R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
485
steps for each decade. The analyzer rounds up any entered value between these 
steps and rounds down values exceeding the maximum bandwidth.   
*RST value 
10 kHz 
SCPI, 
Command 
Types 
Confirmed, command or query 
Example: 
BAND 1.1 
Set a resolution bandwidth of approx. 1.1 Hz for channel 1.  
BAND? 
The analyzer returns the rounded bandwidth of 2 Hz. 
[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution]:SELect FAST | NORMal 
Defines the selectivity of the IF filter for an unsegmented sweep. The value is also used for all segments 
of a segmented sweep, provided that separate selectivity setting is disabled 
([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol OFF).  
<Ch> 
Channel number.  
FAST 
Use IF filter short settling time.  
NORMal 
Use IF filter with higher selectivity but larger settling time. 
*RST value 
FAST  
SCPI, Command Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol 
[SENSe<Ch>:]CORRection...     
This subsystem controls system error correction and recording of correction data.   
[SENSe<Ch>:]CORRection:CDATa 'DIRECTIVITY' | 'SRCMATCH' | 'REFLTRACK' | 
'ISOLATION' | 'LOADMATCH' | 'TRANSTRACK', <port1_no>, <port2_no> 
Writes or reads system error correction data for a specific channel <Ch>, calibration method 
([SENSe<Ch>:]CORRection:COLLect:METHod:DEFine), and port combination <Port1_no>, <Port2_no>. 
The setting command can be used to transfer user-defined correction data to the analyzer; the query 
returns the current correction data set. ASCII or block data can be transferred, depending on the selected 
data transfer format (FORMat[:DATA]). 
The sweep must be stopped to transfer calibration data; see program example for 
[SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault.
For an overview of calibration methods and error terms refer to section Calibration Types.  
<Ch> 
Channel number of the calibrated channel. If unspecified the numeric suffix is set to 
Read pdf metadata java - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
add metadata to pdf file; remove metadata from pdf file
Read pdf metadata java - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
endnote pdf metadata; pdf remove metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
486
1. 
Error term 
parameters 
String parameters describing the different error terms, depending on the current 
calibration method; see table below. Each term contains one complex value (real and 
imaginary part) for each sweep point.  
Range [def. 
unit] 
The error terms consist of dimensionless complex numbers. The parameters must be 
transferred in full length and have the following meaning: 
'DIRECTIVITY' – Directivity at port <port1_no>  
'SRCMATCH' – Source match at port <port1_no> 
'REFLTRACK' – Reflection tracking at port <port1_no> 
'ISOLATION' – Isolation between port <port1_no> and <port2_no> 
'LOADMATCH' – Load match at <port2_no> 
'TRANSTRACK' – Transmission tracking between port <port1_no> and <port2_no> 
*RST value 
The analyzer provides a default calibration corresponding to a test setup which does 
not introduce any systematic errors; see 
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault.
<port1_no> 
Source port number.   
<port2_no> 
Load port number. If the error term is not related to the load port, a dummy number 
can be used; e.g. CORR:CDAT 'REFLTRACK', 1, 0 
SCPI, 
Command 
Types 
Device-specific, command or query  
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault.   
The different calibration types of the analyzer provide the following error terms:  
Calibration type 
Parameters in 
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine 
Available error terms (depending 
on port numbers) 
One-port 
normalization 
(reflection)  
using an open or a 
short standard 
REFL, RSHort 
'REFLTRACK' 
Full one port 
FOPort  
'DIRECTIVITY', 'SRCMATCH', 
'REFLTRACK' 
Two-port 
normalization 
FRTRans 
'TRANSTRACK' 
One path two port 
OPTPort 
'DIRECTIVITY', 'SRCMATCH', 
'REFLTRACK', 'TRANSTRACK' 
TOSM 
TOSM 
'DIRECTIVITY', 'SRCMATCH', 
'REFLTRACK', 'LOADMATCH', 
'TRANSTRACK'  
(at present the 'ISOLATION' terms are 
not included) 
[SENSe<Ch>:]CORRection:CKIT...     
This subsystem controls system error correction and recording of correction data.   
