how to download pdf file in c# windows application : Read pdf metadata control application platform web page azure .net web browser ZVL_Operating_008_V3_2054-part1097

R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
525
<Unused> 
Ignored parameter, should be set to the default value 0.  
<Meas. 
Bandwidth> 
Resolution bandwidth in the segment. See 
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution].
Range [def. unit] 
1.0E-6 Hz to 5 MHz [Hz].  
SCPI, 
Command 
Types 
Device-specific, no query.  
Example: 
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ  
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.  
SEGM2:ADD  
Create a second sweep segment. The frequency range of the second segment 
will be between 1.5 MHz and the maximum frequency of the analyzer.  
[SENSe<Ch>:]SEGMent<Seg>:OVERlap <Boolean> 
Queries whether the analyzer supports overlapping sweep segments.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON | OFF: No effect.  
*RST value 
OFF. If used as a query, the command returns the information that overlapping 
sweep segments are not supported (OFF).
SCPI, Command 
Types 
Device-specific, command (no effect) or query 
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel] <Power> 
Defines the Power of the internal signal source in sweep segment no. <Seg>. At the same time, the 
command activates separate power control in all sweep segments 
([SENSe<Ch>:]SEGMent<Seg>:POWer[:Level]:CONTrol ON).  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Power> 
Internal source power.  
Range [def. unit] 
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to 
the data sheet [dBm].  
*RST value 
–10 dBm. 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and 
thus –10 dBm internal source power.  
SEGM:POW -20 
Read pdf metadata - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
pdf metadata; preview edit pdf metadata
Read pdf metadata - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
modify pdf metadata; change pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
526
Decrease the power to –20 dBm.  
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol <Boolean> 
Qualifies whether or not the Power of the internal signal source can be set independently for each sweep 
segment.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Boolean> 
ON: The power can be set independently for each sweep segment.  
OFF: The power in all sweep segments is equal to the internal source power for 
unsegmented sweeps set via 
SOURce<Ch>:POWer[:LEVel][:IMMediate][:AMPlitude].
*RST value 
OFF. The parameter is automatically switched to ON when a segment level is 
entered using [SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus –
10 dBm internal source power.  
SEGM:POW -20 
Decrease the power to –20 dBm.  
SEGM:POW:CONT OFF  
Couple the powers in all segments and reset the power in segment no. 1 to the initial 
value.   
[SENSe<Ch>:]SEGMent<Seg>[:STATe] <Boolean> 
Activates or deactivates the sweep segment <Seg>. Sweep points belonging to inactive segments only 
are not measured    
<Ch> 
Channel number.  
<Seg> 
Sweep segment number.  
<Boolean> 
ON | OFF - Activates or deactivates the measurement in sweep segment 
<Seg>.  
*RST value 
ON 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings. 
SEGM OFF 
VB.NET PDF Image Extract Library: Select, copy, paste PDF images
Get image information, such as its location, zonal information, metadata, and so on. Extract image from PDF free in .NET framework application with trial SDK
batch update pdf metadata; analyze pdf metadata
How to C#: Basic SDK Concept of XDoc.PDF for .NET
XDoc.PDF for .NET supports editing PDF document metadata, like Title, Subject, Author, Creator, Producer, Keywords, Created Date, and Last Modified Date.
c# read pdf metadata; pdf metadata viewer online
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
527
Disable the measurement in the created sweep segment.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl <Meas. Delay> 
Defines the delay time for each partial measurement in sweep segment no. <Seg> (Meas. Delay). If 
coupling of the segments is switched on ([SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol ON)
the delay is valid for all sweep segments in the current channel. This command is not available on R&S 
ZVL13 network analyzers.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number.  
<Meas. 
Delay> 
Meas. delay before each partial measurement.  
Range [def. 
unit] 
0s to 2.5E+003 s. [s] Changing the delay leaves the number of points unchanged 
but has an impact on the duration of the sweep 
([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME).   
*RST value 
0s
SCPI, 
Command 
Types 
Device-specific, command or query. This default value corresponds to automatic 
sweep time setting in manual control. 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings.  
SEGM:SWE:DWEL 1 MS 
Set the meas. delay in segment no. 1 to 1 ms. 
SEGM:DEF? Response: 300000,8000000000,51,-300,0.056559,0,10000,POS,NORM 
Query the channel parameters for sweep segment 1. The response value for the 
segment sweep time (olive) implicitly contains the defined meas. delay.  
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol <Boolean> 
Qualifies whether or not the Meas. Delay defined via [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl can 
be set independently for each sweep segment.  
<Ch> 
Channel number.   
<Seg> 
Sweep segment number. This suffix is ignored; the command controls the whole 
segmented sweep.  
