R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
455
SCPI, Command 
Types 
Confirmed (with device-specific response), query only.  
Example: 
*RST; MEM:DEF 'SETUP_2' 
Create a setup named 'Setup_2' and make it the active 
setup. 
MEM:CAT? 
Query all setups. The response is 'Set1,Setup_2'.  
M
MEM:STOR:STAT 1, 'C:\R_S\Instr\user\Nwa\RecallSets\Setup_2.nwa'; 
MEM:DEL 'Setup_2.nwa'
Store the active setup Setup_2 to a file, renaming it Setup_2.nwa. Close the 
setup.  
MEMory:DEFine '<setup_name>' 
Creates a new setup <setup_name> using default settings for the traces, channels and diagram areas. 
The created setup becomes the active setup.  
'<setup_name>' 
String parameter to specify the name of the created setup. 
*RST value 
SCPI, Command Types 
Device-specific, no query.  
Example: 
See MEM:CAT?.
MEMory:DELete[:NAME] '<file_name>'  
Closes the specified setup.  
'<setup_name>' 
String parameter to specify the name of the setup to be closed. 
*RST value 
SCPI, Command Types 
Confirmed, no query.   
Example: 
See MEM:CAT?.
MEMory:SELect '<setup_name>' 
Selects a setup as the active setup.  
'<setup_name>' 
String parameter to specify the setup.  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.   
Example: 
*RST; MEM:DEF 'SETUP_2' 
Add metadata to pdf - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
remove pdf metadata online; embed metadata in pdf
Add metadata to pdf - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
pdf metadata online; batch pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
456
Create a setup named 'Setup_2' and make it the active 
setup. 
MEM:SEL 'Set1' 
Activate the default setup Set1.  
M
MEM:STOR:STAT 1, 'C:\R_S\Instr\user\Nwa\RecallSets\Set1.nwa'; MEM:DEL 
'Set1.nwa'
Store the active setup Set1 to a file, renaming it Set1.nwa. Close the setup.  
MMEMory...     
The MMEMory system provides mass storage capabilities for the analyzer.   
Internal and external mass storage 
The mass storage of the analyzer may be internal or external. The internal mass storage device can be 
any section of the internal hard disk (mapped to drive C:\). The external mass storage device can be a 
floppy disk inserted into the drive at the front panel of the instrument, which is mapped to drive a:\ (see 
MMEMory:MSIS), a USB memory stick connected to one of the USB ports (mapped to any free drive 
letter) or a network connection.  
File and directory names 
The <file_name> and <directory_name> parameters are strings. Some commands use a fixed directory; 
for others the <file_name> can contain the complete path including the drive name and all subdirectories, 
e.g. 'C:\TEMP\TRASH\test.txt' for the file named test.txt in the TEMP\TRASH subdirectory of the internal 
hard disk drive C:\. If no complete path is specified, the file location is relative to the current directory 
(queried with MMEMory:CDIRectory?). The file name itself may contain the period as a separator for 
extensions.  
File and directory names can be chosen according to Windows™ conventions; the restrictions placed on 
file names known from DOS systems do not apply. All letters and numbers are allowed, as well as the 
special characters "_", "^", "$", "~", "!", "#", "%", "&", "-", "{", "}", "(", ")", "@" and "`". Reserved file names 
are CON, AUX, COM1, ..., COM4, LPT1, ..., LPT3, NUL and PRN.  
The use of wildcards ? and * is not allowed.   
MMEMory:AKAL:FACTory:CONVersion '<directory_name>' 
Converts the calibration data of the standards in the active calibration unit 
(SYSTem:COMMunicate:RDEVice:AKAL:ADDRess) to Touchstone format and copies it to the specified 
directory.  
'<directory_name>' 
String parameter to specify the directory.  
*RST value 
SCPI, Command 
Types 
Device-specific, command or query (returns the current directory).  
Example: 
MMEM:AKAL:FACTory:CONVersion 'C:\AKAL\Touchstone'
Convert and copy the calibration data of the standards 
VB.NET PDF insert image library: insert images into PDF in vb.net
VB.NET PDF - Add Image to PDF Page in VB.NET. Guide VB.NET Programmers How to Add Images in PDF Document Using XDoc.PDF SDK for VB.NET.
get pdf metadata; pdf metadata editor online
C# PDF Password Library: add, remove, edit PDF file password in C#
C# Sample Code: Add Password to PDF with Permission Settings Applied in C#.NET. This example shows how to add PDF file password with access permission setting.
clean pdf metadata; google search pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
457
to the specified directory.  
