R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
415
Example: 
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold
CALCulate<Chn>:TRANsform:DTFault:PEAK:DATA<nr>? 
Returns the distance and peak response values of the detected fault no. <nr>. If the number of detected 
faults (CALCulate<Chn>:TRANsform:DTFault:PEAK:COUNt?) is smaller than <nr>, an error 
message is returned.  
<Chn> 
Channel number used to identify the active trace  
<nr> 
Current number of the fault. Faults are numbered according to their distance 
(fault no. 1 has the smallest distance)  
Response  
Distance and peak response value of the fault, separated by a comma; see 
example.   
*RST value 
N/a  
SCPI, Command 
Types 
Device-specific, query only 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold 
CALCulate<Chn>:TRANsform:DTFault:PEAK:STATe <Boolean> 
Enables or disables the fault limit check.  
<Chn> 
Channel number used to identify the active trace.  
<Boolean> 
ON | OFF - Limit check on or off.  
*RST value 
OFF  
SCPI, Command Types 
Device-specific, command or query.   
Example: 
See CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold 
CALCulate<Chn>:TRANsform:DTFault:PEAK:THReshold 
Starts the automatic calculation of the number of points.  
<Chn> 
Channel number used to identify the active trace  
Parameters 
N/a 
*RST value 
SCPI, 
Command 
Types 
Device-specific, query only 
Example: 
*RST; :CALC:TRAN:DTF:STAT ON  
Enable distance-to-fault measurement with default settings.  
CALC:TRAN:DTF:PEAK:THR –20  
Specify a fault limit of –20 dB relative tor the absolute maximum of the active trace 
in channel 1.  
C# read pdf metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
batch update pdf metadata; analyze pdf metadata
C# read pdf metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
pdf metadata editor; read pdf metadata online
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
416
CALC:TRAN:DTF:PEAK:STAT ON  
Enable the fault limit check for the active trace.  
CALC:TRAN:DTF:PEAK:COUN?   
Query the number of detected faults. The network analyzer responds with an 
integer number, e.g. 2.  
CALC:TRAN:DTF:PEAK:DATA2?  
Query the distance and peak response values of the second detected fault. 
Possible response: 0.09615384615,-18.07471664 (for a distance of 
0.09615384615 units of length and a peak response value of -18.07471664 dB).  
CALCulate<Chn>:TRANsform:DTFault:POINts 
Starts the automatic calculation of the number of points.  
<Chn> 
Channel number used to identify the active trace 
Parameters 
N/a 
*RST value 
SCPI, Command Types 
Device-specific, no query 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:STATe 
CALCulate<Ch>:TRANsform:DTFault:SELect '<cable name>' 
Selects a standard cable or a previously defined cable identified by its name.   
<Ch> 
Channel number. Note that the cable type is channel-specific (rather than 
trace-specific).   
'<cable name>' 
Cable name used to identify the cable  
*RST value 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:DEFine 
CALCulate<Chn>:TRANsform:DTFault:SPAN <numeric_value> 
Defines the span of the distance-to-fault diagram.  
<Chn> 
Channel number used to identify the active trace   
<numeric_value> 
Distance span   
Range [def. unit] 
1mm to +15 m. The increment (parameters UP or DOWN) is 1 mm. [m or foot 
(ft), depending on the Distance Unit selected in the System Configuration 
dialog]  
*RST value 
10 m 
SCPI, Command 
Device-specific, command or query 
C# PDF Image Extract Library: Select, copy, paste PDF images in C#
such as its location, zonal information, metadata, and so on. Extract image from PDF free in .NET framework with trial SDK components and online C# class source
pdf keywords metadata; view pdf metadata in explorer
How to C#: Basic SDK Concept of XDoc.PDF for .NET
Various PDF annotation features can be integrated into your C# project, such as XDoc.PDF for .NET supports editing PDF document metadata, like Title
pdf metadata online; google search pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
417
Types 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:CENTer  
CALCulate<Chn>:TRANsform:DTFault:STARt <numeric_value> 
Defines the start distance of the distance-to-fault diagram.  
<Chn> 
Channel number used to identify the active trace   
<numeric_value> 
Start distance  
Range [def. unit] 
–5 m to +9.999 m. The increment (parameters UP or DOWN) is 1 mm. [m or 
foot (ft), depending on the Distance Unit selected in the System Configuration 
dialog]  
*RST value 
0m
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:CENTer  
Note: If the start frequency entered is greater than the current stop frequency 
(CALCulate<Chn>:TRANsform:DTFault:STOP), the stop frequency is set to the start 
frequency plus the minimum frequency span (CALCulate<Chn>:TRANsform:DTFault:SPAN).  
