Reflection Measurements 
Operating manual 1303.6538.62 - 05 
45 
leaving an appropriate display margin.  
Refer to section Scaling Diagrams to learn more about the different 
methods and tools for diagram scaling.  
2.1.3  Instrument Calibration  
The analyzer provides sophisticated calibration methods for all types of 
measurements. Which calibration method is selected depends on the expected system 
errors, the accuracy requirements of the measurement, on the test setup and on the 
types of calibration standards available.  
In the following we assume that the calibration kit ZV-Z21 contains an appropriate male 
short standard with known physical properties. With a single short standard, it is 
possible to perform a normalization, compensating for a frequency-dependent 
attenuation and phase shift in the signal path.  
Due to the analyzer's calibration wizard, calibration is a straightforward, menu-guided 
process.  
1.  Unscrew the DUT and connect the male short standard from 
calibration kit ZV-Z21. 
2.  Press the CAL function key to open the calibration menu. 
3.  Activate Start Cal – One Port P1 – Normalization (Short) to open 
the calibration wizard for the selected calibration type.
Pdf metadata editor online - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
remove pdf metadata online; view pdf metadata in explorer
Pdf metadata editor online - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
adding metadata to pdf; remove metadata from pdf
R&S
ZVL 
Getting Started 
Reflection Measurements 
Operating manual 1303.6538.62 - 05 
46 
4.  In the first dialog of the wizard, select the calibration kit (here: ZV-
Z21) and the test port connector (here: N 50  (f), corresponding to 
amale calibration standard), and click Next.
If you have not yet imported the exact cal kit data of your calibration kit, 
you can use the typical data as shown above. Typical data provide an 
approximate description of a calibration kit mode. To import the actual 
(accurate) data of your kit, press Import Kit... and select the appropriate 
cal kit file.   
The next dialog of the calibration wizard shows that only a single 
calibration standard needs to be measured.   
5.  Click the box Short (m)... to initiate the measurement of the 
connected short standard.  
The analyzer performs a calibration sweep and displays a message 
C# HTML5 PDF Viewer SDK to view, annotate, create and convert PDF
RasterEdge C#.NET HTML5 PDF Viewer and HTML5 PDF Editor are professional online PDF manipulation tools, which are compatible with both 32-bit and 64-bit
read pdf metadata; edit pdf metadata acrobat
VB.NET PDF- HTML5 PDF Viewer for VB.NET Project
ASP.NET PDF Viewer; VB.NET: ASP.NET PDF Editor; VB.NET to PDF. Image: Remove Image from PDF Page. Image Link: Edit URL. Bookmark: Edit Bookmark. Metadata: Edit, Delete
online pdf metadata viewer; pdf metadata viewer online
R&S
ZVL 
Getting Started 
Reflection Measurements 
Operating manual 1303.6538.62 - 05 
47 
box with a progress bar. After completing the sweep the analyzer 
generates a short sound and a green checkmark appears in the 
checkbox.   
6.  Click Apply to close the wizard, calculate and store the system 
error correction data and apply them to the current measurement.  
7.  Remove the short standard and connect the DUT again.  
2.1.4  Evaluation of Data  
The analyzer provides various tools to optimize the display and analyze the 
measurement data. For instance, you can use markers determine the maximum of the 
reflection coefficient, and change the display format to obtain information about the 
phase shift of the reflected wave and the impedance of your DUT.  
1.  Press the MKR function key. This places Marker 1 to its default 
position (center of the sweep range). 
Amarker symbol (triangle) appears on the trace. The stimulus value 
(frequency) and response value (magnitude of the reflection coefficient 
converted to a dB value) at the marker position is displayed in the 
marker info field in the upper right corner of the diagram.  
2.  Press the MKR --> function key, open the Marker Search submenu 
and activate Min Search.
The marker jumps to the absolute minimum of the curve in the entire 
sweep range. The marker info field shows the coordinates of the new 
marker position. 
