R&S ZVL 
Preparing for Use 
Putting the Instrument into Operation 
Quick Start Guide 1303.6538.62 - 06
25
If the R&S ZVL is AC supplied the power on/off key switches between the following 
states: 
C
In the switched off state, both LEDs are off. The analyzer is completely 
disconnected from the power supply. The instrument is re-started using the mains 
power switch on the rear panel. 
C
In standby state, the right, amber LED is on. The standby power only supplies the 
power switch circuits, the optional oven quartz (OCXO, 10 MHz reference 
oscillator, option ZVL-B4, order no. 1164.1757.02), the battery (option NIMH 
Battery Pack, R&S FSL–B31), and the fan. The network adapter remains active, 
too. The power consumption of the R&S ZVL is much reduced. In this state it is 
safe to switch off the AC power and disconnect the instrument from the power 
supply. 
C
In ready state, the left, green LED is on, indicating that all modules are power-
supplied. After performing its startup procedure, the analyzer is ready for 
operation.   
The SETUP -> More -> Shutdown Off/Standby softkey determines the behavior of the 
power on/off key:  
C
In the default setting (Shutdown: Off), the standby state is blocked. The power 
on/off key switches the R&S ZVL from ready to off state. Use this instrument 
configuration if you want to avoid any power consumption while you do not use 
your instrument.  
C
If Shutdown: Standby is selected, you can use the power on/off key to toggle 
between standby and ready state. Use this configuration if you want to be able to 
resume your measurements quickly, avoiding the complete startup procedure and 
the system check.  
If the R&S ZVL is DC or battery supplied, the Shutdown:Standby option is not effective, 
see section DC Power Supply and Battery below.   
Power supply in standby mode 
The instrument is still power-supplied while it is in standby mode. 
1.3.10  Replacing Fuses 
The instrument is protected by two fuses (IEC 127 – T 3.15 H / 250 V) located on the 
rear panel at the right side of the AC power switch.  
Add metadata to pdf file - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
adding metadata to pdf files; edit multiple pdf metadata
Add metadata to pdf file - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
pdf metadata online; read pdf metadata online
R&S ZVL 
Preparing for Use 
Putting the Instrument into Operation 
Quick Start Guide 1303.6538.62 - 06
26
Shock hazard 
For fuse replacement, ensure that the instrument is switched off and disconnected from 
the power supply by removing the plug from the AC and DC power connector. 
To replace the fuses 
1.  Open the lid of the AC power connector.  
2.  Lift the fuse holder out of its slot.  
3.  Exchange the two fuses.  
4.  Put the fuse holder back in its slot and close the lid.  
1.3.11  DC Power Supply and Battery 
While the R&S ZVL is disconnected from the AC power supply, it can be supplied either 
by a DC power (option DC Power Supply, R&S FSL–B30) or a battery (option NIMH 
Battery Pack, R&S FSL–B31); see Power Supply Options. With DC power or battery 
supply, the AC power switch on the rear panel is disabled. The power on/off key on the 
front panel is used to switch the analyzer on or off, irrespective of the SETUP -> More -
>Shutdown Off/Standby setting. 
C
While the R&S ZVL is switched off, press the power on/off key to initiate the startup 
procedure until the R&S ZVL enters its ready state. The left, green LED is on. 
C
While the R&S ZVL is switched on (ready), press the power on/off key to switch the 
analyzer off. Both LEDs are off.  
When using the DC power supply or battery, note the safety instructions below. Please 
also refer to the detailed information about DC Power Supply and the Battery Pack.  
C# PDF Password Library: add, remove, edit PDF file password in C#
This example shows how to add PDF file password with access permission setting. passwordSetting.IsAssemble = true; // Add password to PDF file.
pdf metadata viewer online; rename pdf files from metadata
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Document and metadata. All object data. File attachment. Flatten visible layers. VB.NET Demo Code to Optimize An Exist PDF File in Visual C#.NET Project.
pdf metadata editor; pdf remove metadata
R&S ZVL 
Quick Start Guide 1303.6538.62 - 06
27
DC power supply 
The power supply (SELV) that is used must fulfill the requirements for reinforced/double 
insulation for main supply circuits in accordance to DIN/EN/IEC 61010 (UL 61010B–1, 
CSA C22.2 No. 1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). It is 
recommended to fuse the DC power supply appropriately. Before switching on the 
instrument check the connection for correct polarity. 
Battery pack 
The power supply that is used must fulfill the requirements for reinforced/double 
insulation for main supply circuits in accordance to DIN/EN/IEC 61010 (UL 61010B–1, 
CSA C22.2 No. 1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950).  
Note: If the battery is not to be used for a longer time, it is recommended to remove it 
and store it separately. 
