R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
425
'<Ch_name>' 
Channel name, e.g. 'Channel 4'. 
*RST value 
'Ch1'  
SCPI, Command Types 
Device-specific, command or query. 
Example: 
See CONFigure:CHANnel<Ch>:CATalog?
CONFigure:CHANnel<Ch>:NAME:ID? '<Ch_name>' 
Queries the channel number (numeric suffix) of a channel with kown channel name. A channel name must 
be assigned before (CONFigure:CHANnel<Ch>NAME '<Ch_name>'). 
CONFigure:CHANnel<Ch>:CATalog? returns a list of all defined channels with their names.  
<Ch> 
Channel number. This suffix is not relevant and may be omitted (the command 
returns the right channel number).  
'<Ch_name>' 
Channel name, e.g. 'Channel 4'. 
*RST value 
SCPI, Command 
Types 
Device-specific, command or query.  
Example: 
See CONFigure:CHANnel<Ch>:CATalog? 
CONFigure:CHANnel<Ch>[:STATe] <Boolean> 
Creates or deletes channel no. <Ch> and selects it as the active channel. 
CONFigure:CHANnel<Ch>:NAME defines the channel name.  
Achannel created with CONFigure:CHANnel<Ch>[:STATe] ON can be configured but has no 
trace assigned so that no measurement can be initiated. Use 
CALCulate<Ch>:PARameter:SDEFine "<Trace_name>,"<Parameter>" to create a new 
channel and a new trace.   
In remote control it is possible to remove all channels. This is in contrast to manual control where 
at least one channel with one diagram area and one trace must be available.  
<Ch> 
Number of the channel to be created or deleted.   
<Boolean> 
ON - Create channel no. <Ch>. If the channel no. <Ch> exists already, it is not 
modified but selected as the active channel  
OFF - Delete channel no. <Ch>.   
*RST value 
ON for channel no. 1 (created on *RST), OFF for all other channels.    
SCPI, Command 
Types 
Device-specific, command or query.   
Example: 
See CONFigure:CHANnel<Ch>:CATalog? 
<Ch> 
Number of the channel to be created or deleted.   
<Boolean> 
ON - Create channel no. <Ch>. If the channel no. <Ch> exists already, it is not 
modified but selected as the active channel  
Pdf xmp metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
clean pdf metadata; adding metadata to pdf files
Pdf xmp metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
batch pdf metadata; pdf metadata viewer
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
426
OFF - Delete channel no. <Ch>.   
*RST value 
ON for channel no. 1 (created on *RST), OFF for all other channels.    
SCPI, Command 
Types 
Device-specific, command or query.   
Example: 
See CONFigure:CHANnel<Ch>:CATalog? 
CONFigure:TRACe<Trc>:CATalog?  
Returns the numbers and names of all traces in the current setup.  
<Trc> 
Trace number. This suffix is ignored; the command returns a list of all traces.  
Response 
String with comma-separated list of trace numbers and names, see example 
below. If all traces have been deleted the response is an empty string (""). 
*RST value 
SCPI, 
Command 
Types 
Device-specific, query only.  
Example: 
*RST; CALC2:PAR:SDEF 'Ch2Trc2', 'S11'  
Create channel 2 and a new  trace named Ch2Trc2.
CONF:TRAC:CAT?  
Query all traces and their names. As a default trace no. 1 is created upon *RST, 
the response is '1,TRC1,2,Ch2Trc2'.  
CONF:TRAC:NAME:ID? 'Ch2Trc2'  
Query the trace number for the trace named "Ch2Trc2". The response is 2.   
CONF:TRAC2:NAME?   
Query the trace name for trace no. 2. The response is 'Ch2Trc2'.   
CONF:TRAC:CHAN:NAME? 'Ch2Trc2'  
Query the channel name for trace Ch2Trc2. The response is 'Ch2'.   
CONF:TRAC:CHAN:ID? 'Ch2Trc2'  
Query the channel number for trace Ch2Trc2. The response is 2.   