MMEMory:CATalog? '<directory_name>'  
Returns the contents of a specified directory.  
'<directory_name>' 
String parameter to specify the directory.  
Response 
Comma-separated list of all file names in the directory (string parameters). 
*RST value 
SCPI, Command Types 
Confirmed, query only.  
Example: 
MMEM:CAT? 'C:\R_S\Instr\user'  
Response: 
'autolog.txt','autologin.reg','no_autologin.reg'  
MMEMory:CDIRectory '<directory_name>' 
Changes the default directory for mass memory storage.  
'<directory_name>' 
String parameter to specify the directory.  
*RST value 
–(A *RST does not change the current directory)  
SCPI, Command 
Types 
Confirmed, command or query (without parameter, returns the current 
directory).  
Example: 
MMEM:CDIR 'C:\Documents and 
Settings\NetworkService\Application Data'
Change to the specified directory. 
MMEMory:CLEar:ALL 
Deletes all device settings in the current directory. The current directory can be selected with the 
MMEMory:CDIRectory command. The default directory is C:\R_S\instr\user\Nwa. 
*RST value 
SCPI, Command Types 
Device-specific, no query.  
Example: 
MMEM:CLE:ALL 
Delete all device settings files.
MMEMory:CLEar:STATe 1,<file_name> 
Deletes the instrument setting selected by <file_name>. All associated files on the mass memory storage 
are cleared. 
C# PDF Sticky Note Library: add, delete, update PDF note in C#.net
C#.NET PDF SDK - Add Sticky Note to PDF Page in C#.NET. Able to add notes to PDF using C# source code in Visual Studio .NET framework.
analyze pdf metadata; search pdf metadata
C# PDF Library SDK to view, edit, convert, process PDF file for C#
PDF SDK for .NET allows you to read, add, edit, update, and delete PDF file metadata, like Title, Subject, Author, Creator, Producer, Keywords, etc.
read pdf metadata online; extract pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
458
<file_name> 
File name of the settings file. The file extension can be omitted. If the file is not 
stored in the C:\R_S\Instr\user directory, the full path has to be specified. 
*RST value 
SCPI, Command 
Types 
Device-specific, no query. 
Example: 
MMEM:CLE:STAT 1,'TEST.ZVL.DFL' 
Delete file TEST.zvl.dfl. 
MMEMory:COPY '<file_source>','<file_destination>' 
Copies an existing file to a new file or directory.  
'<file_source>', 
'<file_destination>' 
String parameters to specify the name of the file to be copied and the 
name of the new file or directory.  
*RST value 
SCPI, Command Types 
Confirmed, no query.  
Example: 
MMEM:COPY 'C:\USER\DATA\SETUP.ZVL.DFL','E:'
Copy file Setup.zvl.dfl in directory C:\USER\DATA to 
the external storage medium, mapped to drive E:\.  
MMEMory:DATA '<file_name>', <data> 
Loads <data> into the file <file_name>. The delimiter must be set to EOI 
(SYSTem:COMMunicate:GPIB[:SELF]:RTERminator EOI) to obtain error–free data transfer.  
'<file_name>' 
String parameter to specify the name of the file.  
<data> 
Data in 488.2 block data format. The delimiter EOI must be selected to achieve 
correct data transfer.  
*RST value 
SCPI, Command 
Types 
Confirmed, with query. The query form MMEMory:DATA? <file_name> returns 
the associated <data> in block format (the delimiter setting is irrelevant for the 
query).  
Example: 
MMEM:DATA? 'C:\TEMP\TEST01.HCP' 
Query the block data contained in file TEST01.HCP.  
MMEMory:DELete '<file_name>' 
Removes a file from the specified directory. 