CALCulate<Chn>:TRANsform:DTFault:STATe <Boolean> 
Enables or disables the distance-to-fault measurement. In contrast to manual control, the number of points 
is not automatically re-adjusted (CALCulate<Chn>:TRANsform:DTFault:POINts).  
<Chn> 
Channel number used to identify the active trace   
<Boolean> 
ON – Distance-to-fault measurement active  
OFF – Distance-to-fault measurement active  
*RST value 
OFF [–]  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
*RST; :CALC:TRAN:DTF:STAT ON; :SWE:POIN?  
Enable distance-to-fault measurement with default settings. The number of 
sweep points is equal to the default value 201. 
CALC:TRAN:DTF:POIN; :SWE:POIN?  
Adjust the number of sweep points to the default channel settings. For an R&S 
ZVL network analyzer, the result is 519.  
CALC:TRAN:DTF:STAR -1; STOP 5 
Select a distance range between -1 m and +5 m.  
CALC:TRAN:DTF:SEL 'RG214' 
Select a RF214 cable type for the distance-to-fault measurement.  
FREQ:SPAN 2GHz; :CALC:TRAN:DTF:POIN; :SWE:POIN?  
Reduce the frequency span, re-calculate the number of points. The query 
returns 201.  
VB.NET PDF Image Extract Library: Select, copy, paste PDF images
Get image information, such as its location, zonal information, metadata, and so on. Extract image from PDF free in .NET framework application with trial SDK
remove metadata from pdf file; pdf metadata reader
C# PDF Text Extract Library: extract text content from PDF file in
XDoc.PDF ›› C# PDF: Extract PDF Text. C# PDF - Extract Text from PDF in C#.NET. Best C#.NET PDF text extraction library and component for free download.
rename pdf files from metadata; batch edit pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
418
CORR:COLL:METH:DEF 'DTF Cal', FOPort, 2 
CORR:COLL:ACQ:SEL OPEN, 2 
CORR:COLL:ACQ:SEL SHOR, 2 
CORR:COLL:ACQ:SEL MATC, 2 
CORR:COLL:SAVE:SEL 
Perform a full one-port calibration for the distance-to-fault measurement.  
CALCulate<Chn>:TRANsform:DTFault:STOP <numeric_value> 
Defines the stop distance of the distance-to-fault diagram.  
<Chn> 
Channel number used to identify the active trace   
<numeric_value> 
Stop distance  
Range [def. unit] 
–4.999 m to +10 m. The increment (parameters UP or DOWN) is 1 mm. [m or 
foot (ft), depending on the Distance Unit selected in the System Configuration 
dialog]  
*RST value 
+10 m 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
See CALCulate<Chn>:TRANsform:DTFault:CENTer  
Note: If the stop frequency entered is smaller than the current start frequency 
(CALCulate<Chn>:TRANsform:DTFault:STARt), the start frequency is set to the stop 
frequency minus the minimum distance span of 1 mm 
(CALCulate<Chn>:TRANsform:DTFault:SPAN).  
CALCulate<Ch>:TRANsform:VNETworks...     
This subsystem defines the circuit models for single ended and balanced port (de-)embedding and 
activates the (de-)embedding function.  
The circuit models are referenced by means of predefined character data parameters.  
Circuit models for embedding/deembedding: 
Parameter 
Circuit model 
Capacitances 
Resistances 
Inductances 
FIMPort 
File import, no circuit model  
– 
– 
– 
CSL 
Serial C, shunt L  
C1 
R1, R2 
L1 
LSC 
Serial L, shunt C  
C1 
R1, R2 
L1 
CSC 
Serial C, shunt C  
C1, C2 
R1, R2 
– 
C# PDF - Read Barcode on PDF in C#.NET
NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET edit PDF digital signatures, C#.NET edit PDF sticky note, C#.NET print PDF, C#.NET read barcodes from
view pdf metadata; remove pdf metadata
C# PDF Library SDK to view, edit, convert, process PDF file for C#
C#.NET: Edit PDF Metadata. PDF SDK for .NET allows you to read, add, edit, update, and delete PDF file metadata, like Title, Subject, Author, Creator, Producer
extract pdf metadata; preview edit pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
419
LSL 
Serial L, shunt L  
– 
R1, R2 
L1, L2 
SLC 
Shunt L, serial C  
C1 
R1, R2 
L1 
SCL 
Shunt C, serial L  
C1 
R1, R2 
L1 
SCC 
Shunt C, serial C  
C1, C2 
R1, R2 
– 
SLL 
Shunt L, serial L  
– 
R1, R2 
L1, L2 
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARame
ters:C<no> CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>  
Specifies the capacitance values C1, C2 in the different circuit models for deembedding.   