[SENSe<Ch>:]CORRection:CKIT:<std_type> 
.NET PDF SDK | Read & Processing PDF files
of text, hyperlinks, bookmarks and metadata; Advanced document provides royalty-free .NET Imaging PDF Reader SDK Document Imaging SDK and Java Document Imaging
view pdf metadata; pdf metadata reader
DocImage SDK for .NET: Document Imaging Features
viewer application Enable users to add metadata in the TIFF Type 6 (OJPEG) encoding Image only PDF encoding support. Add-on: Able to decode and read 20+ barcode
batch edit pdf metadata; remove pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
487
'<Conn_Name>','<Ckit_Name>','<Std_No>',<Min_Freq>,<Max_Freq>,<El_Length>,<
Loss>,<Z0>,<C0> , <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt | 
MATCh, <Resistance>, <Port_1>, <Port_2>] 
Defines the parameters of a non-ideal 1 port or 2-port calibration standard <std_type>. A particular 
physical standard can be selected by specifying the name of the calibration kit and its serial number. 
Depending on the standard type, only a subset of the parameters may be used; see table below.  
<Ch> 
Channel number. This suffix is ignored because calibration kits are channel-
independent. 
<std_type> 
Standard type. For reflection standards, the first character denotes the gender, e.g.: 
FOPEN, MOPEN: Open (f) or Open (m) standard. 
The following reflection standards are supported: MOPen, FOPen, MSHort, 
FSHort, OSHort, MOSHort, FOSHort, MMTCh, FFTCh, MREFLect, 
FREFLect 
For transmission standards, the first two characters denotes the genders on both 
ends, e.g.: 
FFSNetwork, MFSNetwork, MMSNetwork: Symm. network (ff), symm. network 
(mf) or symm. network (mm) standard.  
For a complete list of standard types refer to the table of standard types below.  
Parameter 
list 
String parameters to specify the configured standard 
('<Conn_Name>','<Ckit_Name>','<Std_No>') and numeric parameters defining its 
properties. See detailed description below.  
The following transmission standards have only 8 parameters (until <Z0>): 
MMTHrough, MFTHrough, FFTHrough, MMLine, MFLine, FFLine 
The following reflection and transmission standards have only 5 parameters 
(until <Max_Freq>): 
MSMatch, FSMatch, MMATten, MFATten, FFATten  
*RST value 
SCPI, 
Command 
Types 
Device-specific, command or query (see examples below)  
Example: 
CORR:CKIT:FOP  'N 50 Ohm','Test Kit',...  
Define the properties of the open (f) standard for a N 50 V connector type in cal kit 
'Test Kit'. See also example for 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>.
CORR:CKIT:FOP?  'N 50 Ohm'  
Query the properties of the open (f) standard for a N 50 V connector type in the 
active cal kit.   
CORR:CKIT:FOP?  'N 50 Ohm','Test Kit'  
Query the properties of the open (f) standard for a N 50 V connector type in cal kit 
'Test Kit'.  
PDF Image Viewer| What is PDF
Promote use of metadata; Used on independent devices and processing SDK, allowing developers to read, write, view and convert word document without need for PDF.
edit pdf metadata online; online pdf metadata viewer
.NET DICOM SDK | DICOM Image Viewing
new style; Ability to read patient metadata from the guide and sample codings; Ability to read DICOM images NET Document Imaging SDK and Java Document Imaging
acrobat pdf additional metadata; pdf metadata viewer
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
488
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> 
'<Ckit_Name>','<Std_No>',<Min_Freq>,<Max_Freq>,<El_Length>,<Loss>,<C0> | 
<L0>,<C1> | <L1>,<C2> | <L2>,<C3> | <L3>, OPEN | SHORt  
Defines the parameters of a calibration standard <std_type> for a specified connector type <conn_type>. 
Aparticular physical standard can be selected by specifying the name of the calibration kit and its serial 
number. Depending on the standard type, only a subset of the parameters may be used; see table below.  
<Ch> 
Channel number. This suffix is ignored because calibration kits are channel-
independent. 
<conn_type> 
Connector type, one of the following identifiers: 
N50, N75: N 50 V or N 75 V connectors  
PC7, PC35, PC292: PC 7, PC 3.5 or 2.92 mm connectors 
USER<no>: User-defined connectors UserConn1, UserConn2 
This command only supports ZVR-compatible connector types. For general 
definitions use [SENSe<Ch>:]CORRection:CKIT:<std_type>.