'<file_name>' 
String parameter to specify the name and path of the file to be removed. If the 
VB.NET PDF Library SDK to view, edit, convert, process PDF file
Various of PDF text and images processing features for VB.NET project. Multiple metadata types of PDF file can be easily added and processed. PDF Metadata Edit.
pdf metadata viewer; batch pdf metadata editor
C# PDF insert image Library: insert images into PDF in C#.net, ASP
C#.NET PDF SDK - Add Image to PDF Page in C#.NET. How to Insert & Add Image, Picture or Logo on PDF Page Using C#.NET. Add Image to PDF Page Using C#.NET.
adding metadata to pdf files; remove metadata from pdf file
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
459
path is omitted, the current directory is used (see MMEMory:CDIRectory). 
*RST value 
SCPI, Command 
Types 
Confirmed, no query.   
Example: 
MMEM:DEL 'C:\TEMP\TEST01.HCP'
Remove TEST01.HCP from the directory C:\TEMP.  
MMEMory:DELete:CORRection <file_name>' 
Deletes a system error correction data set stored in the cal pool (cal group file).   
'<file_name>' 
String parameter to specify the name of the cal group file to be deleted. Cal group 
files must have the extension *.cal. The directory path must not be specified; the 
analyzer always uses the default cal pool directory 
C:\R_S\Instr\user\Nwa\Calibration\Data.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
See  MMEMory:LOAD:CORRection 
MMEMory:LOAD:AUTO 1,'FACTORY | <file_name>' 
Defines which device settings are automatically loaded after the device is switched on. The contents of the 
file are read after switching on the device and used to define the new device state. 
The data set defined as auto recall set can also be restored by the *RST command. 
FACTORY 
Deactivates the startup recall function. 
<file_name> 
Activates the startup recall function and defines the device settings to be 
loaded. 
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
MMEM:LOAD:AUTO 1,'C:\R_S\Instr\user\TEST' 
Define the TEST file as startup recall file and activate the function. 
MMEMory:LOAD:CABLe '<source_directory>' 
Loads all cable description files (*.rsc) from a specified directory. The default source directories for cable 
description files (C:\R_S\Instr\resources\Nwa\CableTypes for predefined, 
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
You can also update, remove, and add metadata. List<EXIFField> exifMetadata = collection.ExifFields; You can also update, remove, and add metadata.
endnote pdf metadata; pdf metadata extract
VB.NET PDF Password Library: add, remove, edit PDF file password
VB: Add Password to PDF with Permission Settings Applied. This VB.NET example shows how to add PDF file password with access permission setting.
pdf metadata editor; pdf remove metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
460
C:\R_S\Instr\user\Nwa\CableTypes for user-defined cables) are not affected; see example.  
'<source_directory>' 
String parameter to specify the source directory. 
*RST value 
SCPI, Command Types 
Device-specific, no query.   
Example: 
See MMEMory:STORe:CABLe 
MMEMory:LOAD:CKIT '<file_name>' 
Loads cal kit data from a specified NWA cal kit file.  
'<file_name>' 
String parameter to specify the name and directory of the cal kit file to be loaded. If 
no path is specified the analyzer searches the current directory, to be queried with 
MMEMory:CDIRectory?.
Note: The loaded file must be a NWA-specific cal kit file with the extension *.calkit. 
ZVR cal kit files can be imported using the 
[SENSe<Ch>:]CORRection:CKIT:INSTall command. Agilent cal kit 
files can be imported manually and converted into *.calkit files.   
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
MMEM:LOAD:CKIT 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit' 
Load the previously created cal kit file New_kit.calkit from 
the default cal kit directory.  
... :MMEM:STOR:CKIT 'New_kit', 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit' 
Store the data for the user-defined cal kit Newkit and 
overwrite the cal kit file New_kit.calkit.  
MMEMory:LOAD:CKIT:SDATa '<conn_name>', '<ckit_name>', MMTHrough | 
MFTHrough | FFTHrough | MMLine | MFLine | FFLine | MMATten | MFATten | 
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen | FOPen | MSHort | 
FSHort | MOSHort | FOSHort | MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | 
FSMatch, '<stdlabel_name>', '<file_name>' [,<port1_no>][,<port2_no>] 
Loads cal kit data for a specific connector type, cal kit, and calibration standard from a specified 
Touchstone file, assigning a label for the cal data. A restriction on the port assignment may be defined in 
addition.  
'<conn_name>' 
String parameters containing the name of the connector type.  