DIAGnostic:SERVice...   
This subsystem provides access to service and diagnostic routines used in service, maintenance and 
repair. In accordance with the SCPI standard all commands are device-specific.  
Service functions are password-protected (SYSTem:PASSword[:CENable]) and should be used by a 
R&S service representative only. Refer to the service manual for more information.  
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
How to Get TIFF XMP Metadata in C#.NET. Use this C# sample code to get Tiff image Xmp metadata for string. // Load your target Tiff docuemnt.
acrobat pdf additional metadata; endnote pdf metadata
DocImage SDK for .NET: Document Imaging Features
a metadata viewer application Enable users to add metadata in the form of EXIF, IPTC, XMP, or COM Type 6 (OJPEG) encoding Image only PDF encoding support.
search pdf metadata; pdf metadata editor online
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
427
DIAGnostic:SERVice:SFUNction '<string>',... 
Activates a service function (for internal use only).  
DISPlay...     
This subsystem controls the selection and presentation of graphical and trace information on the screen.  
Note: Traces are generally identified by a string parameter defining the trace name (e.g. 
CALCulate<Ch>:PARameter:SELect <Trace_Name>). In the DISPlay... subsystem, traces 
are assigned to diagram areas (DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED 
<Trace_Name>). While this assignment is valid, the trace is identified by the numeric suffix 
<WndTr> and the trace name is not needed.  
DISPlay:CMAP<Element>:MARKer[:STATe] <Boolean>  
Displays all markers with the same color or display each marker with the color of the associated trace. The 
colors of all display elements are defined via DISPlay:CMAP<Element>:RGB <red>, <green>, 
<blue>,.... 
<Element> 
Numeric suffix, not used in this command. Implemented for compatibility with 
DISPlay:CMAP<Element>:RGB.  
<Boolean> 
ON – All markers have the same color, to be defined via DISPlay:CMAP6:RGB 
<red>, <green>, <blue>. The marker color is independent of the trace 
colors.  
OFF - Each marker has the color of the associated trace.  
*RST value 
*RST does not affect the color settings; see also description of the Preset 
command.  
SCPI, 
Command 
Types 
Device-specific, command or query   
Example: 
See DISPlay:CMAP<Element>:RGB. 
DISPlay:CMAP<Element>:RGB <red>, <green>, <blue> [,<trace_style>, 
<trace_width>]  
Defines the color of all display elements based on the Red/Green/Blue color model.   
<Element> 
Number of the display element. The display elements corresponding to the 
numbers 1 to 28 are listed below.  
<red> 
<green> 
<blue> 
Red, green, and blue content of the defined color.   
Range [def. unit] 
0(zero intensity, corresponding to a 0 in the 24-bit color model) to 1 (full 
intensity, corresponding to 255 in the 24-bit color model) [–]  
<trace_style> 
Optional trace style, only for traces (<Element> > 12): One of the string 
XDoc.Tiff for .NET, Comprehensive .NET Tiff Imaging Features
types, including EXIF tags, IIM (IPTC), XMP data, and to read, write, delete, and update Tiff file metadata. Render and output text to text, PDF, or Word file.
edit multiple pdf metadata; pdf metadata extract
C# Raster - Raster Conversion & Rendering in C#.NET
RasterEdge XImage.Raster conversion toolkit for C#.NET supports image conversion between various images, like Jpeg, Png, Bmp, Xmp and Gif, .NET Graphics
edit pdf metadata acrobat; remove pdf metadata online
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
428
parameters SOLid | DASHed | DOTTed | DDOTted | DDDotted.  
<trace_width> 
Optional trace width, only for traces (<Element> > 12). 
Range [def. unit] 
1to 20 [–]  
*RST values 
*RST does not affect the color settings; see also description of the Preset 
command.  
SCPI, Command 
Types 
Confirmed (with device-specific numeric suffix and parameters), command or 
query. The query returns three values between 0 and one, separated by 
commas, corresponding to the red, green, and blue color content.   