<Boolean> 
ON: The meas. delay can be set independently for each sweep segment.  
OFF: The meas. delay in all sweep segments is equal to the Meas. Delay 
measurements for unsegmented sweeps set via [SENSe<Ch>:]SWEep:DWELl.
*RST value 
OFF. The parameter is automatically switched to ON when a meas. delay time is 
entered using [SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl.
SCPI, 
Command 
Types 
Device-specific, command or query.  
VB.NET PDF: Basic SDK Concept of XDoc.PDF
XDoc.PDF for .NET supports editing PDF document metadata, like Title, Subject, Author, Creator, Producer, Keywords, Created Date, and Last Modified Date.
rename pdf files from metadata; pdf metadata viewer
C# PDF Image Extract Library: Select, copy, paste PDF images in C#
Scan image to PDF, tiff and various image formats. Get image information, such as its location, zonal information, metadata, and so on.
batch pdf metadata editor; online pdf metadata viewer
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
528
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus 
0s meas. delay.  
SEGM:SWE:DWELl 0.1  
Increase the meas. delay to 0.1 s.  
SEGM:SWE:DWELl:CONT OFF  
Couple the meas. delay in all segments and reset the delay in segment no. 1 to the 
initial value of 0 s.   
[SENSe<Ch>:]SEGMent<Seg>:SWEep:POINts <Points> 
Defines the total number of measurement Points in sweep segment no. <Seg>.  
<Ch> 
Channel number. 
<Seg> 
Sweep segment number.  
<Points> 
Number of points in the segment.  
Range [def. unit] 
1to 2147483647. [1] 1 is allowed if start and stop frequencies are equal.  
*RST value 
51 
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SEGM:ADD  
Create a new sweep segment no. 1 in channel no. 1 using default settings 
and thus 51 sweep points.  
SEGM:SWE:POIN 401 
Increase the number of points to 401.  
[SENSe<Ch>:]SWEep...     
This subsystem provides general settings to control the sweep . Most of the settings are related to the 
sweep time.  
The [SENSe<Ch>:]SWEep... settings apply to non-segmented sweep types. Segmented sweeps are 
selected via [SENSe<Ch>:]SWEep:TPYE SEGMent and configured in the 
[SENSe<Ch>:]SEGMent<Seg>... subsystem.      
[SENSe<Ch>:]SWEep:COUNt <No_of_Sweeps> 
Defines the number of sweeps to be measured in single sweep mode (INITiate<Ch>:CONTinuous 
OFF).  
<Ch> 
Channel number.   
<No_of_Sweeps> 
Number of consecutive sweeps to be measured. 
C# PDF - Read Barcode on PDF in C#.NET
Bookmark: Edit Bookmark. Metadata: Edit, Delete Metadata. Watermark: Add Watermark to PDF. Form Process. Data: Read, Extract Field Data. Data: Auto Fill-in Field
add metadata to pdf; add metadata to pdf file
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
C#.NET. Allow Users to Read and Edit Metadata Stored in Tiff Image in C#.NET Application. C# Overview - View and Edit TIFF Metadata.
remove metadata from pdf acrobat; clean pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
529
Range [def. unit] 
1to 999. [–]  
*RST value 
1
SCPI, Command Types 
Confirmed, command or query 
Example: 
See CALCulate<Chn>:DATA:NSWeep?.
[SENSe<Ch>:]SWEep:DWELl <delay> 
Defines the Meas. Delay time for each partial measurement. This command is not available on R&S 
ZVL13 network analyzers. 
<Ch> 
Channel number.   
<delay> 
Meas. delay before each partial measurement.  
Range [def. unit] 
0s to 3.179551E+002 s. [s] Changing the delay leaves the number of points 
unchanged but has an impact on the duration of the sweep.   
*RST value 
0s
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
FUNC "XFR:POW:S12"  
Activate a frequency sweep and select the S-parameter S12 as measured 
parameter for channel and trace no. 1.  
SWE:TIME?  
Query total sweep time. 
SWE:DWEL 1  
Set a delay of 1 s for each partial measurement. 
SWE:TIME?  
Query total sweep time. The time is extended by the delay times the total 
number of partial measurements per sweep. 
[SENSe<Ch>:]SWEep:POINts <no_points> 
Defines the total number of measurement points per sweep (Number of Points).
<Ch> 
Channel number 1. 
<no_points> 
Number of points per sweep.  
Range [def. unit] 
2to 4001. [1]  
*RST value 
201 
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
FUNC "XFR:POW:S12"  
Activate a frequency sweep and select the S-parameter S12 as measured 
parameter for channel and trace no. 1.  
SWE:TIME?  
