R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
515
The following keywords define the measurement parameter (see SCPI command reference: function 
name): 
POWer:S<Pt
out
><Pt
in
>
S-parameter with output and input port number of the DUT, e.g. S11, S21.
POWer:KFACtor 
POWer:MUFactor<1|2>
Stability factor K 
Stability factors µ1 or µ2 
[SENSe<Ch>:]POWer...     
This subsystem controls the power at the input ports of the analyzer.  
[SENSe<Ch>:]POWer:ATTenuation ARECeiver | BRECeiver, <numeric_value> 
Sets an attenuation factor for the received waves b
1
and b
2
.
<Ch> 
Channel number. If unspecified the numeric suffix is set to 1. 
ARECeiver | 
BRECeiver | 
Test port of the analyzer: 
Test port 1 
Test port 2  
<numeric_value> 
Attenuation factor for the received wave.  
Range [def. unit] 
0dB to 30 dB. UP and DOWN increment/decrement the attenuation in 5 dB 
steps. The analyzer rounds any entered value below the maximum 
attenuation to the closest step.  
*RST value 
+10 dB (0 dB for R&S ZVL13) 
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
POW:ATT AREC, 15
Set an attenuation factor of 15 dB for the waves received at test port 1 and 
channel no. 1. The other test ports and channels are not affected.  
[SENSe<Ch>:]ROSCillator...     
This subsystem controls the frequency reference oscillator.  
[SENSe<Ch>:]ROSCillator[:SOURce] INTernal | EXTernal 
Selects the source of the reference oscillator signal. 
<Ch> 
Channel number. This suffix is ignored in the ROSCillator subsystem and can be 
set to any value.  
INTernal 
Select internal 10 MHz reference oscillator.  
EXTernal 
Select external reference clock. The frequency of the external reference clock is 
specified via [SENSe<Chn>:]ROSCillator:EXTernal:FREQuency.
*RST value 
INTernal  
Metadata in pdf documents - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
pdf xmp metadata editor; read pdf metadata online
Metadata in pdf documents - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
remove pdf metadata; pdf keywords metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
516
SCPI, 
Command 
Types 
Confirmed, command or query 
Example: 
ROSC EXT  
Select external reference oscillator.  
ROSC:EXT:FREQ?  
Query the frequency of the external reference oscillator. The response is 10000000 
Hz, i.e. the frequency of the external reference oscillator must be 10 MHz.  
[SENSe<Ch>:]ROSCillator:EXTernal:FREQuency <numeric_value> 
Specifies or queries the frequency of the external reference oscillator.  
<Ch> 
Channel number. This suffix is ignored in the ROSCillator subsystem and 
can be set to any value.  
<numeric_value> 
Frequency of the external reference clock signal. The frequency must be 10 
MHz; any other setting is ignored.  
Range [def. unit] 
10 MHz [Hz]. 
*RST value 
10 MHz  
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
See [SENSe<Ch>:]ROSCillator[:SOURce]. 
[SENSe<Ch>:]SEGMent<Seg>...     
This subsystem defines all channel settings for a Segmented Frequency sweep. A segmented sweep is 
activated via [SENSe<Ch>:]SWEep:TPYE SEGMent.
The commands in the [SENSe<Ch>:]SEGMent<Seg>... subsystem do not accept the step 
parameters UP and DOWN. Numeric values can be entered directly or using the DEFault, 
MINimum, MAXimum parameters.  
[SENSe<Ch>:]SEGMent<Seg>:ADD 
Inserts a new sweep segment using default channel settings (Insert New Segment). The added segment 
covers the frequency interval between the maximum frequency of the existing sweep segments and the 
stop frequency of the entire sweep range.  
Use [SENSe<Ch>:]SEGMent<Seg>:INSert to create a segment with specific channel settings.  
<Ch> 
Channel number. If unspecified the numeric suffix is set to 1.  
<Seg> 
Sweep segment number. Segment numbers must be sequential. If n segments 
already exist, the added segment must have the segment number n+1.  
*RST value 
SCPI, Command 
Device-specific, no query.  
C# PDF Library SDK to view, edit, convert, process PDF file for C#
NET empowers C# developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. C#.NET: Edit PDF Metadata.
c# read pdf metadata; pdf xmp metadata viewer
C# PDF Print Library: Print PDF documents in C#.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
clean pdf metadata; search pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
517
Types 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ 
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.  