<std_type> 
Standard type. For reflection standards, the first character denotes the gender, e.g.:
FOPEN, MOPEN: Open (f) or Open (m) standard. 
For transmission standards, the first two characters denotes the genders on both 
ends, e.g.: 
FFTHrough, MFTHrough, MMTHrough: Through (ff), Through (mf) or Through 
(mm) standard. 
For a complete list of standard types refer to the table of standard types below.  
Parameter list 
String parameters to specify the configured standard ('<Ckit_Name>','<Std_No>') 
and numeric parameters defining its properties. See detailed description below.  
*RST value 
SCPI, 
Command 
Types 
Device-specific, command or query  
Example: 
CORR:CKIT:N50:FOPEN 'ZV-Z21','',0,1.8E+010,0.0151,0,0,0.22,-0.22,0.0022 
Define the properties of the open (f) standard for the N 50 V connector type 
contained in the ZV-Z21 calibration kit: Assign a valid frequency range of 0 Hz to 18 
GHz, an electrical length of 15.1 mm, 0 dB loss and define the polynomial 
coefficients of the fringing capacitance as 0 fF, 0.22 fF/GHz, –0.22 fF/(GHz)
2
,
0.0022 fF/(GHz)
3
.
The parameters in the [SENSe<Ch>:]CORRection:CKIT:<std_type> and 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type> commands have the following 
meaning: 
Parameter 
Meaning 
Comment/Unit 
'<Conn_Name>' 
Name of the connector type 
String parameter 
'<Ckit_Name>' 
Name of the calibration kit  
String parameter 
'<Std_No>' 
Label (e.g. the serial number) of the standard  
String parameter 
.NET PDF Generator | Generate & Manipulate PDF files
delete any pages in PDFs; Add metadata of a RasterEdge provides royalty-free .NET Imaging PDF Generator of NET Document Imaging SDK and Java Document Imaging
bulk edit pdf metadata; pdf metadata
.NET JPEG 2000 SDK | Encode & Decode JPEG 2000 Images
Able to customize compression ratios (0 - 100); Support metadata encoding and decoding com is professional .NET Document Imaging SDK and Java Document Imaging
batch pdf metadata editor; add metadata to pdf
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
489
<Min_Freq> 
Min. frequency for which the circuit model is valid 
Default unit is Hz 
<Max_Freq> 
Max. frequency for which the circuit model is valid 
Default unit is Hz 
<El_Length> 
Electrical length (offset parameter) of the standard 
Default unit is m or ft, 
depending on the Distance 
Unit setting in the System 
Configuration dialog 
<Loss> 
Loss (offset parameter) of the standard 
To be specified without 
unit (implicit unit is dB) 
<Z0> 
Reference impedance (no unit) 
To be specified without 
unit (implicit unit is V)
<C0> | <L0> 
Polynomial coefficients C
0
or L
0
for the fringing capacitance or residual 
inductance of the standard (load parameter) 
To be specified without 
unit (implicit unit is fF or 
pH) 
<C1> | <L1> 
Polynomial coefficients C
1
or L
1
for the fringing capacitance or residual 
inductance of the standard (load parameter) 
To be specified without 
unit (implicit unit is fF /GHz 
or pH /GHz) 
<C2> | <L2> 
Polynomial coefficients C
2
or L
2
for the fringing capacitance or residual 
inductance of the standard (load parameter) 
To be specified without 
unit (implicit unit is 
fF/GHz
2
or pH/GHz
2
)
<C3> | <L3> 
Polynomial coefficients C
3
or L
3
for the fringing capacitance or residual 
inductance of the standard (load parameter) 
To be specified without 
unit (implicit unit is 
fF/GHz
3
or pH/GHz
3
)
OPEN | SHORt | 
MATCh 
Simplified modelling as an open or short or match standard. The load 
circuit model generally consists of capacitance C which is connected in 
parallel to an inductance L and a resistance R, both connected in series. 