1.3.12  Charging the Battery 
The battery can be charged via the AC or the DC power supply without being removed: 
C
To charge the battery using the AC power supply, connect the instrument to the 
mains supply and switch on the AC power on the rear panel. The power on/off key 
on the front panel can be set either way. 
C
To charge the battery using the DC power supply, connect the instrument to the DC 
power supply. 
If you have more than one battery packs in use, you can also charge them outside the 
instrument using the power supply unit of option R&S FSL–Z4 (DC Supply for FSL–
B31, stock no. 4052.3041.00). An LED with the inscription "Charge" is switched on 
during charging.  
Charging conditions 
Approximate duration 
AC power supply, standby mode 
5 h 
AC power supply, instrument is switched on 
9 h 
DC power supply, instrument is switched off 
5 h 
DC power supply, instrument is switched on 
9 h 
External charging (battery removed) 
5 h 
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Document and metadata. All object data. Flatten visible layers. C#.NET DLLs: Compress PDF Document. Add necessary references: RasterEdge.Imaging.Basic.dll.
batch edit pdf metadata; batch pdf metadata
VB.NET PDF insert image library: insert images into PDF in vb.net
using RasterEdge.XDoc.PDF; Have a try with this sample VB.NET code to add an image to the first page of PDF file. ' Open a document.
remove metadata from pdf online; remove pdf metadata online
R&S ZVL 
Preparing for Use 
Maintenance 
Quick Start Guide 1303.6538.62 - 06
28
1.4  Maintenance 
The network analyzer does not require any special maintenance. Make sure that the 
air vents are not obstructed. The outside of the instrument is suitably cleaned using a 
soft, line-free dust cloth.  
Instrument damage caused by cleaning agents. Cleaning agents contain substances 
that may damage the instrument, e.g. solvent-containing cleaning agents may damage 
the front panel labeling or plastic parts. 
Never use cleaning agents such as solvents (thinners, acetone, etc), acids, bases, or 
other substances. 
The address of our support center and a list of useful contact addresses can be found 
on the Contact page.  
1.4.1  Storing and Packing 
The network analyzer can be stored at the temperature range quoted in the data sheet. 
When it is stored for a longer period of time the instrument should be protected against 
dust.  
The original packing should be used, particularly the protective caps at the front and 
rear, when the instrument is to be transported or dispatched. If the original packing is 
no longer available, use a sturdy cardboard box of suitable size and carefully wrap the 
instrument to protect it against mechanical damage.  
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
Add necessary references: using RasterEdge.XDoc.PDF; Note: When you get the error "Could not load file or assembly 'RasterEdge.Imaging.Basic' or any other
pdf metadata editor online; remove pdf metadata
VB.NET PDF File Split Library: Split, seperate PDF into multiple
1). ' Create output PDF file path list Dim outputFilePaths As New List(Of String) Dim i As Integer For i = 0 To splitIndex.Length outputFilePaths.Add(Program
acrobat pdf additional metadata; pdf metadata
R&S ZVL 
Preparing for Use 
Starting and Shutting down the Analyzer 
Quick Start Guide 1303.6538.62 - 06
29
1.5  Starting and Shutting down the Analyzer  
To start the analyzer, proceed as follows: 
C
If you use the AC power supply, make sure that the instrument is connected to the 
mains supply and switch the power switch on the rear panel to position I (On).  
C
If you use the DC power supply, make sure that the instrument is connected and 
press the power on/off key on the front panel.   
The analyzer automatically performs a system check, boots the Windows XP operating 
system and then starts the analyzer (NWA) application. If the last analyzer session was 
terminated regularly, the NWA application uses the last setup with all instrument 
settings.  
To shut down the analyzer, proceed as follows: 
C
With AC power supply, press the power on/off key to save the current setup, close 
the NWA application, shut down Windows XP and set the instrument to standby or 
off state, depending on the SETUP -> More -> Shutdown Off/Standby settings. 
Refer to section Instrument States with AC Power Supply. From the standby state, 
you can set the AC power switch to position 0 (Off) to turn the analyzer completely 
off..   
C
With DC power supply or battery, press the power on/off key to switch the analyzer 
off.  
Power on/off key, OCXO 
Do not press the power on/off key longer than 3 seconds. 
The AC power switch also interrupts the power supply of the OCXO (option OCXO 
Reference Frequency, R&S FSL–B4). When you switch the instrument back on, be 
sure to comply with the extended warm–up phase specified in the data sheet. 
Saving instrument settings 
When using the AC power supply, it is strongly recommended to press the power on/off 
key on the front panel before disconnecting the power. If you set the power switch on 
the rear panel to 0 while the NWA application is still running, you will lose the current 
settings. Moreover, loss of program data can not be excluded if the application is 
terminated improperly. 