SEGM2:ADD  
Create a second sweep segment. The frequency range of the second segment 
will be between 1.5 MHz and the maximum frequency of the analyzer.  
[SENSe<Ch>:]SEGMent<Seg>:ATTenuation<port_no> <Attenuation> 
Defines the attenuation factors for the received waves b
1
and b
2
in segment no. <Seg>. .  
<Ch> 
Channel number.  
<Seg> 
Sweep segment number.  
<port_no> 
Test port number of the analyzer, 1 (wave b
1
)or 2 (wave b
2
). If unspecified the 
numeric suffix is set to 1. 
<Attenuation> 
Attenuation factor for the received wave.  
Range [def. unit] 
0dB to +30 dB. UP and DOWN increment/decrement the attenuation in 5 dB 
steps. The analyzer rounds any entered value below the maximum attenuation 
to the closest step. 
*RST value 
+10 dB 
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ 
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.  
SEGM:ATT1 5; ATT2 15  
Define an input attenuation of 5 dB for the b
1
wave, 15 dB for the b
2
wave.  
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution] <Bandwidth> 
Defines the resolution bandwidth of the analyzer (Meas Bandwidth) in sweep segment no. <Seg>. At the 
same time, the command activates separate bandwidth setting in all sweep segments 
([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol ON).  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Bandwidth> 
Internal source power.  
Range [def. unit] 
1.0E-6 Hz to 5 MHz [Hz]. UP and DOWN increment/decrement the bandwidth in 
1-2-5 steps for each decade. The analyzer rounds up any entered value between 
these steps and rounds down values exceeding the maximum bandwidth.   
*RST value 
10 kHz  
SCPI, 
Command 
Types 
Device-specific, command or query 
VB.NET PDF Print Library: Print PDF documents in vb.net, ASP.NET
view, Annotate,Convert documents online using ASPX. NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
adding metadata to pdf files; delete metadata from pdf
C# PDF File Compress Library: Compress reduce PDF size in C#.net
compressing control is designed to offer C# developers to compress existing PDF documents in .NET Document and metadata. C#.NET DLLs: Compress PDF Document.
extract pdf metadata; pdf metadata online
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
518
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus 10 kHz measurement bandwidth.  
SEGM:BWID 1 MHZ  
Increase the resolution bandwidth to 1 MHz.  
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol <Boolean> 
Qualifies whether or not the Meas Bandwidth can be set independently for each sweep segment.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON: The bandwidth can be set independently for each sweep segment.  
OFF: The bandwidth in all sweep segments is equal to the bandwidth for 
unsegmented sweeps set via SENSe<Ch>:BWIDth[:RESolution].
*RST value 
OFF. The parameter is automatically switched to ON when a bandwidth is entered 
using [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution].
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 
10 kHz measurement bandwidth.  
SEGM:BWID 1 MHZ  
Increase the resolution bandwidth to 1 MHz.  
SEGM:BWID:CONT OFF  
Couple the bandwidths in all segments and reset the bandwidth in segment no. 1 to 
the initial value.  
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect FAST | NORMal  
Defines the Filter Type (selectivity) in sweep segment no. <Seg>. At the same time, the command 
activates separate filter type settings in all sweep segments 
([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]SELect:CONTrol ON).  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
FAST 
Use IF filter with normal selectivity and short settling time.  
NORMal 
Use IF filter with higher selectivity but larger settling time. 
*RST value 
FAST  
SCPI, Command 
Types 
Device-specific, command or query 
VB.NET PDF File Merge Library: Merge, append PDF files in vb.net
VB.NET Guide and Sample Codes to Merge PDF Documents in VB.NET Project. Batch merge PDF documents in Visual Basic .NET class program.
adding metadata to pdf; pdf metadata reader
VB.NET PDF Library SDK to view, edit, convert, process PDF file
for .NET empowers VB.NET developers to implement fast and high quality PDF conversions to or from multiple supported images and documents. PDF Metadata Edit.
batch pdf metadata; edit pdf metadata acrobat
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
519
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus the Fast filter type.  
SEGM:BWID:SEL NORM 
Activate an IF filter with high selectivity.  
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol <Boolean> 
Qualifies whether or not the Filter Type (selectivity) can be set independently for each sweep segment.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON – The filter type can be set independently for each sweep segment.  