'<ckit_name>' 
String parameters containing the name of a calibration kit available on the 
analyzer.  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
461
Parameters 
Standard type; for a description refer to the table of standard types. 
'<stdlabel_name>' 
String parameter defining a label for the cal kit data. An empty string means 
that no label is defined. 
'<file_name>' 
String parameter to specify the name and directory of the Touchstone file to be 
loaded. A *.s1p file must be used for one-port standards, a *.s2p file for two-
port standards. If no path is specified the analyzer searches the current 
directory, to be queried with MMEMory:CDIRectory?.
<port1_no>, 
<port2_no> 
Port assignment: One port number for one-port standards, two port number for 
two-port standards. If the port numbers are omitted, the cal kit data is valid for 
all ports.  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.   
Example: 
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default Kit',MOPEN,'Test 
data','test.s1p',1 
Load the file Test.s1p from the current directory in order 
to define the properties of an Open (m) standard in the 
cal kit named Default Kit for the N 50 R connector type.
Assign the label Test data and specify that the data 
standard data is only valid for port no. 1.   
Connector and cal kit naming conventions 
Connector and calibration kit names must be entered as string parameters. The strings contain the 
connector and cal kit names used in the Calibration Kits dialog; a V in the name must be replaced by 
'Ohm', e.g.:  
'NewKit1' denotes the user-defined calibration kit NewKit1.
'N 50 Ohm Ideal Kit' denotes the N 50 V Dummy Kit.
'ZV-Z21 typical' denotes the cal kit ZV-Z21 typical.
MMEMory:LOAD:CMAP '<file_name>' 
Loads a color scheme from a specified NWA color scheme file.  
'<file_name>' 
String parameter to specify the name and directory of the cal kit file to be loaded. 
The default extension (manual control) for color scheme files is *.ColorScheme, 
although other extensions are allowed.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
MMEM:LOAD:CMAP 
'C:\R_S\Instr\user\Nwa\ColorSchemes\Test.ColorScheme' 
Load the previously created color scheme file 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
462
Test.ColorScheme from the default color scheme directory.  
D
ISP:CMAP13:RGB 1,0,0; DISP:CMAP14:RGB 0,1,0  
Color the first trace red, the second trace green.  
MMEM:STOR:CMAP 
'C:\R_S\Instr\user\Nwa\ColorSchemes\Test.ColorScheme' 
Store the data for the user-defined cal kit Newkit and 
overwrite the cal kit file New_kit.calkit.  
MMEMory:LOAD:CORRection <Ch>,'<file_name>' 
Applies a system error correction data set stored in the cal pool (cal group file) to channel no. <Ch>.   
<Ch> 
Channel number of an existing channel 
<file_name> 
String parameter to specify the name of the cal group file to be loaded. Cal group 
files must have the extension *.cal. The directory path must not be specified; the 
analyzer always uses the default cal pool directory 
C:\R_S\Instr\user\Nwa\Calibration\Data.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
MMEM:STOR:CORR 1,'Calgroup1.cal' 
Copy the current correction data set of channel 1 to a cal 
group file Calgroup1.cal. 
CONF:CHAN2:STAT ON; :MMEM:LOAD:CORR 2,'Calgroup1.cal' 
Apply the stored correction data to channel 2.  
MMEM:LOAD:CORR:RES 2,'Calgroup1.cal' 
Undo the previous action: Resolve the link, causing channel 2 
to use its previous correction data. 
MMEM:DEL:CORR 'Calgroup1.cal' 
Delete the created cal group file. 
MMEMory:LOAD:CORRection:RESolve <Ch>[,'<file_name>'] 
Resolves the link between channel <Ch> and the correction data set (cal group file) so that the channel 
uses its previous correction data.  
<Ch> 
Channel number 
'<file_name>' 
Optional string parameter to specify the name of the cal group file. Cal group files 
must have the extension *.cal. The directory path must not be specified.
If there is no link between <Ch> and the specified file, the command has no 
effect. If no file is specified, the command resolves any link between <Ch> and 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
463
an arbitrary cal group file.   
*RST value 
SCPI, Command 
Types 
Device-specific, no query.   