Example: 
*RST; DISP:CMAP:MARK ON; CALC:MARK ON  
Create diagram area no. 1 (with default trace showing the S-parameter S
21
)and 
amarker Mkr 1.
CALC:PAR:SDEF 'TRC2', 'S11'; DISP:WIND:TRAC2:FEED 'TRC2'  
Create a new trace named TRC2 and display the trace in diagram area no. 1. 
Note that the new trace automatically becomes the active trace.  
CALC:MARK2 ON 
Assign a marker Mkr 2 to the trace. Both markers are displayed with the same 
color.  
DISP:CMAP13:RGB 1,0,0; DISP:CMAP14:RGB 0,1,0 
Color the first trace red, the second trace green.  
DISP:CMAP6:RGB? 
Query the marker color. The marker color depends on the settings made in 
previous sessions; it is not reset. A possible response is 0,0,0 for black markers.  
DISP:CMAP:MARK OFF 
Change the marker colors: Mkr 1 turns red, Mkr 2 turns green.  
The numeric suffixes <Element> denote the following display elements:  
<Element> 
Display Element 
1
Background 
2
Text 
3
Selected Text 
4
Grid 
5
Reference Line 
6
Same Color for all Markers 
7
Horizontal Line / Vertical Range Lines 
8
Diagram Title 
C# Raster - Image Process in C#.NET
Image Access and Modify. Image Information. Metadata(tag) Edit. Color VB.NET How-to, VB.NET PDF, VB.NET Word process various images, like Jpeg, Png, Bmp, Xmp
c# read pdf metadata; edit pdf metadata online
.NET JPEG 2000 SDK | Encode & Decode JPEG 2000 Images
Able to customize compression ratios (0 - 100); Support metadata encoding and decoding, including IPTC, XMP, XML Box, UUID Boxes, etc.
pdf remove metadata; read pdf metadata java
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
429
9
Limit Fail Trace Color 
10 
Limit Line Type Off 
11 
Limit Line Type Upper 
12 
Limit Line Type Lower 
13 
Trace 1 
14 
Trace 2 
15 
Trace 3 
16 
Trace 4 
17 
Trace 5 
18 
Trace 6 
19 
Trace 7 
20 
Trace 8 
21 
Trace 9 
22 
Trace 10 
23 
Trace 11 
24 
Trace 12 
25 
Trace 13 
26 
Trace 14 
27 
Trace 15 
28 
Trace 16 
DISPlay:CMAP<Element>:TRACe:COLor[:STATe] <Boolean>  
Defines the trace color schemes in different diagram areas.   
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
430
<Element> 
Numeric suffix, not used in this command. Implemented for compatibility with 
DISPlay:CMAP<Element>:RGB.  
<Boolean> 
OFF – Independent color scheme in new diagram area. Moved traces change 
their color.  
ON – Color scheme in new diagram area continues the previous color scheme. 
Moved traces keep their color.  
*RST value 
*RST does not affect the color settings; see also description of the Preset 
command.  
SCPI, Command 
Types 
Device-specific, command or query   
Example: 
*RST; DISP:CMAP13:RGB 1,0,0  
Create diagram area no. 1 (with default trace showing the S-parameter S
21
)and 
color the trace red.  
DISP:CMAP:TRAC:COL OFF; DISP:WIND2:STAT ON  
Select independent color schemes for new diagram areas. Create a new 
diagram area no. 2.  
CALC:PAR:SDEF 'TRC2', 'S11'; DISP:WIND2:TRAC2:FEED 'TRC2'  
Create a new trace named TRC2 and display the trace in a new diagram area 
no. 2. The new trace is red like the first trace.  
DISP:CMAP:TRAC:COL ON; DISP:WIND3:STAT ON  
Continue the same color scheme in new diagram areas. Create a new diagram 
area no. 3.  
CALC:PAR:SDEF 'TRC3', 'S22'; DISP:WIND3:TRAC3:FEED 'TRC3'  
Create a new trace named TRC3 and display the trace in a new diagram area 
no. 3. The new trace is not red.  