<Ch> 
Channel number   
<Ph_pt> 
Physical port number  
<no> 
Number of capacitance in circuit model. If unspecified the numeric suffix is set 
to 1.  
Range [def. unit] 
1to 2 [–] The number of capacitances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
<numeric_value> 
Capacitance C<no> for the specified circuit model.  
Range [def. unit] 
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).  
*RST value 
1pF (10
–12
F) 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the capacitance value for 
the specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC  
Query the default capacitance C2 for the Serial C, shunt C circuit model. The 
response is 1E-012 (1 pF).  
CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12  
Increase the capacitance to 2.2 pF.   
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARame
ters:L<no> CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>  
Specifies the inductance values L1, L2, L3 in the different circuit models for deembedding.   
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<no> 
Number of inductance in circuit model. If unspecified the numeric suffix is set 
to 1.  
VB.NET PDF: Basic SDK Concept of XDoc.PDF
XDoc.PDF for .NET supports editing PDF document metadata, like Title, Subject, Author, Creator, Producer, Keywords, Created Date, and Last Modified Date.
batch pdf metadata editor; remove metadata from pdf online
C# PDF File Compress Library: Compress reduce PDF size in C#.net
Bookmarks. Comments, forms and multimedia. Document and metadata. All object data. File attachment. Flatten visible layers. C#.NET DLLs: Compress PDF Document.
get pdf metadata; pdf xmp metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
420
Range [def. unit] 
1to 3 [–] The number of inductances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
<numeric_value> 
Inductance L<no> for the specified circuit model.  
Range [def. unit] 
–1H to 1 H [H]. The increment is 1 pH (10
–12
H).  
*RST value 
1nH (10
–9
H) 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the inductance value for the 
specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:DEEM:PAR:L1? SLL  
Query the default inductance L1 for the Shunt L, serial L circuit model. The 
response is 1E-009 (1 nH).  
CALC:TRAN:VNET:SEND:DEEM:PAR:L1 SLL, 2.2E-9  
Increase the inductance to 2.2 nH.   
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARame
ters:R<no> CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL,<numeric_value>  
Specifies the resistance values R1, R2, R3 in the different circuit models for deembedding.   
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<no> 
Number of resistance in circuit model. If unspecified the numeric suffix is set to 
1.  
Range [def. unit] 
1to 3 [–] The number of resistances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
<numeric_value> 
Resistance R<no> for the specified circuit model.  
Range [def. unit] 
–10 MV to 10 MV [V]. The increment is 1 mV (10
–3
V).  
*RST value 
0V for all resistances connected in series with an inductance 
10 MV for all resistances connected in parallel with a capacitance 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the resistance value for the 
specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:DEEM:PAR:R1? CSL; R2? CSL  
Query the default resistances for the Serial C, shunt L circuit model. The 
response is 10000000; 0.  
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3  
Increase the resistance R2 to 2.2 kV.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>[:STATe] 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
421
<Boolean>  
Enables or disables the deembedding function. It is allowed to change the circuit model and its 
parameters while deembedding is enabled.  
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<Boolean> 
ON – Deembedding active 
OFF – Deembedding inactive  
*RST value 
OFF [–]  
SCPI, Command 
Types 
Device-specific, command or query  
Example: 
CALC:TRAN:VNET:SEND:DEEM:TND CSL  
Select the Serial C, shunt L circuit model for deembedding.  
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3; CALC:TRAN:VNET:SEND:DEEM ON 
Increase the resistance R2 for the Serial C, shunt L circuit model to 2.2 kV and 
enable deembedding.  
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:TNDefini
tion FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL  
Selects the circuit model for deembedding.   
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
CSL 
SCPI, Command Types 
Device-specific, command or query.   
Example: 
See CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding[:STATe] 
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameter
s:C<no> CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>  
Specifies the capacitance values C1, C2 in the different circuit models.   
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<no> 
Number of capacitance in circuit model. If unspecified the numeric suffix is set 
to 1.  
Range [def. unit] 
1to 2 [–] The number of capacitances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
422
<numeric_value> 
Capacitance C<no> for the specified circuit model.  
Range [def. unit] 
–1mF to 1 mF [F]. The increment is 1 fF (10
–15
F).  
*RST value 
1pF (10
–12
F) 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the capacitance value for 
the specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:EMB:PAR:C2? CSC  
Query the default capacitance C2 for the Serial C, shunt C circuit model. The 
response is 1E-012 (1 pF).  
