FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Maintenance 
6-11 
Caution:
The information in this section does not consti-
tute permission to adjust, manipulate, or remove any 
electrical or other components in the FluoroMax
®
-4. This 
information is provided for reference purposes only. Con-
tact Fluorescence Service in case of system failure or 
suspected failure, before attempting repairs or testing 
with meters. DO NOT ATTEMPT ANY REPAIRS OR IN-
STRUMENT EVALUATION WITHOUT THE EXPRESS 
PERMISSION OF FLUORESCENCE SERVICE. 
Electronics 
In case of the rare chance of system failure, this section is provided to help the user un-
derstand the electronics components. 
Xenon-lamp power supply 
The xenon lamp’s power supply has 
a capacity of 180 W, and is found in 
the center back of the instrument. In 
order to suppress the emf spike 
emitted from the lamp’s power 
supply during arcing and startup of 
the lamp—several thousand volts—
this contains specialized line filters. 
The power supply is preset at the 
factory for the correct current and 
voltage to supply the xenon lamp with 150 W. Poor lamp performance or difficulty 
maintaining a stable lamp arc may be the result of an improperly set power supply. 
Contact Fluorescence Service for more details. 
Monochromator drives and accessory controllers 
The monochromator drives, drive electronics, and accessory electronics are found on 
the underside of the optical platform. Electrical connections for the emission detector 
and the basic system-electronics for AC-power conversion are also underneath the in-
strument. 
To gain access to the drive electronics, remove the bottom cover of the instrument. 
Remove all screws on the bottom cover except those that attach the dust screens for the 
exhaust fans. The leveling feet do not need to be removed before detaching the bottom 
cover: 
Pdf metadata reader - add, remove, update PDF metadata in C#.net, ASP.NET, MVC, Ajax, WinForms, WPF
Allow C# Developers to Read, Add, Edit, Update and Delete PDF Metadata
batch pdf metadata editor; add metadata to pdf file
Pdf metadata reader - VB.NET PDF metadata library: add, remove, update PDF metadata in vb.net, ASP.NET, MVC, Ajax, WinForms, WPF
Enable VB.NET Users to Read, Write, Edit, Delete and Update PDF Document Metadata
embed metadata in pdf; pdf remove metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Maintenance 
6-12 
Caution:
The information in this section does not consti-
tute permission to adjust, manipulate, or remove any 
electrical or other components in the FluoroMax
®
-4. 
This information is provided for reference purposes on-
ly. Contact Fluorescence Service in case of system 
failure or suspected failure, before attempting repairs 
or testing with meters. DO NOT ATTEMPT ANY RE-
PAIRS OR INSTRUMENT EVALUATION WITHOUT 
THE EXPRESS PERMISSION OF FLUORESCENCE 
SERVICE. 
There is a 400906 control board underneath the instrument, which controls the drives, 
slits, shutter, and automated accessories in the monochromators. This board also con-
trols options such as a sample changer. 
Do not detach the 
screens’ screws. 
How to C#: Modify Image Metadata (tag)
XImage.Barcode Reader. XImage.Barcode Generator. C#.NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET
pdf metadata editor online; remove metadata from pdf
VB.NET PDF Library SDK to view, edit, convert, process PDF file
PDF Metadata Edit. Offer professional PDF document metadata editing APIs, using which VB.NET developers can redact, delete, view and save PDF metadata.
pdf xmp metadata viewer; remove metadata from pdf file
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-1 
7: Troubleshooting 
The FluoroMax
®
-4 system has been designed to operate reliably and predictably. If 
there is a problem, examine the chart below, and try the steps on the following pages. 
Problem 
Possible Cause 
Remedy 
Light is not reaching 
the sample. 
Excitation shutter closed.  Using the software, open the shut-
ter. 
Slits are not open to the 
proper width. 
Adjust the slits. 
FluoroMax
®
-4P is in 
phosphorimeter layout. 
Change to a steady-state layout. 
Monochromator is mis-
calibrated. 
Check and recalibrate monochro-
mator. 
Sample turret is not in 
correct position. 
Using FluorEssence™, set the 
position and open the cover to 
verify the position. 
Signal intensity is 
low. 