3.  In the TRACE keypad, press FORMAT and select the Phase of the 
reflection coefficient to be displayed.  
The phase is shown in a Cartesian diagram with a default vertical scale 
of –225 deg to +225 deg. The marker info field shows the frequency and 
phase at the marker position.  
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
C# TIFF - Edit TIFF Metadata in C#.NET. Allow Users to Read and Edit Metadata Stored in Tiff Image in C#.NET Application. How to Get TIFF XMP Metadata in C#.NET.
batch edit pdf metadata; batch update pdf metadata
C# HTML5 PDF Viewer SDK to create PDF document from other file
WPF Viewer & Editor. WPF: View PDF. WPF: Annotate PDF. WPF NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET NET read barcodes from PDF, C#.NET OCR scan PDF
remove metadata from pdf online; acrobat pdf additional metadata
R&S
ZVL 
Getting Started 
Reflection Measurements 
Operating manual 1303.6538.62 - 05 
48 
4.  Still in the FORMAT menu, select Smith.
The Smith chart shows lines of constant real and imaginary part of the 
impedance in the reflection coefficient plane.  
Refer to section Display Formats and Diagram Types in Chapter 3 to 
learn more about the diagram properties. 
2.1.5  Saving and Printing Data  
The analyzer provides standard functions for saving measurement settings and for 
printing the results. You can use these functions as if you were working on a standard 
PC. Moreover you can export your trace data to an ASCII file and reuse it in a later 
session or in an external application.  
Data transfer is made easier if external accessories are connected to the analyzer or if 
the instrument is integrated into a LAN. Refer to sections External Accessories and 
Remote Control in a LAN in Chapter 1 to obtain information about the necessary steps.  
1.  Press TRACE and activate More 1/3 – Import/Export Data – Export 
Data.  
2.  In the Export Complex Data dialog opened, select a file location, 
format and name and activate Save. 
The active trace data is written to an ASCII file.  
Refer to section Trace File Formats to learn more about trace files and 
their use.  
3.  Press the PRINT key to the left of the display; then press Device 
Setup.
4.  In the Hardcopy Setup dialog opened, select Printer. 
5.  Close the dialog and press Print Screen to create a hardcopy of your 
diagram. 
C# PDF Library SDK to view, edit, convert, process PDF file for C#
HTML5 PDF Editor enable users to edit PDF text, image, page, password and so on. C#.NET: WPF PDF Viewer & Editor. C#.NET: Edit PDF Metadata.
analyze pdf metadata; google search pdf metadata
C# TIFF: TIFF Editor SDK to Read & Manipulate TIFF File Using C#.
Tiff File Processing in C#. Refer to this online tutorial page, you will see: 2. Render text to text, PDF, or Word file. Tiff Metadata Editing in C#.
remove pdf metadata; adding metadata to pdf files
R&S
ZVL 
Getting Started 
Transmission Measurements 
Operating manual 1303.6538.62 - 05 
49 
6.  Press Device Setup again and select a file format or Clipboard. 
7.  Close the dialog and press Print Screen again to copy the diagram to 
afile or an external application.   
8.  Open the Nwa-File menu and select Save NWA As...
9.  In the Save As dialog opened, select a file location, format and name 
and activate Save. 
The active setup is stored to a file and can be reused in a later session.  
Proceed as described in section Starting the Analyzer, Shutdown to shut down your 
analyzer. 
2.2  Transmission Measurements  
Atransmission measurement involves the same steps as a reflection measurement. 
Note the following differences:  
J
The test setup for transmission measurements involves two DUT and analyzer 
ports. You can connect the input of your DUT to port 1 of the analyzer, the output to 
port 2. After a preset, the analyzer will measure the forward transmission S-
parameter S
21
.
J
The analyzer provides special calibration types for transmission measurements. 
Use the calibration wizard and select an appropriate type. A TOSM calibration will 
correct the system errors for all transmission and reflection S-parameters.   