CALC:TRAN:VNET:SEND:EMB:PAR:C2 CSC, 2.2E-12  
Increase the capacitance to 2.2 pF.   
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameter
s:L<no> CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>  
Specifies the inductance values L1, L2, L3 in the different circuit models.   
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<no> 
Number of inductance in circuit model. If unspecified the numeric suffix is set 
to 1.  
Range [def. unit] 
1to 3 [–] The number of inductances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
<numeric_value> 
Inductance L<no> for the specified circuit model.  
Range [def. unit] 
–1H to 1 H [H]. The increment is 1 pH (10
–12
H).  
*RST value 
1nH (10
–9
H) 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the inductance value for the 
specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:EMB:PAR:L1? SLL  
Query the default inductance L1 for the Shunt L, serial L circuit model. The 
response is 1E-009 (1 nH).  
CALC:TRAN:VNET:SEND:EMB:PAR:L1 SLL, 2.2E-9  
Increase the inductance to 2.2 nH.   
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameter
s:R<no> CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL,<numeric_value>  
Specifies the resistance values R1, R2, R3 in the different circuit models.   
<Ch> 
Channel number.   
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
423
<Ph_pt> 
Physical port number  
<no> 
Number of resistance in circuit model. If unspecified the numeric suffix is set to 
1.  
Range [def. unit] 
1to 3 [–] The number of resistances depends on the selected circuit model.  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
–(see command type description below) 
<numeric_value> 
Resistance R<no> for the specified circuit model.  
Range [def. unit] 
–10 MV to 10 MV [V]. The increment is 1 mV (10
–3
V).  
*RST value 
0V for all resistances connected in series with an inductance 
10 MV for all resistances connected in parallel with a capacitance 
SCPI, Command 
Types 
Device-specific, command or query. In the query form, the <numeric_value> 
parameter must be omitted. The command returns the resistance value for the 
specified circuit model.  
Example: 
*RST; CALC:TRAN:VNET:SEND:EMB:PAR:R1? CSL; R2? CSL  
Query the default resistances for the Serial C, shunt L circuit model. The 
response is 10000000; 0.  
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3  
Increase the resistance R2 to 2.2 kV.
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>[:STATe] 
<Boolean>  
Enables or disables the embedding function. It is allowed to change the circuit model and its parameters 
while embedding is enabled.  
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
<Boolean> 
ON – Embedding active 
OFF – Embedding inactive  
*RST value 
OFF [–]  
SCPI, Command 
Types 
Device-specific, command or query  
Example: 
CALC:TRAN:VNET:SEND:EMB:TND CSL  
Select the Serial C, shunt L circuit model for embedding.  
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3; CALC:TRAN:VNET:SEND:EMB ON  
Increase the resistance R2 for the Serial C, shunt L circuit model to 2.2 kV
and enable embedding.  
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:TNDefinitio
nFIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL  
Selects the circuit model for sembedding.   
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
424
<Ch> 
Channel number.   
<Ph_pt> 
Physical port number  
Parameters  
Possible circuit models (character data); see circuit models.  
*RST value 
CSL 
SCPI, Command Types 
Device-specific, command or query.   
Example: 
See CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding[:STATe] 
CONFigure...     
This subsystem creates and deletes channels and assigns names to channels. The commands are 
device-specific.  
CONFigure:CHANnel<Ch>:CATalog? 
Returns the numbers and names of all channels in the current setup.  
<Ch> 
Channel number. This parameter is ignored because the command returns all 
channels.   
Response 
String with comma-separated list of channel numbers and names, see example 
below. If all channels have been deleted the response is an empty string (""). 
*RST value 
SCPI, 
Command 
Types 
Device-specific, query only.  
Example: 
*RST; CONF:CHAN2:STAT ON; NAME 'New Channel'  
Create channel 2 and assign the channel name "New Channel".   
CONF:CHAN:CAT?  
Query all channels and their names. As a default channel no. 1 is created on 
*RST, the response is '1,Ch1,2,New_Channel'.  
CONF:CHAN:NAME:ID? 'New Channel'  
Query the channel number for the channel named "New Channel". The response 
is 2.   
CONFigure:CHANnel<Ch>:NAME '<Ch_name>' 
Assigns a name to channel number <Ch>. The channel must be created before 
(CONFigure:CHANnel<Ch>[:STATe] ON). Moreover it is not possible to assign the same name to two 
different channels. CONFigure:CHANnel<Ch>:CATalog? returns a list of all defined channels with their 
names.  
<Ch> 
Number of an existing channel.   