VB.NET PDF Library SDK to view, edit, convert, process PDF file
PDF Metadata Edit. Offer professional PDF document metadata editing APIs, using which VB.NET developers can redact, delete, view and save PDF metadata.
google search pdf metadata; pdf remove metadata
VB.NET PDF Text Extract Library: extract text content from PDF
PDF ›› VB.NET PDF: Extract PDF Text. VB.NET PDF - Extract Text from PDF Using VB. How to Extract Text from PDF with VB.NET Sample Codes in .NET Application.
add metadata to pdf programmatically; get pdf metadata
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
530
Query total sweep time. 
SWE:POIN 2010  
Multiply the (default) number of points by 10. 
SWE:TIME?  
Query total sweep time again. The analyzer estimates a sweep time that is 
also multiplied by 10.  
[SENSe<Ch>:]SWEep:SPACing  LINear | LOGarithmic  
Defines the frequency vs. time characteristics of a frequency sweep (Lin Frequency or Log Frequency).
The command has no effect on segmented frequency sweeps.  
Note: Use [SENSe<Ch>:]SWEep:TPYE to select sweep types other than Lin Frequency or Log 
Frequency. 
<Ch> 
Channel number.  
LINear 
The stimulus frequency is swept in equidistant steps over the frequency range. 
In a Cartesian diagram, the x-axis is a linear frequency axis.  
LOGarithmic 
The frequency is swept in equidistant steps on a logarithmic scale. In a 
Cartesian diagram, the x-axis is a logarithmic frequency axis.  
*RST value 
LINear 
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
FUNC "XFR:POW:S12"  
Activate a frequency sweep and select the S-parameter S12 as measured 
parameter for channel and trace no. 1.  
SWE:SPAC LOG  
Change to sweep type Log Frequency.
[SENSe<Chn>:]SWEep:SRCPort 1 | 2 
Selects a source port for the stimulus signal (Drive Port). The setting acts on the active trace.  
If an S-parameter S
<out><in>
is measured, the second port number index <in> (input port of the DUT = drive 
port of the analyzer) is set equal to the selected drive port: Drive port selection affects the measured 
quantity.   
<Chn> 
Channel number used to identify the active trace. If unspecified the numeric 
suffix is set to 1. 
1... 
Test port number of the analyzer, 1 to 2.  
*RST value 
1(test port 1) 
SCPI, Command 
Types 
Device-specific, with query.  
Example: 
CALC4:PAR:SDEF "Ch4Tr1", "A1"  
Create channel 4 and a trace named Ch4Tr1 to measure the wave quantity a1. 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
531
The trace automatically becomes the active trace.  
SENS4:SWE:SRCP 2 
Select drive port 2 for the active trace.  
[SENSe<Ch>:]SWEep:STEP <step_size> 
Sets the distance between two consecutive sweep points.  
This setting is valid for sweep types with equidistant sweep points only. It does not apply to 
logarithmic and segmented sweeps.  
<Ch> 
Channel number.  
<step_size> 
Stimulus step size.  
Range [def. 
unit] 
Depending on the other channel settings. The minimum step size is equal to the 
sweep span divided by the maximum number of points minus one 
([SENSe<Ch>:]SWEep:POINts), the maximum step size is equal to the sweep span 
([SENSe<Ch>:]FREQuency:SPAN). See also description of manual control and 
program example below. [–]  
*RST value 
The default step size is equal to the default sweep span of the analyzer divided by 
the default number of points minus one.  
SCPI, 
Command 
Types 
Confirmed, with query.  
Example: 
*RST; SWE:STEP?  
Query the default step size. For a 6 GHz analyzer, the response is 29998500. 
29998500 Hz = (6 GHz – 300 kHz) / 200. 
SWE:STEP UP  
Increase the step size. 
FREQ:STOP?; :SWE:POIN? 
Query the stop frequency of the sweep and the number of points. Increasing the 
step size has changed both values.   
[SENSe<Ch>:]SWEep:TYPE LINear | LOGarithmic | SEGMent 
Selects the sweep type (frequency) and the position of the sweep points across the sweep range.  
<Ch> 
Channel number.  
LINear 
Lin. frequency sweep at constant source power 
(SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude]). The 
stimulus frequency ([SENSe<Ch>:]FREQuency:...) is swept in equidistant steps 
over the frequency range. In a Cartesian diagram, the x-axis is a linear frequency 
axis.  
LOGarithmic 
Log. frequency sweep. The frequency is swept in equidistant steps on a logarithmic 
scale. In a Cartesian diagram, the x-axis is a logarithmic frequency axis.  
SEGMent 
Segmented frequency sweep. The sweep range is composed of several continuous 
frequency ranges or single frequency points defined by means of the commands in 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
532
the [SENSe<Ch>:]SEGMent<Seg>... subsystem.  