VB.NET PDF- View PDF Online with VB.NET HTML5 PDF Viewer
NET PDF Viewer; VB.NET: ASP.NET PDF Editor; VB.NET Online. |. Home ›› XDoc.PDF ›› VB.NET PDF: View PDF Online. PDF bookmark, C#.NET edit PDF metadata, C#.NET
clean pdf metadata; edit pdf metadata acrobat
C# PDF Convert to Tiff SDK: Convert PDF to tiff images in C#.net
C#.NET PDF SDK - Convert PDF to TIFF in C#.NET. Online C# Tutorial for How to Convert PDF File to Tiff Image File with .NET XDoc.PDF Control in C#.NET Class.
endnote pdf metadata; edit multiple pdf metadata
R&S
ZVL 
Getting Started 
Basic Tasks 
Operating manual 1303.6538.62 - 05 
50 
2.3  Basic Tasks    
The following sections describe how to solve basic tasks that you will frequently 
encounter when working with the instrument. In particular you can learn how to access 
instrument functions and control dialogs without a mouse and keyboard.  
2.3.1  Control via Front Panel Keys     
Although a mouse and external keyboard simplify the operation of the instrument, you 
can access all essential functions using the keys on the front panel. The following 
examples are intended to make you familiar with front panel key operation.   
To Access a Particular Menu Command ... 
1.  Press the MENU key to the left of the display to access the menu 
bar and open the Nwa-File menu.  
2.  Use the keys in the NAVIGATION keypad or the rotary knob to 
navigate between and within the menus.  
J
Use the Cursor Left and Cursor Right keys to change between the 
different menus in the menu bar. When the first option in a pull-
down menu is a submenu, the submenu will be opened first before 
proceeding to the next option in the menu bar. 
J
Use the Cursor Up and Cursor Down keys to scroll up and down in 
amenu.  
J
ENTER, the Cursor Right key or the rotary knob (if pressed) 
expand a submenu, open a dialog or initiate an action, depending 
on the selected command type.  
J
ESC CANCEL or the Cursor Left key compress the current 
submenu and move the cursor one menu level up or close the 
active dialog, depending on the selected softkey type. 
R&S
ZVL 
Getting Started 
Basic Tasks 
Operating manual 1303.6538.62 - 05 
51 
3.  As soon as you reach the desired menu command (which must not 
be one opening a submenu) press ENTER or press the rotary knob 
to initiate an action or open a dialog.  
After command execution or after closing the dialog, the menu bar is 
deactivated and the cursor returns to the diagram/softkey area.  
To Make a Selection in a Dialog...  
1.  Press a softkey or a menu item followed by three dots to open a 
dialog.  
2.  Use the navigation keys and/or the rotary knob to access the 
controls in the dialog. 
J
Press Left Field or Right Field or the cursor keys to switch between 
the control elements in a dialog. 
J
Press the cursor keys to switch between several entries in a list of 
alternative or independent settings. 
R&S
ZVL 
Getting Started 
Basic Tasks 
Operating manual 1303.6538.62 - 05 
52 
3.  Use the data entry keys or the rotary knob to enter numbers. Use 
the on-screen keyboard for character entry. For more details refer 
to Data Entry. 
4.  Press ENTER, ESC CANCEL or press the rotary knob to close the 
active dialog.  
2.3.2  Data Entry     
The analyzer provides dialogs with various types of input fields where you can enter 
numeric values and character data. Data entry with a mouse and an external keyboard 
is a standard procedure known from other Windows applications. However, there are 
various alternative ways to enter data.    
Using Front Panel Keys  
If no mouse and no external keyboard is connected to the analyzer, you can use the 
data entry keys to enter numbers and units.   
To enter a numeric value 
1.  Place the cursor into a numeric data input field in a dialog or in the numeric entry 
bar.  
2.  Press the data entry keys.  
... 
J
Use 0 to 9 enter the corresponding numbers.  