OPEN means that R is infinite so that the standard behaves like a 
capacitor (no inductance, the polynomial coefficients specify C). SHORt 
means that R is zero so that the standard behaves like an inductance (no 
capacitance, the polynomial coefficients specify L). MATCh means that 
the standard behaves like a match with the specified <Resistance>  
Character data 
<Resistance> 
Optional resistance for the MATCh standard  
Numeric value 
,<Port_1>, 
<Port_2> 
Optional port restriction: one port number for one port standards, two port 
numbers for two port standards  
Integer value 
The different standard types are defined by the following parameters:  
<std_type> 
Meaning 
Parameters 
MOPen | FOPen 
Open (m) or open (f)  
'<Ckit_Name>' ... <C3>[, <Port_1>]  
complete parameter list with capacitance 
coefficients, no OPEN | SHORT | MATCh  
MSHort | FSHort 
Short (m) or short (f)  
'<Ckit_Name>' ... <L3>[, <Port_1>] 
complete parameter list with inductance 
.NET Multipage TIFF SDK| Process Multipage TIFF Files
upload to SharePoint and save to PDF documents. Support for metadata reading & writing. com is professional .NET Document Imaging SDK and Java Document Imaging
google search pdf metadata; pdf xmp metadata
.NET Annotation SDK| Annotate, Redact Images
Automatically save annotations as the metadata of the document. Want to Try RasterEdge.com is professional .NET Document Imaging SDK and Java Document Imaging
add metadata to pdf programmatically; read pdf metadata java
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
490
coefficients, no OPEN | SHORT| MATCh 
OSHort | MOSHort | 
FOSHort 
Offset short (sexless) or offset short (m) 
or offset short (f) 
(for user-defined connector types only) 
'<Ckit_Name>' ... <L3>[, <Port_1>] 
complete parameter list with inductance 
coefficients, no OPEN | SHORT| MATCh 
MMTCh | FMTCh 
Match (m) or match (f) 
'<Ckit_Name>' ... <Max_Freq>[, <Port_1>]  
no offset parameters, no polynomial coefficients, 
no OPEN | SHORt| MATCh 
MREFlect | FREFlect 
Reflect (m) or reflect (f) 
'<Ckit_Name>' ... OPEN | SHORt[, <Port_1>]  
no loss, otherwise complete parameter list 
MMTHrough | MFTHrough | 
FFTHrough 
Through (m - m) or through (m - f) or 
through (f - f) 
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>] 
no load parameters (polynomial coefficients), no 
OPEN | SHORt| MATCh 
MMLIne1 | MFLIne1 | 
FFLIne1 (suffix 1 optional) 
Line1 (m - m) or line1 (m - f) or line1 (f - f) 
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>] 
no load parameters (polynomial coefficients), no 
OPEN | SHORt| MATCh 
MMLIne2 | MFLIne2 | 
FFLIne2 
Line1 (m - m) or line2 (m - f) or line2 (f - f) 
'<Ckit_Name>' ... <Loss>[, <Port_1>, <Port_2>] 
no load parameters (polynomial coefficients), no 
OPEN | SHORt| MATCh 
MMATten | MFATten | 
FFATten 
Attenuation (m - m) or attenuation (m - f) 
or attenuation (f - f) 
'<Ckit_Name>' ... <Max_Freq>[, <Port_1>, 
<Port_2>] 
no offset parameters, no load parameters 
(polynomial coefficients), no OPEN | SHORt| 
MATCh 
MMSNetwork | 
MFSNetwork | FFSNetwork 
Symmetric network (m - m) or symmetric 
network (m - f) or symmetric network (f - 
f) 
'<Ckit_Name>' ... OPEN | SHORt| MATCh[, 
<Port_1>, <Port_2>]  
no loss, otherwise complete parameter list 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect '<ckit_name>'  
Selects the calibration kit to be used for a specified connector type <conn_type>.  
For connector types with arbitrary, user-defined names you can use the command 
[SENSe<Ch>:]CORRection:CKIT:SELect '<conn_name>', '<ckit_name>'.  
<Ch> 
Channel number. This suffix is ignored because calibration kits are channel-
independent. 
<conn_type> 
Connector type, one of the following identifiers: 
N50, N75: N 50 V or N 75 V connectors 
PC7, PC35, PC292: PC 7, PC 3.5 or 2.92 mm connectors 
USER<no>: User-defined connectors UserConn1, UserConn2 
'<ckit_name>' 
String parameters containing the name of a calibration kit available on the 
analyzer.  