DISPlay:MENU:KEY:EXECute '<menu_key>'  
Executes the function of a key with a specified name and switches to the local screen.  
You can use this command to execute part of your measurement task manually; see Combining 
Manual and Remote Control. Menu keys that initiate events are executed immediately, because no 
additional input is necessary. However, DISPlay:MENU:KEY:EXECute can be used as well for manual 
entries in numeric entry bars, dialogs or wizards.   
'<menu_key>' 
Name of the key as shown in the analyzer's softkey bar (case-sensitive string 
variable, may contain blanks but no dots). If the name is not unique, the complete 
menu path can be specified, e.g. 'Trace:Format:Phase'. Menus, submenus and 
menu function must be separated by colons.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query   
Example: 
*RST; DISP:MENU:KEY:EXEC 'S11'; EXEC 'Trace:Format:Phase'  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
431
Assign the S-parameter S
11
to the default trace, open the local analyzer screen, 
then display the phase of the measured quantity. The commands are executed 
immediately without any manual entry.  
DISP:MENU:KEY:EXEC 'Start'  
Open the numeric input bar for the start frequency of the sweep. The frequency 
can be entered manually.  
DISP:MENU:KEY:EXEC 'About Nwa'  
Display information about your network analyzer and the firmware version. The 
info box is closed when you click the analyzer screen or if you send another 
command.  
DISPlay:MENU:KEY:SELect '<menu_key>'  
Activates the menu or submenu of the specified key.  
'<menu_key>' 
Name of the key as shown in the analyzer's softkey bar (case-sensitive 
string variable).  
*RST value 
SCPI, Command 
Types 
Device-specific, no query   
Example: 
*RST; DISP:MENU:KEY:SEL 'S11'  
Open the Trace – Meas menu in order to select a measured quantity.  
DISPlay:PSAVe:HOLDoff <holdoff time> 
Sets the holdoff time for the powersave mode of the display.  
<holdoff time> 
Holdoff time for the powersave mode. 
Range [def. unit] 
1to 60 minutes, the resolution is 1 minute. The entry is dimensionless.
*RST value 
15 
SCPI, Command Types 
Device-specific, command or query 
Example: 
*RST; DISP:PSAV:HOLD 30  
Sets the holdoff time to 30 minutes.  
DISPlay:PSAVe[:STATe] <Boolean> 
This command switches on or off the powersave mode of the display. With the powersave mode activated 
the display including backlight is completely switched off after the elapse of the response time (see 
DISPlay:PSAVe:HOLDoff command). 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
432
This mode is recommended for preserving the display especially if the instrument is exclusively 
operated via remote control. 
<Boolean> 
ON | OFF - Switching the power-save mode on or off.
*RST value 
OFF 
SCPI, Command Types 
Device-specific, command or query   
Example: 
*RST; DISP:PSAVe ON 
Switches on the powersave mode.  
DISPlay[:WINDow<Wnd>]:MAXimize <Boolean>  
Maximizes all diagram areas in the active setup or restores the previous display configuration.  
<Wnd> 
Number of the diagram area to become the active diagram area. 
DISPlay:WINDow<Wnd>:MAXimize acts on all diagrams of the current setup, 
however, the diagram no. <Wnd> is displayed on top of the others.  
<Boolean> 
ON | OFF - Maximize all diagram areas or restore the previous display 
configuration.  
*RST value 
OFF 
SCPI, 
Command 
Types 
Device-specific, command or query (returns whether or not the diagrams are 
maximized).  
Example: 
*RST; DISP:WIND2:STAT ON  
Create diagram areas no. 1 (with default trace) and 2 (with no trace).  
DISP:WIND2:MAXimize ON 
Maximize the diagram areas, placing area no. 2 on top.  
DISPlay[:WINDow<Wnd>]:STATe <Boolean>  
Creates or deletes a diagram area, identified by its area number <Wnd>.  