OFF – The filter type in all sweep segments is equal to the filter type for 
unsegmented sweeps set via SENSe<Ch>:BWIDth[:RESolution]:SELect.
*RST value 
OFF. The parameter is automatically switched to ON when a bandwidth is entered 
using [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect.
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
*RST; SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 
the Fast filter type.  
SEGM:BWID:SEL NORM 
Change the filter type to Normal.
BWID:SEL?  
Query the (default) filter type for unsegmented sweeps. The response is FAST.
SEGM:BWID:SEL:CONT OFF  
Couple the filter type in all segments and reset the filter type in segment no. 1 to the 
unsegmented value FAST..  
[SENSe<Ch>:]SEGMent:CLEar 
Deletes all sweep segments in the channel. The command is equivalent to 
[SENSe<Ch>:]SEGMent:DELet:ALL.
[SENSe<Ch>:]SEGMent<Seg>:COUNt? 
Returns the number of sweep segments in the channel including all segments that are switched off 
([SENSe<Ch>:]SEGMent<Seg>[:STATe] OFF).
<Ch> 
Channel number.   
C# PDF File Merge Library: Merge, append PDF files in C#.net, ASP.
C#.NET PDF Library - Merge PDF Documents in C#.NET. Provide NET components for batch combining PDF documents in C#.NET class. Powerful
read pdf metadata java; pdf metadata editor
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
Our .NET SDKs own the most advanced & comprehensive documents and images reading and has a mature imaging utility which allows Tiff image file metadata to be
rename pdf files from metadata; remove metadata from pdf acrobat
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
520
*RST value 
SCPI, Command 
Types 
Device-specific, query only.  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default 
settings.  
SEGM OFF  
Disable the measurement in the created sweep segment. 
SEGM:COUN?  
Query the number of segments. 
[SENSe<Ch>:]SEGMent<Seg>:DEFine <Start>,<Stop>,<Points>,<Power>,<Point 
Delay>,<Unused>,<Meas. Bandwidth>[,<LO>] 
Re-defines a sweep segment with specific channel settings (Insert New Segment). The segment replaces 
an existing segment <Seg> in the segment list. The modified segment must not overlap with any of the 
existing segments.  
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings 
except <LO> can be changed for existing segments using other commands of the 
[SENSe<Ch>:]SEGMent<Seg>... subsystem.   
Note: Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use 
[SENSe<Ch>:]SEGMent<Seg>:INSert (no query) to insert a new segment into the current 
segment list.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. Segment numbers must be sequential. The specified 
segment number must be smaller or equal to the number of existing segments 
plus 1.   
If segment number <Seg> already exists, it is replaced by the new segment.  
<Start>, <Stop> 
Start and stop frequency of the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP. 
Range [def. unit] 
Anew segment must not overlap with any of the existing segments. Besides the 
frequency range depends on the analyzer model. [Hz]  
<Points> 
Number of sweep points in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts. 
Range [def. unit] 
1to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
<Power> 
Internal source power in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:POWer. 
Range [def. unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to 
the data sheet [dBm]. UP and DOWN increment/decrement the source power in 
1-dB steps.  
<Point Delay> 
Delay for each partial measurement in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting 
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime, this parameter is 
replaced by <Time>. On R&S ZVL13 network analyzers, no point delay can be 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
521
set: the value must be 0.   
Range [def. unit] 
0s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the 
segment, which is equivalent to a meas. delay of 0 s.  
<Unused> 
Ignored parameter, should be set to the default value 0.  
<Meas. 
Bandwidth> 
Resolution bandwidth in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution].
Range [def. unit] 
1.0E-6 Hz to 5 MHz [Hz].  
<LO> 
Position of the local oscillator frequency LO relative to the RF frequency (Spur 
Avoid). In remote control this parameter must be set when a sweep segment is 
created.  
Range [def. unit] 
POSitive: LO > RF 
SCPI, 
Command 
Types 
Device-specific, with query (used to retrieve the channel settings for a particular 
sweep segment).  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings.  
SEGM:DEF?  
Query the channel settings for the new segment.  
[SENSe<Ch>:]SEGMent<Seg>:DELete 
Deletes the specified (single) sweep segment (Del. Selected Segment).
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL deletes all segments in the channel.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. If unspecified, the value is set to 1.  
*RST value 
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus NORMal selectivity.  
SEGM:DEL  
Delete the created segment.
