
NPEak  
Next highest or lowest value among the valid peaks (next peak) 
TARGet 
Target value (see CALCulate<Chn>:MARKer<Mk>:TARget)
RTARget  Next target value to the right of the current marker position  
LTARget 
Next target value to the left 
BFILter 
Bandfilter search. The results are queried using CALCulate<Chn>:MARKer<Mk>:BWIDth. 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult? 
Returns the result (stimulus and response value) of a search started by means of 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute. The search must be executed before the 
command is enabled.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may be set 
to any value.  
Response 
Search result in the format <Stimulus value>, <Response_value>.   
Range [def. 
unit] 
Depending on the measurement parameter. [unit depending on the sweep type and 
the marker format; see CALCulate<Chn>:MARKer<Mk>:FORMat]
*RST value 
SCPI, 
Command 
Types 
Device-specific, query only.  
Example: 
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute.
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect] MAXimum | MINimum | RPEak 
|LPEak | NPEak | TARGet | LTARget | RTARget | BFILter  
Selects a search mode for marker no. <Mk>, which can then be initiated using one of the 
CALCulate<Chn>:MARKer<Mk>:SEARch..., CALCulate<Chn>:MARKer<Mk>:MAXimum or 
CALCulate<Chn>:MARKer<Mk>:MINimum functions. The marker must be created before using 
CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
381
Note: This command is not needed expect for compatibility with ZVR programs. Use 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute to select a search mode and at the same 
time initiate the search. The CALCulate<Chn>:MARKer<Mk>:SEARch..., 
CALCulate<Chn>:MARKer<Mk>:MAXimum or CALCulate<Chn>:MARKer<Mk>:MINimum 
functions also select the search mode. 
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. For a bandfilter search (BFILter) this 
numeric suffix is ignored and may be set to any value because the bandfilter 
search functions always use markers Mkr1 to Mkr4. 
Parameters 
See list of parameters below.  
*RST value 
– 
SCPI, 
Command 
Types 
Device-specific, command or query.  
The analyzer provides the following search modes:  
Mode 
Selected search mode 
MAXimum 
Absolute maximum in the search range (see CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER 
<numeric_value>) 
MINimum 
Absolute maximum in the search range  
RPEak  
Next valid peak to the right of the current marker position  
LPEak   Next valid peak to the left  
NPEak  
Next highest value among the valid peaks 
TARGet 
Target value (see CALCulate<Chn>:MARKer<Mk>:TARget)
RTARget 
Next target value to the right of the current marker position  
LTARget 
Next target value to the left 
BFILter
Bandfilter search. Bandfilter search. The results are queried using CALCulate<Chn>:MARKer<Mk>:BWIDth. 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt 
Sets the beginning (start) of the sweep range equal to the stimulus value of the marker <Mk> on trace no. 
<Chn>.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
Range [def. unit] 
–[–]  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
*RST; CALC:MARK ON   
Create marker 1 in the center of the current sweep range and assign it to 
trace no. 1.   
CALC:MARK:FUNC:STAR  
Divide the sweep range in half, starting at the current marker position. 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
382
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP 
Sets the end (stop) of the sweep range equal to the stimulus value of the marker <Mk> on trace no. 
<Chn>.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
Range [def. unit] 
–[–]  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
*RST; CALC:MARK ON   
Create marker 1 in the center of the current sweep range and assign it to 
trace no. 1.   
CALC:MARK:FUNC:STOP  
Divide the sweep range in half, ending at the current marker position. 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARget <numeric_value> 
Defines the target value for the target search of marker no. <Mk>, which can be activated using 
CALCulate<Chn>:MARKer<Mk>:FUCTion:EXECute TARGet.
Note: This command is the ZVR-compatible equivalent of CALCulate<Chn>:MARKer<Mk>:TARget.
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.   
<numeric_value> 
Target search value of marker no. <Mk>.  
Range [def. unit] 
Depending on the format of the active trace (CALCulate<Chn>:FORMat). For 
adB Mag trace the range is –300 dB to +200 dB, the increment (UP. DOWN) 
is 0.1 dB. [dB, deg etc., depending on the trace format]   
*RST value 
Depending on the trace format; 0 dB for a dB Mag trace. 
SCPI, Command 
Types 
Device-specific, command or query.  
CALCulate<Chn>:MARKer<Mk>:MAXimum  
Selects a search mode for marker no. <Mk> and initiates a maximum search. The marker must be created 
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum.
<Chn> 
Channel number used to identify the active trace.
<Mk> 
Marker number in the range 1 to 10.  
Range [def. unit] 
– [–]  
*RST value 
– 
SCPI, Command Types 
Device-specific, no query.  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
383
CALCulate<Chn>:MARKer<Mk>:MINimum  
Selects a search mode for marker no. <Mk> and initiates a minimum search. The marker must be created 
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of 
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum.
<Chn> 
Channel number used to identify the active trace.
<Mk> 
Marker number in the range 1 to 10.  
Range [def. unit] 
– [–]  
*RST value 
– 
SCPI, Command Types 
Device-specific, no query.  
CALCulate<Chn>:MARKer<Mk>:MODE CONTinuous | DISCrete  
Sets marker no. <Mk> to continuous or discrete mode. The marker doesn't have to be created before 
(CALCulate<Chn>:MARKer<Mk>[:STATe] ON), the mode can be assigned in advance.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10.  
Parameters 
CONTinuous – marker can be positioned on any point of the trace, and its 
response values are obtained by interpolation. 
DISCrete –  marker can be set to discrete sweep points only.   
*RST value 
CONT 
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK:MODE DISC; MARK2 CONT   
Create marker 1 in discrete mode and marker 2 in continuous mode.  
CALC:MARK ON; MARK2 ON  
Display the two markers. Due to the different modes the horizontal position can 
be different. 
CALCulate<Chn>:MARKer<Mk>:REFerence:MODE CONTinuous | DISCrete  
Sets the reference marker to continuous or discrete mode. The marker doesn't have to be created before 
(CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON), the mode can be assigned in advance.  
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may be 
set to any value.  
Parameters 
CONTinuous – marker can be positioned on any point of the trace, and its 
response values are obtained by interpolation. 
DISCrete –  marker can be set to discrete sweep points only.   
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
384
*RST value 
CONT 
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK:REF:MODE DISC; :CALC:MARK2:REF:MODE CONT   
Create the reference marker in discrete mode and marker 2 in continuous 
mode.  
CALC:MARK:REF ON; :CALC:MARK2 ON  
Display the two markers. Due to the different modes the horizontal position can 
be different.  
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] <Boolean> 
Creates the reference marker and assigns it to trace no. <Chn>.   
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may be 
set to any value.   
<Boolean> 
ON | OFF - Creates or removes the marker.  
*RST value 
OFF 
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
Suppose that the active setup contains an active trace no. 1.  
CALC:MARK:REF ON; CALC:MARK ON 
Create the reference marker and marker 1 and assign them to trace no. 1. The 
default position of both markers is the center of the sweep range.  
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE NORMal | FIXed  
Sets the reference to normal or fixed mode. The marker must be created before using 
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.
<Chn> 
Channel number used to identify the active trace. 
<Mk> 
Marker number in the range 1 to 10. This numeric suffix is ignored and may 
be set to any value.  
Parameters 
NORMal – response value changes according to the measurement result. 
FIXed – marker keeps its current response value.   
*RST value 
NORMal 
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
CALC:MARK:REF ON; CALC:MARK:REF:TYPE FIX  
Create the reference markerand display it in the center of the sweep range 