J
Use . and - to enter a decimal point or change the sign of the 
value.  
... 
J
Use GHz / -dBm, MHz / dBm
,
kHz / dB, or Hz / dB.. to multiply 
the entered value with factors of 10
(-)9
,10
(-)6
,10
(-)3
or 1 and/or 
add the appropriate physical unit.  
To enter a character string 
1.  Place the cursor into a character data input field in a dialog.  
2.  Press the data entry keys as if you were writing a short message on your mobile 
phone.  
The different characters assigned to each key are displayed in a popup dialog.  
R&S
ZVL 
Getting Started 
Basic Tasks 
Operating manual 1303.6538.62 - 05 
53 
... 
J
Press 0 to 9 once to enter the corresponding numbers.  
J
Press the keys repeatedly to select one of the other characters 
assigned to the key. 
J
Wait 2 seconds to confirm an entry.  
J
Use . or – to enter a dot or a hyphen.  
J
Use the sign key to change from upper case to lower case and 
vice versa. 
J
Use the checkmark key to enter a space.  
J
Use the BACK key to correct wrong entries, deleting the 
character to the left of the current cursor position.  
J
Press ENTER to complete an entry.  
J
Press ESC CANCEL to close the popup dialog, discarding the 
entries made. 
3.  To enter letters you can also use one of the following methods:  
J
If the active input field has a 
symbol, then use the analyzer's on-screen 
keyboard.  
J
Otherwise, use an external keyboard or a mouse and Windows XP's on-screen 
keyboard.  
Using the Analyzer's On-Screen Keyboard  
The on-screen keyboard allows you to enter characters, in particular letters, without an 
external keyboard. It is available for all character input fields which have a 
symbol.  
Operation with front panel keys 
1.  Place the cursor into a character data input field in a dialog or in the numeric entry 
bar.  
2.  Press ENTER or the Checkmark key to open the on-screen keyboard. 
3.  Use the cursor keys or turn the rotary knob to move the cursor to a character.  
4.  Press ENTER or the rotary knob to select the character for the input string. 
5.  After completing the input string use the Right Field key to move to the OK button.  
6.  Press ENTER or the rotary knob to apply your selection and close the keyboard. 
Operation with a mouse 
1.  Click the keyboard symbol to open the on-screen keyboard.  
2.  Click a sequence of characters and OK to apply your selection and close the 
keyboard. 
R&S
ZVL 
Getting Started 
Basic Tasks 
Operating manual 1303.6538.62 - 05 
54 
You can also access Windows XP's on-screen keyboard from the start menu. Press 
SETUP – General Setup – More – Open Start Menu (or connect an external keyboard 
and press CTRL + ESC) to open the start menu, and click Programs – Accessories – 
Accessibility – On-Screen Keyboard. From the start menu, you can also access other 
useful software accessories.  
2.3.3  Scaling Diagrams     
The analyzer provides several alternative tools for setting the sweep range and 
customizing the diagrams. Pick the method that is most convenient for you.  
Setting the Sweep Range 
The sweep range for all channels is displayed in the channel list across the bottom of 
the diagram area:  
To change the sweep range, use one of the following methods:  
J
Press the CENTER or SPAN function keys on the front panel.  
J
Right-click the start or stop value in the channel list and select Start, Stop, Center, 
Span from the context menu.  
J
Select Start, Stop, Center, Span from the Channel – Center or Channel– Span 
menus.  
J
Use the marker functions (MARKER ––> function key).  
Reference Value and Position 
The analyzer provides three parameters for changing the scale of the vertical 
(response) axis: 
J
Changing the Ref Value or Ref Position shifts the trace in vertical direction and 
adjusts the labels of the vertical axis. Ref Value also works for radial diagrams.   
J
Changing the Scale/Div modifies the value of the vertical or radial diagram 
divisions and thus the entire range of response values displayed. 
J
The Scale/Div and the Ref Value is indicated in the scale section of the trace list.  