[SENSe<Ch>:]SEGMent:DELete:ALL 
Deletes all sweep segments in the channel (Del. All Segments). [SENSe<Ch>:]SEGMent<Seg>:DELete 
deletes a single segment.  
<Ch> 
Channel number.   
*RST value 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
522
SCPI, Command 
Types 
Device-specific, no query.  
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings 
and thus NORMal selectivity.  
SEGM:ALL  
Delete the created segment and all segments in the channel created before. 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:CENTer? 
Returns the center frequency of sweep segment no. <Seg>. 
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
Response 
Center frequency of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz].   
*RST value 
SCPI, Command 
Types 
Device specific, query only. 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz. 
SEGM:FREQ:CENT?  
Query the center frequency of the created segment. The response is 
1.2500000000E+006.  
Note: The frequency range of the sweep segment can be changed via 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN?  
Returns the width of the frequency range of sweep segment no. <Seg>. 
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
Response 
Frequency span of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz].   
*RST value 
SCPI, Command 
Types 
Device specific, query only. 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
523
MHz. 
SEGM:FREQ:SPAN?  
Query the span of the created segment. The response is 
5.0000000000E+005.  
Note: The frequency range of the sweep segment can be changed via 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP.
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt <Start> 
Defines the Start frequency of sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Start> 
Start frequency of the sweep.  
Range [def. unit] 
Depending on the instrument model [Hz]. The increment is 0.1 kHz.  
*RST value 
– (no sweep segment defined after reset) 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz. 
SEGM:FREQ:STAR?  
Query the start frequency of the created segment. The response is 
1.0000000000E+006.  
Note: If the start frequency entered is greater than the current stop frequency 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP), the stop frequency is set to the start 
frequency plus the minimum frequency span of 1 Hz.  
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP <Stop> 
Defines the Stop frequency of sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Stop> 
Stop frequency of the sweep. 
Range [def. unit] 
Depending on the instrument model [Hz]. The increment is 0.1 kHz.  
*RST value 
–(no sweep segment defined after reset) 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ   
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 
MHz. 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
524
SEGM:FREQ:STOP?  
Query the stop frequency of the created segment. The response is 
1.5000000000E+006. 
Note: If the stop frequency entered is smaller than the current start frequency 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt), the start frequency is set to the stop 
frequency minus the minimum frequency span 
([SENSe<Ch>:]SEGMent<Seg>:FREQuency:SPAN).   
[SENSe<Ch>:]SEGMent<Seg>:INSert <Start>,<Stop>,<Points>,<Power>,<Point 
Delay>,<Unused>,<Meas. Bandwidth> 
Inserts a new sweep segment with specific channel settings (Define Segments - Add). The new segment 
must not overlap with any of the existing segments. 
Entry of the first seven numeric parameters is mandatory; no default values are provided. All settings 
except <LO> can be changed for existing segments using other commands of the 
[SENSe<Ch>:]SEGMent<Seg>... subsystem.   
Note: Use [SENSe<Ch>:]SEGMent<Seg>:Add to create a segment with default channel settings. Use 
[SENSe<Ch>:]SEGMent<Seg>:DEFine to change or query all settings of an existing segment.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. Segment numbers must be sequential. The specified 
segment number must be smaller or equal to the number of existing segments 
plus 1. Moreover, segment numbers must selected such that the corresponding 
frequency ranges are in ascending order.  
If one or more sweep segments with segment numbers <Seg> or larger exist in 
the current channel, then all these existing segment numbers are incremented by 
1and the new segment is inserted as segment no. <Seg>.   
<Start>, <Stop> 
Start and stop frequency of the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STARt and 
[SENSe<Ch>:]SEGMent<Seg>:FREQuency:STOP. 
Range [def. unit] 
Anew segment must not overlap with any of the existing segments. Besides the 
frequency range depends on the analyzer model. [Hz]  
<Points> 
Number of sweep points in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts. 
Range [def. unit] 
1to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
<Power> 
Internal source power in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:POWer. 
Range [def. unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to 
the data sheet [dBm]. UP and DOWN increment/decrement the source power in 
1-dB steps.  
<Point Delay> 
Delay for each partial measurement in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting 
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELec SWTime, this parameter is 
replaced by <Time>. On R&S ZVL13 network analyzers, no point delay can be 
set: the value must be 0.  
Range [def. unit] 
0s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the 
segment, which is equivalent to a meas. delay of 0 s.  