<Wnd> 
Number of the diagram area to be created or deleted.  
<Boolean> 
ON | OFF - Creates or deletes diagram area no. <Wnd>.  
*RST value 
SCPI, Command 
Types 
Device-specific, command or query (returns whether or not a particular 
diagram area exists).  
Example: 
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'  
Create channel 4 and a trace named Ch4Tr1 to measure the input reflection 
coefficient S11.  
DISP:WIND2:STAT ON  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
433
Create diagram area no. 2.  
DISP:WIND2:TRAC9:FEED 'CH4TR1' 
Display the generated trace in diagram area no. 2, assigning the trace 
number 9 to it.  
DISPlay[:WINDow<Wnd>]:TITLe:DATA '<string>'  
Defines a title for diagram area <Wnd>.  
<Wnd> 
Number of the diagram area.  
'<string>'
String variable for the title. The length of the title is practically unlimited but 
should be kept short enough to be displayed in the diagrams.  
*RST value 
'' 
SCPI, Command 
Types 
Device-specific, command or query  
Example: 
*RST; :DISP:WIND:TITL:DATA 'S21 Test Diagram'  
Define a title for the default diagram area. The title is displayed below the top of 
the diagram area.   
DISP:WIND:TITL OFF; TITL:DATA?  
Hide the title. The title is no longer displayed but still defined so it can be 
displayed again.  
DISPlay[:WINDow<Wnd>]:TITLe[:STATe] <Boolean>  
Displays or hides the title for area number <Wnd>, defined by means of 
DISPlay:WINDow<Wnd>:TITLe:DATA.
<Wnd> 
Number of the diagram area.  
<Boolean> 
ON | OFF - Displays or hides the title. 
*RST value 
ON 
SCPI, Command Types 
Device-specific, command or query  
Example: 
See DISPlay:WINDow<Wnd>:TITLe:DATA. 
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:DELete  
Releases the assignment between a trace and a diagram area, as defined by means of 
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED <Trace_Name> and expressed by the <WndTr> 
suffix. The trace itself is not deleted; this must be done via CALCulate<Ch>:PARameter:DELete 
<Trace_Name>.
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
434
<Wnd> 
Number of an existing diagram area (defined by means of 
DISPlay:WINDow<Wnd>:STATe ON).  
<WndTr> 
Trace number used to distinguish the traces of the same diagram area <Wnd>.  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'  
Create channel 4 and a trace named Ch4Tr1 to measure the input reflection 
coefficient S11.  
DISP:WIND2:STAT ON  
Create diagram area no. 2.  
DISP:WIND2:TRAC9:FEED 'CH4TR1'  
Display the generated trace in diagram area no. 2, assigning the trace number 9 
to it. 
DISP:WIND2:TRAC9:DELete  
Release the assignment between trace no. 9 and window no. 2. The trace can 
still be referenced with its trace name Ch4Tr1. 
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:EFEed '<trace_name>' 
Assigns an existing trace (CALCulate<Ch>:PARameter:SDEFine <Trace_Name>) to a diagram area 
<Wnd>, and displays the trace. Use DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED to assign the 
trace to a diagram area using a numeric suffix (e.g. in order to use the 
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet command).  
You can open the trace manager (DISPlay:MENU:KEY:EXECute 'Trace Manager') to obtain an 
overview of all channels and traces, including the traces that are not displayed.   
<Wnd> 
Number of an existing diagram area (defined by means of 
DISPlay:WINDow<Wnd>:STATe ON).  
<WndTr> 
Trace number. This suffix is ignored; the trace is referenced by its name.  
'<trace_name>' 
String parameter for the trace name, e.g. 'Trc4'. 
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'  
Create channel 4 and a trace named Ch4Tr1 to measure the input reflection 
coefficient S11.  
DISP:WIND2:STAT ON  
Create diagram area no. 2.  
DISP:WIND2:TRAC:EFE 'CH4TR1'  
Display the generated trace in diagram area no. 2. No trace number is 
Documents you may be interested