Example: 
See MMEMory:LOAD:CORRection 
MMEMory:LOAD:LIMit '<trc_name>','<file_name>'[,'<param_name>', <x_offset>, 
<y_offset>, <type>] 
Loads a limit line definition from a specified file and assigns it to a trace with a specified name. Limit lines 
are created using the CALCulate<Chn>:LIMit... commands.  
Limit lines can be loaded from Touchstone files (*.s<n>p, where <n> denotes the number of ports). 
The optional parameters '<param_name>', <x_offset>, <y_offset>, <type> are only relevant for 
Touchstone files. For *.limit files, no optional parameters can be set.   
'<trc_name>' 
Name of an existing trace in the active setup (string parameter). The imported 
limit line is assigned to this trace, irrespective of the trace information in the limit 
line file.  
*RST value 
'<file_name>' 
String parameter to specify the name and directory of the limit line file to be 
loaded. The default extension (manual control) for limit line files is *.limit, 
although other extensions are allowed. If no path is specified the analyzer 
searches the current directory, to be queried with MMEMory:CDIRectory?. See 
also note on Touchstone files above.  
*RST value 
'<param_name>' 
String parameter, selects an S-parameter from a Touchstone file. The parameter 
must be compatible with the file type (e.g. for one-port Touchstone files *.s1p, 
only the parameter name 'S11' is allowed).  
*RST value 
'S11' (if all optional parameters are omitted) 
<x_offset> 
Stimulus offset for limit lines loaded from a Touchstone file. A 1 GHz offset shifts 
the limit line by 1 GHz in (positive) horizontal direction.    
Range [def. unit] 
Depending on the sweep range of the analyzer. [Hz]   
*RST value 
0
<y_offset> 
Response offset for limit lines loaded from a Touchstone file. A 1 dB offset shifts 
the limit line by 1 dB in (positive) vertical direction.   
Range [def. unit] 
Depending on the measured quantity. [dB]  
*RST value 
0
<type> 
Limit line type : 
LMAX – upper limit line 
LMIN – lower limit line 
OFF – limit line off 
*RST value 
LMAX (if all optional parameters are omitted) 
SCPI, Command 
Types 
Device-specific, no query.   
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
464
Example: 
Assume that the current setup contains two traces named 
Trc1 and Trc2, respectively, and that limit lines have been 
defined for Trc1.  
MMEM:STOR:LIM 'TRC1', 
'C:\R_S\Instr\user\Nwa\LIMitLines\Lim_Trc1.limit' 
Store the limit line definition of Trc1 to a limit line 
file.  
MMEM:LOAD:LIM 'TRC2', 
'C:\R_S\Instr\user\Nwa\LIMitLines\Lim_Trc1.limit' 
Load the previously created limit line file and assign the 
limit lines to Trc2.  
MMEM:STOR:TRAC 'TRC1', 
'C:\R_S\Instr\user\Nwa\LIMitLines\Trc1.s1p' 
Store the current trace data of Trc1 to a limit line file 
in Touchstone format.  
MMEM:LOAD:LIM 'TRC1', 
'C:\R_S\Instr\user\Nwa\LIMitLines\Trc1.s1p', 'S11', 0, 2, 
LMAX 
Load the previously created Touchstone limit line file and 
assign the limit lines to Trc1, applying a response offset 
of 2 dB.  
CALC:LIMit:DISPlay ON 
Show the limit line in the diagram.  
MMEMory:LOAD:NWANalyzer <numeric_value>,'<file_name>' 
Loads configuration data from a specified setup file and sets the analyzer to the corresponding instrument 
state. 
<numeric_value> 
1(the <numeric_value> is used for compatibility with the SCPI standard but is 
ignored).  
<file_name> 
String parameter to specify the name and directory of the setup file to be 
loaded. The default extension (manual control) for setup files is *.nwa, although 
other extensions are allowed. If no path is specified the analyzer searches the 
current directory, to be queried with MMEMory:CDIRectory?.
*RST value 
SCPI, Command 
Types 
Confirmed, no query.   
Example: 
MMEM:STOR:NWAN 1,'C:\R_S\Instr\user\Nwa\Setup_0413.nwa' 
Store the current setup configuration in the file 
Setup_0413.NWA in the default directory for setup files.  
MMEM:LOAD:NWAN 1,'C:\R_S\Instr\user\Nwa\Setup_0413.nwa' 