*RST value 
LINear 
SCPI, 
Command 
Types 
Device-specific, command or query 
Example: 
FUNC "XFR:POW:S12"  
Activate a frequency sweep and select the S-parameter S12 as measured parameter 
for channel and trace no. 1.  
SWE:TYPE LOG  
Change to sweep type Log Frequency.
SOURce<Ch>:...     
This subsystem controls the power of the internal signal source and controls the output signals of the 
instrument.  
SOURce<numeric_suffix>:EXTernal<numeric_suffix>:ROSCillator[:SOURce] 
INTernal | EXTernal 
Switches between external and internal reference oscillator.  
<numeric_suffix> 
1| 2 - The numeric suffixes are ignored. 
INTernal  
Select internal 10 MHz reference oscillator.  
EXTernal 
Select external reference clock. The frequency of the external reference clock 
is specified via [SENSe<Chn>:]ROSCillator:EXTernal:FREQuency.
*RST value 
INT  
SCPI, Command 
Types 
Device-specific, command or query 
Example: 
SOUR:EXT:ROSC EXT  
Switches to external reference oscillator. 
SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude] <numeric_value> 
Defines the power of the internal signal source (channel power). The setting is valid for all sweep types.  
<Ch> 
Channel number.  
<Pt> 
Test port number of the analyzer. This suffix is ignored because the selected 
channel power applies to all source ports used in the active channel.  
<numeric_value> 
Internal source power.  
Range [def. unit] 
–70 dBm to +20 dBm (to 0 dBm for R&S ZVL13 analyzers). The exact range 
for all analyzer models is quoted in the data sheet [dBm]. UP and DOWN 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
533
increment/decrement the source power in 0.1-dB steps.  
*RST value 
–10 dBm  
SCPI, Command 
Types 
Confirmed, command or query 
Example: 
FUNC "XFR:POW:RAT  B1, A2" 
Activate a frequency sweep and select the ratio B1/A2 as measured parameter 
for channel and trace no. 1.  
SOUR:POW -6 
Set the internal source power for channel 1 to -6 dBm. 
STATus...     
This subsystem controls the Status Reporting System. Note that *RST does not influence the status 
registers.   
STATus:PRESet 
Configures the status reporting system such that device-dependent events are reported at a higher level.  
The command affects only the transition filter registers, the ENABle registers, and queue enabling: 
The ENABle parts of the STATus:OPERation and STATus:QUEStionable... registers are set to all 
1's.  
The PTRansition parts are set all 1's, the NTRansition parts are set to all 0's, so that only positive 
transitions in the CONDition part are recognized. 
The status reporting system is also affected by other commands, see Reset Values of the Status 
Reporting System.
*RST value 
SCPI, Command Types 
Confirmed, no query 
Example: 
STAT:PRES  
Preset the status registers.  
STATus:QUEStionable:CONDition? 
Returns the contents of the CONDition part of the QUEStionable register. Reading the CONDition 
registers is nondestructive.  
*RST value 
SCPI, Command 
Types 
Confirmed, query only 
Example: 
STAT:QUES:COND?  
Query the CONDition part of the QUEStionable register to check for 
R&S ZVL 
Command Reference 
Instrument-Control Commands 
Operating Manual 1303.6580.32-05 
534
questionable instrument states. 
STATus:QUEStionable:ENABle 
Sets the enable mask which allows true conditions in the EVENt part of the QUEStionable register to be 
reported in the summary bit. If a bit is 1 in the enable register and its associated event bit transitions to 
true, a positive transition will occur in the summary bit (bit 3 of the STatus Byte).  
<NRf> 
0to 65535 (decimal representation) 
*RST value 
–(see also Reset Values of the Status Reporting System) 
SCPI, Command Types 
Confirmed, command or query 
Example: 
STAT:QUES:ENABle 1536  
Set bits no. 9 and 10 of the QUEStionable:ENABle register
STATus:QUEStionable[:EVENt]? 
Returns the contents of the EVENt part of the QUEStionable register. Reading an EVENt register clears it.  
*RST value 
SCPI, Command 
Types 
Confirmed, query only 
Example: 
STAT:OPER?  
Query the EVENt part of the OPERation register to check whether an event 
has occurred since the last reading. 
STATus:QUEStionable:NTRansition 
Sets the negative transition filter. Setting a bit causes a 1 to 0 transition in the corresponding bit of the 
associated condition register to cause a 1 to be written in the associated bit of the corresponding event 
register.  
<NRf> 
0to 65535 (decimal representation) 
*RST value 
–(see also Reset Values of the Status Reporting System) 
SCPI, Command Types 
Confirmed, command or query 
Example: 
STAT:QUES:NTR 1536  
Set bits no. 9 and 10 of the QUEStionable:NTRansition register