With DC power supply or battery, it is safe to switch off the instrument using the power 
on/off key on the front panel. 
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
by directly tagging the second PDF file to the target one, this PDF file merge function VB.NET Project: DLLs for Merging PDF Documents. Add necessary references
edit pdf metadata online; pdf metadata reader
C# PDF Library SDK to view, edit, convert, process PDF file for C#
PDF SDK for .NET allows you to read, add, edit, update, and delete PDF file metadata, like Title, Subject, Author, Creator, Producer, Keywords, etc.
remove metadata from pdf; endnote pdf metadata
R&S ZVL 
Preparing for Use 
Connecting External Accessories 
Quick Start Guide 1303.6538.62 - 06
30
1.6  Connecting External Accessories 
The equivalent USB ports on the front panel of the analyzer can be used to connect a 
variety of accessories: 
C
Amouse simplifies operation of the instrument using the controls and dialogs of the 
Graphical User Interface (GUI). 
C
Akeyboard simplifies the entry of data.  
C
Aprinter generates hard copies of the screen contents. 
In addition the analyzer provides an interface for network integration:  
C
An external monitor shows the magnified GUI with all diagram areas, 
measurement results and control elements. 
C
ALAN connection can be established in order to access the hard disk or control 
the analyzer from an external PC.  
1.6.1  Connecting a Mouse 
AUSB mouse can be connected to one of the Universal Serial Bus connectors on the 
front panel.   
The mouse is detected automatically when it is connected. It is safe to connect or 
disconnect the mouse during the measurement.  
R&S ZVL 
Quick Start Guide 1303.6538.62 - 06
31
R&S ZVL 
Quick Start Guide 1303.6538.62 - 06
32
1.6.4  Connecting a Monitor 
Astandard DVI monitor can be connected to the DVI-D connector in the lower right 
corner of the rear panel. The monitor displays the magnified R&S ZVL screen with all 
diagram areas, measurement results and control elements. Connection of an analog 
VGA monitor is not supported. 
After connecting the monitor, use the SETUP – General Setup – More – Monitor Int/Ext 
softkey to display the screen contents on the external monitor. To display the screen 
contents on the R&S ZVL and on the external monitor simultaneously, proceed as 
follows:  
1.  Connect an external keyboard to the instrument and press CTRL + ESC to access 
the operating system.  
2.  On the right side of the task bar, click the monitor icon. 
3.  Select Graphics Options – Output To – Intel(R) Dual Display Clone – Monitor + 
Digital Display. 
R&S ZVL 
Quick Start Guide 1303.6538.62 - 06
33
C
Adedicated network connection between the analyzer and a single computer 
made with a cross-over RJ-45 network cable. The computer must be equipped with 
anetwork adapter and is directly connected to the analyzer. The use of hubs, 
switches, or gateways is not needed, however, data transfer is still made using the 
TCP/IP protocol.  
The IP address information is displayed in the Info – Setup Info dialog. 
R&S ZVL 
Preparing for Use 
Remote Control in a LAN 
Quick Start Guide 1303.6538.62 - 06
34
1.7  Remote Control in a LAN 
ALAN connection is used to integrate the analyzer into a home/company network. This 
offers several applications:  
C
Transfer data between a controller and the analyzer, e.g. in order run a remote 
control program.  
C
Control the measurement from a remote computer using the Remote Desktop 
application.   
C
Use external network devices (e.g. printers).  
Virus protection 
An efficient virus protection is a prerequisite for secure operation in the network. Never 
connect your analyzer to an unprotected network because this may cause damage to 
the instrument software. 
To establish the connection proceed as follows: 
1.  Access Windows XP, e.g. by pressing SETUP – General Setup – More – Open 
Start Menu (see section Accessing Windows XP's Start Menu on p. 37).  
2.  Open Windows XP's control panel.  
3.  Select System and open the Remote tab in the System Properties dialog. Enable 
Allow users to connect remotely to this computer.
4.  Assign an IP address to the analyzer following the directions below and connect 
the analyzer to the network as described in Connecting a LAN Cable.
5.  Create a Remote Desktop Connection using the analyzer's IP address. 
Password protection 
The analyzer uses a user name and password as credentials for remote access. In the 
factory configuration, instrument is preset for both the user name and the password. To 
protect the analyzer from unauthorized access, it is recommended to change the 
factory setting. 
1.7.1  Assigning an IP Address 
Depending on the network capacities, the TCP/IP address information for the analyzer 
can be obtained in different ways.   
C
If the network supports dynamic TCP/IP configuration using the Dynamic Host 
Configuration Protocol (DHCP), all address information can be assigned 
automatically.   