CW lamp is not aligned.  Align the lamp. 
Slits are improperly set.  Adjust the slit widths. 
Shutter(s) is(are) not 
completely open. 
Open the shutter(s) in Real Time 
Control. 
Lamp power-supply is 
set to the wrong current 
rating. 
Call the Fluorescence Service De-
partment. 
Lamp is too old. 
Replace lamp. (150-W lamp has 
lamp lifetime 1200–1500 h.) 
Shutter(s) closed. 
Open all shutters in Real Time 
Control. 
Signal intensity is at 
least 10 times lower 
than normal. 
Polarizer is in the light 
path. 
In Real Time Control, move the 
polarizer out of the light path. 
Monochromators are set 
to wrong wavelength. 
Select appropriate wavelength 
based on excitation and emission 
of sample. 
Detectors are saturated. 
Adjust slits. (Signal detector is 
linear to 2 × 10
6
cps in photon-
counting mode. Reference detec-
tor saturates at 200 μA.) 
Optical density effects 
Sample is too concentrated. Dilute 
C# TIFF: TIFF Metadata Editor, How to Write & Read TIFF Metadata
C# TIFF - Edit TIFF Metadata in C#.NET. Allow Users to Read and Edit Metadata Stored in Tiff Image in C#.NET Application. How to Get TIFF XMP Metadata in C#.NET.
read pdf metadata online; read pdf metadata java
C# PDF Library SDK to view, edit, convert, process PDF file for C#
accordingly. Multiple metadata types of PDF file can be easily added and processed in C#.NET Class. Capable C#.NET: Edit PDF Metadata. PDF SDK
pdf xmp metadata; pdf metadata online
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-2 
and self-absorption. 
sample by a factor of 10 or 100 
and retry experiment. 
No change in signal 
intensity. 
Detectors are saturated. 
Reduce slit settings. 
No signal. 
Lamp is not on. 
Bad lamp: change xenon lamp. 
Detectors are saturated. 
Reduce slit settings. 
Erratic signal. 
Lamp unstable. 
Let lamp warm up 20 min before 
use. 
Light leaks. 
Check dark value to determine. 
Sample has particles that 
scatter light irregularly. 
Filter sample, or let particles settle 
before running scan. 
Raman band supe-
rimposed on fluores-
cence scan. 
Aqueous solutions and 
solvents have Raman 
bands. 
Change excitation wavelength to 
move Raman band away from 
fluorescence peak, or run a blank 
scan of the solvent and subtract it 
from the fluorescence spectrum. 
Large off-scale peak 
at twice the excita-
tion wavelength. 
Second-order effects 
from the spectrometer. 
Use cut-on filters to eliminate 2
nd
-
order peak. 
Stray light in emis-
sion scan (also see 
example in this chap-
ter). 
Scattered light off the 
excitation wavelength. 
Place bandpass filters in excita-
tion light path. 
Decrease emission-spectrometer 
slit widths. 
Dirty cuvette. 
Clean the cuvette as described in 
Chapter 5. 
Solid-sample holder in 
sample compartment. 
Rotate the holder to prevent direct 
scatter from entering the emission 
spectrometer. 
Corrected excitation 
spectrum curves up-
ward ~240–270 nm. 
Dark count is divided by 
low reference signal. 
Use Dark Offset checkbox; retry 
scan. 
Noisy spectrum with 
magnetic stirrer. 
Stirring speed is too fast.  Use slower stirring speed. 
Stirring bar is too large; 
light beam is striking it. 
Use a smaller stirring bar (availa-
ble from Bel-Art Products, Pe-
quannock, NJ). 
Communication 
problems between 
computer and in-
strument. 
USB cable is improperly 
connected. 
Check USB cable’s connection. 
VB.NET PDF File Compress Library: Compress reduce PDF size in vb.
Document and metadata. All object data. File attachment. Flatten visible layers. VB.NET Demo Code to Optimize An Exist PDF File in Visual C#.NET Project.
remove pdf metadata; endnote pdf metadata
VB.NET PDF Image Extract Library: Select, copy, paste PDF images
Get image information, such as its location, zonal information, metadata, and so on. Extract image from PDF free in .NET framework application with trial SDK
change pdf metadata creation date; adding metadata to pdf
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-3 
Hardware Init. Error 
appears. 