*RST value 
–(A *RST does not change the assignment between connector types and 
calibration kits.)  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
491
SCPI, 
Command 
Types 
Device-specific, command or query  
Example: 
MMEM:LOAD:CKIT 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit' 
Load the previously created cal kit file New_kit.calkit from 
the default cal kit directory.  
C
ORR:CKIT:N50:SEL 'New_kit'  
Assign the imported kit to the N 50 V connector type (assuming that the cal kit 
name stored in New_kit.calkit reads New_kit).  
Cal kit naming conventions 
Calibration kit names must be entered as string parameters. The string contains the cal kit name used in 
the Calibration Kits dialog; a V in the name must be replaced by 'Ohm', e.g.:  
'NewKit1' denotes the user-defined calibration kit NewKit1.
'N 50 Ohm Ideal Kit' denotes the N 50 V Ideal Kit.
'ZV-Z21 typical' denotes the cal kit ZV-Z21 typical.
[SENSe<Ch>:]CORRection:CKIT:SELect '<conn_name>', '<ckit_name>'  
Selects the calibration kit to be used for a specified connector type <conn_type>.  
The command is suitable for connector types with arbitrary, user-defined names. For standard 
connector types you can use the command 
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect.
<Ch> 
Channel number. This suffix is ignored because calibration kits are channel-
independent. 
'<conn_name>' 
Connector type, e.g. a user-defined connector type (string variable)  
'<ckit_name>' 
String parameters containing the name of a calibration kit available on the 
analyzer.  
*RST value 
–(A *RST does not change the assignment between connector types and 
calibration kits.)  
SCPI, 
Command 
Types 
Device-specific, command or query (the query requires the first string parameter 
only)  
Example: 
MMEM:LOAD:CKIT 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\New_kit.calkit' 
Load the previously created cal kit file New_kit.calkit from 
the default cal kit directory.  
C
ORR:CKIT:SEL 'N 50 Ohm', 'New_kit'  
Assign the imported kit to the N 50 V connector type (assuming that the cal kit 
name stored in New_kit.calkit reads New_kit).  
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
492
[SENSe<Ch>:]CORRection:CKIT:INSTall '<file_name>' 
Loads cal kit data from a specified ZVR cal kit file.  
<Ch> 
Channel number. This suffix is ignored because calibration kits are channel-
independent. 
'<file_name>' 
String parameter to specify the name and directory of the cal kit file to be loaded.  
Note: The loaded file must be a ZVR-specific cal kit file with the extension *.ck. 
NWA cal kit files (*.calkit) can be imported using the MMEMory:LOAD:CKIT 
command. Agilent cal kit files can be imported manually and converted into 
*.calkit files.   
*RST value 
SCPI, 
Command 
Types 
Device-specific, no query.   
Example: 
CORR:CKIT:INST 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\ZCAN.ck' 
Load the previously created ZVR cal kit file ZCAN.ck from the 
default cal kit directory.  
MMEM:STOR:CKIT 'ZCAN', 
'C:\R_S\Instr\user\Nwa\Calibration\Kits\ZCAN.calkit' 
Store the imported cal kit data to a NWA cal kit file 
ZCAN.calkit (assuming that the cal kit name stored in ZCAN.ck 
reads ZCAN).  
[SENSe<Ch>:]CORRection:COLLect...     
This subsystem controls the system error correction and manages calibration kits.   
[SENSe<Ch>:]CORRection:COLLect[:ACQuire] THRough | OPEN1 | OPEN2 | 
OPEN12 | SHORT1 | SHORT2 | SHORT12 | MATCH1 | MATCH2 | MATCH12 | NET | 
ATT | IMATCH12 | REFL1 | REFL2 | SLIDe1 | SLIDE2 | SLIDE12 | LINE1 | LINE2 | 
M1O2 | O1M2 | M1S2 | S1M2 | OSHORT1 | OSHORT2 
Starts a calibration measurement in order to acquire measurement data for the selected standards. The 
standards are reflection or transmission standards and must be connected to port 1 or 2 of the analyzer. 
Use the generalized command [SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected to 
obtain measurement data at arbitrary analyzer ports.  
<Ch> 
Channel number of the calibrated channel. If unspecified the numeric suffix is set to 1. 
THRough ... 