Broken IR sensor in mo-
nochromator. 
Replace IR sensor: Call Fluores-
cence Service. 
Sample turret is not 
operating. 
Software is not enabled.  Check status. 
USB cable is connected 
improperly. 
Check USB cable’s connection. 
FluoroMax
®
-4P re-
sets itself in phospho-
rimeter mode. 
Flash lamp’s life is ex-
pired, drawing excess 
current, causing transient 
emf interference. 
Replace xenon flash lamp. 
“Data file does not 
exist” or “file read 
error” message ap-
pears 
User is not logged into 
Windows
®
XP as admin-
istrator or power user 
Log into Windows
®
XP as admin-
istrator and or power user, and 
restart FluorEssence™ 
How to C#: Basic SDK Concept of XDoc.PDF for .NET
XDoc.PDF for .NET supports editing PDF document metadata, like Title, Subject, Author, Creator, Producer, Keywords, Created Date, and Last Modified Date.
remove metadata from pdf online; acrobat pdf additional metadata
C# PDF - Read Barcode on PDF in C#.NET
C#.NET rotate PDF pages, C#.NET search text in PDF, C#.NET edit PDF bookmark, C#.NET edit PDF metadata, C#.NET edit C#.NET PDF Barcode Reader & Scanner
pdf metadata viewer; c# read pdf metadata
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-4 
Note:
Not  all spectra shown in this  section  were  produced using  the 
FluoroMax
®
-4. The  spectra  are presented to  show  different  possible 
system  or sample  problems, and may not  reflect  the performance of 
your particular FluoroMax
®
-4 
Using diagnostic spectra 
Often the spectrum reveals information regarding the hardware or software parameters 
that should be adjusted. The following spectra occur with explanations about problems 
leading to their appearance. 
Lamp scan 
Running a lamp scan verifies system integrity and indicates whether the correct para-
meters for the best possible trace are being used. The following spectrum shows the 
trace resulting from a lamp scan run with a known good lamp. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-5 
The following lamp scan spectrum shows poor resolution in the area around the peak. 
Wavelength (nm) 
Poor lamp scan of 150-W Xe lamp. Note low resolution in the area near the 467-nm 
peak. 
This lack of spectral resolution appears because the slit widths are set too wide. To re-
solve this problem, narrow the slit widths. 
0.4473 
0.3355 
0.2236 
0.1118 
–Intensity (counts s) 
Xenon-lamp peaks are 
unresolved 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-6 
Water Raman spectra 
Contaminated water 
Running a water Raman scan helps identify abnormalities caused by accessory prob-
lems or miscalibration. The following spectrum is normal: 
Below is a normal water Raman spectrum superimposed on one that exhibits a problem. 
In this instance, the water was contaminated, resulting in a high background. 
If a spectrum similar to this is obtained after running a water Raman scan, 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-7 
Rotate the cuvette 90° and rerun the scan. 
If the problem goes away, then the problem was due to the cuvette surface. Clean or use 
a different cuvette. 
Or 
Clean the cuvette. 
Fill with fresh, double-distilled, de-ionized water. 
If the problem goes away, then the problem was due to contaminated water. 
Light not striking cuvette 
The following graph shows a normal water-Raman scan with a superimposed problem 
scan. 
Here the problem is low intensity of the water signal when compared with the superim-
posed typical water Raman scan. To resolve this problem: 
Make sure the cuvette is filled to the proper lev-
el. 
Light should fall on the sample, and the meniscus should not be in the light 
path. 
FluoroMax
®
-4 & FluoroMax
®
-4P with USB v. B (23 Oct 2009) 
Troubleshooting 
7-8 
Make sure that the excitation and emission slits 
are set to the proper widths. 
Stray light 
In the following diagram, notice the high level of stray-light below 380 nm in the water 
Raman spectrum. 
To correct this problem, 
Inspect the cuvette surface for fingerprints and 
scratches. 
Clean the cuvette or use a new one. 
Verify that the excitation and emission slits are 
set correctly for a water Raman scan. 
Verify that the excitation monochromator is at 
the correct position. 