OSHORT2 
Standard types: Through, Open, Short, Match (MATCH12 and IMATCH12 are 
synonymous), Symmetric Network (NET), Attenuation (ATT), Reflect, Sliding Match 
(SLIDe), Line1 (LINE1 and LINE are synonymous), Line2, Match/Open (M1O2, 
O1M2), Match/Short (M1S2, S1M2), Offset Short (OSHort).
The numbers in the parameters denote the analyzer ports. Two numbers 12 mean that 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
493
two separate calibrations are performed at ports 1 and 2.  
*RST value 
ON 
AUTO | 
<delay | 
phase> 
Optional entry of delay time or phase for UTHRough standard:   
AUTO – The analyzer determines the delay time or phase during the calibration sweep
<delay or phase> – entry of the delay time in ps (for non-dispersive standards) or of an 
estimate of the phase at the start frequency of the sweep in deg (for dispersive 
standards). If an estimate of the start phase is entered, the analyzer uses the 
calculated value which is closest to the estimate.   
*RST value 
AUTO 
SCPI, 
Command 
Types 
Confirmed (with device-specific standards), no query 
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE.  
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe <Boolean> 
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored 
after the calibration is completed. The setting is valid for the current calibration, where it overwrites the 
global setting ([SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault). A new 
calibration deletes the calibration data acquired in previous calibrations.  
<Ch> 
Channel number of the calibrated channel.  
<Boolean> 
ON | OFF - Keep measurement data on or off.   
*RST value 
OFF  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
CORR:COLL:RSAV:DEF ON 
Generally keep the raw measurement data of the standards after the 
calibration is completed.   
CORR:COLL:METH:DEF 'Test',RSHort,1 
Select a one-port normalization at port 1 with a short standard as calibration 
type. 
CORR:COLL:SEL SHOR,1 
Measure a short standard connected to port 1 and store the raw measurement 
results of this standard.  
CORR:COLL:RSAV OFF 
To save disk space, delete the current raw calibration data after the calibration 
is completed.  
CORR:COLL:SAVE:SEL  
Calculate the system error correction data and apply them to the active 
channel. 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
494
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault <Boolean> 
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored 
after the calibration is completed. The setting remains valid for all subsequent calibrations until it is 
changed explicitly. A new calibration deletes the calibration data acquired in previous calibrations.  
<Ch> 
Channel number of the calibrated channel.  
<Boolean> 
ON | OFF - Keep measurement data on or off.   
*RST value 
OFF  
SCPI, Command Types 
Device-specific, command or query 
Example: 
See [SENSe<Ch>:]CORRection:COLLect[:AQUire]:RSAVe. 
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected THRough | OPEN | 
SHORt | MATCh | NET | ATT | REFL | SLIDE | LINE1 | LINE2,<port_no>,<port_no> 
Starts a calibration measurement in order to acquire measurement data for the selected standards. The 
standards are reflection or transmission standards and can be connected to arbitrary analyzer ports.  
<Ch> 
Channel number of the calibrated channel.  
THRough ... 
LINE2 
Standard types: Through, Open, Short, Match, Symmetric Network (NET), 
Attenuation (ATT), Reflect, Sliding Match (SLIDe), Line1 (LINE1 and LINE are 
synonymous), Line2.
<port_no> 
Port numbers of the analyzer. For a transmission standard (through, line, 
attenuation, symmetric network) the input and output port numbers must be 
specified, for reflection standards, only one port number is required.  
*RST value 
AUTO | 
<delay | 
phase> 
Optional entry of delay time or phase for UTHRough standard:   
AUTO – The analyzer determines the delay time or phase during the calibration 
sweep 
<delay or phase> – entry of the delay time in ps (for non-dispersive standards) or of 
the phase at the start frequency of the sweep in deg (for dispersive standards)  
*RST value 
AUTO 
SCPI, 
Command 
Types 
Device-specific, no query 
Example: 
See [SENSe<Ch>:]CORRection:COLLect:SAVE:SELected.  
[SENSe<Ch>:]CORRection:COLLect:AUTO '<file_name>',<port_no>{,<port_no>} 
Selects and initiates a one-port or two-port automatic calibration at arbitrary analyzer and cal unit ports.  
If the test setup contains a high attenuation the analyzer may fail to detect the cal unit ports 
connected to each of its ports. In this case use the extended command 
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs.
If several calibration units are connected, use SYSTem:COMMunicate:RDEVice:AKAL:ADDRess to 